Professional Documents
Culture Documents
Users Manual Elyte-1
Users Manual Elyte-1
Electrolyte Analyzer
Operator’s Manual
Product Information
Product model: E-Lyte Plus
Product name: HTI E-Lyte Plus Electrolyte Analyzer
Manufacturer: High Technology, Inc.
Symbols
Symbols used in the Manual
Notes contain additional information or tips when using the instrument. All
recommendations must be followed carefully to ensure your instrument work correctly
and to avoid unnecessary personal injury.
The symbol means the instrument packaged should not be upended at any time
This means that instrument should handle with care in the course of transportation,
so as not to damage it.
The symbol means that the environment of instruments must be damp-proof in the
course of transport, and instrument must be kept in a dry environment.
This symbol indicate that the analyzer should be protected from overexposure to
sunlight
This symbol indicates the Power Switch. “O” is Power Off and “I” is Power On
RS232 This symbol indicates the data connection for the analyzer
The symbols for “SERIAL NUMBER”, The serial number shall be after or
below the symbol, adjacent to it
The symbol indicates the manufacturer and its address, after which are shown its
name and address
This symbol denotes that this product conforms to the directive 98/79/EC (IVD-
directive)
Declaration
High Technology, Inc. reserves the right to change the product described in this Operating Manual. All information
contained in this Operating Manual is subject to change without prior notice.
Manufacturer’s Responsibility
High Technology, Inc. (HTI) is responsible for the safety, reliability and performance of the device, and will provide
warranty under the following conditions:
Free services apply to those products with applicable items within warranty period. For those beyond the
description of warranty conditions, HTI will charge for service. For all returned goods to HTI for service, customers
should pay for the transportation and any applicable customs fees.
Warranty
The Analyzer is covered by a one-year limited warranty from date of the purchase. The electrodes carry a one-year
limited warranty. Consumables are excluded from this warranty.
Return Procedures
If a return is necessary, take the following steps:
Table of Contents
1. INTRODUCTION ................................................................................................................................................ 1
1.1 OVERVIEW ................................................................................................................................................ 1
1.2 BRIEF INTRODUCTION ............................................................................................................................ 1
1.3 DESCRIPTION OF ELECTRODE.............................................................................................................. 1
1.3.1 Electrode Principles ........................................................................................................................... 2
1.3.2 Measuring Principles .......................................................................................................................... 2
1.4 DESCRIPTION OF REAGENT .................................................................................................................. 3
1.5 SAMPLE COLLECTION AND HANDLING ................................................................................................ 4
1.5.1 Safety ................................................................................................................................................. 4
1.5.2 Serum Samples .................................................................................................................................. 4
1.5.3 Urine Samples .................................................................................................................................... 4
1.6 SAFETY INFORMATION ........................................................................................................................... 5
1.6.1 Danger ................................................................................................................................................ 5
1.6.2 Warning .............................................................................................................................................. 5
1.6.3 Caution ............................................................................................................................................... 5
1.6.4 Attention ............................................................................................................................................. 5
2 DESCRIPTION OF ANALYZER ........................................................................................................................ 6
2.1 EXTERNAL AND INTERNAL FEATURES ................................................................................................ 6
2.1.1 Front Panel ......................................................................................................................................... 6
2.1.2 Rear Panel ......................................................................................................................................... 6
2.1.3 Internal Components .......................................................................................................................... 7
2.2 OPERATING ENVIRONMENT .................................................................................................................. 8
2.2.1 Placement .......................................................................................................................................... 8
2.2.2 Operating Conditions ......................................................................................................................... 8
2.2.3 Influence on other Equipment ............................................................................................................ 8
3 INSTALLATION ................................................................................................................................................. 9
3.1 EQUIPMENT UNPACKING ....................................................................................................................... 9
3.2 INSPECTION OF INSTRUMENT AND ACCESSORIES ........................................................................... 9
