DOCUMENT APPARATUS DATE

Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

TECHNICAL NOTE N° 163

INDEX

1.0 74X0 New LCD Display 2


1.1 Replacing the display when a new one is already installed 3
1.1.1 Procedure: 3
1.2 Replacing a previous display model with the new one 7
1.2.1 Procedure: 8
1.2.2 New monitor setup 18
1.3 Traceability 21

1
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

1.0 74X0 New LCD Display


The LCD display P/N 3900000038 has been replaced by the new one P/N 450000022 for
obsolescence reason.
It is not compatible, as it requires:

1. PC board model M886 (that only works with BACKEND board CI22)
2. SW Release 5.51 or higher

Point n. 1 is true only for MyLab Alpha, as MyLab Gamma started its life with PC board
model 886..

New display Old display

As you may see from the picture, the two displays are a little bit different.
The old display has two different resolutions 1280x768 or 1280x800, instead the new
monitor has a resolution of 1366x768.

2
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

1.1 Replacing the display when a new one is already installed

Follows this procedure only in case MyLab has already mounted the new display P/N
450000022.
In this case what you need is just a new display:

• n. 1 LG LCD Display – P/N 450000022

1.1.1 Procedure:
Please take care in opening the covers without damaging them.
Remove the rubber pins at the sides of the monitor support and unscrew the two screws
below. Open the cover rising up the front cover (pls insert a thin slotted screwdriver where
indicated by the screws, or insert the finger in the internal part of the frame and rise up).

Please keep the rubber pins.


Remove the old LCD display,

Remove the 4 screws around the LCD screen and rotate it, in order to access to the rear part.

3
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

Disconnect the cable from the LCD connector


placed on the small board assembled on the
LCD removing the kapton tape on the LCD
side

Connect the new display and protect the connector applying again some kapton tape.

Warning!
The display must not be touched by fingers (or other) in the areas circled in red
Touching it, it’s possible to damage the screen.

4
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

Rotate the display and place it on


the rear cover.
Partially screw the 4 screws
The bending of cable must M2x4 (pos. 24) + spring washers
be out of display to prevent (pos. 25) where indicated with red
image problems.

Push it light as shown by arrows


direction and tighten the screws
with torque of 0,2Nm (+0;-0,1)

Remove the protection foil

5
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

Place the LCD front panel

Gently press to close

Close the monitor applying the frontal screws and close with the covers.

6
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

1.2 Replacing a previous display model with the new one

In case MyLab mounts an old display model and you need to replace it with the new display
P/N 450000022 you need additional parts:
• n. 1 LCD Spare kit covers – P/N 330001666 (that includes the following two parts)
o n.1 LCD FRONT PANEL for LG Display - P/N 330001029
o n.1 REAR LCD COVER ASSY - P/N 330001667
• n. 1 LG LCD Display – P/N 450000022
• Kapton
• Hot melt glue
The Rear LCD cover assy includes the rear cover with release buttons and other parts.
In particular, as the new LCD Display has eDP protocol, but the unit provides LVDS output,
an LVDS to eDP Converter is necessary (P/N 250007300), that is stuck on the rear cover
and it has its specific connection cable towards the LCD:

The Rear LCD cover assy also includes the antennas, while their cables will have to be
unstuck from the old covers and re-stuck on the new ones.

7
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

In case your MyLab Alpha has not the PC board IC 01 model M886, the operation is
not possible on field and it’s necessary to return the unit to Esaote.

Warning!
The display must not be touched by fingers (or other) in the areas circled in red

1.2.1 Procedure:
Remove the old LCD display, and keep the rubber pins.
Please take care while unsticking the antennas cables and also unstick the magnet that is
stuck on the front panel, that will be re-stuck on the new front panel.

Fit the cable between the


dedicated fins

8
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

Apply on the antenna cables (in


correspondance of four fins) some hot
melt glue.

Connect the flat side of the cable to


the LVDS to eDP converter. Be
careful to the right side of the
connector!!

Flat Side

9
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

Protect the connector with a


piece of Kapton

Adjust the cable pressing it by


finger to have the internal cables
flat and as thin as possible. Apply
some pieces of kapton.

WARNING! The display must not be touched by fingers in some areas (see picture above)

10
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

Connect the cable with LCD


connector and apply the kapton
tape on both connectors

Rotate the display and place it


on the rear cover. Partially
screw the 4 screws M2x4 +
spring washers where
The bending of cable indicated with red
must be out of display to
prevent image problems.

11
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

Push it light as shown by


arrows direction and tighten the
screws with torque of 0,2 Nm
(+0;-0.1)

Connect the ultraminiature coax


connectors

12
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

The cable passes between


the columns

The cable passes


between the columns Assembly the 7400 LCD STRENGHT
RACKETs and tighten the screws with a
torque of (+0; -0.1)

13
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

Hot glue

Fold the cable as showed

Put on the cover


Be careful not to pinch the cables

14
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

Put the screws and tighten them


with torque 0,2 Nm (+0,-0,1)
Be careful the screws don’t
damage the cables

Put on the MAGNET with the “polarity


spot” up as shown

polarity

15
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

Secure the magnet with one or


two drop of hot melt glue

Remove the protection foil

Place the LCD Front panel

16
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

Gently press to close

The monitor must stay correctly


closed

Put the screws and tighten them with


torque 0,2Nm (+0,-0.1)

17
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

Put on the rubbers and gently press


till complete insert

1.2.2 New monitor setup


Once installed the new monitor, pls switch on the unit with the service key installed; you
will get the following message:

18
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

Press the MENU – Service keys

19
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

Look at the monitor setting window and select the model

20
DOCUMENT APPARATUS DATE
Technical Note 163 74X0 25/03/2016

To be applied to: MyLabAlpha MyLabGamma

1.3 Traceability

The S/N of the first MyLabGamma with the new display is 750.
The MyLabAlpha is being manufactured with both the kinds of diplays at the moment, as
some stock of the old kind is still available. The S/N of the first MyLabAlpha after the end
of the stock will be communicated when available.