Clark - Calibration Report (Nexzus 3) 2
Clark - Calibration Report (Nexzus 3) 2
Clark - Calibration Report (Nexzus 3) 2
The following Results where obtain using The Comparison Method on known Simulated Parameters and
Output Readings on Data Analyzers
REFERENCE ACTUAL
CALIBRATION DATA
SETTING READING
Tidal Volume 50mL 48ml
40mL 40ml
20mL 20ml
PEEP 2 2.42
5 5.04
RATE 40 40
60 60
TRACEABILITY
TEST EQUIPMENT & SIMULATOR USED SERIAL NUMBER VALIDITY BY
CERTIFICATE
TSI Certifier FA PLUS 40811901007 TSI Inc.
DIGITAL TESTER
AC/DC Voltage Detector M100
Neonate Test Lung
Neonate Tubings
Adult Tubings for Flow Calibration