3Dhifu (2018斜面)
3Dhifu (2018斜面)
3 Dimension HIFU
F-8003
User manual
Contents
Instrument Part-------------------------------------------------------3
Attentions-------------------------------------------------------------4
User Interface--------------------------------------------------------6
Operating------------------------------------------------------------10
Parameter settings-------------------------------------------------15
Technical parameters----------------------------------------------16
Troubleshooting-----------------------------------------------------17
Instrument principle------------------------------------------------19
1
User manual
③ ④
① ②
① Reset switch
② ON/OFF switch
③ Output socket
④ Power cocket
2
User manual
3
User manual
2,Attentions
2-1 Wanger
2, Don't touch the plug, don't pull out the power cord;
4, Unplug the power line and handle when moving the instrument;
6, Do not place equipment with a strong magnetic field around the instrument;
7, Don't unplug the power line or turn off the power supply when the instrument
is working;
8, When the instrument is abnormal, please turn off the power supply;
4
User manual
2,Attentions
2-2 Be careful
1, Pregnant women;
2, Cardiovascular diseases and diabetes ;
3, Easily infected with bacterial sites;
4, There are patients with pacemakers;
5, Sensitive skin
Be careful
5
User manual
3, User Interface
3-1 Start
3D HIFU ULTRASONIC
_ Width (mm)
10 +
DL-1.5
Use Line: 00010
Total Line: 10500
+ +
Length(mm) Clearance (mm)
_ _
M1 M2
E . J E . J
SAVE L - . mm L - . mm
W - . mm W - . mm
M3 M4 M5
E . J E . J E . J
L - . mm L - . mm L - . mm
Pitch (mm)
_ W - . mm W - . mm W - . mm
Back Enter
2.0 +
_ Power (J)
6
User manual
3, User Interface
3D HIFU ULTRASONIC
_ Width (mm)
10 +
DL-1.5 Handle model
Use Line: 00010
Total Line: 10500
+ +
Length(mm) Clearance (mm) Patient total display
20 2.0 Patient Total 0000 Clear
Patient total clear
_ _
M1 M2
E 1.0 J E 1.0 J
SAVE L 10 - 2 . 0 mm L 10 - 2 . 0 mm
W 20 - 2 . 0 mm W 20 - 2 . 0 mm
M3 M4 M5
parameter set
Pitch (mm)
E 1.0 J
L 10 - 2 . 0 mm
W 20 - 2 . 0 mm
E 1.0 J
L 10 - 2 . 0 mm
W 20 - 2 . 0 mm
E 1.0 J
L 10 - 2 . 0 mm
W 20 - 2 . 0 mm
and save
_
2.0 +
_ Power (J)
2.5 + Start stop Back
Output power
width
pitch
Clearance
Lines=width/pitch+1
Length
Shoot=Lines X
Length
Clearance
7
User manual
3, User Interface
/ 10500
Width Pitch Patient Total Length Clearance
1 Face
Back Enter
Start
Back Stop
2
3D HIFU ULTRASONIC
Power
2.0 J
DL-1.5
/ 10500
Width Pitch Patient Total Length Clearance
waist
Back Enter
Start
Back Stop
2
3D HIFU ULTRASONIC
Power
2.0 J
DL-1.5
/ 10500
Width Pitch Patient Total Length Clearance
Neck
Back Enter
Start
Back Stop
3D HIFU ULTRASONIC
Power
2.0 J
DL-1.5
/ 10500
Width Pitch Patient Total Length Clearance
Buttocks
Back Enter
Start
Back Stop
3D HIFU ULTRASONIC
Power
2.0 J
DL-1.5
/ 10500
Width Pitch Patient Total Length Clearance
Arm
Back Enter
Start
Back Stop
8
User manual
3, User Interface
/ 10500
Width Pitch Patient Total Length Clearance
Chest
Back Enter
Start
Back Stop
3D HIFU ULTRASONIC
Power
2.0 J
DL-1.5
/ 10500
