Professional Documents
Culture Documents
426 1466 1 PB
426 1466 1 PB
Keywords: Android apps, diabetes, mobile medical care, near‑infrared spectroscopy, noninvasive
blood bilirubin monitoring, noninvasive blood glucose monitoring, optical methods, telemedicine,
wireless sensors
126 Journal of Medical Signals & Sensors | Volume 8 | Issue 3 | July - September 2018
Javid, et al.: Blood glucose and bilirubin monitoring
Journal of Medical Signals & Sensors | Volume 8 | Issue 3 | July - September 2018 127
Javid, et al.: Blood glucose and bilirubin monitoring
food was measured using NIR glucose sensor and available the bilirubin content and the second to the oxyhemoglobin
glucometers in the market. It measured diffused reflectance content. The proposed circuit was a dual‑channel circuit.
spectra of the forearm. The signal amplitude decreased after Each of the channels included an LED, an analog circuit for
meal comparing with before meal.[1] Guo et al. in 2015 signal processing, and an analog‑to‑digital converter. The
presented a new noninvasive blood glucose monitoring data from both channels were sent to the PC and processed
method based on four NIR spectra and double artificial by the LabView software (National Instruments). In the
neural network analysis. They used 820, 875, 945, and test phase, the ABL 835 was used to measure bilirubin as
1050 nm wavelengths and have recorded the transmission a reference. The test was performed on seven subjects from
photoplethysmogram signal for four fingers simultaneously.
3 days to 42 years of age. The error level in measurements
After collecting and preproccessing data, a module creates
reached 3.95%.[29] Osman et al. presented a noninvasive
an estimation model that utilizes the wavelet transform and
method for bilirubin measurement in 2014. In this research,
artificial neural network to create this model. Experiments
showed that the RMSE of the prediction is between 0.97 light with a wavelength of 455 nm is transmitted to the
and 6.69 mg/dL and the average of RMSE is 3.80 mg/ sample. To prepare sample, the Sprague–Dawley (SD) rat’s
dL.[24] Tamilselvi and Ramkumar in 2015 proposed a skin is shaved and soaked for 3 min at different artificial
framework for noninvasive blood glucose measurement bilirubin standard solution (ABSS) concentrations. The
using NIR spectroscopy with 940‑nm wavelength. At the reflected light was received by the photodiode, and the
start of experiment, the Global Positioning System module output voltage was measured by the voltmeter. In the next
sends the user’s position coordinates with the message step, the relationship between the received voltage and the
to the physician. A Global System for Mobile modem ABSS concentration was obtained using the MATLAB
is used to send a message through any commercially (MathWorks, Inc) software. The output voltage decreased
available SIM card. However, the proposed method has when the concentration of ABSS increases. In the final
not been evaluated.[25] Pande and Joshi in 2015 in their circuit, the received voltage by the photodiode was sent
article presented a circuit for noninvasive measurement to the Arduino and has been processed to determine the
of the glucose. They concluded that the transmitter with bilirubin concentration. The digitized signal by Arduino was
1450‑nm wavelength is more suitable.[26] Bobade and Patil sent to the PIC microcontroller. At this stage, the digital
in 2016 presented a noninvasive method for measuring signal was processed and the bilirubin level was displayed
blood glucose in diabetics and nondiabetics. In this study,
on the LCD. The degree of jaundice was also determined by
the light with a wavelength of 940 nm is emitted to the
the LED. The method has not been evaluated.[12] Ali et al. in
patient’s finger. The transmitted light is measured. After
2015 presented an optical technique for detecting jaundice
processing the received signal, the results are displayed
on the liquid‑crystal display (LCD) while being able to without using a blood test. In their technique, the blue
send them to the Android app by Bluetooth. Experiments light was emitted to the sample and the light reflected was
are carried out with 48 samples in two phases of fasting captured by the photodiode. In this research, the mock skin
and nonfasting states. The results indicate that there is a of SD rat has been shaved and soaked into fix calibrated
correlation between the intensity level after the transmission bilirubin concentrations and was used to obtain a relationship
and the glucose level. However, the method has not been between the device voltage and the bilirubin concentration.
