Professional Documents
Culture Documents
AcuLight HR Users Manual PB3581110
AcuLight HR Users Manual PB3581110
AcuLight HR Users Manual PB3581110
OPERATOR’S MANUAL
PB 3581110 Rev. B
July 2001
Copyright © 2001 by ESC Medical Systems Ltd.
All rights reserved. No part of this manual may be reproduced or copied in any form by any means
- graphic, electronic or mechanical, including photocopying, typing, or information retrieval
systems - without written permission of aculight or ESC Medical Systems Ltd.
Use of Manual:
The aculight HR system is designed to meet international safety and performance standards.
Personnel operating the unit must have a thorough understanding of the proper operation of the
system.
This manual has been prepared to aid clinicians and other personnel to understand and operate
the system. Do not operate the system before reading this manual and gaining a clear
understanding of the operation of the system. If any part of this manual is not clear, please contact
your aculight representative for clarification.
The information provided in this manual is not intended to replace the professional training on the
clinical use of the aculight HR System. (Such training should include a review of published
literature, seminars, workshops and appropriate preceptorships.) Please contact your aculight
representative for current information on available training. For professional information and
clinical applications, refer to Appendix A of this manual.
This manual should always accompany the unit, and its location must be known to all personnel
operating the unit. Additional copies of this manual are available from aculight or your local
distributor.
ii PB 3581110 Rev. B
PHYSICIAN RESPONSIBILITY
Federal (USA) law restricts prescription medical devices to sale by or on the order
of a physician, or properly licensed practitioner. That physician will be responsible
for the use and operation of the device and for all user qualifications. aculight
makes no representations regarding federal, state or local laws or regulations that
might apply to the use and operation of any medical device. The physician is
responsible for contacting his or her local licensing agencies to determine any
credentials required by law for clinical use and operation of the device.
MAINTENANCE
The aculight HR is a precision, technical medical device that requires routine
service as well as consumable parts. All service must be performed by an aculight
technician and all parts must be purchased from aculight. Failure to obtain service
and parts through aculight voids all warranties, express and implied. Please call
aculight or your local representative for details.
MODIFICATION OF DEVICE
Unauthorized modification of the hardware, software or specifications of the
aculight HR voids all warranties, express and implied. aculight takes no responsi-
bility for the use or operation of such a device.
PB 3581110 Rev. B v
Appendix A Clinical Guide Provides details on
indications, contrain-
dications, pre-treatment care,
treatment procedures and
post-treatment care.
Appendix B Taking Care of Provides details on the
aculight HR correct maintenance of the
Treatment Heads aculight HR treatment heads.
vi PB 3581110 Rev. B
TABLE OF CONTENTS
Introduction
Chapter 2. Safety
2.1 General............................................................................................. 2-1
2.2 Introduction ..................................................................................... 2-1
2.3 Major Cautions and Warnings ......................................................... 2-4
2.4 Optical Safety .................................................................................. 2-7
2.5 Electrical and Mechanical Safety .................................................... 2-8
2.6 Fire Hazards..................................................................................... 2-8
2.7 System Safety Features.................................................................... 2-9
2.8 Equipment Classification............................................................... 2-12
2.9 Compliance with International Standards...................................... 2-12
Chapter 4. Installation
4.1 Equipment List ................................................................................ 4-1
4.2 Facility Requirements...................................................................... 4-2
4.3 Moving the System .......................................................................... 4-5
Chapter 5. Operation
5.1 Screens and Functions ..................................................................... 5-2
5.2 Starting the System .......................................................................... 5-3
5.3 The Treatment Screen...................................................................... 5-6
5.4 Treatment......................................................................................... 5-8
5.5 The Menu Screen........................................................................... 5-13
5.6 Calibration ..................................................................................... 5-13
5.7 Changing the Treatment Head....................................................... 5-19
5.8 Idle Mode....................................................................................... 5-22
5.9 Statistics......................................................................................... 5-23
5.10 Advanced Mode............................................................................. 5-24
