EN - IMMUNOGLOBULIN A (Setting Sheet)

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

e e

IgA, AU400/AU400 /AU640/AU640 Application IgA, AU600 Application


System Reagent: OSR6x171 Reagent ID: 171 System Reagent: OSR6x171 Reagent ID: 171

Specific Test Parameters Specific Test Parameters


General LIH ISE Range
Test No # Test name IgA ∇ Sample type Ser ∇ Page 1/2
Test Name: IgA ∇ < > Type: Serum ∇ Operation: Yes ∇
Sample vol. 2 Dil. Vol 0 μL Min. OD Max. OD
Sample: Volume 2 μL Dilution 0 μL Pre-Dilution Rate: 1 Reagent 1 vol 70 Dil. Vol 115 μL L H
Reagents: R1 Volume 70 μL Dilution 115 μL Min OD Max OD Reagent 2 vol 55 Dil. Vol 10 μL Reagent OD limit
Fst. L -0.1 Fst. H 1.5
R2 Volume 55 μL Dilution 10 μL L H
Wave Main 600 Sub 800 ∇ Lst. L -0.1 Lst. H 1.5
Reagent OD limit: Method END ∇ Dynamic range
Wavelength: Pri. 600 ∇ Sec. 800 ∇ First L -0.1 First H 1.5 Reaction + ∇ L 10* H 700*
Method: END ∇ Last L -0.1 Last H 1.5 Point 1 Fst 0 Lst 27 Correlation factor A 1
Reaction slope: + ∇ Dynamic Range: Point 2 Fst 0 Lst 10 B 0
Measuring Point 1: First 0 Last 27 L 10* H 700* Sample Pre-dil. Rate ¤
Measuring Point 2: First 0 Last 10 Correlation Factor: Linearity Fst % Sec %
Linearity : % A 1 B 0 No lag time ∇ On-board stability period 90
No Lag Time: ∇ On-board stability period: 90 Select using Space key, or select from list displayed by Guide key
Calibration Specific

Test No # Test name IgA ∇


Calibration Specific
General ISE Cal type 13 5AB ∇ Count #
Formula 3 POLYGONAL ∇ Process Conc ∇
Test Name: IgA ∇ < > Type Serum ∇ Selection calibrator
Cal No OD Conc Factor/OD-L Factor/OD-H
Calibration Type: 5AB ∇ Formula: POLYGONAL ∇ Counts: # Process: CONC ∇ Point 1 # ∇ † -0.1 2.5
Point 2 # ∇ † -0.1 2.5
Cal. No. OD CONC Factor/OD-L Factor/OD-H Point 3 # ∇ † -0.1 2.5
Point 1: # † -0.1 2.5 Point 4 # ∇ † -0.1 2.5
Point 2: # † -0.1 2.5 Point 5 # ∇ † -0.1 2.5
Point 3: # † -0.1 2.5 Point 6 ∇
Point 4: # † -0.1 2.5 Point 7 ∇
1-point cal. Point
Point 5: # † -0.1 2.5 MB type factor
Point 6: Calibrator stability period 90
Point 7:
Select the function using the Function key or the Mouse
1-Point Cal. Point: ο with CONC-0 Slope Check: + ∇ Advanced Calibration: # ∇
MB Type Factor: Calibration Stability Period: 90 Data Check Parameters

Test No # Name IgA ∇ Type Ser ∇


Data Check Parameters
Logic check-1 YES Logic Check-2 NO
Check Point-1 11 Check Point-1 0
Test Name: IgA ∇ < > Type: Serum ∇
Check Point-2 17 Check Point Interval
Check Point-3 27 Decision value-1 0.0000
√ Logic Check-1 Logic Check -2 Logic Check-3 Decision value-1 0.0000 Decision value-2 0.0000
Check Point-1: 11 Check Point-1: Check Point-1: Decision value-2 0.7700 Limit Point 1 0
Check Point-2: 17 Check Point Interval: Check Point Interval: Decision value-3 0.0150 Limit Point 2 0
Check Point-3: 27 Decision Value-1: Decision Value-1: Limit Point 1 16
Decision Value-1: 0.0000 Decision Value-2: Decision Value-2: Limit Point 2 27 Logic Check-3 NO
Decision Value-2: 0.7700 Limit Point 1: Limit Point 1: Check Point-1 0
Decision Value-3: 0.0150 Limit Point 2: Limit Point 2: Check Point Interval
Limit Point 1: 16 Decision value-1 0.0000
Limit Point 2: 27 Decision value-2 0.0000
Limit Point 1 0
Limit Point 2 0

# User defined # User defined ¤ Analyser default value


† Serum Protein Multicalibrator Cat. No.: ODR3021 † Serum Protein Multicalibrator Cat. No.: ODR3021
* Values set for working in mg/dL. To work in SI Units (g/L) multiply by 0.01. * Values set for working in mg/dL. To work in SI Units (g/L) multiply by 0.01.

