Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

Detection of Lamination Faults from Rotating

Magnetic Fields

J.C Akiror, Student member, IEEE and P. Pillay, A. Merkhouf, Member, IEEE
Fellow, IEEE Electric Equipment Unit
Department of Electrical and Computer Engineering Hydro-Quebec Research Institute (IREQ)
Concordia University Varennes, Qc, Canada
Montreal, Qc, Canada

Abstract— Detection of lamination faults is key in the fault current fields are measured in faulted laminations using a
prevention of more serious generator core damage. Localized Hall sensor. In [7] the effect of faults in laminations is obtained
stator core faults occur frequently in rotational flux zones. from core loss measurements using pick up coils. The
Therefore, this paper studies the behavior of rotational flux in lamination stacks in both cases are excited with a set of coils
the presence of shorted laminations. Faults result in a and a rotor like structure completes the flux path. The
redistribution of the magnetic field across the sample and hence associated core loss with fault conditions are dependent on the
the flux density distribution. Different sizes and positions of location and number of faulted laminations. In [8], the variation
faults are simulated. Moreover a fault is imposed between two of core loss with the number of shorted laminations is found to
laminations and the associated rotational core losses measured.
be 1.6% for twenty faulted laminations. While in [9] the
Experimental results show a distortion in the measured magnetic
field with a 5% difference between the normal and faulted
detector used measures 50% of the total fault current for forty
rotational core losses samples. faulted laminations and 10% for five faulted laminations.
These references show the importance of the measurement
Keywords—core losses; flux density; hydro generators; sensitivity of the used technique.
lamination faults; rotational core losses; rotating flux The reviewed literature only considers fault detection using
pulsating flux limited to the stator yoke or in the stator teeth.
I. INTRODUCTION Distribution of flux in the machine stator shows that over 60%
Core faults in large hydro generators lead to localized hot of the stator area contains rotational flux [10]. It is therefore
spots in the stator of the machine. The severity of this can lead important to investigate the effect of core faults on rotational
to more serious core damage, winding insulation failure and flux and core losses. Moreover, shorting between laminations
eventual failure of the machine. In [1] and [2], the authors in hydro generators commonly occurs at the back of the slots
discuss factors that initiate core faults, the fault growth [11], where rotational flux exists. Rotational core losses are a
mechanisms and other factors that affect these faults usually potential source of localized heating in the stator core. When
from mechanical or electrical action. Keybars are usually laminations are shorted due to insulation failure, the localized
welded to the stator frame to align lamination segments in the heating of the core increases with the number of shorted
core stack [3], in addition, they ground the core. However this laminations. This paper investigates the effect of shorted
also provides a path for fault currents, where any failure of the laminations on rotational flux density distribution and
lamination insulation forms a closed path for the eddy currents rotational core losses. This provides an understanding of the
to flow between the shorted laminations. flux distribution in the core of stators and transformers if a fault
occurs where rotational flux exists. In section II, a rotational
Different methods have been used to detect faults. Most core loss test rig is modeled for simulations. Section III
methods involve exciting the core and measuring the induced analyses different fault sizes and their effect on the flux
fault current, or using thermal imaging to view the hot spots distribution. Section IV presents the experimental setup and
due to the fault currents. The ring flux test requires the stator to measurement of the faults. Section V concludes the paper.
be excited to 85% of the rated induction, to reproduce the flux
levels similar to a working machine [4]. An alternative to this II. FEA MODELING
is, the high frequency ring flux test [5] where the stator is
excited at a higher frequency to achieve the same fault effects The single sheet rotational test rig in [12] was modeled in
with lower excitation current. The electromagnetic core 3D FEA as shown in Fig. 1. It was designed to measure
imperfection detection method (ELCID) uses a lower energy rotational core losses in a 200mm diameter sample using the
level of 4% rated induction in the core and the fault is detected field metric method. The magnetizing field is provided by two
using a Chattock coil. The above techniques are used on real sets of orthogonal windings in each pole arranged as a
machine stator cores. More recently, other techniques have Halbach array. Two sets of flux density (B) coils are wrapped
been proposed to measure core faults on laminations. In [6]
around the sample in the X and Y directions. The field
This work was funded by the Bourses en milieu de pratique (BMP) -
Innovation Research Scholarship Program offered jointly by Fonds de
recherché du Québec – Nature et technologies (FRQNT) and Natural Sciences
and Engineering Research Council of Canada (NSERC).