5.11 Turning Off the System ................................................................. 5-28
Chapter 7. Accessories
7.1 Safety Eyewear ................................................................................ 7-1
7.2 Coupling Gel.................................................................................... 7-1
7.3 Light Guide Covers ......................................................................... 7-1
7.4 Optional Heads with Other Filters................................................... 7-2
7.5 Ordering Accessories....................................................................... 7-2
Chapter 2. Safety
Figure 2.1 Caution Sign for the Treatment Room ...................................2-3
Figure 2.2 Machine Labels ......................................................................2-6
Figure 2.3 System Safety Features on Control Panel ..............................2-9
Figure 2.4 Circuit Breaker .....................................................................2-11
Chapter 4. Installation
Figure 4.1 aculight HR Physical Dimensions .........................................4-2
Figure 4.2 Location of Line Voltage Label .............................................4-4
Chapter 5. Operation
Figure 5.1 Flowchart of Main Screens ....................................................5-2
Figure 5.2 Current Treatment Head .........................................................5-3
Figure 5.3 Unidentified Head Message ...................................................5-3
Figure 5.4 Initialization Screen ...............................................................5-4
Figure 5.5 Treatment Screen ...................................................................5-4
Figure 5.6 Calibration Warning Message ................................................5-5
Figure 5.7 Fields in Treatment Screen ....................................................5-6
Figure 5.8 Ready Screen ........................................................................5-10
Figure 5.9 Pulse Parameters Window ...................................................5-11
Figure 5.10 Menu screen .........................................................................5-13
Figure 5.11 aculight HR Power Meter ....................................................5-13
Figure 5.12 Location of Socket for Connecting Power Meter ................5-15
Figure 5.13 Calibration Screen ................................................................5-16
Figure 5.14 Calibrating the Treatment Head ...........................................5-16
Figure 5.15 Progress of Calibration .........................................................5-17
Figure 5.16 Recharging between Pulses ..................................................5-17
Figure 5.17 No Power Meter Signal ........................................................5-18
Figure 5.18 Calibration Results Screen ...................................................5-18
Figure 5.19 Calibration Failure ...............................................................5-19
Figure 5.20 Change Head Confirmation .................................................5-20
Figure 5.21 Replacing the Treatment Head .............................................5-20
PB 3581110 Rev. B ix
Figure 5.22 Treatment Head Disconnected .............................................5-21
Figure 5.23 New Treatment Head Identification .....................................5-21
Figure 5.24 Idle Mode Screen .................................................................5-22
Figure 5.25 Statistics Screen ...................................................................5-23
Figure 5.26 Advanced Screen ..................................................................5-25
Figure 5.27 Advanced Screen in User Mode ...........................................5-26
Figure 5.28 Save Parameters Message ....................................................5-27
Figure 5.29 Treatment Screen with User Parameters ..............................5-27
x PB 3581110 Rev. B
1
C H A P T E R
SYSTEM OVERVIEW
Chapter 1 is a general introduction to the aculight HR
system.
Caution
Personnel operating or maintaining this
equipment should read this manual thoroughly
before attempting to operate the aculight HR
system.
SAFETY
2.1 General
This chapter describes general safety issues regarding
the use of the aculight HR, with special emphasis on
optical and electrical safety.
Warning
Use of controls or adjustments, or performance of
procedures other than those specified herein may
result in hazardous radiation exposure.
2.2 Introduction
The aculight HR is an instrument designed for safe and
reliable hair removal. With proper operation and
maintenance, the system can be used safely by trained,
qualified practitioners. The operator and all other
personnel operating or maintaining this equipment
should be familiar with the safety information provided
in this chapter.
The primary considerations should be for the safety of
the client, the operator and other personnel.
Warning
Any intense pulsed light emitting device can
cause injury if used improperly. High voltages are
present inside the aculight HR unit.
Caution
Restricted device. Federal law (USA) restricts this
device to sale by or on the order of a physician or
any practitioner licensed by the law of the state in
which he or she practices to use or order the use
of the device.
The Client
Client safety is primarily assured by a well-trained staff
and a well laid out treatment room. Client education is
also important, including information about the nature
of the treatment.
