Professional Documents
Culture Documents
User Manual of Aigel 300
User Manual of Aigel 300
User Manual of Aigel 300
GROUPING SYSTEM
USER MANUAL
(for Aigel 100/300)
Preface
Introduction
Thank you very much for becoming the user of Automatic Blood Grouping System.
Before performing any work on or with Automatic Blood Grouping System, first read the manual
carefully.
To achieve better results, you have to be familiar with the performance of Automatic Blood Grouping
System before performing any test on it.
Please keep it in good situation after reading, in order to be able to access at any time when needed
Copyright Notice
The copyright of this document is reserved by Ai Kang Medtech Co., Ltd. Any unauthorized copy,
modification and reproduction are forbidden.
◎ 2016-2020Copyright reserve all rights: (China) AiKang Medtech Co., Ltd.
Declaration
Aikang Medtech Co., Ltd. reserves the right of final interpretation of this manual.
In the case meets all of the following requirements, Aikang is considered to be responsible for product
safety, reliability and performance, namely:
■ Assembly operation, expansion, re-adjust, improvement and repair are all performed by professionals
recognized by Aikang.
■ All replacement parts and accessories, consumables are all approved by Aikang.
■Electrical equipment is in line with national standards and the requirements of the manual.
■ Operation is carried out in accordance with this manual.
■This manual is subject to change without notice.
■ Man-made damage;
■ Misemploy;
■ Network voltage is exceeding the range of this product;
■ Irresistible natural disasters
■ Replacing or using parts, accessories, supplies without Aikang-approved or maintenance by non
Aikang-authorized personnel;
I
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Warning
●This system is limited operated by the inspection professionals, doctors or laboratory technician, who is
trained by Aikang or Aikang’s agent.
●If each hospital or institution that use this system cannot achieve a satisfactory repair / maintenance plan,
it will likely result in abnormal equipment failure, and even may endanger human health.
●Make sure that operating this system under the conditions specified in the manual. If beyond conditions of
using, the system may not operate properly, and the measurement results will be unreliable and potentially
damaging some parts of this system or even damage to personal safety.
Notice
This manual is intended for the following laboratory professionals:
●Systematic daily operations personnel;
●System maintenance and troubleshooting personnel;
●Learn the system operation personnel.
II
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
CONTENT
PREFACE ..................................................................................................................................I
CHAPTERⅠOVERVIEW ...................................................................................................... 1
1.1 TERMINOLOGY ........................................................... 1
1.2 SAFETY REGULATION ...................................................... 1
1.3 SYMBOLS .............................................................. 1
1.4 SYSTEM BASE APPLICATION ................................................. 2
1.5 PURPOSE AND LIMITATIONS.................................................. 3
1.6 SAFETY WARNING AND PRECAUTION............................................. 3
1.6.1 Ordinary Precaution ....................................................................................................................... 3
1.6.2 Check and Confirm ........................................................................................................................ 3
1.6.3 Security of Electricity..................................................................................................................... 4
1.6.4 Risk Factors .................................................................................................................................... 4
1.6.5 Mechanical Hazard ......................................................................................................................... 4
1.6.6 Electric Risk ................................................................................................................................... 5
1.6.7 Chemical and Biological Hazards .................................................................................................. 5
1.6.8 Software Precautions ...................................................................................................................... 5
1.6.9 Manufacturer .................................................................................................................................. 5
1.6.10 CE Conformity ............................................................................................................................. 5
CHAPTERⅡ INSTRUMENT INTENDED USE AND OVERVIEW ............................... 7
2.1 BASIC FUNCTION ......................................................... 7
2.2 BASIC CONFIGURATION ..................................................... 8
2.2.1 Module Configuration .................................................................................................................... 8
2.3 WORKING PROCESS ........................................................ 9
2.4 AUTOMATIC BLOOD GROUPING SYSTEM CONFIGURATION ............................... 9
2.5 SYSTEM MODULE ........................................................ 10
2.5.1 Robot Arm Module....................................................................................................................... 10
2.5.2 Wash Station ................................................................................................................................. 11
2.5.3Tube Rack Scanning Module ........................................................................................................ 12
2.5.4 Reagent Shaking Module ............................................................................................................. 12
2.5.5 New Cards for pipette................................................................................................................... 12
2.5.6 Incubator....................................................................................................................................... 13
2.5.7 Centrifuge module ........................................................................................................................ 14
2.5.8 Reader .......................................................................................................................................... 15
2.5.9 Weight Module ............................................................................................................................. 15
CHAPTER III INSTRUMENT INSTALLATION.............................................................. 16
3.1 AUTOMATIC BLOOD GROUPING SYSTEM INSTRUMENT ACCESSORIES ....................... 16
3.2 INSTALLATION REQUIREMENTS ............................................... 16
3.2.1 Environmental Conditions ............................................................................................................ 16
3.2.2 Power............................................................................................................................................ 17
3.2.3 Computer ...................................................................................................................................... 19
3.2.4 Anti-virus software ....................................................................................................................... 19
3.3 CONNECTED TO INSTRUMENT................................................. 19
3.3.1 Power Connection ........................................................................................................................ 19
3.3.2 Interface of Devices and Computers ............................................................................................ 20
3.3.3 Liquid Line Connection................................................................................................................ 22
CHAPTER Ⅳ OPERATION GUIDE ................................................................................ 23
III
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
IV
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
V
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
ChapterⅠOverview
Before performing any work on or with Automatic Blood Grouping System, first read the manual
carefully.
