Professional Documents
Culture Documents
Cancer Biomarker
Cancer Biomarker
Cancer Biomarker
of Cancer Biomarkers
Abstract- The proposed paper incorporates the modeling and There is standard test for breast cancer like mammography
simulation of the highly sensitive photonic crystal-based sensor for and for cervical cancer pep smear test. However, both these
breast and cervical cancer infected cell detection. Two structures, tests will require trained staff, costly equipment. The proposed
regular hexagon, and 180° phase mismatched hexagon are used. photonic crystal sensor can be benchmarked with respect to cost
For the detection of cancer infected cells, the change in refractive and testing procedures because of low cost of production when
index is considered as the basis and corresponding change in produces in bulk. Also, this sensor can be fitted to a handheld
properties of light rays is observed. A comparative study has been so will minimize the cost of maintenance of a bulky system
done for the sensitivity of the designed sensor. The results are
used for mammography. The essence of working of the
observed with the wavelength shift in the transmitted spectrum.
The sensitivity is calculated for both the structures and found that
proposed sensor is similar to the ionic lattice effect on electrons
180° phase mismatch is better than hexagonal shape as regards to in solid. In the same way, the periodic optical nanostructure, the
cervical cancer detection, hexagonal structure fares better than photonic crystal affects the photon movement. Electromagnetic
1800 phase mismatch for breast cancer detection with regards to propagation is affected by a photonic crystal which creates the
performance in sensitivity. bandgap in the layers of periodic dielectric nanostructures.
They have high perspectives ranging from gas sensors, optical
Index Terms - Refractive index, OptiFDTD, Sensitivity. filters, inkless painting, reflective flat display, and photonic
papers, etc. The study of cells in submicron resolution provides
I. INTRODUCTION the ways to RI values of cells. The RI values of normal as well
as cancer cell will be predetermined and is used as a benchmark
According to WHO every 1 out of 6th death is due to cancer. for further comparison. The cells that have been used for the
In India, more than 1300 people die due to cancer in a day. study of cancer affected cell and normal cell condition has the
Blood cancer, lung cancer, breast cancer, cervical cancer, etc. RI as shown in the below table.
are the various types of cancer. This paper mainly deals with
breast and cervical cancer. “Breast cancer is a type of cancer in
TABLE. I
which the cells of breast start growing out of control, beginning Refractive indices used for simulation
with ducts or lobules its can also spread out through vasculature
and lymph vessels, if not diagnosed initially” [1] Cervical
cancer occurs in cells of the lower uterus, growing Sl. No Type of cancer Type of cell RI value
uncontrollably.
1 Breast cancer Normal cell 1.385
So, the reason for the mass destruction of human life due to Cancer cell 1.3999
cancer is due to failure in recognition of the disease in an early
stage. Therefore, the world needs an easy and reliable 2 Cervical cancer Normal cell 1.368
recognition system of cancer. This paper deals with the
identification of breast cancer and cervical cancer using optics. Cancer cell 1.392
It’s a simple technique of comparing the refractive index (RI)
of normal cell and cancer affected cells based on the RI profile. Table I, shows RI values for cancerous and non-cancerous cells
The constituents of cells are plasma membrane(lipid),
(biomarker). From literature survey we found that for cancerous
cytoplasm (protein & water), and nucleus. Each of them has
and non-cancerous cell there is a different RI [2]. After
different RI values. For a cell to be “normal cell” all these
determining the RI value, the crystal structure is designed using
constituents should be optimum whereas an increase in protein
software called OptiFDTD-15.0.1. A two random structure is
content leads to carcinogen cells. [3] The difference in RI
selected to showcase the refraction of the light rays with
values is noted and this technique is used to detect the cancer preferred RI. The structures are Hexagonal shape and the
cells.
180° phase mismatched hexagon structure. The RI values of
II. SURVEY OF METHODS AND TOPOLOGIES normal as well as cancer cell will be predetermined and is used
as a benchmark for further comparison.
Authorized licensed use limited to: Cornell University Library. Downloaded on September 27,2020 at 18:25:45 UTC from IEEE Xplore. Restrictions apply.
III. RESULT AND DISCUSSION normal and cancerous cell.
Fig.4 Relative power of normal cell vs cancerous cell (all dimensions in um)
. .
Sensitivity = = = 2.0134 [1]
Fig.1 Hexagonal model for breast cancer (all dimensions in um) . .
Fig.2 Movement of light rays for normal cell (all dimensions in um)
Fig.3 Movement of light rays for cancerous cell (all dimensions in um)
The above Fig 2 and Fig 3 shows the simulation results for both Fig.5 The 180° phase mismatched hexagon model for breast cancer (all
normal and cancerous cells.it symbolizes the stress value dimensions in um)
associated with the cell. From Fig 2 and 3 we can conclude that
there is a maximum stress value at the right end of the structure.
