Welcome to a new era of petroleum testing

Short Operation Instruction


ERAFLASH – Flash Point Tester
Turning on the analyzer Starting the measurement
Position the analyzer properly on the bench. After connecting Put the sample cup into the opening of the front panel and
the power cable to the back panel of the analyzer, press the close the door. Press the RUN key to start the measurement.
main power switch.
The analyzer will load the firmware. During this process, the As soon as the measurement was started, the screen
LED on the front will flash. The loading takes approximately changes to the measure screen which provides information
15 seconds. After loading, the analyzer will automatically about the status of the measurement.
display the measuring menu:
The entire measurement is fully automated and takes
approximately 5 minutes (depending on the measuring
range)

Measurement settings
Choose the STANDARD, the PROFILE, the SAMPLE ID, the
OERATOR name and enter either the expected flash point or
the start and final temperature. When the measurement is finished, the result is displayed.
Measuring PROFILES can be entered and modified in the At the same time the result is also saved in the result
settings menu. Such Profiles include preset measuring memory.
temperature, appropriate sample names and if required an
offset for perfect correlation to other instruments. Press OK to return to the settings of the measuring screen.
For unknown samples use a wide measuring range entering Clean the electrodes and the sample cup to be ready for the
Tinitial and Tfinal. For routine samples with approximately next measurement.
known flash point simply enter the expected flash point. The
Tinitial and Tfinal is set automatically as required by the LIMS Communication
selected standard method. When ERAFLASH is connected to a LIMS system the results
are automatically transferred.
ASTM D 7094 & D 93 c and D56 c
If you choose these standard methods then the sample cup Results menu
for 2mL sample must be used. Use a pipette to pour the The results are also stored and shown in the Results menu
sample into the sample cup up to the 2mL mark. Put a and ordered by standard method.
stirring magnet into the cup for stirring the sample during the
test. Print results
Select the results to be printed by the line number and then
ASTM D 6450 press the Print button.
This standard method requires the smaller sample cup where
1 mL must be filled only. Fill the sample cup with a pipette Save results on USB memory stick
and put a stirring magnet in. For saving all the results connect a USB memory stick and
press the Save button
www.eralytics.com