Setting to Work Mount Covers Clea Ceiling

LIST OF PAGES & DRAWINGS MODULE CODE NUMBER 4522 981 01913

1 ... 5 (04.0)

Mount Covers Clea Ceiling

Contents

1. INTRODUCTION .....................................................................................................................................2
1.1. Required tools..........................................................................................................................................2

2. MOUNT COVERS....................................................................................................................................3

3. POSITION STICKERS ............................................................................................................................5

4522 981 01913 (04.0) 1


© 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Mount Covers Clea Ceiling Setting to Work

1. INTRODUCTION

This module describes the procedure to mount the covers of the Clea Ceiling Stand.

1.1. REQUIRED TOOLS

Standard Toolkit (TC129).

2 (04.0) 4522 981 01913


© 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Setting to Work Mount Covers Clea Ceiling

2. MOUNT COVERS

NOTE

Mounting procedure and installation materials are exactly the same for systems with I.I. and FD.

• Mount the three covers on top of the L-arm (fitted with Velcro).
• Mount the L-arm side cover (3 countersunk screws, 3 countersunk washers and 3 anti-loss washers)
• Mount the rear cover on the L-arm (6 non-removable screws).
• Insert 6x decorative plastic covers (sealing caps) in the L-arm rear cover.
• Insert 4x sealing cap M16 in the holes of the C-arc junction that have been used for the fixation of the
installation frame.

Figure 1. Mount Covers

4522 981 01913 (04.0) 3


© 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Mount Covers Clea Ceiling Setting to Work

• Mount the metal cover on both sides of the longitudinal carriage (2 screws and 2 washers).
• Mount the rotation cover to the L-arm rotation point (click-construction).
• Insert 4x decorative plastic covers (sealing caps) to bottom side of longitudinal carriage.

Figure 2. Mount Covers

4 (04.0) 4522 981 01913


© 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Setting to Work Mount Covers Clea Ceiling

3. POSITION STICKERS

• Place the Stand in the Park Position (PP).


• Mount the stickers on the rails so they match the arrows on the longitudinal carriage.
(the arrow-stickers are delivered in the blister packaging delivered in the L-arm)
• Repeat the procedure for Work Position 1 (WP1) and Work Position 2 (WP2).

PARK AREA WORK AREA

PP W WP2 WP1 (Can be equal to LEP)

HEP LEP
Rubber Rubber
stops stops

>=100mm >=100mm >=100mm

+100mm

4522 981 01913 (04.0) 5


© 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED