The sparking is made under an argon atmosphere environment in order to prevent the analysis sample's
surface from oxidation.
A closed water cooling circuit maintains the temperature of the table and the sample at an acceptable stable
level.
A high integrity safety system (according to CEI 1010 standard) breaks the power relay (ON/OFF) of the
source through the VDE SAFETY contact of the stand's door. When the door is closed, the mains is present
and the power relay is supplied.
The table is crossed by a water circuit so as to stabilise the table's temperature by thermal conduction. The ta-
ble exchange is very easy.
The heated lens is mounted in a lens holder assembly, and can be easily removed for maintenance purpose.
The stand's enclosure (in fact the sparking chamber) must be aligned on the spectrometer optical axis,
according to the conventional method (dummy grating and target). The S419736 target must be put over the
electrode and the optical axis must be set 1 mm above the electrode (or 2 mm under the sample!).
Most handling and maintenance procedures (removal of the table, cleaning the lens, etc.) are described into
the Reference Manual, chapter 4, Routine Maintenance.