Cautions
• Only aculight authorized personnel may service the
aculight HR system, especially inside its protective
covers. This includes making internal adjustments to
the power supply, cooling system, optics, treatment
head, etc. Dangerous voltages are present inside the
unit.
• Verify that the aculight HR is wired for the appropri-
ate electrical voltage of your country (200 V, 208 V
or 230 V).
• Maintenance should be performed by the user only
when the unit is shut down and disconnected from
power. Performing maintenance procedures with the
system powered up may be hazardous to the user and/
or destructive to the system.
• The light guide must be kept clean at all times and gel
should not be allowed to seep inside the treatment
head. The light guide should be covered with the
provided plastic covers, which should be cleaned
thoroughly or replaced after every client.
• If water is detected leaking from the treatment head,
the aculight HR should not be turned on. If it is
already on, it should be turned off immediately.
Replace the leaking head before starting the system.
CAUTION
HIGH VOLTAGE / HIGH
CAPACITANCE
Warning
Delivering excessive energy to the treatment site
can result in thermal damage to the skin possibly
leading to hypertrophy and/or atrophy and/or
abnormal pigmentation.
Emission Indicator
Key Switch
Power-On Indicator
Emergency Shut-off Knob
Key Switch
The key switch connects and disconnects electrical
power to the system, and is used to switch the system
ON and OFF. Operation of the system is possible only
with the key provided by aculight. This prevents
unauthorized use of the unit.
Do not leave the aculight HR unattended with a key in
it.
Verify that the system is inoperable when the key is
removed.
Emission Indicator
The aculight HR is equipped with a yellow emission
indicator, which is mounted on the control panel.
The emission indicator has two modes of operation:
Idle Mode
The inactivity timer feature switches the unit to Idle
mode whenever it has not been used for ten minutes.
The system can also be switched to this mode by
manually pressing the Idle button. When the system is in
the Idle mode, the capacitors are discharged, the light
guide chiller is turned off and the system remains
inoperable until manually activated.
Circuit Breaker
A circuit breaker located inside the aculight HR, at the
bottom of the rear panel of the console, protects the
system against excessive current consumption. The
circuit breaker disconnects power to the system if the
current consumption exceeds 16 A (for 200/208 V AC)
or 13 A (for 230 V AC). To resume system operation,
lift the circuit breaker handle.
Circuit Breaker
Handle
SYSTEM DESCRIPTION
This chapter provides a general description of the
aculight HR system, including its main components,
controls, and technical specifications.
Treatment
Head
Control
Panel
System
Console
Control Panel
PC
Trigger
button
I2C Card
Over
Energy
Overvoltage
Capacitor Bank
HV Power Supply TE Cooler
Simmer PS
DC PS
Flow Switch
Cooling System
Level Switch
Power Meter
System Console
The system console is the operational heart of the
system. It contains the components required for
operation and control of the system. The console
comprises the following modules:
Name Function
Computer & Interface Controls the operation of the entire system. It
Card (I2C) receives input from all the modules and performs
the appropriate functions.
Power and control module Supplies power to the system modules. It also
shuts down the system in case of any malfunction.
Interface module Serves as an interface between the computer and
system modules.
Switching module Switches the electrical energy into the flashlamp.
High voltage power Charges the capacitor bank.
supply (HVPS)
Capacitor Bank Stores the energy used for the light pulses.
DC Power Supply (DCPS) Provides power for the Simmer power supply, the
Cooling system and the thermo-electric (TE)
cooler.
Simmer Power Supply Ignites the flashlamp and keeps it conducting in
simmer mode.
Cooling System Removes the heat generated in the treatment head.
Emission Indicator
Key Switch
Power On Indicator
Emergency Knob
Treatment Head
The treatment head houses the mechanisms that
generate the light pulse. It is connected to the system
console by an umbilical cord. A glass light guide coated
with a filter is mounted at the bottom of the treatment
head.
The light pulse is activated by pressing the trigger on the
treatment head handle. The light passes through an
aperture and a 34 x 8 mm filter-coated light guide onto
the treatment area. The light guide transmits the light
from the treatment head to the client’s skin. It is
important to keep the light guide clean at all times.
