Presentation Thornton Presentation Conductivity Measurement
Presentation Thornton Presentation Conductivity Measurement
Conductivity/Resistivity pH ORP
Ionic Conductance
~
H+
O H
Na+
O H H
Cl-
Cell Constant
1 cm
1 cm 1 cm
Length Area
1 cm 1 cm2
= 1 cm-1
0.1constant
10.0constant
Resistivity/Conductivity Cell
Concentric Electrodes
OUTER ELECTRODE
INNER ELECTRODE
INSULATOR
Units of Conductivity/Resistivity
Resistance Conductance Resistivity ohm mho, siemens = 1/ohm ohm-cm, megohm-cm, M-cm mho/cm, mho/cm siemens/cm, microsiemens/cm, S/cm
THORNTON / METTLER TOLEDO
Conductivity
Units of Conductivity/Resistivity
Industry Preferences:
Resistivity - Semiconductor ultrapure water Conductivity - Power, Pharmaceutical, Pretreatment stages, Cooling towers, Wastewater Total Dissolved Solids (ppm TDS)
ASTM D1125 Primary Standard and Pure Water at 15, 25, 40C
Thornton Inc.
Conductivity and resistivity are measured at 25C; TDS is expressed as Sodium Chloride (NaCl)
to drain
H+ O H H
Na+
Cl-
~ 2 %/C
Dissociation of Water
H+ H+ O2O2H+ H+
H2O
100
0.01
10
0.1
1 0 10 20 30 40 50 60 70 80 90
1 100
Temperature (C)
THORNTON / METTLER TOLEDO
C onductivity (S/cm)
1730 ppb
1.00
840 ppb 395 ppb 129 ppb
0.10
40 ppb
0 ppb
0.01
0 10 20 30 40 50 60 70 80 90 100
Temperature (C)
Temperature Compensation
Standard -- high purity water with mineral contamination Linear, %/C -- special applications with known coefficient Cation/Ammonia/ETA -- power plant cycle chemistry and semiconductor acid etch rinse operations Alcohol -- special semiconductor rinse operations Glycol -- semiconductor coolant monitoring Light 84 -- same as Standard but with 1984 pure water data
(reads slightly higher resistivity)
Direct % Concentration Readout of HCl, H2SO4, NaOH automatically selects appropriate compensation
Cell Installation
Recommended Cell Installation...
OUTLET
INLET
0.7
HCl
H2SO4
Conductivity (S/cm)
0.6
0.5
0.4
NaOH
0.3
0.2
NaCl
0.1
0 0 10 20 30 40 50 60 70 80 90 100
Concentration (% by weight)
THORNTON / METTLER TOLEDO
Graphite Electrodes
Bottom view
Graphite Electrodes
Bottom view
AC Voltage Measurement
Measuring Electrodes
Energized
Measured
pH Range
pH Measurement
Dissociation of water
H2O
H+ + OH-
Kw =
[H+] [OH-]
pH Scale
pH ~ -log10 [H+] [H+] ~ 10-pH
0 1
3 4 5
6 7
8 9 10 11 12 13 14 pH [H+] [OH+]
10-0 10-1 10-2 10-3 10-4 10-5 10-6 10-7 10-8 10-9 10-10 10-1110-1210-1310-14
10-14 10-13 10-12 10-1110-10 10-9 10-8 10-7 10-6 10-5 10-4 10-3 10-2 10-1 10-0
at 25C
Conductivity vs pH
Conductivity
Acid
Base
pH
THORNTON / METTLER TOLEDO
Conductivity vs pH
pH
8 7 6 0 10 20 30 40
Pure Water
50
60
70
80
90
100
Temperature (C)
pH Electrode System
Glass pH Reference Temperature Compensator To Measuring Circuit
Combination pH Electrode
Electrolyte
RTD (insulated)
Reference junction
Measuring membrane
pH Electrode Output
25C 50C 100C 0C
400 200
mV
-200
-400
9 10 11 12 13 14
pH
pH Temperature Compensation
Electrode mV Temperature
pHT
pH25C
VP Preamplifier
VP pH Electrode
Housing
Combination pH Electrode Buffer solution Calibration Funnel 3-way Valve Sample Inlet to drain RTD
pH Standards
NIST standards--materials to make up buffers at: 1.681, 3.557, 4.006, 6.863, 7.41, 9.180, 10.011, 12.46 pH Commercial standard buffer solutions
NIST recipes Integral pH values
Low volume, stainless steel flow chamber Low sample flowrate with atmospheric discharge Flowing junction reference electrode Solution temperature compensation as well as electrode temperature compensation Convenient disabling of Solution temperature compensation during buffer calibration NIST-traceable buffer solutions
ORP:
Oxidation-Reduction (redox) Potential Oxidation-reduction chemical reactions
Oxidationloss of electrons, higher potential Reductiongain of electrons, reduced potential
Examples
Chlorine, ozone, permanganate can oxidize organics (color, odor, bacteria) Bisulfite or carbon beds can reduce chlorine (protect RO membranes & DI resins); Hydrazine can reduce oxygen
Measuring
Reference
Platinum Disk
Reference Junction