1 s2.0 S008525381559323X Main

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

Kidney International, Vol. 49 (1996), pp.

255—260

TECHNICAL NOTE

Non-invasive, optical measurement of absolute blood volume in


hemodialysis patients
DAVID W. JOHNSON, MIRIAM MCMAHON, SCOTT CAMPBELL, JAMES WILKINSON, NEIL KIME,
GERALD SHANNON, and SIMON J. FLEMING

Departments of Renal Medicine, Physical Sciences, and Haematology, Royal Brisbane Hospital, Brisbane, Queensland, Australia; Optoelectronics
Research Centre, University of Southampton, Southampton, United Kingdom; and Department of Electrical Engineering, University of Queensland,
St. Lucia, Brisbane, Queensland, Australia

During hemodialysis, fluid is removed from or infused into it is predominantly absorbed by hemoglobin (Hb), independently
patients after careful clinical assessment of their hydration states. of its degree of oxygen saturation [18—201. Provided column width
Unfortunately, these assessments are often imprecise, resulting in is constant, the natural log of the attenuated optical signal, [Ln
either fluid overload or intravascular volume depletion. Since (P)], is inversely proportional to hemoglobin concentration ([Hb])
hypovolemia plays an important role in symptomatic hypotension, over the ranges of [Hb] usually encountered in hemodialysis
which complicates up to 25% of dialysis treatments [1], there has patients (50 glliter to 120 glliter) [4]. By collecting blood samples
been increased interest in blood volume (BV) monitoring in to calibrate the optical monitor's output, this received optical signal
recent years. can be converted into a continuous recording of [Hb] [4]. The
Previous authors have suggested monitoring change in BV subsequent calculation of absolute BV was based on a mathematical
during dialysis by examining alterations in optical density [2—5], method described by Schallenberg, Stiller and Mann [3].
optical reflection [6, 71, mass density [8], viscosity [9], protein Assuming ideal mixing of blood and a constant mass of
concentration [10] or electrical conductivity [11—14]. However, hemoglobin within the vascular compartment, [Hb] is inversely
none of these described methods has measured absolute BV, proportional to BV. During a period of ultrafiltration between
which is at least as important a determinant of dialysis-related time 0 minutes and time t minutes, the proportional BV (BV()/
hypotension as the relative change in BV [15, 16]. The best BV(O)) can be calculated as follows:
method currently available is measurement of red cell and plasma
volumes by dilution of radioactively-labeled blood elements. The
time-consuming nature of this method and the exposure of the
patient to radiation makes it unsuitable for routine clinical use,
BV(o)
— IIFlho1
— _____
[Hb(t)]
1 °
especially if a number of studies are required. As fluid is removed by ultrafiltration, oncotic pressure increases,
To our knowledge, there has only been one previous report of resulting in an unknown amount of vascular refilling from the
an attempt to non-invasively measure absolute blood volume in interstitial compartment. The rate of change of BV at time t is
hemodialysis patients, based on plasma density [17], but this was therefore equal to the refilling rate, Qr, less the ultrafiltration rate,
improperly validated against anthropometric calculations derived Q.
from a normal, healthy population.
We have developed a novel, non-invasive method of measuring dBV(t)/dt = Qr — Qu (Eq. 2)
absolute BV by photometric means, which we have previously
validated in Vitro [18]. The purpose of this study was to validate Using subscripts 1 and 2 to denote rates of change of volumes
the technique in ViVO, comparing the results obtained in ten during a defined time period immediately before and after a
does not alter
hemodialysis patients with those obtained using radioisotope change in ultrafiltration rate, and assuming °r
dilution techniques. appreciably during this time, the above equation may be written
for both time periods and the difference found:
Methods
[dBV(t)Idt]2 — [dBV(t)Idt]l = (Or Qu2) (Or — 0)
Physical principle
When light in the near infrared range (810 nm) is passed = Q1 — Q2 (Eq. 3)
through a column of blood within transparent dialysis blood lines,
Both sides of equation 3 can be divided by BV(0), obtaining

Received for publication January 6, 1995 Qul — °u2


__________
[d/dt BV(t)/BV(O)12 — [d/dt BV(t>/BV(O)1I =
and in revised form July 25, 1995 BV(0)
Accepted for publication July 27, 1995
© 1996 by the International Society of Nephrology This may be rearranged and combined with equation I to give:

255
256 Johnson et a!: Optical measurement of absolute BV

Dialyzer

Hemotilter

Pump
Computer
Lilt rat it rate I
(1 Fig. 1. Extracorporeal circuit for measuring a
I
patient's BV using optical techniques.