Trigger
Handle
Thermo-Electric Cooler
Glass Light Guide (inside)
Thermo-Electric Cooler
The light guide is cooled by a thermo-electric (TE)
cooler, which is incorporated in the treatment head. The
cooled light guide reduces skin temperature, thus
reducing client discomfort. The TE cooler has two
cooling levels and can also be turned off.
Environmental Conditions
Operating Temperatures: l0°C– 30°C (50°F – 86°F)
Storage Temperatures: 0°C – 55°C (32°F – 131°F)
Relative Humidity - Operating: Up to 80%
Relative Humidity - Storage: Up to 90%
INSTALLATION
The aculight HR is designed for installation in a clinic
and requires minimal site preparation. When the
aculight HR system is purchased, complete on-site
installation including initial system test and calibration
is included.
Installation is carried out by authorized aculight
personnel, who do the following:
• Unpack the aculight HR system and position it in the
pre-selected location.
• Verify the integrity of the unit and its components.
• Plug the system into a designated electrical outlet.
• Test the system for proper calibration and functional
operation of all components.
Space Requirements
The physical dimensions of the system are:
186 cm
(73")
55 cm
40 cm
40 cm
Electrical Requirements
The aculight HR system is factory pre-wired for the
local line voltage, as ordered by the customer.
Accordingly, the unit will require a separate line supply
of at least:
• Single phase 230 V AC ±10%, 16 A, 50/60 Hz, or
• Single phase 208 V AC ±10%, 20 A, 50/60 Hz, or
• Single phase 200 V AC ±10%, 20 A, 50/60 Hz.
The unit is grounded through the grounding conductor
(yellow/green) in the power cable that is plugged into
the wall power outlet. Proper grounding is essential for
safe operation.
Input power lines should be free of transients, voltage
and current spikes, sags and surges. Consequently the
system power line should not be shared with other
heavy variable loads such as elevators, air conditioning
systems, large motors etc.
It is strongly recommended that the system be
connected to a separate power line with separate circuit
breakers. aculight cannot guarantee adequate
performance unless the system is connected to a
dedicated circuit.
The main fuse located within the unit is a 13 A circuit
breaker (for 230 VAC) or 16 A (for 200 V or 208 VAC).
When the circuit breaker trips, it disconnects power
from the system.
To restore the power after the circuit breaker has
tripped, raise the circuit breaker handle.
Caution
Verify that the aculight HR is wired for the
appropriate line voltage of your country (200 V,
208 V or 230 V) by checking the label located on
the back of the system (see Figure 4.2).
Local
Line
Voltage
Environmental Requirements
Air Quality The aculight HR should operate in a non-corrosive
atmosphere. Corrosive materials such as acids can
damage electrical wiring, electronic components and the
surfaces of optical components.
Air-borne dust particles should be kept to a minimum.
Dust particles absorb light and heat up. Hot particles
located on the optical filters can cause damage.
OPERATION
This chapter contains detailed operating instructions for
the aculight HR.
Warning
Dangerous voltages are present inside the unit!
Keep all panels and covers closed.
Warning
Delivering excessive energy to the treatment site
can result in thermal damage to the skin possibly
leading to hypertrophy and/or atrophy and/or
abnormal pigmentation.
HR Treatment Calibration
Standby
Reset counter Idle
Set pulse type Menu
Menu
Set fluence
Statistics
Ready
Ready to treat Advanced
Treat
If this happens:
1. Press Change Head.
2. Replace the treatment head (see Section 5.7 Changing
the Treatment Head).
Calibration
Calibrating your system on a regular basis ensures the
greatest accuracy of its output, as well as consistent
clinical results.
Calibration is mandatory in the following cases:
• When a new head is connected to the system.
• When the connected head was last calibrated on a
different machine.
• When 3000 pulses were emitted from the current
head without performing calibration. Calibration is
required since the actual fluence output of the
treatment head may otherwise decrease over time,
due to age and wear.
