Professional Documents
Culture Documents
BANDERAS
BANDERAS
Differential false negative flagging issues, if not handled properly, can escalate rapidly into
adverse situations for both the customer and Beckman Coulter. As there are many factors that
affect flagging – the instrument, the flagging algorithm and user-defined setup, and the
limitations of the technology - this Technical Update will provide guidance for bringing false
flagging situations to a quick resolution.
The Instrument
Many of the differential algorithm criteria for both contouring (gating) and flagging are based on
the DC, opacity, and RLS mean and standard deviation (SD) of the individual white cell
subpopulations. Abnormal cell types will typically exhibit a population mean and/or SD outside
of expected ranges and generate an appropriate Suspect flag.
NERLS SD
NEOP SD
NEDC Mean = Neutrophil Volume Mean
NEOP Mean = Neutrophil Opacity Mean
NERLS Mean = Neutrophil Light Scatter Mean NEOP Mean
NEDC SD = Neutrophil Volume Standard Deviation
NEOP SD = Neutrophil Opacity Standard Deviation
NERLS SD = Neutrophil Light Scatter Standard Deviation
NERLS Mean
When white cell populations shift or become diffuse because of instrument related problems, the
means and SDs of populations representing normal cell types may be outside the expected
ranges resulting in an increase in false positive differential flagging and/or incorrect differential
results. The customer complaint will be “too much flagging”. This scenario will typically result in
a service visit to optimize the instrument – bringing the means and SDs back to where they
should be.
CONFIDENTIAL
Reproduction without written permission of Beckman Coulter Incorporated is prohibited.
… Still Counting®
MTS41601-A
TU2000M061 Page 2 of 4
False negatives occur when the customer observes a significant number of abnormal cell types
on the manual differential and the instrument generated no Suspect flags. Be aware that a
number of false negatives can be expected as there is a balance between flagging too much
and flagging too little. Our obligation is to minimize the occurrence of false negatives by
ensuring correct placement of all differential populations.
Example 1: If the DC Mean is low, the second population of lymphocytes commonly seen with
CLL and lymphoma patients will be shifted low and may be gated as non-WBC debris instead of
lymphocytes.
Example 2: If the DC Mean is low, the neutrophil population mean may be too low to trigger the
Imm Ne1 flag as needed.
Example 3: If the RLS Mean is high, the neutrophil population may be falsely gated as
eosinophils.
The Flagging
The GEN·S offers three separate levels of flagging along with the ability to disable the Imm NE
1 flag. Appropriate flagging, specific to the needs of the laboratory, at the onset of instrument
implementation is crucial. The laboratory’s patient population and their sensitivity to abnormal
cell types must be taken into consideration when assisting with flagging limits. In addition,
sufficient sample size and type must be used to test the flagging of the system.
Example: If the laboratory is band sensitive, disabling the Imm NE 1 flag without validation may
result in an unexpected increase of false negatives.
The GEN·S flagging algorithm is unique to the GEN·S. There have been many changes in
regards to contouring and flagging when compared to the STKS/MAXM algorithms. Pre-existing
differential review criteria should be validated prior to use with the GEN·S.
It is extremely important to understand that false positives and false negatives are a
normal outcome of screening tests such as the automated differential. Our obligation is
to minimize these occurrences by instrument optimization and appropriate flagging.
CONFIDENTIAL
Reproduction without written permission of Beckman Coulter Incorporated is prohibited.
… Still Counting®
MTS41601
TU2000M061 Page 3 of 4
2. Explain to the customer that their complaint is being documented, that a Service
Representative will validate the performance of their instrument, optimize the instrument if
needed, and that an Applications Specialists will review their flagging criteria with them to
ensure they are optimally set for their laboratory needs and their patient population.
1. Upon receipt of the Service Order, download the listmode data (both .dat and .inf files) from
the sample(s) in question to a diskette. Retain diskette for future reference.
2. Review Field Trends for proper postional parameters (NEDCMean, NEOPMean, and
NERLSMean). Contact your Hardware Specialist for assistance if necessary.
5. Advise the customer that the instrument has been validated to be within specifications
specific to differential contouring and flagging or optimized to minimize the occurrence of
false negative flagging.
Applications Specialist
1. Were sensitivity and specificty studies performed prior to implementation of the GEN·S?
§ Discuss the need to establish criteria that match the flagging performance of the
GEN·S.
§ Recommend that a study be performed to determine appropriate flagging.
2. Did the orginal studies include a sufficient number of abnormal samples to test the system?
CONFIDENTIAL
Reproduction without written permission of Beckman Coulter Incorporated is prohibited.
… Still Counting®
MTS41601
TU2000M061 Page 4 of 4
3. Did the original studies include samples representative of the laboratory’s typical workload?
§ Discuss the need to meet the laboratory’s flagging needs for pediatrics, oncology,
outpatients, emergency room, etc.
§ Recommend that a new study be performed or additional samples be added to the
existing study.
4. Did the original studies indicate a false positive or false negative rate markedly different than
what is currently being observed?
CONFIDENTIAL
Reproduction without written permission of Beckman Coulter Incorporated is prohibited.
… Still Counting®
MTS41601