Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

Radiol med

DOI 10.1007/s11547-017-0730-1

ABDOMINAL RADIOLOGY

Value of blood flow velocity on color Doppler ultrasonography


for optimization of delay in scanning time on computed
tomography venography in patients with Budd–Chiari syndrome
and inferior vena cava obstruction
Peng‑li Zhou1 · Lei Yan1 · Gang Wu1 · Xin‑Wei Han1,2 · Wen‑Guang Zhang1 

Received: 22 August 2016 / Accepted: 23 January 2017


© Italian Society of Medical Radiology 2017

Abstract  of the delay in scanning time during CT venography to


Purpose To prospectively determine the value of blood ensure good-quality images for the diagnosis of BCS with
flow velocity in the inferior vena cava (IVC) on color Dop- IVC obstruction.
pler ultrasonography for the optimization of the delay in
scanning time after contrast injection during computed Keywords  Budd–Chiari syndrome · Inferior vena cava
tomography (CT) venography in patients with Budd–Chiari (IVC) · Computed tomography (CT) · Color Doppler
syndrome (BCS) with IVC obstruction. ultrasonography · Velocity
Methods We enrolled 122 consecutive BCS patients with
IVC obstruction. All patients underwent color Doppler
ultrasonography, CT venography, and digital subtraction Introduction
angiography (DSA) in that order prior to treatment. The
delay in scanning time during CT venography was set at Budd–Chiari syndrome (BCS) is a clinical disorder arising
120, 180, 240, and 300 s after contrast injection. The cor- as a consequence of the obstruction of the hepatic venous
relation between delay in CT scanning and IVC blood flow outflow [1–3]. In Asia, membranous obstruction of the infe-
velocity on color Doppler ultrasonography was explored. rior vena cava (IVC) and/or hepatic veins (HVs) is the most
Image quality was classified as good, moderate, or poor. common cause of BCS and accounts for up to 60–70% of
Patients with good CT image quality were considered to all BCS patients [4, 5]. Endovascular intervention with bal-
have an optimal delay in scanning time. loon and/or stent is becoming the first choice of treatment
Results  Delays in scanning time of 120, 180, 240, and 300 s for BCS patients with membranous obstruction of the IVC
yielded good-quality images in 2, 7, 49, and 64 patients, and/or HVs in China and Asia [6–10]. Prior to treatment,
respectively. The corresponding IVC blood flow velocities in the type and extent of the obstructive lesion and the status
these patients were 16.10 ± 0.42 cm/s (range 15.8–16.4 cm/s), of the circumambient structures must be clearly displayed
12.90  ± 1.58 cm/s (range, 11–15 cm/s), 7.53 ± 1.35 cm/s using color Doppler ultrasonography, computed tomogra-
(range 5–10 cm/s), and 1.95 ± 1.75 cm/s (range 0–5.5 cm/s). phy (CT), or magnetic resonance imaging (MRI).
Conclusion  IVC blood flow velocity on color Doppler ultra- Currently, CT is the most widely used noninvasive diag-
sonography could serve as a useful tool for the optimization nostic tool for the detection of BCS [11–15]. CT is an
excellent modality to provide information about the liver
parenchyma, cause of BCS, status of the HVs and the sple-
L. Yan is the co-first auhtor. noportal axis, complications of chronic BCS, length of IVC
narrowing, and extra- and intrahepatic collaterals [14, 15].
* Xin‑Wei Han
However, it remains unreliable for the depiction of the IVC
hanxinwei2006@163.com
outline, the circumambient structures, and the exact length
1
Department of Interventional Radiology, The First Affiliated and degree of stenosis due to the use of an inappropriate
Hospital, Zhengzhou University, Zhengzhou, Henan, China interval between contrast injection and the beginning of
2
No.1, East Jian She Road, Zhengzhou 450052, Henan, China scanning during CT venography.

