Professional Documents
Culture Documents
Hemaray 86 Service Manual V1.0e
Hemaray 86 Service Manual V1.0e
Rev: 1.0e
1
Hemaray 86 / 89 Service Manual
Caution
Laser warning
Puncture warning
2
Hemaray 86 / 89 Service Manual
Table of Contents
TABLE OF CONTENTS.......................................................................2
CHAPTER 1 INTRODUCTION..........................................................6
1.1 Introduction....................................................................................................6
1.1.1 Product Name......................................................................................................................6
1.1.2 Model.....................................................................................................................................6
1.1.3 Features................................................................................................................................6
1.1.4 Test Parameters...................................................................................................................6
1.2 Composition and Structure............................................................................7
1.3 Working Principle...........................................................................................9
1.3.1 Overview...............................................................................................................................9
1.3.2 Sample Suction....................................................................................................................9
1.3.3 White Blood Cell (WBC) Testing Principle....................................................................10
1.3.4 Hemoglobin Concentration (HGB) Measurement........................................................12
1.3.5 Red Blood Cell (RBC)/Platelet (PLT) Measurement...................................................12
1.3.6 Parameters.........................................................................................................................13
1.4 Performance................................................................................................14
1.5 Scope of Application....................................................................................16
1.6 Technical Parameters..................................................................................16
1.7 PC Configuration.........................................................................................16
CHAPTER 2 INSTALLATION...........................................................18
2.1 Unpacking....................................................................................................18
2.1.1 Steps of Unpacking...........................................................................................................18
2.1.2 Handling Method...............................................................................................................18
2.2 Installation and Use Environment................................................................18
2.3 Requirements of Power Supply...................................................................19
2.4 Requirements of Space...............................................................................19
2.5 Removal of Auxiliary Fixing Structure for Transportation............................19
2.6 Connection of Reagent and Waste Liquid Containers................................19
2.7 Connecting Power Supply...........................................................................20
2.8 Preparation before Power-on......................................................................20
2.9 Debugging....................................................................................................23
2.10 Software Installation and Upgrade...........................................................27
2.10.1 Installation of XP OS.........................................................................................................27
2.10.2 Installation of Information Processing System Software............................................28
2.10.3 Installation of Main Control Board System Program...................................................29
2.10.4 Installation of Main board application Software...........................................................32
2.10.5 Installation of Driver Board Program..............................................................................32
3
Hemaray 86 / 89 Service Manual
4
Hemaray 86 / 89 Service Manual
5
Hemaray 86 / 89 Service Manual
6.2.10 Dismantling of Sampling Assy Connector Board and Sampling Assy Adapter
Board 123
6.2.11 Replacement of Fuse.....................................................................................................124
6.2.12 Dismantling of Power Supply Assy...............................................................................125
6.2.13 Dismantling of DIFF Bath Assy and Counting Chamber Assy.................................126
6.2.14 Dismantling of Sampling Assy.......................................................................................127
6.2.15 Dismantling of Waste Liquid Pump Assy.....................................................................128
6.2.16 Dismantling of Syringe Assys........................................................................................129
6.2.17 Dismantling of Valves.....................................................................................................131
6.2.18 Dismantling of Liquid Pressure Sensor.......................................................................133
6.2.19 Dismantling of Optical Path Assy..................................................................................133
6.2.20 Dismantling of Board in Optical Assy...........................................................................134
6.2.21 Replacement of Flow Cell..............................................................................................137
CHAPTER 8 MAINTENANCE........................................................153
8.1 Regular Maintenance.................................................................................153
6
Hemaray 86 / 89 Service Manual
7
Hemaray 86/89 Service Manual
Chapter 1 Introduction
.1 Introduction
.1.1 Product Name
Automated Hematology Analyzer
.1.2 Model
Hemaray 86, Hemaray 89
.1.3 Features
This instrument has two sampling modes, i.e. Automatic and Enclosed. The Automatic
sampling mode provides the whole blood test mode.
Hemaray 86: The automatic sampler carries up to 5 tube racks, each of which holds up to
10 samples, total 50 samples.
Hemaray 89: The automatic sampler carries up to 10 tube racks, each of which holds up to
10 samples, total 100 samples.
The Enclosed sampling mode provides the Whole Blood and Pre-diluted test modes. After
sampling, the instrument conducts the test automatically and provides WBC 5-part differentiation
results and scattergram, and parameter results and histograms of RBC and PLT.
.1.4 Test Mode CBC CBC+DIFF Parameters
The instrument WBC WBC has two test modes, i.e.
RBC RBC
CBC and CBC+DIFF. The CBC mode
HGB HGB
provides 14 blood HCT HCT parameters and 3
histograms. The MCV MCV CBC+DIFF mode
provides 28 blood MCH MCH parameters (in which, 4
MCHC MCHC
are study RDW-SD RDW-SD parameters), 2
scattergrams, and 2 RDW-CV RDW-CV histograms.
PLT PLT
Table 1-1 Test MPV MPV Parameter Table
PCT PCT
Parameter
PDW PDW
P-LCR P-LCR
- NEUT%
Caution: (1) “-” - LYMP% indicates that the
current test - MONO% mode does not
- EO%
provide it.
- BASO%
(2) Study - NEUT# parameters are
used - LYMP# for study only and
can’t - MONO# be used as basis
- EO#
for - BASO#
clinical diagnosis.
IG%
Study IG#
Parameter ALY%
.2 Composition and
ALY#
Structure WBC
-
histogram
The instrument RBC RBC mainly includes shell
assy, frame assy, histogram histogram optical system, power
PLT PLT
Chart
histogram histogram
8
Main
-
scattergram
Side
-
scattergram
Hemaray 86 / 89 Service Manual
supply, sampling assy, pump,valve, bath assy, etc. Its structure is shown in the figure below.
1 - Network Port 2 - Earthing Rod 3 - Waste Liquid Sensor 4 - Waste Liquid Pipe Connector 5
- Cleanser 6 - Diluent 7 - 86H Lyse 8 - 86D Lyse 9 - Power Supply Interface (including 2
fuses)
9
Hemaray 86 / 89 Service Manual
.3 Working Principle
.3.1 Overview
This analyzer adopts the Coulter Principle to test the quantity and volume distribution of
WBC, RBC and PLT, adopts the colorimetry to measure hemoglobin concentration, and adopts
the semiconductor laser and flow cell analysis technique to obtain WBC 5-part differentiation.
Based on this, the analyzer calculates the results of other parameters.
10
Hemaray 86 / 89 Service Manual
During the test with the electrical impedance method, the sample being tested, after diluted,
enters the WBC test unit which has a small opening called “test hole(aperture)”. The hole has a
pair of positive and negative electrodes on both sides that are connected to the constant current
power supply. As blood cells have a feature of bad conductor, when the blood cells in the diluted
sample pass through the test hole under the constant negative pressure, the resistance between
electrodes will change, thus forms a pulse signal with the size in proportion to the volume of the
cell on both ends of the electrode. When the cells pass through the hole continuously, a series
of electrical pulses will be produced on both ends of the electrode. The quantity of these
electrical pulses reflects the number of blood cells.
Compare the amplified electrical pulses collected with the channel voltage threshold
corresponding to the normal WBC range, and calculate the number of electrical pulses in the
WBC channel of the electrical pulse amplitude. Thus, all electrical pulses collected are classified
according to the different channel voltage thresholds, and the number of electrical pulses in the
WBC channel is the number of WBCs. The number of cells in each channel range divided based
on the pulse voltage amplitude decides the volume distribution of cells. The two-dimensional
diagram with the x-coordinate representing cell volume and the y-coordinate representing
11
Hemaray 86 / 89 Service Manual
relative cell quantity is the histogram reflecting the cell population distribution.
In the laser scattering method, when a certain amount of diluted blood sample is injected
into the Flow Cell, wrapped by the sheath fluid formed by the diluent, the cells lining up one by
one pass through the center of the Flow Cell. When the blood cells suspending in the sheath
fluid pass through the laser detection area, the blood cells are irradiated by the laser beam, and
the nature of the scattered light produced is relating to the cell size, cytomembrane, and
refractive index of the internal structure of cell. The schematic diagram of laser scattering
method is as follows:
The small-angle, large-angle and side-angle scattered lights in the direction of laser beam
are measured, which reflect the size of blood cell, complexity of cell nucleus, and granularity
respectively. The DIFF channel scattergram is obtained based on these scattered lights, as
shown in the figure below. From the scattergram, the respective proportion of lymphocyte,
monocyte, eosinophil and neutrophil in the total number of WBCs can be obtained.
12
Hemaray 86 / 89 Service Manual
13
Hemaray 86 / 89 Service Manual
distribution of cells. The two-dimensional diagram with the x-coordinate representing cell volume
and the y-coordinate representing relative cell quantity is the histogram reflecting the cell
population distribution.
.3.6 Parameters
The parameters are calculated as follows:
White Blood Cell Quantity (WBC)
The analyzer obtains the WBC quantity by directly measuring the number of electrical
pulses corresponding to WBC.
Neutrophil Percentage (NEUT%)
The analyzer obtains the RBC quantity by directly measuring the number of electrical
pulses corresponding to RBC.
Mean Corpuscular Volume (MCV)
MCV is obtained from the RBC histogram.
Hematocrit (HCT)
14
Hemaray 86 / 89 Service Manual
.4 Performance
For performance indicators not specified in this section, they are applicable to both the
Whole Blood and Pre-diluted modes.
Blank Test
WBC ≤ 0.2 x /L
RBC ≤ 0.02 x /L
HGB ≤ 1g/L
PLT ≤ 10 x /L
HCT ≤ 0.5%
Carry-over
15
Hemaray 86 / 89 Service Manual
Parameter Carry-over, %
WBC 0.5
RBC 0.5
HGB 0.5
PLT 1
Repeatability
Linearity
16
Hemaray 86 / 89 Service Manual
.5 Scope of Application
The analyzer is mainly used to test human blood samples, make qualitative and quantitative
analysis of the visible components of blood, and provide the related information. It is suitable for
testing white blood cell quantity (WBC), red blood cell quantity (RBC), platelet quantity (PLT),
hematocrit (HCT), hemoglobin (HGB), and WBC 5-part differentiation. It is used in experiments
made by medical units, inspection units, disease control centers, scientific research institutions,
etc.
