Advanced Program | Irrigation Tubing Setup Guide

1. Inspect tubing to ensure no visual abnormalities appear in irrigation tubing.

2. Spike the fluid bag and pinch the tubing below the
drip chamber

3. Ensure that arrows on tubing are pointing in same direction as arrows on pump head

4. Mount the tubing and ensure it is within the tubing track, within the
side hooks.

5. Push the tubing into the bubble sensor slot and close the door.

© Medical Device Business Services, Inc., 2020. All rights reserved. 163554-201223
CONFIDENTIAL. FOR INTERNAL USE ONLY. NOT FOR USE WITH ANY CONSUMER OR FOR EXTERNAL DISTRIBUTION.
Advanced Program | Irrigation Tubing Setup Guide

6. Close the door, ensuring the tubing remains within the track and bubble
sensor

7. Press and hold the Flush Button, until you see the image below.
• While flushing make sure to inspect the length of the irrigation line
for air bubbles
• Ensure the drip chamber is flowing once low flow has started.

8. Connect ablation catheter to the cable before attaching fluids


• This ensures pins stay dry and additional errors are avoided

9. Attach tubing to the catheter irrigation lumen.

10. Flush final time to ensure ablation catheter is primed and there are
not bubbles in the irrigation tubing or ablation catheter