Professional Documents
Culture Documents
Cxdi 701g - 701c Service Manual
Cxdi 701g - 701c Service Manual
Cxdi 701g - 701c Service Manual
Confidential
CXDI-701G/701C
Wireless
Service Manual
This service manual consists of nine chapters; “General”, “Installation Manual”, “Functions”,
“Repair Guide”, “Parts Catalog”, “Troubleshooting”, “Service Manual Report”, “Tools” and
“Appendix”. It describes an overview of the product, its functions, product configuration,
installation procedures, dimensions, specifications, and notes.
If the product undergoes a large modification, a revised edition of the service manual will be
sent to you. In other cases, a service manual report will be sent to you to update the manual.
Note 1:
This service manual is the property of Canon Inc. and the company may seek to have it
returned, depending on the circumstances. You are expected to keep it until then.
Note 2:
Your inquiries, suggestions, etc. about the contents of this service manual should be
addressed to:
In order to ensure safety, the best performance, setup, repair, and maintenance work can only
be performed by technicians who have received service training specified by Canon Inc. If
there are order required certificates or restrictions specified by the law or ordinances, those
regulations of the country must be observed.
When removing the cover during maintenance, repair, etc., perform the work after switching
the power off. Never touch the device with wet hands, as there is a risk of electric shock.
The provided ground wire must be connected to the ground terminal indoors. Make sure that
the device is properly grounded.
4. Alternation prohibition
5. Waste control
The service provider is responsible for the disposal of used service parts, packing material,
etc. resulting from the setup, repair, or maintenance of the medical device. However, the
customer is responsible for the disposal of the medical device. Disposal activities must
follow the regulations (especially controlled industrial waste) of the country where the
device is used.
VORSICHT
Befolgen Sie die unten angegebenen Sicherheitsanweisungen.
Mißachtung kann zu erletzungenoder Unfällen führen.
3.Sicherung
Wenn die Sicherung ausgewechselt werden muß, schalten Sie unbedingt die Stromversorgung des
Instruments aus, und beheben Sie die Ursache für das Durchbrennen der Sicherung.
Ersetzen Sie die Sicherung nur durch den vorgeschriebenen Typ.
Anderenfalls kann es zu einem Brand oder elektrischen Schlag kommen.
4.Erdleiter
Erden Sie das Instrument unbedingt an einer Schukosteckdose.
Anderenfalls kann es zu einem Brand oder elektrischen Schlag durch Leckstrom kommen.
5.Umbau
Jeder Umbau des Produktes ist strengstens untersagt, da dies zu einem Brand oder elektrischen Schlag
führen kann.
Caution Regarding the Setup
According to the “IEC 60601-1-1”, equipment installed in the patient environment is restricted
to “medical electric equipment conforming to IEC 60601-1”.
Changes related to installation that occur over the actual lifespan of the equipment must be
evaluated with respect to the requirements in IEC60601-1.
Note that the entire radiography system, including this product, is required to comply with IEC
60601-1-1. The details are as follows: Medical equipment such as an X-ray generator
complying with IEC 60601-1, an image capture computer and a display monitor complying with
IEC 60601-1 or IEC 60950-1.
The image capture computer, optional devices such as network equipment or magnetic card
readers, and display monitors that are classified as other than medical electrical equipment (IEC
60601), for example as information technology equipment (IEC 60950), cannot be installed in
the patient environment.
The appended documentation is to be provided to the user, operator, and installer of the
equipment configuring the system, such as the image capture computer and display monitor,
upon installation of such equipment.
Also, for the installation, the above items must be connected to a common protective earth.
Equipment connected to the image capture computer must satisfy state No. 3 in table BBB.201
of IEC 60601-1-1.
Either connect the image capture computer to a common protective ground, add a dedicated
protective ground, or install an isolation device.
The patient environment described below is an example cited from the “IEC 60601-1-1” – the
measurements are only guidelines. However, the “IEC 60601-1-1” example must be treated as
the standard.
Therefore, the image capture computer, optional devices such as network equipment or
magnetic card readers, and display monitors must be installed in a location separated by at least
the dimensions indicated below (outside the patient environment). In addition, the X-ray
interface box should be installed outside the patient environment.
* The areas through which patients move at times other than X-ray exposure (when
entering and leaving, etc.) are also included in the patient environment. Therefore,
always confirm (discuss) with the customer that the installation location is outside the
patient environment.
Example of patient environment
(1) Due to the risk of electric shock, after turning on the power to each piece of equipment, do
not simultaneously touch the patient and external parts of non-medical equipment other than this
product.
(2) Perform all of the adjustment, cleaning, and sterilization procedures specified in the
operating manual that came with the peripheral parts comprising the system.
(3) Manage the transport and storage / use of the equipment comprising the system, according to
the following same environmental conditions as those of the main unit.
CXDI-701G/701C
Wireless
1. Overview
CONTENTS
1. Overview ............................................................. 1
4. Specifications .......................................................... 8
4.1 Detector specifications ............................................... 8
4.2 Power supply specifications .......................................... 11
4.3 Recommended LAN cable specifications ............................... 11
4.4 Environmental resistance specifications ................................ 11
4.5 Required PC specifications ........................................... 12
4.6 Wireless communication specifications ................................. 12
4.7 Link communication specifications .................................. 19
4.8 Detector LED display specifications ................................. 20
4.9 Ready indicator display and notification tone specifications ............... 20
4.10 Battery and battery charger specifications .............................. 22
4.11 Grid specifications ................................................. 25
4.12 Wiring unit (option) ............................................... 27
4.13 [Supplementary notes] Long-term storage of batteries .................... 28
1. Overview
1. Overview
CXDI-701G/701C Wireless
-1-
1. Overview
-2-
1. Overview
2. Description of system
1000Base-T/1000Base-TX
X-ray interface unit switching HUB
X-ray interface cable AC adapter
X-ray generator
Wireless
access point
Detector
Charger
X-ray generator
Wireless access
point
Ready indicator
USB wireless
adapter
Detector
Battery pack
Grid frame
Control PC
Charger
-3-
1. Overview
1000Base-T/1000Base-TX
X-ray interface unit switching HUB
X-ray interface cable AC adapter
X-ray generator
Det
ecto
r ca
Wiring unit
ble
Detector
Battery pack
Control PC
Grid frame
Status
indicator
Charger
HUB
LAN
Access Point
LAN (20m)
LAN
(2.3m)
(1.8m)
IR Data
Communication Unit
USB
(< 5m)
Operation Unit
-4-
1. Overview
A graphical representation of the non generator connection mode and the basic flow are shown
below.
[Graphical representation] [Basic flow]
Detector now ready
X-ray generator (Exposure timing notified by
ready indicator)
Images stored
Detector
PC
Images transferred from detector
to PC
The hardware requirements do not impose any restrictions, but bear in mind that the
communication performance may deteriorate due to interference arising when wireless channels
are shared.
No such restrictions apply to wired connections.
The following restrictions apply to the connection of external devices to a notebook PC which is
used when the detector is installed in a mobile X-ray system.
Since patients may touch the notebook PC and the detector which is connected to the notebook
PC using the PC connecting cable, the AC adapter of the notebook PC, main network inside the
hospital or clinic or any devices requiring an external power source must not be connected.
For further details, refer to sections 7.2.4 and 7.2.5 on the non generator connection mode
(mobile X-ray systems: wireless/wired) in Chapter 2 Installation Procedures.
-5-
1. Overview
For details of the functions of the various units, refer to the description of functions in
Chapter 3.
-6-
1. Overview
3. CXDI image processing
3.1 Processing flow
Raw images
Pre-processing
• Offset compensation
• Gain compensation
• Defect correction
Original images
Processing from this point Dtstore
onward is handled by the
image capture computer.
Automatic analysis
• Exposure field
• ROI analysis
• Dynamic range analysis
• EI (exposure index)
DICOM output
-7-
1. Overview
4. Specifications
4.1 Detector specifications
CXDI-701G/701CWireless
Item Remarks
(2.4GHz/5GHz)
The wiring unit is required for
wired connection.
Wireless or wired general image
Use form Use of equipment for 14" by 17"
capturing cassette compatible
size film cassettes complying with
the ISO4090 standard enabled.
35.0cm x 42.6cm
Effective imaging range
(14 x 17 in.)
Number of effective pixels 2800x3408
-8-
1. Overview
When the allowed exposure time is
1-second mode and the setting is
Within 15 sec. made for the imaging cycle
(conventional cycle) at installation
or the exposure time is 1-second.
If exposure time is 3 seconds with
Within 30 sec. allowed exposure time: 3 second
mode
1 second mode (Storage time is
Within 15 sec.
Non generator fixed to 1 sec.)
connection mode 3 second mode (Storage time is
Within 35 sec.
fixed to 3 sec.)
During standard
The time from when image capture
synchronization
60 minutes (X-ray exposure) is finished until
Ready mode and external
the detector enters sleep mode,
timeout time power supply
when no operations are being
Other modes 10 minutes performed.
Approximately
Max mode 1200 images 9-second cycle
Standard
(approx. 3 hours)
synchronization
Approximately
mode
Ave mode 140 images 100-second cycle
(approx. 4 hours)
Approximately
Max mode 700 images 15-second cycle
Non generator (approx. 3 hours)
Battery connection mode Approximately
performance Ave mode 110 images 100-second cycle
(approx. 3 hours)
Save mode Approximately 6.5 hours In sleep mode
When installed in the battery
Approximately 3 hours charger
In a 25°C environment
Full charge time
When installed in the detector and
Approximately 6 hours charged via the detector cable
In a 25°C environment
384(W) x 460(D) x 15(H) mm
External dimensions
(15 x 18 x 0.6 in.)
Top cover: Light sky blue
Sides: Aluminum sliver
Colors of detector exterior
Function switches: Medical blue
Rear cover: Titanium white
Detector weight 3.3 kg (incl. battery)
*: Conditions for the preview display and cycle time:
The display and cycle time indicates the performance when using an image capture computer
with the recommended specifications.
*: Battery performance: This is the performance for a new battery which has been 100% charged.
*: Long exposure is possible on non generator connection mode with the following software
combination.
▪ CCS Ver.2.11 and later
▪ Firmware: 0x030b0900 and later, FPGA: 0x02000001 and later
▪ New PCA-REF board (BM7-1212-080) is required.
-9-
1. Overview
CXDI-701G/701C Wireless
Item Remarks
(2.4GHz/5GHz)
7 years
Ambient temperature: +23°C ±2°C
Operation time: 21,000 hours
Relative humidity: 60% ±15% RH
(10 hours a day, 300 days a year,
Atmospheric pressure: 860 to 1060
Service life 7 years)
hPa
Standby time: 61,500 hours
When used under the ambient
(24 hours a day, 365 days a year,
conditions above
7 years)
Momentary Momentary max. value, excluding
Approx. 75 VA or less
max. load inrush currents
Power Assumed
Mean value during continuous
consumption continuous Approx. 20 W or less
operation with exposure at
/heat max. load (72 kJ/h ,17 kcal/h or less)*2
15-second intervals
generation*1 mode
Standby Approx. 10 W or less Sleep mode
Test conditions:
Load applied over entire surface: The detector shall be placed on a
1.47 kN (150 kgf) for 1 minute rigid flat surface with its X-ray
Detector withstand load
Load applied in local areas: 0.98 incidence surface face up, and
operation
kN (100 kgf), 40 mm diameter for images shall be taken with the loads
1 minute on the left brought to bear on the
effective area.
CXDI Control Software NE
Control software
Ver.2.10 and up supported
Photo timer Cannot be incorporated
When the detector panel is detected to
be at 50°C (temperature detector
output +49.4°C), the detector is
prevented from capturing images and
Allowable temperature of is forcefully set to the sleep mode.
Max. +50°C
detector panel When the temperature drops to the
allowable environmental temperature
of 35°C or below (temperature
detector output +48.7°C), image
capturing is allowed again.
*1
: The target unit is a complete CXDI-701G/701C Wireless (2.4 GHz/5 GHz) set (including
detector, X-ray interface box, wiring unit, and status indicator). The values indicated
incorporate a large margin from the actual values or indicate attainable values.
1) Power consumption [VA]: Information for the purpose of designing the power supply
system for the installation site
2) Power consumption [W]: Information for the purpose of designing the power supply
system for the installation site (Note that this is valid only
when the power factor is high [ex. 90% or more].)
3) Heat generation [cal or J]: Information for the purpose of designing the cooling system
for the installation site
*2
: The heat generation is calculated from the power consumption (W) as follows:
Heat generation (kcal/h) = power consumption (W) * 0.23889 * 60 * 60 / 1000
Heat generation (kJ/h) = power consumption (W) * 60 * 60 / 1000
- 10 -
1. Overview
- 11 -
1. Overview
Use of the CXDI-701G/701C Wireless requires an Image Capture Computer which runs
Windows 7 and in which the CXDI Control Software NE (Ver.2.10 or later) or the CXDI
Controller RF (Ver.2.10 or later) made by Canon has been installed, and the sales companies or
agents concerned must provide the hardware that satisfies the required specifications. The
specifications required of the Image Capture Computer can also be found in the service manual
of which the CXDI Control Software NE (Ver.2.10 or later) or the CXDI Controller RF
(Ver.2.10 or later).
The relay unit used for wireless communication between the CXDI-701G/701C Wireless
(2.4GHz/5GHz) and Image Capture Computer is called a wireless LAN access point. The table
below lists the wireless communication specifications.
- 12 -
1. Overview
2400 2410 2420 2430 2440 2450 2460 2470 2480 2490
(MHz)
北米(FCC),台湾
North America 1 1 c11
(FCC), Taiwan h ch
*1: With HT20 (20 MHz band), a frequency of 22 MHz is occupied per channel so the maximum number of channels which can be provided at the same time is limited to 3.
- 13 -
1. Overview
*2: The combinations of the channels which can be used with HT40 (40 MHz band) using channel bonding are listed in the table below.
Primary Secondary channel Secondary Channels which can be used at the same time without
channel offset channel interference
1CH above 5CH HT20: 1 channel among channels 10, 11, 12 and 13
5CH below 1CH NT40: No channels
2CH above 6CH HT20: 1 channel among channels 11, 12, and 13
6CH below 2CH NT40: No channels
3CH above 7CH HT20: 1 channel between channels 12 or 13
7CH below 3CH NT40: No channels
4CH above 8CH HT20: 13 channel
8CH below 4CH NT40: No channels
5CH above 9CH HT20: No channels
9CH below 5CH NT40: No channels
6CH above 10CH HT20: 1 channel
10CH below 6CH NT40: No channels
7CH above 11CH HT20: 1 channel between channels 1 or 2
11CH below 7CH NT40: No channels
As shown above, when HT40 is used in the 2.4 GHz band, the number of channels which can be used without interference is limited to 1 with HT20.
- 14 -
1. Overview
不要 不要 屋内のみ 北米
屋内のみ 不要 不要 屋内のみ 日本
屋内のみ 不要 不要 屋内のみ 欧州
屋内のみ 不要 不要 屋内のみ 豪州
台湾
屋内のみ 不要 不要 屋内のみ 韓国
中国
W52 5170 5189 5190 5209 5210 5229 5230 5249
使用可否
不可
不可
36 40 44 48
5180 5200 5220 5240 (MHz)
DFS*1
-
TPS*2
-
W53 5250 5269 5270 5289 5290 5309 5310 5329
使用可否
屋内のみ
不可
52 56 60 64
可
不明 不明
*1 不明 不要
DFS
要
要
要
-
*2
TPS
要
要
要
-
W56 5490 5509 5510 5529 5530 5549 5550 5569 5570 5589 5590 5609 5610 5629 5630 5649 5650 5669 5670 5689 5690 5709
使用可否
不明
不明
不明
不可
100 104 108 112 116 120 124 128 132 136 140
可
可
可
5500 5520 5540 5560 5580 5600 5620 5640 5660 5680 5700 (MHz)
不明 不明
不明 不明
不明 不明
*1
DFS
要
要
要
*2
TPS
要
要
要
W58 5735 5754 5755 5774 5775 5794 5795 5814 5815 5834
使用可否
不可
不可
*4
可
可
可
可
不要 不要
不要 不要
不要 不要
不要 不要
*1
DFS
-
-
TPS*2
-
-
*1: DFS (Dynamic Frequency Selection): If military or meteorological radar signals are detected, this function ensures that AP changes the channel concerned so that the channel is turned
over to the radar signals. It is required by W53 and W56. Several dozen seconds are required to switch the channel.
*2: TPC (Transmission Power Control): This function automatically adjusts the radio signal output required for transmission between AP and a wireless terminal.
*3: As an exception to the general rule, the DFS function is not required for Taiwan W53.
*4: The DFS function does not support the product specifications so W53 and W56 that require it are not supported.
*5: Up to 161 channels can be used with South Korea W58.
*6: Refer to the following for the combinations of the channels which can be used with HT40 (40 MHz frequency) through channel bonding.
[Reference] The checkpoints regarding radio signals and the usable regions and restrictions are also contained in the operation manual.
- 15 -
1. Overview
Primary channel Secondary channel Secondary channel Channels which can be used at the same time without
offset interference
36CH above 40CH HT20:44,48CH
40CH below 36CH HT40:44-48CH
44CH above 48CH HT20:36,40CH
48CH below 44CH HT40:36-40CH
No channels entered for W53 since it is not supported.
No channels entered for W56 since it is not supported.
149CH above 153CH HT20:157,161,165CH
153CH below 149CH HT40:157-161CH
157CH above 161CH HT20:149,153,165CH
161CH below 157CH HT40:149-153CH
As shown above, when HT40 is used at 5 GHz, the number of channels which can be used without interference in the same frequency band is 2 with HT20 and 1 with HT40 for W52, and
3 with HT20 and 1 with HT40 for W58.
In the 2.4 GHz band, W52 and W58 operations are possible without mutual interference. For instance, the simultaneous operation of three detectors without interference is possible using
2.4 GHz (channel 1 above channel 5) for detector No.1, W52 (channel 36 above channel 40) for detector No.2 and W58 (channel 149 above channel 153) for detector No.3.
- 16 -
1. Overview
Regions where wireless detectors can be used and restrictions on their use
The CXDI-Wireless series of detectors have built-in wireless facilities which have been certified
in accordance with the radio signal laws of the countries concerned, and they must adhere to the
laws governing radio signals in the countries or regions concerned. Outdoor use of the
frequency bands (2.4 GHz/5 GHz) and channels (channels 1 to 13, 36 to 48 and 149 to 165)
established for indoor use may not be allowed under the radio signal laws of the countries
concerned so check the country code table below, and set the frequency band and channels
accordingly.
- 17 -
1. Overview
- 18 -
1. Overview
- 19 -
1. Overview
- 20 -
1. Overview
(1) LED lighting and notification tones when the system has started up and the link has been
established successfully
Item Description
System startup The LEDs at the top and bottom light in the “dim →
medium → bright” sequence of brightness levels in
(When the application in the PC
synchronization with the advisory tone.
acquiring the images is running) The “do-me-so” melody is generated as the advisory tone.
Link established successfully The LEDs light in the “top → bottom → top” sequence in
synchronization with the advisory tone.
The “do-me-ray” melody is generated as the advisory tone.
(2) LED lighting and notification tones which indicate the state displayed as the detector status
Item Description
Imaging mode Detector status Display details
All modes No imaging possible Top LED: Off
Bottom LED: Off
Non generator Imaging Imaging possible time: Top LED: Lights dimly
connection enabled More than 5 minutes Bottom LED: Lights dimly
mode status (ready remaining A short beep (beep!) is generated as
status) the advisory tone when the status
has been switched to this status
from another status.
Imaging possible time: Top LED: Lights dimly
Less than 5 minutes Bottom LED: Lights dimly
remaining Two short beeps (beep, beep!) are
generated as the advisory tone when
the status has been switched to this
status from another status.
Imaging possible time: Top LED: Lights dimly
Less than 1 minutes Bottom LED: Off
remaining Three short beeps (beep, beep,
beep!) are generated as the advisory
tone when the status has been
switched to this status from another
status.
Imaging possible time: Top LED: Blinks dimly at 1-second
Less than 10 seconds intervals
remaining Bottom LED: Off
The series of three short beeps is
generated repeatedly (beep, beep,
beep! … beep, beep, beep! …) as
the advisory tone in synchronization
with the blinking of the LED.
Standard During imaging in the standard Top LED: Lights brightly
synchronization/ synchronization mode or non generator Bottom LED: Lights brightly
non generator connection mode A chime (ding-dong!) is generated
connection as the advisory tone when imaging
mode has been undertaken.
- 21 -
1. Overview
For details on the specifications and handling of the battery and battery charger, refer to the
respective operating instructions.
The table below lists the battery and battery charger specifications.
- 22 -
1. Overview
There is no display in the battery itself which enables its remaining charge level to be checked.
