Usp 35 en

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

Accessed from 128.83.63.

20 by nEwp0rt1 on Sat Dec 03 00:44:32 EST 2011

714 〈1118〉 Monitoring Devices / General Information USP 35

at the end of a chain, and the evaporation of water from ment responsiveness may be defined as the time, t / , re-
1
2

the wick cools the wet bulb thermometer. The temperature quired for a device to read a value of (x + y)/2 after an
difference between the wet and dry thermometers is then instantaneous change in the property being measured from
compared to a table, specific to that psychrometer, based x to y. Measurement responsiveness is typically defined for
on dry bulb temperature, and the relative humidity is deter- the operating range of a device.
mined. The use of a sling psychrometer in a commercial Different levels of responsiveness are needed for different
setting is impractical. monitoring applications. For devices used to monitor stor-
Hair Hygrometer—This type of device is based on the age locations, where the temperature and humidity are un-
fact that the length of a synthetic or human hair increases likely to change rapidly, a t / ≤ 15 minutes may be appropri-
1
2

as a function of the relative humidity. This change is used to ate. For devices used to monitor transport, where more
move an indicator or affect a strain gauge. A hair hygrome- rapid changes are possible, a t / ≤ 5 minutes may be
1
2

ter can be accurate to ±3%, but it is unable to respond to needed.


rapid changes in humidity and loses accuracy at very high Time Accuracy—Most commonly, time accuracy is ex-
or very low levels of relative humidity. pressed as a ± percentage of total duration of the recording
Infrared Hygrometer—This type of hygrometer deter- period. For pharmaceutical applications, a ±0.5% time accu-
mines relative humidity by comparing the absorption of two racy is adequate.
different wavelengths of IR radiation through air. One wave- Validation of Chemical-Based TTIs—This type of device
length is absorbed by water vapor and the other is not. This presents a problem for validation because testing the indi-
type of hygrometer can accurately measure relative humid- vidual device causes its destruction. For this reason, calibra-
ity in large or small volumes of air. It is sensitive to rapid tion of individual chemical-based TTIs against an NIST trace-
changes of humidity and can be integrated with an elec- able standard is not possible. Ideally, chemical-based TTIs
tronic data handling system. would be made using Good Manufacturing Practices, and
Dew Point Hygrometer—This type of device uses a their use in connection with monitoring the storage and
chilled mirror to determine the dew point of an air sample. transport environment of drugs would be appropriately reg-
The dew point is the temperature at which water vapor in ulated. In the absence of those conditions, the performance
the air begins to condense, that is, the temperature at of a batch of these devices may be assessed statistically by
which the relative humidity is 100%. From this measure- subjecting an appropriately sized sample to elevated tem-
ment and an accurate measurement of the ambient temper- perature conditions for a set period of time and observing
ature, the relative humidity can be calculated. The dew the results. Appropriate acceptance criteria should be
point hygrometer is the standard against which most com- adopted.
mercially available instruments are calibrated.
Capacitive Thin-Film Hygrometer—The principle of this THE USE OF HISTORIC TEMPERATURE DATA
type of hygrometer is that the dielectric of a nonconductive
polymer changes in direct proportion to the relative humid- It is clear that the type of temperature monitoring needed
ity. This change is measured as a change in capacitance. is a function of the environmental conditions that can be
This type of hygrometer is accurate to ±3%. expected. Therefore, climatic data are useful when selecting
Resistive Thin-Film Hygrometer—This type of hygrome- the most appropriate local storage conditions and monitor-
ter is similar to the capacitive thin-film type in that it uses ing methods. For example, an inexpensive limit detector
the effect of changing relative humidity on an electrical cir- may be all that is needed when there is a low probability
cuit. In the resistive thin-film hygrometer the sensor is an that excessive temperatures will be experienced. Alterna-
organic polymer whose electrical resistance changes in loga- tively, a data logger may be preferred when it would be
rithmic proportion to the relative humidity. This type of hy- useful to demonstrate that exposure to the highest tempera-
grometer is accurate to ±5%. tures was very brief.
It should be noted, however, that outside temperatures
are not necessarily reliable indicators of the temperatures
VALIDATION OF TEMPERATURE AND experienced by different items in the distribution chain. For
HUMIDITY MONITORING DEVICES example, recent studies reported significant departures from
ambient temperatures on summer days for mailboxes,
Thermometers and hygrometers, used to provide data trucks, and warehouses. Detailed historical temperature data
about the temperature and humidity exposure of a product, are available from the National Oceanic and Atmospheric
must be suitable for their intended use. Specifically, they Administration showing the daily mean maximum and mini-
must be appropriately validated. Validation is a process that mum temperature on any given day of the year in a geo-
assures the user of the monitoring device that the device graphical region of interest (e.g., http://www.cdc.noaa.gov/
has been tested prior to use either by the manufacturer or Usclimate/states.fast.html).
the user, to assess the measurement accuracy, measurement
responsiveness, and time accuracy, where appropriate.
Monitors used in manufacturing, storage, and transport of
drugs should be properly qualified by their users to ensure
that the monitors have been received and maintained in
proper working order. Pharmacies and consumers may ac-
cept the validation performed by the manufacturer of the
device.
〈1119〉 NEAR-INFRARED
Measurement Accuracy—For temperature and humidity
monitoring devices, measurement accuracy refers to the
SPECTROSCOPY
closeness of the value obtained with a particular device to
the true value being measured. In practice, this is deter-
mined by comparison with a device that has been calibrated
against a standard that is obtained from or traceable to the INTRODUCTION
National Institute of Standards and Technology (NIST).
Measurement Responsiveness—Any monitor takes time Near-infrared (NIR) spectroscopy is a branch of vibrational
to respond to a change in the temperature or humidity. The spectroscopy that shares many of the principles that apply
more rapid the response, the clearer the picture of the envi- to other spectroscopic measurements. The NIR spectral re-
ronmental history of a monitored product will be. Measure- gion comprises two subranges associated with detectors

