Professional Documents
Culture Documents
Maintainance of Instruments 2
Maintainance of Instruments 2
jbk@usic.unipune.ernet.in
Laboratory Instruments ?
1. 2. 3. 4. 5. 6. 7. 8. 9.
2
Different types of meters - Voltage, Current, (AC/DC) OHM meter, Analog / Digital Multimeters Power Supplies Regulated / Unregulated, AC / DC, Low Voltage, High Voltage, UPS Signal / Function Generator, Audio / RF Generators Oscilloscopes, Single Channel, Dual Trace, Digital Storage Oscilloscope Digital Frequency Counter, Universal Counter Analytical Instruments like pH mater, Conductivity meter, Colorimeter, Spectrometer Vacuum Gauges Pirani and Penning gauge G. M. Counter Electronic gadgets for specific experiments and many more
Repair and Maintenance of Laboratory Instruments and Equipments
Repairing is done after the failure in the instrument. Maintenance is a periodic activity. Carried out either on scheduled date, or whenever instrument starts deviating form its normal behavior.
4
Repair and Maintenance of Laboratory Instruments and Equipments
Maintenance of Instruments
Observe / monitor the behavior of the instrument to take the corrective action in time. It will save your precious time as well as money.
2. Tighten the mechanical components 3. Check electrical connections and joints 4. Replace faulty cables, sockets and pins etc.
5
Repair and Maintenance of Laboratory Instruments and Equipments
Repairing of Instruments Breakdown / Corrective Maintenance Instrument partially or completely stops working need urgent attention
Ask when, why and how it failed? Gather as much information as you can about last status of the instrument before it went bad Try to analyze the situation and locate the fault Confirm and rectify Recalibrate the instrument at least using the available standard Note down in the log book
6
Repair and Maintenance of Laboratory Instruments and Equipments
10
11
12
13
14
15
16
pH Meter
pH is an indication of the samples acidity, pH is actually a measurement of the activity of hydrogen ions in the sample. pH measurements run on a scale from 0-14, with 7.0 considered neutral. Those solutions with a pH below 7.0 are considered acids, and those above 7.0 are designated bases. The pH scale is logarithmic, so a one unit change in pH actually reflects a ten-fold change in the acidity. For instance, orange juice (pH 4) is ten times more acidic than cottage cheese (pH 5)
17
Repair and Maintenance of Laboratory Instruments and Equipments
pH Meter
18
How pH is measured?
A probe is placed in the sample, and the pH is read directly from the meter. The pH probe measures the millivolts of potential between the reference electrode and the pH electrode (59.16 mV at 25 o C per pH).
19
Repair and Maintenance of Laboratory Instruments and Equipments
Nernst Equation
The meter converts this reading to pH using the Nernst Equation.
E = Eo + (R x T/ n x F) ln(a) R = Universal Gas Constant, T = Absolute temperature, n = Charge of the ion (including sign) F = Faradays constant and a = Activity of the ion.
20
Repair and Maintenance of Laboratory Instruments and Equipments
where
10
21
22
11
Electronics in pH measurement
23
TEMPERATURE ZERO
14.00
24
ADC
12
13
14
29
15
31
16
Colorimeter
Based on the absorption of electromagnetic radiation in visible range by the sample. Consists of a light source (Tungsten lamp), Set of Optical filters, Entrance Slit, Sample holder, a Photodetector and an indicating meter.
Lamp Photodetector
Filter
33
Mechanical Slit
Sample holder
Readout
Spectronic 20
34
17
35
OPERATING INSTRUCTIONS for Spectronic 20 1. Turn the instrument ON by turning the Power Switch knob clockwise (toward the right). Allow the colorimeter to warm up for 15 minutes in order to stabilize the light source and the detector. 2. After the warm-up period, set the desired wavelength with the Wavelength Control knob. For instance, the wavelength for the manganese determination is 525 nm. 3. Set the display mode to transmittance by pressing the MODE control key until the light beside the "Transmittance" is lit. 4. Adjust the display to 0.0% T with the Zero Control knob. Make certain the sample compartment is empty and the cover is closed tightly while make this adjustment. 5. Fill a cuvette with the blank solution to the top of the triangle on the side of the cuvette. Wipe the cuvette with a Kimwipe to remove any liquid and fingerprints on the outside of the cuvette. Both of these will interfere with light transmission and will cause erroneous readings.
36
Repair and Maintenance of Laboratory Instruments and Equipments
18
6. Position the cuvette with the vertical guide line (fiducial line) facing toward your right and insert the tube gently but completely into the cuvette compartment. Next rotate the cuvette by 90 in a clockwise direction in order to align the guide line on the cuvette with the guideline on the sample compartment. This technique prevents scratching of the cuvette in those areas through which the light will pass. Scratches on the cuvette can lead to erroneous measurements. Close the cover of the compartment. 7. Using the Transmittance / Absorbance control knob adjust the display to read 100.0%. 8. Press the MODE control key and switch the Status Indicator light to read Absorbance. If the Transmittance calibration was done correctly the display should read 0.0. No further adjustment is required. If the display does not indicated 0.0 use the Transmittance / Absorbance control knob to adjust it to that value. Using the MODE key switch the display back to Transmittance. 9. Remove the cuvette from the compartment by reversing the previous procedure: rotate the cuvette 90 in a counterclockwise direction then remove it. Fill another cuvette with the solution whose absorbance you wish to measure. Insert it into the compartment in the same manner as before. a. Read the %T value directly from the digital display. b. Using the MODE key select Absorbance and read the A value directly from the digital display. Again select Transmittance 10. Remove the cuvette from the sample compartment by reversing the procedure used to insert the it. Close the lid to the compartment.
37
Repair and Maintenance of Laboratory Instruments and Equipments
POSSIBLE PROBLEMS A. Make sure the cuvettes are clean. B. Fill cuvettes two-thirds full or to the top of the triangular mark. C. Always use the same cuvette for the blank solution and the other one for the sample. D. Make certain that the index mark on the cuvette is lined up with the index mark on the sample compartment before taking a reading. Doing so insures that the cuvette is in the same position for each measurement. E. If the wavelength control is moved in the slightest it is necessary to reset zero and 100 %T. F. If you suspect that your cuvettes are showing differences greater than 3% T, you should search for a pair which has closer agreement. To find such a pair, fill two clean cuvettes with distilled water. Calibrate the instrument using one as a blank. Then read %T on the other. If your reading is less than 97%T or greater than 103%T, fill a third with distilled water and take a new reading. Any pair showing differences greater than 3%T may adversely affect your grade on an experiment which uses the Spectronic 20-D spectrophotometer
38
Repair and Maintenance of Laboratory Instruments and Equipments
19
Thank You!
39
20