In these cases, the following warning message
appears on the screen when you switch on the system,
or after you changed the treatment head.
Instructions/
Messages
Treatment Parameters
Instructions/Messages
This field displays instructions for the operator and
messages on the parameter status.
Note
If the previous use was with user defined pulse
parameters, an additional (third) line appears
indicating that such settings are active (Figure
5.29).
Treatment parameters
From this field the operator selects the treatment
parameters:
• Pulse Type—determines the number of pulses in the
sequence, the pulse duration and the delay between
the pulses. Short, Medium and Long are the available
options. The user may choose either the default
settings or one of his own.
• Fluence—The Treatment screen always appears with
the minimum fluence setting. The operator changes
this setting by pressing the + and - arrows. The
progress bar indicates the position within the
available fluence range.
Pulse Counter
The pulse counter on the Treatment screen indicates the
total number of pulses triggered since the last time the
counter was reset. Each time a pulse is triggered, the
number is incremented by one. The Reset button can be
pressed at any time to reset the counter to zero.
Function Buttons
Ready Pressing the Ready button confirms the selected
treatment parameters and transfers the system to the
Ready mode (Ready button is green). This is the only
mode that enables triggering of the light pulse.
Standby Pressing the Standby button puts the system into the
Standby mode. In Standby mode, the Standby button is
green and light pulses cannot be triggered. All the other
buttons are available and the user can modify treatment
parameters. To return to Ready mode the user presses
the Ready button on the screen.
Change Head Pressing the Change Head button enables changing the
treatment head (see Section 5.7 Changing the Treatment
Head).
Pulse Type The Pulse Type button enables the display of the current
pulse parameters (see Section 5.4 Treatment).
Menu The Menu button displays the Menu screen (see Section
5.5 The Menu Screen).
5.4 Treatment
A typical treatment procedure begins with the system in
Standby mode. The Treatment screen is displayed with
the Standby button green (Figure 5.5).
To begin treatment, proceed as follows:
1. Shave the treatment area to eliminate any surface hair
that could interfere with the treatment.
2. Press the Reset button to reset the Pulse Counter.
3. Select Chiller level by toggling the Chiller button.
4. Set the Pulse Type by pressing Short, Medium or
Long.
The preset pulse type parameters (defaults) for the
available treatment head types are:
Table 5.1 Preset Pulse Type Parameters for 695 nm (standard)
and 645 nm (optional) Treatment Heads
Short 2 5.5 20
Medium 3 3.6 40
Long 3 3.6 100
Short 2 5.5 20
Medium 3 3.6 40
Long 3 3.6 150
Note
The light guide cover should be cleaned
thoroughly or replaced for every new client.
Note
It is recommended that the client return several
weeks after treatment for examination of the
treated site, and continuation of treatment if
necessary.
Note
Changing pulse type parameters is
recommended for advanced users only.
5.6 Calibration
aculight recommends daily calibration of the treatment
head output. Calibration is performed with the supplied
power meter.
Caution
The fluence output of a treatment head decreases
over time. Thus, if calibration is performed after
too long a period of time, the actual fluence
emitted may be higher than before the calibration,
for the same setting. Be sure to perform a test
patch before starting treatment.
Calibration Procedure
To calibrate the treatment head output:
1. Plug the power meter connector into the round socket
(marked) on the rear panel of the aculight HR and
screw the connector closed (Figure 5.12).
Socket
Treatment Head
Power Meter
Number of Pulses
is decremented by
one after each
triggered pulse
• Verify that the light guide and power meter are clean
and free of dust and debris.
• Press the Calibrate button and repeat the calibration
process.
• If the calibration fails a second time, press the
Change Head button and replace the treatment head
(see Section 5.7 Changing the Treatment Head).
Note
The aculight HR automatically enters Idle mode
if it has not been used for ten minutes. The
system shuts down automatically after one hour
in Idle mode.
5.9 Statistics
When the Statistics button is pressed, the Statistics
screen appears (Figure 5.25).