Official from May 1, 2012


Copyright (c) 2011 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 128.83.63.20 by nEwp0rt1 on Sat Dec 03 00:44:32 EST 2011

USP 35 General Information / 〈1119〉 Near-Infrared Spectroscopy 715

used in the initial development of NIR instrumentation. The tion measurements in the NIR are made of scattering sam-
short-wavelength (Herschel or silicon region) extends from ples such as powders and slurries. For such materials NIR
approximately 780 to 1100 nm (12,821–9000 cm–1); and radiation can penetrate a substantial distance into the sam-
longer wavelengths, between 1100 and 2500 nm, compose ple, where it can be absorbed when the wavelength of the
the traditional (lead sulfide) NIR region. Applications of NIR radiation corresponds to a transition between the ground
spectroscopy use spectra displayed in either wavelength or vibrational state of the analyte and either a harmonic of a
wavenumber units. As is the case with other spectroscopy given vibrational mode (an overtone) or the sum of two or
measurements, interactions between NIR radiation and mat- more different modes (a combination band). Nonabsorbed
ter provide information that can be for both qualitative and radiation is scattered back from the sample to the detector.
quantitative assessment of the chemical composition of sam- NIR reflection spectra are accessed by calculating and plot-
ples. In addition, qualitative and quantitative characteriza- ting log(1/R) versus wavelength. This logarithmic form is the
tion of a sample’s physical properties can be made because pseudo-absorbance of the material and is commonly called
of the sample’s influence on NIR spectra. Measurements can absorbance.
be made directly on samples in situ in addition to applica-
tions during standard sampling and testing procedures.
Applications of qualitative analysis include identification of Factors That Affect NIR Spectra
raw material, in-process sample, or finished product. These
applications often involve comparing an NIR spectrum from The following list is not exhaustive, but it includes many
a sample to reference spectra and assessing similarities of the major factors that affect NIR spectra.
against acceptance criteria developed and validated for a Sample Temperature—Sample temperature influences
specific application. In contrast, applications of quantitative spectra obtained from aqueous solutions and other hydro-
analysis involve the development of a predictive relationship gen-bonded liquids, and a difference of a few degrees may
between NIR spectral attributes and sample properties. result in significant spectral changes. Temperature may also
These applications typically use numerical models to quan- affect spectra obtained from less polar liquids, as well as
titatively predict chemical and/or physical properties of the solids that contain solvents and/or water.
sample on the basis of NIR spectral attributes. Moisture and Solvent—Moisture and solvent present in
Vibrational spectroscopy in the NIR region is dominated the sample material and analytical system may change the
by overtones and combinations that are much weaker than spectrum of the sample. Both absorption by moisture and
the fundamental mid-IR vibrations from which they origi- solvent and their influence on hydrogen bonding of the APIs
nate. Because molar absorptivities in the NIR range are low, and excipients can change the NIR spectrum.
radiation can penetrate several millimeters into materials, in- Sample Thickness—Sample thickness is a known source
cluding solids. Many materials, such as glass, are relatively of spectral variability and must be understood and/or con-
transparent in this region. Fiber-optic technology is readily trolled. The sample thickness in transmission mode is typi-
implemented in the NIR range, which allows monitoring of cally controlled by using a fixed optical path length for the
processes in environments that might otherwise be sample. In diffuse reflection mode, the sample thickness is
inaccessible. typically controlled by using samples that are “infinitely
The instrument qualification tests and acceptance criteria thick” relative to the detectable penetration depth of NIR
provided in this chapter may not be appropriate for all in- light into a solid material. Here “infinite thickness” implies
strument configurations. In such cases, alternative instru- that the reflection spectrum does not change if the thick-
ment qualification and performance checks should be scien- ness of the sample is increased.
tifically justified and documented. In addition, validation
parameters discussed in this chapter may not be applicable Sample Optical Properties—In solids, both surface and
for all applications of NIR spectroscopy. Validation parame- bulk scattering properties of calibration standards and ana-
ters characterized for a specific NIR application should lytical samples must be taken into account. Surface mor-
demonstrate suitability of the NIR application for its in- phology and refractive index properties affect the scattering
tended use. properties of solid materials. For powder materials, particle
size and bulk density influence scattering properties and the
NIR spectrum.
Transmission and Reflection Polymorphism—Variation in crystalline structure (poly-
morphism) from materials with the same chemical composi-
The most common measurements performed in the NIR tion can influence NIR spectral response. Different poly-
spectral range are transmission and reflection spectroscopy. morphs and amorphous forms of solid material may be
Incident NIR radiation is absorbed or scattered by the sam- distinguished from one another on the basis of their NIR
ple and is measured as transmittance or reflectance, respec- spectral properties. Similarly, different crystalline hydration
tively. Transflection spectrometry is a hybrid of transmission or solvation states of the same material can display different
and reflection wherein a reflector is placed behind the sam- NIR spectral properties.
ple so that the optical path through the sample and back to Age of Samples—Samples may exhibit changes in their
the detector is doubled compared to a transmission meas- chemical, physical, or optical properties over time. Care
urement of a sample of the same thickness. Transflection is must be taken to ensure that both samples and standards
used to describe any double-pass transmission technique. used for NIR analysis are suitable for the intended
The light may be reflected from a diffuse or specular (mir- application.
ror) reflector placed behind the sample. This configuration
can be adapted to share instrument geometry with certain
reflection or fiber-optic probe systems in which the source INSTRUMENTATION
and the detector are on the same side of the sample.
TRANSMITTANCE, T, is a measure of the decrease in radiation
intensity as a function of wavelength when radiation is
passed through a sample. The sample is placed in the opti- Apparatus
cal beam between the source and the detector. The results
of both transmission and transflection measurements are All NIR measurements are based on exposing material to
usually presented directly in terms of absorbance, i.e., incident NIR light radiation and measuring the attenuation
log10(1/T). of the emerging (transmitted, scattered, or reflected) light.
REFLECTANCE, R, is a measure of the ratio of the intensity of Several spectrophotometers are available; they are based on
light reflected from the sample, I, to that reflected from a different operating principles—for example: filters, grating-
background or reference reflective surface, IR. Most reflec- based dispersive, acousto-optical tunable filter (AOTF),

