Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

Progress in Nuclear Energy 91 (2016) 302e309

Contents lists available at ScienceDirect

Progress in Nuclear Energy


journal homepage: www.elsevier.com/locate/pnucene

Observations and measurements of boric acid precipitation scenarios


Rodolfo Vaghetto*, Mason Childs, Ernie Kee, Yassin A. Hassan
Texas A&M University, USA

a r t i c l e i n f o a b s t r a c t

Article history: During a Loss of Coolant Accident (LOCA) in a Light Water Reactor (LWR), the safety injection system
Received 15 December 2015 injects borated water into the core. The continuous vaporization of the water in the core can increase the
Received in revised form concentration of boric acid in the core, to where the boric acid reaches the solubility limit and pre-
24 May 2016
cipitates. This becomes more likely when the liquid water supply is obstructed due to, for example,
Accepted 25 May 2016
debris accumulation. Questions have been raised regarding this and other ways in which boric acid
Available online 2 June 2016
precipitate in the core can affect the coolant flow. A testbed was designed and constructed for observing
the behavior of borated water under the combined effects of the boiling and the boric acid precipitation
Keywords:
Boric acid concentration and precipitation
(BAP) and for measuring boric acid concentration during different scenarios. The facility consisted of a
Core blockage transparent polycarbonate vertical pipe in which 41 heated rods were installed to enable the water to
Hot leg injection reach saturation temperature and to maintain a specified boil-off rate. The layout and geometry of the
experimental apparatus were designed to emulate a simplified core of a Pressurized Water Reactor
(PWR). The test scenarios varied with respect to thermal power, injection location, and test section
configuration. The observations and measurements conducted demonstrated that, even at concentra-
tions above the solubility limit, no appreciable deposition of the solid precipitate was possible, especially
in the regions at higher power density. Precipitate particles were observed to be relatively smaller than
the channels’ size, and no impact on the coolant behavior was observed. The rate of increase of con-
centration of boric acid was possibly slowed down by the loss of boric acid in entrained liquid droplets
through the exhaust. The results obtained provided a better understanding of the flow behavior in the
core in presence of high concentrations of boric acid and may be of immediate use for utilities and the
research community performing evaluations of the long-term core coolability.
© 2016 Elsevier Ltd. All rights reserved.

1. Introduction order to sustain the required reactivity margin during normal


operation, and to maintain the core at sub-critical during refueling
The Emergency Core Cooling System (ECCS) in a Pressurized and maintenance (Cho and Lee, 1998). During normal reactor
Water Reactor (PWR) is designed to provide the required coolant operation, the specified boron concentration is maintained by the
flow to remove the decay heat from the reactor core during a Loss of volume control system. In case of a LOCA, borated water is injected
Coolant Accident (LOCA). Three injection phases can be identified by the ECCS as previously described. Injection continues into the
during the accident. During the first phase of the LOCA, often same locations (cold legs) during the safety injection phase, and
identified as safety injection phase, the ECCS draws cold borated after the sump switchover (SSO), when water from the contain-
water from the Refueling Water Storage Tank (RWST) and injects it ment sump is recirculated through the primary system after RWST
into the reactor vessel through the cold legs. During this phase, liquid depletion (cold leg recirculation phase). This phase differs in
depending on the pressure of the primary system, cold borated that the coolant is now drawn from the containment sump
water from the accumulators might also be injected into the pri- compartment and recirculated back into the primary system. This
mary system. Boron (in the form of boric acid) is used as a soluble action must be maintained for several days in order to maintain
neutron absorber in the primary coolant. The main functions of adequate core cooling and reach a cold shutdown condition. Hot leg
boron are to compensate for fuel burnup and xenon poisoning in switchover (HLSO) is initiated as a manual action by the operators
at a time determined by plant design to ensure adequate core
* Corresponding author. Department of Nuclear Engineering, Texas A&M Uni- flushing and prevent or mitigate the potential effects of precipita-
versity, 3133 TAMU, College Station, TX 77801, USA. tion of boron in the core. Generally, this involves simultaneous
E-mail address: r.vaghetto@tamu.edu (R. Vaghetto).

