Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

LC•GC Europe - July 2001 GC connections 1

GC Ovens — A Hot Topic


John V. Hinshaw, ChromSource, Franklin, Tennessee, USA.

In this instalment of “GC Connections” John Hinshaw examines various means to heat gas chromatography columns.
Conventional forced-air circulation ovens are most common, but a number of other ingenious heating methods have been
used in the past and recent developments have brought innovations.

In the second Monday of October in thermal environment for GC peak elution. oven, and they cannot account for short-
Canada and on the last Thursday of GC ovens have gone through an evolutionary term oven temperature fluctuations in time
November in the United States, many development as have other GC components. caused by normal electronic and software
people celebrate Thanksgiving Day by Today, most GC ovens use rapidly circulating temperature-control processes. If large
partaking of a traditional turkey dinner with air to transfer heat to and from the column. enough, these spatial and short-term
friends and family. The preparation of this Chromatographers have used other means changes will affect peak shapes and
large indigenous fowl is the topic of much to control the thermal environment resolution: GC oven designers must
debate regarding how to properly surrounding a column, and some of these consider all types of temperature variations.
programme an oven to produce the best- means have been quite ingenious. First, Run-to-run temperature deviations: In
tasting and safest result. Theories abound, though, let’s take a look at what kind of general, I would like to keep run-to-run
ranging from isothermal heating at 160 °C thermal conditions a column requires to retention time deviations within a window
for 10 min/kg to a negative temperature deliver its best possible performance. of 10% of the peak width at base. That
ramp descending ballistically from 230 °C way, there is little ambiguity concerning
to 180 °C that purportedly imbues a GC Oven Requirements peak identification by absolute retention
pleasant hue and texture. I’m sure it has In GC, the relationship between peak time. I can’t assume that other, less
occurred to many gas chromatographers retention and temperature is fundamental. stringent peak-identification schemes, such
that they have a vastly superior device for Changes in temperature cause significant, as identification by relative retention time,
preparing baked meals in their laboratories, exponential retention shifts. In a previous will be used, so I must consider the most
but how many have actually tried to do so? instalment of “GC Connections,” I difficult case. The narrowest practical peak
I personally experienced one such attempt, discussed the dependence of retention widths will put the greatest demands on
although with something much smaller than a times upon column temperature as it the oven temperature repeatability.
turkey. I once attended a holiday fête that took relates to isothermal and temperature- Significantly retained peaks — those with
place adjacent to the gas chromatography (GC) programmed elution (1). In general, retention factors of 2 or greater — are
laboratories where I worked, and employees retention times decrease by a factor of two eluted from relatively short, 10 m 
were encouraged to bring a warm dish. One for every 15–20 °C increase in column 0.20 mm, thin-film capillary columns with
manager took this request unkindly and temperature (Tc). This phenomenon allows base widths greater than or equal to 1.5 s.
contributed one can of baked beans out of chromatographers to control solute This figure assumes that the column
spite. He placed the unopened can in a GC retention times by adjusting the column operates at its optimum average linear
oven and commenced a temperature temperature. This same effect, however, helium carrier gas velocity of roughly
programme from room temperature to 200 °C. causes small run-to-run shifts in the 30 cm/s with the best theoretical efficiency.
After approximately 10 min, a muffled average column temperature to randomize Chromatographers can obtain narrower
report echoed from the laboratory into the peak retention times by a small amount peaks by operating columns at higher than
party. Afterwards, we always knew when from one run to the next. Slightly higher optimum carrier-gas velocities or by
that particular GC oven was being used by oven temperatures cause peaks to be choosing narrower or shorter columns —
the accompanying aroma of burned beans. eluted a little earlier, and lower in other words by performing fast GC. But,
Of course, this use of laboratory equipment temperatures shift retentions to slightly in general, this method is as fast as peaks
violates several safety rules as well as any longer times. These long-term changes in go for most GC systems in operation
decent laboratory results quality programme. average oven temperature are different today. I’ll take a look at one alternative
Naturally, GC ovens were not designed for from spatial fluctuations along the column heating system for fast GC later on in this
cooking but rather to provide a suitable caused by temperature gradients in the GC “GC Connections” column.
2 GC connections LC•GC Europe - July 2001