While the battery is installed in the detector, the remaining charge level information is output
from the management circuit (BMU: battery management unit) incorporated inside the battery,
and a 3-segment scale is displayed using the software in the Image Capture Computer.
The table below shows the remaining battery charge levels.
Wired connection
Remaining Wireless connection No external power supply Status, prohibited
charge Battery drive External power operations, etc.
available
supply available
Battery drive
Full
Full charge
charge
100% to
60%
59% to
9%
8% to 5% Imaging possible
Blinks.
for 6 remaining
Blinks.
images in
100-second cycle
operation
4% to 0% Image operations
Blinks.
Blinks.
prohibited
Operation for 15
minutes in sleep
mode
Immediately after the power of the detector is turned on, this level is read from the battery and
sent to the software in the image capture computer. After this, the level is read inside the
detector from the battery at fixed intervals, but no data is sent until a level below that indicated
by the red underbars in the above table is read.
The figure below shows the correlation between the discharge curves, remaining charge level
prediction curves and posting of the remaining charge level displays.
- 23 -
1. Overview
LV2 8%
Remaining charge level of残容量8%を1回のみ通知
8% posted only once
LV1 4%
Remaining charge level of残容量4%を1回のみ通知
4% posted only once
3V/Cell 0%
2.5V/Cell
Time
* BMU corrects the remaining charge level prediction line from the actual remaining charge
level when the remaining charge level is predicted to be 8%. If the actual remaining charge level
(voltage level) when the remaining charge level is predicted to be 8% is higher than the
prediction, the gradient of the remaining charge level prediction curve is increased. Conversely,
if it is lower, BMU works to correct the gradient so that the prediction curve gradient is reduced.
Therefore, if charging and discharging are repeated at shallow remaining charge levels which
fail to reach the remaining charge prediction of 8%, the compensation described above is not
performed and, along with the deterioration in the battery capacity caused by the repeated
charging and discharging, the accuracy with which the remaining charge level is predicted may
suffer as well. What is ideal is for the battery to be used to below the 8% remaining charge level
(at which imaging is no longer possible) each time the battery is used but, in practical terms, it
is recommended that the battery be used to below the 8% remaining charge level once every
several times the battery is charged and discharged.
- 24 -
1. Overview
Grid needs to be procured by sales companies and agents. The table below lists the
specifications of the grids that can be used.
- 25 -
1. Overview
The CXDI-701G/701C Wireless
does not feature any hardware means
for grid detection. However, the
Grid detection presence/absence and direction of the
grid are identified using software
means by grid stripe suppression
processing.
For a grid used in combination with
the detector, a deviation up to 1° The attachment angle
Grid
from the attachment angle is allowed is the relative angle of
attachment
from the perspective of the the grid stripes to the
angle
performance of the grid stripe detector pixel rows.
suppression processing.
The grid must be used by attaching it
to the grid frame unit.
When the grid is to be adhered for
use, it must be adhered using either 1.1mm < total
Grid adhesion of the adhesive tapes specified below thickness of grid and
(Essential) in such a way that the gap between double sided adhesive
the grid frame unit and grid edges is tape < 3.5mm
a uniform all the way around.
Spacers for adjusting the thickness
Precautions
are not needed.
for attaching
(1) F9473PC (Sumitomo 3M): Use the (1) tape in 2
the grid
Thickness: 0.255 [mm] layers
Adhesive tape (2) Y4930 (Sumitomo 3M):
(recommended) Thickness: 0.64 [mm]
(3) Y4920 (Sumitomo 3M):
Thickness: 0.4 [mm]
Dimension of Width of tape: 8 to 10mm
double sided
Length of tape L: 443±1.5mm
adhesive tape
(Reference) Length of tape M: 367±1.5mm
*Reference example
Magnification ratio 1%
If you use the distance
between X-ray tube
When attaching the grid to an
Collateral and grid is 1000mm,
external stand or table, the allowable
condition when the distance between
magnification ratio for the
attaching the surface of the sensor
recommended grid conditions is: 1%
grid to a stand and the grid should be
or less.
or table 10 mm or less
Also refer to “Grid direction.”
(includes the distance
from the surface of the
sensor and the outer
cover (3.2 mm)).
- 26 -
1. Overview
Output Minimum 0A
conditions Output
Rated 2.5 A
current
Maximum 4A
- 27 -
1. Overview
The characteristics of lithium-ion batteries are such that when the batteries are left standing in
storage for a prolonged period of time, their charging capacity decreases so they may end up
becoming unusable. The higher the temperature in the storage location, the faster the charging
capacity deteriorates. When the batteries become overdischarged, this deterioration progresses
at an even more significant pace.
- 28 -
Printed by Canon Inc.
December 2013 Rev.03
CXDI-701G/701C
Wireless
2. Installation procedures
1. Work-related precautions
Bear the following points in mind when installing the system.
(1) Carry out the work while ensuring that none of the portable units will be dropped or subjected to
impact.
(2) Ensure that the installation location satisfies the following criteria:
2) A location where the air does not contain moisture, dust, salt, sulfur or other such substances and
where the equipment will not be adversely affected by poor ventilation, the air pressure level, the
temperature and other such factors
3) A location which is not tilted or subject to vibration or impact (this includes the locations when the
equipment is conveyed)
4) A location where chemical products are not stored and where no gases are given off
5) A location where the power source providing the correct power line frequency, voltage and power
used by the equipment can be supplied
6) A location where a grounding wire and the 3-pin grounding plug of the power cable can be connected
(3) After installing the equipment, ensure that the product packaging materials and other waste materials
generated at the time of installation are disposed of in an environmentally sound manner.
2. Installation restrictions
(1) Take steps to ensure that the system is not installed near electronic devices or equipment since noise
artifacts occur more readily on the images in environments where strong electromagnetic fields are
present.
Examples of electronic devices and equipment concerned: CRT monitors, X-ray generators, medical
electronic equipment
(2) The system must not be installed near devices or equipment which may interfere with wireless
communication.
Examples of electronic devices and equipment concerned: Thermal therapy devices, laser knives,
cordless telephones
(3) When using a wired connection for the product, install the detector and AC/DC box with a distance
of at least 150 mm between them.
(4) For use, the cables must be routed without bending, pinching or stretching them or subjecting them
to unreasonable loads. Remodeling of the cables is prohibited.
(5) In the case of a network connection, the appropriate network address must be acquired from the
network administrator in the installation location, and it must be set correctly.
(6) When installing the system in a regular radiography room, the wireless LAN access point, switching
hub, battery charger, Image Capture Computer, ready indicator unit and X-ray interface box must be
installed in locations outside the patient environment. For further details, refer to the system connection
diagrams in sections 7.1 and 7.2. (Battery-powered notepad computer can be installed in the locations
inside the patient environment).
-1-
2. Installation procedures
(7) There are no usage restrictions on how the detector is to be positioned. Avoid using the detector
where it will be subjected to extreme bending loads or where it will easily be dropped.
(8) There are some limitations on the operation of the non generator connection mode in the mobile X-
ray system environment. Section 7.8 and 7.9 System connection in the non generator connection mode
(mobile X-ray system environment).
(9) Note that the radio frequency channel configured for indoor use may not be usable in outdoor areas
in the 5 GHz band. For further details, refer to the wireless communication specifications in section 4.6
in Chapter 1 Overview.
3. Installation-related precautions
(1) Bear in mind that there will be a risk of malfunctioning and other trouble if this product is used
together with any equipment which may generates power supply noise, supply voltage fluctuations,
electromagnetic waves and other disturbances.
(2) Before proceeding with the installation work, ensure that the static charged in the bodies of the
installation personnel has been released.
(3) Before connecting or disconnecting the cables, be absolutely sure to turn off the power of the units in
order to ensure safety. In addition, do not touch the power outlets or connectors with wet hands.
Make sure to perform the wireless LAN channel setting according to the channel setting guidelines.
In addition, when trouble has occurred in wireless communication, proceed to ascertain the causes of the
trouble and take the countermeasures on the basis of [When wireless communication trouble has
occurred] in Chapter 6 Troubleshooting.
If the signal strength level in GUI is still 0 or if it takes a significant amount of time to transfer the
images even after taking the countermeasures in 1) and 2) above, explain the signal interference
situation to the user, and ask the user to consider operation based on a wired connection that uses a
wiring unit.
-2-
2. Installation procedures
Guidelines
Wireless LAN channel setting guidelines for adding the DR Wireless system to an existing wireless
LAN environment are described below. These guidelines apply to both the 2.4 GHz and 5 GHz bands.
Basic guidelines
Check out the channels used in the existing wireless LAN system, and set the channels without
interference.
If interference cannot be avoided, select the same channels as the ones used by the existing
wireless LAN system.
Fig. 1 shows the above flow to be followed for establishing the settings.
HT40
(3) Use the same setting for the other wireless LAN system.
primary channel only.*2
(1) Set channels that do not interfere (1) Set channels that do not interfere with
with each other. each other.
(2) Use the same channel setting as (2) Use the same channel setting.
HT20
Reference information:
If the signal strength (dBm) cannot be measured using a software application provided with the wireless
LAN adapter or notebook PC, it is possible to view this strength in a simplified manner using the free
“inSSIDer” tool.
-4-
2. Installation procedures
Fig.1
-5-
2. Installation procedures
4. Product configuration
(1) Product configuration lists
1) CXDI-701G/701C detector
No. Product Quantity Remarks
2 Battery 2
-6-
2. Installation procedures
6) Ready status indicator unit: RI-3A (Option; for CXDI-701G/701C)
No. Product Quantity Remarks
1 Ready Statues indicator unit 1
2 Fastener loop 3
3 Fastener hook 3
4 Documentation - Operation manual, etc.
AC adapter: AIR-PWR-B=
PoE power injector: AIR-PWRINJ4
Legislation dictates that the usage band differs from one country to another so check the format in the
country where this unit is distributed before use. The purpose of this unit is to relay wireless
communications.
-7-
2. Installation procedures
• LAN cable (category 5e or above)
This cable must support Gigabit Ethernet.
A crossover cable or straight cable may be used when the switching hub has an Auto-MDI/MDI-X
function. However, only a straight cable may be used when the switching hub does not have the function.
The maximum cable length is 30 meters.
ケーブル径
Cable
~φ3.5mm
diameter
Up to 3.5 mm
-8-
2. Installation procedures
(2) Configuration
No. 1 No. 2
Product CXDI-701G/701C detector Product Battery pack
Quantity 1 Quantity 2
Remarks Remarks * Common to 70/80
No. 3 No. 4
Product Battery charger Product X-ray interface box
Quantity 1 Quantity 1
* Common to 70/80
However, the Ed.2 (AC adapter
Remarks * Common to 70/80 Remarks model number MPU-15A specified
product)compatible version cannot
be used.
No. 5 No. 6
Product X-ray interface cable Product Power cord set
Quantity 1 Quantity 1
For X-ray interface box
For connection with the X- (Including AC adapter)
Remarks ray generator Remarks However, the Ed.2 compatible
* Common to 70/80 version (AC adapter model number
MPU-15A) cannot be used.
-9-
2. Installation procedures
No. 7 No. 8
Product Grounding wire Product Status indicator
Quantity 1 Quantity 1
For X-ray interface box
Remarks Remarks * Common to 70/80
* Common to 70/80
No. 9 No. 10
Product AC/DC box Product Detector cable
Quantity 1 Quantity 1
For connecting the detector
Remarks Remarks
and ACDC box
No. 11 No. 12
Product Ready indicator unit Product Grid mounting unit
Quantity 1 Quantity 1
Remarks Remarks
- 10 -
2. Installation procedures
No. 13 No.
PC connecting cable
(Scheduled to be launched
Product Product
on the market in or after
July 2013)
Quantity 1 Quantity
Remarks For PC connection Remarks
- 11 -
2. Installation procedures
Plastic bag
Operating
Anti-static bag instructions
Battery packs x 2
Cushions for
protecting long sides
of FPD
FPD
unit
Plastic bag
Cardboard
box for FPD
- 12 -
2. Installation procedures
2) Battery charger package
Plastic bag
Battery charger
Power cord set
Cardboard
box
- 13 -
2. Installation procedures
3) X-ray interface box (option) package
Grounding wire
AC adapter,
Operation manual power cord set
Plastic bag
Cardboard box
- 14 -
2. Installation procedures
4) Battery (option) package
Operation manual
Anti-static
bag
Storage box
Air cap sheet
Cardboard
box
- 15 -
2. Installation procedures
5) Wiring unit (option) package
Z10_WU-3A
Plastic bag inner box
Operating
instructions
18 mm sealing
tape (yellow) Side pads
2 pads used
Plastic bag
Power cable set Status indicator
Wiring unit
Plastic bag
Plastic bag
Detector
cable
Anti-static bubble
wrap bag
- 16 -
2. Installation procedures
6) Ready indicator unit (option) package
Appropriate number of
thin sheets of paper*
Ready indicator
- 17 -
2. Installation procedures
WEEE pamphlet
- 18 -
2. Installation procedures
6.Installation procedure
6.1 List of tools required
Notebook
computer available
Used to check the wireless environment before installation.
on the market
4 1 unit Used to check the wireless environment when trouble with
(incorporating the
wireless communication has occurred.
IEEE 802.11a/b/g/n
specification)
5 Mouse 1 unit
6 Keyboard 1 unit
Must support Gigabit Ethernet.
7 LAN cable 1 unit Used for connecting to the image capture computer and
notebook PC.
Mirror, oil-based
8 1 set Used for adjusting the alignment with the X-ray tube.
marker pen
- 19 -
2. Installation procedures
6.2 System installation procedures
This section broadly divides up the installation procedures by how the installation location (regular
radiography room or mobile X-ray system), operation mode (standard synchronization mode, non
generator connection mode) and communication method (wireless communication, wired
communication or wireless & wired communication) are combined, and it describes each of these
procedures.
The major difference in the installation procedure when it comes to the operation mode (standard
synchronization mode, non generator connection mode) is that the X-ray interface box need not be
connected in the non generator connection mode whereas it is necessary to connect it in the standard
synchronization mode.
Standard synchronization mode and non generator connection mode (regular radiography room:
wireless & wired, wireless)
Procedure Item Conditions and checkpoints
■Checking the LNK Controller settings
• Set the power of the Image Capture Computer to ON.
Check the LNK • Set the Use Link Controller setting to ON using the
1
Controller startup. AP/LNK Configuration Tool of the CCS service tool.
(Click the [SAVE] button.)
■ Connecting the component used for the infrared link
• Connect the ready indicator to one of the USB
Connect the ready connectors of the Image Capture Computer.
indicator to the Image • When using the infrared communication unit, install the
Capture Computer. unit’s driver, and connect the unit to the USB connector
of the Image Capture Computer.
■ Connecting the components used for a wireless connection
Connect the wireless
• This connection must be made through a hub.
LAN access point to the
• This connection is not required when using a USB
Image Capture
wireless LAN adapter.
Computer.
• Turn on the power of the Image Capture Computer, and
select wireless communication from the AP/LNK
Configuration Tool.
• When using the ready indicator for the infrared link,
Set the wireless LAN
select “NOT SET” as the COM port setting.
access point.
• When using the infrared data communication unit for
the infrared link, select the COM port to which the unit is
connected.
(Click the [SEND] button.)
• The access point must be installed outside the patient
Install the wireless LAN environment.
access point. • Install the wireless LAN access point at a distance of
less than 8 meters from the detector.
■ Establishing the infrared link (detector recognition)
• A fully charged battery must have been correctly
Turn on the power of
installed in the detector.
the detector.
• The detector LED must light.
• The distance between the detector and infrared data
communication unit or ready indicator must be less than
30 cm.
Recognize the detector. • Care must be taken not to drop the detector.
• The detector must be recognized in a location outside
the patient environment.
• Wireless communication must be performed properly.
- 20 -
2. Installation procedures
■ Connecting the X-ray interface box (*not necessary for operation in the non generator
connection mode)
Connect the Image • This connection must be made through a hub.
Capture Computer to • Route the LAN cable in such a way that it will not be
the X-ray interface box. subject to any unreasonable loads.
Connect the X-ray
• Route the X-ray interface cable in such a way that it
generator to the X-ray
will not be subject to any unreasonable loads.
interface box.
■ Connecting the wiring unit (*not necessary for operation using a wireless connection only)
Connect the necessary
• Connect the AC power cable, status indicator cable,
cables to the AC/DC
LAN cable and detector cable.
box.
Connect the Image
Capture Computer to • This connection must be made through a hub.
the AC/DC box.
Connect the AC/DC • Remove the connector cap from the external connector
box to the detector area of the detector, and connect the detector cable. The
using the detector cable. detector cable is adsorbed by magnets.
■ Setting the IP address for the wired connection (*not necessary for operation using a wireless
connection only)
• After setting the wired connection IP setting using the
Set the IP address used CCS service tool, establish the infrared link. *Depending
for the wired on the wired link, it is also possible to set the wireless
connection. and wired IP addresses.
For further details, refer to “7.4.2 Wired link.”
■ Checking operation after connecting the components
• Select the generator settings and establish the
CCS settings association with workspace imaging, and set the
necessary items.
Calibration • No errors must be displayed.
Self-diagnosis test • No errors must be displayed.
• Check that the images are transferred to the printer and
Imaging test
storage medium, and check their quality.
Final checks • Carry out these checks as per the check sheet.
Cleaning
- 21 -
2. Installation procedures
Standard synchronization mode and non generator connection mode (regular radiography room: wired)
Procedure Item Conditions and checkpoints
■ Turning on the power of the Image Capture Computer
• Set the Use Link Controller setting to ON using the
Check the LNK
1 AP/LNK Configuration Tool of the CCS service tool.
Controller startup.
(Click the [SAVE] button.)
■ Connecting the component used for the infrared link
• Connect the ready indicator to one of the USB
Connect the ready connectors of the Image Capture Computer.
indicator to the Image • When using the infrared communication unit, install the
Capture Computer. unit’s driver, and connect the unit to the USB connector
of the Image Capture Computer.
■ Connecting the X-ray interface box (*not necessary for operation in the non generator
connection mode)
Connect the Image • This connection must be made through a hub.
Capture Computer to • Route the LAN cable in such a way that it will not be
the X-ray interface box. subject to any unreasonable loads.
Connect the X-ray
• Route the X-ray interface cable in such a way that it
generator to the X-ray
will not be subject to any unreasonable loads.
interface box.
■ Connecting the wiring unit
Connect the necessary
• Connect the AC power cable, status indicator cable,
cables to the AC/DC
LAN cable and detector cable.
box.
Connect the Image
Capture Computer to • This connection must be made through a hub.
the AC/DC box.
Connect the AC/DC • Remove the connector cap from the external connector
box to the detector area of the detector, and connect the detector cable. The
using the detector cable. detector cable is adsorbed by magnets.
■ Establishing the wired link
Link operation using a • With the detector cable connected to the detector, the
wired connection (state power of the detector is turned on by the power link
in which the detector function when the power of the wiring unit is turned on.
cable is connected to The IP address used for the wireless connection is
the detector) allocated automatically.
After establishing the wired IP setting using the CCS
Set the IP address used
service tool, disconnect the detector cable from the
for the wired
detector while the power of the AC/DC box remains ON,
connection.
and then re-connect the cable (re-establish the link).
■ Checking operation after connecting the components
CCS settings • Set the required items.
Calibration • No errors must be displayed.
Self-diagnosis test • No errors must be displayed.
• Check that the images are transferred to the printer and
Imaging test
storage medium, and check their quality.
Final checks • Carry out these checks as per the check sheet.
Cleaning
- 23 -
2. Installation procedures
• Select the “Automatically acquire IP address” setting in
the case of a connection to the main LAN.
• When establishing a wired connection to the USB port
Setting the IP address
using the PC connecting cable, set the IP address to
192.168.101.1/255.255.255.0 using the Ethernet-USB
adapter.
■ Setting the IP address for the wired connection (*not necessary for operation using a wireless
connection only)
• After setting the wired connection IP setting using the
Set the IP address used CCS service tool, establish the infrared link. *Depending
for the wired on the wired link, it is also possible to set the wireless
connection. and wired IP addresses.
For further details, refer to “7.4.2 Wired link.”
■ Checking operation after connecting the components
• Establish the association with workspace imaging in the
CCS settings
generator settings, and set the necessary items.
Calibration • No errors must be displayed.
Self-diagnosis test • No errors must be displayed.
• Check that the images are transferred to the printer and
Imaging test
storage medium, and check their quality.
Final checks • Carry out these checks as per the check sheet.
Cleaning
- 24 -
2. Installation procedures
Non generator connection mode (mobile X-ray system: wireless and wired)
Procedure Item Conditions and checkpoints
■ softAP installation
• Install softAP from the CD provided with the notebook
PC.
• After installing softAP, set the SSID, channel, network
authentication (select WPA2-PSK), data encryption
(select AES) and PSK (network key) on the setting
Installing softAP
screen.