http://dx.doi.org/10.1016/j.pnucene.2016.05.011
0149-1970/© 2016 Elsevier Ltd. All rights reserved.
R. Vaghetto et al. / Progress in Nuclear Energy 91 (2016) 302e309 303

injection through cold and hot legs. Understanding the local


behavior of the borated water under boiling and the possible effects
of the presence of boric acid in the solution is necessary for veri-
fying the manual operation procedures currently in place for LOCA
scenarios of different break sizes and locations. Of particular in-
terest would be scenarios where precipitation of boric acid occurs,
to study the effects of the precipitate on the water cooling capa-
bilities and the overall core coolability during the long-term cooling
phase. Acquiring this understanding could help the resolution of
Generic Safety Issue (GSI) 191, where the combined effect of debris
accumulation and boric acid precipitation can affect core
coolability.

2. Motivation

The industry and the research community have posed questions


on the effect of boric acid on the core and started conducting ex-
periments and analysis on these effects (Umminger et al., 2015;
Tuunanen et al., 1994). Properties of water (particularly viscosity
and density) are expected to change as the concentration of boric
acid increases (Hassan et al., 2015). Were the concentration of boric
acid to reach the solubility limit, it would then precipitate in the
core. The major concern during the long-term cooling phase of a
LOCA scenario is that solid precipitate remaining in the core can Fig. 1. Experimental setupdbeaker.
deposit or accumulate in different regions and affect the coolant
flow and the core coolability. Coolant flow can also be obstructed by
the presence of debris that has penetrated through the sump of boric acid in water at 100  C found in Crapse and Kyser (2011).
strainers, leading to an increase in the cladding temperature. Under The heater was turned on, and the solution was continuously
hypothetical debris-generated core blockage at the bottom of the stirred using a stainless steel stirring rod, allowing the dissolution
core, the coolant can reach the core only from alternative flow of the boric acid as the temperature increased to the water satu-
paths (usually from the top of the core (Vaghetto and Hassan, 2013) ration temperature. This phase of the experiment was recoded with
or lower elevations through the pressure relief holes (Crook et al., a camera placed in front of the experimental setup. Video and
2014)). Change (increase) in density can also change the natural pictures were taken throughout the entire experiment from
circulation patterns in the core and create a stratification where different angles.
denser borated water tends to collect at the bottom of the core, thus The mixture was observed boiling as the liquid level decreased
preventing colder water coming from the alternative flow path to to the point where the boric acid solubility limit was expected to be
reach the bottom of the core. reached. The first visible effect observed at this point was the for-
A very simple boil-off experiment using DI-water with boric acid mation of a very thin layer of precipitate at the free surface of the
was conducted to qualitatively observe the behavior of boiling liquid, as a result of the precipitation of the boric acid near the
water in the presence of high concentrations of boric acid and the vaporization surface. As the quantity of evaporated liquid
possible effects of precipitation on coolant flow patterns. The increased, small white particles of precipitate became visible and
experimental setup consisted of a glass beaker in which a solution easily distinguishable from the vapor bubbles (Fig. 2). These par-
of water and boric acid was left boiling. No liquid water was added ticles were fluctuating in the beaker but mostly entrained by the
during the test to compensate for the water boil-off. This produced rising bubbles toward the top section of the boiling liquid. At this
a continuous decrease of the liquid volume and subsequent in- time, water still appeared clear through the glass beaker.
crease of the boric acid concentration, until boric acid saturation As the evaporation continued, the formation of precipitate at the
was reached. The test was conducted (as shown in Fig. 1) with the surface and the entrainment of other precipitate toward the top of
aid of the liquid contributed to the formation of a growing layer of boric

▪ A 2-liter glass beaker


▪ A small water heater (max power 1 kW)

Temperature was continuously monitored during the experi-


ment using a class 1 k-type thermocouple connected to a Fluke®
52II thermocouple reader (0.3  C accuracy) and periodically
checked with an immerged thermometer.
The test solution was prepared by mixing in the beaker 275 g of
boric acid (Optibor® Orthoboric Acid, H3BO3) in 2 l of DI-water at
room temperature. This initial concentration (C0 ¼ 137.5 g/l) was
imposed in order to reach the saturation limit Cs ¼ 275 g/l when
approximately 1 l of water had evaporated.1 Cs is the solubility limit