The oven is not the only factor that more than adequate carrier-gas control at environment. I’ll take a look at some
contributes to run-to-run retention-time pressures greater than 5 psig (35 kPa) with maximum heating rates for different kinds
variations. The pneumatic system has a the best performance occurring at of ovens later in this column.
significant influence, and the data-handling pressures of 20 psig (69 kPa) or greater. One principal difference between
system introduces other uncertainties, Random run-to-run pressure fluctuations isothermal and programmed elution is the
including determination of the exact run are a less significant contributor to slight time lag between the actual column
start time and the peak apex or centroid. I retention-time variability than normal oven temperature and the set point temperature
assembled a statistical analysis of these temperature variations. At higher pressure of the temperature ramp. As the oven
factors and the GC oven using typical drops, however, changes in atmospheric temperature controller increases the set
values for modern GC instrumentation and pressure can cause long-term retention point temperature according to the
columns, which resulted in too much time drift. In those instances, an electronic demands of the programme ramp, the
material to reproduce in this column. pneumatic system with atmospheric actual oven temperature lags behind by a
Interested readers can refer to my website pressure sensing will compensate for these small amount and the temperature inside
for this supplemental information (see effects. High-output range electronic the column lags a little more. The
http://www.chromsource.com and click on controllers that deliver pressures greater temperature lag is larger at higher
“GC Connections”). In this analysis, I than or equal to 150 psig (1000 kPa) programme rates. This temperature
assumed that the room temperature and should be reserved for columns that differential is a necessary by-product of
GC line voltage supply are stable and do require high pressure drops; they lack the both the nature of temperature-control
not influence the actual oven temperature resolution to deliver adequate low-pressure algorithms and the physics of heat transfer
and column pressure drop. This situation is control. Pressure-controlled pneumatic from the heater to the column. For my
not always the case, of course, and these systems can affect retention-time purposes, it is the temperature in the
problems fall into the category of repeatability significantly when operated at column itself that counts.
instrument environmental difficulties that pressures less than 3 psig (20 kPa), as may Not only must the electronics and
should be resolved before attempting to be required with wide-bore columns software repeat the same
obtain the best retention-time shorter than 20 m. In those instances, a time–temperature profile across the run,
reproducibilities. flow-controlled system is desirable. but the thermal conditions in the oven
must also be the same from the start to
the end of the run. This condition is one
“In GC, the relationship between peak reason temperature-programmed ovens
retention and temperature is fundamental.” require a temperature equilibration period
after cooling down before a new run can
commence. Shortly after the oven cools
Isothermal elution: The analysis indicates Temperature programming: With down, the oven temperature probe may
that the average oven temperature must isothermal elution, peak base widths indicate the initial oven temperature, but a
repeat from run-to-run to within less than increase approximately in proportion to the significant amount of heat remains in the
0.05 °C to hold retention times inside a retention time; however, with oven; the walls and the column can be
window of 10% of peak base widths for programmed-temperature elution, the significantly hotter than the indicated
commonly encountered solutes. Lee, Yang base widths stay approximately constant temperature. This situation is the reverse of
and Bartle (2) cite a similar requirement in across a GC run for peaks that are eluted the temperature lag encountered with
their 1984 book. These figures are for during the temperature ramp. Most peaks oven heating, and good practice dictates
typical test mixture solutes separated at are essentially immobilized at the initial that users allow sufficient time for the
moderate isothermal temperatures on column temperature; as the oven residual oven heat to dissipate and bring
short narrow-bore, thin-film capillary temperature increases, solutes begin to the column as close as possible to the
columns, which represent the more move along the column. Each analyte initial oven temperature. An equilibration
difficult situation. Columns with thicker spends about the same time moving time of 2–4 min is typical for most air-bath
stationary films, longer columns, wide-bore through the column (3), so the peaks have GC ovens with temperature-programming
columns and packed columns generate approximately the same widths as they are capability. With isothermal elution, no
broader peaks and may tolerate a less eluted. For temperature programming, equilibration time is necessary after the
stringent temperature environment and oven temperature ramps must repeat the oven reaches its temperature for the first
still deliver retention-time repeatability time–temperature curve with the same time, because the temperature is not
adequate for peak-identification purposes. precision as required for isothermal elution. changed during the run.
Alternatively, fast GC separations The oven heating system must have Oven gradients: A high degree of run-to-
generated on microbore columns with sufficient power to maintain a linear run average temperature repeatability
inner diameters smaller than 0.2 mm, on programme ramp from the start ensures the best retention-time precision.
columns shorter than 10 m, or with temperature to the final temperature. The GC ovens must also maintain a uniform
elevated carrier-gas velocities greater than maximum rate at which the oven and stable temperature profile across the
100 cm/s may require better temperature temperature can increase linearly is related column during elution. The process of
control to contain retention times within to the thermal mass of the oven cavity, solute partitioning and migration along the
10% of base-width window. including the column, the power that the column averages out temperature
Pneumatic system: In general, modern heater can dissipate into the oven, the gradients so that peaks are eluted as if
pneumatic systems with electronic pressure efficiency of the oven boundary insulation they experienced the average column
controllers or high-precision mechanical and the differential temperature between temperature along the length of the
regulators designed for capillary GC deliver the inside of the oven and the external column. Minor temperature fluctuations of
LC•GC Europe - July 2001 GC connections 3