Official from May 1, 2012


Copyright (c) 2011 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 128.83.63.20 by nEwp0rt1 on Sat Dec 03 00:44:32 EST 2011

716 〈1119〉 Near-Infrared Spectroscopy / General Information USP 35

Fourier–transform NIR (FT–NIR), and liquid crystal tunable demonstrate that no abnormal wavelength shift or change
filter (LCTF). Silicon, lead sulfide, indium gallium arsenide, in detector sensitivity has occurred during ongoing analysis.
and deuterated triglycine sulphate are common detector Characterizing Instrument Performance—Specific pro-
materials. Conventional cuvette sample holders, fiber-optic cedures, acceptance criteria, and time intervals for charac-
probes, transmission dip cells, and spinning or traversing terizing NIR instrument performance depend on the instru-
sample holders are common examples of sample interfaces ment and intended application. Many NIR applications use
for introducing the sample to the optical train of a previously validated models that relate NIR spectral response
spectrometer. to a physical or chemical property of interest. Demonstrat-
The selection of specific NIR instrumentation and sam- ing stable instrument performance over extended periods of
pling accessories should be based on the intended applica- time provides some assurance that reliable measurements
tion, and particular attention should be paid to the suitabil- can be taken from sample spectra using previously validated
ity of the sampling interface for the type of sample that will NIR models.
be analyzed. Wavelength Uncertainty—NIR spectra from sample and/or
reference standard materials can be used to demonstrate an
Near-Infrared Reference Spectra instrument’s suitable wavelength dispersion performance
against target specifications. The USP Near IR System Suita-
NIR references, by providing known stable measurements bility Reference Standard or the National Institute of Stan-
to which other measurements can be compared, are used to dards and Technology (NIST) Standard Reference Material
minimize instrumental variations that would affect the (SRM) 2036 for reflectance measurement and NIST SRM
measurement. 2035 for transmittance measurement can be used for wave-
length verification. Suitable materials for demonstrating
Transmittance—The measurement of transmittance re- wavelength dispersion performance include polystyrene,
quires a background reference spectrum for determining the mixtures of rare earth oxides, and absorption by water va-
absorption by the sample relative to the background. Suita- por for instruments that use an interferometer for wave-
ble transmittance reference materials depend on the specific length dispersion. With appropriate justification, alternative
NIR application and include air, an empty cell, a solvent standards may be used. Wavelength uncertainty typically is
blank, or a reference sample. characterized from a single spectrum (collected with the
Reflectance—The measurement of reflectance requires same spectral resolution to obtain the standard value) using
the measurement of a reference reflection spectrum to de- a minimum of three peaks that cover a suitable spectral
termine the attenuation of reflected light relative to the range of the instrument. Typical tolerances for agreement
unattenuated incident beam. The reflectance spectrum is with standard values are ±1.0 nm from approximately 700
calculated as the ratio of the single-beam spectrum of the to 2000 nm and ±1.5 nm above 2000 nm to approximately
sample to that of the reference material. Suitable reflectance 2500 nm (±8 cm–1 below 5000 cm–1 and ±4 cm–1 from
reference materials depend on the specific NIR application 5000 cm–1 to approximately 14,000 cm–1). Alternative toler-
and include ceramic, perfluorinated polymers, gold, and ances may be used when justified for specific applications.
other suitable materials. Photometric Linearity and Response Stability—NIR spectra
from samples and/or reference standard materials with