— Q2 Every 0.055 seconds, the optical monitor sampled the attenuated


power of the continuously emitted light beam after it passed
BV(o) =
d/dt {[l-ib(o)]I[FTb(t)I}2 d/dt {[l—Ib(o)1/[l—Ib(t)]}l ( 4\ through the blood line and the information was recorded on a
Q1 and Q2 are easily measured. The rate of change of [Hb(o)1/ connected computer. Because of potential variations in the diam-
[Hb(,)1 can be calculated from the slope of the line constructed eters and transparencies of the blood lines between studies, five
using the recorded optical data points [4]. The patient's starting blood samples were collected from the sample port at 7.5 minute
By, BV(0), can therefore be readily calculated. intervals during the 30-minute ultrafiltration phase of each exper-
iment. This enabled the optical signal output to be calibrated
Patients against [HbI, measured using a Coulter S Counter.
Ten stable hemodialysis patients (7 males, 3 females) gave Linear regression analysis was used to determine the relation-
informed consent to enter the study, which was approved by the ship between [HbJ and Ln(P), so that [Hb] could be calculated for
Royal Brisbane Hospital Ethics Committee. All had received each optical signal value measured during the experiment [4].
three to four hours of conventional hemodialysis three times a In a previous in vitro validation study [18] using stored human
week using cuprophane membranes for a median period of 2.7 blood pumped around the above dialysis circuit, commonly used
years (range 0.16 to 20.5 years). Median age was 67.5 years (range blood pump speeds (150 to 300 ml/min) and varying optical
39 to 76 years) and median predialysis weight was 71.1 kg (range monitor positions did not affect optical responses. Varying [Hb] in
50.3 to 85.9 kg). The primary renal diseases included chronic the range of 4 to 12 g/dl did not significantly disturb the linear
glomerulonephritis (4 patients), analgesic nephropathy (3), reno- relationship between [HbI and Ln(P). Accuracy was impaired by
vascular nephrosclerosis (2) and diabetic glomerulosclerosis (1). only more extreme degrees of red cell hemolysis (sufficient to
cause discoloration of ultrafiltrate). However, because high access
Apparatus recirculation could have potentially produced spuriously low BV
The extracorporeal circuit (Fig. 1) included a polyamide 2.0 m2 estimates, pilot in vivo studies were also performed in two patients
hemofilter (Gambro FH88H) and the patient's usual cuprophane with arterial connections to a central venous access, where the
hollow-fiber dialyzer connected in series. Blood was pumped at venous connection was changed between the other lumen of the
200 ml/min from the patient's arteriovenous fistula or subclavian central venous access catheter and an arteriovenous fistula. No
catheter through these two devices by a Cobe CentrySystem 3 appreciable difference was found between BV estimates in the
automatic supply unit and returned to the patient. Anticoagula- same patient despite an obvious and measurable difference in the
tion was by routine systemic heparinization. degree of access recirculation.
During the study period, the dialyzer was set in bypass mode
and its input and output ports clamped, so that neither dialysis nor
ultrafiltration took place. Ultrafiltrate was only removed through Study design
the hemofilter at a rate controlled by a peristaltic pump connected
to the ultrafiltrate port. The ultrafiltrate was collected into a Each study took place immediately prior to a patient's routine
measuring cylinder and its volume continuously monitored. dialysis session. Patients remained supine and fasted for two hours
The optical monitor head, based on the design of Wilkinson et before and throughout the study.
al [4], consisted of a photodetector and a laser diode with a Plasma and red cell volumes were then measured by standard
spectral output at the isobestic point of Hb at 810 nm. It clipped radioisotope dilution methods [21], using '311-labeled albumin
around the "arterial" blood line proximal to the hemofilter. Two and 51Cr-labeled red cells, respectively. BV(o) was calculated by
clear perspex inserts held the PVC blood tubing firmly in place to the summation of these volumes. To conduct a cross-check on the
minimize tubing diameter variations due to pulsatile blood flow. validity of this measurement, a whole blood/venous hematocrit
An optical filter was included to remove ambient light effect. ratio was performed.
Johnson et a!: Optical measurement of absolute BV 257

—600
Start of ultrafiltration

—700

—800 -

—900 -

-—1000 -

—1100 -

End of ultrafiltration
—1200
I• F 'I r F I
0 500 1000 1500 2000 2500 3000 3500 Fig. 2. Typical tracing of received optical power
versus time in one pilot study patient before, during
Time, seconds and after a period of isolated ultrafiliration.