BSSA6x171.01
2009-08
IgA, AU2700/AU5400 Application IgA, AU680 Application
System Reagent: OSR6x171 Reagent ID: 171 System Reagent: OSR6x171 Reagent ID: 171
Parameters Specific Test Parameters
Specific Test Parameters General LIH ISE HbA1c Calculated Test Range
General LIH ISE Range
Test Name: IgA ∇ < > Type: Serum ∇ Operation Yes ∇
Test Name: IgA ∇ < > Type: Serum ∇ Operation: Yes ∇

Sample: Volume 1.6 μL Dilution 0 μL Pre-Dilution Rate: 1 Sample Volume 1.6 μL Dilution 0 μL OD Limit
Pre-Dilution Rate 1 ∇ Min.OD Max.OD
Reagents: R1 Volume 56 μL Dilution 92 μL Min OD Max OD
Rgt. Volume R1(R1-1) 56 μL Dilution 92 μL Reagent OD Limit
R2 Volume 44 μL Dilution 10 μL L H
First Low -0.1 High 1.5
Reagent OD limit:
Last Low -0.1 High 1.5
Wavelength: Pri. 600 ∇ Sec. 800 ∇ First L -0.1 First H 1.5
R2(R2-1) 44 μL Dilution 10 μL
Method: END ∇ Last L -0.1 Last H 1.5
Dynamic Range Low 10* High 700*
Reaction slope: + ∇ Dynamic Range: Common Rgt. Type None Name Correlation Factor A 1 B 0
Measuring Point 1: First 0 Last 27 L 10* H 700* Wavelength Pri 600 ∇nm Sec. 800 ∇nm Factor for Maker A 1 B 0
Measuring Point 2: First 0 Last 10 Correlation Factor: Method END ∇
Linearity : % A 1 B 0 Reaction Slope + ∇ Onboard Stability Period 90 Day 0 Hour
No Lag Time: ∇ On-board stability period: 90 Measuring Point1 First 0 Last 27 LIH Influence Check # ∇
Measuring Point2 First 0 Last 10 Lipemia +++++ ∇
Linearity Limit % Icterus +++++ ∇
Lag Time Check ∇ Hemolysis +++++ ∇
Calibration Specific
General ISE Parameters Calibration Parameters
Calibrators Calibration Specific STAT Table Calibration
Test Name: IgA ∇ < > Type: Serum ∇ General ISE

Calibration Type: 5AB ∇ Formula: POLYGONAL ∇ Counts: # Process: CONC ∇ Test Name: IgA ∇ < > Type Serum ∇ ο Use Serum Cal.

Calibration Type: 5AB ∇ Formula: POLYGONAL ∇ Counts: # ∇


Cal. No. OD CONC Factor/OD-L Factor/OD-H
<Calibrator Parameters> OD Range
Point 1: # † -0.1 2.5
Calibrator OD Conc Low High Slope Check + ∇
Point 2: # † -0.1 2.5
Point 1: # ∇ † -0.1 2.5
Point 3: # † -0.1 2.5
Point 2: # ∇ † -0.1 2.5 Allowance Range Check
Point 4: # † -0.1 2.5
Point 3: # ∇ † -0.1 2.5
Point 5: # † -0.1 2.5 Point 4: # ∇ † -0.1 2.5 ο Reagent Blank
Point 6: Point 5: # ∇ † -0.1 2.5 ο Calibration
Point 7: Point 6:
1-Point Cal. Point: ο with CONC-0 Slope Check: + ∇ Advanced Calibration: # ∇ Point 7 ∇ Advanced Calibration
MB Type Factor: Calibration Stability Period: 90 Point 8 ∇ Operation Yes ∇
Point 9 ∇
Point 10 ∇ Interval (RB/ACAL) Lot/Lot ∇
Data Check Parameters <Point Cal. For No. of Correction Points ∇ Use Master Curve ∇ ο Lot Calibration
Master Curve> OD Range
Calibrator OD Conc Low High Stability
Test Name: IgA ∇ < > Type: Serum ∇ Point-1 ∇ Reagent Blank 90 Day 0 Hour
Point-2 ∇ Calibration 90 Day 0 Hour
MB Type Factor: 1-Point Calibration Point None ∇ ο with Conc-0
√ Logic Check-1 Logic Check -2 Logic Check-3
Check Point-1: 11 Check Point-1: Check Point-1:
Parameters Misc
Check Point-2: 17 Check Point Interval: Check Point Interval:
Check Point-3: 27 Decision Value-1: Decision Value-1:
CheckedTests Contamination Data Check
Decision Value-1: 0.0000 Decision Value-2: Decision Value-2:
Parameters Parameters
Decision Value-2: 0.8000 Limit Point 1: Limit Point 1:
Decision Value-3: 0.0230 Limit Point 2: Limit Point 2: Test Name: IgA ∇ < > Type: Serum ∇
Limit Point 1: 16
Limit Point 2: 27 √ Logic Check-1 Logic Check -2 Logic Check-3
Check Point-1: 11 Check Point-1: Check Point-1:
Check Point-2: 17 Check Point Interval: Check Point Interval:
Check Point-3: 27
Decision Value-1: 0.0000 Decision Value-1: Decision Value-1:
# User defined Decision Value-2: 0.7700 Decision Value-2: Decision Value 2:
† Serum Protein Multicalibrator Cat. No.: ODR3021 Decision Value-3: 0.0160
* Values set for working in mg/dL. To work in SI Units (g/L) multiply by 0.01.
Limit Point 1: 16 Limit Point 1: Limit Point 1:
Limit Point 2: 27 Limit Point 2: Limit Point 2:
Check Pattern: Pattern1 ∇