Qualification of NIR Instruments known relative transmittance or reflectance can be used to
demonstrate a suitable relationship between NIR light atten-
uation (due to absorption) and instrument response. For re-
Qualification—Qualification of an NIR instrument can be flectance measurements, commercially-available reflectance
divided into three elements: Installation Qualification (IQ); standards with known reflectance properties are often used.
Operational Qualification (OQ); and Performance Qualifica- Spectra obtained from reflection standards are subject to va-
tion (PQ). For further discussion, see the proposed general riability as a result of the difference between the experimen-
information chapter Analytical Instrument Qualification tal conditions under which they were factory calibrated and
〈1058〉. those under which they are subsequently put to use. Hence,
Installation Qualification—The IQ requirements help en- the reflectance values supplied with a set of calibration stan-
sure that the hardware and software are installed to accom- dards may not be useful in the attempt to establish an “ab-
modate safe and effective use of the instrument at the de- solute” calibration for a given instrument. Provided that (1)
sired location. the standards do not change chemically or physically, (2)
Operational Qualification—In operational qualification, an the same reference background is also used to obtain the
instrument’s performance is characterized using standards to standard values, and (3) the instrument measures each stan-
verify that the system operates within target specifications. dard under identical conditions (including precise sample
The purpose of operational qualification is to demonstrate positioning), the reproducibility of the photometric scale will
that instrument performance is suitable. Because there are be established over the range of standards. Subsequent
so many different approaches for measuring NIR spectra, measurements on the identical set of standards give infor-
operational qualification using standards with known spec- mation on long-term stability. Photometric linearity is typi-
tral properties is recommended. Using external traceable ref- cally characterized using a minimum of four reference stan-
erence standard materials does not justify omitting the in- dards in the range from 10% to 90% reflection (or
strument’s internal quality control procedures. As is the case transmission). NIR applications based on measuring an ab-
with any spectroscopic device, wavelength uncertainty, pho- sorbance larger than 1.0 may require standards with reflec-
tometric linearity, and noise characteristics of NIR instru- tivity properties between 2% and 5% reflection (or transmis-
ments should be qualified against target specifications for sion) for characterizing instrument performance at low
the intended application. reflectance. The purpose is to demonstrate a linear relation-
Performance Qualification—Performance qualification ship between NIR reflectance and/or transmittance and in-
demonstrates that the NIR measurement consistently oper- strument response over the scanning range of the instru-
ates within target specifications defined by the user for a ment. Typical tolerances for a linear relationship are 1.00 ±
specific application; it is often referred to as system suitabil- 0.05 for the slope and 0.00 ± 0.05 for the intercept of a
ity. Performance qualification for NIR measurements can in- plot of the measured photometric response versus standard
clude comparing a sample or standard spectrum to previ- photometric response. Alternative tolerances may occur
ously recorded spectra. Comparisons of spectra taken over when justified for specific applications.
time from identical and stable samples or reference standard Spectroscopic Noise—NIR instrument software may in-
materials can form the basis for evaluating the long-term clude built-in procedures to automatically determine system
stability of an NIR measurement system. The objective is to noise and to provide a statistical report of noise or S/N over