13
Radiol med

BCS with IVC obstruction is a rare form of hepatic venous was used for enhanced CT venography with the following
outflow obstruction at the suprahepatic IVC. Due to the parameters: 0.625 mm × 64 slice, 120 kVp, 100–600 mA,
obstruction, the blood flow in the proximal IVC is slow, tur- and 1.375:1 helical pitch. For enhanced CT venography, a
bulent, or reversed. Thus, the blood flow velocity in the IVC standard dose of 0.7 ml/kg body weight of contrast mate-
differs from person to person. To obtain clear images of the rial in 40 ml normal saline was administered using a power
IVC on CT venography, the technique must be optimized such injector at a rate of 4 ml/s through a 22-gage needle into
that the interval between contrast injection and the beginning the antecubital vein. When the concentration of the con-
of scanning is appropriate for the blood flow velocity in the trast medium in the ascending aorta reached 100 HU, the
IVC for a given patient. However, optimization of the delay CT device automatically scanned an area spanning the
in scanning time after contrast injection according to the lower third of the heart to the inferior margin of the kid-
blood flow velocity in the IVC, as determined on color Dop- ney. CT images were obtained during the hepatic arterial
pler ultrasonography, has been rarely reported in the literature. phase (25 s after administering the nonionic contrast mate-
Hence, the purpose of the study was to determine the value of rial), portal venous phase (60 s), and delayed phase (120,
IVC blood flow velocity measured using color Doppler ultra- 180, 240, and 300 s). The acquired image data sets were
sonography for the optimization of the delay in scanning time then transferred to a workstation (GE AW4.3, GE Medical),
during CT venography in BCS patients with IVC obstruction. where three-dimensional (3D) image reconstruction was
performed with a 732 × 732 matrix. The reconstruction
included the development of oblique, coronal, and sagittal
Materials and methods maximum-intensity projections (MIPs), multiplanar recon-
struction (MPR), and 3D volume-rendered (VR) images of
Study design the thoracic and abdominal vascular structures.

The institutional review board of our hospital approved DSA


the study protocol, and all patients or their family mem-
bers provided informed consent before participation in the DSA was performed by an interventional radiologist fol-
study. From July 2012 to December 2015, 122 consecu- lowing the CT examinations. The access routes for angiog-
tive BCS patients with IVC obstruction were prospectively raphy were the trans-femoral approach and/or trans-jugular
enrolled in our study. All patients underwent color Doppler approach. IVC angiography was performed using a 5-Fr
ultrasonography followed by CT venography before endo- straight catheter with multiple side holes (Cook, Bloom-
vascular treatment. In addition, digital subtraction angiog- ington, USA) to characterize the morphology of the IVC
raphy (DSA) was performed after CT venography was used and the extent of the obstruction. Two observers, who were
to confirm the diagnosis of BCS. Patients with severe liver blinded to all clinical and previous imaging results, identi-
failure or terminal hepatic carcinoma were excluded from fied and analyzed all the DSA data together.
the study. Patients were divided into two groups accord-
ing to the DSA findings: partial IVC obstruction group and Image review
complete IVC obstruction group.
After reconstruction, MIP and VR images were interac-
Color Doppler ultrasonography tively evaluated by two senior radiologists working inde-
pendently. Both raters were unaware of the clinical out-
Prior to CT venography, the blood flow velocity in the IVC comes of the patients, and the DSA and color Doppler
was measured in all patients using color Doppler ultra- results. They independently analyzed all CT venography
sonography (Vivid 7 Pro, GE, USA). Blood flow velocity datasets on an offline workstation from multiple on-screen
was measured in the post-hepatic IVC segment avoiding viewing angles. The source images and the MIP, MPR,
both the orifices of the HVs and collateral branches. The and VR images were presented on screen, thus allowing
blood flow velocity index was quantified as the average for adjustment of the appropriate threshold of the window
peak velocity. If the blood flow in the IVC was reversed, width and level. In the presence of rater disagreement in
the velocity was recorded as zero. the detection of BCS with IVC obstruction, a consensus or
majority decision was obtained.
CT venography
Quantitative analysis
All CT venography examinations were performed on a
64-row CT scanner (LightSpeed VCT or Discovery CT750 Image quality was evaluated on a per-patient basis and clas-
HD, GE Healthcare). The smart preset scan technique sified as good (diagnostic quality with vascular structures

13
Radiol med

Table 1  Baseline characteristics of the 122 BCS patients included in the study


Characteristic All patients Patients with partial Patients with complete P value
(n = 122) IVC obstruction IVC obstruction
(n = 53) (n = 69)

