Professional Documents
Culture Documents
Tanaka2012-Técinica RQA
Tanaka2012-Técinica RQA
DOI 10.1007/s12194-012-0196-8
Received: 17 July 2012 / Revised: 11 December 2012 / Accepted: 11 December 2012 / Published online: 29 December 2012
Ó Japanese Society of Radiological Technology and Japan Society of Medical Physics 2012
Abstract Real-time tumor tracking in external radio- lag were 1.1 mm and 10.0 %, respectively, in all combi-
therapy can be achieved by diagnostic (kV) X-ray imaging nations of imaging parameters examined in this study.
with a dynamic flat-panel detector (FPD). It is crucial to Blurred boundaries and changes in pixel value due to
understand the effects of image lag for real-time tumor image lag were estimated under various imaging conditions
tracking. Our purpose in this study was to develop a lag with use of the simulation system. Our system would be
simulation system based on the image lag properties of an helpful for a better understanding of the effects of image
FPD system. Image lag properties were measured on flat- lag in fluoroscopic images.
field images both in direct- and indirect-conversion
dynamic FPDs. A moving target with image lag was sim- Keywords Image lag Simulation Target tracking
ulated based on the lag properties in all combinations of Image-guided radiotherapy Flat-panel detector (FPD)
FPD types, imaging rates, exposure doses, and target
speeds, and then compared with actual moving targets for
investigation of the reproducibility of image lag. Image lag
was simulated successfully and agreed well with the actual 1 Introduction
lag as well as with the predicted effect. In the indirect-
conversion FPD, a higher dose caused greater image lag on Real-time tumor tracking has been facilitated by the use of
images. In contrast, there were no significant differences diagnostic (kV) X-ray imaging with a dynamic flat-panel
among dose levels in a direct-conversion FPD. There were detector (FPD) in external radiotherapy [1–3]. Kilovoltage
no relationships between target speed and amount of image X-ray imaging is expected to solve the problems of low
blurring in either type of FPD. The maximum contour image contrast and poor image quality seen in megavoltage
blurring and the rate of increase in pixel value due to image imaging. However, several factors remain that could
reduce the accuracy of target tracking in external radiation
therapy. In terms of image quality, it is thought that image
R. Tanaka (&) K. Ichikawa K. Matsubara S. Sanada lag, ghosting, image noise, and image resolution affect the
Department of Radiological Technology, School of Health accuracy of tracking. In particular, a great deal of attention
Sciences, College of Medical, Pharmaceutical and Health
Sciences, Kanazawa University, 5-11-80 Kodatsuno, must be paid to image lag—i.e., the carryover of the image
Kanazawa 920-0942, Japan charge generated by previous X-ray exposure into sub-
e-mail: rie44@mhs.mp.kanazawa-u.ac.jp sequent image frames [4, 5]—because it may induce
blurring on the contours of a moving target.
H. Kawashima
Division of Health Sciences, Graduate School of Medical Previous studies indicated that primary a-Se type of
Science, Kanazawa University, 5-11-80 Kodatsuno, FPDs showed a lag in the first frame after an X-ray
Kanazawa 920-0942, Japan exposure of \5 % of the original signal [6–9]. Modifica-
tions in a-Se detector technology appear to have resulted in
H. Kawashima H. Iida
Department of Radiological Technology, Kanazawa University marked decreases in both lag and ghosting effects in more
Hospital, 13-1 Takara-machi, Kanazawa 920-8641, Japan recent systems [10–12]. However, the effect of image lag
274 R. Tanaka et al.
on the accuracy of target tracking has not been fully 0.05, 0.1, 0.2 mR; indirect-conversion type: 0.01, 0.02,
addressed except for one investigation carried out under 0.04 mR). The exposure dose was determined according to
restricted conditions of imaging parameters by use of a the IEC62220-1-3 standard [15], although with differences
direct-conversion FPD [13]. among FPD types due to the difference in imaging mode
In real-time tumor tracking, it is necessary to optimize (direct-conversion type: dynamic imaging mode; indirect-
the imaging parameters in each patient to keep the patient conversion type: fluoroscopic imaging mode). The matrix
dose as low as possible while maintaining tracking size was 1024 9 1024 pixels, the pixel size was 300 9
accuracy. In addition, it is crucial to understand the 300 lm, the field of view (FOV) was 30.4 9 30.4 cm, and
amount of blurring on the contours of a moving target due the grayscale range of the images was 16 bits in the direct-
to image lag in order to set an appropriate margin around conversion FPD. The matrix size was 1024 9 1024 pixels,
a target when planning radiotherapy. A previous study the pixel size was 130 9 130 lm, the FOV was 15.1 9
showed that the amount of image lag could be changed 15.1 cm, and the grayscale range of the images was 16 bits
depending on the exposure dose and the type of FPD [14]. in the indirect-conversion FPD.