.6 Technical Parameters
Testing WBC/RBC/PLT: Electrical impedance method; HGB: Colorimetry;
Principle: WBC 5-part differentiation: Laser scattering method
Measuring
≤ 60sample/hour
Speed:
Display: External PC
Connector: Ethernet port
Work Temperature 15C~30C; RH 30%~85%; Air Pressure 70 kpa-
Environment: 106kpa
Storage
Temperature 0C~40C; RH ≤ 85%; Air Pressure 50 kpa-106kpa
Environment:
Transport Temperature -20℃~55℃; RH ≤ 93%; Atmospheric Pressure 50
Environment: kPa~106 kPa
Mains Input: a.c.100V-240V, 50Hz/60Hz
Input Power: ≤ 600VA
Instrument
Standby ≤ 60dB; Running ≤ 66.86DB
Noise:
.7 PC Configuration
This instrument uses the PC terminal software control instrument, and the operational
software is installed on the PC. The recommended PC configuration is as follows:
CPU: 1.6GHz or above; Memory: 512MB or above, HDD Capacity: 160GB or above. Best
Display Resolution: 1366*768; OS: Microsoft Windows XP or above. The PC has at least 2
network ports (one is used to connect to the instrument, and the other to the LIS/HIS) and 2
USB ports.
Before using the instrument, set the IP address of the PC to the fixed address 192.168.1.64
and the subnet mask to 255.255.255.0. For the specific operations, refer to the Help file of
Windows.
To connect to the LIS, in System Setup – Communication Setup, set the IP address and
port of the LIS. If the LIS is installed on the same PC, the IP address is set to 127.0.0.1. If the
LIS is not installed locally and the IP address is in the same network segment as “192.168.1.64”,
17
Hemaray 86 / 89 Service Manual
the address of the instrument needs to be changed to one not in the same network segment as
“192.168.1.64”, that is, the third number “1” in the address “192.168.1.64” is changed to N (a
number from 2 to 250) (i.e. “192.168.N.64”), and the fourth number “64” can be changed
between 1 and 250.
If the OS is Microsoft Windows XP, in order to get the best display effect of the analysis
software, the Microsoft YaHei font needs to be installed.
18
Hemaray 86/89 Service Manual
Chapter 2 Installation
.8 Unpacking
.8.1 Steps of Unpacking
Unpack the instrument and remove the materials for transport. Keep the packing case
and packing materials properly for future repacking.
1) Put the packing case upright and ensure the arrow on the packing case points upward.
Open the packing case with tools, take out the accessories, and check the items
against the packing list.
2) Take out the upper buffer. At least two persons carefully take the instrument from the
packing case by grasping the carrying handles at the bottom on both sides of the
instrument and put it on a level operation desk.
19
Hemaray 86 / 89 Service Manual
20
Hemaray 86 / 89 Service Manual
1. Connect the Hemaray 86\89 to the patch board or receptacle using power supply cable.
2. Connect the Hemaray 86\89 to PC using LAN cable.
3. Connect the PC to the patch board or receptacle using power supply cable.
4. Connect the printer to the patch board or receptacle using power supply cable and to the
PC using the signal cable.
Again confirm that all the devices are correctly connected and respectively turn the power
supply outlet switch, Hemaray 86\89 instrument, and PC switch ON. Confirm that the IP
address of the PC is“192.168.1.64”.Install Hemaray 86\89 system software on the PC.
(Default installation address: D:\Program Files\Hemaray. Ensure no other software is
installed and no other files are stored in D: disk.) . After the installation, open the “Automated
Hematology Analysis Software” on the desktop of the PC, and input the username “5x3041”
and the corresponding password. Click “OK” and immediately press the key combination
“ALT+F4” on the keyboard to directly enter the Test interface. The initialization will be
conducted, during which, if the dialog box “The configuration of the PC is inconsistent with
that of the instrument” is pops up, be sure to use the configuration on the instrument. Then
1) Enter “Menu → Service → Mechanical Check” as shown in the following figure:
Check each part. It's important to note that for the syringe and sampling probe tests, after
each part is tested, the software will give prompt. Do not test other parts before the Finish
prompt appears to avoid faults; for the valve test, the valve is normal only when 25 valve
switching sounds are heard, or the valve can be tested one by one.
21
Hemaray 86 / 89 Service Manual
2) Enter “Menu → Service → Adjustment → Probe Pos.” to Detect the various positions
of the sampling probe.
The Hemaray 86/89 interface is as shown in the following figure:
You do not need to conduct any Detect and just confirm the number of steps is the
corresponding value。
Vertically, the value of “Upper Pos.” is 0;
the value of “Isolation Bubble Pos.” is 100;
the value of “Autoloader Sampling Pos.” is 800;
the value of “closed Sampling Pos.” is 800;
the value of “First Piercing Pos.” is 400.
[1] When Detecting the various positions, be sure not to click the “Adjust” button at will,
otherwise, the probe position may be wrong. Please bear this in mind。
[2] Before the Detect, please conduct “Assy Init.”, and doing so is required after each position
is Detected.
[3] Firstly, Detect whether the horizontal position is accurate. For example, to Detect “DIFF
Bath Pos.”, click the “Detect” button under the DIFF bath position and judge whether the
sampling probe assy moves above the DIFF bath. The judgment method is: click the
“V.Motor Power-off” button, move down the sampling probe with hand, and observe whether
the sampling probe can be inserted into the DIFF bath near the middle of the bath. If it can’t
enter or the deviation is big, firstly check whether the horizontal belt is loose. If it is loose,
adjust the belt and conduct the Detect again. Be sure to conduct “V.Motor Power-on”
(corresponding to the aforesaid “V.Motor Power-off”) and then “Assy Init.” before the next
Detect, otherwise, the initialization will go wrong. If the horizontal belt is not loose, the
position of the probe needs to be adjusted. The adjustment method is as follows: horizontally,
click the “Adjust” button under “DIFF Bath Pos.” to pop up “Move the probe to the appropriate
position, and click the ‘OK’ button!” prompt box, and move the sampling probe horizontally
22
Hemaray 86 / 89 Service Manual
with hand to the middle of the DIFF bath; when the movement is finished, click “OK” on the
prompt box, and the sampling probe will be reset horizontally; click “Read”; click the “Setup”
button. When the position adjustment is finished, conduct confirmation Detect of the results
of adjustment. For the method, see Steps [2] and [3].
[4] The detect steps of other positions are the same as Step [3].
[5] For vertical detecting, the standard for judging whether the position is correct is: the tip
of the sampling probe is about 2mm away from the bottom of the “Bath” or “Tube”.
For horizontal Detecting, the standard for judging whether the position is correct is:
the sampling probe is near the middle of the bath.
3) Enter “Menu → Service → Adjustment → Autoloader” to Detect the various positions of the
automatic sampler.
Detecting the resetting position of “Mix Motor”: Click the "Initialize" of “Mix Motor”
and then the grip is at horizontal position, not inclined. When the grip moves upwards, it
must be 2 mm or more from the sampling assembly. Otherwise, adjust the "Reset and
Correct" quantity behind the “Detect Mix Motor”. Click the "Initialize" of this motor and
again conduct Detecting until the requirements are satisfied. Please note that it is
necessary to click "Setup" for effective adjustment of parameters after the resetting and
correction quantity is adjusted.
“Elevating motor: Detect the resetting position: Push one sample holder to the
Detect tube position. Click the "Initialize" behind the “Elevating Motor” and then
manually pull the grip over the sample holder. It is required that the lower gripping edge
is about 3 mm from the sample. If this condition fails to be met, adjust the parameters
behind the “Elevating Motor” and again Detect until the requirements are met. Please
note that it is necessary to click "Setup" for effective adjustment of parameters after
adjustment.
Detecting the Detect gripping: Push the sample holder containing two Detect tubes
to the Detect gripping position. Click the "Detect" at this position and then the Detect
tube is just positioned on the center of the grip. When clicking the "Detect" behind the
“Elevating motor”, the Detect tube is able to smoothly leave away and be put into the
23
Hemaray 86 / 89 Service Manual
sample holder. Otherwise, adjust the parameters behind the "Pinch Pos.” Click the
"Detect" at this position until the above-mentioned requirements are met. Please note
that it is necessary to click "Setup" for effective adjustment of parameters after
adjustment.
Detecting other parameters: The defaulted parameters in the above figure are
used for others.
Detecting scanner position: Push a sample holder containing empty tubes under
the scanning rotary motor. Click the "Detect" behind the "Feeding Steps" and then click
the "Turn on scanner’s Laser". It is required that the light emitted from the scanner is
just directed the center of the Detect tube. If not, adjust the installation position of the
scanner. Upon completion, click the "Turn off scanner’s Laser".
4) Check the pipes connection, confirm there are no pipes detached, enter “Menu → Service
→ Daily Maintenance → Fluidics Maint.”, and conduct “Prime” for the Fluidics respectively.
.16 Debugging
1. Blank Test
Blank Test in Automatic-Whole Blood Mode:
Click “Sample Analysis” on the top of the software interface, click the “Next(N)” button on
the “Sample Analysis” interface, and make the setup shown in the following interface:
Click the “OK” button to enter the Test interface in the Enclosed-Whole Blood mode and
conduct the blank test. If the blank is not up to standard, conduct the blank test again. If it is
still not up to standard, clean the related pipes (the easiest method is: enter “Menu →
Service → Daily Maintenance → Clean” to conduct “Fluidics Clean”), and measure the blank
till the blank meets the requirements listed in Table 2-2.
Blank Test in Enclosed-Whole Blood Mode:
24
Hemaray 86 / 89 Service Manual
Click “Sample Analysis” on the top of the software interface, click the “Next(N)” button on
the “Sample Analysis” interface, and make the setup shown in the following interface:
Click the “OK” button to enter the Test interface in the Enclosed-Whole Blood mode and
conduct the blank test. If the blank is not up to standard, conduct the blank test again. If it is
still not up to standard, clean the related pipe (the easiest method is: enter “Menu → Service
→ Daily Maintenance → Clean” to conduct “Fluidics Clean”), and measure the blank till the
blank meets the requirements listed in Table 2-2.
25
Hemaray 86 / 89 Service Manual
Conduct sample analysis on the interface with normal-level Quality Control, observe
whether the positions of the “Lymphocyte Peak” and “RBC Peak” are within the target value
deviation range corresponding to the peak given in the Quality Control Target Value Table
and whether “HGB Blank AD” is between 2500 and 2900.
If “Lymphocyte Peak” is on the large side, decrease the value of “WBC Gain”; if
“Lymphocyte Peak” is on the small side, increase the value of “WBC Gain”.