1
Due to the boric acid carryover in the vapor, the estimation of the volume of
water to be evaporated to reach the desired concentration Cs is approximate. Fig. 2. Early stagedformation of precipitate.
304 R. Vaghetto et al. / Progress in Nuclear Energy 91 (2016) 302e309

acid on the liquid surface (Fig. 3a). The presence of a larger amount 2) The mechanisms of deposition of the precipitate (the large open
of precipitate in the liquid water increased the opacity of the water, space in the beaker might not allow precipitate to settle due to
as can be readily seen in Fig. 3b. As the process continued, larger bubble formation at the base of the beaker, in contrast to the
pieces of boric acid were observed to be released from the growing regions between vertical adjacent rods, where deposition might
layer, further affecting the water opacity (Fig. 3c). In a later phase, as occur).
the liquid level drastically decreased, the precipitate entrainment
developed the following cyclic behavior: These basic considerations motivated the design and con-
struction of the experimental facility discussed in the next sec-
▪ As the precipitate particles entrained by rising bubbles moved tion. While other experimental facilities have been used to study
toward the liquid top level, the water became clearer, and the the effects of the crystallization of boric acid on the core cool-
top layer started growing (Fig. 4a). ability (Umminger et al., 2015; Tuunanen et al., 1994), the main
▪ Water reached its maximum clarity, and larger pieces of pre- scope of the proposed experimental facility is to combine the
cipitate (approximately 0.5 cm in diameter) appeared. The top concentration measurements with a “real time” visualization of
layer appeared detached from the liquid level, perhaps due to the behavior of the boric acid solution and the precipitate
the increased pressure under the layer (Fig. 4b). particles.
▪ The top layer reached a maximum thickness and collapsed
suddenly, releasing a large amount of boric acid into the liquid
3. The precipitation and stratification test apparatus (PASTA)
water and rendering it opaque (Fig. 4c).

The Precipitation and Stratification Test Apparatus (PASTA) is


This was observed even at low quantity of liquid (approx.
designed for observing a boiling solution of water and boric acid
300 ml). The experiment was terminated after a few cycles were
and measuring boric acid concentrations in a test environment that
recorded.
replicates the key geometric properties of a typical PWR core.
The behavior of the water can be summarized as follows:
A detailed view of the facility, the test section, and the other
components is shown in Fig. 5a. The test facility consists of a
▪ When approaching its solubility limit, the boric acid was
polycarbonate pipe (which is analogous to the core barrel of a
observed to precipitate at the liquid free surface.
typical PWR design). Two polycarbonate flanges were welded
▪ Small precipitate particles were released and/or formed in the
to the cylindrical section to create the facility test section (Fig. 5a,
liquid water. The amount of the boric acid in the liquid phase
b, and c). Two stainless steel plates were fabricated that
could be judged by the water opacity. The increasing amount of
would allow the heated rods to be mounted in a given pattern
precipitate as the liquid level decreased did not seem to
(bottom plate) and would keep the rods parallel during the ex-
appreciably affect the overall liquid behavior in the beaker, even
periments (top plate). The top plate was also equipped with
during the later phase of the experiment.
venting holes to allow flow through the top region (Fig. 5c). It did
▪ At very high concentrations, a thick layer of precipitate formed
not seem necessary for the design of these plates to precisely
at the top of the liquid. This layer detached from the liquid
reflect that of the upper and lower plates of a PWR core. Forty-
volume as a consequence of the vapor overpressure at the bot-
five threaded holes were machined on the bottom plate to
tom of the layer, and its collapse released a large amount of boric
allow the placement of heater rods (Fig. 5e). The four corner
acid into the liquid. This cyclic behavior was observed until the
holes were used for instrumentation (two k-type thermocouples
end of the experiment.
placed to monitor the temperature inside the test section at two
▪ Large pieces of precipitate were observed to float during the
locations) and water sampling (two sampling ports shown in
later phase of the experiment, but no appreciable deposition of
Fig. 5d to collect samples of boric acid solutions for concentration
particles was observed at the bottom of the beaker, due to the
measurements). The thermocouples were connected to a Fluke®
turbulent motion of the rising bubbles.
52II thermocouple reader (Fig. 5a). One of the two sampling
ports was used as an injection port during bottom-injection
These observations prompted questions regarding the
experiments.
following:
The test section was connected at the top end to a polycarbonate
cylindrical extension where four ports equipped with valves were
1) The behavior of the solution under a different geometry and/or
installed (Fig. 5a). These ports were mainly used for injection of
heat configuration (heat from the bottom horizontal plate might
borated solution during the top injection experiment, simulating
induce different boiling/recirculation patterns and transport the
injection from the top of a PWR core. The heated rods (41 in total)
precipitate in ways different from heated vertical rod bundles in
were connected at the bottom section of the test facility. This
a typical LWR).
number of rods was selected also to allow a more uniform mixing in

