Troubleshooting Guide
Stand - Table
Artis one with VA10 SW
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed. Please contact your local Siemens
office to order a current version or refer to our website http://www.health‐
care.siemens.com.
Disclaimer
Siemens provides this documentation “as is” without the assumption of any liability under
any theory of law.
The content described herein shall be used by qualified personnel who are employed by
Siemens or one of its affiliates or who are otherwise authorized by Siemens or its affiliates
to use such documents.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are not entitled to use this documen‐
tation without prior written authority.
Copyright
“© Siemens, 2014” refers to the copyright of a Siemens entity such as:
Siemens Healthcare GmbH - Germany
Siemens Aktiengesellschaft - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany
1 General Information 4
1.1 Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.1 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2.1 Required Aids and Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
5 Error handling 17
6 Procedures 18
1.1 Notes
WARNING
Improperly established protective ground conductor connections during installation,
maintenance, or service.
Personal injury due to electric shock.
Protective ground conductor tests shall be performed in accordance with the in‐
structions in the technical documentation prior to initially switching on the device/
system, as well as after finishing all types of service / installation activities and prior
to handing over the system to the customer.
hm_serv_ProtectiveEarthMeasurement
CAUTION
Suspending, or motorized system parts (e.g. table or C-arm) start moving unintention‐
ally when accessing eletromechanical parts, e.g. while replacing drives or brakes.
Risk: Personal injury
Press the service or emergency stop switch, and secure suspending parts (e.g. ta‐
ble or C-arm) before starting critical service activity.
hm_serv_UnintentMovePart
WARNING
Misdiagnosis and incorrect treatment because of incorrect results from the quantifica‐
tion software.
This can lead to death or serious bodily injury!
Perform the geometry adjustment carefully.
hm_serv_GeometryAdjust
WARNING
Room size is incorrectly configured.
Injury to personnel or damage to the system possible.
Following installation and following service work, check the collision zones.
hm_serv_CheckStandPositions
1.2 Requirements
The function description for the Artis one also includes an overview of the stand and table
components.
a) Move the C-arm and table in the axes displayed in the list (use SCM / TCM).
b) Press Go.
The values are read out.
Fig. 2: Position check, C-Arm and Table, test result
c) Check to make sure the deviation between sensor and motor is within tolerance.
d) Continue with the System Troubleshooting Guide and motion control trouble‐
shooting.
a) Move the stand in the axes displayed in the list (use SCM / TCM).
b) Press Go.
The values are read out.
Fig. 3: Position Check, stand, test result
c) Check to make sure the deviation between sensor and motor is within tolerance.
d) Continue with the System Troubleshooting Guide and motion control trouble‐
shooting.
The test results would need to be evaluated by the developer. This is only necessary if
the CSE needs to escalate the issue.
4. Continue with the System Troubleshooting Guide and motion control troubleshoot‐
ing.
4. Continue with the System Troubleshooting Guide and motion control troubleshoot‐
ing.
The tube cover and pivot base rotation proximity switch checks are combined.
Both will be shown as active even if only one of them is pressed.
4. Continue with the System Troubleshooting Guide and motion control troubleshoot‐
ing.
4. Continue with the System Troubleshooting Guide and motion control troubleshoot‐
ing.
4. Continue with the System Troubleshooting Guide and motion control troubleshoot‐
ing.
4. Continue with the System Troubleshooting Guide and motion control troubleshoot‐
ing.
Detailed instructions for dealing with movement problems are given in the System Trou‐
bleshooting Guide.
Please refer to the System Troubleshooting Guide ( Troubleshooting with the VA10
SW / AXA6-000.840.01).
Detailed instructions for error handling are given in the System Troubleshooting Guide.
Please refer to the System Troubleshooting Guide ( Troubleshooting with the VA10
SW / AXA6-000.840.01).
Chapter Changes
Stand and Table Tests Updated to conform to the user interface changes.