If “RBC Peak” is on the large side, decrease the value of “RBC Gain”; if “RBC Peak” is
on the small side, increase the value of “RBC Gain”.
If “HGB Blank AD” is on the large side, decrease the value of “HGB Gain”; if “HGB Blank
AD” is on the small side, increase the value of “HGB Gain.
3. Standard Particle Test
Enter the “Menu” → “Service” → “Adjustment” → “DIFF Gain (Particle)” interface.
26
Hemaray 86 / 89 Service Manual
Dilute 7um standard particles into liquid with the total quantity of 3000~5000 for testing.
Requirements: the CV values of FS and FL are below 7%.
If the CV value of FS or FL does not meet the requirements, the inner and outer walls of
the sheath flow cell need to be cleaned.
Note: The method for diluting standard particles is: add about 1ml deionized water to the
bullet-shaped container and then add 2 drops of standard particles, and blend them for
testing.
4. Fresh Blood Test
Enter the “Menu” → “Service” → “Adjustment” → “DIFF Gain (Blood)” interface.
27
Hemaray 86 / 89 Service Manual
Use at least 3 tubes of normal fresh blood for testing. The requirements are as follows:
LOWER
PARAMETER LIMIT UPPER LIMIT
CENTER OF
CENTER OF
CENTER OF
28
Hemaray 86 / 89 Service Manual
29
Hemaray 86 / 89 Service Manual
3. Click the “Browse” button, create a Hemaray folder under the root directory of D: disk, and
select the new folder as the installation directory. Click the “Next” button, and the program will
begin to install the software till the installation is ended.
30
Hemaray 86 / 89 Service Manual
4. Click the “Finish” button to finish the installation, and an “Automated Hematology Analyzer”
shortcut will be created on the desktop.
.17.3 Installation of Main Control Board System Program
.17.3.1 Programming FPGA
FPGA Programming Tool: USB-Blaster Connection Method: Connect the USB port to the PC,
and the other to the port J5 on the main control board.
1. Click “Start” and, from the “Programs” on the “Start” menu, run the programming software
(Quartus II 12.osp2 Programmer), as shown in the figure below:
2. On the window popped up, click “Hardware Setup”, as shown in the figure below:
3. From the “Currently selected hardware” option, select “USB-Blaster[USB-0]” and click
“Close”.
31
Hemaray 86 / 89 Service Manual
4. From the “Mode” option, select “Active Serial Programming” and click “Add File”. On the
window popped up, find the folder where the file to be programmed is located.
6. On the window popped up, tick the options shown in the above figure according to the
prompt and click “Start” to begin programming.
32
Hemaray 86 / 89 Service Manual
33
Hemaray 86 / 89 Service Manual
2. Select the Main board application Software to be upgraded with the “。。。” button,
check that the PC and the instrument have been connected to the network, and click the
“Download” button. The progress bar will prompt the downloading progress. After the file is
downloaded, the Main board application Software will be turned off automatically, and the
new version software will run again. You can check the version information of the main board
application in Service -> Version Information.
34
Hemaray 86 / 89 Service Manual
5. In the download setup, click the dropdown box “Part” and select “C8051F340” from the
dropdown box. In the “Settings” check box, select the “Erase code space before
programming” option. In the “Hex File Location(s)” option box, click the “Open File” button on
the top right corner. Select to open the BootLoad.hex boot loader file. Configure
Programming Settings are shown in the figure below:
Note: After the “Configure Programming Settings” dialogue box is opened, when the “Debug”
option box is not in the format of TSxxxxxxxx (such as TS00014857), it indicates that the
download cable is not connected properly. Here, you should close
MCUProductionProgrammer.exe, reconnect the USB properly, and open
MCUProductionProgrammer.exe again for setup.
6. After the Configure Programming Settings are finished, click “Accept Settings” botton to
return to the main interface of the downloader, shown in the figure below:
35
Hemaray 86 / 89 Service Manual
7. On the main interface of the downloader, click the “Program Device” button to begin to
download the boot loader. When the programming prompt box displays “Device Erased,
Programmed and Verified”, it indicates the programming is finished. Turn off the power of the
board and unplug the download cable from the board.
36
Hemaray 86 / 89 Service Manual
37
Hemaray 86 / 89 Service Manual
According to the mechanical molding design and the whole machine layout design, its
overall hardware structure is divided into three assys, i.e. electrical circuit unit, power supply
assy, and shell assy.
Driver board
Main control
board
38
Hemaray 86 / 89 Service Manual
Burning board
Forward small-angle
preamplification
board
Sampling assy
Sampling assy interface board
adapter board
39
Hemaray 86 / 89 Service Manual
The boards are directly fixed to the corresponding position on the respective assy with
screws. Seven M3 screws are used for the driver board; six for the main control board; four
for the burning board; three for the sampling assy adapter board; four for the sampling assy
Interface board; four for the laser driver board, forward small-angle preamplification board,
forward large-angle preamplification board, and side scatter preamplification board; four
screws are used for the automatic sampling driver board. There is a partition between the
main control board and the driver board as a shield.
40
Hemaray 86 / 89 Service Manual
24V power
5V power
Power switch
Filter
LED display
board
Key board
41
Hemaray 86 / 89 Service Manual
1) Embedded Platform
The embedded platform uses ARM9 as the core. The peripheral circuits include real-time
clock (RTC), SDRAM, NAND FLASH, Ethernet, and UART circuits. In which, ARM9, SDRAM,
and NAND FLASH constitute the basic running platform of the software; RTC provides the
calendar and real-time clock; Ethernet provides the 10/100Mbps adaptive network; UART
provides the RS-232C circuit.
2) Digital Logic Processing Unit.
The digital logic processing unit uses FPGA as the core. The auxiliary circuit includes
serial FLASH download and 2.5V DC-DC power supply circuits. In which, the serial FLASH
mainly stores the FPGA logic; the 2.5V DC-DC power supply provides 2.5V DC current to the
FPGA phase-locked loop circuit.
3) Analog Conditioning and ADC
Analog conditioning and ADC is divided into DIFF and CBC analog conditioning. In
which, the DIFF conditioning is divided into blocking, buffer, digital gain adjustable amplifying,
active power filter, and buffer and pipelined ADC processing circuits, while the CBC
conditioning is divided into blocking, one-stage amplifying, two-stage amplifying, RC filter,
three-stage digital gain adjustable amplifying, buffer, and pipelined ADC circuits.
.21.2 Principle
The specific functional block diagram is as follows:
42
Hemaray 86 / 89 Service Manual
43
Hemaray 86 / 89 Service Manual
44
Hemaray 86 / 89 Service Manual
45
Hemaray 86 / 89 Service Manual
(PWR_SW);
PIN No. Assignment Description
1 PWRDWN When the software is shut down, the
power supply Off signal is controlled.
2 NC Blank
3 GND Digital ground wire
46
Hemaray 86 / 89 Service Manual
(LED&AUTO_KEY);
PIN No. Assignment Description
1 +12VA +12V power supply
2 LED_LASER Indication signal of laser indicator
3 +3.3V +3.3V power supply
4 GND Digital ground wire
5 LED_RUNNING Indication of normal system running of
LED indicator
6 GND Digital ground wire
7 LED_WARNING Indication of system warning of LED
indicator
8 GND Digital ground wire
47
Hemaray 86 / 89 Service Manual
48
Hemaray 86 / 89 Service Manual
49
Hemaray 86 / 89 Service Manual
50
Hemaray 86 / 89 Service Manual
small-angle signals from the optical path and converts weak current
51
Hemaray 86 / 89 Service Manual
large-angle signals from the optical path and converts weak current
52
Hemaray 86 / 89 Service Manual
scatter signals from the optical path and converts weak current
(1) Laser Slow Start: As an impulse current is produced when the power switch is turned on,
53
Hemaray 86 / 89 Service Manual
in order to prevent such current from destroying the laser diode, the driving source of the
laser must be designed as Slow Start. The circuit is mainly composed of 2 NPN triodes and
multiple capacitors, and the voltage increases slowly through capacitor charging and triode
turn-on.
(2) Constant-current Driving: The light output of laser diode relates to the drive current. Laser
is produced only when the drive current passing through the laser is above the threshold.
(3) Automatic Power Control (APC): In the constant-current driving mode, the temperature
changes influence the laser output. APC circuit must be added to reduce the influence of
temperature on the output to a certain extent. The circuit is mainly composed of the reference
voltage output, signal amplification, etc. The laser output must be stabilized through
regulation of the drive current according to the signal value at the feedback end of the laser
diode.
(4) Current Detection: For more convenient debugging, the current detection function is
added.
54
Hemaray 86 / 89 Service Manual
terminal
6 \ \
55
Hemaray 86 / 89 Service Manual
9 \ \
(welded) J2 (welded)
2 (GND) GND
output terminal
control signal
Point
56
Hemaray 86 / 89 Service Manual
Interface Section: Including Interface of sensor signal input, drive circuit, and control
section
Drive Section: Composed of stepper motor drive and pump valve drive
The driver board is composed of 5 MCUs, one of which is used as the dispatching center
and the other 4 are used as parts control software to form a master-slave structure.
commands from the main board application, issue the commands to the parts
control software units, and return the execution results to the main board
application;
MCU2 (U20) realizes control of sampling syringe motor, diluent syringe motor,
Waste Syringe motor and V23 valve and testing of 1 air pressure sensor;
MCU3 (U22) realizes control of level motor, vertical motor, valve, and pump of
57
Hemaray 86 / 89 Service Manual
MCU4 (U23) realizes control of sample syringe motor, sheath fluid syringe motor
and sample syringe motor and testing of 5 liquid sensors, 1 liquid pressure sensor,
Optocoupler Sensor Testing: Optocoupler signals after shape correction are input to
the IO port of the FPGA and provide access by the corresponding CPU;
Liquid Sensor Testing: Signals after processing are input to the analog input port of
Air Pressure Sensor and Liquid Pressure Sensor Testing: Signals after processing
and amplification are input to the analog input port of the corresponding CPU.
Pump Valve and Heating Film Drive: Provides at least 1.5A current drive.
No.