less than 1 °C along the column do not problem of transient thermal gradients. substance (4). This technique of
affect retention times seriously, as long as They had it easier than present-day temperature regulation had been used in
these gradients are stable and repeatable. chromatographers, however, because they various related disciplines before gas-
Unstable gradients that change from run had to accommodate packed columns and partition GC, and it was a natural step to
to run, however, will impose additional metal or thick-walled glass capillary tubing, use it for the then-new technique.
uncertainty on retention times because but not fused-silica tubing. Thin-walled, Operators could change the temperature
they affect the average temperature that fused-silica tubing responds more rapidly of the vapour bath by regulating the
solutes experience. to oven temperature changes and is more pressure inside a sealed oven container.
Some temperature gradients are normal susceptible to the effects of rapidly p-Xylene, for example, boils at 138 °C at
and necessary. The column inlet and outlet changing oven gradients. The newest 1 atm, and temperatures approximately
connect to the injector and detector, which generation of GC ovens includes features 30 °C lower were attainable at reduced
are normally at elevated temperatures designed to maximize performance with pressures (5). Changing the oven
relative to the oven. Most modern capillary currently available columns. temperature outside this limited range was
GC systems position the column ends I’ll divide oven heating schemes into two inconvenient, because the pure substance
inside the inlet and detector, which creates broad categories: those that rely on a fluid had to be exchanged for another with a
a negative temperature gradient from the to transfer heat to and from the column higher or lower boiling point. This
inlet into the column and a positive and those that use solid-state materials for arrangement produced a fairly stable
gradient into the detector. The net effect heating purposes. Today, of course, air is the thermal environment for the column, as
on retention is negligible, because solutes heat-transfer fluid of choice for most, but long as the atmospheric pressure didn’t
will be retained weakly, if at all, in the that was not always the situation. change too much. Other problems arise
short sections of heated column ends, Chromatographers have experimented with with decomposition of the boiling
which act more like connecting tubing. various vapours and liquids as well. Solid- substance at higher temperatures and the
Gradients also exist close to the oven state ovens have been used for temperature flammability of the vapours, and
boundaries, because of heat losses or gains programming with metal capillary columns, chromatographers abandoned this kind of
from cooler oven walls or warmer heated and small completely solid-state ovens are system fairly early.
inlets and detectors. In a conventional found primarily in portable instruments Oil baths were also used as oven
air-bath oven, the column is normally because of their low power requirements thermostats (6). After the operating
positioned in the centre of the oven to and good long-term stability. Columns temperature was reached, a very uniform
immerse it in the most thermally uniform installed inside resistively heated metal thermal environment was possible if the
area. These kinds of stable temperature tubing that is suspended inside a heating vessel was well insulated and
gradients do not affect performance conventional air-bath oven deliver very fast efficiently stirred. Temperature control
seriously if they are stable during a run and programming rates for high-speed GC. circuitry provided fairly good temperature
from run to run in the long term. Fluid ovens: The earliest report I could find stability. These baths could be temperature
Gradients that fluctuate from run to run of a fluid oven for GC dates back to 1952 programmed but at a limited rate because of
will affect retention time stability, but and involves the suspension of the column the large heat capacity of the oil. A main
gradients that fluctuate rapidly during a inside a vapour bath over a boiling pure drawback of oil-bath ovens was the
run will affect peak shapes. A poorly sealed
air-bath oven can generate this type of
Warm air
gradient. A controlled amount of air is vent Detector
normally allowed to circulate through an Inlet system
air-bath oven to prevent the build up of
explosive hydrogen gas. This circulation is
performed in a way that prevents the
generation of significant temperature
gradients. But if the oven door or vent isn’t
sealed properly, significant amounts of cold
air can enter and blow directly on the
column to create a rapidly fluctuating
localized temperature gradient. For fused-
silica columns, with their thin walls and
rapid thermal time constants, fluctuating
gradients will alternately trap and release Smart
portions of solute bands as they pass by, doors
which will distort the peak shapes and can
even split peaks into apparent multiples.

