TRACEABLE
V–range:
U In USP <857> (42th edition, update December 2019),
Potassium dichromate solutions are used over the two possible procedures are described for checking
intended operational range up to 3 Abs. limit of stray light.
Vis–range:
Neutral density glass filters are used over the intended Procedure A: the already given and discribed method.
operational range up to 3 Abs.
Procedure B: The analysts can measure the stray light
New supplement in USP <857> filter with a path length of 10 mm against the reference
UV range: 0 – 1 Abs: filter filled with pure water (10 mm path length), with
Potassium dichromate solutions with concentrations the exception of acetone, which is measured against
from 20 – 60 mg/l, or Nicotinic acid solutions with con- air. The measured absorbance value per filter should
centrations from 6 – 24 mg/l may be used. be greater than 2.0 at the recommended wavelength
UV range: 1 – 3 Abs: listed in the table:
Potassium dichromate solutions with concentrations
from 80 – 200 mg/l may be used. Filter/Solution Spectral Recom-
Range (nm) mended
WL (nm)
3 Control of photometric linearity Potassium chloride (12 g/l) 190 – 210 198
Sodium iodide (10 g/l) 210 – 270 220
New supplement in USP <857> Potassium iodide (10 g/l) 210 – 270 220
“Verification of photometric response (linearity) is Acetone (pure) 250 – 330 300
required and it should be evaluated using a standard Sodium nitrite (50 g/l) 300 – 400 340
type appropriate for the wavelength(s) required, where
at least three different absorbance levels appropriate 5 Resolution:
to and spanning the required operational range are
measured” to date:
UV–range: A liquid filter of Toluene in Hexane and a Hexane
Nicotinic acid solutions 6 – 60 mg/l or Potassium reference filter are used to check the resolution of the
dichromate solutions 20 – 200 mg/l may be used. spectrophotometer.
Vis–range:
Neutral Density Glass Filters may be used. No Changes!
DOING SO ENSURES CONFORMITY WITH USP <857>
UPDATE FROM DECEMBER 2019!
Supplement your existing USP reference materials:
Please check to ensure that your reference materials comply with the updated regulations of USP <857>.
1 Control
of wavelength accuracy
Filters used today
667005USP UV5: Holmium in perchloric acid 241, 250, 278, 287, 333, 345,
361, 385, 416, 452, 468, 485,
536, 640
Supplement if necessary
667035 UV35: Rare Earth (Cerium) in perchloric acid 201, 211, 222, 239, 252
667025EPUSP UV25: Didymium in perchloric acid 512, 732, 740, 794, 801, 864
5 Control of resolution
Filters used today
667200 Filter set with Toluene in Hexane and Hexane reference filter Scan from 265 to 270
slit width: 0.5, 1.0, 1.5, 2.0, 3.0
No supplement necessary
BUY A NEW BASIS FILTER SET COMPLIANT
WITH NEW UPDATED USP <857>
667857V3 New
Basis Filter set according to updated USP <857>
(for wavelength range 240 – 640 nm and absorbance range 0.2 – 2.0)
!
