Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Technical Specifications ARTIS Q Ceiling

BRAND / MANUFACTURER SIEMENS


MODEL / TYPE ARTIS Q Ceiling
COUNTRY OF ORIGIN GERMANY

Gantry
C-arm depth 94 cm
Gantry rotation ±135°
C-arm oblique projection 330° in the rotational direction and ± 100° in the orbital direction
Max. Rotation speed 25°/sec
Source Image Distance 90 - 120 cm
Isocenter to floor distance 108 cm
Programmable positions Yes

Detectors
Detector Size* We have two options, diagonal 20" large detector and diagonal 10"
Small detector
* Rumah sakit/ Pembeli wajib memilih ukuran detektor yang diinginkan pada saat proses pembelian di laman e-katalog
Synchronized FD/collimator rotation Automatic synchronous rotation of the detector and collimator unit to
compensate for image rotation at different examination positions of
the support stand

X-ray tubes
Type GIGALIX
Cathode technology Flat emitter technology
Flat emitter technology offers the highest fluoroscopy power and also
a smaller point source that will improve the spatial resolution of the
image.
Tube performance - Pulsed Fluoroscopy current: 250 mA
- Anode heat storage capacity: 5.2 MHU
- Anode cooling rate: 1.52 MHU/ min
- Max Tube Current : 1000 mA
- Filter levels: 0.1, 0.2, 0.3, 0.6 and 0.9 mm Cu
Collimator Rotatable collimator

Patient table
Type Floor-mounted patient table for all angiographic examinations and
interventions
Table top Translucent Free-floating tabletop

Generator
Model Polydoros A100 Plus
- Converter Frequency: 100kHz
- Max. generator power : 100kW
- Tube voltage : 40 to 125 kV
- Tube Current : 15 to 1000 mA

Dose management features


Dose features / package Combined Applicactions to Reduce Exposure (CARE)
Automatic calculation and optimization of exposure CAREmatic
data based on fluoroscopic data
Display of the measured dose area product and CAREwatch using measurement chamber for most accurate reading.
the calculated patient entry dose on live monitor

Pre-filtration for lower skin dose. Automatic CAREfilter with automatic positioning
placement and responding in real-time to SID,
patient thickness, C-arm and table positions

Unrestricted
Pulsed fluoroscopy CAREvision
(0.5 p/s to 4 p/s, 9 settings) including standard pulsed fluoroscopy:
7.5, 10, 15, 30 p/s
Collimation without radiation using LIH CAREprofile
Positioning of patient without radiation CAREposition
Real-time patient dose monitoring with visual and CAREguard
audio warnings for exceeding pre-set patient DAP
with choice of 3 different settings
Real-time patient dose monitoring with display on CAREmonitor: Shows accumulated peak skin dose based on current
live monitor of accumulated skin dose based on projection of C-arm. Any change in C-arm, table, SID, zoom or
actual C-arm angulations, table positions and SID collimator prompts the system to automatically update the calculation

Clinical Applications Software


analysis and measurement of cardiac vessels QCA, (Quantitative coronary analysis)
analysis and measurement of left ventricle LVA, (Left ventricular analysis)
Coronary stent assessment CLEARstent
Archived as DICOM image
3D rotational Dyna 3D
Vascular navigation with fluoroscopy on the Roadmap
monitor in the exam room.
Rotational Angiography DYNAVISION

Patient monitoring and reporting Yes, Using Sensis Hemodynamic and Reporting.

Accessories :
Intercom Yes
LED Exam light Yes
Upper body protection Yes
Lower body radiation protection Yes

Other Accessories :
Lead glass Yes, 100 cm x 120 cm
Injector Yes, Injector

Unrestricted

You might also like