3 Power Supply, 24 V
5 Power Supply, 24 V
7 Power Supply, 24 V
9 Power Supply, 24 V
11 Power Supply, 24 V
13 Power Supply, 24 V
58
Hemaray 86 / 89 Service Manual
15 Power Supply, 24 V
17 Power Supply, 24 V
19 Power Supply, 24 V
21 Power Supply, 24 V
23 Power Supply, 24 V
25 Power Supply, 24 V
27 Power Supply, 24 V
29 Power Supply, 24 V
31 Power Supply, 24 V
33 Power Supply, 24 V
35 Power Supply, 24 V
37 Power Supply, 24 V
39 Power Supply, 24 V
2 V1 Control Signal
3 Power Supply, 24 V
59
Hemaray 86 / 89 Service Manual
4 V2 Control Signal
5 Power Supply, 24 V
6 V3 Control Signal
7 Power Supply, 24 V
8 V4 Control Signal
9 Power Supply, 24 V
10 V5 Control Signal
11 Power Supply, 24 V
12 V6 Control Signal
13 Power Supply, 24 V
14 V7 Control Signal
15 Power Supply, 24 V
16 V8 Control Signal
17 Power Supply, 24 V
18 V9 Control Signal
19 Power Supply, 24 V
21 Power Supply, 24 V
23 Power Supply, 24 V
25 Power Supply, 24 V
27 Power Supply, 24 V
29 Power Supply, 24 V
31 Power Supply, 24 V
33 Power Supply, 24 V
60
Hemaray 86 / 89 Service Manual
35 Power Supply, 24 V
37 Power Supply, 24 V
39 Power Supply, 24 V
2 24V
3 24V
4 24V
5 GND
6 GND
7 GND
8 GND
2 Signal-
3 GND
4 Signal+
2 3.3V
3 GND
4 Signal
2 3.3V
3 GND
4 Signal
61
Hemaray 86 / 89 Service Manual
2 3.3V
3 GND
4 Signal
2 3.3V
3 GND
4 Signal
2 3.3V
3 GND
4 Signal
3 TXD
4 GND
2 Signal
3 3.3V
sensor 2 2 Signal
3 3.3V
diluent syringe
2 GND
3 5V
5 GND
6 5V
62
Hemaray 86 / 89 Service Manual
Sample Syringe
8 GND
9 5V
11 GND
12 5V
Lyse Syringe
14 GND
15 5V
Sampling Syringe
17 GND
18 5V
Horizontal Motor of
Sampling Assembly
20 GND
21 5V
22 NC
Assembly
2 GND
3 5V
Assembly
2 GND
3 5V
63
Hemaray 86 / 89 Service Manual
2 Motor Signal A-
3 Motor Signal B+
4 Motor Signal B-
3 Motor Signal B+
4 Motor Signal B-
2 Motor Signal A-
3 Motor Signal B+
4 Motor Signal B-
2 Motor Signal A-
3 Motor Signal B+
4 Motor Signal B-
2 Motor Signal A-
3 Motor Signal B+
4 Motor Signal B-
2 Motor Signal A-
3 Motor Signal B+
4 Motor Signal B-
3 Motor Signal B+
4 Motor Signal B-
2 Motor Signal A-
64
Hemaray 86 / 89 Service Manual
4 Motor Signal B-
Measuring
Points
GND GND
Syringe
Syringe
Liquid Syringe
Liquid Syringe
65
Hemaray 86 / 89 Service Manual
(2) DC-DC Conversion: With the DC-DC chip as the core, including peripheral booster circuit.
The main working principle is to realize DC 110V constant voltage output through
(3) Output Control: Mainly includes relay, optocoupler isolation, and MOS control circuits.
Switching between high voltage burning and signal transmission is realized through the
66
Hemaray 86 / 89 Service Manual
67
Hemaray 86 / 89 Service Manual
J6 B4B-PH-KS
1 GND \
5V Power Supply
2 5V \
3 GND \
24V Power Supply
4 24V \
J5 B3B-PH-KS
2 2 GND GND
signal
J2 B3B-XH-A JST
electrode
68
Hemaray 86 / 89 Service Manual
J4 B3B-XH-A JST
electrode
J1 B3B-XH-A JST
electrode /110V
J3 B3B-XH-A JST
electrode /110V
69
Hemaray 86 / 89 Service Manual
70
Hemaray 86 / 89 Service Manual
71
Hemaray 86 / 89 Service Manual
72
Hemaray 86 / 89 Service Manual
power supply
13 HOME_05# Optocoupler 05# reset signal
14 OPT05#_SPY Optocoupler 05# transmitting terminal
power supply
15 GND Digital ground wire
16 GND Digital ground wire
17 HOME_06# Optocoupler 06# reset signal
18 OPT06#_SPY Optocoupler 06# transmitting terminal
power supply
19 HOME_07# Optocoupler 07# reset signal
20 +5V +5V power supply
21 GND Digital ground wire
22 GND Digital ground wire
23 HOME_08# Optocoupler 08# reset signal
24 OPT08#_SPY Optocoupler 08# transmitting terminal
power supply
25 HOME_09# Optocoupler 09# reset signal
26 OPT09#_SPY Optocoupler 09# transmitting terminal
power supply
27 GND Digital ground wire
28 GND Digital ground wire
29 HOME_010# Optocoupler 010# reset signal
30 OPT010#_SPY Optocoupler 010# transmitting terminal
power supply
73
Hemaray 86 / 89 Service Manual
74
Hemaray 86 / 89 Service Manual
75
Hemaray 86 / 89 Service Manual
1. Basic Functions
Uses the LED lighting sets and indicates the work status of the
2. Principle
76
Hemaray 86 / 89 Service Manual
77
Hemaray 86 / 89 Service Manual
1. Basic Functions
2. Principle
78
Hemaray 86 / 89 Service Manual
1. Basic Functions
Key testing.
2. Principle
79
Hemaray 86 / 89 Service Manual
1. Basic Functions
2. Principle
80
Hemaray 86 / 89 Service Manual
81
Hemaray 86/89 Service Manual
82
Hemaray 86 / 89 Service Manual
83
Hemaray 86/89 Service Manual
84
Hemaray 86 / 89 Service Manual
M a i n C o n t r o l B o a r d J 1 7
1-5-082-027-10 RT861.85.23 Door Open/Close Cable
Suction and Door Microswitches
Main Control Board J18-Key Board
1-5-082-029-10 RT861.85.25 Key Board Connection Cable
and LED Board
M a i n C o n t r o l BFooar rwda rJ d7 -
R T 8 6 1 . 8 5 . 2 6 F o r w a r d S m a l l - a n g l e
1-5-082-030-10 S m a Pl rl e- aa nm gp
Cable
Board FS
Main C o n t r o l BFooar rwda rJ d8 -
R T 8 6 1 . 8 5 . 2 7 F o r w a r d L a r g e - a n g l e S
1-5-082-031-10 L a r gP er -e aa nm g p
Cable
Board FL
R T 8 6 1 . 8 5 . 2 8 S i d e S c a t Mt ea r ii nn g C
A no g nl et rSSoii gld ne aB l o a
1-5-082-032-10
Cable Preamplification Board SS
M a i n C o n t r o l L B
a so ea r d J 1
1-5-082-033-10 RT861.85.29 Laser Driver Cable
Driver Board J1 and Laser Lock
R T 8 6 1 .H 8o 5r .i 3z M1o on t t Coa orl n n e Hc ot ri i oz nM
o no tt afo olr S
r a m p l i n g
1-5-082-035-10
Cable Assy-Driver Board J7
L y s e S y r i n Lg Ye - DM r oi tv oe r r
1-5-082-036-10 RT861.85.32 Lyse Syringe Cable
Board J3
Sh e at h F l uid Sy r in gS
e HM- o t o r
1-5-082-037-10 RT861.85.33 Sheath Fluid Syringe Cable
Driver Board J2
D i l u e n t S y r i n gD eI - M
D ro it vo er r
1-5-082-038-10 RT861.85.34 Diluent Syringe Cable
Board J1
S a m p S
l i yn r gi n g e A MS oP t - o r
1-5-082-039-10 RT861.85.35 Sample Suction Syringe Cable
Driver Board J4
S a m p l e S y r i nMgoet o rS P - D r i v e r
1-5-082-040-10 RT861.85.36 Sample Syringe Cable
Board J5
D r i v eB ro a r Jd 8 - S a m p l i n g Assy
1-5-082-041-10 RT861.85.37 Vertical Motor Adapter Cable
Adapter Board J3
D r i v eB
r o a r dJ 1 - S a m p l i n g A s s y
1-5-082-044-10 RT861.85.40 Vertical Position Sensor Cable
Adapter Board J13
Driver Board J10-Position Sensors
1-5-082-046-10 RT861.85.42 Position Sensor Connection Cable (DI\SH\SA(SP)\WA\HE(LY),
AS(ASP)\HO (Horizontal))
D r i B
v eo ra Jr 2d 9 - D I F F B
1-5-082-047-10 RT861.85.43 Heater Cable
Heating Film
1-5-082-048-10 RT861.85.44 Pump Connection Cable Driver Board J16-Pump
D r Bi vo Jea R 3rr o 9d o- m
1-5-082-049-10 RT861.85.45 Room Temperature Sensor Cable
Temperature Sensor
S a m p l i n g A s s y C o n nB e co taorrd
R T 8 6 1 . 8 5 . 4 6 S a m p l i n g A s s y T r a n s m i s s i
1-5-082-050-10 J 4 -S a m p l i n g A s s y A d a pBt eora r d
Cable (Enclosed)
J2
1-5-082-051-10 RT861.85.47 Valves V1-V13 Driver Board J37-Valves V1-V13
1-5-082-052-10 RT861.85.48 Valves V21-V32 Driver Board J38-Valves V21-V32
85
Hemaray 86 / 89 Service Manual
86
Hemaray 86 / 89 Service Manual
System, main board application, and parts control software. The parts control software is mainly
responsible for drive control of mechanical components, and has the functions of temperature
and pressure testing. The main board application realizes storage of instrument configuration
parsing of sequential action, collection of signal data, etc. The Information Processing System is
responsible for interaction with users, operating and controlling instrument running, displaying
main board application and parts control software will not be described.
87
Hemaray 86 / 89 Service Manual
88
Hemaray 86 / 89 Service Manual
89
Hemaray 86 / 89 Service Manual
I n w h i c h P , of w
o re r “ I on fsf t r u m e n t ” , p o n w leyr - t oh f ef
procedure is executed, and you will not exit the Information
Processing System analysis software . For “Exit System”, after
t h e s a m e o p e r a t i o n aP so wt he ar t Ionff o
sf tr r u
“ ment” is
executed, the system will prompt whether to exit the Information
Processing System software.