Fig. 3. Middle stagedgrowing layer and reduced transparency.


R. Vaghetto et al. / Progress in Nuclear Energy 91 (2016) 302e309 305

Fig. 4. Late stagedcycling pattern.

the core expected in larger bundles (Tuunanen et al., 1994). The 3.1. Scaling, facility dimensions, and other characteristics
flow added to the test section through the injection ports (top or
bottom) was controlled by a metering pump (MityFlex 913) located The facility dimensions and other important features were
under the test section (Fig. 5a). The suction pipe of the pump was defined and calculated based on simple scaling requirements. The
immerged into a metered glass beaker (reservoir) containing the size of the test section enclosure (polycarbonate pipe) was driven
boric acid solution to be added during the experiment. The use of a by the core aspect ratio (the ratio between the diameter and height
glass beaker allowed the injection rate to be estimated during the of the core, Dcore/Hcore). For a typical PWR core, this ratio was
experiment, to compensate for the evaporation rate and maintain assumed equal to 0.78, which was reflected in the design of the test
the liquid inventory in the test section approximately constant. The section. Other engineering considerations (including polycarbonate
solution in the beaker was maintained at an approximately con- pipe and flange availability and total electric power to be installed)
stant temperature using a heater (mentioned in section 2). An led to a test section inner diameter of ID ¼ 15.24 cm (600 ID standard
immerged thermocouple connected to a secondary thermocouple polycarbonate pipe). The active height of the test section was then
reader was used to monitor the temperature of the solution in the defined to preserve the aspect ratio:
reservoir.
The top of the extension section was closed by a polycarbonate
Hcore
lid connected to the top flange. During the experiments, steam was Htest ¼ Dtest ¼ 19:38 cm: (1)
discharged through a 100 hose exhaust connected to the top lid and DCore
diverted upward toward a filtration system. The diameter of the electrically heated rods was selected to
The heated rods were connected to power supplies that could reproduce that of a typical PWR fuel pin. The diameter of the rods
be powered independently, in anticipation of future work that installed in the facility was Drod ¼ 0.95 cm. The free volume of the
might test the effect of different radial power distributions. Each test section (volume occupied by the coolant) was
power supply was equipped with a gauge to monitor its power Vtest ¼ 2.97103 m3 (2.97 l). The total heat transfer area was
output. The experiments used a uniform power configuration Atot ¼ 0.24 m2.
where the total electric power was uniformly distributed among The total power to be installed in the experimental facility was
the heater rods. estimated based on the following:

Fig. 5. PASTA facility: a. Overview; b. Test section; c Top view of the test section; Sampling port; e. Electric rod.
306 R. Vaghetto et al. / Progress in Nuclear Energy 91 (2016) 302e309

