- User ’s manual -
Model:RL-K09
Thank you very much for purchasing this beauty instrument.
Please read this instruction manual carefully before using the product to ensure correct use. In particular, be sure to read "Safety
Precautions" before use to ensure safe use. After reading, please keep it in a safe place for inspection when necessary.
Subject to technical changes without notice.
1
CONTENTS
Chapter 1 Foreword, safety matters
1.1 Fo r e w o r d - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3
1.2 Machine maintenance safety - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3
1. 3 La s e r sa f e t y - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3
1.4 Operational safety - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4
Chapter 5 Parameter - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 14
Chapter 6 Precautions - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 15
Chapter 8 Troubleshooting - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 17
2
Foreword, safety matters
1.1 Foreword
The laser system (hereinafter referred to as therapeutic instrument) is an intelligent laser
medical device launched by our company. The therapeutic apparatus has the characteristics of
excellent design, reasonable structure, beautiful appearance, reliable performance and convenient use.
The function and technical index have reached the international advanced level, and it is the current
upgraded product of similar instruments in China. Products are mainly sold to the United States,
Europe, Japan, South Korea and India.
The machine is a vertical structure, using a low-order mode carbon dioxide laser, good laser
mode, small focusing spot, high power density, good wavelength biological effect, with fast cutting
speed, good coagulation, less intraoperative bleeding, clear surgical vision . Flexible and convenient
to use, large operation surface. The laser power has high stability, the size can be preset at random,
and it has status self-diagnosis indication. It can be used in wards, beauty salons and private clinics.
Pigmented moles, spots, warts, nasal polyps, turbinate hypertrophy, snoring, hemangioma,
Treatment scars, hemorrhoids, axillary odor, cervicitis, cervical erosion, vulvar leukoplakia, removal of
foreskin, tattoos, condyloma acuminatum, etc.
Various pigmented hemorrhoids, plaques, acne, acne, syringoma, lipoma, hemangioma, scars,
Beauty
blephar ophthalmos, tattoos, assisted eyebrow cutting, wrinkle removal, vaginal, etc.
Irradiation with beam expander (optional) can treat skin numbness, pericervitis, periarthritis,
Physiotherapy
arthritis
3
1.3 Laser safety
1.The treatment system is classified into 4 types of laser products according to the laser
radiation safety classification!
2.The laser is invisible infrared light, and its wavelength is 1064nm. All personnel in the laser
operation area (including guests) must wear protective glasses!
3.The laser window is located at the end of the hand cavity head aiming. Never look directly at
the end port of the aiming and any reflected laser beam!
4.Flammable and explosive narcotic drugs, liquids, gases (such as alcohol, ether, oxygen, etc.)
are far away from the laser radiation area, and have protective measures!
5.Metal instruments can cause laser reflections. Operators should avoid direct exposure to
these instruments and try not to use reflective instruments!
6.A laser radiation warning sign should be affixed to the entrance of the laser treatment room.
It is forbidden for people to enter and leave at will during the use of the laser!
7.The company recommends that you equip a special laser operation room. Any reflective
equipment is prohibited in the room. The laser operation should be divided into a dedicated area for
the operator and the recipient.
4
System operation principle
5
2.4 Product advantages
Sufficient penetration depth to tighten effectively;
The effect lasts for a long time, and a course of treatment lasts for 1-1.5 years;
No pain, no holiday, no special care;
Safe and avoid cross infection;
The operation is simple and takes only 15 minutes;
Very high return on investment.
6
2.7 Introduction to Operation Head
7
Instrument operation guide
3.1.2 Connect the articulated arm: Take out the articulated arm, connect it to the articulated arm
seat of the machine, and tighten it with the screw ring.
3.1.4 Check the input voltage: check whether the input power voltage is consistent with the
power voltage of the machine before turning on the power, so as to avoid unnecessary danger.
3.1.5 Connect the power cord: plug one end of the corresponding power cord into the power
socket of the machine and the other end into the power outlet of the network. (Note: The network
power supply must have a good grounding, and the recommended grounding is medical grade).