Age (years) 42.09 ± 10.80 43.57 ± 9.04 40.96 ± 11.92 0.187


(range 12–74) (range 28–63) (range 12–72)
Female/Male (no.) 45/77 20/33 26/43 0.995
Interval between clinical symptoms 79.28 ± 101.04 (0.25–480) 74.54 ± 93.99 (0.25–360) 82.92 ± 106.67 (0.25–480) 0.652
and treatment (month)
Length of obstruction (mm) 13.78 ± 14.70 (0.5–56) 16.53 ± 15.53 (0.5–50) 10.20 ± 12.82 (0.8–56) 0.018
Presence of thrombosis (no.) 8 (6.6) 1 (1.9) 7 (10.1) 0.050
IVC blood flow velocity (cm/s) 5.05 ± 3.95 (0-16.4) 8.18 ± 3.10 (1-16.4) 2.65 ± 2.63 (0-9) <0.001
Symptoms at diagnosis, no. (%)
 Abdominal distention 36 (29.5) 20 (37.7) 16 (23.2) 0.081
 Abdominal pain 9 (7.4) 4 (7.5) 5 (7.2) 0.950
 Varices in abdominal wall 60 (49.2) 25 (47.2) 35 (50.7) 0.697
 Ascites 35 (28.7) 18 (34.0) 17 (24.6) 0.259
 Jaundice 10 (8.2) 4 (7.5) 6 (8.7) 0.819
 Hepatic encephalopathy 2 (1.6) 1 (1.9) 1 (1.4) 0.850
 Hepatomegaly 45 (36.9) 19 (35.8) 26 (37.7) 0.835
 Splenomegaly 91 (74.6) 40 (75.5) 51 (73.9) 0.845
 GI bleeding 10 (8.2) 3 (5.7) 7 (10.1) 0.371
 Lower limb edema 49 (40.2) 23 (43.4) 26 (37.7) 0.523
 Varices in lower limbs 28 (23.0) 10 (18.7) 18 (26.1) 0.347
 Pigmentation of lower limbs 43 (35.2) 16 (30.2) 27 (39.1) 0.305
 Ulcer in lower limbs 10 (8.2) 5 (9.4) 5 (7.2) 0.662
 Child–Pugh score 6.45 ± 1.36 (5–11) 6.57 ± 1.40 (5–11) 6.35 ± 1.32 (5–11) 0.393

Values shown are expressed as mean ± standard deviation and ranges, unless otherwise indicated
BCS Budd–Chiari syndrome, IVC inferior vena cava, GI gastrointestinal

clearly visualized on MIP, MPR, and VR images), moder- κ  = 0.21–0.40; moderate, κ  = 0.41–0.60; substantial,
ate (vascular structures mildly limited on 3D reconstruction κ  = 0.61–0.80; and almost perfect, κ  = 0.81–1.00.
due to contrast bolus or motion, but diagnostic quality with Kappa statistics with 95% confidence intervals (CIs)
moderate confidence), or poor (vascular structures poorly were used to assess agreement between the angio-
visualized, diagnostic quality with low confidence). graphic results.

Statistical analysis

All data are presented as means ± standard deviations. Results


Dichotomous and categorical data were reported as num-
bers and percentages. The χ2 test was used for categorical Patient population
variables, and the unpaired t test was used for continuous
variables. Statistical analyses were performed using SPSS From July 2012 to December 2015, a total of 122 patients
statistical software (version 13.0 for Windows, SPSS, Chi- were enrolled in our study. They included 77 male
cago, IL, USA). P < 0.05 was considered to indicate a sig- patients with a mean age of 42.78 ± 11.87 years (range
nificant difference. 12–72 years) and 45 female patients with a mean age of
The kappa statistic was used for comparison of 40.96  ± 8.75 years (range 14–63 years). The baseline
observer performance in the assessment of image characteristics of the patients are summarized in Table 1.
quality, IVC outline, and the upper and lower mor- There were no significant differences in clinical signs and
phology of the IVC obstruction. Interobserver agree- symptoms between patients with partial and complete
ment was classified as follows: slight, κ  ≤ 0.2; fair, IVC obstruction.