Short 2 or 3 3 - 6 @ 0.1 10 - 30 @ 5
695
Medium 2 or 3 3 - 6 @ 0.1 35 - 80 @ 5
(Standard)
Long 2 or 3 3 - 6 @ 0.1 85 - 150 @ 5
Short 2 or 3 3 - 6 @ 0.1 10 - 30 @ 5
755
Medium 2 or 3 3 - 6 @ 0.1 35 - 80 @ 5
(Optional)
Long 2 or 3 3 - 6 @ 0.1 85 - 150 @ 5
Short 2 or 3 3 - 6 @ 0.1 10 - 30 @ 5
645
Medium 2 or 3 3 - 6 @ 0.1 35 - 80 @ 5
(Optional)
Long 2 or 3 3 - 6 @ 0.1 85 - 150 @ 5
Short
Medium 25–45@1 25–45@1 20–35@1
Long
Current
Parameter Values
Parameter
Values
Window
Note
For each Head Type, each Pulse Type (Short,
Medium or Long) has one set of default
parameters. The user can modify these settings
and save them as user-defined parameters.
Note
Only one set of user defined parameters can be
saved. Pressing the Save button saves the
current values that appear on the screen as the
user defined parameters.
Warning
Maintenance by the user should be performed
only when the unit is shut down and disconnected
from power. Performing maintenance procedures
with the system ON can be hazardous to you and/
or destructive to the system.
Caution
Never immerse the light guide or any other part of
the treatment head in water, nor hold it under
running water.
Open
Top Cover
Connectors for
filling water tank
Calibration Check
The calibration check compares the fluence reading on
the aculight HR screen to that of a calibrated power
meter.
The calibration check is performed by aculight service
personnel, using a calibrated power meter made by
Ophir Optronics Ltd. (An equivalent calibrated power
meter may be used.)
To check the power meter calibration:
1. Using the aculight HR power meter, the Field Service
Engineer performs standard output calibration (see
Section 5.6 Calibration).
2. The Field Service Engineer then selects a fluence
value on the Treatment screen and measures the
output of the calibrated machine with the calibrated
Ophir power meter. He then compares the measured
fluence with the value on the Treatment screen.
Calibration Procedure
If the fluence reading on the screen and that of the Ophir
power meter differ by more than 10%, the Field Service
Engineer should calibrate the aculight HR power meter.
6.7 Troubleshooting
If one of the following symptoms appears, follow the
recommended action as listed below:
Troubleshooting Guide
Caution
If an error message appears, write down its
contents and contact aculight service.
The text on the error message window describes
the machine’s status and may be helpful to the
Field Service Engineer in diagnosing the problem
and knowing how to solve it
Type of
error
OK
ACCESSORIES
The aculight HR accessories are:
• Safety eyewear
• Coupling gel
• Light guide covers (plastic)
• Optional heads with other filters
Part
Product
Number
Optical
AX 0000010
density 5
Safety eyewear
Optical
AX 0000011
density 3
Coupling gel 0.25 liter GL 0000006
Coupling gel 1 liter GL 0000007
Treatment head, 695 nm Standard SA 3507000
Treatment head, 645 nm Optional SA 3509000
Treatment head, 755 nm Optional SA 3508000
Light guide cover Set of 3 MI 3557030
A
P P E N D I X
CLINICAL GUIDE
CHAPTER0
General
During the client’s first visit, the clinician (or an
authorized staff member) should:
1. Exclude from treatment anyone who was exposed to
sun or artificial tanning during the last month.
2. Take a detailed client history, including previous
treatment modalities, and determine the suitability
for treatment with the aculight HR.
3. Determine why the client is seeking treatment and
understand his/her expectations.
4. Discuss the aculight HR treatment plan with the
client.
Counseling
During the first visit, the clinician (or an authorized
member of the staff) should inform the client of the
following:
• There may be some discomfort or pain associated
with treatment.
• Transient erythema/edema may appear immediately
following treatment.
• Hair removal may take a number of treatments.
• There is a small risk of adverse reactions such as
changes in the texture and pigmentation of the skin.
These are usually transient.
Eye Protection
It is imperative that all people present in the treatment
room (client and personnel) during treatment protect
themselves from retinal damage by wearing aculight
recommended safety eyewear whenever the aculight
HR is in use.