evaluated.[27] The reflected light produced a voltage. This voltage was
processed by the Arduino, and the result was displayed
The second part of literature review is assigned for the
on the LCD. Arduino was communicated with the Visual
bilirubin measurement. Baharuddin et al. in 2010 presented
a cost‑effective and portable bilirubin measurement device. Basic software (Microsoft) to store results in the SQLyog
In the proposed system, the visible light in the range of database. Results were displayed on the website. The doctor
380–760 nm has been emitted to the blood serum and the and nurses could monitor the babies jaundice level through
reflected light has been measured. The sample which is a online system.[11]
yellow liquid with different concentration has been placed in In this paper, in addition to combining devices for
a black box. The outputs of system consisted of LCD display, measuring glucose and bilirubin, both transmittance
red and green LED, and a buzzer to show normal, mild, and
and reflectance modes have been investigated. Android
critical conditions. The output voltage of the device decreased
application for processing and sending information has also
with increasing bilirubin concentrations. The method has not
been developed.
been evaluated on the body.[28] Penhaker et al. presented
an article in 2013 in which they described the design of an Methods
electronic device for measuring bilirubin by optical method.
They used two LEDs with the wavelengths of 455 and 575. In this section, implementation of our proposed system
The concentration of bilirubin was found from the difference for noninvasive glucose and bilirubin measurement using
in their absorbance. The first LED corresponded mainly to optical method is explained.
128 Journal of Medical Signals & Sensors | Volume 8 | Issue 3 | July - September 2018
Javid, et al.: Blood glucose and bilirubin monitoring
Hardware configuration 457 nm and 473 nm.[11] Therefore, in this study, light with
a wavelength of 470 nm was used to measure bilirubin.
The NIR spectroscopy is used to measure the blood
Furthermore, an optical receiver in the range of 330–720 nm
glucose, and an optical method incorporating the visible
was proposed to receive an optical signal.
light is selected to measure the amount of bilirubin.
The wavelength selection of light is very important for The circuit for our proposed noninvasive blood glucose and
optical methods. As mentioned in Introduction Section, bilirubin monitoring system is shown in Figure 3.
three bands exist in the NIR range; combination In our system, we propose a suitable microcontroller,
overtone band (2000–2500 nm), first overtone LEDs, transmitter and receiver, and wireless and mobile
band (1400–2000 nm), and second or higher overtone capabilities. The AVR ATMega32 microcontroller is used
band (750–1400 nm).[8] to control the transmitters and the receiver which are
Glucose has maximum absorption at wavelength of 939, connected to Port D pins of this microcontroller. The Pande
970, and 1197 nm in higher overtone regions; wavelength of and Joshi[22] circuit is added for filtering, amplification, and
1408, 1536, 1688, and 1925 nm in the first overtone region; signal isolation. This circuit includes a 1 μF capacitor that
and wavelength of 2100, 2261, and 2326 nm in combination is a couplage capacitor and remove DC [Box A, Figure 3].
region.[1] In this study, NIR light with wavelength of Then, an op‑amp creates a low‑pass filter to remove
940, 1550, and 1650 nm is used to measure glucose. The the high‑frequency noise and the probable leakage
wavelengths 1550 and 1650 nm have maximum absorption currents [Box B, Figure 3]. In the next step, the second
of glucose. Although glucose absorption at wavelength 1 µF capacitor with 68 kΩ resistance creates a high‑pass
940 nm is lower compared to the first overtone and filter to remove the low‑frequency noise [Box C, Figure 3].
combination overtone, it has minimum optical signal The second op‑amp amplifies the signal and it has created
attenuation by other components such as water. Therefore, a low‑pass filter to probable leakage currents [Box D,
the intended depth of penetration could be achieved at this Figure 3]. Finally, the transistor decreases the result on the
wavelength.[1] An infrared receiver in the range of 800– loading effect of microcontroller. It can isolate the circuit
output from the microcontroller circuit [Box E, Figure 3].
1700 nm was used to receive the optical signal. In the same
circuit, another LED for bilirubin measurement is added. The circuit output is connected to pin A0 which are
The bilirubin will absorb wavelength of light between analog ports. Microcontroller converts the analog signal to
Figure 3: The proposed system for non‑invasive blood glucose and bilirubin measurement
Journal of Medical Signals & Sensors | Volume 8 | Issue 3 | July - September 2018 129
Javid, et al.: Blood glucose and bilirubin monitoring
130 Journal of Medical Signals & Sensors | Volume 8 | Issue 3 | July - September 2018
Javid, et al.: Blood glucose and bilirubin monitoring
in the option selection screen. These buttons include Figures 9 and 10 display the test on glucose and the
(i) viewing the glucose or bilirubin chart, (ii) sending SMS artificial bilirubin solutions in the transmittance mode.
or e‑mail to doctor, (iii) sending SMS or E‑mail to family,
To obtain the relationship between the voltage and the
and (iv) logging out. Figure 8 (i) shows the screen for
solution’s concentrations, a scatter diagram was drawn and
drawing glucose or bilirubin chart.
the relationship were obtained by the linear regression.
