Operating Instruction

一、 NOTES
CONTENTS 1. The UV lamp can't direct skin, eyes and other parts of the body in the
process of irradiation。
一、 Notes............................................................................... 2
2. To fasten on irradiation head, use light button or foot switch control
二、 System Working Surroundings..................................... 2
head work。
三、 System Configuration.................................................... 2
3. Are not allowed to budge irradiation head front lens, otherwise it will
四、SYSTEM OPERATION INTERFACE................................... 3
affect the output of uv energy。
1. Operation Interface Diagram......................................... 3
4. Please use dry towel to wipe the equipment。
2. Connection Terminal Diagram....................................... 3
5. Ensure that equipment is maintained in a dry environment。
3. Control Interface DB15 Terminal Definition................. 4
6. The equipment is in a state of irradiation, please don't set parameters。
4. LCD Screen Display Content Annotation....................... 5
7. Power or power outages soon, please don't touch the light head。
5. Instruction of Buttons.................................................... 5
8. When pull out plug plug, please ensure that the equipment switch in
五、 System Installation.........................................................6
the off state。
1. Irradiation head connection.......................................... 6
9. Without permission to remove the equipment, no warranty。
2. Foot switch connection.................................................. 6
10.Wash at least once a week irradiation head, drive outlet, in case of
3. Power connection........................................................... 6
heat dissipation is not good。
六、 Steps of the system operation...................................... 7
二、 SYSTEM WORKING SURROUNDINGS
1. Starting system............................................................... 7
1. Environment temperature:0℃--40℃
2. Irradiation........................................................................ 7
2. Environment humidity:20%--80%
2.1 SYSTEM RUNNING OPERATION..................................... 7
3. Power connection:220VAC
2.2 IRRADIATION PARAMETER SETTINGS.............................. 8
三、 SYSTEM CONFIGURATION
3. System Settings............................................................... 9
3.1 BEEPER SETUP................................................................ 10
Configuration item Model Qty/set
3.2 TEMPERATURE SENSOR SETTING.................................... 10
Controller 1 PCE
4. The total irradiation time query.................................. 11
UV Head 1 PCE
5. Irradiation temperature query.................................... 11
UV Head Connecting 1 PCE
6. System Password Setting............................................. 11
Power line 1 PCE
7. Foot switch control Irradiation.................................... 12
Food Swith 1 PCE
8. SHUTDOWN SYSTEM..........................................................12
Operating Instruction 1
EQUIPMENT WARRANTY............................................................13

1 2
Operating Instruction

四、System operation interface 3. CONTROL INTERFACE TERMINAL DEFINITION

1. OPERATION INTERFACE DIAGRAM

LCD

Terminal Terminal

sequence Definition and Function sequence Definition and Function

(Down) (Up)

2 terminal is defined as the

relays normally open normally ,


Power Switch Operation Buttons 1 1 NC
1and 2 terminal normally open,

2. CONNECTION TERMINAL DIAGRAM Alarm conduction

1 terminal is defined as the


2 2 NC
relays normally open

2 3 Common ground 3 NC

State of the output, when the alarm


4 Common ground 4
output is 24 v

Artificial control interface,


3 1 5 Common ground 5
connection switch signal

Automatic control interface,

6 Common ground 6 irradiation time is the machine


1:Control signal interface time
2:RS232 interface Automatic control interface,
3:220V AC interface 7 Common ground 7 irradiation time is the machine

time

Automatic control interface,


8 Common ground 8
irradiation time is manual control

3 4
Operating Instruction

4. LCD SCREEN DISPLAY CONTENT ANNOTATION 五、 SYSTEM INSTALLATION

Name Notes Name Notes 1. IRRADIATION HEAD CONNECTION


Worktype Settings Super PW Super password

System Settings password Password Irradiation head interfaces are connected at one end(see Figure 1), and the other

Time Time query Set PW Password Setting end connection controller interface(see Figure 2)

Temp Temp query P Power

Buzzer Settings T Time

Sensor Settings M Manual control

Valid settings A Auto control Figure 1 Figure 2

Invalid settings Voltage On-off control 2. FOOT SWITCH CONNECTION


Terminal 5 to 8 to terminal for the foot switch terminal blocks, each pair of

terminal connected to a pedal switch. 4 the terminal connection in the same way, for
5. INSTRUCTION OF BUTTONS
example as follows:

The 6 pair terminal connection with foot switch:


Buttons name Icon Function
Around the sixth of the terminal 2 terminals have a wiring hole and a fixed wires for
Confirm Confirm setting
the yellow button, press the yellow button terminal on the left, insert the foot switch
Cancel Back To Previous Menu
is one of the wires on the left side of the connection hole, then loosen the button,
Menu Enter the main menu, auxiliary channel Settings
the left terminal wiring is completed; Similarly, press the yellow button on the right
Up Buttons Choose option Settings, auxiliary channel
side of the terminal, insert another wire pedal switch to the right of the connection
Settings
hole, then loosen the button on the right side of the terminal wiring.
Down Buttons The same
Foot switch line does not distinguish between positive and negative.
Left Buttons The same
If need to disconnect, just press the yellow button, pull out the wire
Right Buttons The same