3.3 REAGENT PACK ..................................................................................................................................... 10
3.4 TUBING .................................................................................................................................................... 10
3.5 ELECTRODES ......................................................................................................................................... 11
3.5.1 Filling or Refilling Electrode Internal Solution .................................................................................. 11
3.5.2 Filling or Refilling the Reference Electrode ...................................................................................... 12
3.5.3 Electrode Installation ........................................................................................................................ 13
3.6 PRINTER PAPER .................................................................................................................................... 15
3.7 CONNECTING AC POWER .................................................................................................................... 15
3.8 TURNING ON THE UNIT ......................................................................................................................... 16
3.9 ACTIVATION ............................................................................................................................................ 16
3.9.1 Priming Reagent Pack ..................................................................................................................... 16
3.9.2 Activating Electrodes ........................................................................................................................ 18
3.9.3 Running QC for the First Time ......................................................................................................... 19
4 OPERATION .................................................................................................................................................... 20
4.1 MAIN MENU ............................................................................................................................................. 20
4.1.1 ISE Calibration ................................................................................................................................. 20
4.1.2 Quality Control.................................................................................................................................. 21
4.1.2.1 Test ................................................................................................................................................................... 22
4.1.2.2 Factor ................................................................................................................................................................ 23
4.1.2.3 QC Calibration................................................................................................................................................... 24
4.1.2.4 Manual QC Calibration ...................................................................................................................................... 25
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL I
Document Number OM-E-ELT-P-10
Revision Level Rev.13
High Technology, Inc.
Prepared By J.S. Bolio
Approved By S. Titov
Effective date 23Oct17
1. INTRODUCTION
1.1 Overview
The purpose of this manual is to provide operating, maintenance, and repair information to users. This manual
illustrates detailed performance indexes, routine installation, operation, maintenance procedures, and
important safety information. To ensure safety standards and performance functions carefully read this
manual prior to using the analyzer.
The instrument is intended for use in hospitals and labs to measure K+, Na+, Cl-, Ca2+, Li, Mg2+ ions, as well
as pH values, serum, and urine solution (diluted). This product has multi-parameter functions which can be
selected, combined and configured by the user according to the hospital or labs specific requirements. The
instrument has 7 combination types:
“Ion-Selective Electrode” means that each electrode is only sensitive to one type of ion. For example, the
Na electrode is only sensitive to Na+, but not sensitive to other ions. The key component of the electrode is
the ion-selective membrane. The two sides of the membrane get contact from a sample and the internal
Electrode Internal Solution respectively. The side touching the sample is responding to the ion
concentrations change. The other side touches the internal Electrode Internal Solution, the conversion from
ion conduction to electron conduction is carried out by Ag/AgCl inner electrode. The reference electrode
provides the reference potential to complete the measuring circuit. This electric potential does not change
along with ion concentration, therefore providing a standard reference for measuring potential difference.
1 Measured Solution
2 Conjunctive Point
3 Reference Electrode
ISE is sensitive to the ion activity in the sample. When the ion concentration in the sample is below 10-4M
and activity coefficient is close to 1, the difference between ion activity and concentration can be ignored
(when concentration goes over 10-4M, the activity coefficient decreases, and the difference increases.)
When the ISE contacts the measured solution, the measured ion in the sample goes to the ISE membrane
due to the diffusion effects of the concentration difference, which creates a potential between the measure
electrode and the reference electrode. The potential that an ideal ISE creates for the “X” ion can be described
by the Nernst formula:
When the activity coefficient is 1 and temperature is 25oC (298°K), the electrode potential is proportional to
the logarithm of ion concentration “C”, which can be described as the following formula:
59.12
E=E0 + log10 C(x)
Z
It is manifested that before applying ISE to perform the measurement, the values of potential (E0) and slope
(59.12/Z is the theoretical value) must be determined first through the two-point calibration by using two
standard solutions. Moreover, the ion activities in the standard solution and the testing sample must be kept
close enough to ensure the accuracy of measurement. The following figure shows the measurement
principle:
MV
Calibrating Curve
MV. B
MV. S
MV. A
Cleaning Solution For cleaning the analyzer measuring system before shutdown.
