A-Dec DV Dry Vacuum Piping Layout: Pre-Installation Guide

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 12

Pre-Installation Guide

A-dec DV Dry Vacuum Piping Layout


Contents
Glossary of Terms..................................................................................................1
General Guidelines............................................................................................... 2 1" Branch
PVC Fitting Examples........................................................................................... 2 All Locations
Main Trunk Line Sizing......................................................................................... 3
Main Trunk Line Sizing Example......................................................................... 4 1.5"

Below Grade Details for Branch Connection to Main Trunk............................. 5


Overhead Details for Branch Connection to Main Trunk................................... 6
Nitrous Oxide (N2O) Scavenging...........................................................................7

Note

1.5"
This document contains recommendations for facility vacuum piping
layout. This guide is to serve as a reference only. A-dec makes no claim
to knowing local codes, local installation practices or specific facility
construction. A-dec does not guarantee performance.

2"

1.5"
2"
2" 1.5"
2.5"

1" Branch
All Locations

86.0917.00 Rev A
Copyright Regulatory Information and Warranty
© 2021 A-dec, Inc. All rights reserved. For required regulatory information and the A-dec warranty, see
the Regulatory Information, Specifications, and Warranty document
A-dec, Inc. makes no warranty of any kind with regard to this material,
(p/n 86.0221.00) available in the Resource Center at www.a-dec.com.
including, but not limited to, the implied warranties of merchantability
and fitness for a particular purpose. A-dec, Inc. shall not be held liable
for any errors contained herein or any consequential or other damages Product Service
concerning the furnishing, performance or use of this material. The
Product service is available through your local authorized A-dec dealer.
information in this document is subject to change without notice. If you
For service information, or to locate an authorized dealer, contact A-dec
find any problems in the documentation, please report them to us in
at 1.800.547.1883 in the USA and Canada or 1.503.538.7478 worldwide,
writing. A-dec, Inc. does not warrant that this document is error-free.
or visit www.a-dec.com.
No part of this document may be copied, reproduced, altered, or
transmitted in any form or by any means, electronic or mechanical,
including photocopying, recording, or by any information storage and
Product Models and Versions Covered in This Document
retrieval system, without prior written permission from A-dec, Inc.
Model Versions Description
DV5, DV7, n/a Dental Vacuum
Trademarks and Additional Intellectual Property Rights DV10, DV12
A-dec, the A-dec logo, A-dec Inspire, Cascade, Century Plus, Continental,
Decade, ICX, ICV, Performer, Preference, Preference Collection,
Preference ICC, Radius, and reliablecreativesolutions are trademarks of
A-dec, Inc. and are registered in the United States and other countries.
A-dec 500, A-dec 400, A-dec 300, A-dec 200, and EasyFlex are also
trademarks of A-dec, Inc. None of the trademarks or trade names in this
document may be reproduced, copied, or manipulated in any manner
without the express, written approval of the trademark owner.
Certain touchpad symbols and icons are proprietary to A-dec, Inc. Any
use of these symbols or icons, in whole or in part, without the express
written consent of A-dec, Inc., is strictly prohibited.
A-dec DV Dry Vacuum Piping Layout Pre-Installation Guide Glossary of Terms