90
Hemaray 86 / 89 Service Manual
.30.2 Syringes
91
Hemaray 86 / 89 Service Manual
92
Hemaray 86/89 Service Manual
Sampling probe
sucks blood
sample 20μL
Diluent Diluent
615μL 2008.9μL
5D Lyse 375μL
5D Lyse 375μL
93
End of Test
Hemaray 86 / 89 Service Manual
and the liquid pump P1 drains the base solution from the DIFF bath.
2. Spitting diluent, distributing blood sample, and adding 86D lyse to DIFF bath: Power
94
Enter Flow Chamger
DIFF Bath WBC Bath
WBCBath
WBC Test
Bath RBC Bath Blending
WBC 5-part
BlendingDifferentiation
(dilution WBC
Flow Chamger Cleaning/Burning
HGB
SecondaryMeasurement
Blending RBC Bath
(dilution ratio
Constant Temp Measurement
ratio 1:250) Blending (dilution
End of Test Sampling probe sucks initial diluent 2668.7μL
Cleaning 5H Lyse 200μL RBC and PLT
Cleaning/BurningTests
Diluent
(dilution ratio 1:750)
ratio 1:698) 1:24000)
Hemaray 86 / 89 Service Manual
on and open the SV11 valve, and the 86D DIL syringe spits diluent to the DIFF bath. The
sampling probe moves to the DIFF bath, and the ASP syringe distributes 10ul blood
sample from the sampling probe to the DIFF bath. At the same time, power on and open
the SV10 valve, and the 86D LY syringe spits lyse to the DIFF bath.
3. DIFF bath sample bubble blending: Power on and open the SV30 and SV28 valves,
and the Waste Syringe establishes positive pressure (5.5~6.5Kpa) and adds bubble to
95
Hemaray 86 / 89 Service Manual
Schematic Diagram of Spitting Diluent, Distributing Blood Sample and Spitting Lyse to DIFF
Bath
96
Hemaray 86 / 89 Service Manual
the diluted and blended sample from the DIFF bath. After the suction, the pinch valve
SV29 (pipe stop valve) is powered on, and the sample is stored in the pipe.
2. Sheath flow formation: Power on and open the SV13, SV6, and SV25 valves, and
the SH syringe spits diluent to the Flow Cell at a certain speed to form a stable sheath
flow.
3. Rapid sample pushing: Power on and open the SV4 valve, and the sample is rapidly
pushed to the Flow Cell with the bypassing effect of the sample pushing of the SH
syringe. At the same time, before the SV4 valve is closed, the SP syringe pushes the
4. DIFF channel test: After the sample reaches the test area of the Flow Cell, the SP
syringe pushes the sample to the Flow Cell at a low speed. The sample, wrapped by the
sheath flow, passes through the test area of the Flow Cell steadily, and the test begins
97
Hemaray 86 / 89 Service Manual
14S.
98
Hemaray 86 / 89 Service Manual
99
Hemaray 86 / 89 Service Manual
100
Hemaray 86 / 89 Service Manual
for 2S for cleaning the Flow Cell and the Flow Cell outlet related pipes.
2 Back flushing sample probe in Flow Cell: Power off and close SV25 and power off
SV29 (the pipe is through), and the SH syringe pushes the diluent to back flush the
sample probe in the Flow Cell. At the same time, power on and open the SV1 valve, and
the DIL syringe pushes the diluent to clean the sample preparation pipe.
3 Cleaning sample preparation pipe: Power on and open the SV1 valve, and the DIL
4 Adding base solution to DIFF bath: After the DIFF bath is drained, power on and
open the SV1 valve, and the DIL syringe pushes 1.2ml diluent to the DIFF bath to soak
101
Hemaray 86 / 89 Service Manual
102
Hemaray 86 / 89 Service Manual
103
Hemaray 86 / 89 Service Manual
2. Draining RBC bath: Power on the SV26 and SV32 valves, and drain the RBC bath
3. Draining WBC bath: Power on the SV27 and SV32 valves, and drain the WBC bath
4. Spitting diluent to RBC and WBC baths: Power on and open the SV13 and SV7
valves, and the SH syringe will spit diluent to the RBC bath. Open the SV9 valve, and
104
Hemaray 86 / 89 Service Manual
105
Hemaray 86 / 89 Service Manual
bath position, and the ASP syringe distributes 6ul blood sample to the WBC bath.
2. Initial mixture blending in WBC bath: Power on and open the SV30 and SV27
valves, and the WA syringe will establish positive pressure (5.5~6.5Kpa) and add bubble
106
Hemaray 86 / 89 Service Manual
2. Secondary blending in WBC bath: Power on and open the SV30 and SV27 valves,
and the WA syringe will establish positive pressure (4.5~5.5Kpa) and add bubble to the
WBC bath.
107
Hemaray 86 / 89 Service Manual
RBC bath position, and the ASP syringe spits the initial diluted mixture sucked from the
2. Blending in RBC bath: Power on and open the SV30 and SV26 valves, and the WA
syringe will establish positive pressure (4.5~5.5Kpa) and add bubble to the RBC bath.
108
Hemaray 86 / 89 Service Manual
pressure (-23~-24Kpa).
2. Flushing back bath before testing: Power on and open the SV24 and SV21 valves,
and the testing pipes of the back baths of the WBC and RBC baths will be flushed.
3. WBC, RBC/PLT test: Power off and close the SV21 valve, and the SV24 valve is
kept on and opened. The CBC channel test begins after the liquid flow becomes stable
12S.
109
Hemaray 86 / 89 Service Manual
110
Hemaray 86 / 89 Service Manual
drain the liquid from the front bath of the WBC bath with the waste liquid pump. Power
on and open the SV13, SV7, and SV9 valves, and the SH syringe will spit 3.5ml diluent
2. Adding RBC base solution: Power on and open the SV26 and SV32 valves, and
drain the liquid from the front bath of the RBC bath. Power on and open the SV13 and
SV7 valves, and the SH syringe will spit 3.5ml diluent to the RBC bath to soak the RBC
bath.
3. Adding WBC base solution: Power on and open the SV13, SV7, and SV9 valves,
and the SH syringe will spit 2.5ml diluent to the WBC bath to soak the WBC bath.
4. Flushing back bath pipe: The WA syringe establishes negative pressure (-30~-
35Kpa). Power on and open the SV21 and SV24 valves and, after 2.5S, flush the testing
111
Hemaray 86 / 89 Service Manual
5. Burning Ruby Aperture: Start high pressure to burn the Ruby Aperture for 4S.
.33.3 Clean
1. Clean WBC bath: The front and back baths of the WBC counting chamber are
cleaned, in which, the front bath is cleaned with the initial bubble.
2. Clean RBC bath: The front and back baths of the RBC counting chamber are
cleaned, in which, the front bath is cleaned with the initial bubble.
3. Clean DIFF bath: The DIFF bath is cleaned with the initial bubble.
4. Fluidics Clean: The Fluidics cleaning is executed to clean all testing pipes and
counting chambers.
5. Clean probe: The inner and outer walls of the sampling probe module are cleaned,
6. Clean Flow Cell: The DIFF channel testing pipes is cleaned, including cleaning of
pipes for sample preparation, sample pushing, sheath flow, Flow Cell, etc.
.33.4 Upkeep
1. Unclog: The Ruby Aperture of the WBC/RBC counting chamber is zaped; positive
pressure is established to back flush the Ruby Aperture of the WBC/RBC counting
chamber, after which, the front and back baths of the WBC/RBC counting chamber are
cleaned.
2. Zap Apertures: The Ruby Aperture of the WBC/RBC counting chamber is zaped for
112
Hemaray 86 / 89 Service Manual
8S.
3. Back Flush: Positive pressure is established to back flush the Ruby Aperture of the
4. DIFF bath soak: The DIL syringe sucks 1.5ml concentrated cleanser through the
sampling probe. The DIFF testing channel related pipes (including pipes for sample
preparation, sample pushing, sheath flow, Flow Cell, etc.) are soaked for 5 minutes with
the mixture automatically diluted with the diluent spit at the ratio of 1:3 in the machine.
The pipes soaked with the cleanser are fully flushed with the diluent.
5. WBC/RBC bath soak: The DIL syringe sucks 1.8ml concentrated cleanser through
the sampling probe. The front and back baths of the WBC and RBC counting chambers
are soaked for 5 minutes with the mixture automatically diluted with the diluent spit at the
ratio of 1:3 in the machine. The front and back baths of the WBC and RBC counting
6. Fluidics conc. Cleanser soak: The DIL syringe sucks 3.3ml concentrated cleanser
through the sampling probe. The DIFF testing channel related pipes and the front and
back baths of the WBC and RBC counting chambers are soaked for 5 minutes with the
mixture automatically diluted with the diluent spit at the ratio of 1:3 in the machine. The
pipes of the whole machine soaked are rinsed with the diluent.
7. Cleanser soak: The cleanser is sucked to soak the DIFF testing channel related
pipes and the front and back baths of the WBC and RBC counting chambers for 10
minutes (adjustable). The pipes of the whole machine soaked are rinsed with the diluent.
2. Empty RBC bath: The liquid in the RBC bath is emptied independently.
3. Empty DIFF bath: The liquid in the DIFF bath is emptied independently.
of service for more than 3 days. Firstly, empty the liquid in all pipes; then perfuse distilled
water into all pipes; and then drain the liquid in the pipes again.
113
Hemaray 86 / 89 Service Manual
7. Prime: All pipes of the analyzer are primed, including perfusion of diluent, cleanser,
Function: Spitting 180μl diluent for peripheral blood dilution and quantification.
2. Power-on after Normal Power-off
Function: After daily leak detection of related pipes and pipe cleaning upon normal
power-on of the analyzer, the background test is executed once. If the testing result does not
meet the background range, the background is tested once again.
3. Power-on after Abnormal Power-off
Function: After the leak detection of related pipes and pipe cleaning upon power-on after
the last abnormal power-off of the analyzer (the cleaning intensity is higher than that for
normal power-on), the background test is executed once. If the testing result does not meet
the background range, the background is tested once again.
4. Power-on after Drain and Power-off
Function: After the leak detection, automatic perfusion and automatic cleaning of pipes
upon power-on after the analyzer is drained, the background test is executed once. If the
testing result does not meet the background range, the background is tested once again.
5. Power-off Maintenance
Function: Daily power-off maintenance. The pipes of the counting chamber and DIFF
channel testing pipes of the analyzer are soaked with the cleanser. The pipes of the Fluidics
soaked are rinsed.
6. Sleep
Function: The laser is turned off; the WBC/RBC bath is topped up with reagent for
soaking; the instrument is in the sleep mode (is standby).