This Section includes the safety precautions about operating on Automatic Blood Grouping System 's.
In this manual, it will in detail talk about the safe specific operation. In the process of reading this manual,
reader have to pay attention to the warning sign labeled content, and must strictly comply with all the safety
instructions of this manual.
Before performing any work on or with Automatic Blood Grouping System, please read the following
safety tips and guidance, well understand and implement all the precautionary measures, to ensure the
proper use of this system, so that user can play its best features, and ensure operator safety.
1.1 Terminology
In this manual:
“Instrument”indicates Automatic Blood Grouping System.
“User Software”indicates the operating software of Automatic Blood Grouping System.
“Data base”indicates all information of user software and testing results of Automatic Blood Grouping
System.
1.3 Symbols
There are Signs and symbols following of Automatic Blood Grouping System in this manual and
packaging. Their meanings are as follows:
Warning: General Warning, please refers to the detailed instructions on the manual, and
carefully operates.
Warning: There is an electric shock hazard; there is a risk of causing injury or damage to
the equipment. Please refers to the detailed instructions on the manual, and carefully
operate.
Warning: There is a high temperature burn hazards. Please refers to the detailed
1 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Warning: danger of toxic chemicals may cause operator injury or equipment damage.
Please refers to the detailed instructions on the manual, and carefully operate.
Warning: There are biologically harmful substances, which may cause operator injury or
equipment damage. Please refers to the detailed instructions on the manual, and carefully
operate.
Warning: There is a laser shot, it may cause damage to the operators’ eyes, please do not
look directly into the light emitted from the laser scanning head.
Caution: In the absence of a well understanding of this manual, it will cause damage or
data loss to this system; it is strongly recommended that the operator need to read the
relevant sections of this manual.
Temperature limits: Temperature Limit that transport package should comply with.
Fragile: Transport package contents fragile, handle with care when transporting.
Stacking Limit: The maximum stacking layers of the same package, 4 is the number of
limit.
2 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Automatic Blood Grouping System is applicable to all hospitals, blood banks, the CDC and any other
institutions, that need automatic detection of positive and negative stereotypes ABO blood group,
cross-matching of blood, Rh (-D) and irregular antibody screening.
It can effectively improve the efficiency of the experiment via automated batch processing .
All steps are performed by standardized to reduce the possibility of error handling or subjective bias
error when analyzing, increase the credibility
It can reduce the risk of the operator being infected via reducing the operator exposure to the sample
and reagents.
Limitations of this device are:
The instrument can only be used for immunohematology testing, but not for other nominal test
species.
The instrument can only use cards with gel pores
If there are doubted or uneven reactions, the operator need to acknowledge.
3 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Only consumables or original parts supplied by Aikang can match with Automatic Blood Grouping
System, such as dilution plate, waste container, the system fluid bottles etc.
In the course of handling a single module or move the whole instrument, please make sure stable, do not
roll or inverted, and take the necessary protective measures.
To ensure the security and measurement data integrity during operation, please make Uninterruptible
Power Supply (UPS) available.
1.6.3 Security of Electricity
The instrument is grounded through the ground power line, in order to avoid electric shock, the power
ground wire must be grounded.
AC power must be stable; it is prohibited to share power with other high-power instruments.
During operation or maintenance, please don’t touch the power connector on the rear of the instrument pod,
or it may have a risk of electric shock.