3.1.6 Connect the foot switch: Align the plug of the foot switch power cord with the
corresponding foot switch socket on the instrument. When you hear a "click" sound, it means that it
is connected normally.
3.1.7 Trial operation: Turn the power switch clockwise to turn on the machine, and let the
machine run for ten minutes. (Rotate in the opposite direction to turn off the power)
8
3.2 Instrument operation method
When the emergency switch is released and the key switch is turned on, the following screen appears.
A. Fractional mode: mainly aimed at acne scars on the face and diminish scars;
B. Gynae mode: mainly for vaginal tightening and vulva redness;
C. Pulsed mode: mainly cut some moles protruding from the epidermis, syringoma, etc;
D. Vulva mode: mainly for vaginal tightening and vulva redness.
1) After selecting the power mode, use the + and-keys at the bottom of the screen to set the power.
Setting range: 1W ~ 60W point energy PE is the energy size of each point
2) After selecting the continuous (time) mode, use the + and-keys at the bottom of the screen to set the point
dwell time.
Setting range: 0.1ms ~ 10ms
9
3) After selecting the pitch (time) mode, use the + and-keys at the bottom of the screen to set the point interval
time.
Setting range: 1ms ~ 5000ms
4) After selecting the dot pitch (density) mode, use the + and-keys at the bottom of the screen to set the
dot-to-dot spacing distance.
Setting range: 0.1mm ~ 2.6mm, the dot coverage rate PE is the percentage of dot coverage.
5) After selecting the shape (graphic) mode, set the emission form. (Change form every time you press)
Launch form: straight line, square, rectangle, triangle, regular hexagon, circle and ellipse
6) Use the right side of the screen, select the "diagonal" key, and press the UP and DOWN keys to adjust the
size of the graphic.
Setting range: 0.1mm ~ 20mm
7) After selecting the scan (scanning mode) mode, set the emission mode (change the mode each time you
press it). There are 3 types in total.
Launch mode: out-of-order play, play in order based on one side, and play according to the maximum distance
between two points
8) Use the save (save) key to save the set value.
First save the value you just set in the specified number key. For example, press the number 1 in the set value,
and then press the save (save) key. When you press the number 1 key in the future, the stored setting will be
automatically entered. Value.
9) After the setting is completed, switch the emission standby (standby) button to the ready (permitted) mode.
In the ready (ready) mode, press the foot switch to emit laser.
10) While emitting laser light, press the (Preview) button to display the preview light preview effect.
11) After the transmission is completed, press the return key at the top of the screen to the standby screen.
Shape Choose according to the actual scar shape, etc., generally square.
(The above parameters are for reference only, please adjust according to the actual situation)
10
Pulsed mode interface
1) After pressing the power key, use the + and-keys to set the transmission energy.
Setting range: 1W ~ 60W
2) After pressing the mode key, use 1, 2, 3 keys to set the transmission mode.
Transmit mode: single (1), continuous (2), pulse (3)
3) Pulse width, multiple action time, frequency, adjust according to left and right.
Predetermined area:
In a single shot, only the pulse width is effective, and the adjustment range is: 10ms ~ 1000ms
Continuous laser, long laser on, no adjustment time
When pulsed, there is only multiple action time, frequency is active, adjustment range: 10ms ~ 1000ms
4) After setting, put STANDBY (standby) into READY (preparation) state, then step on the foot to have laser
output.
Power: (8 ~ 15) W, according to the size of moles and fat particles, generally 9W.
Single mode Dwell time: (1.0 ~ 2.5) ms, generally 1.5ms. The longer the dwell time, the longer the laser acts on
the skin, otherwise, the shorter.
Continuous
Power: (6 ~ 9) W
mode
Power: (6 ~ 12) W, generally 8W, according to the size of the skin mole to be cut
Pulse width: (1.0 ~ 2.0), generally 1.0ms is selected, the greater the pulse width, the greater the
Pulse mode energy, and vice versa, the smaller.