Results Figures 11 and 12 show the graphs. From the results, it can
be inferred that the output voltage increases by increasing
In this study, two types of tests have been conducted to the concentration of the glucose solution and decreases by
evaluate the system: (1) in vitro experiments on the glucose increasing the concentration of artificial bilirubin solutions
solution and (2) in vivo experiments on the human body. In in the transmittance mode.
the following, each of these tests will be described.
In the reflectance mode, the transmitters and the receiver
In vitro experiments were placed on the one side and the cuvette on the other
At this stage, 22 solutions with different glucose side, which is different from the transmittance mode. Here,
concentrations in the range of 0–800 mg/dL were obtained we only measure the reflected light, but in transmittance
by dissolving glucose or dextrose monohydrate powder mode, we measure the signal passed through the solution.
in distilled water. Moreover, 15 solutions with different Figures 13 and 14 display the experiment in the reflectance
concentration in the range of 2–30 ml/dL were obtained mode.
from dissolving methyl red powder in acetic acid to test To obtain the relationship between the voltage and the solution
bilirubin. The range of wavelength peak absorption in concentrations, a scatter diagram was drawn and the relationship
this solution is similar to the real bilirubin serum, and it
was obtained by linear regression. Figure 15 and 16 show
provides a good alternative to the bilirubin serum. After
the graphs. It can be inferred from the results that the output
preparing all the desired concentrations, each solution
voltage decreases by increasing the concentration of glucose
was poured to a cuvette. Cuvette is a container used for
solution and increases by increasing the concentration of
spectroscopic experiments, in which a beam of light is
artificial bilirubin solution in the reflectance mode.
passed through the sample to measure the absorbance or
transmittance. Experiments were conducted in transmittance It was observed that the relationship between the
and reflectance mode in the darkroom to decrease other concentration and the output voltage in the transmittance
beams of light. A power supply of 9 V voltage was used mode is the inverse of the relationship between
to provide the circuit’s input voltage. The test was repeated the concentration and the output voltage in the reflectance
twice for each solution, and the output voltage was mode. This is because, by increasing the transmitted light,
recorded. In transmittance mode, the cuvette containing the the reflected light decreases. In the following section, the
solution was placed between the transmitters and receiver. experiments performed on the body are mentioned. It was
Journal of Medical Signals & Sensors | Volume 8 | Issue 3 | July - September 2018 131
Javid, et al.: Blood glucose and bilirubin monitoring
a b c
d e f
132 Journal of Medical Signals & Sensors | Volume 8 | Issue 3 | July - September 2018
Javid, et al.: Blood glucose and bilirubin monitoring
Journal of Medical Signals & Sensors | Volume 8 | Issue 3 | July - September 2018 133
Javid, et al.: Blood glucose and bilirubin monitoring
battery was used. In transmittance mode, the finger is All the steps that have been executed in the transmit mode
placed between the transmitter and the receiver. Figure 17 were repeated in the reflectance mode. Figure 22 shows the
shows the appropriate body configuration for experiment’s scatter diagram for 19 training data.
transmittance mode.
After entering the obtained regression formula to the
The blood glucose of 19 training data was measured using Android software [Figure 23], the blood glucose of five
the glucometer. The sensor output voltage of sensor for these test data was obtained. Figure 24 shows the diagrams and
data was recorded. To obtain the relationship between the relationships obtained for the 940, 1550, and 1650 nm
voltage and the glucose concentrations, scatter diagram was wavelengths in the reflectance mode.
used by producing linear regression line. Figure 18 shows
Another option that has been investigated here is the
the relationship between the blood glucose concentration
calculation of the final result’s average obtained in the
and the output voltage on the body in transmittance mode.
transmittance mode and the reflectance mode. It is shown
The obtained formula from Figure 11’s linear regression in Eq. 4 where t is glucose concentration obtained from
was selected and entered in the Android application. The transmittance mode and r is glucose concentration obtained
formula converts the voltage to its corresponding glucose from reflectance mode. This will help reduce the impact of
concentration. Android code is shown in Figure 19. noise data. Table 1 represents the results obtained from the
invasive and noninvasive methods. The percentage error was
Then, to evaluate the proposed device, five people’s
calculated using Eq. 5. The RMSE was calculated by Eq.
blood glucose was measured with the designed sensor in
6, where xi is the estimated blood glucose by noninvasive
the transmittance mode. To study the behavior of each
method, yi is the measured glucose concentration by
transmitter in transmittance mode, the scatter diagram
invasive method, and n is number of samples.
was drawn and the relationship between the voltage and
the concentration was obtained by the linear regression. Average glucose concentration = average (t , r ) (4)
Figure 20 shows diagrams and relationships for 940, 1550, Percentage error (%) =
and 1650 nm transmitter in the transmittance mode.