13
Radiol med

Fig. 1  Budd–Chiari syndrome
(BCS) with complete inferior
vena cava (IVC) obstruction in
a 55-year-old man. a A color
Doppler sonogram prior to treat-
ment shows complete obstruc-
tion of the IVC with a blood
flow velocity of 1.5 cm/s in the
post-hepatic IVC. Note also an
irregular filling defect in the
IVC (arrow). b Coronal muli-
planar reconstruction with good
image quality was obtained
using an optimal delay in scan-
ning time of 300 s; the image
shows complete obstruction
of the IVC and a thrombus in
the IVC (arrow). c An inferior
vena cavogram via the trans-
femoral approach demonstrates
complete obstruction of the
IVC and the thrombus (arrow).
The lower morphology of the
IVC obstruction on digital
subtraction angiography (DSA)
corresponded almost exactly
with the morphology depicted
on computed tomography (CT)
venography. d An inferior vena
cavogram after rupture of the
membrane shows that the extent
of the IVC obstruction on DSA
corresponds almost exactly with
that depicted on CT venography

DSA patients, 5–10 cm/s in 39 patients, 11–15 cm/s in 6 patients,


and > 15 cm/s in 3 patients. The IVC blood flow velocity in
According to the reference standard, a partial IVC obstruc- patients with complete IVC obstruction was < 5 cm/s in 56
tion was observed in 53 BCS patients and a complete patients, 5–10 cm/s in 10 patients, 11–15 cm/s in 3 patients,
IVC obstruction was observed in 69 patients. The mean and > 15 cm/s in 0 patients.
length of the IVC obstruction in all 122 patients was
13.78 ± 14.70 mm (range 0.5–60 mm). The length of the CT venography and color Doppler ultrasonography
obstruction was significantly smaller in patients with par-
tial IVC obstruction than in patients with complete IVC Patients with good image quality on CT venography were
obstruction (P < 0.05, Table 1). considered to have an optimal delay in scanning time for
an accurate diagnosis of BCS with IVC obstruction. Delay
Color Doppler ultrasonography times of 120, 180, 240, and 300 s after contrast injection
yielded good-quality images in 2, 7, 49, and 64 patients,
The mean blood flow velocity in the IVC on color Doppler respectively. The corresponding mean IVC blood flow
ultrasonography was 5.05 ± 3.95 cm/s (95% CI 4.35–5.76; velocities in these patients were 16.10 ± 0.42 cm/s (range
range 0–16.4 cm/s). The IVC blood flow velocity was sig- 15.8–16.4 cm/s), 12.90 ± 1.58 cm/s (range 11–15.3 cm/s),
nificantly greater in patients with partial obstruction than in 7.53 ± 1.35 cm/s (range 5–10 cm/s), and 1.95 ± 1.75 cm/s
patients with complete obstruction (P < 0.05, Table 1). (range 0–5.5 cm/s; Fig. 1).
The IVC blood flow velocity was < 5 cm/s in 61 Delay times of 120, 180, 240, and 300 s after contrast
patients, 5–10 cm/s in 49 patients, 11–15 cm/s in 9 patients, injection yielded good-quality images in 2, 7, 37, and 7
and > 15 cm/s in 3 patients. The IVC blood flow velocity patients with partial IVC obstruction, respectively. The
in patients with partial IVC obstruction was < 5 cm/s in 5 corresponding mean IVC blood flow velocities in these

