2021 International Conference on Radar, Antenna, Microwave, Electronics, and Telecommunications

High-Flow and High-Pressure Oxygen Mixing


for Ventilator System
2021 International Conference on Radar, Antenna, Microwave, Electronics, and Telecommunications (ICRAMET) | 978-1-6654-2816-3/21/$31.00 ©2021 IEEE | DOI: 10.1109/ICRAMET53537.2021.9650464

Eko Joni Pristianto Deni Permana Kurniadi Hana Arisesa


Research Center for Electronics and Research Center for Electronics and Research Center for Electronics and
Telecommunications Telecommunications Telecommunications
National Research and Innovation Agency National Research and Innovation Agency National Research and Innovation Agency
of Indonesia of Indonesia of Indonesia
Bandung, Indonesia Bandung, Indonesia Bandung, Indonesia
ekojonip5@gmail.com dennypermana69@gmail.com hana003@lipi.go.id

Dayat Kurniawan Erry Dwi Kurniawan Yusuf Nur Wijayanto


Research Center for Electronics and Research Center for Electronics and Research Center for Electronics and
Telecommunications Telecommunications Telecommunications
National Research and Innovation Agency National Research and Innovation Agency National Research and Innovation Agency
of Indonesia of Indonesia of Indonesia
Bandung, Indonesia Bandung, Indonesia Bandung, Indonesia
daya004@lipi.go.id erry.2k@gmail.com ynwijayanto@gmail.com

Budi Prawara
Research Center for Electrical Power and
Mechatronics
National Research and Innovation Agency
of Indonesia
Bandung, Indonesia
budi006@lipi.go.id

Abstract— The treatment of patients with acute respiratory A ventilator, specifically a medical ventilator, is one of
failure due to COVID-19 needs ventilator machine support to the medical devices needed to support the healing of
provide oxygenation and ventilation processes. The most COVID-19 patients. The medical instrument serves to
critical control in the machine is how to provide precision- provide ventilation for airflow that can be pressurized, flow-
inspired oxygen fraction (FiO2) to the patients during the regulated, volume-controlled and can be enriched with
treatment. Generally, the low-flow nasal cannula can provide 4- oxygen gas needed for COVID-19 patients who have
6 liters per minute of supplemental oxygen with approximately difficulty breathing. With the help of using a ventilator, it is
37 - 45% FiO2. In this paper, we present high-flow and high- hoped that COVID-19 patients will maintain their breath
pressure oxygen mixing to provide up to 100% FiO2 and 60
until the patient's health recovers.
liters per minute. The proposed system also employs the close-
loop feedback control to achieve a desirable FiO2 set point. It There are various types of ventilators that have different
can be used for a Continuous Positive Airway Pressure (CPAP) functions. Some ventilators are purely mechanical, some
and Bilevel Positive Airway Pressure (BiPAP) ventilator ventilators use electronics, and some are somewhere in
systems with oxygen as well as a High Flow Nasal Cannula between. In operation, some ventilators still use mechanical
(HFNC) system for acute respiratory patient’s syndrome. switch interfaces, and some use digital as a whole, some are
a combination of the two. In terms of appearance, some
Keywords—C-BiPAP ventilator, FiO2, high flow, high
ventilators have a very minimal appearance, so only limited
pressure, oxygenation.
parameters are displayed. There are modern types of
I. INTRODUCTION ventilators that have used a complete screen so that various
controlled parameters can be displayed, and graphs are added
Overcoming the COVID-19 pandemic caused by the to determine the patient's vital signs.
severe acute respiratory syndrome coronavirus 2 (SARS-
CoV-2) virus has become one of the main priorities. The Adequate and precision oxygenation are essential factors
virus attacks human respiration and can cause the emergence to support patient recovery. Excessive inspired oxygen
of other comorbidities that can lead to death. The researcher fraction leads to the intoxication of the patient. Hence, the
is still working on the medicine on patients infected with the mixin g process needs to be precisely maintained. Several
virus. Some new antiviral drugs have been deemed effective methods have been reported. Automated Close Loop Control
in clinical trials study. However, regulators are moving has been implemented in several works employing several
swiftly to consider the general use of the drugs. It is needed algorithms such as PID, Fuzzy, and neuro-Fuzzy, etc [1-4].
to complete medical studies to determine whether any of
In the High-Flow Nasal Cannula (HFNC) therapy, stable
these medications are effective against COVID-19. One of
and precise FiO2 is needed for the treatment such as infant
the medical devices used for COVID-19 sufferers is a
and neonatal patients [5, 6] as well as COVID-19 patients [7,
ventilator that helps pump oxygen into the lung to support
8]. The exactly inspired oxygen fraction is also substantial to
the breathing system. Breathing problems can become severe
be accurately determined such in the Non-Invasive Positive
enough to require treatment for the patient COVID-19.
Pressure Ventilation (NIPPV) and Continuous Positive