Oven Heating Schemes Cold air


Over the years, GC researchers have intake
envisioned and built a wide variety of Temperature
Stepper
column heating instruments. Early workers Heater shield Fan sensor
motor
were well aware of the requirements
imposed by the sensitivity of GC retention Figure 1: Modern GC oven schematic. (Courtesy of PerkinElmer Analytical Instruments,
times to temperature changes and by the Norwalk, Connecticut, USA.)
4 GC connections LC•GC Europe - July 2001

“The development minimizes gradients in the cylindrical insulated with asbestos tape. Variable
column area in the oven centre. A heater voltage provided rudimentary temperature
of GC ovens has shield prevents radiated heat from heating control (7). Another approach connected a
helped foster the the column or the temperature sensor. A
pair of smart doors opens under stepper-
source of electrical current across the metal
packed column itself, and at least one
acceptance of motor control according to the cooling commercial gas chromatograph
demand of the oven programme. Cool air manufactured in the West used this
new GC column is forced into the lower door with a fan method (8).
technologies.” (not shown) and hot air exists through the A current product line uses another
upper door. approach to solid-state column heating by
inconvenience of working with high This type of oven can deliver retention putting a column and a separate heater
molecular weight oils that would coat the time repeatabilities well within the criteria into a monolithic block. One or more metal
column and connections. The slightest leak cited earlier in this column. Modern GC capillary columns are wound with a
would permit the oil to enter and destroy ovens are capable of linear temperature heating element around a core, and the
the column or, at the least, drastically modify programming rates as high as 75 °C/min entire assembly is encased in thermoplastic
its retention characteristics. Changing the with a standard 2 kW heater, when the resin. The column tubing terminates
column was difficult and messy. instrument software permits it. The outside the oven assembly in adjacent
Air-bath ovens: GC researchers maximum linear programme rate declines valves, inlet systems, and detectors and the
understood from the onset that an air-bath to approximately 20–30 °C/min at entire assemblage fits inside an insulated
oven could be superior to liquid- or vapour- temperatures higher than 175 °C, because enclosure. This type of construction is
bath designs. The temperature in an air bath of increasing heat losses across the attractive for portable isothermal gas
could be varied rapidly across a wide range temperature gradient from the oven to the chromatographs because once heated, it
without changing the working fluid, and it surrounding air. The more powerful doesn’t require much power to maintain
would be much easier to change columns. optional heaters available with some temperature, and because it can be very
However, early air-bath oven designs from instruments can yield programming rates small. However, changing the column
the 1950s suffered from large temperature as high as 120 °C/min that are suitable for requires a complete exchange of the inner
gradients and a significant lag time between high-speed GC separations. oven assembly.
the oven set point and actual column The fastest GC total run times require An alternative solid state–oven
temperatures. Special fans and better not only fast heating but also rapid cooling arrangement used a flat, helically coiled
insulation helped to remedy the situation, to the initial oven temperature. The use of metal column sandwiched between two
and by the mid-1960s the air-bath oven an extra fan — either separately or heated metal plates. This system was the
was almost universally accepted. The coaxially on the main fan motor shaft — Perkin-Elmer model 226 gas chromatograph
subsequent introduction of microprocessor that forces cooling air through the oven (Norwalk, Connecticut, USA), which was
temperature control in the 1970s and the vents helps to speed cool-down time. marketed from 1962 to 1967. The
adoption of a number of other important Cryogenic cooling is another option that instrument could perform 50 °C/min linear
features — including improvements made minimizes cool down, but it is not always temperature-programming ramps that
specifically for fused-silica columns — convenient or available. were quite fast compared with other
resulted in modern GC oven designs, such Solid-state ovens: GC experimenters also instruments of the time, and it provided
as the one shown in Figure 1. developed several solid-state column retention time and results reproducibilities
The oven door (not shown) covers the heating schemes over the years. An early comparable to those obtained using
front. The impeller fan at the rear of the design used heating wire that was simply modern GC instrumentation (9). Figure 2
oven creates a rapid, turbulent airflow that wrapped around a packed column and shows a photograph of this system’s
column oven assembly. The flat helical
metal capillary column was sandwiched
between two aluminium plates, with exit
Column sandwich