1.6.4 Risk Factors
Warning:
Warning: Biological Movement parts
Harmfulness (pipetting channel) (Robot Arm)
Warning: Biological
Harmfulness Power Switch
(Working Platform)
Scram Button
Fig.1.1 Warning flag
1.6.5 Mechanical Hazard
Automatic Blood Grouping System is automation equipment, it may cause personal injury when moving
parts are on work.
Please note the following possible mechanical hazards:
Robot arm may cause squeeze harm.
Permanent pipetting needle may cause stabbing harm.
When Automatic Blood Grouping System is in work process, it is strictly prohibited to stretch any part of
body into the working cabin, unless the system has been suspended and a dialog box reminding manual
operation. If not, it may cause injury to the operator, and also Automatic Blood Grouping System
damage.
4 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Note:Please pay attention to warning labels, and follow safety guidelines from reagent
vendors. Laboratory chemical effects and ultraviolet radiation etc. of environmental factors
may cause centrifuges, reader damage. In order to ensure the continued use of the instrument,
please operate strictly in accordance with Laboratory Practices specification.
1.6.9 Manufacturer
5 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
98/79/EC on in vitro diagnostic medical devices. With the declaration of conformity the manufacturer
declares conformity with the provisions of the Directive.
6 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Automatic Blood Grouping System sets liquid distribution, incubation, reagent mixing, puncture,
centrifugation, analysis and other functions, that required in analysis experiments, in one interpretation,
with intelligent process optimization, to ensure the accuracy of the test results. It is intended to use for
Reverse& Forward tests, Cross Match tests and Antibody screen tests based on Gel cards before Blood
transfusion in Diagnostics laboratories.
Wash Station
Reader
Liss Liquid
Tubes Rack
7 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
It can automatically perform the whole process of Gel card Blood grouping tests on Automatic Blood
Grouping System, also separately as required pipetting, incubation, centrifugation, reading and other
activities.
Automatic Blood Grouping System main features include:
Sample barcode scanning
Sample distributing
Reagent distributing
Incubation
Centrifugation
Reading
User software controls all modules’ movement, processing test results, test methods setting, and instrument
maintenance. Its Key features include:
controlling instrument operation
statistical test results
generating test results
creating, editing, test methods
Instrument System Maintenance
8 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Reader
9 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Start switch
Front door
handle
Light switch
Scram button
Fig.2.3 External View
Communication port
power input
10 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
11 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Tube rack scanning module’s main function is to place the specimen and scan the specimen barcode.
Specimen are placed in a single tube rack, each rack can be placed up to 16 samples. In the beginning of the
experiment, the tube rack is pushed into the tube rack carrier one by one according to prompts; the scanner
will automatically scan the barcodes.
Warning:During scanning, there is laser beam ejecting from the scanner, please don’t stare at
the scanner.
Reagent bottle will be placed on this shaking module to mix reagent. When pipetting is not performing, this
module will shake to mix reagents; when performing, it will stop shaking and return back to balance to
ensure well pipetting.
2.5.5 New Cards for pipette
Before running test,please Tear the aluminum film on the gel card After starting test, the gripper will get
the new card to scan, and then place on the original position. Start pipetting……
12 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
2.5.6 Incubator
Fig.2.10 Incubator
After liquid dispensing, the gel card will be incubating. Incubating temperature and time can be set via user
software according to the reagent specification.
This incubator can place up to 24 gel cards at the same time. If some cards need incubation ,the gripper will
holding a gel card gets to the incubator, Then close the lid began to incubation.
13 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Fig.2.11 Centrifuge
The main function of this module is to centrifuge the gel card, in order to deposit the un- agglutination red
blood cells in the bottom of gel card.
This module can hold up to 24 gel cards to centrifuge.
When centrifuging, gel card tray must be trimmed. User software can calculate based on the number of test
sample, and grip the new card to make it balance.
14 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
2.5.8 Reader
Fig.2.12 reader
The reader of Automatic Blood Grouping System, is researched and developed by Aikang, with high
reading speed, high reliability.
Reader is to read reacting card and transport the results and the original image to the system database.
2.5.9 Weight Module
15 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
16 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Dimensions Weight
Model
Length × width × height(mm) (kg)
Aigel 100 1038×781×1039 145
Table 3.2 Instrument Dimensions and weight
Depending on the configuration of the instrument,working audience should be set aside area to place
auxiliary equipment.