Official from May 1, 2012


Copyright (c) 2011 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 128.83.63.20 by nEwp0rt1 on Sat Dec 03 00:44:32 EST 2011

USP 35 General Information / 〈1119〉 Near-Infrared Spectroscopy 717

the instrument’s operating range. In addition, it may be de- should be based on sound scientific judgment and can in-
sirable to supplement such checks with measurements that clude materials similar in visual appearance, chemical struc-
do not rely directly on manufacturer-supplied procedures. ture, or name.
Typical procedures involve measuring spectra of traceable Quantitative—Quantitative applications of NIR spectros-
reference materials with high and low reflectance. Toler- copy typically involve establishing a mathematical relation-
ances for these procedures should demonstrate suitable S/N ship between NIR spectral response and a physical or chem-
for the intended application. ical property of interest. Demonstrating specificity against a
HIGH-FLUX NOISE—Instrument noise is evaluated at high- physical or chemical property of interest is based on inter-
light flux by measuring reflectance or transmittance of the preting both NIR spectral attributes and chemometric pa-
reference standard, with the reference material (e.g., 99% rameters in terms of the intended application and may in-
reflection standard) acting as both the sample and the back- clude the following:
ground reference. • Spectral regions in the calibration model can be corre-
LOW-FLUX NOISE—The same procedure may be used with a lated to a known NIR spectral response associated with
lower-reflectivity reference material (e.g., 10% reflectance the property of interest.
standard) to determine system noise at reduced light flux. • Wavelengths used by regression analysis for the calibra-
The source, optics, detector, and electronics make signifi- tion (e.g., for multiple linear regression [MLR] models)
cant contributions to the noise under these conditions. or the loading vector for each factor (e.g., for partial
least squares [PLS] or principal component regression
[PCR] models) can be examined to verify relevant spec-
METHOD VALIDATION troscopic information that is used for the mathematical
model.
• Variation in spectra from samples for calibration can be
Introduction examined and interpreted as expected spectral
observations.
The objective of NIR method validation, as is the case • Variation in material composition and sample matrix
with the validation of any analytical procedure, is to demon- may be shown to have no significant effect on quantifi-
strate that the measurement is suitable for its intended pur- cation of the property of interest within the specified
pose. NIR spectroscopy is somewhat different from conven- method range.
tional analytical techniques because validation of the former Linearity—Quantitative NIR methods generally attempt
generally is achieved by the assessment of chemometric pa- to demonstrate a linear relationship between NIR spectral
rameters, but these parameters can still be related to the response and the property of interest. Although demonstrat-
fundamental validation characteristics required for any ana- ing a linear response is not required for all NIR applications,
lytical method. the model chosen, whether linear or not, should properly
Data pretreatment is often a vital step in the chemometric represent the relationship.
analysis of NIR spectral data. Data pretreatment can be de- Validation of linearity in NIR methods may be accom-
fined as the mathematical transformation of NIR spectral plished by examining a plot of NIR spectral response versus
data to enhance spectral features and/or remove or reduce actual or accepted values for the property of interest. Many
unwanted sources of variation prior to using the spectrum. statistical methods are available for evaluation of the good-
Calibration is the process of developing a mathematical rela- ness of fit of the linear relationship. Other applicable statis-
tionship between NIR spectral response and properties of tics and graphical methods may be as appropriate.
samples. Many suitable chemometric algorithms for data The correlation coefficient, r, may not be an informative
pretreatment and calibration exist; the selection should be measure of linearity. The square of the (Pearson) correlation
based on sound scientific judgment and suitability for the coefficient is a measure of the fraction of the data’s variation
intended application. that is adequately modeled by the equation. Linearity de-
pends on the standard error of the calibration equation (and
hence the reference method) and on the range of the cali-
Validation Parameters bration data. Thus, although values very near 1.00, such as
0.99 or greater, typically indicate a linear relationship, lower
Performance characteristics that demonstrate the suitabil- values do not distinguish between nonlinearity and variabil-
ity of NIR methods are similar to those required for any ity around the line.
analytical procedure. A discussion of the applicable general Range—The specified range of an NIR method depends
principles is found in Validation of Compendial Procedures on the specific application. The range typically is established
〈1225〉. These principles should be considered typical for by confirming that the NIR method provides suitable meas-
NIR procedures, but exceptions should be dealt with on a urement capability (accuracy and precision) when applied to
case-by-case basis. For qualitative NIR methods, see chapter samples within extreme limits of the NIR measurement.
Data Elements Required for Validation, Category IV 〈1225〉, as- Controls must be used to ensure that results outside the
says. For quantitative NIR methods, see chapter Data Ele- validated range are not accepted. In certain circumstances,
ments Required for Validation 〈1225〉, Category I and Category it may not be possible or desirable to extend the validated
II assays. Specific acceptance criteria for each validation pa- range to include sample variability outside the validated
rameter must be consistent with the intended use of the range. Extending the range of an NIR method requires dem-
method. The samples for validation should be independent onstration of suitable measurement capability within the
of the calibration set. limits of the expanded range. Examples of situations in
Specificity—The extent of specificity testing depends on which only a limited sample range may be available are
the intended application. Demonstration of specificity in NIR samples from a controlled manufacturing process and in-
methods is typically accomplished by using the following process samples. A limited method range does not preclude
approaches: the use of an NIR method.
Qualitative—Identification testing is a common applica- Accuracy—Accuracy in NIR methods is demonstrated by
tion of qualitative NIR spectroscopy. Identification is showing the closeness of agreement between the value that
achieved by comparing a sample spectrum to a reference is accepted as either a conventional true value or an ac-
spectrum or a library of reference spectra. The specificity of cepted reference value. Accuracy can be determined by di-
the NIR identification method is demonstrated by obtaining rect comparison between NIR validation results and actual
positive identification from samples coupled with negative or accepted reference values. Suitable agreement between
results from materials that should not meet criteria for posi- NIR and reference values is based on required measurement
tive identification. Materials to demonstrate specificity capability for a specific application. The purpose is to

Official from May 1, 2012


Copyright (c) 2011 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 128.83.63.20 by nEwp0rt1 on Sat Dec 03 00:44:32 EST 2011