• Make a note of the SSID, channel and PSK (network
key) settings selected here because they must be set on
the Configuration tab of the AP/LNK Configuration
Tool.
■ Setting the notebook PC internal LAN port
Setting the LAN port • The network address must be an address other than
incorporated in the 192.168.100.1 or 255.255.255.0.
notebook PC • Turn off the power of the notebook PC.
■ CCS service tool AP/LNK Configuration Tool settings (to be selected before establishing the
infrared communication link)
Settings selected for the • Using the AP/LNK Configuration Tool, select the
access point on the ESSID, PSK and Channel settings on the Configuration
Configuration tab tab.
• Select the Access Point and Link Module settings on
Settings selected on the
the Connection tab using the AP/LNK Configuration
Connect tab using the
Tool.
AP/LNK Configuration
• Click the [SAVE] button. (There is no need to click the
Tool
Send button.)
■ Connecting the component used for the infrared link
Connect the ready
indicator to the Image • Connect it to one of the USB connectors on the Image
Capture Computer Capture Computer (notebook PC).
(notebook PC).
■ Establishing the infrared link (detector recognition)
• A fully charged battery must have been correctly
Turn on the power of
installed in the detector.
the detector.
• The detector LED must light.
• The distance between the detector and infrared data
communication unit or ready indicator must be less than
30 cm.
Recognize the detector. • Care must be taken not to drop the detector.
• The detector must be recognized in a location outside
the patient environment.
• Wireless communication must be performed properly.
■ Ethernet-USB adapter settings (LAN port in the case of a connection to the main LAN)
Install the driver of the • For details of the installation procedure, refer to the
Ethernet-USB adapter. manual provided.
• Select the “Automatically acquire IP address” setting in
the case of a connection to the main LAN.
• When establishing a wired connection to the USB port
Setting the IP address
using the PC connecting cable, set the IP address to
192.168.101.1/255.255.255.0 using the Ethernet-USB
adapter.
- 25 -
2. Installation procedures
■ Setting the IP address for the wired connection (*not necessary for operation using a wireless
connection only)
• After setting the wired connection IP setting using the
Set the IP address used CCS service tool, establish the infrared link. *Depending
for the wired on the wired link, it is also possible to set the wireless
connection. and wired IP addresses.
For further details, refer to “7.4.2 Wired link.”
■ Checking operation after connecting the components
• Select the generator settings and establish the
CCS settings association with workspace imaging, and set the
necessary items.
Calibration • No errors must be displayed.
Self-diagnosis test • No errors must be displayed.
• Check that the images are transferred to the printer and
Imaging test
storage medium, and check their quality.
Final checks • Carry out these checks as per the check sheet.
Cleaning
* Refer to the following sections for the system diagrams.
7.2.4 Non generator connection mode (mobile X-ray system: wireless)
7.2.5 Non generator connection mode (mobile X-ray system: wired)
* Check the following sections for the detailed procedures.
7.9 System connections in the non generator connection mode (mobile X-ray system
environment: outside Japan)
- 26 -
2. Installation procedures
7. Installation
7.1 System connection diagrams (standard synchronization mode)
7.1.1 Standard synchronization mode (regular radiography room: wireless and wired)
To be procured by
Infrared data the sales companies
communication unit
Standard configuration
cables
HUB
Optional cables
Capture
Computer
Wireless LAN Ethernet cables
X-ray access point procured by the sales
companies
interface box
Inside Outside
the radiography room the radiography room
- 27 -
2. Installation procedures
Battery pack
Standard unit
To be procured by
Infrared data the sales companies
communication unit
Standard configuration
cables
HUB
Image Capture Optional cables
Inside Outside
the radiography room the radiography room
Patient Outside
environment the patient environment
- 28 -
2. Installation procedures
To be procured by
Infrared data the sales companies
communication unit
Standard configuration
cables
HUB
Optional cables
Capture
Computer
Ethernet cables
procured by the sales
X-ray companies
interface box
Inside Outside
the radiography room the radiography room
Patient Outside
environment the patient environment
- 29 -
2. Installation procedures
To be procured by
Infrared data the sales companies
communication unit
Standard configuration
cables
AC/DC box
HUB
Image Capture Optional cables
Computer
medical facility
Wireless LAN Ethernet cables procured
LAN inside
access point by the sales companies
Inside Outside
the radiography room the radiography room
Patient Outside
environment the patient environment
- 30 -
2. Installation procedures
To be procured by
Infrared data the sales companies
communication unit
Standard configuration
cables
HUB
Optional cables
Capture
Computer
Wireless LAN Ethernet cables procured
access point by the sales companies
Inside Outside
the radiography room the radiography room
Patient Outside
environment the patient environment
- 31 -
2. Installation procedures
To be procured by
Infrared data the sales companies
communication unit
Standard configuration
cables
HUB
Optional cables
Capture
Computer
Ethernet cables procured
by the sales companies
Inside Outside
the radiography room the radiography room
Patient Outside
environment the patient environment
- 32 -
2. Installation procedures
Battery pack
Dedicated charger
Standard unit
Optional unit
Ready indicator unit
To be procured by
the sales companies
Standard configuration
X-ray cables
generator Detector
Optional cables
Ethernet cables
procured by the sales
companies
Image Capture Computer
(The notebook PC specified by Canon Inc.
must be used.) Cables procured
USB wireless LAN adapter to be procured by the sales companies
(item specified by Canon Inc. must be
used; in Japan only)
Restrictions:
With this installation, the following restrictions apply to the connection of external devices to a notebook PC.
If there is a possibility that a patient might touch the notebook PC, do not connect the AC adaptor of the
notebook PC, the main network inside the hospital or clinic, or any devices that require an external power
supply.
The need may conceivably arise to connect an AC adapter in order to charge the battery of the notebook PC
or to initiate connection to the hospital’s main network for the purpose of capturing study orders or
transferring the acquired images to an external system, but these activities are to be restricted because
electrical safety compliant with the 60601-1 standard is not guaranteed with either the AC adapter or the
hospital’s main network, and patients will be exposed to the hazard of electric shocks.
- 33 -
2. Installation procedures
Battery pack
Dedicated charger
Standard unit
Optional unit
Ready indicator unit
To be procured by
X-ray Detector the sales companies
generator
Standard configuration
cables
Optional cables
Restrictions:
With this installation, the following restrictions apply to the connection of external devices to a notebook PC.
If there is a possibility that a patient might touch the notebook PC, do not connect the AC adaptor of the
notebook PC, the main network inside the hospital or clinic, or any devices that require an external power
supply.
The need may conceivably arise to connect an AC adapter in order to charge the battery of the notebook PC
or to initiate connection to the hospital’s main network for the purpose of capturing study orders or
transferring the acquired images to an external system, but these activities are to be restricted because
electrical safety compliant with the 60601-1 standard is not guaranteed with either the AC adapter or the
hospital’s main network, and patients will be exposed to the hazard of electric shocks.
- 34 -
2. Installation procedures
Turn on the power of the Image Capture Computer, and check that LNK Controller is running.
• Start the AP/LNK Configuration Tool from the service tool, select [ON] for “Use Link
Module” on the Connect tab, and press the [Save] button to start LNK Controller. By selecting
this setting, LNK Controller is started automatically when the PC is started up.
• For further details, refer to Appendix 6 AP/LNK Configuration Tool in 4. Setting Reference of
the CXDI Control Software NE Service Manual.
Insert the USB cable (procured by the sales companies) from the cable fitting area (circled in
red) of the ready indicator unit (optional product).
• Only the USB cable specified by Canon Inc. may be used.
USB cable
- 35 -
2. Installation procedures
Connect the USB cable to the connector area of the ready indicator unit.
• Connect the cable in such a way that the connector is not inserted the wrong way round.
USB cable
• In addition to its function for displaying the imaging enabled status (ready status) of the
detector, the ready indicator unit has an infrared communication function.
Infrared communication area
When an infrared data communication unit (procured by the sales companies) has been
purchased, install its driver from the CD provided.
For details of the driver installation procedure, refer to the Manual and Quick Guide contained
on the CD provided.
When the ready indicator unit (optional product) is used, the infrared data communication unit
will not be needed.
- 36 -
2. Installation procedures
Connect the ready indicator unit (optional product) or the infrared data communication unit
(procured by the sales companies) to the USB connector on the Image Capture Computer.
Check the COM port if the infrared data communication unit has been connected.
Even when both the ready indicator unit and infrared data communication unit are concurrently
connected, they will function without any problems.
Infrared communication can be performed by using either the ready indicator unit or infrared
data communication unit.
Infrared data
communication unit
Ready indicator unit
6) Connect the access point to the Image Capture Computer through the hub.
• The Hub/LAN cables to be procured by the sales companies must support Gigabit Ethernet.
• Only the access point specified by Canon Inc. may be used.
Access point
Image
LAN cables Capture
Computer
- 37 -
2. Installation procedures
8) Start the AP/LNK Configuration Tool from the service tool, input the setting items on the
Connect tab and Configuration tab, and click the [Send] button.
• For further details of the setting items of the Connect tab and Configuration tab, refer to
section 4.3.2 “In the case of Ferry protocol and wirelessly connected detectors” under 4.
Essential settings in 2. Installation Manual of the CXDI Control Software NE Service Manual.
- 38 -
2. Installation procedures
• COM port setting
Bear in mind that the setting for COM port on the Connect tab differs depending on whether the
ready indicator unit is used or the infrared data communication unit is used.
* COM port setting when using only the ready indicator unit:
Select “Not set.” (The COM port setting need not be selected.)
* COM port setting when using only the infrared data communication unit:
Select the COM port to which the infrared data communication unit is connected.
* Setting when using both the ready indicator unit and infrared data communication unit:
Select the COM port to which the infrared data communication unit is connected.
9) Install the access point in such a way that the detector is less than 8 meters away with the
access point as the starting point.
• Do not place any structures between the detector and access point which will block
communication.
Access point
- 39 -
2. Installation procedures
Using infrared communication, establish the communication line between the detector and Image
Capture Computer to get ready for imaging. In this Service Manual, the establishment of this
communication line is referred to as “link” or “sensor recognition.”
Note:
* The status LED of the battery charger lights up green if the battery
is fully charged.
* The battery must be installed in the detector when either wired
Battery
connections or wireless connections are used.
Rear panel of detector
How to remove the battery
1) With the battery held in place, slide the lock switch to the UNLOCK side.
2) The battery now pops up, and can be removed.
1) Press the power switch on the side of the detector for at least 0.5 seconds to turn on the power.
Note:
* Check that the blue LED on the detector lights.
* The link (recognition/connection) must be established with no monitors or reflective objects in the
vicinity.
* To turn off the power of the detector, hold down the power switch for 3 seconds.
Infrared communication
area
2) Ensure that the distance between the infrared communication area on the side of the detector and the
signal sensing area on the infrared data communication unit is less than 30 cm.
Note:
* Make sure that the power of the image capture computer has been turned on and that link Controller
has started up.
- 40 -
2. Installation procedures
Note:
* The distance between the signal sensor of the infrared data communication unit and the infrared
communication area on the detector must be less than 30 cm and the incident angle of the signals must
be less than 10 degrees. The work environment must not be close to the monitor or an environment
where signals are reflected.
* The LED (blue-green) on the detector does not light when the detector is not registered or when the
communication circuit is not established.
4) Connection to the access point is made automatically, and the communication circuit is established.
• Two LEDs (blue and blue-green) on the detector light up. (The green LED goes off.)
Note:
* If the communication circuit is not established, the blue-green LED on the detector will not light.
* Do not place any structures that will block communication between the detector and access point.
- 41 -
2. Installation procedures
5) Start the service tool, and check that the detector which has registered the detector is displayed.
Select [Service Tool] → [Connection] → [Detector], and then open the [Ferry Static Detector] screen.
• If, after the detector has been registered, the detector is not reflected on the Connection tab of the
service tool, start the service tool again.
• For details, refer to “4. Essential Settings” in Chapter 2. Installation Manual of the CXDI Control
Software NE Service Manual.
(3) When using wireless and wired connection with one detector
1) Select [Service Tool] → [Connection] → [Detector], and then open the [Ferry Static Detector] screen.
3) From the [IP] tab, set [Use wired option] of [Wired] to ON, set the target IP address, and then register
the detector again.
• For details, refer to “Appendix6 AP/LINK Configuration Tool” in Chapter 4. Setting Reference of the
CXDI Control Software NE Service Manual.
Note:
* Allocate different IP addresses when using both wireless and wired connections with one detector.
Example: Serial no. 19000001
IP address for wireless connection: 192.168.100.11 (assigned automatically)
IP address for wired connection: 192.168.100.12 (changed manually)
- 42 -
2. Installation procedures
1) Remove the two screws of each of the two side panels (a total of four screws) and the two screws on
the bottom panel of the X-ray interface box.
* Take care not to strip the screw heads of the side panel and bottom panel.
2) Slide the top cover toward the connector connection side of the X-ray interface box, and raise it to
remove it.
* Do not tilt the top cover too far since this may cause the inside of the X-ray interface box to be
damaged.
3) Remove the top cover. The box now looks like what is shown in the figure below.
- 43 -
2. Installation procedures
(2) Cable connections
1) Remove the skin-top spiral piece and lock nut of the X-ray interface cable.
Lock nut
2) As shown in the figure below, pass the X-ray interface cable through the X-ray interface box cabinet,
and loosely tighten the lock nut.
3) Pass the X-ray interface cable through the cable clamp, and connect the cable connector to the X-ray
interface PCB. Do not subject the cable to any stress and also, in order to prevent the excess length from
touching the devices on the circuit board, adjust the cable length and tighten the skin-top spiral piece in
such a way that no gap will be left as shown in the figure below.
Connectors
Cable clamp
No gap to be
left here
Skin-top spiral piece
4) Using a wrench, fasten the skin-top spiral piece and lock nut which were loosely tightened.
- 44 -
2. Installation procedures
5) Upon completion of the connections, attach the covers of the X-ray interface box.
* Take care not to strip the screw heads of the side panel and bottom panel.
* The screws used for mounting must be tightened up in the following sequence: bottom panel (screws
with external teeth) → both side panels (brazier head screws).
AC power cable
7) Connect the supplied grounding wire (functional ground) to the ground terminal on the X-ray
interface box. The grounding wire is used for functional ground, but not protective earth. To prevent the
equipments from being damaged, connect to other than potential equalization bus (grounding wire of the
distribution panel or iron frame).
Grounding wire
8) Connect the LAN cable and DC plug (of the AC adapter) to the connector connection area of the X-
ray interface box.
* Be absolutely sure to use the AC adapter and AC power cable which come supplied as accessories
with this product.
DC plug
X-ray interface cable
Power LED
Grounding wire
LAN cable
- 45 -
2. Installation procedures
9) Connect the LAN cable, which has been connected to the X-ray interface box, to the hub.
* The hub must be connected to the image capture computer.
* Always setup the X-ray interface box outside of the patient environment.
- 46 -
2. Installation procedures
The detector is equipped with an external connector which enables a wired connection using the
wiring unit.
These connections need not be undertaken when the detector will be used with a wireless
connection only.
The wiring unit is used when a wired connection is used, and it supports imaging while still
installed in the Bucky unit of the reclining table or upright stand.
By establishing a wired connection, communication is performed and power is supplied through
the detector cable so this kind of connection can be used if the communication line is unstable
with a wireless connection or as a backup connection when the battery does not have a sufficient
charge.
It is necessary to install the battery in the detector even when a wired connection has been
established.
AC/DC box
Detector
[Description of functions]
If the detector cable has been connected while the detector is operating with a wireless
connection (driven by the battery) and while the power of the AC/DC box is on, operation will
switch to the wired connection (the detector will be powered from an external source). At the
same time, the charging of the battery installed in the detector will start when the detector is
powered from an external source. However, the battery charging is stopped temporarily during
radiography.
Operation will switch to the wired connection (the detector is driven by the battery) if the power
of the AC/DC box is turned off while the detector is operating with a wired connection (while it
is powered from an external source). At the same time, the charging of the battery installed in
the detector will stop.
If the detector cable has been disconnected from the detector while the detector is operating
with a wired connection (while it is powered from an external source) and while the power of
the AC/DC box is on, operation will switch to the wireless connection (the detector will be
driven by the battery).
*For a description of the functions which are available during switching between wired and
wireless connections, refer to “s12034-007 Automatic wireless/wired switching” in the
Description of New Functions V2.10 for CXDI Control Software NE.
- 47 -
2. Installation procedures
<Wired and wireless icon displays>
Wired (power supplied from external source)
- 48 -
2. Installation procedures
A wired link can be established.
Link operations can be performed using a wired cable with this product. For further details of
the wired link, refer to “7.4.3 Wired link.”
AC
adapter
Wireless
Wired link AP
Detector
Control PC
1) Connect the AC power cable, status indicator cable, LAN cable and detector cables to the
connector area of the AC/DC box.
* For the AC power cable, the one provided with this product must be used without fail.
* Check the direction in which the detector cable connector is to be inserted, and then insert the
connector.
* Check the direction in which the status indicator cable connector is to be inserted, and then
insert the connector.
* Route the cables without forcibly bending, pinching or bridging them.
AC/DC box
AC power cable
Detector cable
- 49 -
2. Installation procedures
2) Connect the LAN cable, which has been connected to the AC/DC box, to the hub.
* The hub must be connected to the Image Capture Computer.
3) Remove the connector cap from the external connector area of the detector, and connect the
detector cable.
* The detector cable is adsorbed by magnets. When disconnecting the detector cable, pull the
connector area of the cable up or down to disconnect the cable.
* Align the interlocking parts with the guide pins. Take care not to leave any gaps when
attaching the detector cable.
Detector
External connector
Detector cable
connection area
Connector area
Magnet
Guide pin
Magnet
* The user can select the direction in which to pull out the detector cable. (It does not matter in
which the direction the cable points.)
- 50 -
2. Installation procedures
(2) Setting the IP address used for a wired connection
1) Select the X-ray generator/detector setting on the service tool menu, and then click Detector.
2) On the Detector screen, first select [Ferry Static Detector], and then select the Wireless
detector, whose IP address to be used for the wired connection is to be registered, from among
the Wireless detectors registered on the [Registered Detectors] list.
3) On the [IP] tab screen, select [Wired], select ON for [Use wired option], input the target IP
address, and then click the [OK] button.
Note:
* A different IP address from the ones of the other registered Wireless detector must be set as
the target IP address to be used for the wired connection.
- 51 -
2. Installation procedures
Start the AP/LNK Configuration Tool from the service tool, input “192.168.100.10” for “Local
IP address” on the Connect tab, select ON for “Use Link Module,” and click the [Save] button.
• If only a wired connection is going to be used for operation, an access point is not installed so
leave the default settings in place for “Target IP address” and “Subnet mask.” However, do not
click the [Send] button to send the settings to the access point.
Connect the Image Capture Computer and X-ray interface box through the hub.
The Hub/LAN cables to be procured by the sales companies must support Gigabit Ethernet.
For further installation details, refer to “7.3.3 Connecting to the X-ray interface box (XB-1A).”
Connect the Image Capture Computer and AC/DC box through the hub.
The Hub/LAN cables to be procured by the sales companies must support Gigabit Ethernet.
For further installation details, refer to “7.3.4 Wiring unit (WU-3A) connections.”
6) Establish the wired link, and allocate the wireless IP addresses. Then set the wired IP
addresses, and re-establish the wired link.
Refer to * Case 1 in “7.4.2 Wired link.”
- 52 -
2. Installation procedures
7.4.2 Wired link
In addition to link operations using infrared communication (ready indicator unit and infrared
data communication unit), this product allows link operations to be performed using a wired
connection. The wired connection can be established regardless of whether the AC/DC box
(WU-3A) is present or not.
The wired link operation procedure differs depending on the configuration of the units in the
system so refer to the reference cases (case 1 to case 3) described below.
* Case 1: Wired link operation procedure (without an access point or ready indicator)
1) Start the Image Capture Computer, and start the AP/LNK Configuration Tool from the
service tool.
2) Input “192.168.100.10” for “Local IP address” on the Connect tab, and select ON for “Use
Link Module.”
3) Click the [Save] button, and check that LNK Controller starts.
4) Connect one end of the detector cable to the AC/DC box (WU-3A).
5) Connect the other end of the detector cable to the wired cable connection socket on the
detector, and turn on the power of the WU-3A.
The power of the detector is turned on by the power link function.
6) The wired link (sensor registration) is established automatically. Allocate the IP addresses for
a wireless connection.
7) Select the X-ray generator/detector setting on the service tool menu, and then click Detector.
On the Detector screen, first select [Ferry Static Detector], and then check that the detector
whose wired link has been established has been registered on the [Registered Detectors] list.
8) Select the detector which has been registered by the wired link, and set the IP address used
for the wired connection from the [IP] tab.