Glossary of Terms collect and remove excess gases to prevent them from being vented back into the
operating room. Flow requirement is approximately 1.5 CFM or 1/2 user.
Air Velocity - Velocity refers to how fast the air is moving in a pipe, in distance, per
unit of time. Dry Vacuum systems use air velocity to move liquids and solids through Treatment Rooms- A highly specialized space configured to deliver dental treatment
the piping system. to patients while supporting all associated tasks performed by the dentist and
auxiliaries.
Backflow - An unwanted flow of suction liquid in the reverse direction. Prevented by
proper piping design. Overhead - Vacuum piping system where the main trunk is suspended from the
ceiling and runs above treatment rooms.
Below Grade - Piping that is below ground level either buried in a slab or in a
basement. PVC Pipe - Abbreviation for polyvinyl chloride. White plastic pipe commonly used for
plumbing and drainage.
CFM - Cubic Feet per Minute (cu ft/min). Unit that represents the flow of air.
Slope - Has the same meaning as pitch. It is generally accepted that 1/4" per 10' of
Cleanout - A plumbing cleanout provides a convenient place to access a building’s pipe run is the minimum for proper slope on a Dental Vacuum Piping System.
vacuum piping system to clear clogs and debris.
SE (Saliva Ejector) - A narrow tubular device providing suction to draw saliva, blood,
Double 45° - Using (2) 45° fittings to create a “sweep” 90°. and debris from the mouth of a dental patient in order to maintain a clear operating
Dry Vacuum - A type of vacuum generating equipment that does not use water field. Flow requirement is approximately 1/2 User.
during operation. Sweep - DWV elbows are usually long-radius (“sweep”) types. To reduce flow
DWV - Pipe fittings designed specifically for Drain, Waste, and Vent applications. resistance and solid deposits when the direction of flow is changed, they use a
These are the only type of fittings accepted for use in a vacuum system. shallow curve with a large radius of curvature.

HVE - A High Volume Evacuator is a suction device that draws a large volume of air Trunk - In a complete Dental Vacuum Piping System, the main supply lines, the
over a period of time. Flow requirement is approximately 1 user. “trunks,” provide suction to the general area where it will be used. Smaller-diameter
tubing, the “branch” lines, provide connections to the point of use.
InHG / "HG / Inches of Mercury - Inch of mercury is a unit of measurement for
vacuum. It is widely used in the dental industry. It is the vacuum level required Wet Vacuum - Wet dental vacuum systems use water to create vacuum pressure.
to vertically lift a column of mercury 1 inch (25.4 mm) in height at the standard They require a lot of water to operate.
acceleration of gravity.
Inverted P-Trap - Branch lines connecting to overhead trunk tie into the top of the
trunk to prevent backflow.
Junction Box - Connection box for all utilities to dental delivery unit. Generally floor
or wall mounted. Details provided by equipment provider.
NFPA99 - The National Fire Protection Association (NFPA) establishes criteria
for levels of health care services or systems based on risk to the patients, staff,
or visitors in health care facilities to minimize the hazards of fire, explosion, and
electricity. Dental Vacuum piping is covered under NFPA99 Level 3.
Nitrous (N2O) Scavenging - A scavenging system, simply defined, is a means to

86.0917.00 Rev A 1
A-dec DV Dry Vacuum Piping Layout Pre-Installation Guide General Guidelines

General Guidelines PVC Fitting Examples


• Vacuum piping to be PVC or Type M copper per local code. Drain Waste Vent Wye
• Main trunk lines and headers to have a minimum slope of 1/4" per
10'.
• For main trunk lines longer than 50', increase in pipe size. Drain Waste Vent Offset Wye
• No split wye on main trunk lines, use only a branch wye.
• Branch lines should be of equal length, where possible.
• Use DWV piping and fittings.
Drain Waste Vent Sweep 90° Elbow
• If DWV fittings are not available, use double 45° elbow.
• Do NOT use tee, 4-way cross, or short elbow connectors.
• Do NOT use standard PVC 90° elbows.
Standard PVC Double 45° Elbows to Make 90°

Cleaning Vacuum Lines


Use only dental-specific vacuum line cleaner made for dry vacuum
systems. Use only a non-foaming cleaner. No other cleaning agents,
such as ultrasonic fluid, should be used in your vacuum lines.
Do Not Use

NOTE  Failure to comply with vacuum line cleaning


guidance can decrease vacuum performance, including a Tee
complete loss of vacuum. Problems arising from the use of
proper cleaning agents can affect the warranty.

