A Burr-Brown converter, type VFC32KP is used for the measurement of the profile. It converts the pulses
generated by the IVFC board into a voltage proportional to the frequency of these pulses. The conversion
ratio is of 1 V for 50 kHz.
P6
+12 V 25 KΩ -12 V
12 KΩ
10 MΩ
27 KΩ
1 nF
P
12 1 4
8 10 KΩ
Profile
= fin 10 13 Vout
100 pF
VFC 32 KP
TP4 TP3
11 5 14
2.2 KΩ 100 pF
Figure 2.26
Using an IVFC board and an adjustable current source, create a profile signal close to zero (several tenths of
Hz) and adjust P6 so as to obtain, between TP4 and TP3, the corresponding voltage according to the relation
50 kHz = 1 V (for example 100 Hz will give 2 mV).
In the same way, inject a high frequency signal (~200 kHz) and adjust P8 so as to obtain the corresponding
voltage (200 kHz will give 4 V).