[6]
(5)
|Glucose Levelinvasive − Glucose Levelnon-invasive |
To study the reflectance mode, the transmitters and the Glucose Levelinvasive
× 100%
receiver were placed on the one side of the finger. Figure 21
shows the finger’s configuration in the reflectance mode.
∑
n
RMSE = i =1
( xi − yi ) 2 (6)
n
Figure 18: The relationship between the blood glucose concentration and Figure 19: Android codes to convert voltage to glucose concentration in
the output voltage on the body in transmittance mode transmittance mode
Table 1: The average glucose obtained from the combination of the transmittance mode and the reflectance mode
Percentage Glucose concentration measured by noninvasive method using (mg/dl) Glucose concentration Sample
error calculation the average of final results obtained in transmittance mode and measured by invasive
reflectance mode method (mg/dl)
12.3377 67.5 77 1
10.1974 167.5 152 2
3.57143 217.5 210 4
11.2069 322.5 290 6
4.0146 427.5 411 7
134 Journal of Medical Signals & Sensors | Volume 8 | Issue 3 | July - September 2018
Javid, et al.: Blood glucose and bilirubin monitoring
Figure 20: The relationship between the glucose concentration and the output voltage for 940, 1550, 1650 nm transmitters in transmittance mode
Journal of Medical Signals & Sensors | Volume 8 | Issue 3 | July - September 2018 135
Javid, et al.: Blood glucose and bilirubin monitoring
Furthermore, the scatter diagram was drawn for 19 training By calculating the average of results obtained in the
data in the reflectance mode. By linear regression, the transmittance and reflectance mode after weighing,
relationship between the glucose concentration and the the final results were obtained for the five test data
voltage was obtained. The blood glucose of five test data [Table 2].
was calculated in the reflectance mode using the new
formula. Figure 28 displays the relationship between the As you can see, the results are not improved. The average
glucose concentration and the voltage after the weighing in percentage error in this mode is 28.4 and RMSE is 65.12.
the reflectance mode. This is because when the light passes through the body’s
tissue, the optical wavelength that receiver absorbs may
vary with the transmitter’s wavelength. Therefore, the
possible changes due to the wavelength and wave velocity
and the side effects of the testing case can cause errors.
Consequently, by accepting this error, the results of the
weighting are reported.
Glucose measurement could be affected by the intervention
of physiological parameters such as variation in body
temperature, chemical parameters, triglyceride and albumin
levels, and intervention of environmental variations such
as changes in temperature, humidity, carbon dioxide,
and atmospheric pressure, which are the drawbacks of
NIR spectroscopy.[8] In this research also, due to the
low sensitivity of the receiver and because of their high
Figure 23: Android codes to convert the voltage to glucose concentration expenses, each transmitter does not have a separate
in reflectance mode receiver. The patient’s hand tremor, skin thickness, and not
Figure 24: The relationship between the glucose concentration and the output voltage for 940, 1550, and 1650 nm wavelengths transmitters in reflectance mode
Table 2: The average glucose obtained from the combination of the transmittance mode and the reflectance mode after
weighing
Percentage Glucose concentration measured by noninvasive method Glucose concentration measured Sample
error using (mg/dl) calculation the average of final results obtained in by invasive method (mg/dl)
transmittance mode and reflectance mode after weighing
40.26 46 77 1
25.66 113 152 2
28.1 151 210 3
22.41 225 290 4
25.55 306 411 5
136 Journal of Medical Signals & Sensors | Volume 8 | Issue 3 | July - September 2018
Javid, et al.: Blood glucose and bilirubin monitoring
Figure 25: Clarke error grid analyses between reference and measured Figure 26: The spectral response of the receiver[30]
blood glucose
Journal of Medical Signals & Sensors | Volume 8 | Issue 3 | July - September 2018 137
Javid, et al.: Blood glucose and bilirubin monitoring
Table 3: The comparison between the proposed system and previous articles
Research Providing Cost Glucose Bilirubin RMSE
mobile
communications
Our proposed system Low 18.52 mg/dL
Müller et al.[17] ‑ High ‑ 18.4‑33.8 mg/dL
Danzer et al.[18] ‑ High ‑ 36 mg/dL
Araujo-Andrade et al.[19] ‑ High ‑ 16 mg/dL
Xu et al.[20] ‑ High ‑ 15‑20 mg/dL
Guevara and González[21] ‑ High ‑ 21.96 mg/dL
Srivastava et al.[22] ‑ High ‑ Not evaluated
Pavithra et al.[23] ‑ High ‑ Not reported
Yadav et al.[1] ‑ High ‑ Not reported
Guo et al.[24] ‑ High ‑ The average of RMSE=3.80 mg/dL
Tamilselvi and Ramkumar[25] Low ‑ Not evaluated
Pande and Joshi[26] ‑ High ‑ Not evaluated
Bobade and Patil[27] Low ‑ Not reported
RMSE – Root mean square error
weighing the result of different transmitters, and checking 10. Govada A, Renumadhavi C, Ramesh K. Non‑invasive blood
the effect of pulse width are some offers to improve the glucose measurement. Int J Adv Res Comput Commun Eng
2014;3:5122‑5.