9) With the power of the AC/DC box on, first disconnect the detector cable from the detector,
and then re-connect it (re-establish the link).
Detector Control PC
Wired link
- 53 -
2. Installation procedures
* Case 2: Wired link operation procedure (with an access point and ready indicator)
1) Start the Image Capture Computer, and start the AP/LNK Configuration Tool from the
service tool.
2) Input “192.168.100.10” for “Local IP address” on the Connect tab, select ON for “Use Link
Module.”
Select the COM port to which the infrared communication unit (existing IrDA) has been
connected as the COM port setting.
When using the ready indicator unit (RI-3A), select “Not set” as the COM port setting. Next,
click the [Save] button, and check that LNK Controller starts.
3) Select the settings on the Configuration tab, and click the [Send] button to complete the AP
settings.
4) Connect one end of the detector cable to the AC/DC box (WU-3A).
5) Connect the other end of the detector cable to the wired cable connection socket on the
detector, and turn on the power of the WU-3A.
The power of the detector is turned on by the power link function.
6) The wired link (sensor registration) is established automatically. Allocate the IP addresses for
a wireless connection.
7) Select the X-ray generator/detector setting on the service tool menu, and then click Detector.
On the Detector screen, first select [Ferry Static Detector], and then check that the detector
whose wired link has been established has been registered on the [Registered Detectors] list.
8) Select the detector which has been registered by the wired link, and set the IP address used
for the wired connection from the [IP] tab.
9) With the power of the AC/DC box on, first disconnect the detector cable from the detector,
and then re-connect it (re-establish the link).
Switching HUB
X-ray interface
X-ray interface unit
cable AC adapter
X-ray generator
AC adapter
Wireless
AP
Wired link
Detector Control PC
- 54 -
2. Installation procedures
*Start the wired link by connecting the detector cable to the wired cable connection socket of
the detector.
- 55 -
2. Installation procedures
The installation steps from the system connections in the non generator connection mode
(wireless and wired) to the connection of the wiring unit (WU-3A) are the same as for the
standard synchronization mode so refer to the sections listed below.
* The installation procedure is different for operation in a mobile X-ray system environment
(wireless) in the non generator connection mode so refer to “7.8 System connections in the non
generator connection mode (mobile X-ray system environment: wireless).”
The installation steps for the system connections in the non generator connection mode (wired)
are the same as for the standard synchronization mode so refer to the sections listed below.
* The installation procedure is different for operation in a mobile X-ray system environment
(wired) in the non generator connection mode so refer to “7.9 System connections in the non
generator connection mode (mobile X-ray system environment: wired).”
- 56 -
2. Installation procedures
7.5.3 Overview of non generator connection imaging
Starting with CXDI Control Software NE V2.10, non generator connection imaging is supported
in environments where only detectors (CXDI-701G/701C) that support non generator
connection are used.
When the detector is operated with non generator connection imaging, use of the ready indicator
is recommended.
X-ray generator
Ready indicator
X-rays
Detector PC
- 57 -
2. Installation procedures
The imaging flow in the non generator connection mode is shown below.
(1) Exposure now ready
The ready indicator notifies the operator by light and sound of the transition to the Ready status
and informs the operator of the exposure timing. The status is also displayed on GUI.
(2) Exposure
The X-ray exposure time in the non generator connection mode is 1 second.
(3) Automatic detection of X-ray exposure
(4) Storage of images
(5) Transfer of images from detector to PC
<Caution>
In the image capturing flow (1) above, you must ensure X-ray preparation is complete (the
Ready state) before performing X-ray exposure.
*The following points must be borne in mind during operation in the non generator connection
mode.
▪ For a description of the detailed settings of the non generator connection mode, refer to
“Appendix N3. Non generator connection operation in Setting Reference of the CXDI
Control Software NE Service Manual”.
▪ For details on the operation methods during operation in the non generator connection
mode, refer to “Automatic X-ray imaging” in the User Operation Manual.
▪ When using a 34 lp/cm grid and 40 lp/cm grid in the non generator connection mode, the
grid must be installed in such a way that the grid stripes are parallel to the longer sides of
the 701G/701C detector.
▪ During operations in the non generator connection mode, only a detector (CXDI-
701G/701C) which supports non generator connection can be used, and it is not possible to
use a detector which does not support this. When a detector which does not support non
generator connection has been set for the workspace, an error (F040300053) will result
when CCS is started.
▪ The maximum X-ray exposure time in the non generator connection mode is as follows.
1000 ms or less (Standard setting)
3000 ms or less (Long exposure is set)
▪ If the exposure dosage is insufficient, artifacts may occur on the images.
In cases like this, increase the tube current. If it is not possible to increase the tube current,
the tube voltage can be increased instead to resolve the problem.
▪ When the exposure field has been narrowed down and images have been taken (such as
when the image of one finger has been taken), artifacts may occur at the top and bottom
edges of the exposure field.
- 58 -
2. Installation procedures
In cases like this, set to an exposure field which is wide enough for the diagnosis area
concerned.
▪ When the detector is exposed to the X-rays before it is set to exposure now ready (Ready
status) or immediately before it transitions to the ready status, the X-rays may not be
detected and the images may not be displayed after they have been taken.
The electrical charge will remain accumulated in the detector when the detector is not in
the Ready status and exposure is undertaken by mistake so when the next images are taken,
the “effects retained on the images” or “X-rays cannot be detected” event may occur. In
this case, it is necessary to refresh the accumulated charge by pressing the Refresh button
on the GUI. (For details on the Refresh button, refer to 7.5.5 Refresh button.)
1) Turn on the power of the Image Capture Computer, and start the service tool.
2) Click X-ray generator/detector settings on the service tool menu > X-ray generator.
3) The X-ray generator setting screen now appears. Set the “Non Generator Connection Mode”
under the exposure settings to on, and enable non generator connection.
(Default setting: Generator Connection Mode)
- 59 -
2. Installation procedures
4) After taking step 3), click [OK] or [Apply] to save the settings.
*When [OK] or [Apply] is clicked, the setting change check message is displayed.
When [OK] is clicked, “Show image analysis error” and “Show grid warning” on examination
screen 1 accessed by setting System Setting followed by Application setting are set to off, the
settings are saved, and the menus screen is restored.
When [Cancel] is clicked, the setting change check message is closed.
<Reference>
If an attempt has been made to close the service tool while non generator connection imaging is
enabled by the X-ray generator settings and an existing detector (which does not support non
generator connection) incapable of non generator connection imaging is associated with the
workspace, the warning message shown below will be displayed.
When [OK] is clicked and the service tool is closed at this point, the following fatal error is
displayed when CCS is started up.
- 60 -
2. Installation procedures
7.5.5 Refresh button
By enabling the non generator connection setting, the Refresh button is added at the top right of
the examination screen, and this button can now be operated.
[Header Menu (the lower of the two bars) of the examination screen
when the non generator connection setting is enabled]
For details on the imaging procedure and operation methods in the non generator connection
mode, refer to “Automatic X-ray imaging” in the User Operator Manual.
- 61 -
2. Installation procedures
7.6 Calibration during non generator connection operations
When non generator connection settings are enabled, the user cannot perform
calibration/performance testing. Therefore, generally a service technician must perform
calibration and performance testing at the site where non generator connection operation is
taking place.
For calibration, when non generator connection settings are enabled, the operator performs X-
ray imaging by manually coordinating the timing. Using the CXDI-701G Wireless as an
example, the calibration screen transitions are described below. Note that the same is true for
performance testing when non generator connection settings are enabled.
Shows the calibration preparation screen when non generator connection settings are enabled.
If you press the start button on the preparation screen, the detector preparation notification
screen is displayed.
- 62 -
2. Installation procedures
(2) Detector preparation notification screen
It takes approximately 50 seconds for the detector preparation notification screen (Detector
preparations underway) to change to the detector preparation notification screen (Detector now
ready).
The countdown time settings [seconds] are displayed in the detector preparation notification
screen.
The accumulate time [seconds] is displayed in the detector preparation notification screen.
If you press the “Cancel” button, the detector preparation notification screen closes, and the
screen returns to the main screen of the QC tool.
- 63 -
2. Installation procedures
(3) Countdown screen
[Countdown screen]
The countdown screen performs a countdown until exposure is ready, and it displays the
progress.
Starts the countdown screen display and the countdown.
When counting is finished (remaining time is 0), the countdown screen changes to the exposure
enabled notification screen
When there are 3 seconds remaining in the countdown, the “Cancel” button is disabled.
The exposure enabled notification screen notifies the user that exposure is enabled and prompts
the user to start exposure.
When the accumulate time has passed, the exposure enabled notification screen closes, and the
image check results screen appears.
- 64 -
2. Installation procedures
(5) Image check result screen
Click [Next], and the screen below appears after a total of 4 exposures. Click [End] to complete
the operations.
When the non generator connection setting has been enabled, the service technician must
perform performance testing as it is for calibration. Conduct the performance test by following
the same operation procedure as in “7.6.1 Calibration operation procedure.”
- 65 -
2. Installation procedures
[Details]
1. Overview describing non generator connection
2. Description of how to operate non generator connection
3. Description of the detection performance of the non generator connection function
4. Detection performance in the characteristic image capturing procedure
5. Confirmation method under actual usage conditions
(1) Detection is easier when there is a larger quantity of X-ray exposure per unit of time.
(2) Detection is sometimes difficult when the X-ray exposure time is extremely short.
(3) Detection is easier when the exposure field is wider.
<Other considerations>
Due to differences in the object of the image being captured as well as the output properties of
the X-ray generator, the detectable range cannot be specified uniformly according to the setting
conditions of the X-ray generator.
(Be sure to check actual usage conditions before actually capturing images)
There is no difference in detection performance between GOS products and Csl products.
(They have equivalent detection sensitivities with respect to exposure conditions that achieve
equivalent image quality.)
- 66 -
2. Installation procedures
4. Detection performance in the characteristic image capturing procedure
In the reference example below, by using a human phantom, a normal image was captured using
low-dose radiation in a range that is not harmful to detectability or post-calibration image
quality.
◆ Common conditions:
Manufacturer of X-ray generator:
Hitachi DHF-158H II (nominal maximum electrical output 80 kW)
Exposure distance (source to image distance [SID]): 130[cm]
CXDI-701 C (Csl panel)
Grid: Not provided
a. Front of chest
Exposure conditions: 100[kV], 63[mA], 2[msec] (0.12 mAs)
Exposure field: Full radiation
- 67 -
2. Installation procedures
b. Fingers
Exposure conditions: 45[kV], 50[mA], 10[msec] (0.5 mAs)
Exposure field: 6 cm×10 cm
- 68 -
2. Installation procedures
5. Confirmation method under actual usage conditions.
For the selected image capturing procedure, check the operation of the non generator connection
function in advance, based on the image capturing conditions at your hospital or clinic.
Select the image capture Prepare the following in order to simulate the selected actual usage
procedure to be checked conditions.
- Phantom that simulates the object of the image
- Actually-used X-ray generator
- Normally-used
Capture images usingX-ray
normalsettings
image capture conditions Also, in order to
check the detection capability of non generator connection in a manner that
Actual image capture in non
accounts for variations occurring during actual image capture, implement
generator connection mode
the following.
- Decrease exposure
- Make the simulation phantom thicker
For the target procedure, capture For the target procedure, you can use non For the target procedure, you
images in the usual manner in generator connection mode by cannot capture images in non
non generator connection mode. appropriately adjusting conditions. generator connection mode.
The following is a trouble example of “b. Image artifacts” when checking the actual usage
conditions in the flow described above.
<Avoidance method>
Increase the tube current. If the tube current cannot be increased, increase the tube voltage.
Increase the exposure field so that it is wide enough.
<Avoidance method>
Increase the exposure field so that it is wide enough.
Increase the tube current. If the tube current cannot be increased, increase the tube voltage.
- 69 -
2. Installation procedures
7.8 System connections in the non generator connection mode
(mobile X-ray system environment: Japan)
7.8.1 System connections (wireless and wired)
• Refer to the connection diagram for the non generator connection mode (mobile X-
ray system: wireless) in section 7.2.4.
• Refer to the connection diagram for the non generator connection mode (mobile X-
ray system: wired) in section 7.2.5.
The notebook PC with a wireless LAN incorporated and the USB wireless LAN adapter are procured
by the sales companies, and the system is constructed as shown in the figure entitled “Installation case 1”
below. Furthermore, when connecting the detector and notebook PC by PC connecting cable, the system
is constructed as shown in the figure entitled "Installation case 2".
[Installation case 1] When connecting the detector and notebook PC by wireless connection
Main LAN
基幹LAN
Mobile X-ray system
回診車 TOUGHBOOK
Let’s note CF-C2(win CCS)
(Panasonic)
Internal
softAP 内蔵無線
Wireless AP
Detector (planex) LAN LAN
Restrictions:
制限事項:
The network added of (1) cannot be changed.
①のネットワークアドレスは変更不可
(GW-USDual300 set the address automatically)
(GW-USDual300が自動でアドレスを設定する)
Canon Inc. has inspected the operation of the notebook PC with a wireless LAN incorporated
and the USB wireless LAN adapter made by the manufacturers listed below and verified that
these units operate properly.
Notebook PC with wireless LAN incorporated <<The built-in softAP can be used only in
2.4-GHz band>>
• Manufacturer: Panasonic
• Model: TOUGHBOOK CF-C2
USB wireless LAN adapter <<softAP (Planex) can be used in the 2.4 GHz and 5 GHz bands
>>
• Manufacturer: Planex
• Model: GW-US Dual 300
To connect the main LAN, select either wireless connection using internal wireless LAN or
wired connection through the internal LAN port according to the environment of installation site.
You can also use Ether-USB adapter instead of the internal LAN port. (Refer to 7.8.5 for the
details of Ether-USB).
*Remarks
In the event that a user requests that a USB wireless LAN adapter or a notebook PC with a
wireless LAN incorporated made by other manufacturers (or different models of the same
manufacturers) be used, please address the request to the Medical Equipment PLM Product
Planning Dept. 2 of the Medical Equipment Development Center 2.
- 70 -
2. Installation procedures
[Installation case 2] When connecting the detector and notebook PC by PC connecting cable.
Restrictions:
制限事項:
The network added of (1) cannot be changed.
①のネットワークアドレスは変更不可
(GW-USDual300 set the address automatically)
(GW-USDual300が自動でアドレスを設定する)
When a internal LAN port is to be used for wired connection to main LAN, Ether-USB can be
used to connect a PC connecting cable.
7.8.2 Restrictions
The following restrictions apply to the connection of external devices to a notebook PC operating in a
mobile X-ray system environment.
1) In cases where patients may touch the notebook PC or the detector which is connected to the
notebook PC using the PC connecting cable, the AC adapter, or any devices requiring an external
power source must not be connected to the notebook PC. Main network inside the hospital or clinic
must to be connected, either.
2) The softAP (Planex) network address must be 192.168.100.1 or 255.255.255.0 only (no other
addresses are permitted).
3) When using a wireless connection for the FPD in the 5 GHz band, a USB wireless LAN adapter
(GW-US Dual 300) is required.
4) When a setting for internal LAN port and Ether-USB adapter is made separately, please do not
misuse them. (Do not use the internal LAN port which is set to connect to a detector to connect a
main LAN.)
- 71 -
2. Installation procedures
7.8.3 Setting procedure
This section describes the procedure for enabling communication with the detector using the notebook
PC with a wireless LAN incorporated and the USB wireless LAN adapter as the access point.
This setting procedure assumes that [Installation case 2] described above is used for the system.
For installation case 1, refer to the necessary points.
Operating environment
Notebook PC with wireless LAN incorporated
Model: TOUGHBOOK CF-C2” made by Panasonic
Operating System: Windows 7 Professional 64-bit version
CPU: Intel® Core™ i5--3427U vPro CPU @1.80GHz
Wireless: Intel® Centrino® Advanced-N 6235
Memory installed: 4 GB
7.8.3.1 Settings to invalid the unnecessary items in the notebook PC (TOUGHBOOK CF-
C2) internal LAN port
2.From the control panel, go to program and open the program uninstall
4.In the Intel®Proset/Wireless Wifi software setup window, select Change from the Change, restore or
delete the installation.
5. In the Intel®Proset/wireless Wifi software setup window, select “Intel ®MyWifi technology” in the
custom setup.
Assuming that a wired connection is established for the detector using the PC connecting cable, set the
IP address of the LAN port incorporated in the notebook PC to an address other than 192.168.100.1
/255.255.255.0.
If, as in [Installation case 1], the LAN port incorporated in the notebook PC is to be connected to the
main LAN, select the “Obtain an IP address automatically” setting.
1) Select Control Panel > View network status and tasks > Change adapter settings, and check that
- 72 -
2. Installation procedures
“Intel® 82579” is displayed.
2) Select and right-click Intel® 82579, then open Local area connection properties from Properties, and
select the Internet protocol version concerned. After making the selections, select Properties, and set
the IP address for the wired connection.
• IP address: 192.168.101.1
• Subnet mask: 255.255.255.0
• If, as in [Installation case 2], the LAN port incorporated in the notebook PC is to be connected
to the main LAN, select the “Obtain an IP address automatically” setting.
- 73 -
2. Installation procedures
7.8.3.3 Wireless LAN adapter (GW-US Dual 300) settings (Japan only)
Note: The wireless LAN adapter (GW-USDual300\) can only be used in the mobile X-ray system in
Japan only. Therefore, this section is only a reference for overseas sales companies.
1) Connect the wireless LAN adapter (GW-US Dual 300) before starting the notebook PC.
2) Set the switch used to turn ON/OFF the wireless communication function originally provided on the
notebook PC to OFF.
(This is to prevent contention with the GW-US Dual 300 which is to be connected.)
3)-(1) For the steps to be taken from the software installation to the device connections, refer to the
manual of the manufacturer.
- 74 -
2. Installation procedures
3)-(2) For the steps to be taken for the access point settings, refer to the manual of the manufacturer.
• The SSID, channel, network authentication (select WPA2-PSK), data encryption (select AES) and
PSK (network key) settings are selected here.
• Make a note of the SSID, channel and PSK (network key) settings selected here because they must be
set on the Configuration tab of the AP/LNK Configuration Tool.
Setting items
Network name
Channel
Network key
- 75 -
2. Installation procedures
3)-(3) Change the service to manually startup.
From the control panel, select “System and security”, “control tool”, “service” and then double click
“Realteck11nCU”.
Change the type of startup to “Manual” and press the “Stop” button. Then press the “OK” button.
- 76 -
2. Installation procedures
After the GW-US Dual 300 access point settings have been completed, “GW-US Dual300 MAC0” is
displayed as a result of selecting Control Panel > View network status and tasks > Change adapter
settings.
The 192.168.100.1 address is used automatically by the GW-US Dual 300. This means that the IP
address setting on the connection properties screen shown below must not be changed.
[Checkpoints]
• Timing at which the wireless LAN adapter (GW-US Dual 300) is to be connected
The wireless LAN adapter (GW-US Dual 300) must be connected before starting the notebook PC. If
the wireless LAN adapter has been connected after the notebook PC has been started up, the power of
the PC must be shut down and then turned back on again after the access point settings have been
selected.
- 77 -
2. Installation procedures
• When the communication with detector cannot be established after shutting down the PC and turning
onf the PC.
Connect the AC adapter first and then start the notebook PC.
(2) Start the text editor (such as notepad), and input the characters indicated below. Then name the file
“boot.bat,” and save it in the script folder. (Create a file with the .bat extension.)
netsh wlan set autoconfig enabled = yes interface=” wireless network connection 3”
ping -n 5 localhost>nul
net start Realtek11nCU
*Alphanumeric characters and symbols (=, ", >): one-byte / Strings of wireless network connecting:
two-byte / Space: one-byte.
*If “GW-USDual300 MAC0” is wireless connection 2, input the “wireless connection 2” characters
into the text.
- 78 -
2. Installation procedures
(3) Start the text, and input the characters indicated below. Then name the file “shutdown.bat,” and
save it in the script folder. (Create a file with the .bat extension.)
netsh wlan set autoconfig enabled = yes interface=” wireless network connection 3”
*Alphanumeric characters and symbols (=, ", >): one-byte / Strings of wireless network connecting:
two-bite / Space: one-byte.
*If “GW-USDual300 MAC0” is wireless connection 2, input the “wireless connection 2” characters
into the text.
*In order to confirm that the batch file (boot.bat&shutdown.bat) created above is correct, start the
command prompt and make sure that an error does not occur when you execute drag & drop ⇒
Enter. (Batch file confirmation time: Approximately 150 sec)
(5) Under [Local computer policy] go to [User configuration], and under [Windows settings] select
[Script (logon/logoff)].
- 79 -
2. Installation procedures
(6) Double-click logon, and click “Add.”