Heater block Injector

Inlet and outlet


tubes

Carrier-gas
flow

Figure 2: Perkin-Elmer model 226 column oven assembly. (Courtesy of PerkinElmer Analytical Figure 3: Metal column configuration for the
Instruments, Norwalk, Connecticut, USA.) PerkinElmer model 226 gas chromatograph.
LC•GC Europe - July 2001 GC connections 5

holes for the column inlet and outlet. The packages, such as the Agilent References
sandwich was clamped between two flat Technologies’ Method Translation software (1) J.V. Hinshaw, LC•GC, 18(10), 1040–1047
(2000).
heater blocks, and an insulated cover (Wilmington, Delaware, USA), which is
(2) M.L. Lee, F.J. Yang and K.D Bartle, Open
finished the assembly. Forced air or available free of charge at Tubular Gas Chromatography: Theory and
cryogenic fluid cooled the oven after http://www.chem.agilent.com/ Practice (John Wiley & Sons, Inc., New York,
temperature programming. cag/servsup/usersoft/main.html, can help USA, 1984), 106.
(3) J.C. Giddings, J. Chem. Educ., 39, 569 (1962).
Figure 3 illustrates the column determine appropriate column and elution (4) A.T. James and A.J.P. Martin, Biochem. J., 50,
configuration itself. The column’s outer condition adjustments. 679–690 (1952).
diameter determined the number of The column runs coaxially inside a coiled, (5) A.I.M. Keulemans, Gas Chromatography,
(Reinhold Publishing Corp., New York, USA,
column turns and, thus, the maximum externally insulated, tubular, metallic 1957), 59–61.
length — longer columns with smaller heater that is suspended on a cage in an (6) D.H. Desty and B.H.F. Whyman, Anal. Chem.,
diameters could be wound into the air-bath GC oven. The heater tube is 29, 320–329 (1957).
(7) L.S. Ettre, J. Chromatogr. Sci., 15, 90–110
available space. The heater coils above and insulated and is coiled to the same nominal (1977).
below the column were wound in a similar diameter as conventional fused-silica (8) L.S Ettre, Am. Lab., 31(14), 30–33 (1999).
fashion. Columns were available with 0.010 columns. Inlet and detector covers over the (9) H.A Gill and W. Averill, “Design Considerations
and Performance of a Linear Programmed
in inner diameters and 300 ft lengths and heater and pneumatic connections insulate
Temperature Gas-Liquid Chromatograph for
with 0.020 in inner diameters and 200 ft the column ends both electrically and Golay and Packed Columns,” paper presented
lengths. Packed columns were available with thermally. An external controller senses the at the 1962 Pittsburgh Conference on
0.085 in inner diameters and 15 ft lengths. average heater temperature and regulates the Analytical Chemistry and Applied Spectroscopy,
Pittsburgh, Pennsylvania, USA, 5 March 1962.
More recently, researchers heated electrical power accordingly. The device can (10) S.R. Lipsky and M.L. Duffy, J. High Resolut.
aluminium-coated, fused-silica columns achieve ramp rates as high as 1200 °C/min, a Chromatogr., 9‚ 376–382 (1986).
directly by passing a current through the maximum temperature of 400 °C, and (11) S.R. Lipsky and M.L. Duffy, J. High Resolut.
Chromatogr., 9‚ 725–730 (1986).
metallic outer layer (10,11). This device cool-down times of 1 min or less from