Width Pitch Patient Total Length Clearance
Abdomen
Back Enter
Start
Back Stop
3D HIFU ULTRASONIC
Power
2.0 J
DL-1.5
/ 10500
Width Pitch Patient Total Length Clearance
Thigh
Back Enter
Start
Back Stop
3D HIFU ULTRASONIC
Power
2.0 J
DL-1.5
/ 10500
Width Pitch Patient Total Length Clearance
Legs
Back Enter
Start
Back Stop
9
User manual
4, Operating
4-1 Professional
3D HIFU ULTRASONIC
_ Width (mm)
10 +
DL-1.5
+ +
Length(mm) Clearance (mm)
_ _
M1 M2
E 1.0 J E 1.0 J
SAVE L 10 - 2 . 0 mm L 10 - 2 . 0 mm
W 20 - 2 . 0 mm W 20 - 2 . 0 mm
M3 M4 M5
E 1.0 J E 1.0 J E 1.0 J
L 10 - 2 . 0 mm L 10 - 2 . 0 mm L 10 - 2 . 0 mm
Pitch (mm) W 20 - 2 . 0 mm W 20 - 2 . 0 mm W 20 - 2 . 0 mm
_
2.0 +
_ Power (J)
2.5 + Start stop Back
width
pitch
_ Width (mm)
10 +
+ +
Clearance
20 2.0
Length
_ _
Pitch (mm)
_
2.0 +
SAVE
M1
E 1.0 J
L 10 - 2 . 0 mm
M2
E 1.0 J
L 10 - 2 . 0 mm
M2
W 20 - 2 . 0 mm W 20 - 2 . 0 mm
E 1.0 J power
M3
E 1.0 J
M4
E 1.0 J
M5
E 1.0 J
Length L 10 - 2 . 0 mm Clearance
L 10 - 2 . 0 mm
W 20 - 2 . 0 mm
L 10 - 2 . 0 mm
W 20 - 2 . 0 mm
L 10 - 2 . 0 mm
W 20 - 2 . 0 mm
Width W 20 - 2 . 0 mm Pitch
Use Line: 00010 The number of lines that have been used
Total Line: 10500 Set the maximum number of lines that can be used
10
User manual
4, Operating
1 1
2
_ Width (mm)
10 +
4
+ +
Length(mm) Clearance (mm)
3 20 2.0 4
_ _
3
Pitch (mm)
_
2.0 +
2 Ready
Lines=width/pitch+1
1
Length
Shoot=Lines X
Clearance
2
11
User manual
4, Operating
_ Width (mm)
Press M1-M5 10 +
M1 M2
E 1.0 J E 1.0 J
SAVE L 10 - 2 . 0 mm L 10 - 2 . 0 mm + +
W 20 - 2 . 0 mm W 20 - 2 . 0 mm Length(mm) Clearance (mm)
20 2.0
M3 M4 M5
E 1.0 J E 1.0 J E 1.0 J
L 10 - 2 . 0 mm L 10 - 2 . 0 mm L 10 - 2 . 0 mm _ _
W 20 - 2 . 0 mm W 20 - 2 . 0 mm W 20 - 2 . 0 mm
Pitch (mm)
_
2.0 +
1 Press M1-M5;
2 the corresponding area is yellow;
M1 M2
3 Adjust parameters “width,pitch,length,clearance”; SAVE E 1.0 J
L 10 - 2 . 0 mm
E 1.0 J
L 10 - 2 . 0 mm
W 20 - 2 . 0 mm W 20 - 2 . 0 mm
4 Press “save”. 1
M3 M4 M5
E 1.0 J E 1.0 J E 1.0 J
L 10 - 2 . 0 mm L 10 - 2 . 0 mm L 10 - 2 . 0 mm
W 20 - 2 . 0 mm W 20 - 2 . 0 mm W 20 - 2 . 0 mm
_ Width (mm)
10 +
+ +
Length(mm) Clearance (mm)
M1 M2
20 2.0 SAVE E 1.0 J E 1.0 J
L 10 - 2 . 0 mm L 10 - 2 . 0 mm
W 20 - 2 . 0 mm W 20 - 2 . 0 mm
_ _
2
M3 M4 M5
E 1.0 J E 1.0 J E1.0 J
L 10 - 2 . 0 mm L 10 - 2 . 0 mm L 10 - 2 . 0 mm
W 20 - 2 . 0 mm W 20 - 2 . 0 mm W 20 - 2 . 0 mm
3
Pitch (mm)
_
2.0 +
M1 M2
E 1.0 J E 1.0 J
SAVE L 10 - 2 . 0 mm L 10 - 2 . 0 mm
W 20 - 2 . 0 mm W 20 - 2 . 0 mm
M3 M4 M5
4
E 1.0 J E 1.0 J E 1.0 J
L 10 - 2 . 0 mm L 10 - 2 . 0 mm L 10 - 2 . 0 mm
W 20 - 2 . 0 mm W 20 - 2 . 0 mm W 20 - 2 . 0 mm
12
User manual
4, Operating
Shoot display
3D HIFU ULTRASONIC
Power
2.0 J Output power
DL-1.5
00000 / 10500
width
pitch
Start Clearance
Back Stop
Length
13
User manual
4, Operating
3D HIFU ULTRASONIC
Power
2.0 J
Lines=width/pitch+1
DL-1.5
Length
00000 / 10500
Shoot=Lines X
Clearance
Width Pitch Patient Total Length Clearance
width
pitch
Start
Back Stop
Clearance
Length
2 Ready
Start Start
Stop Stop
14
User manual
5, Parameter settings
DL-1.5mm
15-20 1.5-2.0 8-10 1.5-2.0 1.0-1.5 DL-3.0mm