1) The reactor power is the decay power at the time of the sump using DI water to verify the validity of the calibration method, by
switchover for a Double-Ended Guillotine break for a typical confirming that the original mass was recovered after the drying
PWR system. The sump switchover time can be assumed to be procedure, and to estimate the drying time.
approximately 30 min from the break event. The decay power at After the samples were let dry, the trays for each calibration
the time of switchover was estimated to be 75 MW. solution were again weighed using the Acculab® VI-350 (0.01 g
2) Two scaling parameters were considered: reactor power density accuracy) scale. The average mass of boric acid from the 10 samples
Pd (W/m3), defined as the ratio of the reactor power to the total was calculated for each calibration solution, along with the stan-
core free volume (volume occupied by the coolant); and the heat dard deviation, and is recorded in Table 2.
flux q” (W/m2), defined as the reactor power per unit of core Samples from the calibration solution were collected and dried
heat transfer surface area. The reactor reference values used for to provide concentration measurements. Uncertainty was calcu-
this evaluation were Pd ¼ 3.7 MW/m3 and q” ¼ 11.5 kW/m2, both lated as 2s for the 10 samples of each calibration solution.
estimated assuming a reactor power equal to decay power In Fig. 6, the mean dried mass for each calibration solution is
described above. plotted against the expected mass for a 16 ml sample (no loss, solid
line).
The reference power to be installed in the facility was fixed at From the dry mass of the samples, it was determined that
the greater of the following: the power required to preserve the approximately 15.2% of the boric acid in solution was lost from each
heat flux and that required to preserve the power density: sample during the drying process. The losses were found to follow a
linear relationship. The linear relationship results in a multiplica-
Ref
Ptest ¼ maxðq00 Atest ; Pd Vtest Þ ¼ 11 kW: (2) tion coefficient of 1.18 as a fraction of grams in solution to meas-
ureable grams after drying, which was used to determine the
Based on the characteristics of the electrically heated rods concentration of boric acid in DI water of samples taken from the
readily available, rods of 500 W nominal power were selected. This PASTA facility during experimentation.
Inst ¼ 20:5 kW, enabling the test
provided a total installed power Ptest
section to support higher power if required by future experiments.
The dimensions of the test facility and other key parameters are 5. Experimental procedure
summarized on Table 1.
Although the test rod diameter was representative of a typical Each test was conducted according to written procedures which
PWR pin, the rod bundle pitch/diameter ratio, and consequently were developed and refined during the facility shakedown.
the center-to-center pin distance, differed from the reference value The facility was initially loaded with a solution prepared with
(1.33 for a typical PWR fuel assembly), due to the shape and size of 660 g of boric acid (Optibor® Orthoboric Acid, H3BO3) in 2 l of warm
the rod heads (1.27 cm threaded head, Fig. 5e) and its mounting DI-water (approximately 50  C). This initial solution was poured
configuration at the bottom plate. into the test section from the top. An additional quantity (0.5 l) of
DI-water at the same temperature was used to clean the boric acid
accumulated at the top flange. When the test was initiated, the final
4. Boric acid concentration measurementsdcalibration and solution in the test section contained 660 g of boric acid in 2.5 l of
error estimation DI-water (C0 ¼ 264 g/l). This concentration was chosen to be
slightly below the solubility limit of boric acid in water at 100  C
The following methodology determined the concentration of (Cs ¼ 275 g/l), so that the saturation could be reached shortly after
the boric acid in solution inside the experimental facility. Drying the water started boiling and evaporating from the test section. The
trays were prepared with absorbent paper sheets and then total initial volume corresponded to a liquid level right above the
weighed. The experiment was then run, after which the liquid top plate. The temperature of 50  C for the initial solution allowed
sample containing boric acid was collected in the tray and dried the dissolution of most of the boric acid, limiting thermal stresses
until only boric acid remained. The mass of the boric acid was to the test facility (which was usually slowly pre-heated). The
determined by weighing the tray after drying and subtracting the heated rods were turned on when the solution in the test section
predetermined weight of the tray with the paper sheets. was quiescent (Fig. 7), with the excess of boric acid deposited at the
To perform the calibration, five solutions of known boric acid bottom of the test section.
concentrations were prepared. A solution of 100 g/l boric acid in DI The injection solution was prepared using similar techniques
water was prepared at 100  C. From this calibration solution, a described above, with a concentration of Ca ¼ 60 g/l. This concen-
16 ml sample was collected and immediately transferred to a dry- tration is not prototypical for any plant conditions and was selected
ing tray. The tray was placed on a surface to dry at a constant arbitrarily based on solubility of boric acid at the preparation
temperature of 50  C. This process was repeated to produce 10 temperature. Additional prepared solution was added to the
samples for error estimation. The same procedure was applied for reservoir beaker when the remaining solution reached the 300 ml
calibration solutions of 250 g/l, 275 g/l, 280 g/l, and 300 g/l of boric mark. The addition of each batch was recorded during the
acid in DI water at 100  C. Control samples were also prepared experiment.
Periodically throughout the experiment, 16 mL samples were
collected from the port at the bottom of the facility test section,
Table 1
Main features of the experimental facility.
Table 2
Parameter Value Unit Expected mass, mean dried mass, and uncertainty of calibration solutions samples.
Test section inner diameter 15.24 cm
Concentration (g/L) Target mass per 16 mL (g) Mean dried mass (g)
Test section height 19.38 cm
Test section diameter/height ratio 0.785 e 100 1.60 1.49 ± 0.06
Test section free volume 2.97 l 250 4.00 3.41 ± 0.10
Rod diameter 0.95 cm 275 4.40 3.70 ± 0.13
Rod bundle pitch/diameter ratio 1.75 e 280 4.48 3.72 ± 0.12
Power installed 20.5 kWe 300 4.80 4.11 ± 0.17
R. Vaghetto et al. / Progress in Nuclear Energy 91 (2016) 302e309 307