© IEEE 2021. This article is free to access and download, along with rights for full text and data mining, re-use and analysis.
201

horized licensed use limited to: NUST School of Electrical Engineering and Computer Science (SEECS). Downloaded on September 13,2022 at 04:23:37 UTC from IEEE Xplore. Restrictions ap
Pressure (CPAP) for obstructive sleep apnea syndrome A filter and oxygen sensors are also installed to filtrate
patients (OSAS), with acute respiratory failure breathing [9, and monitor the oxygen flow. An oxygen flow sensor is used
10]. This work reported the method to provide inspired to provide the oxygen feedback as the input to the closed-
oxygen fraction applicable for high-flow and high-pressure loop feedback control system to achieve a desirable oxygen
oxygen air mixing, which is suitable not only for Continuous- flow. The second source, ambient air from free air, is filtered
Bilevel Positive Airway Pressure (C-BiPAP) but also High- and blown by a motor-controlled blower and air filter. A
Flow Nasal Cannula (HFNC) therapy. C-BiPAP ventilator brushless centrifugal DC motor blower (OLOEY WM7040-
can deliver a single pressure as CPAP (Continuous Positive 24) is used to supply adequate ambient airflow and pressure,
Airway Pressure) mode or deliver two pressures: an inhale provides high pressure (up to 7.5 KPa) and high airflow (up
pressure and an exhale pressure as BiPAP (Bilevel or two- to 240L/min) to deliver airflow and pressure into the oxygen
level Positive Airway Pressure) mode. These two pressures baffle.
are known as inhalation positive airway pressure (IPAP), and
exhalation positive airway pressure (EPAP). The In-Line Flow Restrictor (ILFR) valve is used to
control the airflow to the mixer (oxygen baffle). Then, the
II. METHODS oxygen and air are mixed and circulated in the oxygen baffle
to provide precision-inspired oxygen fraction (FiO2) to the
Fig. 1 shows a block diagram of the proposed ventilator’s patients during the treatment.
pneumatic system which consists of several functional
modules. The ventilator system was developed by the After that, the Pressure Release Valve (PRV) valve is
Research Center for Electronics and Telecommunication, controlled to provide the desired high-flow and high-pressure
National Research and Innovation Agency of Indonesia, oxygen mixing. The outlet of this system is added air filter
named Smart & Innovated Ventilator Indonesia (Si- before the exhaled air from the patient is thrown into the free
Venesia). air. An oxygen fraction flow and pressure sensor are used to
provide the feedback as the input to the closed-loop feedback
The gas source consists of two inputs, which are input control system to achieve a desirable oxygen fraction flow
source from an oxygen inlet and an ambient air inlet. The two and pressure.
sources are used to provide an inspired oxygen fraction in the
mixing module before delivering to the patient. In the first There is a minimum of two important parameters for the
source, oxygen gas will be regulated by a gas regulator to outputs in ventilator, that is oxygen fraction and pressure. To
maintain the pressure around 3-5 bar. maintain those two parameters, we set the desired set point
using databases. The database consists of the data of oxygen
Then, a proportional valve is used to control the oxygen flow, ambient air flow, and opening/closing valve. while the
flow to the mixer. The module proportional valve was FIO2 sensor is used to monitor the level of the oxygen
developed by us and is responsible for providing the required fraction produced. All the modules are controlled by a
oxygen flow. The DC motor controlled proportional valve microprocessor in the data and signal processing unit.
based on a self-locking gearbox (Indonesian Patent
Application P00202105266) maintains the oxygen flow for Table I shows specific medical parameters of the
controlled inspired oxygen fraction production. High rpm developed Si-Venesia ventilator. Where, this is a standard
DC motor and high ratio gearbox used (JGY370 - 3RPM), parameter that must be met on a ventilator.
lead to the high-resolution flow oxygen as 0.1 L/min can be
achieved.