7. Quit Sleep 1
Function: Used when the sleep time of the instrument is less than 1 hour. The counting
chamber and Flow Cell related pipes are cleaned.
8. Quit Sleep 2
Function: Used when the sleep time of the instrument is greater than 1 hour and less
than 3 hours. The counting chamber and Flow Cell related pipes are cleaned.
9. Quit Sleep 3
Function: Used when the sleep time of the instrument is greater than 3 hours. The
counting chamber and Flow Cell related pipes are cleaned.
10. Quit Sleep 4
Function: Used when the sleep time of the instrument is greater than 6 hours. The
counting chamber and Flow Cell related pipes are cleaned.
11. Negative Pressure Self-test
114
Hemaray 86 / 89 Service Manual
Function: The sequence is executed during power-on to conduct leak detection of DIFF
testing channel pipes. The detection positions include sheath flow pipe, sample pipe, and
Flow Cell.
14. WBC/RBC Channel Pipe Leak Detection
Function: The sequence is executed during power-on to conduct leak detection of CBC
channel pipes. The detection positions include drain pipe and CBC testing pipe.
115
Hemaray 86/89 Service Manual
116
Hemaray 86 / 89 Service Manual
117
Hemaray 86 / 89 Service Manual
118
Hemaray 86 / 89 Service Manual
119
Hemaray 86 / 89 Service Manual
120
Hemaray 86 / 89 Service Manual
121
Hemaray 86 / 89 Service Manual
122
Hemaray 86 / 89 Service Manual
2. Closing
Reverse the steps of opening the top shell to close it.
.35.3 Dismantling of Shell of Automatic Sampler
1. Removal
As shown in the figure, open the left and right doors and screw out the screws (at total 4
positions) indicated by the arrow with a cross screwdriver.
123
Hemaray 86 / 89 Service Manual
2. Mounting
Reverse the steps of removal to mount it.
.35.4 Dismantling of Automatic Sampling Driver Board
124
Hemaray 86 / 89 Service Manual
125
Hemaray 86 / 89 Service Manual
4. Mounting
Reverse the steps of removal to mount it.
126
Hemaray 86 / 89 Service Manual
sampler, and screw out the screws (at total 6 positions) indicated by the arrow with a cross
screwdriver.
127
Hemaray 86 / 89 Service Manual
2. Mounting
Reverse the steps of removal to mount it.
.35.6 Dismantling of Top Plate
1. Removal
As shown in the figure, screw out the 4 screws indicated by the arrow with a cross
screwdriver.
128
Hemaray 86 / 89 Service Manual
2. Mounting
Reverse the steps of removal to mount it.
2. Mounting
Reverse the steps of removal to mount it.
129
Hemaray 86 / 89 Service Manual
.35.8 Dismantling of Cooling Fan, Main Control Board, and Driver Board
1. Removal
Open the rear door, as shown in the figure:
1) To remove the cooling fan, screw out the 4 screws indicated by the arrow with a
cross screwdriver.
2) To remove the main control board, screw out the 6 screws indicated by the arrow
with a cross screwdriver.
3) To remove the driver board, screw out the 7 screws indicated by the arrow with a
cross screwdriver.
2. Mounting
Reverse the steps of removal to mount them.
130
Hemaray 86 / 89 Service Manual
2. Mounting
Reverse the steps of removal to mount it.
131
Hemaray 86 / 89 Service Manual
2) To remove the sampling assy adapter board, screw out the screws (at total 3
positions) indicated by the arrow with a cross screwdriver.
2. Mounting
Reverse the steps of removal to mount it.
132
Hemaray 86 / 89 Service Manual
133
Hemaray 86 / 89 Service Manual
2. Mounting
Reverse the above steps to mount the power supply assy.
134
Hemaray 86 / 89 Service Manual
2. Mounting
Reverse the above steps to mount the DIFF bath assy or counting chamber assy.
[2] Open the right door, as shown in the figure, and screw out the 4 screws with the
mark as indicated by the arrow with a cross screwdriver.
135
Hemaray 86 / 89 Service Manual
136
Hemaray 86 / 89 Service Manual
Open the right door, as shown in the figure, and screw out the 4 screws indicated by the
arrow with a cross screwdriver.
2. Mounting
Reverse the above steps to mount the waste liquid pump assy.
137
Hemaray 86 / 89 Service Manual
4) To remove the sheath fluid syringe, screw out the 4 screws indicated by the arrow with a
cross screwdriver, remove the sheath fluid syringe, and unplug the terminal blocks on its
motor and optocoupler.
5) To remove the lyse syringe, screw out the 4 screws indicated by the arrow with a cross
screwdriver, remove the lyse syringe, and unplug the terminal blocks on its motor and
optocoupler.
6) To remove the Waste Syringe, screw out the 4 screws indicated by the arrow with a
cross screwdriver, remove the Waste Syringe, and unplug the terminal blocks on its motor
and optocoupler.
138
Hemaray 86 / 89 Service Manual
2. Mounting
Reverse the above steps to mount the syringe assys.
139
Hemaray 86 / 89 Service Manual
Open the right door, as shown in the figure, and screw out the 2 screws that lock the
valve as indicated by the arrow with a cross screwdriver.
Open the left door, as shown in the figure, and screw out the 2 screws that lock the
valve as indicated by the arrow with a cross screwdriver.
140
Hemaray 86 / 89 Service Manual
2. Mounting
Reverse the above steps to mount the valves.
.35.18 Dismantling of Liquid Pressure Sensor
Open the left door, as shown in the figure, and screw out the 2 screws indicated by
the arrow with a cross screwdriver.
141
Hemaray 86 / 89 Service Manual
2. Mounting
Reverse the above steps to mount the optical assy.
.35.20 Dismantling of Board in Optical Assy
Remove the top plate and open the right door, as shown in the figure. Screw out the 8
screws indicated by the arrow with a cross screwdriver and remove the optical path cover.
142
Hemaray 86 / 89 Service Manual
Screw out the 5 screws indicated by the arrow with a cross screwdriver and remove the
optical path cartridge, as shown in the figure.
143
Hemaray 86 / 89 Service Manual
As shown in the figure, screw out the screws (at total 2 positions) indicated by the arrow
with an Allen wrench and screw out the screws (at total 2 positions) indicated by the arrow
with a cross screwdriver. Remove the side shielding shell, screw out the screws (at total 4
positions) indicated by the arrow with an Allen wrench, and replace the side preamplification
board.
144
Hemaray 86 / 89 Service Manual
2. Mounting
Reverse the above steps to mount the boards relating to the optical system.
Adjustment of laser driver board:
To adjust the laser drive current, make a trial first to determine the adjustment
direction, then slowly adjust the variable resistor VR1, and adjust the potential of the
test port LD_CURT to 3.3V. If the adjustment is too fast or the amplitude of
adjustment is too big, the laser may be damaged.
Requirements:
Testing Voltage: Position No. LD_CURT;
Voltage Range: 3.3V±0.05V;
Conversion to Laser Drive Current (mA):
I=LD_CURT/50/0.76 (mA)
The drive current of the laser is about 85mA;
The radiant power of the laser is about 15mW-22mW (test after focusing)
.35.21 Replacement of Flow Cell
1. Unplug the pipe of the Flow Cell. Prevent the reagent in the pipe from touching the
external surface of the Flow Cell.
2. Screw out the 4 screws shown in the figure.
145
Hemaray 86 / 89 Service Manual
146
Hemaray 86 / 89 Service Manual
.35.21.2 Replace with a new Flow Cell and conduct the position test by following the steps
described below
1. Adjust the position of the Flow Cell to make the light spot reflected by the Flow Cell
deviate from the slit for about 2mm, and tighten the screws.
The screw
The screw
2. Adjust the central axis of the Flow Cell into the light spot of the laser, as shown in the
figure below. Loosen the adjusting nut, adjust the bolt, and make the detection area of
147
Hemaray 86 / 89 Service Manual
Loosen the
adjusting nut
3. Observe the scattered light spots on both sides of the forward diaphragm for preliminary
judgment of whether the Flow Cell is adjusted in place.
4. If there
is an
oscilloscope, the Flow cell can be adjusted by capturing the presignal of the standard
particles. When the signal amplitude is a maximum, it indicates that the Flow cell is at the
most reasonable position.
148
Hemaray 86 / 89 Service Manual
If there is no oscilloscope, the Flow Cell can be adjusted with method of successive
approximation only by testing the standard particles or the sample.
5. At last, tighten the adjusting nut. Do not rotate the screw any more when tightening the
adjusting nut.
Otherwise, adjust the gains of FS and FL till the above requirements are met.
149
Hemaray 86 / 89 Service Manual
150
Hemaray 86 / 89 Service Manual
151
Hemaray 86 / 89 Service Manual
Add 50 to the reading after the input boxes for the cleanser sensor, diluent sensor, 86H
lyse sensor, and 86D lyse sensor, and input them to the respective input box.
Then click the “Setup” button under it.
.36.2 Debugging of Sampling Probe Position
1. Power on the machine and enter the “Menu” → “Service” → “Adjustment” → “Probe Pos”
interface.
2. DIFF Pos. (Horizontal) Debugging: Click the “Adjust” button, when “Move the probe to the
appropriate position,then click the ‘Detect’ button!” is popped up, put the tool of the DIFF bath
on the DIFF bath, and push the sampling probe into the hole of the tool manually. Then, click
“Detect” on the prompt box, and the sampling probe will be reset horizontally. Click “Read”,
152
Hemaray 86 / 89 Service Manual
and the edit box there will display the corresponding number of steps. Click the “Setup”
button.
3. Debug the horizontal positions of the autoloader piercing position, WBC bath position,
RBC bath position, and enclosed puncture position by referring to Step 2.
4. Upper Position Debugging: Directly set to “0” and click the “Setup” button.
5. Isolation Bubble Establishment Position Debugging: Directly set to “100” and click the
“Setup” button.
6. WBC (RBC) Position (Vertical) Debugging: Click the “Adjust” button, when “Move the
probe to the appropriate position and click the ‘Detect’ button!” is popped up, put the tool of
the WBC bath on the WBC bath, and pull the sampling probe to the bottom of the tool
manually. Then, click “Detect” on the prompt box, and the sampling probe will be reset
vertically and the edit box there will display the corresponding number of steps. Add 400 to
the value in the edit box and input the new value to the edit box. Click “Setup” to finish the
debugging.