The maximum stress value for cancerous cell is 0.07 and for
normal cell is 0.633. Hence easily we can differentiate between
Authorized licensed use limited to: Cornell University Library. Downloaded on September 27,2020 at 18:25:45 UTC from IEEE Xplore. Restrictions apply.
1) Observation: easy identification of cancer affected cells. The peak value of
the normal cell graph corresponds to the value of 1.69 and that
of a cancer cell is 13.24. So, this makes the clear difference
between normal and cancer cells.
2) Numerical calculation:
. .
Sensitivity = = = 0.6711 [2]
. .
Fig.6 Movement of light rays for normal cell (all dimensions in um)
Fig.9 Movement of light rays for normal cell (all dimensions in um)
Fig.7 Movement of light rays for cancerous cell (all dimensions in um)
Fig.11 relative power of normal cell vs cancerous cell (all dimensions in um)
Fig.8 relative power of normal cell vs cancerous cell (all dimensions in um)
The green curvature represents the relative power vs distance
of cancer cells whereas the red curvature represents relative
power vs distance of normal cell. The amplitude variation can
The green curvature represents the relative power vs distance be observed between normal and cancer cells which helps in
of cancer cells whereas the red curvature represents relative easy identification of cancer affected cells. The peak value of
power vs distance of normal cell. The amplitude variation can the normal cell graph corresponds to the value of 12.47 and that
be observed between normal and cancer cells which helps in
Authorized licensed use limited to: Cornell University Library. Downloaded on September 27,2020 at 18:25:45 UTC from IEEE Xplore. Restrictions apply.
of a cancer cell is 8.69. So, this makes a clear difference 1) Observation
between normal and cancer cells.
2) Numerical analysis:
. .
Sensitivity = = = 1.25 [3]
. .
Fig.15 relative power of normal cell vs cancerous cell (all dimensions in um)
2) Numerical analysis:
. .
Sensitivity = = = 4.166 [4]
. .
Fig.12 180° phase mismatched hexagon model for cervical cancer (all The idea proposed here is majorly about determining the cancerous
dimensions in um) cell using photonics. The observation made here is that the sensitivity
of cells has changed drastically. Additionally, by applying this
technique, it can be easily distinguishable between the normal and
cancer cell condition. Here in this work sensitivity is used to measure
the efficiency of the detection. If the sensitivity is more than the
efficiency of detecting the cancerous cell is more. The minimum or
maximum value of sensitivity is not a benchmark. The sensitivity is
used only to know how efficient the detection is. There is no need for
benchmarking the minimum value of sensitivity. In this paper, the
comparative study has been done for the sensitivity by considering
hexagonal and 180° phase miss-match. After sensitivity calculation it
is observed that for cervical cancer detection 180° phase mismatch is
better as its sensitivity is 4166nm/RIU compared to a hexagonal shape
having sensitivity of 1250nm/RIU. On the contrary, for breast cancer,
hexagonal structure is better with its sensitivity is 2013nm/RIU
compared with the 180° phase mis-match having sensitivity of
671nm/RIU. The designed structure has high sensitivity value hence
Fig.13 Movement of light rays for normal cell (all dimensions in um) can be profoundly used to separate cancerous and normal cells. If there
is a need of minimum value of sensitivity required using the proposed
technique to differentiate between normal and cancerous cells then
671nm/RIU can be taken as the basis. The future scope of extending
this work includes testing using more samples with different age group
and fabrication for a nano scale device.
V ACKNOWLEDGEMENTS
VI REFERENCES
Fig.14 Movement of light rays for cancerous cell (all dimensions in um [1] Cogliano VJ, Baan R, Straif K, et al. 2011. Preventable exposures
associated with human cancers. J Natl Cancer Inst. 103:1827–1839.
Authorized licensed use limited to: Cornell University Library. Downloaded on September 27,2020 at 18:25:45 UTC from IEEE Xplore. Restrictions apply.
[3] Li Q, Jiang Y. Top classic citations in pancreatic cancer
research. 2016;14:298.https://doi.org/10.1186/s12957-016-1061-8.
[5] Lee S-G, Choi J-S, Kim J-E and Park H Y 2008 Reflection
minimization at two-dimensional photonic crystal interfaces
[15] Ge, D., Shi, J., Zhang, Y., Zhang, L., & Yang, P. (2019, February
11). Numerical simulation of a novel bilayer photonic crystal slab
biosensor with hexagonal lattice
Authorized licensed use limited to: Cornell University Library. Downloaded on September 27,2020 at 18:25:45 UTC from IEEE Xplore. Restrictions apply.