Fig. 1. System block diagram ventilator.

© IEEE 2021. This article is free to access and download, along with rights for full text and data mining, re-use and analysis.
202

horized licensed use limited to: NUST School of Electrical Engineering and Computer Science (SEECS). Downloaded on September 13,2022 at 04:23:37 UTC from IEEE Xplore. Restrictions ap
TABLE I. MEDICAL PARAMETERS OF THE DEVELOPED SI-VENESIA

Parameter Value
140

Positive end-expiratory pressure (PEEP) 5 – 10 cmH2O 120


Pressure support (PS) 1 – 35 cmH2O
100

Airflow (lpm)
Inspiratory positive airway pressure (IPAP) 5 – 40 cmH2O
Expiratory positive airway pressure (EPAP) 5 – 20 cmH2O
80
Inspiratory: Expiratory ratio (I:E) 1:1; 1:2 ; 1:3
Oxygen fraction (FiO2) 21 – 100% 60

40
Experimental Data
Fitting Curve
20
for Calibration

0
0 20 40 60 80 100 120 140 160 180
PWM Value of Motor Blower

Fig. 3. Measured air flow.

100

Oxygen Fraction (%) 90

80

70
Fig. 2. Test setup.
60
III. EXPERIMENTAL SETTING 50
Fig. 2 shows the setup for testing and measurement in this
40
study. The ventilator needs an oxygen source from a portable
oxygen tank attached to the oxygen inlet. Meanwhile, the air 30 Experimental Data
inlet is used to fetch ambient air from the environment. The Regression Fitting
oxygen inlet can receive the oxygen pressure from 3 – 5 bar. 20 for Calibration

The output-inspired oxygen fraction is attached to the Fluke 10


biomedical VT650 gas flow analyzer. The personal computer 0 10 20 30 40 50 60
(PC) is used to control and process the data.
Output Voltage (mV)
The inspired oxygen fraction is produced by mixing the
oxygen flow with the ambient air inside the mixer using an Fig. 4. Measured inspired oxygen fraction (FiO2).
oxygen baffle (BiPAP Vision PH13606). The amount of the
oxygen flow and ambient airflow automatically using IV. DISCUSSION
controlled closed-loop feedback control to achieve a Fig. 3 shows the experiment results of measured airflow
desirable FiO2. The module proportional valve was from ambient. The flow of the air was controlled by changing
developed by ourselves and is responsible for providing the the speed of the motor blower. A microcontroller is used to
required oxygen flow, meanwhile, a brushless centrifugal control the speed using Pulse Width Modulation (PWM)
DC motor blower (OLOEY WM7040-24) is used to supply from 0 - 255. As shown in Figure 3, we can find that the
adequate ambient airflow and pressure. relation of PWM and airflow is quite linear. Using those data,
The DC motor controlled proportional valve based on a we can control the output flow of FiO2 precisely by mixing
self-locking gearbox (Indonesian Patent Application the airflow from ambient combined with from oxygen
P00202105266) maintains the oxygen flow for controlled source.
inspired oxygen fraction production. The closed-loop The inspired oxygen fraction detection use FiO2 flow and
feedback control applied in the system adjusts the DC motor pressure sensor (SIARGO FS6122). The sensor can measure
speed to open and close the valve. High rpm DC motor and a uni- or bi-directional flow and pressure from inspiration
high ratio gearbox used (JGY370 - 3RPM) lead to the high- and expiration processes. A pressure up to 250 SLPM and
resolution flow oxygen as 0.1 L/min can be achieved. pressure up to 40 gauge can be provided. The accuracy of
The three-phase brushless DC motor blower (OLOEY pressure reading is 1%. The data and control using I2C
WM7040-24) provides high pressure (up to 7.5KPa) and connection with the output is linear DC voltage. Fig. 4 shows
high airflow (up to 240L/min) to deliver airflow and pressure
into the oxygen baffle.