The patient was then connected to the extracorporeal circuit Results


after the priming volume of saline had been discarded. This was
followed by a 15-minute equilibration period in which blood was Effect of fluid removal on received optical power
pumped through the circuit without ultrafiltration taking place. A typical tracing of received optical power versus time is shown
Approximately 900 ml of ultrafiltrate was then withdrawn at a in Figure 2. Prior to the commencement of ultrafiltration, when
constant rate over 30 minutes. A further 15-minute equilibration equilibration was taking place, there was little overall change in
period without fluid removal followed. The ultrafiltrate port of the received optical power. During ultrafiltration, the received signal
hemofilter was then clamped and the patient received his or her intensity decreased as BV contracted. However, after the first five
usual dialysis. minutes, the rate of decline of received optical power decreased
Two separate photometric determinations of BV(o) were made due to a progressive increase in vascular refilling rate. This
using data obtained when ultrafiltration rate abruptly changed at refilling continued for a short period of time after the completion
the start (BV(O)start) and end (BV(O)end) of the ultrafiltration of ultrafiltration, leading to a small rebound increase in the signal.
phase. At the start of ultrafiltration, Q equalled 0 mi/mm and The optical tracings of all patients revealed sizeable moment-
Q2 equaled the measured ultrafiltration rate. At the end of to-moment signal variations, which effectively translated into
ultrafiltration, Q equaled the measured ultrafiltration rate and fluctuations in [HbI of up to 3.5%. The magnitude of these signal
fluctuations was generally greater toward the end of ultrafiltration
QU2 equaled 0 mI/mm. The sampling time period, t, for calculating
each BV(o) was set at 4.58 minutes (or 5000 data points) before and greater in the eight study patients with intact spleens com-
and after the change in ultrafiltration rate on the basis of preliminaiy pared with two study patients who were asplenic.
studies. d/dt {[Hb](o)/[HbI(t)}l and d/dt {[Hb](o)/[l-Ib(t)}2 were cal-
culated from the slopes of the regression lines of [Hb] versus time
Ln (P) versus [Hb]
over the 4.58 minutes before and after the change ir ultrafiltration Calibration plots of [Hb] versus Ln(P) in individual patients
rate. BV(o) was then calculated (Eq. 4). were all associated with highly statistically significant (P < 0.01)
BV measurements by the radioisotope and photometric meth- correlation coefficients in excess of 0.97 (Fig. 3).
ods were performed by different observers who were blinded to
each other's results. When all ten patients had been studied, the Comparison of BV measurement by photometric and radioisotope
paired results were compared using linear regression analysis. dilution techniques
Radioisotope BVs (BVRJ) ranged between 4.077 liter and 6.211
Statistical analysis liter, with a median value of 4.9 19 liter. In each case, the whole
Statistical analysis was performed using the software package, blood/venous hematocrit ratio was within the anticipated range of
Sigmastat version 1.01. Data were initially tested for normality 0.84 to 0.95.
(Kolmogorov-Smirnov test), homoscedacity and sequential corre- Photometrically-derived BVs calculated from optical signals
lation of adjacent error terms (Durbin-Watson statistic). Compar- collected around the start of ultrafiltration (BV(O)start) ranged
isons between radioisotope and photometric BVs were carried out between 1.817 liter and 10.05 liter (median 5.127 liter). The data
using the Mann-Whitney rank sum test. The relationship between passed tests for normality (P > 0.2) and homoscedacity (P =
the two methods was then compared using linear regression 0.296), and the Durbin-Watson statistic was 2.456. A Mann-
analysis. The 0.05 level of significance was used for all analyses. Whitney rank sum test revealed no significant difference in outcomes
258 Johnson et al: Optical measurement of absolute BV

50
45

-z 30

-J 25
20 Fig. 3. Typical correlation plot obtained in one of
8.5 9 9.5 10 the study patients between [HbJ and Ln (P) during
isolated ultrafiltration (y = 111.045 — 7.957x; r =
Hemoglobin conc entration [Hb], g/dI —0.973; N = 5; P = 0.008; SEE = 0.236).

