Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

From sheet 1

Output terminal of solid no Replace solid relay


relay and relay K1 can or relay K1
work?

yes

no
Voltage of U1-2 is
Check inductor T1
> 5mV

yes

no
Voltage of U1-1
Check U1,R1 ,C4
is >2V?

yes

no
Base of triode Q1
Check R2,R6,C5
>0.7V?

yes

no
Base of triode Q1
Repalce triode Q1
>0V?

yes

Connet pin 1 (ssr-ctr) and pin3 (relay-ctr) on J8 to GND


make two control signals high level

To sheet 3

FLOW CHART 5.11 HEAT SYSTEM FAILURE (SHEET 2 OF 4)

I2 5-19
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

From sheet 2

yes
Base of triode Q1,Q3 Check triode Q2,Q3
is 0V? and resistor R8, R9

no

Is there +12V on yes


Replace triode
control terminal of solid
Q2 or Q3
relay and relay K1

no

no
Voltage U1-2 is
Check inductor T1
0V?

Yes

no
Voltage U1-1 is
Check U1,R1,C4
0V?

Yes

no
Base of triode Q1
Check R2, R6, C5
is 0V?

yes

no
Base of triode Q1
Replace triode Q1
is 5V?

yes

To sheet 4

FLOW CHART 5.11 HEAT SYSTEM FAILURE (SHEET 3 OF 4)

5-20
I2
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

From sheet 3

Connect power board and main board

Switch off during measuring

The pins on porer no


board J8-1, 3, 4 and Replace connecting wire
J2-1, 3, 4 on main
board works?

yes Element number from here is on


mainboard

Connect pins U8-2, 11 to VCC power

no
Pin of U8-9, 18 is
Check U8
5V?

Yes

Measure the voltage of U15-2 for self-checking

yes

no
Voltage is changing
from low level to high Replace U15
level?

yes yes
Still alarming after
Replace main board replacing U15

FLOW CHART 5.11 HEAT SYSTEM FAILURE (SHEET 4 OF 4)

I2 5-21
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

Power Failure Alarm

Is there AC power on No
Check outside power input
input socket?

Yse

No Check the connecting wire of


Initial trans for mer is
trans for mer and ower cord,
AC power?
and Fuse F1, F2, F3 on AC
entry board
Yes

No
Secondary output of
transformer is 9V, 15V? Replace transformer

Yes

On power board Pin 3 on U2, U3 is +5V? Check U2, U3 and prefix circuit

Yes

Pin4 of U4 is No
On power board Check U4 and prefix circuit
+12V?

Yes

Cross point of R17, No


On power board R18 is +2.5V? Check R17, R18

Yes
To sheet 2

FLOW CHART 5.12 POWER FAILURE ALARM (SHEET 1 OF 2)

5-22
I2
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

From sheet 1

Crosspoint of No
On power board R19, R20 is 2.5V? Replace R19, R20

Yes

Crosspoint of No
On power board R21, R22 is Replace R21, R22
2.1V?

Yes

No
Pins7, 1 of U6
On power board Replace U6
is high level?

Yes

No
Pin5 of U8 is low
On main board Check the connection of power board
level?
and main board

Yes

No
Pin4 of U8 is
On main board Replace U8
low level?

Yes

Replace main board

FLOW CHART 5.12 POWER FAILURE ALARM (SHEET 2 OF 2)