Professional Documents
Culture Documents
Equ211-02 RXL Calibarion SOP
Equ211-02 RXL Calibarion SOP
Approved
By
SOP
Annual
Review
Distributed
Copies to
Associated Forms:
I acknowledge that I have read, understand and agree to follow this SOP.
Name (print) Signature Date
PURPOSE
This standard operating procedure (SOP) covers the Calibration/Verification for the RxL
Max Dimension system.
SCOPE
This procedure applies to the Calibration/Verification on the RxL Dimension. It is vital to perform
proper calibration on the analyzer as per manufacturer instructions so as to ensure accurate results
are been obtained. You should calibrate also if any of the following occur.
Calibration frequency:
Most methods required to be calibrated within 90 day for a complete calibration frequency for all
methods please refer to Appendix A
1.2 Calibration of the Urine Drugs of Abuse Methods in the Qualitative Mode
Step Action
1. From Operating Menu
2. Press F1: Enter data and enter
patient name: calibrator name
mode: sample cup
priority: QC
sample no: lot #
fluid: 3rd QC (serum QC3)
tests: request DAU method in triplicate
3. Place sample cup with the calibrator in the indicated sample wheel position
4. Press F2: Process single
5. Process negative and positive controls
IF……… THEN……
Tolerance Limits If result fall within assay – Proceed with
specific specification. analysis
And QC values are within
acceptable limits
If result falls out-side assay Troubleshoot the
specific specification, and assay and/or
QC values are out of instrument and
Acceptable limits. repeat calibration
1.3 Documentation
After calibration all calibration and QC run should be attached to calibration log
Sheet and must be documented on calibration log book.
3.0 REFERENCE
4.0 Appendix