7
Q)

>
w5 .
0
0
(1)4
0 Fig. 4. Comparison of BVs derived by the new
optical method and those derived by standard
4, radioisotope dilution techniques in ten
0 hemodialysis patients (y = 3.707 + 0.225x; r =
0 2 4 6 8 10 12 0.822; P = 0.004; SEE = 0.403). 95%
confidence intervals for the regression line are
Optically measured By, liters represented by dotted lines.

Table 1. Predicated BVs for 10 hernodialysis patients calculated by Discussion


applying the regression equation, y = 3.707 + 0.225x, to optically We have described a new optical method for measuring abso-
measured BVs
lute By, which is quick, inexpensive and non-invasive (above and
Calculated BV liters 95% Confidence intervals beyond the degree of invasiveness required for routine hemodi-
5.456 5.025—5.886 alysis). The method is based on the difference in rates of change
4.686 4.368—5.004 of optical density of blood during a brief time period before and
5.971 5.297—6.642 after a change of ultrafiltration rate. By assuming that vascular
5.097 4.782—5.411 refilling is constant during the brief time period, this unknown
5.108 4.791—5.425
4.704 4.390—5.018
quantity can be cancelled out of the calculations, allowing deter-
4.434 4.039—4.828 mination of absolute By. Photometric determinations of BV
4.116 3.584—4.648 obtained when ultrafiltration rate abruptly changed from 0 mI/mm
5.018 4.716—5.320 to 30 ml/min at the commencement of ultrafiltration (BV(O)start)
4.407 4.002—4.812
correlated strongly (r = 0.822, P = 0.004) with BV measurements
using the standard technique of radioisotope dilution (BVRI).
This enabled calculation of predicted BVs with 95% confidence
limits of approximately 400 ml. A poorer correlation was seen
between BV()rt and BVRI (P> 0.1). Moreover, photometrically- between BVRI and photometric BV measurements obtained when
derived BVs correlated strongly with radioisotope BVs (r = 0.822; ultrafiltration was abruptly ceased (r = 0.035).
P = 0.004; SEE 0.403; y = 3.707 + 0.225x; Fig. 4). The 95% There are two principal potential sources of error when apply-
confidence limits for the predicted BVs derived from the regres- ing the technique around the start of ultrafiltration. The first is
sion equation, y = 3,707 + 0.225x, were of the order of 400 ml due to the possible occurrence of significant change in vascular
(Table 1). refilling rate immediately after the alteration in ultrafiltration
By contrast, photometrically-derived BVs calculated from op- rate, thus invalidating the assumptions used in deriving Equation
tical signals collected at the end of ultrafiltration (BV(())end) 3. An appreciable increase or decrease in vascular refilling rate
showed much greater variability (median 6.726 liter, range 3.8 to during the measurement period could lead to an overestimation
24.32 liter). Although rank sum testing failed to detect a signifi- or underestimation of absolute By, respectively.
cant difference in test outcomes by the two methods, optically The second potential error can arise from calculating the rate of
measured BVs were poorly correlated with radioisotope BVs (r = change of [HbI, (and hence, By), when there are marked
0.035; P = 0.929). moment-to-moment fluctuations in the received optical signal.
Johnson et al: Optical measurement of absolute BV 259

