Professional Documents
Culture Documents
Physics13 117
Physics13 117
For discussion purposes, consider an image display monitor with a 512 3 512 pixel
resolution. The image acquisition matrix in MRI is often lower, and some interpola-
tion is typically performed automatically. A straightforward approach is to place an
empty pixel between pixels (step 1 in Fig. 117.1) and assign the amplitude of this pixel
to be the average value of the adjacent pixels (linear interpolation, step 2 in Fig. 117.1).
Fig. 117.1
An advanced solution is the use of a bicubic spline interpolation. In this case, the value
of not only the adjacent pixel is taken into account, but also that of the next pixel. The
result is smoother, masking the truly measured spatial resolution. For Fourier encoding,
a low spatial resolution ma-
trix corresponds to a small
k space (limited dimension
in the readout and phase
encoding directions). In any k
space matrix, the outer data
points (the high spatial fre-
quency components) contain
the information concerning
the detailed structure in the
image. Consider a homoge-
neous phantom that fills the
entire field of view. There
is no detailed structure, re-
sulting in zero values for
the outer data points in the
k space matrix. Reflecting
upon this situation, doubling
the raw data matrix size and
filling the missing (outer)
data points with a value of
zero will not improve spatial
resolution, but will mimic
in some ways the measure-
ment of a higher resolution
Fig. 117.2 matrix. Zero filling is simply
246