5 to induce greater mixing in the test section solution, which would


create a more uniform concentration throughout the facility.
No Loss
Opening the facility lid for the duration of an experiment was
4 Mean dried mass (g) hypothesized to decrease the overall concentration of the solution
Dried weighed mass (g)

in the facility, due to additional losses in liquid droplet entrainment


Linear (Mean dried mass (g))
through exhaust. The rapid boiling of the facility solution was also
3 thought to cause splashes of solution out of the facility which could
y(x) = 0.848x
also entrain boric acid out of the facility. Each of the hypotheses was
tested.
2

1
7. Analytical solution

0 The rate of change of mass of the boric acid in solution can be


0 1 2 3 4 5 analytically calculated by solving the mass conservation equation
Target Sample Mass (g) expressed in Eq. (3).

Fig. 6. Boric acid dry mass against expected mass.


dMB
¼ m_ in  m_ out ; (3)
dt
while recording the time of collection. Each sample collected was In Equation (3), the injection mass rate, m_ in (g/s), can be deter-
dried in the same manner as described above. From the samples’ mined by the concentration, Ca (g/l), and constant volumetric flow
dried mass, the concentration of the solution in the test section rate, Qin (l/s), of the solution that was injected into the facility as
facility was determined with associated uncertainty. Concentration
of the solution in the test section was plotted as a function of m_ in ¼ Q in Ca : (4)
elapsed time since the initial injection from the reservoir. Test
repeatability was confirmed during the facility shakedown by The boric acid loss rate, m_ out (g/s), was calculated from the
repeating the test several times under the same test conditions, calibration data to determine the loss of boric acid due to evapo-
while samples were collected. ration, El (g/L). The fractional loss was calculated for each concen-
tration of boric acid per the sample collected. Solution volume in
6. Test matrix and conditions the test section was held constant, which permitted the assumption
that the volume of solution injected was equal to the volume of
To determine differences in the operation of the experiment, solution evaporated, Qout ¼ Qin. These assumptions allowed for a
various features were altered for full experiments and compared mass loss rate to be determined as
with one another. Features that were altered included the sealing of
the top lid and injection location (from either the top or bottom of m_ out ¼ Q in El : (5)
the facility). Other operational parameters were maintained unal-
If the loss of boric acid due to evaporation is neglected, the rate
tered during the experiments, including the total power applied,
of change of boric acid can be approximated as
radial power distribution, concentration of injected solution, and
initial concentration of facility solution. Table 3 summarizes the
dMB
conditions of the experiments. ¼ m_ in ; (6)
Test 1 was used as a baseline, as it presented the easiest setup dt
and operation of the facility. Injection from the top of the facility Equations (3) and (6) were divided by the test section volume,
was hypothesized to create a gradient of boric acid concentration in Vtest to express the rate of change in boric acid concentration. The
which the bottom of the test section near the sampling port would analytical solution of (3) and (6) was compared to the concentra-
be lower than near the top. Injection from the bottom was thought tion of the test section as determined from the samples collected.

Fig. 7. Test preparationdinjection of the initial solution (660 g of H3BO3 þ 2.5 l of DI-water).
308 R. Vaghetto et al. / Progress in Nuclear Energy 91 (2016) 302e309

Table 3
Operational parameters of experiments.