- 80 -
2. Installation procedures
(8) Under [Local computer policy], go to [User configuration], and under [Windows settings] select
[Script (logon/logoff)].
- 81 -
2. Installation procedures
1) Start the service tool and, using the AP/LNK Configuration Tool, select the ESSID, PSK and
Channel settings on the Configuration tab.
Select the SSID, PSK (network key) and Channel settings established in the GW-US Dual 300 access
point for ESSID, PSK (network key) and Channel on the Configuration tab, which is accessed by
selecting Service tool > AP/LNK Configuration Tool, and then click the [Save] button. (There is no
need to click the Send button.)
- 82 -
2. Installation procedures
2) Set the Connect tab using the AP/LNK Configuration Tool, and click the [Save] button.
(There is no need to click the Send button.)
<Access Point>
• Local IP address: 192.168.101.1
(In this case, IP address is set to 192.168.101.1 for the built-in LAN port used for wired connection
with a detector.)
• Target IP address: Default
• Subnet mask: Default
<Link Module>
・Use Link Module:ON
・COM port:
* When using the ready indicator unit:
Select “Not set.” (The COM port setting need not be selected.)
* When using the infrared data communication unit
Select the COM port to which the infrared data communication unit is connected.
4) In the case of a wired connection with the detector using the PC connecting cable, select “Detector”
under “X-ray generator/detector settings” on the service tool menu, establish the link and select the
detector registered. Then set the IP address used for the wired connection from the [IP] tab. After
selecting the setting, establish the link again.
• Set 192.168.101.11 or 255.255.255.0 as the IP address used for the wired connection.
- 83 -
2. Installation procedures
7.8.5 Ethernet-USB adapter settings
This section describes the setting when the notebook PC is connected to the detector using the USB
connector and when the Ether-USB adapter is used for a wired connection to the main LAN.
Ethernet-USB adapter
• The adapter must support Gigabit Ethernet.
• The adapter is to be tested and its operation is to be verified by the sales companies, and then used.
- 84 -
2. Installation procedures
7.8.6 Setting for internal wireless LAN (for a wireless connection to the main
LAN)
The following setting is required only for a wireless connection between the notebook PC and
the main LAN. The settings in this section is not necessary when a wireless connection between
the notebook PC and the main LAN is not used.
1) From Start menu, select “Control Panel”, “System security” and “System” select “Device
manager”.
2) Select “Network adapter” and right click “Intel @ Centrino ® Advanced-N 6235” and click
properties.
3) Open “detailed setting” in “Intel® Centriono® Advanced-N 6235 properties” and select
“wireless mode” from the list in the properties.
4) From the pull down list displayed on the right side, select “802.11a/b/g”.
1) From Start menu, select “Control Panel”, and click “network and internet”.
5) Click the “add” button and select “Create the network profile manually (M)”.
6) Set the following items according to the main network of installation site.
Network name
Type of security
Type of encrypted
Security Key
- 85 -
2. Installation procedures
7.9 System connections in the non generator connection mode
(mobile X-ray system environment: outside Japan)
7.9.1 System connections (wireless and wired)
• Refer to the connection diagram for the non generator connection mode (mobile X-
ray system: wireless) in section 7.2.4.
• Refer to the connection diagram for the non generator connection mode (mobile X-
ray system: wired) in section 7.2.5.
The notebook PC with a wireless LAN incorporated is procured by the sales companies, and the system
is constructed as shown in the figure below. In the case of a wired connection, connect the detector to the
notebook PC using the PC connecting cable.
A system such as the one shown below can be constructed even in Japan, but it is not recommended
owing to the various restrictions which would apply.
Main LAN
Internal Internal
wireless LAN LAN port
Restrictions:
• Use is possible only in 2.4 GHz frequency areas.
• A wireless connection with the main LAN is not possible.
Canon Inc. has inspected the operation of the notebook PC with a wireless LAN incorporated made by
the manufacturer listed below and verified that this unit operates properly. Compared with the operation
with SoftAP (Planex), display of 1st preview will be delayed for about 1 second. There is no difference
in image cycle time.
Notebook PC with wireless LAN incorporated <<The built-in softAP can be used only in 2.4-GHz
band.>>
• Manufacturer: Panasonic
• Model: TOUGHBOOK CF-C2
- 86 -
2. Installation procedures
7.9.2 Restrictions
1) In cases where patients may touch the notebook PC or the detector which is connected to the
notebook PC using the PC connecting cable, the AC adapter, main network inside the hospital or
clinic or any devices requiring an external power source must not be connected to the notebook PC.
2) Users in some areas cannot buy the USB wireless LAN adapter whose operation was verified by Inc.
They must use the softAP built in the notebook PC. However, usually the built-in softAP can be used
only in 2.4-GHz band. (It cannot be used in 5-GHz band.)
3) The wireless LAN incorporated inside the notebook PC cannot be used for the wireless connection
with the main LAN because it is used by the wireless connection with the detector. This means that a
wired connection using the internal LAN port is used as the connection with the main LAN.
4) In the case of a wired connection with the detector, an Ethernet-USB adapter is required.
7.9.3 Settings
1) Disabling softAP(intel)
(1) From the Start menu, open the Control Panel.
(2) From the Control Panel, open Program and then click Uninstall Program.
(3) Select “Intel®Proset/Wireless Wifi software”.
(4) In the Intel®Proset/Wireless Wifi software setup window, select “Change(C)” from
“Change, restore or delete the installation”.
(5) From the intel®Proset/Wireless Wifi software setup window”, disable “intel®MyWifi
technology” from “Custom Setup”. (x mark)
(6) Press “Next” and complete the change of installation.
2) Settings to start softAP (Microsoft) and stop the wireless slave unit search function.
If the wireless slave unit search function is not stopped, there is a possibility that the
connection with detector unit is disconnected. The wireless slave unit search function must
be stopped by following the steps below.
(1) From the “Control Panel”, open “View network status and task”, select “Change the
adapter setting”.
(2) Check the wireless network connection number of the device named“Intel ® Centriono
® Advanced-N 6235”.
(The procedures are on the assumption of wireless network connection 3.)
(3) Under the C drive, create a script folder.
(4) Start text editor (such as notepad), enter the following strings by the keyboard, name
the file name boot.bat and save it under script folder.
netsh wlan set autoconfig enabled=no interface= “Wireless Network Connection 3”
* For the wireless network connection, enter the connection number of “Intel ® Centriono ®
Advanced-N 6235” you checked in (2) above. (If there is no number at the end, leave it blank.)
* For the area of the strings in red, specify SSID and PSK you want to specify.
(Restrictions: For SSID, from 1 character to 32 characters. For PSK, from 8 characters to 63
characters)
* Alphanumeric characters and symbols (=, ", >): one-byte / Strings of wireless network connecting:
two-bite. (If English OS is used, enter “Wireless Network Connection” in one-bite characters.)
- 87 -
2. Installation procedures
(5) To check the created batch file (boot.bat) is correct, start the command prompt by the
administration right, drag & drop the boot.bat, execute Enter and make sure an error
does not occur.
(6) From Start, open “Search program and file” and enter “gpedit.msc”.
(7) In the “Local computer policy” , “User configuration” and “Windows setting”, select
“Script (logon/logoff)” and double click “logon”.
- 88 -
2. Installation procedures
(1) From “Control Panel”, “View network status and task”, select “change the adapter
setting”.
(2) For the IP address of the device named “Microsoft Virtual WIFI Miniport…”, set the
IP address.
*There is no restrictions for the IP address, however, unless specified, set 192.168.100.10
/255.255.255.0 for the wireless IP address of softAP (Microsoft).
Wireless IP address for the softAP (Microsoft) must be set to wireless connection of the
“Microsoft Virtual WIFI Miniport…”. It must not set to the local area connection or
other wireless network.
4) Other setting
(2) In the case of connecting a built-in LAN port to the main LAN set up the built-in LAN
port of a notebook PC. For details, refer to “7.8.3.2 Settings for the internal LAN port
in notebook PC (TOUCHBOOK CF-C2).”
- 89 -
2. Installation procedures
(1) The Ethernet-USB adapter is required in the case of a wired connection with a
notebook PC and the detector using the PC connecting cable so refer to “7.8.4
Ethernet-USB adapter settings.”
Set the IP address used for the wired connection to 192.168.101.1 / 255.255.255.0.
(2) At the same time, select “Detector“ under “X-ray generator/detector settings” on the
service tool menu, establish the link and select the detector registered. Then set the IP
address used for the wired connection from the [IP] tab. After selecting the setting,
establish the link again.
Set the IP address used for the wired connection to 192.168.101.11 / 255.255.255.0.
- 90 -
2. Installation procedures
We describe the setup steps of the USB wireless LAN adapter (WLAN-LCUSB2458:L-com) which
can be procured in Europe and USA. Users in Europe and USA can procure this USB wireless LAN
adapter and build a system like the following figures: [installation case 1] or [installation case 2].
In addition, if this USB wireless LAN adapter is procurable outside Europe and USA, then this system
is applicable.
[Installation case 1] When a detector and a notebook PC are connected in wireless condition.
Main LAN
Mobile X-ray System
Internal
Wireless
LAN
Internal
LAN Port
Limitation:
You cannot change (1) network address.
(WLAN-LCUSB2458 sets up address automatically.)
The Inc operation verification ensures the normal operation of the wireless LAN notebook PC that is built
in a main part, and the USB wireless LAN adapter of the following maker.
♦ a wireless LAN notebook PC that is built in a main part << the built-in softAP can be used only in 2.4
GHz band >>
- Maker : Panasonic
- Part number : TOUGHBOOK CF-C2
♦ USB wireless LAN adapter <<Useable in 2.4 GHz band & 5 GHz band>>
- Maker : L-com
- Part number : WLAN-LCUSB2458
For the main LAN connection, you must choose the wireless connection using a built-in wireless LAN or
the wired connection (Ether connection) via a built-in LAN port, depending on the environment of an
installation site. You can use an Ether-USB adapter instead of a built-in LAN port.
(For an Ether-USB adapter, refer to 7.8.5).
- 91 -
2. Installation procedures
[Installation case 2] When it corresponds to wireless connection & wired connection (A PC connecting
cable is used)
Internal
Wireless
LAN
Ether-USA adapter
IO-data ETG3-US2
Etc/
Limitation:
You cannot change (1) network address.
(WLAN-LCUSB2458 sets up address automatically.)
When you have set up a built-in LAN port to be used for the wired connection to the main LAN, you can
use also an Ether-USB adapter as a connection destination of a PC connecting cable.
7.9.5 Limitations
There are the following limitations about external device connection of the notebook PC mounted in
mobile type equipment.
1) When a patient may touch the notebook PC or the detector that is connected to the notebook PC with a
PC connecting cable, then the notebook PC must not be wired to an AC adapter or any devices that
require external power supply. Moreover, the notebook PC must not be wired to the trunk-line network in
a hospital, either.
2) The network address of a USB wireless LAN adapter (WLAN-LCUSB2458) shall be "Only
192.168.123.1 / 255.255.255.0 is allowable."
3) When connecting a USB wireless LAN adapter (WLAN-LCUSB2458) to a PC, you are recommended
to connect it to the port of "USB2.0."
* When the USB wireless LAN adapter (WLAN-LCUSB2458) was connected to the USB 3.0 port of
the PC once, the phenomenon that the communication with the detector becomes unstable (such
as the detector was not recognized, or the wireless connection was not restored automatically
after the USB wireless LAN adapter was removed and inserted.)
The phenomenon, however did not occur with the USB 3.0 port of another PC. In addition, the
phenomenon did not recur after that at all. As a result, we are not sure whether the problem was
caused by the USB 3.0 or not. However, considering the risk of recurrence of this phenomenon, if
you need to use the USB 3.0 as the connection port on site, please be very careful about the
behavior.
4) In order to perform wireless connection of a detector in 5 GHz band, you need a USB wireless LAN
adapter (WLAN-LCUSB2458).
5) When you have set up a built-in LAN port and an Ether-USB adapter individually, be careful not to
mix them.
(Do not connect the main LAN to the built-in LAN port that was set up to be connected to a detector.)
- 92 -
2. Installation procedures
7.9.6 Setting up of various factors
We describe the procedure to use a wireless LAN notebook PC built in a main part and a USB wireless
LAN adapter, as an access point, and to enable communicate with a sensor.
In addition, these setup procedures address the [installation case 2] mentioned above.
For [the installation case 1], refer to only the portion required.
1) After a PC has been started, a USB wireless LAN adapter shall be connected to the port of USB2.0 at
the CF-C2 back.
USB2.0(PC back)
USB3.0(PC side)
2) Shortly after connection of a USB wireless LAN adapter, the message is displayed to indicate of failure
of installation of driver. Then, click the Close button.
- 93 -
2. Installation procedures
3) Insert a CD that is attached to a USB wireless LAN adapter, and execute "\windows\setup.exe." A
setup type screen is displayed. Choose "an installation driver and a Ralink WLAN LAN utility". Click
"next" and the screen will change to an installation screen. Click "installation" and start installation.
1) AP mode
This setup uses a USB wireless LAN adapter as a main unit of an access point.
(Cordless handset it is switched to a main unit)
Right-click the Ralink Wireless icon in a red circle of a system tray. Choose "Switch to AP mode" in pop-
up.
Right-click
- 94 -
2. Installation procedures
Switching to AP mode will change a Ralink Wireless icon to the following figure.
When the share setting screen of the network is displayed as shown below, then cancel it by upper right x.
Put a check.
- 95 -
2. Installation procedures
Choose a region code in accordance with the law about the electric wave of a country or a region.
Choose and set up a channel with little radio wave interference. When using channel bonding, put a check
in "Enable 40 MHz Bandwidth".
- 96 -
2. Installation procedures
4) Advance setup
Click the icon in a red circle of a Ralink screen, and display a setting screen, and apply the following
setup.
- Between wireless network cards, a packet cannot pass. Remove a check.
- Beacon interval : 100 milliseconds
- Transfer power : 100%
- Idle time : 300 seconds
- 97 -
2. Installation procedures
5) IP address
After completing a setup, choose "a change of an adapter setup” from "a control panel > network status
and task display.” Then, the following screen will be displayed.
However, a USB wireless LAN adapter uses the IP address "192.168.123.1" automatically. So, you
must not change the IP address setting on the following connection properties.
It is set up automatically.
[Remarks]
- 98 -
2. Installation procedures
In order to start SoftAP in CCS shell mode, perform start-up processing in the following procedures.
2) Start text editors (notepad etc.) and enter the following characters by a keyboard. After that, give it the
name, boot.bat. Store it in a script folder. (Create a .bat file.)
* The above-mentioned batch file (boot.bat) created must be correct. In order to confirm its correctness,
a command prompt shall be started by administrator authority. Drag and drop a batch file (boot.bat)
=>Enter. Confirm that it does not cause any error.
- 99 -
2. Installation procedures
4) [Local Computer Policy] ― [User Configuration] ― Choose [Script (logon/logoff)] in [Windows
Setting].
- 100 -
2. Installation procedures
1) Start a service tool and use AP LNK Configuration Tool. Set up ESSID, PSK, and Channel range of a
Configuration tab.
SSID, PSK (network key), Region code and Channel were set up in 7.9.6.2 Setup of various
factors. They shall be set up to a service tool >AP/LNK Configuration Tool>Configuration tab,
ESSID, PSK (network key), and Channel. Push the "Save" button. (You need not depress the
Send button.)
- 101 -
2. Installation procedures
2) Set up a Connect tab in AP LNK Configuration Tool, and push the "Save" button.
(You need not depress the Send button.)
<Access Point>
- Local IP address:192.168.101.1
(When the IP address of the built-in LAN port used by a detector and wired connection is set to
192.168.101.1)
- Target IP address : default
- Subnet mask : default
<Link Module>
- Use Link Module: ON
- COM port:
* When a ready indicator unit is used:
=> Choose “Not set”.(You need not set up COM port.)
* When an infrared ray data communication unit is used:
=> Choose the COM port to which infrared-ray-communication unit is connected.
- 102 -
2. Installation procedures
6) When a PC connecting cable is used for wired connection with a detector, then a service tool menu X-
rays generator/sensor setup > From a sensor, choose a LINK registered detector. From [IP] tab, set up the
Wired IP address. After setup, perform the infrared rays LINK again.
- Local IP address : 192.168.101.1
(When the IP address of the built-in LAN port used by a detector and wired connection is set to
192.168.101.1)
- Target IP address : 192.168.101.11
- 103 -
2. Installation procedures
Set up the Ether-USB adapter. This adapter is used when the USB connection connector of a notebook PC
is used for wired connection with a detector, or when it is used for wired connection with the main LAN.
* For details, refer to 7.8.5 Ether-USB adapter settings.
7.9.6.6 Setting up built-in wireless LAN (when wireless connection is performed to the
main LAN)
Only when wireless connection is performed between a notebook PC and the main LAN, then the
following setup shall be performed. When wireless connection is not performed between a notebook PC
and the main LAN, then you need not perform this item.
* For details, refer to 7.8.6 Setting for internal wireless LAN (for a wireless connection to a Main
LAN).
- 104 -
2. Installation procedures
7.10 Using the detector in multiple radiography rooms
Only in an environment in which the service technician has installed this product (701 series detector) in
a multiple number of radiography rooms can the user carry the detector into each of these rooms and use
it. For this to happen, a link (recognition/connection) must have been established with the Image
Capture Computer installed in each radiography room and the communication line must have been
established ahead of time. (At the same time, the settings essential for CXDI Control Software NE must
also be selected.)
* When the detector is to be taken from X-ray room 1 into X-ray room 2 and used in
room 2, first its power must be turned off and then turned back on, then the link
(recognition/connection) must be established for the ready indicator unit in X-ray
room 2, and finally the settings selected for the detector must be changed. When the
detector is to be taken back to X-ray room 1 from X-ray room 2 and used in room 1,
first its power must be turned off and then turned back on, and then the link
(recognition/connection) must be established.
* The settings will be retained even when the power of the detector has been switched off. The
link (recognition/connection) need not be established when the detector is to be used by the
same system. However, the link (recognition/connection) will need to be established again if
the settings have been erased due to trouble of some kind.
Detector
Hub Hub
Operation Operation
room room
Image Capture Image Capture
Computer Computer
- 105 -
2. Installation procedures
This DR Wireless system (CXDI-701C/701G) can be used while existing DR Wireless systems (such as
the CXDI-70C/70G/80C) and other Ferry-compatible systems (401G/401C, 401G COMPACT/401C
COMPACT, 501G/501C), and existing CXDI systems (CXDI-40EG/40EC, 50G/50C, 40G COMPACT,
55G/55C and 60G/60C) are also being used. However, care must be taken since the following restrictions
apply.
Restrictions:
Since none of the existing models except for the CXDI-701 series of detectors support the non
generator connection mode, the non generator connection mode cannot be used in a system
environment where an existing model is connected. Consequently, use is restricted to operations in
the standard synchronization mode only.
CXDI Control Software NE Ver.2.10 and up must be used. Furthermore, check the software
compatibility list and, if required, update the firmware to the version which is compatible with CXDI
Control Software NE Ver.2.10 or up.
[Installation flow: When using the detector with an existing CXDI system]
(1) Put together the existing CXDI system (example: CXDI-50G system).
(2) Connect the CXDI-50G system to the Image Capture Computer via the switching hub.
(3) Turn on the power of the Image Capture Computer, and install CXDI Control Software NE.
(4) Start up the service tool, and click the [ENROLL] button.
(5) Check that the serial number of the 50G detector is displayed on the Connection tab.
(6) Check the firmware of the 50G detector and, if required, update it.
(7) Put together the DR Wireless system, and establish the link (recognition/connection).
(8) Check that the serial number of the Wireless detector is displayed on the Connection tab.
- 106 -
2. Installation procedures
RX_REQ
X-ray exposure signal (SW level 2 ON)
Is imaging possible?
Generator
time setting
RX_COM X-ray exposure
permission signal ON
X-ray exposure
X-ray exposure signal OFF permission signal OFF
- 107 -
2. Installation procedures
● For normal imaging
TPRE_DELAY
< 100 ms
X-ray exposure
signal (RX_REQ)
Timeout: Normally 1100 ms
TPREP Long periods of time: Up to 3100 ms
< 300 ms
X-ray release signal
(RX_COM)
TPRE_DELAY TPOST_DELAY
< 100 ms < 100 ms
- 108 -
2. Installation procedures
7.12.2 Signals and their functions for connection with the X-ray generator
- 109 -
2. Installation procedures
Connection conditions
1. The X-ray exposure signal line (including the switching mechanisms) must be insulated, and the
total impedance must be less than 100 Ω. The device delay time must be 1 ms or less.
2. The maximum contact voltages of the X-ray exposure permission signal line must be AC 250 V
and DC 30 V whereas the current must be 10 mA to 2 A.
Furthermore, only an insulated secondary power supply can be connected.