Contains Set of Potassium dichromate filters includes: 235, 257, 313, 350
everything UV20, 20 mg/l in perchloric acid (0.1 – 0.3 Abs)
you need to UV60, 60 mg/l in perchloric acid (0.3 – 0.9 Abs)
conform with UV0140, 140 mg/l in perchloric acid (0.7 – 2.0 Abs)
USP <857>! UV14: Perchloric acid (reference filter)
(Minimum
configuration)
Control of absorbance accuracy and photometric linearity in the Vis range
Set of Neutral Density Glass filters includes: 440, 465, 546.1, 590, 635
F2 (0.25 Abs)
F4 (1.0 Abs)
F203 (2.0 Abs)
Control limit of stray light
UV19, Acetone* (pure), ≥ 2 Abs at 300 nm Spectral range: 250 – 330
Control of resolution
Filter set with Scan from 265 to 270
UV6 Toluene in Hexane, slit width: 0.5, 1.0, 1.5, 2.0, 3.0
UV9 Hexane (reference filter)
*) Measurement against air
TS
EXPER
TIP:
Birgit Kehl,
Compliance Manager
ISO / IEC 17025
Absorbance 667020 UV20, Potassium dichromate 20 mg/l, (0.1 – 0.3 Abs) 235, 257, 313, 350
accuracy
667040 UV40, Potassium dichromate 40 mg/l, (0.2 – 0.6 Abs) 235, 257, 313, 350
UV range
667060 UV60, Potassium dichromate 60 mg/l, (0.3 – 0.9 Abs) 235, 257, 313, 350
Potassium
667080 UV80, Potassium dichromate 80 mg/l, (0.4 – 1.2 Abs) 235, 257, 313, 350
Dichromate
6670100 UV0100, Potassium dichromate 100 mg/l, (0.5 – 1.45 Abs) 235, 257, 313, 350
6670120 UV0120, Potassium dichromate 120 mg/l, (0.6 – 1.7 Abs) 235, 257, 313, 350
6670140 UV0140, Potassium dichromate 140 mg/l, (0.7 – 2.0 Abs) 235, 257, 313, 350
6670160 UV0160, Potassium dichromate 160 mg/l, (0.8 – 2.3 Abs) 235, 257, 313, 350
6670180 UV0180, Potassium dichromate 180 mg/l, (0.9 – 2.6 Abs) 235, 257, 313, 350
6670200 UV0200, Potassium dichromate 200 mg/l, (1.0 – 3.0 Abs) 235, 257, 313, 350
667014 UV14, Perchloric acid (reference filter) 235, 257, 313, 350
667307 Linearity set UV307, includes Potassium dichromate filters with 20, 40, 235, 257, 313, 350
60, 80, 100 mg/l and reference filter
Absorbance 667506 UV506, Nicotinic acid (Niacin) 6 mg/l, 0.25 Abs) 213, 261
accuracy
667512 UV512, Nicotinic acid (Niacin) 12 mg/l, 0.5 Abs) 213, 261
UV range
667518 UV518, Nicotinic acid (Niacin) 18 mg/l, 0.75 Abs) 213, 261
Niacin
667524 UV524, Nicotinic acid (Niacin) 24 mg/l, 1.0 Abs) 213, 261
667536 UV536, Nicotinic acid (Niacin) 36 mg/l, 1.5 Abs) 213, 261
667548 UV548, Nicotinic acid (Niacin) 48 mg/l, 2.0 Abs) 213, 261
667599 UV599, Hydrochloric acid (reference filter) 213, 261
667S501 Linearity set UVS501, includes Niacin filters with 6, 12, 24 mg/l and 213, 261
reference filter
667350 Linearity set UV350, includes Niacin filters with 6, 12, 18, 24 mg/l and 213, 261
reference filter
Stray Light Limit of Stray Light according to USP <857> procedure A (Measurement with 10 mm and 5 mm cell)
667100H Set includes 10 and 5 mm cell filled with Potassium Chloride Spectral range: 190 – 210
667101H Set includes 10 and 5 mm cell filled with Sodium Iodide Spectral range: 210 – 270
667119H Set includes 10 and 5 mm cell filled with Acetone Spectral range: 250 – 330
667102H Set includes 10 and 5 mm cell filled with Sodium Nitrite Spectral range: 300 – 400
667105H Set includes 10 and 5 mm cell filled with Potassium Chloride, Sodium Spectral range: 190 – 400
Iodide, Acetone and Sodium Nitrite
Limit of Stray Light according to USP <857> procedure B (Measurement against water as reference)
667001 UV1, Potassium Chloride 12 g/l, ≥ 2 Abs at 198 nm Spectral range: 190 – 210
667010 UV10, Sodium Iodide 10 g/l, ≥ 2 Abs at 220 nm Spectral range: 210 – 270
667023 UV23, Potassium Iodide 10 g/l, ≥ 2 Abs at 220 nm Spectral range: 210 – 270
667019 UV19, Acetone* (pure), ≥ 2 Abs at 300 nm Spectral range: 250 – 330
667011 UV11, Sodium Nitrite 50 g/l, ≥ 2 Abs at 340 nm Spectral range: 300 – 400
667012 UV12, Pure Water (reference filter) 198, 200, 300, 400
Resolution 667006 UV6, Toluene in Hexane slit width: 0.5, 1.0, 1.5, 2.0, 3.0
667009 UV9, Hexane (reference filter)
667200 Filter set with Toluene in Hexane and Hexane reference filter Scan from 265 to 270
slit width: 0.5, 1.0, 1.5, 2.0, 3.0