Placement of the instrument must ensure that the operator can easily operate and maintain the instrument,
and ensure good ventilation.
Besides,Automatic Blood Grouping System installation location must also meet:
The plant operator can hear the alarm device, can see the indicator signal. When the device occurs an
exception in need of emergency stop, the operator can easily press the red emergency stop button in the left
corner of the front of the device, so that disconnect the power supply of all modules of the equipment in a
timely manner.
Notice:
In the dry environment, especially the existence of artificial materials (synthetic fabrics,
carpet, etc.) ,when using this equipment, may cause damage of electrostatic discharge, resulting
in abnormal operation.
Beside the strong radiation sources shall be forbidden to use the device, otherwise it may
interfere with the normal work of the equipment.
evaluate the electromagnetic environment before Install the equipment.
3.2.2 Power
Automatic Blood Grouping System suitable for 220V AC power,the power supply requirements are as
follows:
Power outlet should not be located at the back of the instrument, ,Shall be placed in a position that
convenient operation,current supply of the equipment should be implemented voltage control, and make
overload protection.
Automatic Blood Grouping System power as show in the following table:
Model Power Fuse Specifications
17 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Fig.3. 1fuse
Note:The operator is not allowed to contact or replace with the fuse, If you need to replace the
fuse, to be operating by professionals.
Warning:There is high-voltage in the interface main power box of the instrument. When the
repair or maintenance of equipment, please cut off the power first.
Aikang MedTech Co.,Ltd recommends that you use UPS (3KVA) power. When experimental power
outage occurs, the machine can continue to run,to ensure experimental data and experimental processes
information is not lost, the equipment potential damage caused by power outage when running can also be
prevented.
UPS power specifications:
Technical Details Recommended values
Maximum power
3000W
consumption
Average power consumption 500-650W
Average power consumption
~300W
of the computer
18 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
In order to ensure safe operation of laboratory personnel,Aikang MedTech Co.,Ltd requires the equipment
grounding. Automatic Blood Grouping System is grounded through the power line ground. To avoid
electrical shock, the power supply must be connected to earth ground.
3.2.3 Computer
Install and run Automatic Blood Grouping System requires a separate computer (computer), Please do
not share your computer with other devices or systems.
Automatic Blood Grouping System equipped with a computer, fully able to meet the instrument software
to install, use requirements. For a self-made computer, the computer's configuration requirements of not
less than:
Configuration Name Parameter
CPU 2.4GHz
Memory 1GB
Monitor Color, Resolution:1280×1024
Operating system Windows XP/7, RecommendWindows7
Serial ports:RS 232
Board Connector USB:4×USB2.0
Computer I slot:2×computer I Express
Table 3.5 Computer minimum
19 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Follow the signs, use USB cable to connect the instrument to the computer corresponding port.
20 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
CAN1,CAN2 Bus,
Close to the motherboard
side is CAN1
Fig. 3.4 Instrument bus, video output lines connect computer interface
21 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Fig. 3.6 The connecting port of system liquid, waste liquid pipe
22 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Double-click the icon of the Auto Blood software into the software. Click on the device
menu to initialize the machine, to view whether there are error messages or warnings to ensure that the
device is properly connected.
The system parameters in the system parameters must be checked whether they are set correctly before
use, to ensure that the test is edited correctly.
Note:The software that Aikang provides does not allow non-professionals to make any
changes.
4.2 Power On
Users need to start with the computer equipment in the following order:
1) Press the instrument power switch to turn on the Automatic Blood Grouping System
instrument.
2) Start the computer that controls the Automatic Blood Grouping System instrument.
Double-click in the path D:\ Across to open the login screen, as show as Figure 4.1:
23 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Run the test. The button will be invalid during test running.
24 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Pause test.
Quit software.
In operation interface, click to enter the reagent loading interface, as show as Figure 4.3:
Put the corresponding reagent bottle into the reagent carrier, click , the scan head lamp lights,
then slowly push the reagent carrier to the end position. This time reagent bottle barcode scanning results
will be displayed in the list shown in Figure 4.3. Reagent scan module was optional for user.
If you need to manually enter the barcode,manually enter the barcode corresponding to the reagent bottle
in the input box, entered the position number that reagent bottles are
25 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
positioned into the box before scanning; scan results are displayed in the input box,
and then placed the Liss liquid bottle into the position of Liss liquid. Enter the position number in the
Position box to specify the location that the Liss liquid is placed.