718 〈1119〉 Near-Infrared Spectroscopy / General Information USP 35

demonstrate a linear relationship between NIR results and • Identification of previously unknown critical attribute(s)
actual values. Accuracy can be determined by agreement of material(s)
between the standard error of prediction (SEP) and the stan- Revalidation of a quantitative model may be necessary as
dard error of the reference method for validation. The error a result of the following:
of the reference method may be known on the basis of • Changes in the composition of the test sample or fin-
historical data, through validation results specific to the ref- ished product
erence method, or by calculating the standard error of the • Changes in the manufacturing process
laboratory (SEL). Suitable agreement between SEP and SEL is • Changes in the sources or grades of raw materials
based on required measurement capability for a specific • Changes in the reference analytical method
application. • Major changes in instrument hardware
Precision—The precision of an NIR method expresses the Outliers—Sample spectra that produce an NIR response
closeness of agreement between a series of measurements that differs from the qualitative or quantitative calibration
under prescribed conditions. Two levels of precision should model may produce an outlier. This does not necessarily
be considered: repeatability and intermediate precision. The indicate an out-of-specification result; but rather an outlier
precision of an NIR method typically is expressed as the indicates that further testing of the sample may be required
relative standard deviation of a series of NIR method results and is dependent on the particular NIR method. If subse-
and should be suitable for the intended application. Demon- quent testing of the sample by an appropriate method indi-
stration of precision in NIR methods may be accomplished cates that the property of interest is within specifications,
using the following approaches: then the sample meets its specifications. Outlier samples
Repeatability—Repeatability can be demonstrated by the may be incorporated into an updated calibration model
following: subsequent to execution and documentation of suitable vali-
• Statistical evaluation of a number of replicate measure- dation studies.
ments of the sample without repositioning the sample
between each individual spectral acquisition, or
• Statistical evaluation of multiple NIR method results, Method Transfer
each result from a replicate analysis of a sample subse-
quent to re-positioning between spectral acquisitions Controls and measures for demonstrating the suitability of
NIR method performance following method transfer are
Intermediate Precision—Intermediate precision can be similar to those required for any analytical procedure. Ex-
shown by the following: ceptions to general principles for conducting method trans-
• Statistical evaluation of a number of replicate NIR meas- fer for NIR methods should be justified on a case-by-case
urements of the same or similar samples in the basis. The transfer of an NIR method is often performed by
Repeatability study by different analysts on different using an NIR calibration model on a second instrument that
days. is similar to the primary instrument used to develop and
Robustness—NIR measurement parameters selected to validate the method. When a calibration model is trans-
demonstrate robustness will vary depending on the applica- ferred to another instrument, procedures and criteria must
tion and the sample’s interface with the NIR instrument. be applied to demonstrate that the calibration model meets
Critical measurement parameters associated with robustness suitable measurement criteria on the second instrument.
often are identified and characterized during method devel- The selection of an appropriate calibration model transfer
opment. Typical measurement parameters include the procedure should be based on sound scientific judgment.
following:
• Effect of environmental conditions (e.g., temperature,
humidity, and vibration) GLOSSARY
• Effect of sample temperature
• Sample handling (e.g., probe depth, compression of ABSORBANCE, A, is represented by the equation:
material, sample depth/thickness, sample presentation)
• Influence of instrument changes (e.g., lamp change, A = –log T = log (1/T)
warm-up time)
where T is the transmittance of the sample. Absorbance is
also frequently given as:
Ongoing Method Evaluation
A = log (1/R)
Validated NIR methods should be subject to ongoing per-
formance evaluation, which may include monitoring accu- where R is the reflectance of the sample.
racy, precision, and other suitable method parameters. If BACKGROUND SPECTRUM is used for generating a sample
performance is unacceptable, corrective action is necessary. spectrum with minimal contributions from instrument re-
It involves conducting an investigation to identify the cause sponse. It is also referred to as a reference spectrum or back-
of change in method performance and may indicate that ground reference. The ratio of the sample spectrum to the
the NIR method is not suitable for continued use. Improving background spectrum produces a transmittance or reflec-
the NIR method to meet measurement suitability criteria tance spectrum dominated by NIR spectral response associ-
may require additional method development and documen- ated with the sample. In reflection measurements, a highly
tation of validation experiments demonstrating that the im- reflective diffuse standard reference material is for the meas-
proved method is suitable for the intended application. The urement of the background spectrum. For transmission
extent of revalidation required depends on the cause of measurement, the background spectrum may be measured
change in method performance and the nature of corrective with no sample present in the spectrometer or using a cell
action required in order to establish suitable method perfor- with the solvent blank or a cell filled with appropriate refer-
mance. Appropriate change controls should be imple- ence material.
mented to document ongoing method improvement CALIBRATION MODEL is a mathematical expression to relate
activities. the response from an analytical instrument to the properties
Revalidation of a qualitative model may be necessary as a of samples.
result of the following: DIFFUSE REFLECTANCE is the ratio of the spectrum of radiated
• Addition of a new material to the spectral reference light penetrating the sample surface, interacting with the
library sample, passing back through the sample’s surface, and
• Changes in the physical properties of the material reaching the detector to the background spectrum. This is
• Changes in the source of material supply

Official from May 1, 2012


Copyright (c) 2011 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 128.83.63.20 by nEwp0rt1 on Sat Dec 03 00:44:32 EST 2011