If the amount of image blurring on the contours of a Image lag can be defined as the carryover of trapped
moving target could be estimated based on the image lag charges generated by X-ray exposure into subsequent
properties of an FPD system, it would be useful for a frames acquired with no X-ray exposure. The lag in the nth
margin setting and for parameter selection in radiotherapy frame (Ln) was calculated as follows [4]:
planning of real-time target tracking. Thus, we launched a
Sn B
simulation system for understanding the lag effect in Ln ¼ 100; ð1Þ
Sn B
fluoroscopic images. We performed this study to develop
a lag simulation system that provided image blurring for where S0 is the average pixel values measured in the region
all parameter settings, such as the exposure dose (mR), of interest (ROI) in the frame acquired just before X-ray
target speed (mm/s), imaging rate (frame per sec; fps), cutoff, and Sn is that acquired in the nth frame after X-ray
and type of FPD. cutoff. The term B is the pixel value measured in the
background image, which was obtained without any radi-
ation. The details were described elsewhere [14]. Figure 1
2 Materials and methods shows one of the image lag properties which was installed
in our simulation system as initial data. These data were
2.1 Measurement of lag properties also prepared for the other imaging parameters, i.e., three
dose qualities, three dose levels, and three imaging rates.
Lag measurements were performed with a direct- and an
indirect-conversion dynamic FPD, SONIALVISION 2.2 Simulation of a moving target with image lag
Safire2 (Shimadzu, Japan) and AXIOM Luminos dRF
(Siemens, Germany), according to the IEC 62220-1-3 2.2.1 Target settings
standard [15]. Fluoroscopic images were obtained at three
standard X-ray spectra (IEC RQA3, RQA5, RQA7), at Our system was designed to provide two types of target,
three imaging rates (3.0, 7.5, 15 fps), and at three exposure i.e., a digital phantom simulating a round target and a real
dose levels at the detector surface (direct-conversion type: target cut out from a real radiograph, as shown in Fig. 2.
The digital phantom was created based on a Gaussian 2.2.2 Lag simulation
distribution as follows:
1 A moving target in the nth frame, Tlag(n), was simulated by
f ðx; y; nÞ ¼ pffiffiffiffiffiffi exp use of the results of the lag measurement as follows:
2pr
0 qffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi1 X
K
ð0 x\SÞ; ð0 y\SÞ;
ðx n rÞ2 þ ðy n rÞ2 Tlag ðnÞ ¼ TðnÞ þ fTðn kÞ Lk g; ð4Þ
@ A
2r2 k¼1
where k is the frame number going back from the nth frame
ð2Þ
of interest, and Lk is the image lag (%) calculated by
where x and y are the distance from the center of Eq. (1). The computation was repeated, for Lk was greater
the Gaussian distribution, which determine a target size S. than 1.0. Here, T(0) means a target without lag, equivalent
The n and r are the number of frames processed and the to a static target.
movement rate in units of pixels/frame, respectively, which
control the mean displacement of the target, and r is a 2.3 Calculation of predicted value
parameter that controls the edge gradient of the target.
The target moving rates were prepared in three patterns, We used the digital phantom, as shown in Fig. 2a, to cal-
i.e., 10, 20, and 30 mm/s, covering the moving rates of culate predicted values. Image lag was simulated by use of
lung tumors in clinical practice [16]. The target motion was the known image lag properties. The effect of image lag
given a user-specified trajectory, which was digitally was estimated based on the simulation results, at each
superimposed onto a flat-field image. The target in the nth exposure dose, type of FPD, and target speed.
frame T(n) is described as
TðnÞ ¼ P f ðx; y; nÞ; ð3Þ 2.4 System evaluation
where P is the pixel value at the center of the target, for 2.4.1 Evaluation items
which any value could be set according to the average pixel
value of the insert position. The superimposed image was The real target, as shown in Fig. 2b, was used for system
created by addition of T(n) to a plain or original image. We evaluation. The system was validated in terms of the
used the digital phantom to calculate predicted values of reproducibility of image lag, indicating increases in pixel
image lag in this study (Fig. 2a). value and image blurring on the contours of a moving
Meanwhile, our system provided a real target by clip- target. Previous studies indicated that image lag properties
ping out the ROI from a real radiograph. In this study, a are dependent on the exposure dose and FPD type without
simulation lung nodule with a diameter of 1 cm (paraffin relation to the imaging rate and exposure quality. Thus, the
mixture) on a 15-cm-thick acrylic plate was imaged using validation was conducted in terms of FPD type (direct or
the FPD systems under the same conditions as in lag indirect), exposure dose (direct-conversion type: 0.05, 0.1,
measurements (RQA5, 15 fps, SID = 150 cm). The ROI, 0.2 mR; indirect-conversion type: 0.01, 0.02, 0.04 mR),
40 mm on a side, was located on the nodule area and clip and target speed (10, 20, and 30 mm/s). The target speed
out for use in a system evaluation in this study (Fig. 2b). was set at 10 mm/s for investigation of differences in
exposure dose. In the validation of differences in target
speed, the exposure dose was set at reference dose levels,
0.2 mR in the direct-conversion type and 0.04 mR in the
indirect-conversion type.