• For serum samples, DO NOT use any surfactant, anticoagulants, etc., otherwise
it will cause incorrect results, or even destroy the sensor.
1.6.1 Danger
There is no safety information on levels of danger.
1.6.2 Warning
Improper and unsafe use of the analyzer can result in death, serious injury or material damage to the
operator.
• Only health professionals should use this analyzer for clinical diagnostics and under certain
conditions.
• Before using this analyzer, users must check the device and its accessories to ensure its proper
function and safety.
• Reagent should be disposed of per the regulations of the hospital / clinic.
• Do not use this analyzer in an environment where flammable substances are present to prevent
fire and explosion.
• Do not open the analyzer case as it may cause electric shock. Any services or upgrades should
be conducted by authorized personnel only.
• The packaging material should be disposed of per local regulations and should be kept in a place
where no children can reach.
• Connect this analyzer only to a power outlet with protective grounding. Do not use any power
outlets without grounding.
1.6.3 Caution
Failure to comply with the following and unsafe use of the analyzer can lead to injury, malfunction of the
instrument, damage, or other material damages.
• Use of calibration solutions or electrodes not manufactured by High Technology, Inc. could void
the warranty.
• The waste container holds human body fluids which may be potentially infectious; handle with
appropriate care to avoid skin contact or ingestion.
• Dispose the analyzer and consumables per local regulations. If there is any doubt, contact
manufacturer.
• Magnetic fields can affect the performance of the analyzer. Any equipment in use near the
analyzer must comply with EMC Standards. Cell phones, X-ray, or MRI devices are possible
interference sources, as they emit high electro-magnetic radiation.
• Before connecting the analyzer to power, make sure the voltage and frequency are in accordance
with the requirements on the label or in the manual.
• Install and transport the analyzer properly to prevent any damages by dropping, impact, shakes,
or other mechanical forces.
1.6.4 Attention
Some operations require extra attention. The following recommendations offer options to improve the
methods of operation of the instrument.
2 DESCRIPTION OF ANALYZER
2.1 External and Internal Features
The following describes the external and internal components of the analyzer.
1 2 1 LCD Screen.
2 Internal Printer
3 Sample Probe
3
5 4 Wash Cup
4
5 Reagent Pack
6 5 Main Door
2 Grounding Connection
3 AC Power Connection
1 2
1 Detector
3 Peristaltic Pump
4 Distribution Valve
K Potassium Electrode
Na Sodium Electrode
Ca Calcium Electrode
pH pH Electrode
Cl Chloride Electrode
The analyzer should be placed on a stable surface at an appropriate height for its use. Secure the
analyzer so it will not fall and/or cause injury.
There should be power with grounding within reach of the analyzers power cable.
• The instrument must have good grounding. Only connect power to an outlet that
has sufficient grounding.
If installed or not used per the instruction of this manual, it may cause or receive interference from nearby
equipment.
If the analyzer receives unstable readings, interruption of operation without obvious reasons, or if
malfunctions occur, it may be due to such interference. Keep the analyzer away from high efficiency
equipment such as an X-ray device to avoid interference.
3 INSTALLATION
3.1 Equipment Unpacking
Open the package per the marks on the box. Carefully remove the analyzer and its accessories (Please Note:
Sample Tray is optional and may not be included depending on model) per the pictures listed. Save all
original packaging for future storage and/or shipping needs.
• PLEASE NOTE: Sample tray is an optional component and may not be included
depending on model
• Electrode serial number is the proof of warranty; please keep electrode and
packaging as well.
3.4 Tubing
Pump Tubing will need to be secured for the pump to operate. Follow the instructions below to secure the
tubing and the peristaltic pump.
3.5 Electrodes
3.5.1 Filling or Refilling Electrode Internal Solution
Rotate to open the head of electrode. Use clean syringe aspirate Electrode Internal Solution, then inject into
electrode chamber. The ISE Electrode Internal Solution should be filled up to about 2/3 - 3/4 of the electrode
chamber. Follow the steps listed below to properly fill or refill the solution in each electrode.