4-Way Cross

Short Elbow

86.0917.00 Rev A 2
A-dec DV Dry Vacuum Piping Layout Pre-Installation Guide Main Trunk Line Sizing

Main Trunk Line Sizing


IT IS THE ARCHITECT/CONTRACTOR'S RESPONSIBILITY TO VERIFY Main Trunk Sizing
THAT ALL CODE RESTRICTIONS & CLEARANCES ARE MET. NEW WALLS TO BE CONSTRUCTED
Number of Treatment Minimum Size of Line
Rooms
4 1.5"
THIS PLAN IS FOR THE SOLE PURPOSE OF SPECIFICATION AND
INSTALLATION OF THE DENTAL EQUIPMENT SUPPLIED BY HENRY

8 2"
SCHEIN DENTAL. THE SPACE PLANNING AND CONCEPTUAL

1" Branch DESIGN HAS BEEN PROVIDED BY OTHERS AND THE IDEAS
HEREIN ARE NOT NECESSARILY RECOMMENDED BY THE HENRY
SCHEIN INTEGRATED DESIGN STUDIO .
TM

All Locations 12 2.5"

LOCKERS
CLOSET/ FUTURE
TRE. ROOM

16 3"
STAFF

For offices designed with more than 16


FUTURE WALL LOUNGE
REF.

treatments rooms, with extra large length runs


TRE. TRE.
ROOM 3 ROOM 1

I.T.
2" WASHER/
DRYER
or multiple branch trunk lines, contact A-dec
CLOSET
Customer Service for piping size and layout
CAD/CAM MILL
LAB
advisement.
HYG.
ROOM 4

HENRY SCHEIN EQUIPMENT REP: ADAM BRIDGEFORD


DR. MONTY BULZOMI
TRE.
ROOM 4 ADA
A

Calculating Sizing
STERILIZATION
GLASS WALL • Sizing applies to Overhead and Below Grade
layouts.
1.5"

MECHANICAL
HYG.
ROOM 3

REST
RESTROOM

MECHANICAL
• Using table above, count the number of
EXISTING
WOMEN'S
OME
G

treatment rooms starting with the furthest from


RESTROOM
RES

Nitrous Scavenging to be
Separate 3/4" Branch into
the Mechanical Room.
KID'S

Main Trunk at Each Location AREA

HYG.
CONSULT. BUSINESS
• Every 4 Chairs, step up in size.
OFFICE

1" Branch ROOM 2

All Locations • Route Main Trunk for equal branch lengths (as
much as possible, see #5 Branch in example.
Extending slightly will not significantly impact
EXISTING
G

performance).
MEN
MEN'S
CHECK-OUTT
Pipe Legend: RESTROOM
RES

3”
• Reduce to 1.5" in Mechanical Room.
1.5"

HYG. WALL MTD. 2.5”


ROOM 1 COMPUTER
USABLE SQ FT:
2”
3363
IMAGING
PRIVATE
OFFICE
1.5” • Dual Vacuum: 2" connection
SCALE:
in mechanical
1” Branch
room.
ALCOVE

3/4” Branch for N20 NTS


DATE:
• Optional: Add cleanout access at end of header.
02/13/2019
Plumber to choose access method.
DESIGNED BY:
DMS/TJH
PROJECT #:
P:\_ALL CENTERS\Fargo\Bulzomi, Monty\BULZOMI-1H c 2019 by HENRY SCHEIN, INC. 19-0252

86.0917.00 Rev A 3
A-dec DV Dry Vacuum Piping Layout Pre-Installation Guide Main Trunk Line Sizing Example

Main Trunk Line Sizing Example Main Trunk Sizing


Number of Treatment Minimum Size of Line
Rooms
4 1.5"
1" Branch
All Locations 8 2"
12 2.5"
1.5"
16 3"
For offices designed with more than 16
Pipe Legend: treatments rooms, with extra large length runs
3” or multiple branch trunk lines, contact A-dec
2.5” Customer Service for piping size and layout
2”
1.5” advisement.
1” Branch
3/4” Branch for N20

Multi-Branch
1.5"

• Join Branch Trunk with Main Trunk using only


DWV fittings.
• Using the table above, count the number of
treatment rooms starting with the furthest from
2"

the mechanical room. Each Branch has its own


1.5"

count, adding to the total.


2"

• Every 4 chairs, step up in size.