13
Radiol med

patients were 16.10 ± 0.42 cm/s (range 15.8–16.4 cm/s), Usually, CT venography images are obtained at 180 s
12.90 ± 1.58 cm/s (range 11–15.3 cm/s), 7.65 ± 1.27 cm/s after administering the contrast medium. An inappropriate
(range 5.4–10 cm/s), and 4.0 ± 1.43 cm/s (range delay in scanning during CT venography may result in the
1–5.4 cm/s). Delay times of 120, 180, 240, and 300 s after following pitfalls: (1) The contrast medium may flow along
contrast injection yielded good-quality images in 0, 0, 12, the IVC wall, resulting in laminar flow due to stasis of the
and 57 patients with complete IVC obstruction, respec- blood flow, and this phenomenon might be erroneously
tively. The corresponding mean IVC blood flow velocities identified as IVC thrombolysis. (2) The enhancement of the
in these patients were 0, 0 cm/s, 7.18 ± 1.57 cm/s (range IVC may be insufficient due to the lack of an optimal delay
5–9 cm/s), and 1.70 ± 1.62 cm/s (range 0–5.0 cm/s). in scanning, resulting in poor quality of 3D reconstructed
The kappa value for the assessment of CT image quality images. (3) Multiple scans will be required to obtain good
for the detection of IVC obstruction was 0.91, suggesting image quality, which increases the exposure of the patient
excellent interobserver agreement. The kappa value for the to radiation.
assessment of CT image quality for the detection of partial To date, few studies have evaluated the clinical implica-
IVC obstruction and complete IVC obstruction was 0.89 tions of IVC blood flow velocity on the optimization of the
and 0.92, respectively. delay in CT scanning time to obtain good-quality images
for the diagnosis of BCS. In this study, we explored the
CT venography and DSA optimal delay in scanning time on CT venography accord-
ing to the IVC blood flow on color Doppler ultrasonogra-
CT venography had a good correlation with DSA for the phy. The results showed that the optimal delay in scanning
diagnosis of BCS with IVC obstruction, with an accuracy time was 120 s in patients with an IVC blood flow velocity
of 100%. The extent and morphology of the IVC obstruc- of > 15 cm/s; 180 s in patients an IVC blood flow velocity
tion on CT venography corresponded almost exactly with of 10–15 cm/s; 240 s in patients with an IVC blood flow
those on DSA. velocity of 5–10 cm/s; and 300 s in patients with an IVC
blood flow velocity of ≤ 5 cm/s.
It should be noted that not all patients with an IVC
Discussion blood flow velocity of > 15 cm/s had an optimal delay in
scanning time of 120 s; in one patient, the optimal delay
This prospective study was designed to test the hypothesis was 180 s. Such exceptions also occurred in patients with
that good image quality on CT venography can be obtained IVC blood flow velocities of 10–15, 5–10, and ≤ 5 cm/s.
by optimization of the delay in scanning time according to Owing to differences in the examination method, manner,
the IVC blood flow velocity on color Doppler ultrasonog- and time, the status of the IVC blood flow velocity may
raphy. Although the optimal delay in scanning time for vary, so the above exceptions are understandable. Because
good image quality on CT venography did not correspond the scanning time of CT venography is very short, the
exactly with the IVC blood flow velocity on color Doppler optimal delay in scanning time for obtaining good-quality
ultrasonography, good image quality could be obtained images may not be exactly 120, 180, 240, or 300 s, but
in the majority of the BCS patients with IVC obstruction may be some time within the 120–180, 180–240, or 240–
using multiple CT venography scans. Moreover, the exten- 300 s intervals. In this situation, the quality of the images
sion and morphology of IVC obstruction on 64-row CT obtained may be between good and moderate. In this
venography images corresponded almost exactly with the study, there were only 7 patients (6.3%) in this situation;
angiographically depicted extension and morphology of however, the CT results correlated with the DSA findings
IVC obstruction. in all patients.
BCS is not an uncommon cause of hepatic venous out- This study had some limitations. First, this was a single-
flow obstruction leading to liver disease in the Far East, center study, and no control group was included. Second,
the Middle East, India, and Africa [16]. IVC obstruction, the number of patients was relatively small; a larger sam-
with or without the involvement of the HVs, is the cause ple size is required to reach definitive conclusions. Third, to
of BCS in the majority of cases in these geographical areas ensure that we obtained good-quality images, we used four
[17]. In BCS patients with IVC obstruction, the blood flow different scan times for CT venography, and this may have
velocity in the IVC varies according to the degree of steno- exposed the patients to a greater amount of radiation.
sis. In patients with partial IVC obstruction, the blood flow In conclusion, our findings suggest that the IVC blood
may be relatively fast, while in patients with complete IVC flow velocity on color Doppler ultrasonography could serve
obstruction, the blood flow may be very slow or reversed. as a useful tool for the optimization of the delay in scanning
Hence, it is difficult to select an optimal delay in scanning time during CT venography to obtain good-quality images
time for obtaining good-quality CT venography images. for the accurate diagnosis of BCS with IVC obstruction.