© IEEE 2021. This article is free to access and download, along with rights for full text and data mining, re-use and analysis.
203

horized licensed use limited to: NUST School of Electrical Engineering and Computer Science (SEECS). Downloaded on September 13,2022 at 04:23:37 UTC from IEEE Xplore. Restrictions ap
160 160 V. CONCLUSION
Flow Output (FiO2)

140 Flow O2
140 High-flow and high-pressure inspired oxygen fractions in
Air Flow
Output Pressure
the Si-Venesia ventilator have been described. The mixing
120 120 process relied on the amount of oxygen and ambient airflow

Pressure (cmH2O)
to achieve a determined oxygen fraction. The inspired
Flow (lpm)

100 100
oxygen fraction (FiO2) can be achieved from 20 to 100% (0.2
80 80 to 1). The flow FiO2 as high as 60 to 120 lpm can be reached,
Min output flow spec instantaneously mixed from Oxygen flow from 0 up to 80
60 60
lpm and ambient airflow from 65 until 120 lpm. The output
40 40 pressure can be provided from 20 ~ 40 cmH2O (0.28 ~ 0.57
psi). These results are suitable for obstructive sleep apnea
20 Min output pressure spec 20
syndrome patients using Continuous Positive Airway
0 0 Pressure (CPAP) treatment with oxygenation feature as well
20 30 40 50 60 70 80 90 100 as High-Flow Nasal Cannula (HFNC) therapy.
Oxygen Fraction (%) ACKNOWLEDGMENT
Fig. 5. Measured flow and pressure This research is supported partly by funding from
Konsorsium Riset dan Inovasi untuk percepatan Penanganan
the sensor’s output voltage corresponds to the oxygen
fraction (FiO2). The graph shows that the output sensor and Corona Disease (Covid-19) Lembaga Pengelola Dana
the oxygen fraction have a good correlation. The sensor Pendidikan (LPDP) 2020-2021 also partly supported by
works that FiO2 from 20 ~ 100% can be detected, which is annual grant DIPA LIPI 2020.
suitable for monitoring bidirectional flow and pressure Authors thanks to Mr. Hery Soepriyadi for his kind
oxygen fraction. supports and helps during the development and realization of
the Si-Venesia ventilator.
Fig. 5 describes the measured flow of the Oxygen (blue
dot line), ambient air (pink dot line), inspired oxygen fraction REFERENCES
(red dot line), and pressure (black start line), which [1] C. Nelson and B. Eduardo, “Automated Closed Loop Control of
correspond to the inspired oxygen fraction (FiO2). Minimum Inspired Oxygen Concentration,” in Respiratory Care, Jan. 2013, vol
output flow specification is determined at 60 lpm (red dash 58 no 1, pp. 151-161.
line), and minimum output pressure is 20 cmH2O. Based on [2] Dani C. “Automated control of inspired oxygen (FiO2) in preterm
the requirement, the Continuous Positive Airway Pressure infants: Literature review”, in Pediatric Pulmonology. 2019, pp. 1–6,
(CPAP) therapy need up to 40 cmH2O flow FiO2, meanwhile doi: https://doi.org/10.1002/ppul.24238.
above 50 cmH2O flow FIO2 can be used for High-Flow Nasal [3] Chanyagorn et al., “Automatic Control of Fraction of Inspired Oxygen