978-1-4673-7151-3/15/$31.00 ©2015 IEEE 6102


intensity (H) coils are placed above and below the sample in a The poles where modeled as a block with a stacking factor
50mm by 50mm measurement area at the center of the sample. of 0.96 considering the stacking in the z-direction. This
The measurement of the field quantities averages the material simplifies the entire model from being drawn as separate
effects within the sample measurement area. The 3D model of lamination layers because only the samples are modelled as
Fig. 1 has two samples, with each having a thickness of individual laminations. Moreover reducing the memory
0.3556mm and a spacing of 0.05mm between them to account requirement, meshing and simulation time. Phase windings are
for surface insulation on the samples. The material assigned to designed as bars whose height is slightly longer than the poles.
both the poles and the sample is non oriented M15G29. Non- Each bar is given a current direction and the same number of
linear B-H curve and core loss measured data were imported conductors to model the turns per coil before assigning it to a
for the material properties. The material permeability, winding. This further reduces the simulation time compared to
magnetic coercivity, conductivity, permittivity are assumed to when the end windings and individual number of turns are
be isotropic. It is possible to consider anisotropic materials modelled. The effect of the end windings and the winding
with separately defined tensors relative to the coordinate impedance can be added in the external circuit. In the
system. However this requires material property data in all the simulations, currents are directly assigned to the windings. A
axes that is currently unavailable. fine mesh of 2mm for the samples was sufficient for both
computation and simulation time. Eddy current effects were
also enabled in the samples to include the actual effects in the
material. A nonlinear transient solver is selected for a
simulation run of one cycle, with a time step of 5.2e-4s.
The experimental setup was designed to hold only one
lamination. Therefore adding another lamination causes the
samples to be unsymmetrically centered with respect to the
pole depth. Simulations were necessary to compare the effect
of adding another sample. The flux density plots in the middle
of the sample along measurement lines in both the X and Y
Fig. 1. 3D model of the test rig with two laminations, the red shows the y
directions are used for comparison. Fig 2 shows that the flux
coils and the blue the x coils density profile is the same in the measurement region in both
1.6 directions. The magnitude of the flux density with two
samples was reduced by 0.47% compared to that of one
sample. An additional sample requires more exciting current
to achieve the same flux density level as in the single sample.
Flux density (T)

A comparison of the flux density distribution in both the upper


1.5 and lower samples was also done to ensure uniformity of the
1 lamination - X
flux density and field intensity in both samples. Simulation
1 lamination - Y
results in Fig 3 showed that the addition of another lamination
2 laminations - X
does not affect the uniformity in both samples. Both samples
2 laminations - Y had the same magnitude and distribution of flux density with
1.4 no effect of the z-offset on the second sample. Measurements
0 40 80 120 160 200
Sample diameter (mm) of H above and below both samples also showed no effect of
Fig. 2. Comparison of flux density distribution across the sample in the X stacking
and Y directions using one and two samples
1.6 III. LAMINATION FAULT ANALYSIS
Localized faults were modeled by creating a short between
the two samples in the location of the fault. Different positions
and sizes of the faults as shown in Fig. 4 were simulated.
Flux density (T)

Contact resistance between the laminations is a function of the


1.5
area of contact. In the FEA simulation contact resistance was
2 laminations - X not included mainly because at the point of fault there is
2 laminations - Y continuity between the laminations. This negates the need to
define contact resistance. Fault A with a 20mm diameter is
Upper sample - X
implemented on the x and y axes at 30mm from the middle of
Upper sample - Y the sample on x and 70mm from the middle of the sample on y.
1.4 This simulates the effect of fault position either near or further
0 40 80 120 160 200 from the center of the sample. Fault B shows the effect of
Sample diameter (mm) having the fault away from the measurement axes in addition to
Fig. 3. Comparison of the flux density distribution across the sample in the X
and Y direction in the lower and upper samples