Introduction
The aculight HR is an intense pulsed light device for
hair removal. aculight HR eliminates hair, employing a
method known as selective photothermolysis. This
involves disabling hair regrowth mechanisms by raising
the temperature of the hair follicle high enough to
damage the follicle’s germinative cells without
damaging the epidermis and the surrounding tissue. The
germinative cells are those cells which divide and
differentiate into the cells comprising the hair.
Following photoepilation, the debris of these cells and
the shattered melanin are eliminated from the tissue by
phagocytosis.
Hair on different parts of the body is of different depths.
For example, upper lip or mustache hair follicles are 1 to
2.5 mm deep, while axillae or pubic hair follicles are 3.5
to 4.5 mm deep.
The aculight HR is furnished with a set of default Pulse
Type settings. These defaults are based on successful
results obtained by experienced users. In addition, the
user can select and set his own parameter values.
The aculight HR treatment parameters consist of light
pulse parameters (Pulse Type) and light energy
(Fluence).
For each treatment with the aculight HR, these
parameters are to be set, either by selecting one of the
three default settings, or by entering a User setting of
your own, and then selecting the treatment fluence.
Definitions
Pulse Type
Choosing the correct pulse type is essential for
optimizing the treatment.
Cooling time of an object is a function of its size.
Cooling time is shorter for small objects and longer for
larger objects.
The aculight HR is designed to take advantage of this
thermal selectivity, i.e. longer cooling time of the
relatively large hair follicles and shorter cooling time of
the thinner epidermis. This is accomplished by dividing
the total energy delivered during treatment into several
sequential pulses:
The delay should be long enough for the epidermis to
cool down between pulses, by losing its heat to the cold
gel and cooled light guide. But the delay should also be
shorter than the cooling time of the follicle, so that the
follicle retains its heat and its temperature increases
with each successive pulse. For most clients short delays
are suitable. Dark skin absorbs more light and heats to a
higher temperature. Consequently, longer delay times
are necessary.
The aculight HR provides three preset pulse sequences,
labeled Short, Medium and Long (the Advanced mode
enables customized settings as well).
The operation of selecting presets, setting parameters
and ranges of parameter values is detailed in Chapter 5,
Operation.
Fluence
Fluence measures the light energy delivered to the skin,
in units of energy per unit area (Joules/cm2). The
aculight HR’s intense light pulses pass through the light
guide and the coupling gel onto the client’s skin. With a
large spot size (8 x 34 mm) the scattering effect is
minimized, resulting in a constant fluence and deeper
penetration (unlike the smaller spot sizes used with
lasers). The greater the fluence, the higher the
Coupling Gel
Before the light pulse is triggered, apply a thin layer of
cold coupling gel to the treatment site.
Note
Treatment of Skin Type VI
Note
aculight does not recommend treating hair that
is lighter than the skin!
A.7 Treatment
1. Press the Reset button to reset the Pulse Counter.
2. Set the Pulse Type by pressing Short, Medium or
Long.
A minimum fluence value appears on the screen.
To change the fluence value, press the plus or minus
arrows. It is recommended to start with a low
fluence value
3. Shave the treatment area to eliminate any surface hair
that could interfere with the treatment.
4. Ensure that the light guide is clean and free of frost.
5. Cover the light guide with the provided cover.
6. Apply a thin layer of cold coupling gel to the
treatment site. This aids skin cooling during the light
pulse and improves coupling of the light into the
skin.
Light pulses should be triggered on cold gel, so
spread gel only on immediate treatment area and treat
right away, then spread gel on next area.
7. Place the treatment head perpendicular to the skin.
The light guide should gently touch the skin.
8. Trigger a test pulse. (The system recharges
automatically.)
9. Wipe off the gel and examine the treatment site.
The smell of burnt hair and/or a brownish color in the
gel may be detected. Swelling around the hair
follicles should appear within about five minutes
after treatment.
Darker skin may take longer to react than lighter skin
(from a few minutes to 24 hours).