3. The protective ground of the X-ray generator must have the same potential as the system itself. The
device delay time must be 20 ms or less.
- 110 -
2. Installation procedures
(3) Performance
Item Description
Operation time Max. 20 ms
Reset time Max. 10 ms
Maximum Mechanical 18,000 times/hour
opening/closin
g frequency Rated load 1,800 times/hour
Between coil contacts: AC 2,000 V for 1 minute
Withstand voltage Between same pole contacts: AC 1,000 V for 1
minute
Mechanical life More than 50 million times
Service life Electrical life More than 100,000 times (5 A, 250 V, AC)
(resistive load) More than 500,000 times (5 A, 30 V, DC)
–40 to +70 ºC
Operating ambient temperature (However, there must be no formation of ice or
condensation.)
Maximum operation frequency 50 times/second
- 111 -
2. Installation procedures
2) PC1 (photocoupler)
(Note 1)
Pulse forward current IFP 1A
DC reverse voltage VR 5V
Voltage from collector to VCEO 80 V
Light-sensing
Voltage emitter
from emitter to VECO 7V
side
collector
Collector current IC 50 mA
Collector loss (1 circuit) PC 150 mW
Allowable loss (1 circuit) PT 200 mW
Dielectric strength (Note 2) BVs 2500 Vms
Operating temperature Topr -55 to 100 ℃
Storage temperature Tstg -55 to 125 ℃
Note 1
Pulse width: less than 100 µs; frequency: 100 Hz
Note 2
The AC voltage is applied to all the LED pins and all light-sensing pins, respectively, for 1 minute at a
relative humidity of ≦60 %.
- 112 -
2. Installation procedures
Breakdown voltage
from emitter to V(BR)ECO IE=0.1 mA 7 - - V
collector
VCE=48 V 0.01 0.1
-
Luminance (100 lux) (2) (10)
Dark current ICEO µA
VCE=48 V, Ta=85 °C 2 50
-
Luminance (100 lux) (4) (50)
Rise time tr - 2 -
VCC=10 V
Fall tim tf - 3 -
characteristics
IC=2 mA µs
Turn-on time tON - 3 -
Transfer
RL=100 Ω
Turn-off time tOff - 3 -
Turn-on time tON VCC=5 V - 2 -
Storage time tS IF=16 mA - 25 - µs
Turn-off time tOFF RL=1.9 kΩ - 40 -
Note 3: The structure is such that the leakage current may increase as a result of extraneous
light.
- 113 -
2. Installation procedures
The various settings are established using CXDI Control Software NE. For details on these
settings, refer to the “CXDI Control Software NE Service Manual.”
[Reference information]
(1) IP addresses of image capture computer (Local IP addresses)
Item Setting value
Local IP address 192.168.100.10
Subnet mask 255.255.255.0
(2) IP address (target IP address) allocated to the 701 series detector (PCA-FE board)
Item Setting value
701 series detector IP address 192.168.100.11
192.168.100.12– (digits in fourth column
701 series detector IP address
changed)
Subnet mask 255.255.255.0
* When the link (recognition/connection) has been established for the first time,
192.168.100.11 is allocated automatically as the address.
* When using wireless and wired connections with one detector, allocate different IP
addresses.
* When using two detectors in the same room, allocate different IP addresses.
(3) IP addresses registered in X-ray interface box (PCA-XIF board) (factory defaults)
Item Default
X-ray interface box IP address 192.168.100.96
Subnet mask 255.255.255.0
(6) Settings
A number of other settings are required depending on the configuration of the equipment at
the location where the system is installed and on the requirements of the user.
• User settings
• Protocol creation
• RIS, PACS, printer connection settings
• Other necessary items
- 114 -
2. Installation procedures
(1) Objective
To check the final quality of the images.
Step/item Description/conditions
Center line at
right angle
X-ray tube
focal point
Detector
- 115 -
2. Installation procedures
Step/item Description/condition
Image output
5. Output the captured images to the Dicom storage or
Dicom printer.
Image quality
checks
* Concerning calibration
Calibration by the users is prohibited when the non generator connection setting has been
enabled. This means that at any site where non generator connection operations are performed,
calibration must be carried out by the service technician while referring to section 7.6 on
calibration during non generator connection operations.
- 116 -
2. Installation procedures
- 117 -
2. Installation procedures
Positioning rib
(Volume control
button)
Hook and loop fastener
- 118 -
2. Installation procedures
3) After securing the ready indicator unit, check the following points.
*It must be possible to operate the volume control button from the operator’s side.
*It must be possible to check the infrared communication sensor area (IrDA window) and
LED from the operator’s side.
(Infrared communication
area)
(Volume
button)
(Notebook PC monitor)
[A wall or pillar: When the positioning of the unit relative to the Image Capture
Computer is virtually constant as in a regular radiography room]
1) Place the edge of the fastener loop up against the positioning rib on the back of the ready
indicator unit (where the speaker holes are located), and adhere it in place.
(In the figure below, the positioning rib is highlighted in orange for the sake of
convenience, but the rib itself is actually white.)
Positioning rib
(Speaker holes)
2) Adhere the fastener hook to a wall or pillar where the unit is to be secured.
*The fastener loop and fastener hook can be interchanged between the ready indicator
and the wall/a column. However, use the fastener hook and fastener loop as a pair.
3) Adhere the fastener loop on the ready indicator unit and the fastener hook on the wall or
pillar together, and anchor the ready indicator unit.
4) After securing the ready indicator unit, check the following points.
*It must be possible to operate the volume control button from the operator’s side.
*It must be possible to check the infrared communication sensor area (IrDA window)
and LED from the operator’s side.
- 119 -
2. Installation procedures
- 120 -
2. Installation Procedures
8. Outline drawings
1) 701G/701C detector
Connector cap
- 121 -
2. Installation Procedures
Weight: 1.0 kg
Dimensional unit: mm
Dimensional tolerance: ±1
(Scale 1: N)
- 122 -
2. Installation Procedures
3) Battery charger
- 123 -
2. Installation Procedures
4) Battery
Weight: 0.25 kg
Dimensional unit: mm
Dimensional tolerance: ±1
(Scale 1: N)
- 124 -
2. Installation Procedures
- 125 -
2. Installation Procedures
Weight: 0.27 kg
Dimensional unit: mm
Dimensional tolerance: ±1
(Scale 1: N)
- 126 -
2. Installation Procedures
- 127 -
2. Installation Procedures
Weight: 0.3 kg
Dimensional unit: mm
Dimensional tolerance: ±1
(Scale 1: N)
- 128 -
Printed by Canon Inc.
April 2013 Rev.01
CXDI-701G/701C
Wireless
3. Description of
Functions
CONTENTS
1. Detector .......................................................................................................... 1
1.1 Wiring diagram of detector ...................................................................... 1
1.2 Loading and outputting the data of the images taken ............................... 2
1.3 Detector .................................................................................................... 3
1.3.1 PCA-FE-M255 board ...................................................................... 3
1.3.2 FPC-UIF board ................................................................................ 3
1.3.3 PCA-WM-DB board ........................................................................ 4
1.3.4 PCA-CN-M255 board...................................................................... 4
1.3.5 PCA-DCDC-M255 board ................................................................ 4
1.3.6 PCA-BAT board .............................................................................. 4
1.3.7 PCA-REF-M255 board ..................................................................... 4
2. X-ray interface box......................................................................................... 5
2.1 Wiring diagram of X-ray interface box .................................................... 5
2.2 X-ray interface box................................................................................... 6
2.2.1 PCA-XIF board ................................................................................ 7
3. Wiring unit (option) ........................................................................................ 8
3.1 Wiring diagram of wiring unit................................................................... 8
3.2 Wiring unit ................................................................................................ 9
3.2.1 PCA-PB-M255 board ..................................................................... 10
3.2.2 AC/DC power supply ..................................................................... 10
3.2.3 Power cable assembly ..................................................................... 10
3.2.4 Detector cable ................................................................................. 10
3.2.5 Status indicator unit ........................................................................ 12
4. Grid attachment unit (optional) .................................................................... 13
4.1 Overview ................................................................................................. 13
4.2 Grid attachment (essential)...................................................................... 13
4.2.1 Precautions ..................................................................................... 13
4.2.2 Recommendations for when attaching the grid .............................. 13
4.3 Exterior .................................................................................................... 14
5. Ready indicator unit (option) ....................................................................... 15
5.1 Overview ................................................................................................ 15
5.2 Functions ................................................................................................ 15
5.3 Exterior ··································································································· 15
6. Handle unit (option) ..................................................................................... 16
6.1 Overview ................................................................................................ 16
3. Description of Functions
1. Detector
1.1 Wiring diagram of detector
PCA-D-M255
Item name
-1-
3 Description of Functions
-2-
3. Description of Functions
1.3 Detector
The detector comes in the form of a flat panel detector (FPD) for obtaining the X-ray images.
The images obtained are transferred to the image capture computer using Gigabit Ethernet by
wireless or wired means. The X-rays are first converted into visible light using a fluorescent
screen, and then stored electrically in the detector. The stored electrical signals are then loaded
from the detector, and A/D conversion is performed.
The power switch provided on the detector can be used to perform the operations of power
ON/OFF and infrared communication so that the detector is recognized by the image capture
computer. The functions concerned with the detector power are described below.
(2) Power ON
When the main power control circuit detects that the power switch of the detector has been
pushed down all the way (tap the switch for infrared communication; push it down all the way
to turn on the power) for one second, power begins to be supplied to the other circuits, and the
system operation mode is established.
-3-
3. Description of Functions
the shutdown process is initiated, after which the power is turned off.
The power OFF function described above works both when power is supplied by the battery and
when power is supplied by the wiring unit.
-4-
3. Description of Functions
To X-ray generator
Image
Capture
Switching hub
Computer
PCA-XIF
-5-
3. Description of Functions
The X-ray interface box monitors the statuses of the X-ray generator and detector to control the
X-ray exposure and imaging. It consists of the PCA-XIF board, AC adapter, X-ray interface
cable, grounding wire and covers. Even when a multiple number of detectors are in use, only
one detector per X-ray interface box can be specified from the image capture computer to take
images.
Exposure switch
Hub/wireless AP
Image
Capture
X-ray
Computer
TCP/IP connection
interface box Detector selection
Imaging condition
setting
Imaging data transfer
Wired LAN
*Either a wireless or wired format can be used for the LAN connection of the detector.
In the imaging sequence described below, the items printed in bold represent the control
operations performed by the X-ray interface box.
(1) The user presses the exposure switch.
(2) The exposure enable request signal from the X-ray generator is monitored by the X-ray
interface box.
(3) The X-ray interface box makes an Enable judgment for the exposure enable request
signal, and sends the trigger signal to the detector.
(4) When the imaging preparations of the detector are completed, the imaging enable signal is
sent from the detector to the X-ray interface box.
(5) The X-ray interface box monitors the imaging enable signal from the detector.
(6) The X-ray interface box makes an Enable judgment for the imaging enable signal, and
sends the exposure enable signal to the X-ray generator.
(7) With the detector now in the imaging status, X-ray exposure is commenced.
(8) When the detector has finished imaging, the images are transferred to the image capture
computer.
(9) This completes the imaging sequence.
-6-
3. Description of Functions
The PCA-XIF board is located inside the X-ray interface box, and has the main functions which
are described below.
(1) It has a function for interfacing with X-ray generator. In addition, the board employs a relay
circuit for the output signals and a photo-coupler for the input signals to protect the internal
circuitry.
(2) It has an exposure control function which monitors the statuses of the X-ray generator and
detector to determine whether exposure is possible and, if it is possible, to enable it.
(3) It has a function for interfacing with the detector.
(4) It uses LEDs to display the power supply status.
LED
* The dedicated AC adapter is used to supply power to the PCA-XIF board. It does not have a
power switch because the X-ray interface box and the AC adapter are supported to be used by
continuous energization. The PCA-XIF board must be turned on and off by connecting or
disconnecting the plug on the wall.
* For safety concerns, do not disconnect the cable from the X-ray interface box or disconnect
the AC adaptor. It would expose the electrically-activated part. To avoid this, please
disconnect the plug on the wall or turn off the commercial power supply.
* The use of table tap, which causes a change in electrical safety, is not allowed according to the
electrical safety standards.
* The X-ray interface box must be placed outside the patient environment.
-7-
3. Description of Functions
To switching
スイッチング HUBhub
へ
Status indicators
ステータスインジケータ
電源コードセット
Power cord set
Detector
センサケーブル cable
AC ソケットケーブル
PCA-PB-M255
AC socket cable
Detector
撮影部
電源ケーブルアセンブリ
Power cable assembly
AC/DCACDC
power
電源 supply
AC/DC box
-8-
3. Description of Functions
Power cable
assembly
PCA-PB
-9-
3. Description of Functions
The PCA-PB-M255 board is located inside the AC/DC box, and it has the connectors to which
the detector cable, LAN cable and status indicator unit are connected. Its main functions are
described below.
(1) It provides interfacing for supplying power from the AC/DC power supply (DC 12V) to the
detector.
(2) It provides interfacing for Ethernet communication (IEEE 802.3u standard) with the image
capture computer through the hub.
(3) It provides interfacing for LED control communication from the detector to the status
indicator.
The AC/DC power supply is located inside the AC/DC box, and it generates DC power from the
AC input. The DC 12 V power is supplied through the PCA-PB board for use by the detector
and status indicator unit. The board has a toggle switch for turning the power ON and OFF and
a blue LED to verify the power-on status.
Magnets
- 10 -
3. Description of Functions
By making the connector pin layout symmetrical at the left and right, it is possible to select
the direction in which to pull out the detector cable.
- 11 -
3. Description of Functions
When connected to the AC/DC box, this unit functions to provide the same displays as the LED
display on the detector. It contains the PCA-EXTS board in which are mounted the LEDs.
Blue/white
Blue
Red/white
Red
- 12 -
3. Description of Functions
4. Grid attachment unit (option)
4.1 Overview
This is used to position the grid against the detector, and it is used when placing the grid on the front
surface of the detector to eliminate the scattering radiation that occurs when X-rays pass through the
object of the photograph. This unit functions only by using it with the detector and each type of grid.
*The grid is a product that is procured by sales companies, so only the required grid specifications are
presented.
(Refer to 4.11 Grid specifications in 4. Specifications of Chapter 1. Overview.)
4.2.1 Precautions
• The grid must be used by attaching it to the grid frame unit.
• When the grid is to be adhered for use, it must be adhered using either of the adhesive tapes specified
below in such a way that the gap between the grid frame unit and grid edges is a uniform all the way
around. The total thichness of the adhesive souble-sided tapes and the grid must be less than 3.5mm.
(1.1mm < allowable total thickness < 3.5 mm).
• Ensure that the double-sided adhesive tape does not protrude from the edge of the grid attachment unit
when attaching.
• Clean the attachment surface so that no oil, etc. adheres to it.
• The grid is structurally weak, so be careful when handling it.
• The grid attachment unit is not guaranteed against dropping impact or overloading.
[Double-sided tape]
Width: 10 mm
Length: 443 mm±1.5 mm
[Double-sided tape]
Width: 10 mm
Length: 367 mm±1.5 mm
- 13 -
3. Description of Functions
Attach a grid by fitting the edge to the inside wall of hook for a detector.
4.3 Exterior
(1) External dimensions: 463 [mm] x 387 [mm]x 16.8 [mm] Tolerance;±1[mm]
(2) Weight: 0.3±0.1 [kg]
- 14 -
3. Description of Functions
5.1 Overview
The ready indicator unit is connected using a USB connector to the control PC, and it notifies
the operator of the imaging enable status using light and sound.
5.2 Functions
The main functions of this unit are described below.
(1) The operator is notified by light and sound that the detector is in the ready status.
(2) The operator is notified by light and sound of the time that the detector will remain in the
ready status and that this time has elapsed.
(3) The operator is notified by light that automatic X-ray detection has been achieved
successfully.
(4) The unit serves as an interface for linking to the system.
(5) The unit is connected to the PC through a micro USB cable purchased on the market.
(6) The volume level of the tones can be set to one of three levels using the volume button and
the sound can be muted using the mute switch.
For further details of the ready indicator displays and notification tones, refer to section 4.9
“Ready indicator display and notification tone specifications” in Chapter 1 Overview.
5.3 Exterior
(1) External dimensions: 35 ±0.2 (W) x 140 ±0.2 (D) x 20 ±0.5 (H) mm
(2) Weight: 55 g ±10%
- 15 -
3. Description of Functions
6.1 Overview
The handle unit used for this detector is not made commercially available.
For wireless operation, the handle unit used for the CXDI-70 series of Wireless detectors can be
installed; however, for wired operation, it cannot be installed because the shape of the connector
of the detector cable at the detector end differs from that used for the CXDI-70 series of
Wireless detectors.
Cable cover
グリップ
Handle ロック(操作部)
Lock lever ケーブルカバ ロックストッパ
Lock stopper
ロック(爪部)
部 Latch ー
リフター
Grid frame
lifter
Opening
開口部 (バッテリ)
(for changing the
battery)
Opening
開口部(無線) (for wireless communication)
Side cushioning
サイドクッション
引掛け爪部
Catch メインフレーム
Base frame Bottom cushioning
ボトムクッション
- 16 -
Printed by Canon Inc.
March 2013 Rev.01
CXDI-701G/701C
Wireless
4. Repair Guide
CONTENTS
1. Restrictions on disassembly and repair work ............................................................. 1
2. Precautions ................................................................................................................. 1
2.1 Repair and maintenance inspections .................................................................... 1
2.2 Removal of external covers .................................................................................. 1
2.3 Confirmation after PCB replacement ................................................................... 1
• Parts and areas not described in the repair guide must not be disassembled or repaired.
• The discrete parts (such as the electrical boards) described in the parts catalog must not be
disassembled or repaired.
2. Precautions
Before attempting to perform the repair work, the static charged in the bodies of the individuals
undertaking the work must be discharged, and only after this must the work be carried out.
Similarly, if the PCBs and cable connectors are to be touched when replacing the detector cable,
the static in the bodies of the individuals undertaking the work must be discharged, and only
after this must the work be carried out.
-1-
4. Repair Guide
3.1 Detector
In principle, the detector should be returned to Canon Inc. for repairs.
However, this does not apply to those sales companies whose representatives have received
service training in replacing the external covers of the detector and replacing its internal parts.
1) Remove the two screws (XB2-7300-606) on the bottom panel of the X-ray interface box, and
remove the six M3L6 brazier head screws (XA9-2033) on the side panels.
* When re-attaching the top cover, tighten up the screws in the following sequence:
bottom panel (screws with external teeth) → both side panels (brazier head screws).
Note
In actual fact, the X-ray interface
cable is connected to the PCA-XIF
board. This cable must be
disconnected after the top cover has
been removed.
XB2-7300-606
screws (x 2)
Note
The two M3L6 brazier head screws
(XA9-2033) at the bottom of the side
panels need not be removed since
they are provided for securing the
box to external mounting panels
when the box is to be mounted inside
a health screening van rather than for
securing the top cover.
Reference:
Tightening torque: 78.4 N·cm (8 kgf·cm) for side panel M3L6 brazier head screws
Tightening torque: 117.6 N·cm (12 kgf·cm) for bottom panel screws
2) Slide the top cover of the X-ray interface box toward the side where the connectors of the box
are connected, and lift to remove it.
* Do not incline the top cover too far since doing so may damage the inside of the X-ray
interface box.
-2-
4. Repair Guide
Top cover of X-ray interface box
Ensure that the protruding bent part (shown on
orange) at the bottom of the cover does not catch on
the base.
1) Remove the eight screws (XB1-2300-606) used to secure the PCA-XIF board.
XB1-2300-606 screws (x 8)
LED
AC adapter connector
LAN cable connector
2) In order to pull the LAN cable connector, AC adapter connector (DC inlet) and LED of the
PCA-XIF board out through the holes in the base, lift the PCA-XIF board in the direction
shown by the red arrows in the figure, and remove it.
PCA-XIF board
-3-
4. Repair Guide
1) Insert the LAN cable connector, AC adapter connector (DC inlet) and LED of the PCA-XIF
board through the holes in the base, and lay the PCA-XIF board flat.
* The shielded plate that covers the LAN cable connector is made electrically
conductive with the edge by pushing it into the hole in the base.
PCA-XIF board
2) Using the DC inlet position as reference, adjust the position of the PCA-XIF board.
* The end of the DC inlet must protrude by about 0.1 mm from the base wall.
DC
XB1-2300-606 screws (x 8)
Reference:
Tightening torque: 58.8 N·cm (6 kgf·cm)
-4-
4. Repair Guide
When replacing or removing the bottom surface PAD, attach the bottom surface PAD using the
specified screws (XB2-8301-006), and then coat the screw locking glue.