USP 35 General Information / 〈1119〉 Near-Infrared Spectroscopy 719

the component of the overall reflectance that produces the in which Ai is the absorbance for each data point; A is the
absorption spectrum of the sample. mean absorbance over the spectral segment; and N is the
FIBER-OPTIC PROBES consist of two components: optical fibers number of points per segment.
that may vary in length and in the number of fibers and a SPECTRAL REFERENCE LIBRARY is a collection of spectra of
terminus, which contains specially designed optics for exam- known materials for comparison with unknown materials.
ination of the sample matrix. The term is commonly used in connection with qualitative
INSTALLATION QUALIFICATION is the documented collection of methods of spectral analysis (e.g., identification of
activities necessary to establish that an instrument is deliv- materials).
ered as designed and specified, is properly installed in the SPECULAR (SURFACE) REFLECTANCE is the reflectance of the front
selected environment, and that this environment is suitable surface of the sample.
for the instrument’s intended purpose. STANDARD ERROR OF CALIBRATION (SEC) is a measure of the ca-
INSTRUMENT BANDWIDTH OR RESOLUTON is a measure of the pability of a model to fit reference data. SEC is the standard
ability of a spectrometer to separate radiation of similar deviation of the residuals obtained from comparing the
wavelengths. known values for each of the calibration samples to the val-
MULTIPLE LINEAR REGRESSION is a calibration algorithm to re- ues that are calculated from the calibration. SEC should not
late the response from an analytical instrument to the be used as an assessment tool for the expected method
properties of samples. The distinguishing feature of this al- accuracy (trueness and precision of prediction) of the pre-
gorithm is the use of a limited number of independent vari- dicted value of future samples. The method accuracy should
ables. Linear-least-squares calculations are performed to es- generally be verified by calculating the standard error of
tablish a relationship between these independent variables prediction (SEP), using an independent validation set of
and the properties of the samples. samples. An accepted method is to mark a part of the cali-
OPERATIONAL QUALIFICATION is the process by which it is bration set as the validation set. This set is not fully inde-
demonstrated and documented that an instrument performs pendent but can be used as an alternative for the determi-
according to specifications and that it can perform the in- nation of the accuracy.
tended task. This process is required following any signifi- STANDARD ERROR OF CROSS-VALIDATION (SECV) is the standard
cant change such as instrument installation, relocation, or deviation calculated using the leave-one-out method. In this
major repair. method, one calibration sample is omitted from the calibra-
OVERALL REFLECTANCE is the sum of diffuse and specular tion, and the difference is found between the value for this
reflectance. sample calculated from its reference value and the value ob-
PARTIAL LEAST SQUARES (PLS) is a calibration algorithm to re- tained from the calibration calculated from all the other
late instrument responses to the properties of samples. The samples in the set. This process is repeated for all samples in
distinguishing feature of this algorithm is that data concern- the set, and the SECV is the standard deviation of the differ-
ing the properties of the samples for calibration are used in ences calculated for all the calibration samples. This proce-
the calculation of the factors to describe instrument dure can also be performed with a group of samples. In-
responses. stead of leaving the sample out, a group of samples is left
PERFORMANCE QUALIFICATION is the process of using one or out. The SECV is a measure of the model accuracy that one
more well-characterized and stable reference materials to can expect when measuring future samples if not enough
verify consistent instrument performance. Performance qual- samples are available for the SEP to be calculated from a
ification may employ the same or different standards for dif- completely independent validation set.
ferent performance characteristics. STANDARD ERROR OF THE LABORATORY (SEL) is a calculation
PHOTOMETRIC LINEARITY, also referred to as photometric verifi- based on repeated readings of one or more samples to esti-
cation, is the process of verifying the response of the photo- mate the precision and/or accuracy of the reference labora-
metric scale of an instrument. tory method, depending on how the data were collected.
PRINCIPAL COMPONENT REGRESSION (PCR) is a calibration al- STANDARD ERROR OF PREDICTION (SEP) is a measure of model
gorithm to relate the response from an analytical instrument accuracy of an analytical method based on applying a given
to the properties of samples. This algorithm, which ex- calibration model to the spectral data from a set of samples
presses a set of independent variables as a linear combina- different from but similar to those used to calculate the cali-
tion of factors, is a method of relating these factors to the bration model. SEP is the standard deviation of the residuals
properties of the samples for which the independent vari- obtained from comparing the values from the reference lab-
ables were obtained. oratory to those from the method under test for the speci-
PSEUDO-ABSORBANCE, A, is represented by the equation: fied samples. SEP provides a measure of the model accuracy
expected when one measures future samples.
A = –log R = log (1/R) SURFACE REFLECTANCE, also known as specular reflection, is
that portion of the radiation not interacting with the sample
where R is the diffuse reflectance of the sample. but simply reflecting back from the sample surface layer
REFERENCE SPECTRUM—See Background Spectrum. (sample–air interface).
REFLECTANCE is described by the equation: TRANSFLECTION is a transmittance measurement technique
in which the radiation traverses the sample twice. The sec-
R = I/IR ond time occurs after the radiation is reflected from a sur-
face behind the sample.
in which I is the intensity of radiation reflected from the TRANSMITTANCE is represented by the equation:
surface of the sample and IR is the intensity of radiation
reflected from a background reference material and its in- T = I/I0 or T = 10A
corporated losses due to solvent absorption, refraction, and
scattering. in which I is the intensity of the radiation transmitted
ROOT-MEAN-SQUARE (RMS) NOISE is calculated by the equation: through the sample; I0 is the intensity of the radiant energy
incident on the sample and includes losses due to solvent

Official from May 1, 2012


Copyright (c) 2011 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 128.83.63.20 by nEwp0rt1 on Sat Dec 03 00:44:32 EST 2011

720 〈1119〉 Near-Infrared Spectroscopy / General Information USP 35

absorption, refraction, and scattering; and A is the optic probes should be used with caution and with refer-
absorbance. ence to appropriate government regulations regarding lasers
and laser classes.
In addition to “normal” Raman spectroscopy, there are
several more specialized Raman techniques. These include
resonance Raman (RR), surface-enhanced Raman spectros-
copy (SERS), Raman optical activity (ROA), coherent anti-
Stokes Raman spectroscopy (CARS), Raman gain or loss
〈1120〉 RAMAN SPECTROSCOPY spectroscopy, and hyper-Raman spectroscopy. These tech-
niques are not widely employed in pharmaceutical laborato-
ries, and are not addressed in this general information
chapter.