These signal fluctuations have also been reported with techniques tendency of Hb mass to change with time (such as with bleeding,
that measure BV change on the basis of blood viscosity [9] and erythropoietin therapy, etc.).
plasma protein concentration [10]. They were of similar magni- Additional potential future applications of this method in the
tude and may therefore predominantly represent rapid fluid shifts hemodialysis setting may include guiding the prescription of
between different compartments within the vascular space. A certain treatments (such as fluid removal on dialysis and exchange
small proportion of the oscillations may also be accounted for by volumes for plasma exchange) and monitoring the BY effects of
splenic mobilization of small boluses of red cells into the circula- various treatments (such as erythropoietin therapy). It may fur-
tion, given that the magnitude of oscillations appeared less ther prove to be a useful research tool to investigate the physiol-
marked in two asplenic patients. More studies are needed to ogy of BY responses to ultrafiltration. In a previous in vitro
distinguish between these possibilities. validation study [18], we have demonstrated that the technique
The potential error due to an alteration in vascular refilling rate can be adapted to measure the rate of fluid refilling from the
is minimized by the use of a small sampling period after the interstitial compartment to the vascular compartment during
ultrafiltration rate is changed. However, errors due to moment- ultrafiltration. Inadequate vascular refilling is suspected to be a
to-moment signal variations become more significant when short major cause for dialysis-related hypotension [17], but has not been
time periods are selected, due to the collection of a smaller readily measurable previously.
number of data points for calculation of the regression line. This In summary, the optical technique described in this paper is
especially applies to data collected at the end of ultrafiltration, quick, simple, inexpensive and noninvasive. Photometric BYs
where signal variations were more marked, and this probably measured at the commencement of a period of isolated ultrafil-
accounts for the poor correlation of BV(O)efld with BVRI. tration correlated strongly with BYs derived by the standard
A time period of 4.58 minutes (5000 data points) was therefore technique of radioisotope dilution. Taken in conjunction with
selected after preliminary studies to attempt to minimize both clinical assessment, the method may enable a more accurate
sources of error. The curtailment of these errors at the start of assessment of intravascular hydration to be performed. Future
studies are planned to assess the ability of continuous photometric
ultrafiltration was supported by the failure of the Mann-Whitney
BY measurements to predict and avoid the development of
rank sum test to find any evidence that BV(O)start systematically
over- or underestimated BVRI. symptomatic hypotension due to hypovolemia.
Potential errors arising from intertreatment variations in opti-
Reprint requests to Dr. David Johnson, Department of Medicine, Level 3,
cal path length and tubing transparency were circumvented by Wallace Freeborn Professorial Block, Royal North Shore Hospita4 St.
calibrating the received optical power with [Hb] for each patient. Leonard's, Sydney, NSW, Australia 2065.
We also avoided errors induced by an effect of changing erythro-
cyte mean corpuscular volume (MCV) on received optical power References
by removing fluid during isolated ultrafiltration, where plasma
osmolarity, and hence MCV, does not alter [22]. However, further 1. HENDERSON LW, Kocu KM, DINARELLO CA, SHELDON S: Haemodi-
work is needed to assess the validity of the photometric method in alysis hypotension: The interleukin hypothesis. Blood Punf 1:3—8,
certain uncommon clinical circumstances, such as patients with 1983
hemoglobinopathies or excessive Rouleaux formation. 2. STEUER RR, HARRIS DH, CONNIS JM: A new optical technique for
monitoring haematocrit and circulating blood volume: Its application
For any given dialysis session, once initial absolute BV is in renal dialysis. Dial Transplant 22:260—265, 1993
established photometrically, the same equipment can be used to 3. SCHALLENBERG U, STILLER S, MANN H: A new method of continuous
continuously measure absolute BV during subsequent ultrafiltra- haemoglobinometric measurement of blood volume during hemodi-
tion in the same dialysis session. Use of the regression equation alysis. Life Support Systems 5:293—305, 1987
4. WILKINSON JS, FLEMING Si, GREENWOOD RN, CATrELL WR: Contin-
y = 3.707 + 0.225x enables optically measured BY to be trans- uous measurement of blood hydration during ultrafiltration using
lated into isotopically measured BY with a 95% confidence limit optical methods. Med Biol Eng Comput 25:317—323, 1987
of 400 ml. This represents a potential variation of 8.1% in the 5. STILLER 5, THOMMES A, KONIGS F, SCHALLENBERG U, MANN H:
median value of 4.919 liters for the radio isotopically measured Characteristic profiles of circulating blood volume during dialysis
therapy. ASAJO Trans 35:530—532, 1989
BYs. However, given that the isotope method itself can incur 6. DR VRIES JPPM, OLTHOF CG, VISSER V, 1(00W PM, DE VRIES PMJM:
error rates as high as 10% [23, 24], a significant portion of the Continuous measurement of blood volume using light reflection:
measured variation between the two methods will be due to errors Method and validation. Med Biol Eng Comput 31:412—415, 1993
in the reference method itself. 7. DE VRIES JPPM, Kouw PM, VAN DER MEER NJM, ET AL: Nun-
invasive monitoring of blood volume during hemodialysis: Its relation
Nevertheless, the size of the variation could be potentially large
with post-dialytic dry weight. Kidney mt 44:851—854, 1993
enough to limit the clinical utility of the photometric method. A 8. KENNER T: The measurement of blood density and its meaning. Bas
future clinical study is therefore planned to monitor photometric Res Cardiol 84:399—401, 1989
measurements of absolute BY continuously in individual patients 9. GREENWOOD RN, ALDRIDGE C, CATrELL WR: Serial blood water
estimations and in-line viscusimetry: The continuous measurement of
during symptomatic hypotension. It is our expectation that com- blood volume during dialysis procedures. Clin Sci 66:575—583, 1984
parisons performed on an individual rather than a population 10. SCHNEDITZ D, POGGLITSCH H, HORINA J, BINSWANGER U: A blood
basis will further improve sensitivity and accuracy, and thereby protein monitor for the continuous measurement of blood volume
assist clinicians in detecting and preventing symptomatic hypoten- changes during hemodialysis. Kidney mt 38:342—346, 1990
sion related to hypovolemia. The concept of "dry weight" could 11. DE VRIES PMJM, Kouw PM, MEIJER JH, OE LP, SCHNEIDER H,
DONKER JM: Changes in blood parameters as determined by conduc-
hopefully therefore be supplemented with "dry BY." This is tivity measurements. ASAJO Trans 34:623—626, 1988
superior to the use of target [Hb], as has been proposed by Steuer, 12. MAEDA K, SHINZATO T, GOSHIDA F: Newly developed laboratory
Harris and Connis [2] and de Yries et al [6, 7], because of the blood volume-monitoring system and its implications for measuring
260 Johnson et al: Optical measurement of absolute BY