Point time: (1.0 ~ 2.0) ms, generally 1.0ms.
Interval: (1.0 ~ 2.0) ms, generally 1.0ms.
(The above parameters are for reference only, please adjust according to the actual situation)
11
Gynae mode interface
1) Laser pulse mode (Laser pulse mode) This function is the order in which the laser fires at 360 degrees,
which is divided into three types: sequential, out of order, and maximum distance. Use the plus and minus keys on
the screen to set.
2) The number of shots (Shot counter) is set using the plus and minus keys on the screen, up to 10 times, and
the common value is 1 time.
3) Size (Shap size) This function is the size of the laser scanning onto the vertebrae. Use the plus and minus
keys in the screen to set it. The usual value is a fixed value, generally between 4.5mm-5mm. No need to adjust it
often.
4) Pulse time (Duration), this function is the laser burning time of a single point, which is an important
parameter to determine the amount of energy, the highest value is 10, use the plus and minus keys in the screen to
set, the common value is 2 -5.
5) Laser power (Power), this function is the laser output power, which is also an important parameter to
determine the energy size. The maximum value is 40. Use the plus and minus keys in the screen to set, generally set
at 33-40 between.
6) The number of laser spots in a single circle (Spot size), this function is the number of laser spots in a
360-degree week, the highest value is 100, which is adjusted in units of 4, using the plus and minus keys in the
screen It is generally set between 12-28. This value cannot be too large, too large, which is equivalent to peeling,
and the vagina can't stand it.
7) Frequency of laser scanning (Frequency), this function is the number of laser scanning circles, the highest is
10, and the common value is 1 or 2.
8) 12345 is to save five commonly used parameters, click Save to save.
9) Complete the setting preparation (Ready), complete the parameter setting, and press the foot switch to
work.
10) Parameter setting status (Stand by).
11) Back to the previous level (Back)
12) The plus sign and minus sign set the parameter adjustment, upward or downward.
12
Gynae mode parameter reference table
Internal vulva treatment head: 10 ~ 20, according to the degree of skin pain and self-repair
ability, generally choose 15.
Power (W)
Vulva treatment head: 6 ~ 9, according to the degree of skin pain and self-repair ability,
generally choose 7.
Residence time 1.0 ~ 2.0, 1.0 is generally selected. The longer the residence time, the longer the laser will act on
(ms) the skin, otherwise, the shorter.
Interval time 1.0 ~ 2.0, 1.0 is generally selected. The greater the interval time, the faster the laser light output
(ms) speed, otherwise, the slower.
Point spacing 1.0 ~ 2.0, 1.0 is generally selected. The larger the point spacing, the more sparse the points of
(mm) action of the laser on the skin, otherwise, the denser.
(The above parameters are for reference only, please adjust according to the actual situation)
Accessories table
Standard configuration, slightly different according to customer requirements
User's Guide 1 set
13
Parameter
Wavelength: 10600nm
Beam transport device: Articulated arm with 7-joint and balance weight
Output Mode: Continuous, single pulse, interval pulse and super pulse
Energy: 1- 40W
Distance: 0.1-2.6mm
Weight: 33KG
Dimension: 98*62*34CM
Power: 220V/110V
14
Precautions
Laser light is not allowed to directly illuminate human eyes or healthy skin at any time.
In order to prevent human eyes or skin from being irradiated by laser, the laser is not allowed to
shine on smooth surfaces, so as to prevent the laser from forming reflection or diffuse reflection,
such as: stainless steel surface, mirror surface, etc.
After the relevant parts of the instrument are cleaned and disinfected with 75% alcohol, the laser
must be used after the alcohol has completely evaporated to avoid burning or explosion. When
using this instrument, flammable anesthetics or oxidizing gases such as nitrous oxide (N2O) and
oxygen must be avoided.