• Route Main Trunk and Branch Trunks for
equal branch lengths (as much as possible;
2" extending slightly will not significantly impact
1.5"
performance).
2.5"
• Single Vacuum: Reduce to 1.5" in mechanical
room.
1" Branch
All Locations • Dual Vacuum: 2" connection in mechanical
room.
• Optional: Add cleanout access at the end of the
header. Plumber to choose access method.

86.0917.00 Rev A 4
A-dec DV Dry Vacuum Piping Layout Pre-Installation Guide Below Grade Details for Branch Connection to Main Trunk

Below Grade Details for Branch Connection to Main Trunk


Rules for Below Grade Piping
• Vertical connection to Vacuum system to be at
45° angle.

B • Complete piping system to be sloped


1/4" per 10' towards mechanical room.

C
r
me l)
u mp usto tica
P C er
u um -dec 8' V
c A r
Va lt ove
Complete piping system to slope To onsu e if
1/4" per 10' towards the (C rvic
Se
mechanical room.

45°Vertical Rise from Main Trunk


to Vacuum System.
Detail B
Scale 1:5 45° Vertical Rise
from Main Trunk to
Vacuum System

pe
Pi

pe
5"

Pnik
1.

5Tr"u
1n.
e

ai
em ro ip
M
st e f " P

R pe
Sy Ris 1.5
m
1.5" Pipe

P i
1.5" Pipe

Sy Tru ise
Detail C

o
cu M tic "

em t
5
.

Va om Ver 1.

st nk
Detail A Scale 1:5

um in l a
um cal
Scale 1:5

cu rti
2" or2"Greater
Pipe
Pipe

a
Va Ve

fr °
to 5°

45
4
Reducing Bushing

86.0917.00 Rev A 5
A-dec DV Dry Vacuum Piping Layout Pre-Installation Guide Overhead Details for Branch Connection to Main Trunk

Overhead Details for Branch Connection to Main Trunk

A
Sloped 1/4"/10'
DWV Fittings Towards Vacuum Pump

Pitch towards
the vacuum. 1/2" PVC
10' Max

Floor B

DETAIL A DETAIL B Rules for Overhead Piping


• Overhead Main Trunk height must be kept
as low as possible.
• Vertical Branch must be 1/2" PVC.
• Branch must connect into Main Trunk per
All turns to be DETAIL A. Use DWV fittings and do not tee.
two 45° elbows.
• Operators must be trained to allow air to
move liquid before closing HVE/SE.

Use overhead riser to


connect to header.
(Inverse P-trap
to prevent backflow.)

86.0917.00 Rev A 6
A-dec DV Dry Vacuum Piping Layout Pre-Installation Guide Nitrous Oxide (N2O) Scavenging

Nitrous Oxide (N2O) Scavenging

Rules for Nitrous Oxide Scavenging


• Must have a separate branch dedicated to N2O Scavenging
• Applies to Below Grade or Overhead piping

Below Grade Piping Overhead Piping


General Dental Suction Branch

Main Trunk Line

N2O Connection Branch

General Dental Suction Branch

N2O Connection Branch

Main Trunk Line

86.0917.00 Rev A 7
A-dec DV Dry Vacuum Piping Layout Pre-Installation Guide Nitrous Oxide (N2O) Scavenging

86.0917.00 Rev A 8
A-dec DV Dry Vacuum Piping Layout Pre-Installation Guide Nitrous Oxide (N2O) Scavenging

86.0917.00 Rev A 9
A-dec Headquarters A-dec St. Louis
1601 Manufacturers Drive
2601 Crestview Drive
Newberg, Oregon 97132 Fenton, MO 63026 USA ÍvÈ.Ç)1È.00@Î
United States
Tel: 1.800.547.1883 within USA/CAN 86.0917.00 Rev A
Tel: +1.503.538.7478 outside USA/CAN Date of Issue: 2021-02-04
Fax: 1.503.538.0276 Copyright 2021 A-dec, Inc.
www.a-dec.com All rights reserved.