changes in blood volume during ultrafiltration. Artif Organs 10:452— blood volume during hemodialysis using optical techniques. Med Eng
459, 1986 Phys (in press)
13. BONNIE E, STILLER S, MANN H: The influence of fluid overload on 19. ANDERSON NM, SEKEU P: Studies on light transmission of non-
vascular refilling rate in hemodialysis: Continuous on-line measure- haemolysed whole blood. Determination of oxygen saturation. J Lab
ment with the conductivity method. Prog Artif Organs 85:135—136, Clin Med 65:153—166, 1965
1986 20. MENDELSON Y, CHEUNG PW, NEUMAN MR: Non-invasive transcuta-
14. STILLER S, MANN H, BYRNE T: Continuous monitoring of blood neous monitoring of arterial blood gases. IEEE Trans Biomed Eng
volume during dialysis. Proc Eur Soc Artif Organs 7:167—171, 1980 31:792—800, 1984
15. LIN5 LE, HEDENBORG G, JACOBSON H, SAMUELSON K, TEDNER B, 21. INTERNATIONAL COMMITFEE FOR STANDARDIZATION IN HAEMATOL-
ZETITERHOLM UB, LJUNGOVIST 0: Blood pressure reduction during
OGY: Recommended methods for measurement of red-cell and
hemodialysis correlates to intradialytic changes in plasma volume. plasma volumes. JNucl Med 21:793—800, 1980
Clin Nephrol 37:308—313, 1992
16. KIM ICE, NEFF M, COHEN B, SOMERSTEIN M, CI-IINITZ J, ONESTI G, 22. FLEMING SJ, WILKINSON JS, ALDRIDGE C, GREENWOOD RN, MUG-
SWARTZ C: Blood volume changes and hypotension during hemodi- GLESTON SD, BAKER LRI, CAYrELL WR: Dialysis-induced change in
alysis. ASAJO Trans 16:508—514, 1970 erythrocyte volume: Effect on change in blood volume calculated from
17. SCHNEDITZ D, R00B J, OSWALD M, POGGLITSCH H, MOSER M, packed cell volume. Clin Nephrol 29:63—68, 1988
KENNER T, BINSWANGER U: Nature and rate of vascular refilling 23. RAWLE PR, SEELEY HF: Assessing blood volume, blood loss and
during hemodialysis and ultrafiltration. Kidney mt 42:1425—1433, 1992 blood replacement. Br J Hosp Med 38:554—557, 1987
18. MCMAHON MP, CAMPBELL SB, SHANNON GF, WILKINSON JS, FLEM- 24. NAJEAN Y, CACCHIONE R: Blood volume in health and disease. Clin
ING SJ: A non-invasive continuous method of measuring absolute Haematol 6:543—561, 1977

You might also like