Total Buttons Manual control buttons, press work starts for


3. POWER CONNECTION
the first time, press the light end again

Discrete Button 1 Four functions of automatic control buttons are


Connect the power line interface with power
Discrete Button 2 the same. Press any one of four buttons, the
interface corresponding(see Figure 3)
Discrete Button 3 system will be set in accordance with the

Discrete Button 4 exposure time and power


Figure 3

5 6
Operating Instruction

六、STEPS OF THE SYSTEM OPERATION (2)Press any channel button, the system controlled by the set parameters, 4
channel display lamp all in red(See Figure 6),LCD display irradiation time change

1. STARTING SYSTEM (See Figure 8).Irradiation time reaches the set value, the system automatically

Put through power supply, press the controller at the back of the red button(See stop.Indicator light is all green,LCD display by the automatic control of radiation

Figure 3),system start to finish. parameters.

2. IRRADIATION Among them, the figure 7 M is the abbreviation of "Manual" form, namely manual control,

2.1system running operation "18" for the irradiation time. Figure 8 A is the abbreviation of "AUTO" in the form,namely

After startup, the system to enter the main menu interface(See Figure 4),there the automatic control

are 4 set options for“Work type”、“System”、“Time”、“Temp”,By the direction Simple operation process:
key to select up and down,press“OK”confirm options.The system default options for (1)Manual Control:
“Work type”。 Starting up (see figure 4)→Use the up and down the direction key choose the

Work type→OK(see figure5)→MULTI(see figure 6,7)→System running→MULTI(see figure5)

→The system stop illuminate

(2)Automatic Control:
Starting up (see figure 4)→Using the up and down the direction key choose the

Figure 4 Main menu interface Figure 5 Work type→OK(see figure5)→MULTI(see figure 6,7)→System running→MULTI(see figure5)

After selecting“Work type”option,Press OK button,the system will appear factory →The system stop illuminate

default Settings(See Figure 5),Irradiation power for 100%,Irradiation time for 2.2 Irradiation parameter Settings
20s.Press the "MULTI" or any channel button, system can begin to illuminate. If you want to set the irradiation power and irradiation time, and can be set by

(1)Press the "MULTI", the system is not affected by the set time limit,4 channel the following method.When leaving the factory, the system has carried on the password

display lamp all in red(See Figure 6),LCD display irradiation time change(See Figure Settings,So set parameters for the first time need to enter the password.The initial

7).Press the "MULTI" again, the system will stop,Indicator light is all green,LCD display password is 666666.

by the automatic control of radiation parameters(See Figure 5) Simple operation process:


Starting up(see figure4)→MENU(see figure9)→ Using the four direction key input

initial password→OK→Using the up and down the direction key choose the Work type

→OK→Using the four direction keys set power value→MENU(Power setting has been

completed)→Using the up and down direction keys move the cursor to the time value

→MENU→Using the four direction keys set time value→OK(see figure10)→Setup

Figure 6 Figure 7 Figure 8 completed(flashing cursor stop)

7 8
Operating Instruction

Settings option,press“OK”confirm option,The system will save the Settings.

If set to "Open", each time you start and end illuminate will have a buzzer beep;

If set to "Close", each time start and end illuminate buzzer will not sing.

No matter what kind of situation is the setting of the buzzer, the LED don't work

Figure 9 Figure 10 properly or irradiation head overheat, buzzer will call the police.

Show in Figure10,P:55% said irradiation power set to 55%;T:55555S said irradiation If detected LED don't work properly, buzzer alarm, interface shown below:
time set to 55555s.Time set range 0~59999s,power set range 0~100%.The system default LED err
time is 20s,power is 90%. At this point, press "MENU" button to return to the main MENU.

After completion of the power set, if found by the power value does not meet the If the operating temperature of the detected radiation head more than set the alarm

requirements, the need to modify, is the first to use the cursor up and down arrow keys temperature,the buzzer will alarm,and the interface shown in figure 13.At this point,

to power value, then press "MENU" key, then the numerical changes.Finally,press‘OK’ press the "OK" or "ECS" button to return to the main menu.

to confirm numerical,namely modify complete.

NOTES:
(1)After setup complete,press "ECS" return to the main menu interface(Namely exit
the Settings state, enter the running state),According to the operation specified in

section 2.1 again,The system according to the parameters that are set to run.

(2)When the system is in operation, press the MULTI or any channel button, system Figure 12 Figure 13 Figure 14
will illuminate;If before irradiation, press the MENU key, the system into the setup 3.2 Temperature sensor setting
state, press the MULTI or any channel button at this time, the system does not illuminate. After entering the system Settings interface, select the "Sensor" option,then enter

(3)If the reboot the system, the system will run according to the last set parameters. setup interface of the temperature sensor(see figure14)Using the up and down direction

When need to modify the parameters, it can press the MENU, according to the above method key choose Settings option,press“OK”confirm option.

to set parameters. After selected the "Valid", appear alarm temperature Settings interface ( see

3. SYSTEM SETTINGS figure14).

Enter the main menu, select the "System" option, then enter Using the four direction keys set temperature values(setup range 0~75℃)
,press

the interface of System setup(see figure11),it have “OK”button confirm setup,The system will save the Settings.