results.
11. Ali N, Muji S, Joret A, Amirulah R, Podari N, Risep ND.
Financial support and sponsorship Optical technique for jaundice detection. ARPN J Eng Appl Sci
2006;10:9929‑33.
None. 12. Osman Z, Ahmad A, Muharam A. Rapid prototyping of
Conflicts of interest neonatal jaundice detector using skin optics theory. 2014 IEEE
Conference on Biomedical Engineering and Sciences (IECBES).
There are no conflicts of interest. Kuala Lumpur, Malaysia: IEEE; 2014.
13. Kudavelly S, Keswarpu P, Balakrishnan S. A simple and
References accurate method for estimating bilirubin from blood.
2011 IEEE Instrumentation and Measurement Technology
1. Yadav J, Rani A, Singh V, Murari BM. Near‑infrared LED based Conference (I2MTC). Binjiang, China: IEEE; 2011.
non‑invasive blood glucose sensor. 2014 International Conference
14. Ali N, editor. A Review of Non_Invasive Jaundice detection
on Signal Processing and Integrated Networks (SPIN). Noida,
using Optical Technique in Neonates. Conference in Advances in
India: IEEE; 2014.
Computing, Electronics and Electrical Technology‑CEET; 2014.
2. Zhou F, Yang HI, Álamo JM, Wong JS, Chang CK. Mobile
15. Ng H, Sim M, Tan CM, Wong C. Wireless technologies for
personal health care system for patients with diabetes. telemedicine. BT Technol J 2006;24:130‑7.
International Conference on Smart Homes and Health Telematics.
16. Goyal S, Morita P, Lewis GF, Yu C, Seto E, Cafazzo JA, et al.
Berlin, Heidelberg: Springer; 2010.
The systematic design of a behavioural mobile health application
3. Quinn CC, Gruber‑Baldini AL, Shardell M, Weed K, Clough SS, for the self‑management of type 2 diabetes. Can J Diabetes
Peeples M, et al. Mobile diabetes intervention study: Testing a 2016;40:95‑104.
personalized treatment/behavioral communication intervention
17. Müller UA, Mertes B, Fischbacher C, Jageman KU, Danzer K.
for blood glucose control. Contemp Clin Trials 2009;30:334‑46.
Non‑invasive blood glucose monitoring by means of near
4. Krushinitskaya O. Osmotic Sensor for Blood Glucose Monitoring infrared spectroscopy: Methods for improving the reliability of
Applications. Vestfold University College; 2012. the calibration models. Int J Artif Organs 1997;20:285‑90.
5. Oliver NS, Toumazou C, Cass AE, Johnston DG. Glucose 18. Danzer K, Fischbacher C, Jagemann K, Reichelt K. Near‑infrared
sensors: A review of current and emerging technology. Diabet diffuse reflection spectroscopy for non‑invasive blood‑glucose
Med 2009;26:197‑210. monitoring. LEOS Newslett 1998;12:9.
6. Buda RA. A Portable Non‑Invasive Blood Glucose Monitoring 19. Araujo-Andrade C, Ruiz F, Martínez-Mendoza J, Terrones H.
