CALIBRATION REPORT

CUSTOMER : Medical City Clark DATE 6-Sep-2021


ADDRESS: 100 Gatwick Gateway Clark Global City, Clark Freeport, Pampanga NEXT DUE DATE
EQUIPMENT TYPE : Ventilator

BRAND: PNEUMOVENT HOUR READING


MODEL No. : NEXZUS 3 INVENTORY No
SERIAL Number: V6-10703059 DEPARTMENT NICU

The following Results where obtain using The Comparison Method on known Simulated Parameters and
Output Readings on Data Analyzers

REFERENCE ACTUAL
CALIBRATION DATA
SETTING READING
Tidal Volume 50mL 48ml
40mL 40ml
20mL 20ml

FiO2% 100% 100%


tolerance +/- 2 -4 % 90% 91%
30% 30%

PEEP 2 2.42
5 5.04
RATE 40 40
60 60

ELECTRICAL CONDITIONS READING


GROUND RESISTANCE 0.01-500 ohms 0.11 ohms
CHASSIS LEAKAGE 50-500uA 50
BATTERY CHARGING ( AC unplug) OK/PASSED

TRACEABILITY
TEST EQUIPMENT & SIMULATOR USED SERIAL NUMBER VALIDITY BY
CERTIFICATE
TSI Certifier FA PLUS 40811901007 TSI Inc.
DIGITAL TESTER
AC/DC Voltage Detector M100
Neonate Test Lung
Neonate Tubings
Adult Tubings for Flow Calibration

BIOMEDICAL TECHNICIAN -ENGINEER

JAYSON C. ANTONIO, ECE

SIGNATURE OVER PRINTED NAME / DATE