"Buzzer", "Sensor" 2 Settings,By the up And down direction After setting the alarm temperature, when the controller detected temperature of

key to select,press“OK”button confirm option,The system UV head exceeds the alarm temperature,It will close the illuminate of ongoing process,

default options for“Buzzer”. Figure 11 and after irradiation closed, irradiation head continue air cooling 35s.

3.1 Beeper setup


After entering the system Settings interface, select "Buzzer" option,then enter

setup interface of the buzzer(see figure12)Using the up and down direction key choose

9 10
Operating Instruction

4. THE TOTAL IRRADIATION TIME QUERY 7. FOOT SWITCH CONTROL IRRADIATION


Enter the main menu interface, select “Time” option, If you want to control system of the irradiation conditions through the foot switch,

As shown in figure 15 interface:among them,the H said can through the following operations.

hours(the abbreviation of Hour),M said points First of all, according to the connection method of the foot switch connected to

(the abbreviation of Minute),The following values is the it.

irradiation time,Figure 15 shows the total irradiation Figure 15 Fifth terminal connects the foot switch:after the connection,starting up,Stepped

time is 22 minutes. on the foot switch doesn't loosen,the system began to illuminate;Loosen the pedal swith,

the system will be stop illuminate.

5. IRRADIATION TEMPERATURE QUERY Sixth and seventh terminal connects the foot switch:after the connection,starting

Enter the main menu interface, select “Time” option, up,Stepped on the foot switch,system began to illuminate,To set up after the irradiation

As shown in figure 15 interface.The figure 16 shows time, the system automatically stops.If before have not reached by the irradiation time,

Irradiation temperature is 28.5℃ Figure 16 stepped on the foot switch, the system will also stop illuminate.

6. SYSTEM PASSWORD SETTING Sixth and seventh terminal connects the foot switch:after the connection,starting

At the same time press "ESC",up and down arrow keys ,There will be appear super up,Stepped on the foot switch,system began to illuminate;Stepped on the foot switch,

password Settings interface(see figure17)


,Using the four direction key input then the system stop illuminate.The illuminate of system is completely controlled by

password is 020928,Press "OK" to enter the interface shown in figure 18,using manual, not affected by the irradiation time and the power of the setup.

‘↑’、‘↓’select“YES”,Press "OK" to enter the password Settings(see figure19), Four pair foot switch control signal, the sixth and seventh signal control effect

Using the direction keys to enter the password,press“OK”confirm setup,The system is equivalent to 1, 2, 3, 4 channel buttons;The eighth signal control effect is equivalent

save Settings. to MULTI function key.

After the password, if need to set up the system running parameters, need to enter the 8. Shutdown system
password before setting. At this time will be shown in figure 19 interface: Close the controller at the back of the red power switch, unplug the power supply,

closed system.

9. Announcements
1.Irradiation process, cannot use radiation head direct illuminate human skin,
eyes and other parts of the body.

Figure17 Figure 18 Figure 19 2.Fixed irradiation head working position,using the button control irradiation
Simple operation process is as follows: head work.

At the same time press "ESC",up and down arrow keys(see figure17)→enter super 3.Do not allow the budge irradiation head front lens, otherwise it will affect
password→OK(see figure 18)→select “YES”→OK(see figure19)→enter the output of uv energy.

password→OK→setting completed 4.In the case of a power outage, please use dry towel scrubbing equipment.

11 12
Operating Instruction

5.To ensure dry work environment. 三、Fee Service


6.When equipment outage or energization soon,Please don't touch the irradiation For more than free warranty period and was not damaged by natural products need to

head. collect fees.Our company only charge cost (including shipping costs or transportation

7.Please clean up once a week irradiation head outlet and drives the dust off costs).

the fan cover,Otherwise the dust influence heat dissipation effect.

Equipment Warranty

一、 Normal warranty service


Malfunction under normal use, customers can enjoy the warranty service.So-called

failure under normal use, it is to point to in the normal service condition, with no

artificial factors intentionally or negligently caused by the natural damage, which is

not human improper operation, not to the user to remove, repair by the damage caused.

High power LED light source device, since the date of purchase, one year free warranty.

Enjoy lifelong maintenance, more than a year later, to take the paid service, the company

charge a cost (including shipping costs or transportation costs).

二、Users in the use process, if appear the following situations,


the company assumes no obligation to provide warranty service,
customers can choose premium services.
1、In violation of the conditions of use cause damage;

2、Subsequent alteration or maintenance lead to damage;

3、Beyond allows the use of the environment (such as high temperature, corrosion, etc.)

lead to damage;

4、Due to abnormal external forces (such as falling, extrusion, knock against, impact,

etc.) caused by the damage of product;

5、Natural disaster or other force majeure (such as fire, flood, earthquake, lightning,

etc.) caused by the damage of product;

6、Counterfeit products or unable to provide effective product purchase vouchers.

13 14