After using the private instrument, the treatment head and dilator must be disinfected with
boiling water and alcohol.
In order to prevent contamination of the focusing lens of the cutter head, and to ensure a clean
surgical environment, it is recommended to use a smoking device during surgery. Clean the
cutter head and focusing lens at least every 3 months.
When the instrument is used, in the area of 35m, looking directly at the laser will cause eye
damage. The instrument operator must wear safety glasses when working.
High voltage will be generated inside the instrument. Non-professionals are strictly prohibited to
open the instrument case to avoid electric shock.
Once the machine emits abnormal odor or produces abnormal sound, the power supply should be
cut off immediately, the power cord should be unplugged for inspection, and it is forbidden to
turn on the power before troubleshooting.
The normal temperature range of the machine is 5 ℃ ~ 40 ℃, and the humidity range is
10% ~ 80%.
In order to prevent the laser tube from getting wet and freezing, the water in the water tank
should be emptied before the instrument is moved.
When the life of the parts on the instrument is over, don't throw it away at random, please handle
it according to the relevant regulations of the local environmental protection department.
When the instrument is not in use, the key on the lock switch must be pulled out, and the key is
kept by a special person.
The operating room should be equipped with dust removal or smoking devices to avoid the
pollution of biological tissue exhaust dust.
The laser tube is made of glass, be careful to avoid damage.
15
Maintenance
The power supply and laser tube of this instrument will generate high voltage electricity. Ask a
professional to maintain the instrument to avoid electric shock.
6.1 Lens cleaning
After the instrument is used for half a year, the output energy of the instrument will slightly
decrease, which may be caused by the contamination of the focusing lens of the cutter head. Wipe
the lens a few times with a damp cotton ball, being careful to damage the lens.
6.2 Chassis cleaning
If there is dust inside the case, clean it with a damp cotton cloth dipped in detergent or
toothpaste. The cloth must not be too wet to prevent the instrument from short-circuiting.
6.3 Power calibration
The laser output power should be calibrated every year. Standard laser power meter should be
used for calibration, and the calibration personnel should have corresponding qualifications.
6.4 Fuse replacement
Turn off the power and unplug the power cord before replacing the fuse. When replacing, use a
screwdriver to unscrew the fuse mount, remove the old fuse, install a new fuse, install the fuse mount
and tighten, note that the corresponding fuse specifications should be consistent with the nameplate.
6.5 Disinfecting the cutter head
The cutter head must be disinfected after use.
16
Troubleshooting
If you encounter problems when using the instrument, please try the following solutions.
a. The foot switch plug is not a. Connect the foot switch plug
properly connected; correctly according to the
b. The parameter setting on the instructions;
control board is incorrect; b. Reset the parameters on the
The machine is operating normally, c. When using the machine for the control board according to the
but there is no laser output first time, the two nuts are not instructions;
tightened after filling the water tank; c. Tighten the nut of the water tank;
the key of the switch is not inserted; insert the key of the lock switch and
the joint arm assembly is not place it in the "1" position; tighten
tightened. the components on the joint arm.
Note: The operator is not allowed to adjust the following components: laser tube, articulated arm, semiconductor
parallel beam and microcomputer control board.
17
Quality assurance and after-sales service
This product is a high-quality laser treatment instrument with excellent design and convenient
operation. It can exert its excellent performance under the premise of normal use and maintenance of
users. Within one year from the date of purchase, the company provides free warranty service for any
damage caused by manufacturing or component defects (users must bear parts transportation fees
and professional door-to-door service fees).
During the warranty period, the instrument is damaged due to improper use, such as the use of
mismatched power supplies, mismatched parts, damage caused by installation, operation or
maintenance that is not in accordance with the instructions, and damage caused by handling the
machine. The free warranty service is immediately terminated.
* Product warranty period is 1 year. No warranty for man-made damage.
** Exceeding the warranty period, you will be charged for consumables
MODEL SER.NO.
ORIGIN Guangzhou, China
RL-K02
ADDRESS
18