Technical Service Bulletin # 3
Devices affected: All units with software version 2.00 and 2.01.
Additional Info: After installation of the new Sensor PCB and the new software version
calibrate the pressure sensor in the DrägerService mode and carry out
a complete inspection of the Oxylog 2000 according to the PMS
procedure.
In the future a third sensor version will be available. This sensor has
shape and can be used with all software versions.
Software version 2.02 and higher, and 3.n supports all versions of the
Sensor PCB.
Oxylog 2000 units with SW 1.n are not available in the USA and
Canada.