If you need to manually enter the barcode, just enter the barcode in , enter the position
number in the Position box to specify the location that the Liss liquid is placed.
26 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
In operation interface, click to enter micro gel card loading interface, as show as Figure 4.5:
Enter the code in , enter the position number of the micro-gel cartridge is
27 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
In operation interface, click to enter the test project and the sample count loading
interface, as show as Figure 4.6:
28 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Once there is one barcode un-scanned, please slide out corresponding sample rack, check
the integrity of sample barcode and re-scan that rack.
If there are still some barcodes not scanned perfectly, user could double click corresponding
wells and type in barcode manually.
Attention: 1. All sample barcodes direction should be same with location barcode
direction.
2. Pause a little bit at the edge of sample rack holder when sliding samples
in to trigger the laser scanner to light up, then slide the tube rack in
smoothly.
3. Please scan the sample as a uniform speed.
Warning: There is laser ejected from scanner, please do not stare at lens.
29 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Warning:
1. For correct operation, sample barcode must be stick on sample tube and it’s not
manuscript or missed.
2. Using haemolytic, chylus, clotted or visible fibrous protein may lead to false
positive result.
to F column.
One sample could be selected to do multiple items, as show as figure 4.9. All selected test
item must be edited before test running. For item editing, please take the contents in appendix 3
as reference.
30 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
When “Complete Cross Match Major & Auto” is chosen, click to use patient cursor
for selecting patient, left click in blank area in list according to sample barcode, the “P*” will be
marked after selection. Click to use donor cursor for selecting donor, left click in blank
area in list according to sample barcode, the “D*” will be marked after selection. One patient
could be matched with several donors synchronously in cross match item. Click to
select second patient, this frame will become cross 2 which means
second patient, then select corresponding sample as shown above.
If other item (not cross match) such as “Forward&Reverse ABO with DVI-/DVI+” is
selected, tick the corresponding position directly. Multiple options could be selected by
dragging mouse.
If any mistake happened, use to cancel selected sample. Left click the
31 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Warning:
For proper operation of the instrument, barcode must be firmly affixed in the sample tube wall
and cannot be handwritten or missing.
Use Hemolytic, chyle samples or with clot, visible fibrins samples may lead to interpretation of
false-positive results occur.
Note:Experiments should be used the centrifuged sample.
Close the device front door, click to run the current test.
Note:Before the test run the front door of the equipment should be closed.
32 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Click then pause screen appears and suspend the current test. Click“continue”button to
continue running test.
Check the "Enable Abort button" and click on the "Abort" button to terminate the current test.
Click to exit the test program. It cannot exit when the process test is running, you
can exit after terminate the test.
33 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Chapter V Maintenance
In order to ensure the accuracy and reliability Automatic Blood Grouping System running experiment,
the user must be a regular maintenance and repair for the instrument.
Automatic Blood Grouping System equipment maintenance is divided into three different levels:
Daily maintenance
Weekly Maintenance
Regular annual maintenance
Warning:To safely perform maintenance work, please wear a lab coat, wear disposable rubber
gloves and protective glasses.
Turn on the instrument power and open user software, click to initialize the system.
Observe whether each pipetting channel operation is normal during the initialization process. If there are air
bubbles in the pipeline,Please re-initialize the system and flick pipeline with your hand until bubbles
cleared during rinsing.
If the bubbles in the pipeline cannot be excluded, to check the interface in the pipeline whether there is
leakage, etc. and phone contact engineers to solve problems.
5.1.1.2 Maintenance after Test
1) Clean the tip
34 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
After closing the instrument, with a lint-free paper or swabs moistened with alcohol to gently wipe the
outer wall of the tip to remove a blood clot or reagents crystallization may exist.
2) Clean the wash station
With a cotton swab dipped in alcohol to wipe the inside wall of the wash station, then washed it with dilute
molar lotion and other special lotion. Finally, use distilled water rinsing.
3) Empty waste liquid bucket
Keep the waste liquid into the designated location and make the necessary disinfection.
Warning:
Waste liquid container and pipeline with biohazardous, please take precautions : Wear
protective clothing; wear protective rubber gloves and protective glasses.
After maintenance is completed, please wash your hands with disinfectant to prevent microbial
infection.
4) Cleaning Platform
Remove the test tube rack, reagent bottles and other accessories from the work cabin platform of the
instrument. Use a wet cloth with distilled water to carefully clean the surface of work cabin platform and
each module, and then wipe them with 75% alcohol solution.