7. DIFF Bath Position (Vertical) Debugging: Click the “Adjust” button, when “Move the probe
to the appropriate position and click the ‘Detect’ button!” is popped up, put the debugging tool
of the DIFF bath on the DIFF bath, and pull the sampling probe to the bottom of the
debugging tool manually. Then, click “Detect” on the prompt box, and the sampling probe will
be reset vertically and the edit box there will display the corresponding number of steps. Add
460 to the value in the edit box and input the new value to the edit box. Click “Set up” to finish
the debugging.
8. Autoloader Sampling Position (Vertical) and closed Sampling Position (Vertical)
Debugging: Directly set to “800” respectively and click the “Setup” button.
9. First Piercing Position (Vertical) Debugging: Directly set to “400” and click the “Setup”
button.
.36.3 Waste Syringe Pressure Calibration
Enter the “Menu” → “Service” → “Adjustment” → “Pressure Calib.” interface; connect the
pressure gauge to the pipe joint on the top of the Waste Syringe.
153
Hemaray 86 / 89 Service Manual
1. Positive Pressure Calibration: Click the “Reset” button on the interface, and input 100 to
the edit box with the unit of “AD”. Click the “Build” button, read the reading on the pressure
gauge connected to the pipe, and write it to the edit box with the unit of “KPa”. Click the “Add”
button. According to the above steps, test the pressure values 200, 400, 600, and 800
respectively. Then click the “Fit” button and click “Save” to finish the positive pressure
calibration.
2. Negative Pressure Calibration: Set the initial number of steps of negative pressure
establishment to “3500”. Click the “Reset” button on the interface, and input 100 to the edit
box with the unit of “AD”. Click the “Build” button, read the reading on the pressure gauge
connected to the pipe, and write it to the edit box with the unit of “KPa”. Click the “Add”
button. According to the above steps, test the pressure values 200, 400, 600, and 800
respectively. Then click the “Fit” button and click “Save” to finish the negative pressure
calibration.
1. Input “30” to the first edit box and “0” to the second, and click “Set up”. Insert the
temperature sensor of the thermometer into the distilled water in the DIFF bath. Wait 5
minutes, and the reading of the thermometer should be “30±0.1” degrees.
2. If the reading is out of the range of “30±0.1”, the test value should be reduced by 30. Input
the calculation result to the second edit box and click “Setup” till the reading of the
thermometer is “30±0.1” degrees. Suck the distilled water in the DIFF bath with the pipette.
154
Hemaray 86 / 89 Service Manual
1. Test a tpipe of QC blood or fresh human blood on the standard Analyzer three, calculate
the mean of the peak value of lymphocyte, MCV of RBC, and MPV of platelet as the
calibration reference value.
2. Conduct the test on the debugging instrument, and the test result should be basically
consistent with the calibration reference value (deviation of lymphocyte Peak: ±1.5FL;
deviation of RBC Peak: ±1FL). At the same time, the light intensity of HGB blank should be
between 2500-2900.
3. If the above results do not meet the requirements, adjust the gain values of the
corresponding items and click “Setup”. Retest the sample till the four items all meet the
requirements. Note: When the gain value increases, the light intensity of HGB, lymphocyte
peak, and RBC peak will increase.
155
Hemaray 86 / 89 Service Manual
1. Dilute the standard particles into liquid with the total quantity of 3000~5000 for testing.
Requirements: the CV values of FS and FL are below 7%; the PK.AD value of FS is between
386~398; the PK.AD value of FL is between 423~439.
2. If the CV value of FS or FL does not meet the requirements, the inner and outer walls of
the sheath flow cell need to be cleaned.
3. If the PK.AD value of FS or FL does not meet the requirements, adjust he gains of the
corresponding items and click “Setup”. Retest the sample till the PK.AD values of FS and FL
meet the requirements.
Enter the “Menu” → “Service” → “Adjustment” → “DIFF Gain (Blood)” interface.
4. Test a pipe of QC blood or fresh human blood on the standard analyser thrice (the
differentiation charts of the various cells must be separated), calculate the mean of the center
of gravity in the FS direction of neutrophil, center of gravity in the FL direction of lymphocyte,
and center of gravity in the SS direction of neutrophil as the calibration reference value.
156
Hemaray 86 / 89 Service Manual
5. Conduct the test on the debugging instrument, and the test result should be basically
consistent with the calibration reference value (deviation of center of gravity in all directions.
±10).
6. If the center of gravity value in the FS, FL or SS direction does not meet the requirements,
adjust the gain values of the corresponding items and click “Setup”. Retest the sample till the
values in the FS, FL and SS directions all meet the requirements.
.37 Test
.37.1 Functional Test
1. Earthing Check
Use the multimeter turned to 200Ω to test whether the resistance between all screws
and the ground terminal of the power supply on the rear side is less than 0.5 Ω. If it is greater
than 0.5, this item doesn’t meet requirement.
2. Software Version Check
Check whether the versions of the Information Processing System Application software,
main board software, driver board software, Autoloader board software, and Fluidics
sequence comply with the BOM (or contract review). If the versions are not compliant, this
item doesn’t meet requirement.
3. Sampling Probe Position Check
Enter the “Menu” → “Service” → “Adjustment” → “Probe Pos.” interface.
Horizontal Position Check:
Automatic Puncture Position Check: The parameter should be “0”. If it is not “0”, this item
doesn’t meet requirement. After the check, click the “Assy Initialization” button.
DIFF Bath Position Check (Horizontal): Put the inspection tool into the hole in the middle
of the DIFF bath. Click the “Detect” button there and, when the sampling assy stops motion,
click the “V.motor Power-off” button. Pull down the sampling probe assy with hand, and the
sampling probe should to be inserted into the small hole of the tool. If it can’t be inserted into
the small hole, this item doesn’t meet requirement. After the check, click the “Assy Init.”
button.
WBC Bath Position Check: Put the inspection tool into the hole in the middle of the WBC
bath. Click the “Detect” button there and, when the sampling assy stops motion, click the
“V.motor Power-off ” button. Pull down the sampling probe assy with hand, and the sampling
probe should to be inserted into the small hole of the tool. If it can’t be inserted into the small
hole, this item doesn’t meet requirement. After the check, click the “Assy Init.” button.
RBC Bath Position Check: Put the inspection tool into the hole in the middle of the RBC
bath. Click the “Detect” button there and, when the sampling assy stops motion, click the
“V.motor Power-off” button. Pull down the sampling probe assy with hand, and the sampling
probe should to be inserted into the small hole of the tool. If it can’t be inserted into the small
hole, this item doesn’t meet requirement. After the check, click the “Assy Init.” button.
Upper Position: The parameter should be “0”. If it is not “0”, this item doesn’t meet
requirement.
Isolation Bubble Position Establishment Check: The parameter should be “100”. If it is
not “100”, this item doesn’t meet requirement.
WBC (RBC) Position Check: The parameter should be from 660-720. If it is outside the
157
Hemaray 86 / 89 Service Manual
158
Hemaray 86 / 89 Service Manual
Test a pipe of QC blood or fresh human blood on the standard analyzer thrice (the
differentiation charts of the various cells must be separated), calculate the mean of the center
of gravity in the FS direction of neutrophil, center of gravity in the FL direction of lymphocyte,
and center of gravity in the SS direction of neutrophil as the calibration reference value.
Compare the mean of the test results to the calibration reference value. The deviation of the
center of gravity in all directions should be ±10. If the above results do not meet the
requirements, this item doesn’t meet requirement.
11. Pressure Check
Enter the “Menu” → “Service” → “Adjustment” → “Pressure Calib” interface; connect the
pressure gauge to the pipe joint on the top of the Waste Syringe.
Positive Pressure Detect: Input “5” to the input box for positive pressure. Click “Reset”
and then “Detect”. The reading on the pressure gauge should be “5±1”Kpa. If it is outside the
range, this item doesn’t meet requirement. Negative Pressure Test: Input “15” to the input
box for negative pressure. Click “Reset” and then “Detect”. The reading on the pressure
gauge should be “-15±1”Kpa. If it is outside the range, this item doesn’t meet requirement.
12. Reagent Lack Alarm Function Check
Enter the “Menu” → “Service” → “Daily Maintenance” → “Replace Reagent” interface.
Remove the diluent pipe from the rear side of the instrument and click “Diluent”, and the
instrument will conduct diluent perfusion, during which, the status bar at the bottom of the
instrument will display the diluent lack. If there is no such prompt, this item doesn’t meet
requirement. After the check, connect the diluent pipe to the instrument and click “Diluent”
again, and the instrument will conduct diluent perfusion. Detect whether the cleanser, 86D
lyse, and 86H lyse are normal by referring to the above steps.
13. Waste Liquid Alarm Function Check
During the test, when the floating ball on the waste liquid sensor is pushed to the top, the
Information Processing System Software will prompt full waste liquid. If there is no such
prompt, this item doesn’t meet requirement.
14. Quantity Check of Optical Path Particle Test
Conduct the test on the test interface with the fresh human blood. When the test is
finished, click “Orig.Info.(O)”. The number of Non-ghost particles should be greater than 65%
of the number of WBC channels. If it is less than 60%, this item doesn’t meet requirement.
15. Prompt Sound and Alarm Sound Check
On the test interface, press the Liquid Suction Key for normal test. When the sample
suction is finished and the sampling probe is withdrawn, there should be prompt sound. If
there is no prompt sound or the prompt sound is abnormal, this item doesn’t meet
requirement.
.37.2 Performance Test
1. Blank Test: On the test interface, select the CBC+DIFF Whole Blood mode and conduct
the blank test. Three tests in row should meet these requirements: WBC ≤ 0.2, RBC ≤ 0.02,
PLT ≤ 10, and HGB ≤ 1. (Up to 10 tests can be conducted. If the requirements are not me et,
please treat the machine as one with faults).
2. Carry-over Test: On the test interface, select the CBC+DIFF Whole Blood mode, test the
fresh human blood or high-level QC thrice, and record the test results. Conduct 3 blank tests
and record the test results. When the tests are finished, calculate the carry-over with the
159
Hemaray 86 / 89 Service Manual
carry-over formula. Requirements: WBC ≤ 0.5%, RBC ≤ 0.5%, PLT ≤ 1%, and HGB ≤ 0.5%.
Note: Requirements of high-level sample: WBC>15, RBC>6, PLT>200, and HGB>300. For
low-level samples, conduct blank tests.
3. Repeatability Test: On the test interface, select the CBC+DIFF Whole Blood mode. Test
the normal fresh human blood more than 1mL 10 times in a row and record the test results.