detector surface. The motion speed was 10, 20, or 30 mm/ 3 Results
s, covering the range of lung tumor movement during
normal breathing [16]. 3.1 Calculation of predicted value
2.4.3 Measurement of lag effect Figures 3 and 4 show the process of creating a target with
image lag, and Fig. 5 shows the resulting images. The
ROIs of 40 9 40 mm2 were located on the lung nodule motion speed of all targets in Fig. 3 was set at 10 mm/s.
area and trimmed. Profile curves were obtained horizon- The exposure dose of all targets in Fig. 4 was set at ref-
tally across the moving target, and the half-value width of erence levels, 0.2 mR in the direct-conversion type and
the profile curve was measured for quantification of image 0.04 mR in the indirect-conversion type. Differences in the
blurring on the contours of a moving target as the differ- effect of image lag were estimated according to the known
ence in width measured between the static and moving image lag properties as follows:
targets. In addition, increases in pixel value due to image (i) In terms of radiation dose, the effect of image lag
lag were calculated as pixel values relative to the maxi- increases with increases in exposure dose in an indirect-
mum original value measured on the static nodule. The conversion FPD, because the lag effect is dependent on the
results were compared among simulation and actual mov- dose level. In contrast, there is little or no difference among
ing targets, and against each predicted value. different dose levels in a direct-conversion FPD, because of
Fig. 3 Profile curves of a target with image lag at each dose level in the case of a digital phantom simulating a lung tumor 1 cm in diameter. T(0)
means a target without lag, which is equivalent to a static target, moving at 10 mm/s
Simulation system for understanding the lag effect 277
Fig. 4 Profile curves of a target with image lag at each target speed static target at reference dose levels, 0.2 mR in direct-conversion type
in the case of a digital phantom simulating a lung tumor 1 cm in and 0.04 mR in indirect-conversion type
diameter. T(0) means a target without lag which is equivalent to a
the independence of image lag on the dose level (Figs. 1, 3; the same region, whereas faster target motion affects a
Table 1). larger area (Fig. 4; Table 1).
(ii) In terms of target speed, the lag effect shows no (iii) In terms of FPD type, almost the same effect is
trend, because of the offset between two factors—slower observed in direct- and indirect-conversion FPDs, because
target motion results in accumulation of more image lag in of the offset between large image lag (%) in a direct-
278 R. Tanaka et al.
Table 1 Predicted image blurring on the contours of a moving target Table 2 Maximum image blurring found on the contours of a
at each dose level and target speed moving target at each dose level (FPD type A)
Dose level 1/4 Ref. 1/2 Ref. Ref. Ref. Ref. 0.05 mR 0.1 mR 0.2 mR
Target speed 10 mm/s 10 mm/s 10 mm/s 20 mm/s 30 mm/s
Actual 0.8 0.7 0.7
Direct 0.9 0.9 0.9 0.6 0.6 Simulation 0.8 0.7 0.7
Indirect 0.5 0.7 1.2 1.4 0.9 Prediction 0.9 0.9 0.9
Reference dose levels (Ref.) were 0.2 mR in direct-conversion FPD
and 0.04 mR in indirect-conversion FPD
Unit: mm Table 3 Maximum image blurring found on the contours of a
moving target at each dose level (FPD type B)
conversion FPD and continuous image lag (%) in an 0.01 mR 0.02 mR 0.04 mR
indirect-conversion FPD (Fig. 1). Actual 0.3 0.7 0.7
Simulation 0.7 0.7 0.8
3.2 System evaluation
Prediction 0.5 0.7 1.2
Fig. 6 Profile curves measured horizontally across actual static and moving targets and a simulated motion target, obtained at three different
dose levels (Upper direct-conversion FPD, lower indirect-conversion FPD). The target speed was set at 10 mm/s
Simulation system for understanding the lag effect 279
Fig. 7 Profile curves measured horizontally across actual static and conversion FPD, lower indirect-conversion FPD). The exposure dose
moving targets and a simulated motion target obtained at imaging was set at reference dose levels, 0.2 mR in direct-conversion type and
rates of a 10 mm/s, b 20 mm/s, and c 30 mm/s (Upper direct- 0.04 mR in indirect-conversion type