If the reagent or other liquid residue on the platform, please scrubbed with cotton dipped in alcohol. Finally,
wipe it with alcohol.
Used dedicated scanning head wipes to clean the lens of the specimen scanning head, ensure that the mirror
clean and dust-free.
5.1.2 Weekly Maintenance Methods and Procedures
1) Cleaning system liquid bucket
After the experiment is completed, empty the system liquid in the system liquid bucket. Use of distilled
water that is not higher than 50 ℃, Cleaned it with a brush Then washed three times with distilled water.
2) Soak pipeline
According to the proportion to formulate Moore lotion and loaded into the system liquid bucket. Click the
"rinse needles" under the "action" menu bar in the user software to rinse the pipetting channel three times,
each time let Moore lotion soaking in the pipeline 15 minutes.
Finally, rinse pipeline 3 times with system liquid.
3) Backup Auto Blood Folder
Exit Auto Blood user software,compressed backup Auto Blood folder which is under "D: \".
4) Check the condition of tip
After closing the instrument, check whether there is an obvious scratch on the tip of liquid handling system.
If the tip coated layer is damaged, please contact engineers to replace tip.
5) Wash all reagent bottles, dilution plates
Empty all reagent bottles, dilution plates(Excluding the single-use bottle or dilution plate), Wash them
with distilled water, and then dry naturally.
5.1.3 Regular Annual Maintenance
Regular annual maintenance performed by the engineers of Aikang MedTech Co.,Ltd or authorized agent
engineer, perform precision calibration of the instrument and thorough cleaning and maintenance.
35 / 64
AUTOMATIC BLOOD GROUPING SYSTEM USER MANUAL
Engineer to check the user's situation of daily maintenance and weekly maintenance. If find that user
did not perform a complete maintenance cycle, it is need to retraining instrument operator and stressed the
importance of the maintaining.
Asks the user the existing problems in used, record and solve them.
Check and clean the system solution bucket. First use water that is not higher than 50 ℃to clean, then
disinfected with disinfectant(or 75% ethanol),after soaking for 15 minutes, washed with distilled water.
Check whether the operation of each module of pipetting arm and handler are normal. Check whether
the screws are loose or rusty; clean the two robotic arms rail. Prohibited use alcohol or other solvents to
clean the rails.
Check whether there are damages of washable tip, etc.,and regulate verticality of the tip.
Check whether the centrifuge tray rotor is flexible and do necessary cleaning lubricated.
Clean-up the sample pump. Filled with system solution to the system solution bucket. Enter Auto
Blood user software to rinse tip, emptying all air bubbles in the pipe.
Test whether probe module function of each tip is normal.
Calibration accuracy of the liquid handling system.
Calibration accuracy of the reader.
Clean the entire platform of the equipment and its surface.
Back up all motors and pumps hardware parameters and hardware version information.
Check the Ghost of the system whether exist. If not, redo system Backup.
After the instrument maintenance engineer need to fill < Regular Annual Maintenance Record Form >
carefully.
5.1.4 Maintenance Record Form
36 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Maintenance
personnel Signature
Please carefully complete the periodic maintenance procedures specified in this table, and complete maintenance records. If you find equipment malfunction, please
immediately contact service engineer that Aikang authorized. Telephone service: 800 999 0929.
37 / 64
ACROSS AUTO SYSTEMUSER MANUAL
38 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Soak pipeline
Maintenance
personnel
Signature
Please carefully complete the periodic maintenance procedures specified in this table, and complete maintenance records. If you find equipment malfunction, please
immediately contact service engineer that Aikang authorized. Telephone service: 800 999 0929.
39 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Replacement pipeline of
the liquid handling system
Check whether the operation
of each module of pipetting
arm and handler are normal.
Check whether the screws are
loose or rusty, clean the two
robotic arms rail
Check whether there are
damages of washable tip, etc.,
and regulate verticality of the
tip
40 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Maintenance
personnel Signature
Equipment
Authorized Signature
Please carefully complete the periodic maintenance procedures specified in this table, and complete maintenance records. If you find equipment malfunction, please
immediately contact service engineer that Aikang authorized. Telephone service: 800 999 0929.
41 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Note:When transporting the instrument, use only the original packing material.
Warning:
Prohibited crawling robotic arms of the Instrument, glass plate or platform trunk to lift the
equipment.
