HORIZONTAL F LAMMABILITY (UL94) TEST SHEET Page 1 of 1

Sample Identification #: J384035 Date: 11-13-20 Conditioning: 23/50/48 70°C / 168 hours Time: 13:00

Specimen Dimensions: inches millimeters


Temperature:
18.6°C
Humidity Level: Technician: Length: 5.042 Radius Verification: Pass Fail
Corner Radius Range: < 1.3 mm Color: Black
45% Bobc
Light Reading: Width: 0.496 Back Pressure: 4 mm
lux
Thickness: 0.063 Flow Rate: 110 ml/min

Category Designation
Material shall be classified HB if:
 Material does not ignite or the flame front does not reach the 100mm reference mark
 If the flame front reaches the 100mm reference mark:
* The linear burning rate cannot exceed 40mm/minute for 1/8” specimens
* The linear burning rate cannot exceed 75mm/minute for 1/16” and 1/32” specimens
Specimen Burn to the 25mm Burn to the 100mm Time, Average Burn
Total Burn Length Calculated Burn Rate
# reference mark? reference mark? seconds Rate

1 Yes Yes 75mm 148.8 30.24

29.86mm/min
2 Yes Yes 75mm 186.8 24.09
1.18 in/min

3 Yes Yes 75mm 127.7 35.24

Does Material meet HB Classification: Yes No


Calculation :

B = (60 * D) / T
B = Burn rate in millimeter / minute
D = Length the flame travels in millimeter
T = Time, in seconds, for the flame to travel “D”
Note:
 Only report the linear burn rate for specimens that the flame front reaches the 100mm reference mark
 If the flame front passes the 25mm mark but does not reach the 100mm reference mark, only report the burned length and
elapsed time
 If flame front reaches the 100mm reference mark, burn length equals 75mm

Flammability Terminology (Results of Test) Symbol

DOES NOT IGNITE. Material does not support combustion during or after ignition. DNI
SELF-EXTINGUISHING. Material ignites but does not burn to the timing zone. SE
Comments:

Revision: 10/05/16 FORM Owner: Laboratory Technical Manager