• The length of syringe probe should be 8-10mm. The probe should not enter the
electrode to avoid damaging the electrode membrane
• The Ag/AgCl inner electrode must always be immersed in reference solution.
• Add Electrode Internal Solution to reference electrode from the orifice on top
right corner of the electrode.
• Flick the electrode bottom repeatedly. Make sure no air bubbles are present next
to the electrode membrane.
• Do not use expired, mildewed or damaged reagent.
• Immediately seal the reagent bottle after taking out the required reagent.
• Do not leave the reagent bottle opened for long periods of time.
Once all electrodes are in place make sure that, they are all flush with each other.
3.9 Activation
3.9.1 Priming Reagent Pack
After checking the analyzers fluid lines, the screen will display:
If QC ranges are unacceptable or if Quality Control fails, you will need to run QC Calibration.
• Follow the instructions in Chapter 4.1.2.3 to run QC Calibration.
4 OPERATION
This chapter will cover all the functionality of the analyzers software. All software menus and sub-menus will
be explained.
4.1.2.1 Test
Users should perform a “Quality Control Test” every day. From the QC Analysis Menu select “TEST”, the
screen will display:
• Each QC test needs to be run individually. Run QC1, once complete repeat for
both QC2 and QC3
4.1.2.2 Factor
The Factor screen will display the Slope and Intercept factors from the last QC Calibration that was
performed. Users can manually enter these factors by entering the Slope and Intercept values calculated
by using the formula and instructions from section 4.1.2.4. From the QC Analysis Menu select FACTOR”,
the screen will display the current factors:
4.1.2.3 QC Calibration
The QC Calibration screen is used to calculate new factors for Slope and Intercept for each parameter. The
user will enter the target (or mean) values for the Low and High controls on the Insert Sheet from the Quality
Control material. Select “QC CALIBRATION”, the screen will display:
• The analyzer will not automatically calibrate and calculate factors for Li or Mg.
Calibration for Li and Mg need to be run manually. Use the following steps to run calibration to get the Slope
and Intercept values for.
Using the control insert sheet, your calculated values and the formulas below to calculate the Slope and
Intercept values for Li and Mg. The following formulas describe how the Slope ( m) and the Intercept (b)
are calculated:
• From the Main Menu select “QC”, and then select “FACTOR”.
• Enter the calculated values for Slope and Intercept into the corresponding boxes and press “OK”.
4.1.2.5 Statistics
The QC Statistics screen will allow the user to review saved QC data for the previous 30 days. QC Statistics
data consists of the Mean, Standard Deviation (SD), and Coefficient Variation (CV) from the last saved test
of the day, for the previous 31 days for all tested parameters. Users can review and print the selected data
for all 3 levels of QC. To review or print results select “STATISTICS”, the screen will display:
4.1.3 Samples
This section will describe how to process samples, the analyzer will test the following sample types, Serum
and Pre-diluted urine. Sample ID’s will consist of 4 digits (e.g. 0001). Sample ID’s will count in succession if
continuous samples are run. Conversely a unique ID can be entered for each sample. The analyzer can
store data for up to 7100 samples and if connected to and LIS or computer will automatically send data after
sample analysis is complete. From the Main Menu select “SAMPLES” to start sample analysis.
4.1.3.1 Manual
Follow the instructions below to perform sample analysis.
4.1.4 Settings
The settings menu will allow the user to adjust the analyzers settings for Date & Time, Parameters, Print
Setup, Normal Ranges, Test Speed and Calibration Frequency. From the Main Menu select “SETTINGS” to
set make adjustment to the system settings
4.1.4.2 Parameters
Parameters will allow the user to turn on and off selected parameters for testing. Turning off a parameter
means that it will not be tested or calibrated. For system models that do not support a specific parameter
these will be off by default. From the System Settings Menu select “PARAMETERS”, the screen will display:
4.1.5 Maintenance
The instrument has four system maintenance programs, Deproteinize, Activate Electrode, Flushing and
Replace Reagent Pack. This section will describe the functionality of each menu. From the Main Menu select
“MAINTENANCE”, the screen will display:
4.1.5.1 Deproteinize
Deproteinizing is used to help keep the tubing in the analyzer clean and free of debris. The software is
preprogramed to request deproteinization every 100 samples. It is recommended to run deproteinization
2 times per week. Before running Deproteinize you must prepare the solution:
• Take caution not to have the probe touch the bottom of the mixed solution. The
probe touching the bottom of the cup can cause an aspiration error.