Test number C0 (g/L) Ca (g/L) Total power (kW) Power profile shape Lid status Injection orientation

1 264 60.0 4 kW Flat Closed Top


2 264 60.0 4 kW Flat Open Top
3 264 60.0 4 kW Flat Closed Bottom

8. Results in the facility. The determined concentration of the solution was


plotted as a function of time after the injection was initiated for
After the electric heaters were turned on, the temperature was each experiment. Equations (3) and (6) were used to calculate the
found to increase until reaching the boiling point. The temperature theoretical concentration in the facility and were plotted as a
achieved and the high turbulence in the test section due to the function of time according to the injection rate during experi-
bubble formation and migration to the top produced a very rapid mentation. The analytical approximations of facility concentration
dissolution of the entire quantity of boric acid previously deposited along with the experimental data from various experiment runs are
at the bottom of the test section. When complete dissolution was plotted in Fig. 10.
achieved, no visible traces of boric acid between rods or in the Experimental data indicate a linear trend for the facility solution
region between the outer rods and the test enclosure were concentration, which is between the bounds of the analytical so-
observed. At this point, the metering pump was turned on and set lution (Eqs. (3) and (6)). The lower rate of change of the boric acid
to the rate needed to compensate for the boil-off from the test concentration in the test section compared to Eq. (6) is an indica-
section for the selected power. The setting of the metering pump tion of a loss of boric acid from the test section. These losses are
was defined during previous shakedown testing and manually fine- hypothesized to be due to entrainment of liquid droplets out the
tuned during the experiment. The observations conducted showed exhaust vent. The losses appear less significant than what was
the following main features: expected from the drying method. Boric acid, however, is certainly
lost from the facility, as the collected data are well below the
▪ Water remained clear during the early stage of the experiment, lossless approximation (Eq. (6)). The vapor was not collected or
even after the solubility limit was expected to be achieved analyzed for boric acid content to determine concentration. No
(Fig. 8a). appreciable difference can be deduced from the comparison of the
▪ Small visible particles started appearing in the solution but three tests conducted, indicating that the injection location and the
mainly entrained or maintained in suspension by violent bub- presence of the test section lid did not affect the rate of change of
bles departing from the surface of the rods, preventing any the concentration. The concentration of the solution in the test
deposition at this time. section was found to be greater than the concentration of satura-
▪ As the mass of boric acid in the liquid increased, the turbidity of tion, Cs, for the majority of the experiment. The high concentration
the water visibly increased while a layer of boric acid on top of of boric acid in the facility did not have a visible effect on the flow
the mixture level was observed to form and grow (Fig. 8b). characteristics of the boiling, even when particles of boric acid were
visible throughout the test section.
At very high concentrations, bigger particles (approximately
0.5 mm in diameter based on observation of the particles at the test
9. Discussion
section wall visible in Figs. 8b and 9) were observed while solid
precipitate started accumulating at the following regions inside the
An experimental apparatus was constructed to conduct exper-
test section:
iments with water and boric acid, to observe and study the flow
behavior, and measure the boric acid concentration in boiling wa-
▪ The rod bundle periphery, between the outer rods and the en-
ter. The following phenomena were observed during the
closure’s wall
experiments:
▪ The test section walls.

 High concentrations of boric acid and the presence of solid


Each collected sample was dried, and the measured mass of
precipitate did not appreciably affect the flow behavior in the
boric acid was used to determine the concentration of the solution
test section, which was mostly driven by the vapor bubbles.

Fig. 8. Experimental results (a. early stage; b. middle stage).


R. Vaghetto et al. / Progress in Nuclear Energy 91 (2016) 302e309 309

 The injection location and the presence of the lid during ex-
periments did not seem to affect the overall concentration in-
crease in the test section of the facility.
 The comparison with the analytical solution confirmed that a
fraction of the boric acid was lost from the test facility. This loss
was essentially related to the liquid droplets entrained in the
vapor phase (Bo €hlke et al., 2008). The fraction released
contributed toward reducing the rate of increase of boric acid
concentration in the solution.
 The fraction of boric acid lost from the test section was found to
be lower than that estimated during the calibration procedure.
This might be essentially due to condensation in the upper
plenum and through the test exhaust and the subsequent reflux
of the liquid into the test section.

