Artis Q Ceiling Spek E-Katalog
Artis Q Ceiling Spek E-Katalog
Gantry
C-arm depth 94 cm
Gantry rotation ±135°
C-arm oblique projection 330° in the rotational direction and ± 100° in the orbital direction
Max. Rotation speed 25°/sec
Source Image Distance 90 - 120 cm
Isocenter to floor distance 108 cm
Programmable positions Yes
Detectors
Detector Size* We have two options, diagonal 20" large detector and diagonal 10"
Small detector
* Rumah sakit/ Pembeli wajib memilih ukuran detektor yang diinginkan pada saat proses pembelian di laman e-katalog
Synchronized FD/collimator rotation Automatic synchronous rotation of the detector and collimator unit to
compensate for image rotation at different examination positions of
the support stand
X-ray tubes
Type GIGALIX
Cathode technology Flat emitter technology
Flat emitter technology offers the highest fluoroscopy power and also
a smaller point source that will improve the spatial resolution of the
image.
Tube performance - Pulsed Fluoroscopy current: 250 mA
- Anode heat storage capacity: 5.2 MHU
- Anode cooling rate: 1.52 MHU/ min
- Max Tube Current : 1000 mA
- Filter levels: 0.1, 0.2, 0.3, 0.6 and 0.9 mm Cu
Collimator Rotatable collimator
Patient table
Type Floor-mounted patient table for all angiographic examinations and
interventions
Table top Translucent Free-floating tabletop
Generator
Model Polydoros A100 Plus
- Converter Frequency: 100kHz
- Max. generator power : 100kW
- Tube voltage : 40 to 125 kV
- Tube Current : 15 to 1000 mA
Pre-filtration for lower skin dose. Automatic CAREfilter with automatic positioning
placement and responding in real-time to SID,
patient thickness, C-arm and table positions
Unrestricted
Pulsed fluoroscopy CAREvision
(0.5 p/s to 4 p/s, 9 settings) including standard pulsed fluoroscopy:
7.5, 10, 15, 30 p/s
Collimation without radiation using LIH CAREprofile
Positioning of patient without radiation CAREposition
Real-time patient dose monitoring with visual and CAREguard
audio warnings for exceeding pre-set patient DAP
with choice of 3 different settings
Real-time patient dose monitoring with display on CAREmonitor: Shows accumulated peak skin dose based on current
live monitor of accumulated skin dose based on projection of C-arm. Any change in C-arm, table, SID, zoom or
actual C-arm angulations, table positions and SID collimator prompts the system to automatically update the calculation
Patient monitoring and reporting Yes, Using Sensis Hemodynamic and Reporting.
Accessories :
Intercom Yes
LED Exam light Yes
Upper body protection Yes
Lower body radiation protection Yes
Other Accessories :
Lead glass Yes, 100 cm x 120 cm
Injector Yes, Injector
Unrestricted