would have very fast heating and cooling 325 °C to 35 °C. Figure 4 shows the coaxial
rates, but this attractive concept was column–heater arrangement of this system. “GC Connections” editor John V. Hinshaw is
hindered by two principal difficulties. First, The heater insulation, the close contact president and principal scientist of
the outer metallic coating was uneven, between the column tubing and the heater, ChromSource, Franklin, Tennessee, USA and a
which created unwanted temperature and its placement in a well-regulated air- member of the Editorial Advisory Board of
gradients, and second, each column coil bath oven minimize thermal gradients. LC•GC Europe.
had to be carefully insulated, thermally and The EZ Flash GC system has an interesting Direct correspondence about this column
electrically, from adjacent coils as well as history. It was first applied to explosives to “GC Connections,” LC•GC Europe,
from the column supports. Any contact detection for airport security in the mid- Advanstar House, Park West, Sealand Road,
created a large thermal gradient and 1980s. A dual-column configuration Chester CH1 4RN, UK,
shorted out the heater circuit. delivered a combined sample preparation e-mail: dhills@advanstar.com
A hybrid oven: A recent commercial and analysis time of 18 s for explosives For an ongoing discussion of GC issues with
system places a solid-state heating system residues in a wide variety of gas, solid or John Hinshaw and other chromatographers,
inside a conventional air-bath oven. The EZ liquid materials. In 1996, a multidimensional visit the Chromatography Forum discussion
Flash GC system (ThermoOrion, Beverly, GC instrument based on the EZ Flash GC group at http://www.chromforum.com.
Massachusetts, USA) uses conventional concept won the Pittcon Editor’s Gold
5 or 10 m fused-silica columns with inner Award in recognition of the best new
diameters as large as 0.53 mm that are products. The current add-on single column
connected to standard inlet and detector system was introduced in 1998.
ports with short lengths of deactivated
fused-silica tubing. By using short columns Conclusion
with smaller inner diameters in Gas chromatographers have invented and
combination with fast heating, this device adapted various methodologies for
can deliver the resolution of longer, larger- controlling column temperature. Far from
diameter columns in less time. Software simple turkey baking requirements, the
physics and chemistry of GC separation and
elution determine the requirements for
  temperature control. The need for application
Heater tube flexibility or specificity determines what kind
of heating system is appropriate. A broadly
applicable system, such as the conventional
air-bath oven provides the greatest flexibility
5–10 m at the expense of portability, power
consumption, and space. Dedicated systems
Fused-silica in which the column and heater are
column permanently mated yield low power
consumption, portability and compactness.
Figure 4: Heating schematic of the EZ Flash To a great extent, the development of GC
gas chromatograph. (Courtesy of Thermo ovens has helped foster the acceptance of
Orion.) new GC column technologies.

You might also like