Although crystallization of boric acid and deposition of boric


acid was clearly visible during the experiments, no impact on the
Fig. 9. Experimental results (end of the test).
flow in the test section was observed. In particular, the boric acid
precipitation did not produce any blockage of the flow channels. As
 Violent bubble generation, departure, and migration to the top the blockage occurrence is affected by different parameters
of the test section prevented deposition of precipitation be- (Tuunanen et al., 1994), conditions for blockage were not achieved
tween the rods and along the heated test section. In particular, in the experiments conducted.
no appreciable deposition of solid precipitate was observed
between the rods inside the bundle or on the rods’ surface. Nomenclature
 Change in water properties caused by increased concentration
did not caused stratification. Instead uniform flow mixing was A Surface Area
observed, allowed by the boiling in the test section. BAP Boric Acid Precipitation
 Deposition of the precipitate occurred in regions of lower power C0 Initial Concentration
density (the rod bundle periphery) or at “colder” surfaces (the Ca Concentration of Added Solution
test section walls). Cs Concentration at Saturation (Solubility Limit)
D Diameter
DI De-Ionized
10. Conclusions ECCS Emergency Core Cooling System
El Loss due to Evaporation
The conclusions related to the measurements of the boric acid H Height
concentration can be summarized as follow: HLSO Hot Leg Switchover
LOCA Loss Of Coolant Accident
 Concentrations of boric acid were found to increase linearly LWR Light Water Reactor
with time reaching the saturation limit very early in the ex- PWR Pressurized Water Reactor
periments, when no major effect on the flow was observed. Qin Experimental Volumetric Flow Rate
 Due to the uniform mixing in the test section, no appreciable RWST Refueling Water Storage Tank
difference in the concentration trend was observed between the SSO Sump Switchover
two injection locations selected (top and bottom injection). V Volume
Uniform distribution of the precipitate particles when visible
was observed. This proves that the concentration of boric acid is References
maintained uniform in the test section due to the high mixing, €hlke, S., Schuster, C., Hurtado, A., 2008. About the volatility of boron in aqueous
Bo
similarly to what was found in other experiments (Tuunanen solutions of borates with vapour in relevance to BWR-Reactors. In: International
et al., 1994). Conference on the Physics of Reactors Nuclear Power: a Sustainable Resource,
Switzerland, September 14e19.
Cho, H.R., Lee, S.K., 1998. Establishment of a procedure to prevent boron precipi-
450
tation during post-LOCA long term cooling for WH 3-loop NPPs. J. Korean Nucl.
Test 1 Soc. 30 (I), 47e57.
Test 2 Crapse, K.P., Kyser III, E.A., 2011. Literature Review of Boric Acid Solubility Data.
400 Test 3 Savannah River National Laboratory Report, SRNL-STI-2011e00578.
H3BO3 concentration (g/L)

Eq. (3) Crook, T., Vaghetto, R., Vanni, A., Hassan, Y.A., 2014. Sensitivity analysis of a PWR
Eq. (6) response during a loss of coolant accident under hypothetical core blockage
350 scenario using RELAP5-3D. Paper 31272. In: Proceedings of the 2014 22nd In-
ternational Conference on Nuclear Engineering ICONE22 July 7-11, 2014, Pra-
gue, Czech Republic.
300 Hassan, Y.A., Osturk, S., Lee, S., 2015. Rheological characterization of buffered boric
acid aqueous solutions. Prog. Nucl. Energy 85, 239e253.
Tuunanen, J., Tuomisto, H., Raussi, P., 1994. Experimental and analytical studies of
250 boric acid concentrations in a VVER-440 reactor during the long-term cooling
period of loss-of-coolant accidents. Nucl. Eng. Des. 148, 217e231.
Umminger, K., Schoen, B., Schollenberger, S.P., 2015. Conclusions on boron precip-
200 itation following a large break loss of coolant accident. In: International Topical
0 10 20 30 40 50 60 70 80 Meeting on Nuclear Reactor Thermal Hydraulics, Chicago, IL, August 30 e
Time (m) September 4.
Vaghetto, R., Hassan, Y.A., 2013. Study of debris-generated core blockage scenarios
Fig. 10. Experimental and analytical boric acid concentrations. during loss of coolant accidents using RELAP5-3D. Nucl. Eng. Des. 261, 144e155.

You might also like