13
Radiol med

Compliance with ethical standards  6. Furui Sl, Sawada S, Irie T et al (1990) Hepatic inferior vena cava
obstruction: treatment of two types with Gianturco expandable
metallic stents. Radiology 176:665–670
Conflict of interest statement  Peng-li Zhou declares that he has no
7. Xu K, He FX, Zhang HG et al (1996) Budd–Chiari syndrome
conflict of interest.
caused by obstruction of the hepatic inferior vena cava: immedi-
Gang Wu declares that he has no conflict of interest. ate and 2-year treatment results of transluminal angioplasty and
Xin-wei Han declares that he has no conflict of interest. metallic stent placement. Cardiovasc Interv Radiol 19:32–36
Lei Yan declares that he has no conflict of interest. 8. Ding PX, Han XW, Wu G et al (2010) Outcome of a retrieval
Wen-Guang Zhang declares that he has no conflict of interest. stent filter and 30-mm balloon dilator for patients with Budd–
Chiari syndrome and old inferior vena cava thrombosis: a pro-
Ethical approval All procedures performed in studies involving spective pilot study. Clin Radiol 65:629–635
human participants were in accordance with the ethical standards of 9. Meng QY, Sun NF, Wang JX et al (2011) Endovascular treatment
the institutional and/or national research committee and with the 1964 of Budd–Chiari syndrome. Chin Med J (Engl) 124:3289–3292
Helsinki declaration and its later amendments or comparable ethical 10. Han G, Qi X, Zhang W et al (2013) Percutaneous recanaliza-
standards. tion for Budd–Chiari syndrome: an 11-year retrospective study
on patency and survival in 177 Chinese patients from a single
center. Radiology 266:657–667
Informed consent  Informed consent was obtained from all individual
11. Vogelzang RL, Anschuetz SL, Gore RM (1987) Budd–Chiari
participants included in the study.
syndrome: CT observations. Radiology 163:329–333
12. Virmani V, Khandelwal N, Kang M et al (2009) MDCT venogra-
phy in the evaluation of inferior vena cava in Budd–Chiari syn-
drome. Indian J Gastroenterol 28:17–23
References 13. Camera Ll, Mainenti PP, Di Giacomo A et al (2006) Triphasic
helical CT in Budd–Chiari syndrome: patterns of enhancement
1. Janssen HL, Garcia-Pagan JC, Elias E, Mentha G, Hadengue A, in acute, subacute and chronic disease. Clin Radiol 61:331–337
Valla DC (2003) Budd–Chiari syndrome: a review by an expert 14. Cai SFl, Gai YH, Liu QWl (2015) Computed tomography angi-
panel. J Hepatol 38:364–371 ography manifestations of collateral circulations in Budd–Chiari
2. Murad SD, Plessier A, Hernandez-Guerra M, Fabris F, Eapen syndrome. Exp Ther Med 9:399–404
CE, Bahr MJ, Trebicka J, Morard I, Lasser L, Heller J, Hadengue 15. Cho OKl, Koo JH et al (1996) Collateral pathways in Budd–Chi-
A (2009) Etiology, management, and outcome of the Budd–Chi- ari syndrome: CT and venographic correlation. AJR Am J Roent-
ari syndrome. Ann Intern Med 151:167–175 genol 167:1163–1167
3. Menon KN, Shah V, Kamath PS (2004) The Budd–Chiari syn- 16. De BKl, De KK, Sen S et al (2000) Etiology based prevalence
drome. N Engl J Med 350:578–585 of Budd–Chiari syndrome in eastern India. J Assoc Phys India
4. Xu K, Feng B, Zhong H et al (2003) Clinical application of inter- 48:800–803
ventional techniques in the treatment of Budd–Chiari syndrome. 17. Eapen CE, Mammen T, Moses V et al (2007) Changing pro-
Chin Med J (Engl) 116:609–615 file of Budd–Chiari syndrome in India. Indian J Astroenterol
5. Qiao T, Liu CJ, Liu C et al (2005) Interventional endovascular 26(1):77–81
treatment for Budd–Chiari syndrome with long-term follow-up.
Swiss Med Wkly 135:318–326

13

You might also like