Face
DL-10mm
20-25 2.0-3.0 10 1.0-2.0 2.5-3.0 DL-13mm
waist
Neck
DL-10mm
20-25 1.0-2.0 10 1.0-2.0 2.5-3.0 DL-13mm
Buttocks
DL-6.0mm
20-25 1.5-2.0 8-10 1.0-2.0 2.0-2.5 DL-8.0mm
Arm
DL-10mm
20-25 1.0-2.0 10 1.0-2.0 2.0-2.5 DL-13mm
Abdomen
DL-10mm
20-25 1.0-2.0 10 1.0-2.0 2.5-3.0 DL-13mm
Thigh
Legs
The form provides a reference, please select the parameters according to the actual situation
15
User manual
6, Technical parameters
Operation voltage: 110V-240V
50Hz/60Hz
Working frequency: 4MHz
MAX power:100VA
16
User manual
9 Troubleshooting
1、No display on the on the apparatus interface:Please check if the
power plug is inserted well or the power switch is turned on.
2、The apparatus operates normally but the press key malfunctions:
Please check if the power supply is connected well.
17
User manual
This apparatus shall be used with enough caution in the following cases,
suggestions from doctors or professionals are necessary prior to use, some
details are listed as follows:
1 Clean the skin at physiotherapeutic part before using the apparatus to
prevent dirt from entering into the skin with the ultrasound or compromising
the penetrability of the ultrasound. Please remove all metallic objects before
the treatment lest unexpected situation impacts curative effects.
2 It is strongly recommended to use medium with certain viscosity to facilitate
better fusion between ultrasound and skin and prevent any reflection arising
from gap from hampering acoustic and energy absorption.
3 Generally each physical therapy period shall be maintained for 15 to 20
minutes, over long time or improper use will cause skin fatigue or bad
absorption. The user may select appropriate duration considering the size of
physiotherapeutic part.
4 The degree of heat at the probe cannot be treated equally as the ultrasonic
power output; it is not advisable to penetrate low-concentration agent or
drug directly, which will lead to dry skin.
5 When this apparatus is being used, the probe must be kept away from the
eye and prohibited from passing the eyeball, the pregnant and patients with
severe heart disease.
6 Unless the doctor’s permission is obtained, the patients in the course of
treatment shall be cautious enough to use this apparatus.
7 Patients with malignant tumor, hemophilia or severe bleeding;
8 Patients with the major disease, skin disease or infectious disease shall be
cautious enough to use this apparatus.
9 Enough caution must be given when such apparatus is used in the heart,
brain, neck nerve centre, spine, eyeball, bleeding and injection.
10 Patients with the hyperthyroidism, asthma and hypertension;
11 Ultrasonic scalpel treatment is not allowed within 15 days after injection
of hyaluronic acid.
12 Please do not wash you face with hot water and don’t massage in one week,
please apply mask every day during the first week, please pay attention to
sun protection,please don’t operate other beauty equipment on treatment
skin in one month,such as RF\Vacuum\ ultrasound\mechanical roller ect.
18
User manual
touch screen
Power suppy
AC 110-240V