2:47
4.1.5.2 Activate
Activation of electrode is required for when new electrodes have been installed, or for when there is a need
if ISE calibration fails. From the Maintenance Menu select “ACTIVATE”, the screen will display:
4.1.5.3 Flushing
The Flushing function is used for flushing the tubing in the analyzer of liquid. The analyzer will use
Calibration Solution A from the pack to perform flushing. From the Maintenance Menu select “FLUSHING”,
the screen will display:
4.1.6 Other
This menu will allow users to Review and Upload Data, check Calibration Data used Advanced Setup
features on the instrument. From the Maintenance Menu select “OTHERS” the screen will display:
QC Data Review
QC Data review is for the use to review QC data from previous saved QC results. Users can review
date from a specific date, review on screen and print selected data.
• Detector Enabled
The Detectors functionality is used for detecting liquid in the tubing it serves as the key component for
sensing samples and calibration solution from the reagent pack. The default setting is on. If turned off
the analyzer can still function when the fluidic line adjustment is performed to account for enough liquid
reaching though the electrode housing
• Activation (minutes)
Activation time setting is for the number of minutes that the Activate function runs when new electrodes
are installed or for when the Activate function is performed. The default setting is 30 minutes, but users
can adjust the time from 30 to 120 minutes.
• Deprotein Counter
The Deprotein counter setting is used for alerting the user when deproteinization needs to be
performed. Deproteinization should be performed weekly and when requested. The default setting is
every 100 samples and can be adjusted by the user from 50 to 300 samples
• Sample Type
The objective or sample type setting is used for changing the type of sample that can be tested by the
analyzer. Currently the supported sample types are Serum and Urine (diluted). The default objective is
Serum and users can change this setting by using this function
• Sample Volume
The Sample volume setting is used to change the volume amount that is aspirated when samples are
requested. Users can change this setting to the following volumes, 65μ, 100 μ, and 140μ. The default
setting is 140μ
• Pumping A 10ts • Use to pump (10 times per) Solution A from Reagent Pack into analyzer tubing
• Pumping B 10ts • Use to pump (10 times per) Solution B from Reagent Pack into analyzer tubing
• Detector Adjust
The Detector Adjust function is used to calibrate the detector. Performing this test will test the mV
values of the detector with and without liquid. The detector will be calibrated based on the values of the
mV readings. This function should be performed by trained Service Technicians only.
• Fluid Adjust
The Fluid Adjust function is for adjusting the liquid position in the analyzer tubing. Performing this test
will adjust the position of the liquid while sample analysis is being performed. This function should be
performed by trained Service Technicians only
• Pumping A 10ts
The Pumping A 10ts function is for pumping Solution A from the reagent pack. Performing this function
will allow the user to manually prime reagents for troubleshooting purposes.
• Pumping B 10ts
The Pumping B 10ts function is for pumping Solution B from the reagent pack. Performing this function
will allow the user to manually prime reagents for troubleshooting purposes
• Electrode Test
The Electrode Test function is for testing the stability of the electrodes. Performing this function will test
the mV values for all electrode over time. This function should be performed by trained Service
Technicians only
5 MAINTENANCE
5.1 Inspection
Before using the system, perform the following inspection:
If any signs of damage is found and evidenced, the analyzer should not be used for any measurements.
Contact the service department or a local representative. Inspect the analyzer’s functions and safety every 6-
12 months or after service by authorized personnel. All safety and maintenance inspections should be done
by qualified maintenance personnel. Contact High Technology, Inc. for information on local
maintenance/service centers.