Note:
Please keep the packaging material after disassembly, so that packaging equipment to the next.
42 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Warning:The heavier the weight of the instrument, while handling should pay attention to
safety.
Note:Do not turn on the main power immediately after placement instrument; please operate
according to this manual follow.
When removing the packaging film, to handle with care, it can not scratch.
6.4 Packaging
When packing equipment, we recommend that you use the original foam blocks and boxes. If the original
carton and foam block lost or damaged, please contact your supplier or Aikang MedTech Co.,Ltd to
re-order.
43 / 64
ACROSS AUTO SYSTEMUSER MANUAL
6.4.1 Cleaning
Before packaging, transportation equipment, instruments should first conduct a comprehensive clean.
Warning: When cleaning, there may be a potentially dangerous liquid spilled from the
instrument process, including infectious clinical sample, biological hazard.
Cleaning and decontamination equipment and modules before transport packaging, please
wear protective clothing, gloves and glasses to prevent infection.
System liquid pot installed with distilled water, for all pipetting channels and pipeline
flushing operation.
With a cotton swab dipped in 75% alcohol to clean the ends of each pipetting tip.
With a cotton swab dipped in 75% alcohol to wipe the inside and outside of grip shrapnel.
With a cotton swab dipped in 75% alcohol to clean equipment work cabin platform
6.4.2 Packaging pipetting arm
Together pipetting tip adapters for each pipetting channel, and moved to the middle position
in the Y direction of pipetting arm.
With thicker foam blocks to protect pipetting tip adapter, and use PE packaging film fixed
on the cover of pipetting arm.
Move the pipetting arm to the left, close to the left side plate of the instrument work cabin,
and plus foam board apart between the left side plate of work cabin and pipetting arm.
Below the ends of the Y direction in the pipetting arm plus foam blocks to support,Prevent
pipetting arm sliding along the rails.
Note:Use PE packaging film to fixed pipetting channel together,Z direction should use foam
board fixed to the home position, Otherwise, may cause pipetting channel vertical deformation,
thus affecting the pipetting accuracy.
6.4.3 Packaging work cabin platform
Remove the test tube rack module, incubate module, kit, dilution plate and carrier,
centrifuge, reader on the work cabin platform, wrapped with foam packaging film to protect,
place into the special box to packaging.
Covering a thin sheet of foam on the platform, then tightly wrapped with PE packaging film
fasten it to the floor on the table.
6.4.4 Packaging equipment
Installation the equipment acrylic glass and wrapped with PE packaging film.
Transport handle fitted to the instrument, lift the equipment,carefully placed on the wooden
44 / 64
ACROSS AUTO SYSTEMUSER MANUAL
pallet of packing crate, let the structural frame in the bottom of the device fall on the foam
blocks on the wood care of the box.
Remove the transport handle, packaged devices to save.
6.5 Transport
When transporting the instrument, the transportation Precautions on the instrument package must be
meeting. When handling the instrument, it must be placed vertically upwards, prohibit rollover, collision or
fall and to prevent rain.
Note:During transport, even moving short distances, pipetting arm of the instrument must be
fixed in the side panels of the work cabin. Meanwhile, pipetting channel should be placed in
the origin position of Z direction.
Before transport, if you do not take the right package of instruments, the damage caused will
not warranty.
45 / 64
ACROSS AUTO SYSTEMUSER MANUAL
46 / 64
ACROSS AUTO SYSTEMUSER MANUAL
47 / 64
ACROSS AUTO SYSTEMUSER MANUAL
8.2 Contacts
Company Name:Aikang MedTech Co.,Ltd
Registered Address:Building B, Zhangkeng Industrial Park, Minkang Road, Minzhi, Bao’an District,
Shenzhen, Guangdong, China
Production Address:Building B, Zhangkeng Industrial Park, Minkang Road, Minzhi, Bao’an District,
Shenzhen, Guangdong, China
Phone:0755-25871250 Toll-free:800-999-0929
Fax:0755-25871281
Technical Support:Service@aikang-medical.com
Product Sales:Sales@aikang-medical.com
8.3 Qualification
Manufacturing Enterprise License for medical device No.:粤食药监械生产许20030835号
Registration certificate for medical device No.:
Product Standard No.:YZB/粤 2100-2013
Online Website:http://www.aikang-medical.com
Instant Messaging
Phone:0755-25871250 Toll-free:8009990929
Fax:0755-25871281
48 / 64
ACROSS AUTO SYSTEMUSER MANUAL
In the historical results interface can view the historical results data and the run log information,as shown
as figure 9.1:
Click to go into historical results view interface, can filter the results of history by date or
project, and can,And can printout etc., as shown as figure 9.2:
49 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Select one result and double click, the coming window will represent detailed information
and result picture of current sample. User can check patient information, reanalyze, save result,
print result and send result in this page as show as figure 9.3.