* Tightening torque: 78.4 N·cm (8 kgf·cm)
PAD×4
XB2-8301-006×4
* Tighten the screws of the bottom surface PAD at the specified torque, and then coat the screw
locking glue from the inside of the X-ray interface box base.
* Exercise caution as to prevent the screw locking glue from adhering to the PAD.
* Use the specified screw locking glue (Manufacturer: Threebond, Model name: 1401C or 1401B)
-5-
4. Repair Guide
Loosen the two screws (XB2-7300-606) on the rear panel of the AC/DC box, remove the four
M3L6 brazier head screws (XA9-2033) on the side panels, and then remove the top cover of the
AC/DC box.
* Take care not to deform the top cover which has been removed.
* When re-attaching the top cover, tighten up the screws in the following sequence:
connector panel (screws with external teeth) → both side panels (brazier head screws).
XB2-7300-606×2
XA9-2033×4
Reference:
Tightening torque: 78.4 N-cm (8 kgf-cm) for side panel M3L6 brazier head screws
Tightening torque: 117.6 N-cm (12 kgf-cm) for bottom panel screws
-6-
4. Repair Guide
Remove the four M3L6 brazier head hexagon socket bolts (XA9-2036) on the front panel of the
AC/DC box, and remove the cosmetic cover of the box.
* By removing the cosmetic cover, it is possible to access the screws used to make the
connection between the AC/DC power supply and AC/DC box base.
Cosmetic cover
* The screws are secured through two 3 mm spacers with a 3 mm diameter (XZ1-1300-
305) only on the left side as seen from the front of the cosmetic cover so take steps to
ensure that these parts are not lost or misplaced.
Cosmetic cover
XZ1-1300-305 x 2
(3 mm spacers with a 3 mm diameter)
Reference:
Tightening torque: 58.8 N-cm (6 kgf-cm)
-7-
4. Repair Guide
*When replacing the cosmetic cover, either the light pipe (XZ9-0695) must be re-used or a
new light pipe must be mounted on the cosmetic cover.
The light pipe has been inserted from the front of the cosmetic cover so it can be removed
by pushing it out from the back.
-8-
4. Repair Guide
1) Disconnect the AC socket cable connector from the AC/DC power supply.
2) Detach the power cable assembly that connects the AC/DC power supply and PCA-PB board,
disconnect the connectors at both ends of the cable, and remove the cable.
3) Remove the screw (XB1-2300-806) used to fasten both the cable clamp and the PCA-PB
board.
* Do not pull the cables free in order to disconnect the connectors.
Power cable assembly
View A
XB1-2300-806
XB1-2300-806(x 1) screw (x 1)
Cable clamp
AC socket cable
-9-
4. Repair Guide
* When mounting the AC socket cable, tie it using the cable tie from above the protective
tube at a position 10 ±3 mm from the connector with a view to preventing strain on the
cable arising from the connector becoming disconnected.
* When mounting the power cable assembly, fasten the cable to the cable clamp.
- 10 -
4. Repair Guide
PCA-PB board
XB1-2300-606×4
As a basic rule, mount the PCA-PB board by following the procedure for its removal in reverse.
Pass the detector connector part of the PCA-PB board through the hole in the base, and secure
the PCA-PB board using the four screws (XB1-2300-606).
Reference:
Tightening torque: 58.8 N-cm (6 kgf-cm)
- 11 -
4. Repair Guide
1) Remove the two screws (XB1-2300-606) on the front panel of the AC/DC box which are
used to secure the AC/DC power supply.
AC/DC powerACDC 電源
supply
XB1-2300-606 screws (x 2)
2) Remove the five screws (XB1-2300-606) used to secure the AC/DC power supply.
3) Slide the AC/DC power supply in the direction of the red arrow shown in the figure, draw the
switch part out from the hole in the base, and remove the power supply.
XB1-2300-606 screws (x 5)
- 12 -
4. Repair Guide
1) Slide the AC/DC power supply in the direction of the red arrow shown in the figure, and
insert the switch part into the hole in the base.
2) While checking the clearance between the hole in the base and area around the switch part,
adjust the position of the AC/DC power supply so that the clearance is uniform on the left and
right, and loosely secure it using the five screws (XB1-2300-606).
XB1-2300-606 screws (x 5)
Loosely tighten these screws.
AC/DC
Insert the switch part into power supply
the hole in the base.
3) Attach the two screws (XB1-2300-606) on the front panel of the AC/DC box used to secure
the AC/DC power supply.
XB1-2300-606 screws (x 2)
Tighten up only the screws on the
far left and far right.
Reference:
Tightening torque: 58.8 N-cm (6 kgf-cm)
- 13 -
4. Repair Guide
3.3.6 Replacing the B-type device mark label
When the B-type device mark label (BA5-3401-000) is to be replaced, adhere it at the position
which is designated in the figure below.
Position at which label is to be
adhered
Ensure that the mark is oriented
and angled correctly.
(Angle deviation: ±10° or less)
When the bottom panel pads have been replaced or removed, use the specified screws (XB2-
8301-006) to secure the pads, and then apply a thread-locking adhesive.
Tightening torque: 78.4 N-cm (8 kgf-cm)
PAD×4 XB2-8301-006×4
* After tightening the bottom panel pad screws using the specified torque, the thread-locking
adhesive must be applied from the inside of the AC/DC box base.
* Take care not to apply any of the thread-locking adhesive to the pads.
* The designated product must be used as the thread-locking adhesive (product name: 1401C or
1401B made by ThreeBond Co., Ltd.)
- 14 -
4. Repair Guide
Operation now starts up using the backup firmware and backup FPGA. When the power is
turned on with the boot switch at the backup mode setting, the backup code firmware is loaded
into the SDRAM, and operations are performed. With the backup firmware, the FPGA config
data for backup is written directly to the FPGA internally, and operation starts up.
NOTE: During startup from the backup code, the LINK LED (blue-green) blinks, its
brightness changing gradually. The LED stops blinking when startup is complete, so it
is necessary to wait at least 30 seconds.
The firmware or FPGA updating can be performed using the service tool while in the backup
mode boot status.
- 15 -
4. Repair Guide
After updating has been completed in the backup mode boot status, follow the steps below for
booting in the normal mode.
1) Hold down the power switch of the detector for at least five seconds to turn the unit’s power
OFF.
2) With a wired connection using the wiring unit, turn off the power of the AC/DC box.
3) Remove the battery.
4) Return the boot switch to the normal mode setting.
5) Connect a battery with a full charge again.
6) With a wired connection using the wiring unit, turn on the power of the AC/DC box.
7) Press the power switch of the detector for at least a second to turn on the unit’s power.
When the power is turned on with the boot switch at the normal mode setting, the normal code
firmware is loaded into the SDRAM, and operations are performed.
* When booting the normal mode after the backup mode booting, check the versions of the
firmware and FPGA from the Ferry type FPD maintenance tool and perform update to
have the right combination of them.
Boost switch
Normal mode
Backup mode
Battery holder
- 16 -
4. Repair Guide
5.1.1 Introduction
The PCA-XIF firmware incorporates an http server to enable its management and control to be
operated via a web browser. This section described the operating procedure for updating the
firmware from the web browser.
The PCA-XIF board has two code areas—one for the normal codes and one for backup code—
as areas where the firmware is stored. The backup codes are positioned at the head of the ROM
while the normal codes are stored in an intermediate area of the ROM. When the power is
turned on, the backup code initializing routine boots, and the backup code body boots when the
normal codes have been erased for some reason or when the BOOTSEL switch on the board has
been pressed. Unless other specific action is taken, this routine boots the normal code body.
For regular updating, operation is performed in such a way that only the normal codes are
updated. This has the effect of preventing the board from failing to boot since the backup codes
at the ROM head are preserved even if a power failure should occur during updating, for
example. The only time when the backup codes are updated is in special situations such as when
a fatal bug is found in the backup codes.
This section describes the procedure for updating both the backup codes and normal codes, but
unless otherwise specified, the backup codes should not be updated.
The PCA-XIF management screen can be accessed by inputting the IP address of the PCA-XIF
board in the URL field of the web browser. When the board is shipped from the manufacturing
plant, the IP address is 192.168.100.96 so this is the address which is normally input. When the
address has been changed from this default setting, input the new address.
If the IP address which has been set is no longer known, turn on the power while pressing the
BOOTSEL switch on the board (or press the RESET switch): The backup codes are forcibly
booted, and the IP address is set to 192.168.100.96.
* This operation is referred to as the “forced backup boot operation.” For details on the
actual operations and the precautions to be heeded, refer to “5.2.2 Forced backup boot
operation.”
When the screen has been accessed successfully by the web browser, the top page is displayed.
When the link on the page is clicked, the management and control pages are displayed. If, when
an attempt has been made to display a page using the web browser, a screen such as the one
shown below requesting that the password be input has appeared, input the user name and
password given below.
- 17 -
4. Repair Guide
When the “status view” link is clicked on the top page of the web browser, information on the
firmware version now operating, boot reason and network configuration settings are displayed.
(Commentary)
In the example of the screen display shown above, version 0.0.6 is written in both the backup
area (Flash #0) and normal area (Flash #1). When the power is turned on, the normal booting
operation is executed and the codes in the normal area are executed.
- 18 -
4. Repair Guide
When the “firmware update” link is clicked on the top page of the web browser, the screen on
which to perform the operations for updating the firmware appears.
To update the normal codes, do not make any changes in the “ROM Area” input field, and click
the “Browse” button to select the firmware file. Then click the UPDATE button.
* The file of the firmware used has the “.ehex” extension.
* After clicking the UPDATE button, it will take 10 to 20 seconds for the updating to be
completed.
When updating has been completed successfully, the “XIF firmware update was successfully
done” screen appears.
- 19 -
4. Repair Guide
5.1.5 Checking after updating that the updating has been reflected
Return to the top page of the web browser, click the “status view” link, check the firmware
version (running), and check that the version in the normal area (Flash #1) has been updated.
* At this stage, however, the original version of the firmware is still operating (the
“running” operation remains unchanged). Operation is started by the updated firmware
after the power is next turned on (or the RESET switch is pressed).
The procedure used to update the backup codes is virtually the same as the procedure for the
normal codes. To update the backup codes, except for inputting “flash0” (half-size lower-case
characters with no spaces) in the “ROM Area” input field, follow the same procedure as for the
normal codes. When the updating has been completed successfully, check that the version in the
backup area (Flash #0) has been updated.
* However, since the only time when the backup codes are updated is in special situations
such as when a fatal bug is found in the backup codes, the backup codes should not be
updated unless otherwise specified.
* Even when operation has been booted from either the backup area or normal area, the
firmware has the ability to rewrite the ROM data by itself. Therefore, if the firmware
ends up booting, it can be rewritten by communication between the PC and PCA-XIF
board without the use of a special tool (programmer). However, since a board whose
backup area is empty or whose data has been damaged cannot be booted, it will be
necessary to write the firmware using a programmer.
The Ethernet interface of the PCA-XIF board does not have a function for automatic
crossover/straight LAN cable adjustments (AUTO-MDI/MDI-X function) so the switching hub
with an AUTO-MDI/MDI-X function must be used between the PC and PCA-XIF board. If the
board is to be connected directly to the PC, a crossover cable must be used.
The default IP address of the PCA-XIF board is 192.168.100.96/255.255.255.0, and the IP
address of the PC used to communicate with the PCA-XIF board must have the IP address that
belongs to the same network. (Recommended address: 192.168.100.10/255.255.255.0)
Subsequent to Windows Vista, telnet which is the PC application for logging in to the PCA-XIF
board is not installed under the standard setup. However, a telnet connection with the PCA-XIF
can still be established by using TeraTerm whose installation is recommended using the
software (CXDI Control Software NE or CXDI Controller RF) setup.
When TeraTerm is to be used, first start up TeraTerm. Now, on the setting screen, set TCP/IP as the
connection method, the IP address of the PCA-XIF board as the host address, Telnet (change from the
default of SSH) as the Service, and 23 (default) as the port number. Then press OK to initiate the telnet
connection.
When the telnet client is to be used, select [Control Panel] → [Programs] → [Windows functions enable
or disable], and enable [telnet client]. Then start the command prompt, input the “IP address of the telnet
PCA-XIF board”, and establish the telnet connection.
The login prompt can be input simply by pressing return without entering the login name or password.
When XIF_GUEST> appears, login is completed. To log out, use the “bye” command.
- 20 -
4. Repair Guide
5.2.1 Introduction
It is possible for the firmware in the normal area to be erased or damaged if the updating has
failed because a communication error occurred while the normal area firmware was being
updated or because the power of the X-ray interface box was turned off.
If the normal area firmware has been erased, the backup code body can be booted by the backup
code initializing routine when the power is turned off and then back on so the firmware can be
updated again.
However, if the normal area firmware has been damaged, the backup code body will not be
booted simply by turning the power off and then back on, and it will be necessary to forcibly
boot the backup codes.
This section describes how to perform the forced backup booting operation and also the
precautions to be heeded.
If the normal area firmware has been damaged, the firmware in the backup area can be forcibly
booted by operating the switches on the board. The operations undertaken for forced backup
booting are described below.
1) Start pressing the RESET switch and BOOTSEL switch on the PCA-XIF board at the same
time.
2) With the BOOTSEL switch still held down, release the RESET switch first.
3) Wait for one second, and then release the BOOTSEL switch.
When the firmware has been booted as a result of taking the steps above, the IP address and
MAC address are set to the default values.
[Default settings]
IP address : 192.168.100.96
Subnet mask : 255.255.255.0
Default gateway : 0.0.0.0
MAC address : 00:1E:8F:CA:87:80
* Since the IP address and MAC address (separate values different from 00:1E:8F:CA:87:80)
written in the flash memory are displayed as the IP address and MAC address indicated in the
network settings on the Status View screen for the Web browser, the default values will not be
displayed even if forced backup start is implemented.
* Even after forced backup start has been implemented, the original IP address and separate
MAC address are stored on the flash memory so the original addresses will be restored when
normal start is initiated.
- 21 -
4. Repair Guide
Since a separate MAC address is written for the PCA-XIF board, care is required when
proceeding with forced backup booting.
* Since the operating system of the PC remembers the separate MAC address of the PCA-
XIF board, the PC tries communicating with the default MAC address of the PCA-XIF
board while the separate MAC address remains valid, and communication may fail. For
details on how to avoid this kind of situation, refer to “5.2.3 Concerning ARP clear for
the PC.”
On the PC, open the command prompt, and execute the following command:
The above arp command has the effect of clearing the separate MAC address of the PCA-XIF
board stored in the memory of the PC’s operating system and of re-initiating the inquiry. Unless
this step is taken, even though the MAC address of the board is switched to the default value by
the forced backup boot operation, the PC will still try communicating to the separate MAC
address used before the switch, and no further communication may be possible by the telnet
terminal or web browser.
* The operating system of the PC stores the IP address and MAC address correspondence
table in the memory, and reset is initiated when the PC is rebooted or there has been no
communication for a certain period of time. This means that it is not always necessary to
execute the ARP command. It is necessary only when communication is to be performed
immediately to a different MAC address using the same IP address. It may also be
necessary after the PCA-XIF board has been replaced when the IP address matches and
communication is to be tried immediately.
In a PC using Windows Vista or a later operating system, an error results when the above
command is simply executed even if the user is the administrator. To execute the command
correctly, the command prompt for executing the arp command must be booted in the “Run as
administrator” mode.
To boot the command prompt in the “Run as administrator” mode, proceed as follows: Select
[All Programs] → [Accessories] → right-click [Command Prompt] → then select “Run as
administrator” on the menu that appears.
- 22 -
4. Repair Guide
The IP address of the PCA-XIF board can be changed or checked using the “ipadr” command in the
telnet connection status.
(Commentary)
After “ipadr-ip 192.168.100.97” has been input, the “New setting takes effect after power cycling”
message is displayed. Before turning the power back on, input “ipadr,” and when the current status is
checked, “192.168.100.96” will be displayed for Current and “192.168.100.97” for the flash memory.
At this point in time, it is clear that only “192.168.100.97” has been written in the flash memory and that
the new setting has not been reflected.
- 23 -
4. Repair Guide
When the PCA-XIF board is now accessed from the Web browser and the details of the network
settings are checked, a display similar to the one shown below appears, and it is clear that the setting in
the flash memory has been changed, and the display now reflects this change.
Screen on which to check the change in the IP address using the web browser
The PCA-XIF board is provided with a 3-megabyte log memory. In situations where trouble has
occurred, for instance, the log of the PCA-XIF board can be obtained by clicking on the “log download”
link on the web browser top screen of the PCA-XIF board.
Note:
When the “log download” link is clicked, the “Open/Save” dialog box is displayed. If too much time is
now taken to input the storage location, only the initial part of the log details and not enough information
may be obtained. The time permitted for the input operation is about 30 seconds. If the amount of the
log information which has been saved is correct, the file size indicated on the “Open/Save” dialog box
and the size of the file actually saved will match. Therefore, when obtaining the log, it must be checked
whether these two file sizes match.
Reference:
When the log memory becomes full, the log information is successively erased starting with the oldest,
and the latest information is written in its place. This means that the log memory can always be expected
to be nearly full (2.8 to 3 megabytes) with the log of the PCA-XIF board log which is actually used.
- 24 -
4. Repair Guide
(2) When the X-ray interface box and PCA-XIF board have been replaced
• Change the IP address (but only when it has been changed from the default setting).
• Check the firmware versions (and update them if necessary).
• Check that booting is performed correctly.
• Check that the operations for taking and obtaining the images and for transferring the data can
be performed normally.
(3) When the AC/DC box unit, PCA-PB board and AC/DC power supply have been replaced
• Check that the power is supplied to the detector with the wired connection.
• Check that the battery installed in the detector can be charged with the wired connection.
• Check that CCS is booted correctly with the wired connection.
• Check that the operations for taking and obtaining the images and for transferring the data can
be performed normally with the wired connection.
- 25 -
Printed by Canon Inc.
April 2013 Rev.01
CXDI-701G/701C
Wireless
5. Parts Catalog
CONTENTS
1
Rev.01
3
2
2
Rev.01
1 6 2 3
4 5
3
Rev.01
17
4 3 6 7 8
5 2
9 12
13
14
10 11 16 15
4
Rev.01
1 2
5
Rev.01
2
3
6
Printed by Canon Inc.
April 2013 Rev.01
CXDI-701G/701C
Wireless
6. Troubleshooting
CONTENTS
1. Primary response
This work is done at the hospital or clinic by the service engineer of the sales company or medical
engineer (ME).
The work of checking out the detector involves checking the wireless communication, checking
the wiring connections and collecting the data*1 on the factors involved in the trouble that has
occurred. When it has been confirmed that the trouble is in the detector by the primary response,
the unit is exchanged for an detector owned by the sales company which is loaned out, and after
the data required has been collected*1, the detector with the trouble is sent to the sales company.
When trouble has occurred in the wired connection, the detector cable is replaced.
*1
For details on the data collected, refer to “4. Collecting data when trouble has occurred.”
2. Secondary response
This work is done at the office of the sales company.
The service engineer of the sales company checks on-site that there is something wrong with the
functions or performance of the detector based on checking whether the data collected at the time
of the primary response is sufficient and based on the work which has been carried out.
As soon as it has become clear to the sales company that the trouble is in the detector, details of
the detector concerned, the collected data and work carried out up to the point where it was
determined that the trouble that occurred was in the detector are entered on the inquiry sheet, and
this sheet is sent to Canon Inc.
However, only sales companies that have received repair training for the detector (internal parts &
external cover replacement) are allowed to repair the detector. If the detector is repaired in the
office of a sales company, refer to the separately-issued internal parts replacement
manual/external cover replacement manual.
3. Tertiary response
This work is done by Canon Inc.
On-site checks are carried out based on the data collected and work carried out up to and
including the secondary response.
Repairs are undertaken as soon as it has been confirmed that the trouble is in the detector.
In order to ensure accuracy in identifying in which unit the trouble occurred, what must be
done without fail if the trouble has been resolved by replacing the unit is to replace the new
unit with the original unit and verify that the trouble recurs.
-1-
6. Troubleshooting
1. Troubleshooting flow
NG
• Detector
• Collected data
• Details of work carried out
NG
• Detector
• Collected data
• Details of work carried out
-2-
6. Troubleshooting
2. Primary response
The work done at the hospital or clinic constitutes the primary response. Only service engineers or
medical engineers (ME) are allowed to do this work. When service engineers are to pay service
calls to hospitals or clinics, it is recommended that they take with them the tools listed below and
a detector which the sales company owns and which can be loaned out. What must be done
without fail if the trouble has been resolved by replacing the unit is to replace the new unit with
the original unit and verify that the trouble recurs.