6 TROUBLESHOOTING
Please refer to the following troubleshooting techniques if you come across any problems while operating the
analyzer. This topic covers the most common problems that can occur. Please refer to the service manual if
the topics are not covered.
• The pump will run one more cycle automatically if it continues to fail, please
• Detector Fails
refer to the Service Manual
• Sample Probe • Turn off the analyzer, grab the sample probe (line up the probe with the sample
Fails tray) and gently slide it up to down.
• Turn off the analyzer, check the connection to make sure it is secure. Re-
connect the sample tray if necessary.
• Sample Tray Fails
• Use extreme caution when connecting sample tray, the probe sample tube is
extremely close to the connector and could be damaged if care is not taken.
• Test Results drift • Use Pipe Adjust in the Advanced Settings menu (PW 123678) to adjust the
out of Range liquid to the correct position (about 1 inch past the electrode housing)
• Check for blockage from salt crystallization in the tubing, most commonly
around distribution valve. Clean/flush with syringe using DI water.
• Liquid Path Error
• Check electrodes for missing/damaged O-Rings, replace if necessary
(A) or (B)
• Use pumping A/B 10ts in the advanced settings menu (PW 123678) to check is
fluid is flowing properly from Reagent Pack
7.1 Installation
Push the plug on the side of sample tray into the socket on the left of analyzer
1 2
• •1 Sample Tray Connector (male)
• •
• Use extreme caution when connecting sample tray, the probe sample tube is
extremely close to the connector and could be damaged if care is not taken.
• •
2:47
There are three kinds of protocol formats for the E-Lyte Plus analyzer, Real-Time Testing Data, Batch Testing
Data, and QC Testing Data.
12 digits should be kept between items no matter if there are one, two or three numbers in each item’s value
Sample*Report** Sample*Report**
Date**00-00-00** Date**00-00-00**
Time**00:00:00** Time**00:00:00**
PAT*******000** PAT*******000**
Reel*******000** Reel*******000**
Cup*******000** Cup*******000**
Barcode***????????????????????** Barcode***????????????????????**
K******0.00** K******0.00**
Na******0.00** Na******0.00**
Cl******0.00** Cl******0.00**
iCa******0.00** iCa******0.00**
TCa******0.00** TCa******0.00**
nCa******0.00** nCa******0.00**
pH******0.00** pH******0.00**
Mg******0.00** Mg******0.00**
Li******0.00** Li******0.00**
End End
QC Data
QC*Report*
Date:YY-MM-DD*
QC?*
Lot*No:??????* ? stands for 6-digit number
K*:????* ? stands for numbers or decimal point without limited digits
Na:????*
Cl:????*
iCa:????*
Li:????* Or Mg:????* Only one choice between them
End
The relationship between nCa, TCa and complexed calcium is described below:
TCa = nCa + Complexed calcium
(When pH = 7.40, TCa ≈ 2·nCa)
The electrolyte analyzer measures ion concentration by analyzing a serum sample. After serum is separated,
due to centrifugal role and time extend, contact with ambient air will cause a loss of CO2 in the serum sample
and the subsequent rise in pH will cause a reduction in ionized calcium. In this case, the iCa value in the
measuring result is not clinically significant.
The pH value displayed on screen is on the high side because of CO2 emission. Based on this actual pH
value, the instrument calculates the nCa value (equal to ionized calcium in human body) according to a specific
formula (nCa = iCa + (pH-7.4)/1.89). The nCa value is the main basis for clinical diagnosis when testing serum.
During sample analyzing, the system displays only nCa on screen, but it will print out iCa, nCa and TCa in the
measuring results, therein nCa is used for clinic diagnosis, iCa and TCa are only for reference.
Because CO2 emission is irregular, the pH value is only for reference in serum analyzing, it mainly works to
correct calcium ion.
When testing using a QC solution or other samples (with the exception of serum) which have no CO2 emission,
the iCa and displayed pH value on screen should be the basis for clinical diagnostics