INTRODUCTION QUALITATIVE AND QUANTITATIVE RAMAN


Raman spectroscopy shares many of the principles that
MEASUREMENTS
apply to other spectroscopic measurements discussed in There are two general classes of measurements that are
Spectrophotometry and Light-Scattering 〈851〉. Raman is a vi- commonly performed by Raman spectrometry: qualitative
brational spectroscopic technique and is therefore related to and quantitative.
infrared (IR) and near-infrared (NIR) spectroscopy. The
Raman effect itself arises as a result of a change in the po-
larizability of molecular bonds during a given vibrational Qualitative Raman Measurements
mode and is measured as inelastically scattered radiation.
A Raman spectrum is generated by exciting the sample of Qualitative Raman measurements yield spectral informa-
interest to a virtual state with a monochromatic source, typi- tion about the functional groups that are present in a sam-
cally a laser. Light elastically scattered (no change in wave- ple. Because the Raman spectrum is specific for a given
length) is known as Rayleigh scatter and is not of interest in compound, qualitative Raman measurements can be used as
Raman spectrometry, except for marking the laser wave- a compendial ID test, as well as for structural elucidation.
length. However, if the sample relaxes to a vibrational en-
ergy level that differs from the initial state, the scattered
radiation is shifted in energy. This shift is commensurate Quantitative Raman Measurements
with the energy difference between the initial and final vi-
brational states. This “inelastically scattered” light is referred For instruments equipped with a detector that measures
to as Raman scatter. Only about one in 106–108 photons optical power (such as Fourier transform [FT]-Raman spec-
incident on the sample undergoes Raman scattering. Thus trometers), quantitative Raman measurements utilize the fol-
lasers are employed in Raman spectrometers. If the Raman- lowing relationship between signal, Sν, at a given
scattered photon is of lower energy, it is referred to as wavenumber, ν, and the concentration of an analyte, C:
Stokes scattering. If it is of higher energy, it is referred to as
anti-Stokes scattering. In practice, nearly all analytically use- Sν = Kσν(νL − νβ)4P0C
ful Raman measurements make use of Stokes-shifted Raman
scatter. in which K is a constant that depends on laser beam diame-
The appearance of a Raman spectrum is much like an ter, collection optics, sample volume, and temperature; σν is
infrared spectrum plotted linearly in absorbance. The inten- the Raman cross section of the particular vibrational mode;
sities, or the number of Raman photons counted, are plot- νL is the laser wavenumber; νβ is the wavenumber of the
ted against the shifted energies. The x-axis is generally la- vibrational mode; and P0 is the laser power. The Raman
beled “Raman Shift/cm–1” or “Wavenumber/cm–1”. The cross section, σV, is characteristic of the nature of the partic-
Raman shift is usually expressed in wavenumber and repre- ular vibrational mode. The sample volume is defined by size
sents the difference in the absolute wavenumber of the peak of the focus of the laser beam at the sample, the optic
and the laser wavenumber. The spectrum is interpreted in being used for focusing, and the optical properties of the
the same manner as the corresponding mid-infrared spec- sample itself. Spot sizes at the sample can range from less
trum. The positions of the (Raman shifted) wavenumbers for than 1 µm for a microprobe to 6 mm for a large area sam-
a given vibrational mode are identical to the wavenumbers ple system. For Raman spectrometers that measure the
of the corresponding bands in an IR absorption spectrum. number of photons per second (such as change-coupled de-
However, the stronger peaks in a Raman spectrum are often vice [CCD]-Raman spectrometers) the corresponding equa-
weak in an IR spectrum, and vice versa. Thus the two spec- tion is:
troscopic techniques are often said to be complementary.
Raman spectroscopy is advantageous because quick and Sν = KσννL(νL − νβ)3P0C
accurate measurements can often be made without destroy-
ing the sample (solid, semisolid, liquid or, less frequently, From the above equations, it is apparent that peak signal is
gas) and with minimal or no sample preparation. The directly proportional to concentration. It is this relationship
Raman spectrum contains information on fundamental vi- that is the basis for the majority of quantitative Raman
brational modes of the sample that can yield both sample applications.
and process understanding. The signal is typically in the visi-
ble or NIR range, allowing efficient coupling to fiber optics.
This also means that a signal can be obtained from any FACTORS AFFECTING QUANTIFICATION
medium transparent to the laser light; examples are glass,
plastics, or samples in aqueous media. In addition, because
Raman spectra are ordinarily excited with visible or NIR radi- Sample-Based Factors
ation, standard glass/quartz optics may be used. From an
instrumental point of view, modern systems are easy to use, The most important sample-based factors that deleteri-
provide fast analysis times (seconds to several minutes), and ously affect quantitative Raman spectrometry are fluores-
are reliable. However, the danger of using high-powered la- cence, sample heating, absorption by the matrix or the
sers must be recognized, especially when their wavelengths sample itself, and the effect of polarization. If the sample
are in the NIR and, therefore, not visible to the eye. Fiber- matrix includes fluorescent compounds, the measured signal

Official from May 1, 2012


Copyright (c) 2011 The United States Pharmacopeial Convention. All rights reserved.

You might also like