• Notebook computer purchased on the market (incorporating the IEEE 802.11 a/b/g/n
specification): Used to check the wireless environment when trouble with wireless
communication has occurred
• Large storage media (at least 100 GB recommended): To save the collected data
• Detector
• Wiring unit (WU-3A)
• Ready indicator unit (or infrared data communication unit)
• Wireless LAN access point
Described below are the symptoms of the trouble which may conceivably occur at hospitals and
clinics and the suspected causes of the trouble.
Symptom (1):
The detector was recognized using infrared communication, but the three LEDs on the detector
failed to light (for about 1 second).
Suspected causes:
• Defective connection between the image capture computer and ready indicator unit
• Problem in work environment (the unit must not be located close to a monitor or installed in an
environment where signals are reflected)
• LNK Controller does not start up
• AP/LNK configuration tool is improperly set (settings for the COM port of the connect tab, etc.)
• Trouble with ready indicator unit (or infrared data communication unit)
• Trouble with FPC-UIF board (board inside detector)
• Trouble with PCA-WM-DB board (board inside detector)
Conduct checks based on the suspected causes listed above or replace the unit.
Symptom (2):
• The three LEDs of the detector lighted up (for about 1 second) with infrared communication and
the detector was recognized successfully, but the blue-green LED failed to light.
Suspected causes:
• The power of the access point has not been turned on.
• Incorrect setting of the access point using the AP/LNK configuration tool
-3-
6. Troubleshooting
Suspected causes:
• The X-ray imaging room is not tightly sealed (or the door to this room is open, etc.)
• The distance between the access point and detector is too great.
• A structure is blocking communication between the access point and detector.
• There is a thermal therapy device, laser knife, cordless telephone, etc. near the detector.
• There is interference with the signals (extraneous signals) of another wireless LAN device.*1
• DR Wireless SYSTEM are installed in two adjacent X-ray imaging rooms, and the images
taken are being transferred by both systems at the same time.*1
• Trouble with PCA-WM-DB board (board inside the detector)
• Trouble with access point*2
Conduct checks based on the suspected causes listed above or replace the unit.
*1
Boot the notebook computer (incorporating the IEEE 802.11 a/b/g/n specification), and check
whether the signals (extraneous signals) of another wireless LAN device are detected. If
another wireless LAN signal is detected, make channel settings based on the Channel setting
guidelines.*3
*2
If the wireless communication trouble has been resolved by replacing the access point, replace
the new unit with the original access point to verify that the trouble recurs, and if it does recur,
move on to the secondary response.
* If the wireless communication trouble is not resolved even when the channel setting has been
changed, the following information must be gathered without fail. It is recommended that this
information be recorded as the installation history.
• Wireless environment inside X-ray imaging room (existing wireless LAN environment)
• Signals of other wireless LAN devices detected by the notebook computer inside the X-ray
imaging room
• Surrounding environment (adjacent rooms, corridors, etc.)
• Whether or not DR Wireless SYSTEM are installed in adjacent rooms
• Information of settings established by AP/LNK configuration tool (including the setting
information of any DR Wireless SYSTEM installed in an adjacent room)
• Frequency at which the wireless communication trouble occurs
Reference: If there is interference with strong signals (extraneous signals), it is possible that
X-ray exposure cannot be performed because the RX_REQ signal from the X-ray
generator cannot be recognized.
-4-
6. Troubleshooting
Wireless LAN channel setting guidelines for adding the DR Wireless system to an environment
where another wireless LAN is already in use are described below. These guidelines apply to both
the 2.4 GHz and 5 GHz bands.
other.
HT40
(2) Use the same setting for the (2) Use the same setting for the primary
primary and secondary channels. channel as the channel setting of the other
(3) Use the same setting for the wireless LAN system.
primary channel only.*2
(1) Set channels that do not interfere (1) Set channels that do not interfere with each
with each other. other.
HT20
(2) Use the same setting as that of the (2) Use the same channel settings.
primary channel of the other
wireless LAN system.
*1
: In the 2.4 GHz band it is not possible to specify two noninterfering channels in HT40.
*2
: In the 5 GHz band the channel pairs are predetermined and there is no way to set the same
primary channel only. In contrast, in the 2.4 GHz band it is possible to specify either of two
channels for each primary channel, such as 6CH below 2CH or 6CH above 10CH.
Symptom (1): Dealing with system errors occurring while CCS is running
(1) Refer to the error guide in the CXDI Control Software NE Service Manual.
(2) Based on the comments in the cause column of the error guide, use the suspected causes as a
reference.
(3) Start up Collection Tool of Control Software NE_Service Tool.exe, and execute*4 file
collection.
*4
Refer to “4. Collecting information when trouble has occurred” and collect the information and
files based on the system error items.
If the error has recurred after the above action has been taken, replace the detector.
If the error is resolved after the detector has been replaced, replace the new unit with the original
unit, verify that the trouble recurs, and then send the detector to the office of the sales company
which will proceed with the secondary response.
If the error is not resolved even after the detector has been replaced, consider replacing the image
capture computer.
-5-
6. Troubleshooting
* When requesting a secondary response, the following information and data must be sent without
fail.
• Installation environment information (CCS version, PC specifications and hardware
configuration, wireless setting information, etc.)
• Files collected by Collection Tool
• Status of detector exterior
• Frequency at which the trouble occurs and report detailing the work which has been done
If the error has recurred after the above action has been taken, replace the detector.
If the error is resolved after the detector has been replaced, replace the new unit with the original
unit, verify that the trouble recurs, and then send the detector to the office of the sales company
which will proceed with the secondary response.
* When requesting a secondary response, the following information and data must be sent without
fail.
• Installation environment information (CCS version, PC specifications and hardware
configuration, wireless setting information, etc.)
• Files collected by Collection Tool
• Status of detector exterior
• Frequency at which the trouble occurs and report detailing the work which has been done
If the trouble has recurred after the above action has been taken, replace the detector.
If the trouble is resolved after the detector has been replaced, replace the new unit with the
original unit, verify that the trouble recurs, and then send the detector to the office of the sales
company which will proceed with the secondary response.
When abnormal image trouble has occurred, be absolutely sure to proceed with the imaging in
step (1), and collect the files resulting from this imaging.
* When requesting a secondary response, the following information and data must be sent without
fail.
• Installation environment information (CCS version, PC specifications and hardware
configuration, information on surrounding environment, etc.)
• Files collected by Collection Tool
• Status of detector exterior
• Frequency at which the trouble occurs and report detailing the work which has been done
-6-
6. Troubleshooting
Suspected causes:
• Wireless communication trouble
• Insufficient battery charge
It is recommended that a wired connection (using the equipment owned and loaned by the sales
company) be used for the firmware and FPGA updating. If updating is to be performed using a
wireless connection, proceed after ensuring that the battery has a full (100%) charge.
Suspected causes:
• Wireless communication trouble*6
• Incorrect X-ray interface IP address setting established by AP/LNK configuration tool
• Improper connection of X-ray interface cable
• Defective PCA-XIF board (inside the X-ray interface box)
• When X-ray exposure is accidentally performed during non generator connection mode.*7
Conduct checks based on the suspected causes listed above or replace the unit.
*6
If there is interference with signals of another wireless LAN device, it may be a case where the
RX_REQ signal from the X-ray generator cannot be recognized and X-ray exposure is disabled.
In such cases, use a wired connection with the wiring unit, and if X-ray exposure is enabled, it
means that trouble in the wireless communication is responsible for the problem.
*7
For details, refer to “7.5.3 Overview of non generator connection imaging” in Chapter 2.
Installation procedures.
-7-
6. Troubleshooting
3. Secondary response
This work (on-site checks, investigation and checks of collected data) which is done at the office
of the sales company constitutes the secondary response.
3.1 Detector
Check whether too much or too little data was collected at the primary response. If there is too
little data, ask for more data to be collected.
The sales company connects a system using its loan equipment (Wireless Detector, access point,
infrared communication unit and image capture computer), and checks whether the operation of
the loan equipment is problem-free. If it is problem-free, it replaces the detector (in which the
trouble occurred) whose problem was not resolved by the primary response with the detector in
the above system.
(1) Connect the detector (in which the trouble occurred) to the DR Wireless SYSTEM on loan.
(2) Using link operation, ensure that the detector is recognized.
(3) Check whether the trouble symptom (wireless communication trouble, error or abnormal
images) recurs.
If the symptom recurs, proceed with step (5).
(4) If the trouble symptom did not occur, consult with Canon Inc.
(5) If the trouble symptom recurred, send the detector to Canon Inc. (for the tertiary response).
* If the trouble does not occur frequently, perform tests to check that the trouble recurs at least
twice of the number of trouble occurrences.
* When requesting a tertiary response to errors or abnormal image trouble, the following
information and data must be sent without fail.
• Installation environment information (CCS version, PC specifications and hardware
configuration, wireless setting information, etc.)
• Files collected by Collection Tool (both at the primary response and secondary response stages)
• Status of detector exterior
• Frequency at which the trouble occurs
• Submission of inquiry sheet (report detailing the work carried out both at the primary response
and secondary response stages)
-8-
6. Troubleshooting
(1) Connect the access point (in which the trouble occurred) to the DR Wireless SYSTEM on
loan.
(2) Start the AP/LNK configuration tool, and set the access point.
(3) Using link operation, ensure that the detector is recognized.
(4) Check whether the wireless communication trouble which occurred where the system was
installed recurs.
(5) Make inquiries with CISCO on the CISCO support page.
-9-
6. Troubleshooting
◎: This indicates essential information and files; as a basic rule, all the files must be collected.
However, in the case of images, each image file is large. Therefore, as a basic rule, all the
trouble images in question and the 3 or 4 images immediately before and after the trouble
images must be collected.
In the case of an image processing error, as a basic rule, all the trouble images in question and
the images successfully processed in the same imaging areas (same imaging method) must be
collected.
- 10 -
6. Troubleshooting
○: This indicates essential files; all the files must be collected if this is possible.
If difficulties are posed in collecting all the files because the files are large, then three files
must be collected, namely, the file in the trouble neighborhood, the file at the time of
installation, and the latest file.
No.1
CCS version, PC specifications and hardware configuration, peripheral device connection
information
No.2
After Collection Tool has been started up and the [DB backup data] category has been selected,
one of the following 7 databases can be selected as the information to be obtained.
• STORE (storage output data)
• PROTOCOL (protocol data)
• PRINT (printer output data)
• MPPS (patient information acquisition/progress management notification data)
• LOG (CXDI Control Software NE operation logs)
• EXAM (examination data)
• CLIENT (log-on user data)
No.3
After Collection Tool has been started up and the [Exposure image] category has been selected,
the taken images to be collected can be selected using refined-search conditions.
Unlike data files, each image file is large. Therefore, as a basic rule, all the trouble images in
question and the 3 or 4 images immediately before and after the trouble images must be collected.
In the case of an image processing error, as a basic rule, all the trouble images in question and the
images successfully processed in the same imaging areas (same imaging method) must be
collected.
The successfully processed images are collected in order to check the exposure fields and the
differences in the subjects.
No.4
After Collection Tool has been started up and the [QC result] category has been selected, check
the required items on the QC history list to collect the calibration, performance test and
self-diagnosis files.
It is strongly recommended that the calibration, performance test and self-diagnosis imaging be
carried out in a state in which the image trouble has been reproduced.
If difficulties are posed in collecting all the files because the files are large, then three files must
be collected, namely, the file in the trouble neighborhood, the file at the time of installation, and
the latest file.
- 11 -
6. Troubleshooting
No.5
If Collection Tool is started up and the [Other Information] category is selected, the screwcap.log,
cxdcap.log and other log files can be collected.
No.6
If Collection Tool is started up and the [Other Information] category is selected, [SnsPBInfo] is
generated automatically, and the DP files and detector logs and X-ray interface box logs can be
collected.
No.7
If Collection Tool is started up and the [Other Information] category is selected, [elog] is
generated automatically, and the following four event viewer logs can be collected.
C:¥Windows¥System32¥winevt¥Logs
• Application
• Security
• Setup
• System
2) Sensor logs
• Sensor logs must be required when the system error, wireless communication failure and/or image
problem occurs.
• Start Ferry Type FPD Maintenance Tool, select Target IP Address of the detector of which logs you
are going to collect and press the Read Log button.
• Sensor logs are saved in the SnsPBInfo folder.
• For details on the procedures, please refer to CXDI Control Software NE Maintenance Manual.
To collect the logs of X-ray interface box and sensor logs by the Collection Tool, be sure to collect
them and save in the SnsPBInfo folder manually.
- 12 -
6. Troubleshooting
Calibration is a function that performs dose checking as well as generating and updating
correction data for adjusting the sensor in order to maintain normal sensor characteristics.
The performance test function exposes an X-ray image with no subject present and compares the
result with the calibration screen data to evaluate sensor performance and determine whether or
not it meets the setting standards.
The self-diagnostic test function can be used to perform a self-test of the sensor only, without
making an exposure using X-rays. It includes test functions such as internal device check, test
pattern check, and offset correction test. Note that the device check function is not available for
still image sensors without Ferry support (CXDI-55G, CXDI-55C, etc.); it is only available on
still image sensors with Ferry support (CXDI-70C, CXDI-80C, etc.).
Listed below are the precautions to be heeded during performance test and self-diagnosis
imaging.
[Performance test]
The performance test only performs Mode 1 (which corresponds to the former Test 1).
• This must be carried out after checking that the system is operating normally after installation.
• Performance test imaging must be implemented in order to check the conditions resulting after a
system has been newly installed and after the photo timer adjustments have been carried out.
When this is done, the resulting files must be taken back together with the files produced from
the self-diagnosis imaging and stored in a safe place.
* Recording of performance test or self-diagnostic test files can be enabled by starting Collection
Tool from among the service tools, selecting the [QC result] category, and checking the desired
items on the QC history list.
• The entire detector surface must be exposed without a subject to the X-rays under the same
conditions as the calibration conditions where the system is installed.
• When any kind of error has occurred, the calibration must be performed again.
• If the error persists even after performing calibration again, the following must be checked while
referring to the images:
* The entire surface of the detector is appropriately exposed to X-rays.
* The X-ray exposure must be performed properly (there must be no shading anywhere on the
entire detector surface).
* As when performing calibration, the grid is not attached.
* It must be checked that the defects have been rectified properly.
- 13 -
6. Troubleshooting
[Notes on use of a photo timer]
Using a photo timer with a stand mounting, etc., may effect the performance test and cause an
error, depending on the type and mounting configuration of the photo timer. The following may
be to blame for these errors.
• An image of the photo timer may be captured due to geometrical misalignment of the X-ray
generator during calibration.
• The mounting fixture of the photo timer or wires extending from it may be captured in the
calibration image, resulting in a low S/N determination for those portions only.
* Extraneous images, as described above, were captured by conventional SF systems as well, but
the characteristics of the film were such that these images were not visible. Therefore, there is
no problem with the sensor itself. If this kind of error has occurred, the calibration images,
performance test images and also the self-diagnosis images taken at the time concerned must
also be checked.
If the error persists after all of the above checks have been completed, check the following and
then contact Canon Inc.
• Confirm that the phantom image, etc., appears in film or the output of a high-resolution monitor
when images are processed in the normal manner, and clarify whether or not artifacts are being
observed. For example, artifacts can manifest themselves as vertical or horizontal lines, or as
rectangular granularity that obscures a portion of the sensor field.
[Self-diagnosis]
The self-diagnostic test function includes Test A device check, Test B test pattern check (which
corresponds to the former Test 2), and Test E offset correction test (which corresponds to the
former Test 5).
• This must be carried out after checking that the system is operating normally after installation.
• Self-diagnosis imaging for checking the conditions resulting after a system has been newly
installed and after the photo timer adjustments have been carried out must be implemented.
When this is done, the resulting files must be taken back together with the files produced from
the performance test imaging and stored in a safe place.
* Recording of performance test or self-diagnostic test files can be enabled by starting Collection
Tool from among the service tools, selecting the [QC result] category, and checking the desired
items on the QC history list.
Test E: If errors have been detected in connection with the offset image inspection
• Detector panel trouble or extraneous noise may be to blame.
* If errors have been detected, the calibration images, performance test images and also the
self-diagnosis images taken at the time concerned must also be checked.
If there are many abnormal or defective pixels in the self-diagnosis images, investigate how this
came about.
In the case of extraneous noise, change the position of the detector, and check for changes in the
extent of the resulting effect and the sources of the noise.
If the error persists, the following must be checked, and then Canon Inc. must be contacted.
- 14 -
6. Troubleshooting
• Confirm that the phantom image, etc., appears in film or the output of a high-resolution monitor
when images are processed in the normal manner, and clarify whether or not artifacts are being
observed. For example, artifacts can manifest themselves as vertical or horizontal lines, or as
rectangular granularity that obscures a portion of the sensor field.
- 15 -
Printed by Canon Inc.
April 2013 Rev.01
CXDI-701G/701C
Wireless
7. Service Manual
Report
CXDI-701G/701C
Wireless
8. Tools
CXDI-701G/701C
Wireless
9. Appendix
<Inspection Items>
ITEM Category NO. Item Characteristics values Test means Record
Note
Wireless communication PC(IEEE
1 External
must be possible 802.11n
incorporated)
Wireless It must be possible to
1 2 Access point Visual check
environment operate
Channel bonding
Access point On or Off
3 Visual check
setting Frequency band
2.4 GHz or 5 GHz
It must be possible to
1 Battery status Visual check
2 Battery charge
2 Battery cycle Up to 300 times Investigate
Battery It must be possible to Visual check
1
installation lock or release the lock Touch
Sensor LED
Blue LED must light when
2 (Blue, Green, Visual check
the power is ON
Blue-green)
3 Firmware Ver. - Visual check Ver.
4 FPGA Ver. - Visual check Ver.
3 Sensor unit
Sensor unit
5 (wireless) IP - Visual check
address
Sensor unit
6 (wired) IP With wiring sensor unit Visual check
address
Number of sensor
7 - Visual check
units installed
Start LNK Controller Visual check
Infrared Sensor recognition must
4 Sensor recognition 1 communication Visual check
be successfully done
link or wired link LED on the detector must
Visual check
not show error status
Connectors must be Visual check
1 Connectors
connected touch
Visual check
2 Earth It must be grounded
X ray I/F box touch
(when using in Cable must not be
Power Supply Visual check
5 standard 3 disconnected or plug
cable/adaptor touch
synchronization must not be damaged
mode) Visual check
4 POWER LED It must be turned on
touch
X ray I/F box IP Visual check
5
address touch
Connectors must be Visual check
1 Connectors
connected touch
It must be turned on
2 POWER LED when the power supply is Visual check
on.
Wiring unit (when
6 using with the wired LED status changes
connection) 3 Status indicator according to the Visual check
operation status.
Detector cable LED must
be turned on when Visual check
4 Detector cable
external power supply is touch
available.
9. Appendix Page 2 of 3
Inspection Check Sheet
<Inspection Items>
ITEM Category NO. Item Characteristics values Test means Record
Detector cable It must be routed in a Visual check
5
wiring proper way. touch
It must start normally
Image capture after the power is on. It
1 Visual check
computer must terminate normally
after the power is off.
7 System connection
CXDI control
2 It must start without error Visual check
software
Each Status It changes according to
3 Visual check
Lamp the operation status
8 The conditions under kV
1 X-ray generator which calibration is
Visual check mA
settings implemented must be
Calibration noted. msec
Calibration processing
Calibration
2 must completes Visual check
processing
successfully.
9 □TEST A
□TEST B-1
Self-diagnosis □TEST B-2
Self-diagnosis 1 TESTS must be passed. Visual check
pass/fail □TEST B-3
□TEST B-4
□TEST E-1
10 Performance test
Performance Test 1 Test must be passed. Visual check □Mode 1
Pass/fail
11 There must be no
Image quality Phantom
1 artifacts, shading, grid Visual check
exposure
stripes, etc.
12 □DB
backup
data
It must be collected by External □QC result
1 Retention of data
the Collection Tool. media □Other
□SnsPBInf
o
Control PC □elog
Date and Time must be
2 Date and Time Visual check
correct
Control Software
3 - Visual check Ver
Ver.
Remaining
4 battery charge on It must be fully charged Visual check
GUI
There must be no dirt or
1 Units Visual check
damage
13 System exterior
Cables must be wired
2 Cables Visual check
properly
14 Hookup with RIS, It must be possible to
1 patient data transmit and receive the Visual check
terminals data correctly.
Exposure
information, There must be no errors
Final checks 2 Visual check
patient data in in the information or data.
film server
X-ray generator, The equipment must be
3 RIS and other the same as before the Visual check
equipment inspection work.
9. Appendix Page 3 of 3
Inspection Check Sheet
<Inspection Items>
ITEM Category NO. Item Characteristics values Test means Record
Existing wireless
equipment
Adjacent environment
information
Notes on
Whether the DR Wireless
4 installation Visual check
SYSTEM installed in the
information
next room or not.
AP/LNK configuration tool
setting information
(such as channel setting)