Cannula (HFNC) therapy as well. in Neonatal Oxygen Therapy using Fuzzy Logic Control”, in IEIE
Transactions on Smart Processing and Computing, Apr. 2016, Vol. 5,
Both flow oxygen and airflow are used to produce the No. 2, doi: http://dx.doi.org/10.5573/IEIESPC.2016.5.2.107.
oxygen fraction flow. Here we simplified the formula by: [4] Hanifah Rahmi Fajrin et al. “Mixed Pressure Control Innovation of
Oxygen and Air Pressure” in Ventilator with Safety Valve, J. Phys.:
( )= ( _ )+ ( ) (1) Conf. Series. 2021, Vol 1933 (2021) 012106.
[5] C. Yusuke, O. Saki, and N. Masaji, “Humidity and Inspired Oxygen
with the assumption of no leakage in the system. There is no Concentration During High-Flow Nasal Cannula Therapy in Neonatal
correlation in the amount of both flows with the output flow and Infant Lung Models”, in Respiratory Care, May 2017, Vol. 62,
FiO2. To reach, above 100 lpm flow FiO2, the required flow No. 5, pp. 532-537.
oxygen and ambient airs are 20 ~ 60 lpm and 115 ~ 85 lpm [6] Mayfield, S., Jauncey-Cooke, J., Hough, J. L., Schibler, A., Gibbons.
respectively. This can produce an oxygen fraction of 20 ~ 60 “High-flow nasal cannula therapy for respiratory support in children”,
% and pressure remain flat at 20 cmH2O. However, below in Cochrane Database of Systematic Reviews. March 2014, Issue 3,
doi: https://doi.org/10.1002/14651858.
100 lpm flow output (oxygen fraction), increasing flow
oxygen is needed (40 ~ 80 lpm) with decreasing flow [7] Abhimanyu Chandel et all, “High-Flow Nasal Cannula Therapy in
COVID-19: Using the ROX Index to Predict Success”, in Respiratory
ambient air 120 ~ 0 lpm, this led to the oxygen fraction from Care, 66 (6) 909-919, Jun 2021, doi:10.4187/respcare.08631.
60 up to 100% and pressure stable at 40 cmH2O. [8] Vianello A, Arcaro G, Molena B, et al. “High-flow nasal cannula
Due to varying input flow oxygen and ambient air to oxygen therapy to treat patients with hypoxemic acute respiratory
failure consequent to SARS-CoV-2 infection”, in Thorax, 2020, pp.
produce the output flow, it is difficult to use manual 998-1000.
proportional mixing. Here, the mixing process based on
[9] D. Samolskia et al. “Inspired oxygen fraction achieved with a portable
calibrated values is used. These values are measured and ventilator: Determinant factors”, in Respiratory Medicine, Dec 2005,
calibrated on the normal operation ventilator using the Fluke Vol. (2006) 100, pp. 1608–1613.
VT650 gas flow meter. All calibrated possible values were [10] Coudroy R, Frat J, Girault C, et al. “Reliability of methods to estimate
saved in the memory of the data and signal processing unit. the fraction of inspired oxygen in patients with acute respiratory failure
Whenever the required flow, pressure, oxygen fraction is breathing through non-rebreather reservoir bag oxygen mask”, in
needed, the system generates the suitable value on the system Thorax, 2020, Vol. 75, pp. 805-807.
to achieve that. With this method, all possible target
parameters output can be easily reached.