10.If the skin shows no adverse effects and there are no
visible changes in the follicles, increase the Fluence
and treat an adjacent site.
If the skin shows adverse effects (such as excessive
redness), reduce the Fluence and/or choose a longer
Pulse Type. Treat an adjacent site.
11.Wipe off the gel and examine the treatment site. If
there are still no adverse effects or changes in the
follicles, increase the Fluence and treat an adjacent
site.
Do not overlap treatment sites by more than ~1 mm.
General
It is recommended that cold (not frozen) packs be
applied immediately after treatment, to cool the
treatment site, reduce swelling and ease discomfort.
Chemical cold packs are not recommended, if their
temperature is below 4 degrees C. Frozen 4 X 4 gauze,
previously moistened with water and inserted into small
plastic bags (sandwich wrap) can be used.
Blisters or ulcerated skin can be treated with antibiotic
ointment or burn treatment creams. Low concentration
steroid cream can be used to reduce swelling.
Care should be taken to prevent trauma to the treated
area for the first four or five days following treatment:
avoid hot bath, aerobic exercise, massage, etc.
If scabs appear after blistering, they should be kept soft
with a lubricating cream such as petroleum jelly.
Exposure to Sunlight
Clients should use highest factor sunscreen and protect
the treated area from exposure to sunlight for at least
one month following treatment. Tanning after treatment
sessions may enhance melanin regeneration, which may
result in hyperpigmentation.
Makeup
Some users allow their clients to apply makeup
immediately after treatment. However, they advise
clients to notify them, and to stop wearing makeup, if
the treatment site scales or cakes.
Others take a more conservative approach, advising
against the use of makeup for four or five days
following treatment. The skin is sensitive during this
period, and removing makeup, especially if it is difficult
to remove, may damage the skin and predispose the site
to infection.
Other Post-treatment
Recommendations
If the treatment site could be exposed to dirt routinely,
cover it with a dressing.
Clients should be advised not to participate in rough
sports or similar activities for several days following
treatment, until the skin returns to its normal condition.
Follow-up
Refer to the following recommendations for follow-up,
and then determine your own suitable regime.
The client should return several weeks after treatment (4
to 8 depending on anatomical site treated), for
examination of the treatment site. If hair is visible on
that site, an additional treatment should be performed.
If no additional treatment is necessary, the client should
return for an additional re-examination three to four
weeks later, preferably when new hair has grown in the
treated area.
If there has been partial hair clearance, treatment should
be continued using the same parameters, and the client
should return after 4 to 8 weeks for examination and for
additional treatment, if necessary.
If no change is noted, the fluence for the next treatment
should be increased. With multiple treatments, increase
the time intervals between treatment sessions (after the
second one), to allow new hair to grow in the treatment
area. This depends on the body area and on the
individual client (male or female, hormonal problems,
etc.).
Concluding Treatment
Determining when treatments should be concluded is
left to the user’s discretion. In general, treatment ceases
when the client is satisfied with the amount of hair
removed.
Note
This section is presented for informational
purposes only and should not be interpreted as
an indication or guarantee of specific results that
could be expected from use of the aculight HR.
N=24
76% 84%
59%
62% (24) (24)
(18)
100% 56% (21)
44% (21)
Average clearanc
(23)
80%
60%
40%
20%
0%
0 2 4 6 8 10 12 14 16 18 20 22 24
Follow -up (m onths)
Note
The number in parentheses for each time period
represents the number of anatomical sites
recorded at that time.
The average clearance value is indicated
above the vertical lines, representing the
standard deviation (SD) for each follow-up.
0 weeks represents the immediate clearance
rate.
Skin Type VI
No data on the effects of multiple treatments on skin
type VI is available at this time.
HEADS
B.1 Introduction
aculight HR treatment heads have a built-in integral
filter that controls the spectral output of your
aculight HR system. The filter is coated on the inner
side of the light guide. The integral filter is not
interchangeable, therefore proper care is important.
Note
Remember to clean the integral filter
immediately after each client.
PB 3581110 Rev. B 1
2 PB 3581110 Rev. B