Device. Universiti Teknologi Malaysia; 2014. Non-invasive in vivo blood glucose levels prediction using near
7. Vashist SK. Non‑invasive glucose monitoring technology in infrared spectroscopy. AIP Conference Proceedings. Guanajuato,
diabetes management: A review. Anal Chim Acta 2012;750:16‑27. Gto. (Mexico): AIP; 2004.
8. Yadav J, Rani A, Singh V, Murari BM. Prospects and limitations 20. Xu K, Qiu Q, Jiang J, Yang X. Non‑invasive glucose
of non‑invasive blood glucose monitoring using near‑infrared sensing with near‑infrared spectroscopy enhanced by optical
spectroscopy. Biomed Signal Proc Control 2015;18:214‑27. measurement conditions reproduction technique. Optics Lasers
9. Tura A, Maran A, Pacini G. Non‑invasive glucose monitoring: Eng 2005;43:1096‑106.
Assessment of technologies and devices according to quantitative 21. Guevara E, González F. Joint optical‑electrical technique for
criteria. Diabetes Res Clin Pract 2007;77:16‑40. noninvasive glucose monitoring. Rev Mex Fís 2010;56:430‑4.
138 Journal of Medical Signals & Sensors | Volume 8 | Issue 3 | July - September 2018
Javid, et al.: Blood glucose and bilirubin monitoring
22. Srivastava A, Chowdhury MK, Sharma S, Sharma N. Blood 26. Pande MC, Joshi A. Non‑invasive blood glucose measurement.
glucose monitoring using non invasive optical method: Design Int J Comput Eng Res 2015;5:26‑8.
limitations and challenges. Int J Adv Res Electr Electron Instrum 27. Bobade CD, Patil MS. Non‑invasive blood glucose level
Eng 2013;2:615-20. monitoring system for diabetic patients using near‑infrared
23. Pavithra A, Karan DM, Kumar DA, Anu Shalin P. Non invasive spectroscopy. Am J Comput Sci Inf Technol 2016;4.
technique to measure glucose and haemoglobin level in blood 28. Baharuddin H, Sulong MS, Joret A, Ab Rahman T, Ismail N. Bile
using NIR‑occlusion spectroscopy. Int J Sci Res Manag pigments detection via IR sensor. 3rd Engineering Conference on
2014;2:756‑62. Advancement in Mechanical and Manufacturing for Sustainable
24. Guo D, Shang Y, Peng R, Yong S, Wang X. Noninvasive blood Environment. Malaysia: 2010. pp. 1-3.
glucose measurement based on NIR spectrums and double ANN 29. Penhaker M, Kasik V, Hrvolová B. Advanced bilirubin
analysis. J Biosci Med 2015;3:42. measurement by a photometric method. Elektronika
25. Tamilselvi M, Ramkumar G, editors. Non‑invasive tracking and Elektrotechnika 2013;19:47‑50.
monitoring glucose content using near infrared spectroscopy. 30. InGaAs Photodetector PRELIMINARY SD012‑151‑011.
Computational Intelligence and Computing Research (ICCIC). Available from: http://lunainc.com/wp-content/
2015 IEEE International Conference. Madurai, India: IEEE; 2015. uploads/2016/06/012151011.pdf. Last accessed on 2018 Apr 20].
BIOGRAPHIES
Bahareh Javid obtained her BSc. in professor at Department of Computer Engineering and
Information Technology Engineering from Information Technology and the director of IOT and Smart
Shahrood University of Technology in 2012 Environments Research Group in University of Qom. Her
and her MSc. in Information Technology research interests include IOT, Smart Home, Assistive
Engineering (e-commerce) from University Living and Deep Learning.
of Qom in 2016. Her research interests
include Bioinformatics- Wearable Sensors - Email: f-fotouhi@qom.ac.ir
Data Mining - Artificial Intelligence.
Fahime Sadat Zakeri obtained her MSc. in
Email: bahareh.javid@gmail.com Biomedical Engineering from Iran
University of Science and Technology in
Dr. Faranak Fotouhi-Ghazvini has 2009 and her PhD in Biomedical Engineering
received her MEng degree in from Tehran University in 2018. She
Telecommunication Engineering from currently lectures in University of Qom. Her
King's College London (UK) in 2001 (1st research interests include Bioinformatics,
class), and her PhD degree from Bradford Image processing, Machine Learning.
University (UK), Department of
Informatics in 2011. She is an Assistant Email: zakerifahimeh@gmail.com
Journal of Medical Signals & Sensors | Volume 8 | Issue 3 | July - September 2018 139