50 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Click to go into the Longview interface; user can open history running log, save the
current log, etc. as below:
51 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Select "Device" in the left column to go into the instrument operation and maintenance interface, as shown
below:
Note:After the first turn on, it request on this to initialize the machine.
Click can open the drawer,users need to manually close the drawer. This button is grayed
during test running, and it can not be manipulated.
Click can open rinse tip interface,as shown below. After setting the appropriate parameters
can rinse the tip.
52 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Note:Maintenance Engineer has been complete the calibration for the platform position,
non-professionals here can not do any changes for the platform position information.
53 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Select "System" in the left column to go into the instrument system parameter interface, as shown below:
54 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Here, you can create, delete, and modify the parameters of various types of liquid. In the parameters
interface you can set the detailed parameters of suction, liquid injection and rinse. These parameters setting
will affect the absorption, liquid injection precision, as well as the degree of sample cross-contamination.
So, these liquid parameters were tested and verified by experiment should be set to the default parameters
that cannot be altered.
55 / 64
ACROSS AUTO SYSTEMUSER MANUAL
You can create, delete, modify and save the Experimental projects in this interface.
1.2.1 Micro Gel Immunoassay Definition
Fig.
11.4 Blood Card information definition
Ref NO: Delimit the micro gel card number that the test project uses.
Product barcode:Delimit barcode rules of micro-gel card.
Card Name:Delimit the name of micro-gel card that is the name of the test project.
Test/Card :Delimit the number of detected specimens of such micro-gel card.
Cell Definition:Setting the holes mark of micro-gel cards and use“|”to block off them.
56 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Full in pipetting volume of reagents in ,select liquid parameters that used for
57 / 64
ACROSS AUTO SYSTEMUSER MANUAL
If the test requires centrifugal,fill in the centrifugation speed, centrifugation time that the test required in
the centrifugal parameters setting interface, and then click to add centrifugal parameters to the list.
You can also check the centrifugal parameters in the list and click the "Delete" button to delete them.
In this interface,you can create new reagent types and information, or to delete the existing reagent types.
As follow there are meanings of each reagent information in the formula bar:
Ref.NO:Reagents type number.
Reagent Name:Delimit reagents name.
Barcode:Delimit the barcode rules of reagents.
Volume(ml):Delimit the maximum capacity of this type of reagent bottles.
Type:To delimit the type of this reagent belongs.
Token Color:Set the color of the reagent displays in the interface.
Click ,fill in the appropriate information in the reagent information bar,and then click ,
you can complete the definition of an agent.
Selected a reagent in the left column and then click to delete this reagent.
58 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Default value is 1000,this parameter is called by system when device is washing probe
during initialization period.
59 / 64
ACROSS AUTO SYSTEMUSER MANUAL
The prompt option is shown as below, the options in this setting page are prompt for test
preparing or test running.
Centrifuge speed and centrifuge time should be set according to specification of gel card.
60 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Across System has excellent error-correction system, when there is error occur, system will
apply user to give order to correct current faulty movement and continue to work remaining test.
Note: During pipette Arm Z motor initialization, the blocking sheet failed to move out from
optical sensor.
61 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Note: During initialization, optical sensor doesn’t sense the blocking sheet.
62 / 64
ACROSS AUTO SYSTEMUSER MANUAL
If user needs to deal with this sample, user could move the cursor and left click any other
area in this hint window, the countdown will be aborted, after sample dealing, click “retry” to
continue test.
Example 2: If there is nothing detected or not enough liquid detected when probe is
detecting reagent, system will show this warning:
Click , probe will detect again, if enough liquid is detected, system will
continue to work, if no, this hint window will come out again.
63 / 64
ACROSS AUTO SYSTEMUSER MANUAL
Click , big gripper will skip this position and move to next card and catch.
Click , gripper will ignore this error and continue next step.
Click “OK”, big gripper will transfer the cards in centrifuge to original position and
continue next batch of sample.
64 / 64