BSSA6x171.01
2009-08
IgA, AU480 Application
System Reagent: OSR6x171 Reagent ID: 171
Parameters Specific Test Parameters
General LIH ISE HbA1c Calculated Test Range

Test Name: IgA ∇ < > Type: Serum ∇ Operation Yes ∇

Sample Volume 1.6 μL Dilution 0 μL OD Limit


Pre-Dilution Rate 1 ∇ Min.OD Max.OD
Rgt. Volume R1(R1-1) 56 μL Dilution 92 μL Reagent OD Limit
First Low -0.1 High 1.5
Last Low -0.1 High 1.5
R2(R2-1) 44 μL Dilution 10 μL
Dynamic Range Low 10* High 700*
Correlation Factor A 1 B 0
Wavelength Pri 600 ∇nm Sec. 800 ∇nm Factor for Maker A 1 B 0
Method END ∇
Reaction Slope + ∇ Onboard Stability Period 90 Day 0 Hour
Measuring Point1 First 0 Last 27 LIH Influence Check # ∇
Measuring Point2 First 0 Last 10 Lipemia +++++ ∇
Linearity Limit % Icterus +++++ ∇
Lag Time Check ∇ Hemolysis +++++ ∇

Parameters Calibration Parameters


Calibrators Calibration Specific STAT Table Calibration
General ISE

Test Name: IgA ∇ < > Type Serum ∇ ο Use Serum Cal.

Calibration Type: 5AB ∇ Formula: POLYGONAL ∇ Counts: # ∇


<Calibrator Parameters> OD Range
Calibrator OD Conc Low High Slope Check + ∇
Point 1: # ∇ † -0.1 2.5
Point 2: # ∇ † -0.1 2.5 Allowance Range Check
Point 3: # ∇ † -0.1 2.5
Point 4: # ∇ † -0.1 2.5 ο Reagent Blank
Point 5: # ∇ † -0.1 2.5 ο Calibration
Point 6:
Point 7 ∇ Advanced Calibration
Point 8 ∇ Operation Yes ∇
Point 9 ∇
Point 10 ∇ Interval (RB/ACAL) Lot/Lot ∇
<Point Cal. For No. of Correction Points ∇ Use Master Curve ∇ ο Lot Calibration
Master Curve> OD Range
Calibrator OD Conc Low High Stability
Point-1 ∇ Reagent Blank 90 Day 0 Hour
Point-2 ∇ Calibration 90 Day 0 Hour
MB Type Factor: 1-Point Calibration Point None ∇ ο with Conc-0

Parameters Misc

CheckedTests Contamination Data Check


Parameters Parameters
Test Name: IgA ∇ < > Type: Serum ∇

√ Logic Check-1 Logic Check -2 Logic Check-3


Check Point-1: 11 Check Point-1: Check Point-1:
Check Point-2: 17 Check Point Interval: Check Point Interval:
Check Point-3: 27
Decision Value-1: 0.0000 Decision Value-1: Decision Value-1:
Decision Value-2: 0.7700 Decision Value-2: Decision Value 2:
Decision Value-3: 0.0150
Limit Point 1: 16 Limit Point 1: Limit Point 1:
Limit Point 2: 27 Limit Point 2: Limit Point 2: # User defined
Check Pattern: Pattern1 ∇ † Serum Protein Multicalibrator Cat. No.: ODR3021
* Values set for working in mg/dL. To work in SI Units (g/L) multiply by 0.01.