Professional Documents
Culture Documents
MDR - Clinical Evaluation
MDR - Clinical Evaluation
1. Clinical Evaluation
One of the main purposes of the MDR was to amend the current clinical evaluation
procedure under the directives in order to improve the quality, safety and reliability of
medical devices. While mentioned only 23 times in the AIMDD and the MDD together, the
term ‘clinical evaluation’ is given attention 140 times under the MDR.
In other words, the objective of a clinical evaluation is, based on clinical data, to
provide sufficient clinical evidence to demonstrate conformity to the relevant General Safety
and Performance Requirements (GPRS); to evaluate undesirable side-effects; and to judge
the acceptability of the benefit-risk-ratio.
Consequently, risk management files and clinical evaluations should not be separate,
stand-alone documents under the MDR. The new Regulation expects a careful alignment of
the risk management system with the clinical evaluation (i.e. clinical risks should be
addressed in clinical investigations, clinical evaluations, and Post-Market Clinical Follow-Up)
Notice also that a clinical evaluation is mandatory for all devices and should be
proportionate to their risk class (e.g. for Class I devices, for instance, clinical evidence might
not be necessary on the basis of a justification). A clinical evaluation must follow a certain
procedure based on either:
Due to the laissez-faire attitude of the Commission until very recently (MEDDEV 2.7.1
rev 4 published on July 1, 2016 in the OJEU), using clinical justifications based on device
equivalence has been a standard practice under the MDD for decades. Although still
possible under the MDR, the equivalence argument is going to be less accepted, particularly
for higher risk devices. Even with lower risk devices, equivalence justifications may be
harder to make and must always be endorsed by the manufacturer’s NB. According to Part A
of Annex XIV, more specifically, a clinical evaluation may be based on clinical data from an
equivalent device only if they both have the same technical, biological, and clinical
characteristic. As a result, more clinical data will need to be obtained from clinical
investigations of the device in question.
Another consequence of the easier ‘search literature route’ getting much harder, is
that we can expect many manufacturers to use PMCF data to fulfil the clinical evaluation
requirements. This clearly poses a challenge for new manufacturers with no approved
medical devices on the market under the MDD.
The goals of a clinical investigation are to demonstrate that the device achieves the
performances specified by the manufacturer; to establish and verify the clinical benefit of the
device as predicted by the manufacturer; and to assess whether any side effect constitute
acceptable risks when weighed against the benefits.
1
Other categories of devices could need a clinical investigation in the absence of other means to
show clinical evidence (e.g. literature review relating to the safety, performance, design characteristics
and intended purpose of a proven equivalent device)
Glòria Macià Muñoz
In addition, no obligation of clinical investigation for ‘simple’ class III devices such as
sutures, staples, dental fillings, screws or plates if their clinical evaluation is based on
sufficient clinical data.
2. Clinical Investigation
In cases where the sponsor is the same medical device manufacturer, the obligations
listed in Chapter III of Annex XV will apply to the later too.
Finally, one exciting change for sponsors under the MDR is that in the case of clinical
trials taking place in more than one Member State, they will be able to submit a single
2
It should be possible both for the manufacturer and for another natural or legal person to be the
sponsor taking responsibility for the clinical investigation. Similarly to ARs, if the sponsor of a clinical
investigation is not established in the EU, a natural or legal person established in the EU must act as
its legal representative.
Glòria Macià Muñoz
● Improvement on the quality, safety and reliability of medical devices ⎼rebuilded trust
and reputation of the industry.
● The equivalence argument and the literature route will become unavailable for many
manufacturers ⎼need for other sources of clinical data (e.g. clinical investigations or
PMCF).
● Clinical data definition excludes some sources of valid but unpublished data such as
registries or patient feedback.
● New equivalence definition may lead to unnecessary clinical investigations.
● Clinicians may be reluctant to conduct studies where the outcome data is commonly
anticipated
● Manufacturers may use PMCF data to comply with the new MDR clinical evaluation
requirements ⎼clinical investigations may be the only valid option for new
manufacturers and existing manufacturers who aim to innovate.
● Clinical investigations will be required more often ⎼mandatory for implantable devices
and Class III with some envisaged exemptions.
● New possibility for manufacturers of implatable and Class III to receive voluntary
early scientific advice from an expert panel.
● Stricter requirements for clinical investigations ⎼greater emphasis on patients rights,
safety, dignity and well-being.
● Clinical investigations will take more time and will be more expensive ⎼increased
R&D costs.
● Potential for cash-flow problems due to delay in releasing a new product or
re-certifying an existing one.
● Potential increased costs for NB review of clinical evaluations.
● Strengthening of the transparency of information regarding clinical investigations.
Previously confidential documents will become publicly available in the Eudamed
⎼manufacturer’s sensitive information may be exposed.
● Faster serious adverse events reporting in the Eudamed.
● Easier registration of clinical trials that take place in more than one Member State.
● … any other remarks?
Which are going to be the requirements regarding clinical evaluations (AND clinical
investigations) for....
3
Notice though that coordinated assessments of clinical investigation applications will be of voluntary
participation for each Member State. They will not become compulsory until after 7 years of the MDR
Application Date.