When the tests are finished, calculate the CV value with the repeatability formula. Test 3
tubes of human blood in a row. The mean of CV values of each tube of blood sample should
meet these requirements: WBC ≤ 2%, RBC ≤ 1.5%, PLT ≤ 4%, HGB ≤ 1.5%, MCV ≤ 1%,
MPV ≤ 4%, NEUT%±4, LYMP%±3, MONO%±2, EO%±1.5, and BA%±0.8, (If the repeatability
does not meet the requirements, conduct the set of tests again. If the repeatability still does
not meet the requirements, please treat the machine as one with faults).
4. Instrument Calibration: On the Manual Calibration interface, set the calibration parameters
in the Whole Blood mode and Pre-diluted mode to “1”. Test the normal-level QC blood more
than 2mL on the standard analyzer thrice and calculate the mean of WBC, RBC, HGB, PLT,
MCV and MPV as the target value of Quality Control.
① Calibration in Whole Blood Mode: On the Automatic Calibration interface, select the Whole
Blood mode, and input the target value of normal-level QC blood in WBC, RBC, HGB, PLT,
MCV, and MPV. Test the normal-level QC on the machine 5 times in a row. The CV values of
the 5 results must meet these requirements: WBC ≤ 2%, RBC ≤ 1.5%, PLT ≤ 4%, HGB ≤
1.5%, and MCV ≤ 1%. Save the results. If the CV values of the 5 results do not meet the
requirements, conduct the calibration again.
② Calibration in Pre-diluted Mode: On the Automatic Calibration interface, select the Pre-
diluted mode, and input the target value of normal-level QC blood in WBC, RBC, HGB, PLT,
MCV, and MPV. Click “OK” and then “Spit Diluent”. Put the tube under the sampling probe,
click the Liquid Suction Key 4 times, and the instrument will spit diluent. Take 80uL normal-
level QC blood with the 200uL pipette, wipe the outer wall of the head of the pipette clean,
add the QC to the tube with diluent spit by the instrument, and blend the contents well.
Conduct the test on the machine 5 times. The CV values of the 5 results must meet these
requirements: WBC ≤ 2%, RBC ≤ 1.5%, PLT ≤ 4%, HGB ≤ 1.5%, and MCV ≤ 1%. Save the
results. If the CV values of the 5 results do not meet the requirements, conduct the
calibration again.
5. Instrument Comparability Test: Test the QC blood more than 1mL on the standard analyzer
5 times and calculate the mean of the items as the reference value.
① On the test interface, select the CBC+DIFF Whole Blood mode, test the sample thrice,
and calculate the mean of the items. The deviation percentage between the mean of the
items and the reference value should meet these requirements: WBC±3%, RBC±2%,
HGB±2%, PLT±5%, and HCT±2%.
② On the test interface, select the CBC+DIFF Pre-diluted mode. Click “OK” and then “Spit
Diluent”. Put the tube under the sampling probe, press the Liquid Suction Key 4 times, and
the instrument will spit diluent. Take 80uL of the sample with the 200uL pipette, wipe the
outer wall of the head of the pipette clean, add the sample to the tube with diluent spit by the
instrument, and blend the contents well. Conduct the test on the machine thrice and calculate
the mean of the items. The deviation percentage between the mean of the items and the
reference value should meet these requirements: WBC±3%, RBC±2%, HGB±2%, PLT±5%,
and HCT±2%.
160
Hemaray 86 / 89 Service Manual
161
Hemaray 86 / 89 Service Manual
Chapter 8 Maintenance
Cleanser”.
Daily
When the accumulated number of samples tested reach 100pcs, the
Cleanser”.
Weekly every one week (7 days), the system will execute the operation of
Wipe the surface of the sampling probe, particularly the probe tip, with
Sampling
Monthly soft tissue manually at least once a month to remove the stubborn
probe
stains on the surface of the sampling probe.
Sampling Re-tension the horizontal and vertical belt pulleys to ensure the
Clean the bloodstain on the surface of the Rinse part and remove the
Semi- Rinse part
wear debris of the belt pulleys on the periphery of the Rinse part.
annually
Waste
Apply (special) lubricant on the plunger of the Waste Syringe.
Syringe
All syringes Apply lubricant on the 6 syringe drive screws of the whole machine.
On Counting When the counting chamber is contaminated, the “Clean *** Counting
Flow Cell range or the CV value of the particle test is greater than 7%, the “DIFF
Ruby When the Ruby aperture is blocked, on the Daily Maintenance interface,
162
Hemaray 86 / 89 Service Manual
Out of than 3 days, execute the “Pack” operation in the “Service” menu to
service clean and drain the Fluidics of the analyzer, wipe the analyzer dry, and
163
Hemaray 86 / 89 Service Manual
Flag Description
::::: Hole blockage
The parameter is invalid or can’t be
-----
calculated.
+++++ The parameter is out of the display range.
.39.2 Parameter result with alarm flag
The parameter value alarm flag and the parameter result appear at the same time. The
specific definitions are show inTable 4.
Flag Description
+ The parameter is outside the measurement range.
* A suspicious flag of parameter indicates the credibility of the
parameter result is not high. The suspicious flags of the
various parameters are as follows:
Parameter Possible Causes
WBC Poor hemolysis
Nucleated red blood cells, large PLT / PLT
clumps interference
Electrical noise and bubble interference
Hole blockage
DIFF5% Poor hemolysis of DIFF channel
Low-level white blood cell
Nucleated red blood cells, large PLT / PLT
clumps interference
WBC excessive dissolved
Electrical noise and bubble interference
Dirty or blocked Flow Cell
HGB The instrument has not been calibrated.
High white blood cell interference, high
lipid blood sample
Abnormal hemoglobin voltage
RBC Large PLT / PLT clumps interference
Small red blood cell interfered by platelet
Abnormal red blood cell distribution
Electrical noise and bubble interference
164
Hemaray 86 / 89 Service Manual
Hole blockage
PLT Large PLT / PLT clumps
Small red blood cell
Abnormal platelet distribution
Electrical noise and bubble interference
Hole blockage
Note: DIFF5% refers to the 5-part differentiation parameters, i.e. NEUT%, LYMP%, MONO%, EO%,
and BASO%.
165
Hemaray 86 / 89 Service Manual
.40 Histogram
This section gives some WBC histograms frequently seen in actual applications.
1. Normal histogram
2. The histogram is too narrow (Abnormal sample? Small WBC gain? Aged sample?)
166
Hemaray 86 / 89 Service Manual
3. There is interference at the front end of histogram (Poor hemolysis, nucleated red
blood cells, large PLT / PLT clumps, electrical noise, etc.)
167
Hemaray 86/89 Service Manual
168
Hemaray 86 / 89 Service Manual
Sheath
fluid Left door
1012 1034 1056 Autoloader module action abnormal
pipeline open
leaks
Sample
Right door
1013 pipeline is 1035 1057 Mix module action abnormal#1057
open
blocked
Sample
Front
1014 pipeline 1036 1058 Mix module action abnormal#1058
shell open
leaks
Temperatur
F l o w c e el le x c e e d s
1015 1037 1059 Mix module action abnormal#1059
blocked working
range
DIFF
External
reaction
drain
1016 1038 bath 1060 Mix module action abnormal#1060
pipeline
temperatur
blocked
e abnormal
Liquid Probe
drain horizental
1017 1039 1061 Mix module action abnormal#1061
pipeline action
leaks abnormal
Probe
B a c k b a t h
vertical
1018 pipeline 1040 1062 Unloading tube rack error
action
leaks
abnormal
Positive
pressure
1019 1041 1063 Unloading tray is full
establishmen
t failed
Negative
pressure
1020 1042 1064 Scan action abnormal
establishmen
t failed
1021 No cleanser 1043 1065
1022 No diluent 1044 1066
169
Hemaray 86/89 Service Manual
Pressure of 1. Blockage of Flow 1. Check whether the valve ports of the valves
1010
sheath fluid Cell SV5, SV6, SV7 and SV13 and the related pipes
170
Hemaray 86 / 89 Service Manual
171
Hemaray 86 / 89 Service Manual
172
Hemaray 86 / 89 Service Manual
173
Hemaray 86 / 89 Service Manual
174
Hemaray 86 / 89 Service Manual
3. Fault of driver
board
1. Click the “Clear fault” button to reconfigure
the temperature of the DIFF bath. If the
1. The temperature temperature of the DIFF bath is too high,
setting of the DIFF execute the “Clean DIFF Bath” operation
bath is abnormal. several times. If the temperature is still
DIFF
2. Fault of DIFF abnormal, turn off the power of the instrument
reaction
bath temperature and restart it later.
1038 bath
sensor or 3. If the fault still exists, unplug the two-way plug
temperature
temperature connector of the heating wire at the bottom of
abnormal
protection switch the DIFF bath and measure whether the
3. Fault of driver heating element is normal with a multimeter (if
board the resistance is infinite, it indicates that the
heating element of the DIFF bath is damaged).
4. Check the related section of the driver board.
1 Fault of
1. Click the “Clear fault” button to execute assy
optocoupler test
initialization.
2. Fault of horizontal
2. If the fault still exists, check whether the
driving of sampling
Probe horizontal optocoupler test section of the
assy (fault of motor,
horizental sampling assy is normal.
1039 big resistance of
action 3. If the fault still exists, check whether the
horizontal driving
abnormal horizontal motor drive and mechanical drive
block, wear of belt
sections of the sampling assy are normal.
pulley)
4. Check whether the related section of the
3. Fault of driver
driver board is normal.
board
1. Click the “Clear fault” button to execute assy
1. Fault of initialization.
optocoupler test 2. If the fault still exists, check whether the
Probe 2. Fault of vertical vertical optocoupler test section of the sampling
vertical driving of sampling assy is normal.
1040
action assy (fault of motor, 3. If the fault still exists, check whether the
abnormal wear of belt pulley) horizontal motor drive and mechanical drive
3. Fault of driver sections of the sampling assy are normal.
board 4. Check whether the related section of the
driver board is normal.
175
Hemaray 86 / 89 Service Manual
176
Hemaray 86 / 89 Service Manual
177
Hemaray 86 / 89 Service Manual
178
Hemaray 86 / 89 Service Manual
4 RBC bath 1. The waste liquid pipe 1. Firstly, check whether the waste liquid pipe of the
179
Hemaray 86 / 89 Service Manual
180
Hemaray 86 / 89 Service Manual
181
Hemaray 86 / 89 Service Manual
182
Hemaray 86 / 89 Service Manual
183
Hemaray 86 / 89 Service Manual
184
Hemaray 86 / 89 Service Manual
185