6103
around the fault. The size of the area affected by the fault is
also dependent on the magnetization direction and if there are
other faults close to each other as in Fig 5 b). This is because as
C the magnetization direction changes, the flux density
distribution around the fault changes and an interaction
between two adjacent faults can cause a redistribution of the
A
flux within the sample. A large fault as in Fig 5 c) has the same
Y effect as the smaller faults with one side of the fault having a
B
higher flux density and the other having a lower flux density.
The flux density across the sample was measured using a
line in the middle of the sample while field intensity was
measured on the surface of the sample. The flux density in the
middle of the lower sample is plotted in Fig. 6 showing both
X
the X and Y directions for each of the simulated faults to
Fig. 4. Sample with location and size of different faults; A is 20mm
quantify their effect on flux density. The figures are plotted at
diameter; B is 40mm diameter; C is 140mm one time instant when the flux vector is in the x direction.
Each of the faults show unique flux density signatures in the
considering an increase in the fault size, while fault C shows fault locations. In general a fault lowers the flux density in its
the effect of the size of the fault. position. For fault B, which is not aligned with any axes also
Fig. 5 shows the B distribution on the samples when the shows an effect on the flux density on both the x and y - B
flux vector is aligned with the X and Y axes for different fault plots. The flux density profile plotted therefore is located
locations and size. All figures are plotted on the same scale. Fig outside the fault region. As noted in the flux distribution in Fig.
5 a) has no fault and shows the B distribution across the sample 5, the fault alters the distribution of the flux in the areas around
under normal operation. It can be seen that the flux density is the fault hence the observed variation in B in fault B shown in
mostly uniform across the sample except for the edges with Fig.6.
lower flux density. Fig 5 b) and c) show fault A in two H on the surface of the sample is for all the fault cases is
locations and fault C respectively. The presence of a fault plotted in Fig. 7. It was observed that the presence of a fault
causes the flux density distribution to be non-uniform within results to spiking of H along the fault edges as shown by fault.
and around the fault area. As the B vector rotates, these regions A and C. It should be noted that the traces are at one time
of non-uniformity also rotate therefore affecting the area in and instant in the cycle. An average of H for one cycle shows no

a) b) c)
Fig. 5. Flux density distribution in the sample with the flux vector aligned to; the – X axis for figures above and aligned to the
Y axis for figures below a) No fault in the sample, b) With fault A on the X and Y axis 30 and 40mm from the center of the
sample, c) With fault aligned to the X axis

6104
900 160
No fault
Fault A on X
Fault B
800 120
Flux density (mT)

Fault C

Current density (kA/m2)


700 80

No fault
Fault A on X
600 40
Fault B
Fault C
500 0
0 40 80 120 160 200
0 50 100 150 200 Sample diameter (mm)
Sample diameter (mm)
a) Fig. 8. Flux density magnitude in the middle of the lower sample with the
flux vector in the – X direction a) In the X direction b) In the Y direction
900
40.9kA/m2, 23kA/m2, 75.2kA/m2 for fault A, fault B and fault
C respectively. The main limitation in simulation analysis is
800 that the average flux density across the samples was not
controlled which is usually the case with rotational core loss
Flux density (mT)

experimental measurements.
700
No fault IV. EXPERIMENTAL MEASUREMENT
Fault A on Y From the numerical results, faults are significant but are
600
Fault B difficult to measure using coils that average the field
Fault C quantities. Smaller faults may not appear in the captured field
measurements hence only fault C was considered. M15G29
500
0 50 100 150 200 samples are cut to 200mm diameter for measurement. Because
Sample diameter (mm) of the power requirement and rating of the test rig windings,
b) the maximum number of laminations that can be stacked is 2.
Fig. 6. Flux density magnitude in the middle of the lower sample with the
flux vector in the – X direction a) In the X direction b) In the Y direction
Two sets of samples are prepared, one with no fault and the
1.2 other with a fault imposed. The fault was modeled by
No fault removing the surface insulation as shown in Fig. 9. The
Fault A on Y samples are then brought into contact and twenty turns of B-
Magnetic field intensity (kA/m)

Fault B
Fault C coils wrapped around them in the X and Y directions. Twenty
0.8 turns of B-coils are also wrapped around another set of two
laminations without a fault and around one lamination. These
are used for comparison on the effect of the fault. The 240
turns of the H-coils are wrapped around a former to cover a 50
0.4 by 50mm area and placed above and below the samples. Fig
10 shows the samples placed for measurement in the Halbach
array test rig. The test setup consists of two amplifiers for
driving the x and y windings whose input signals are
0 controlled from the computer via DSpace software as shown
0 50 100 150 200
Sample diameter (mm) in Fig. 11. Bx,y and Hx,y sensors are fed back to the computer
Fig. 7. Flux density magnitude in the middle of the lower sample with the via DSpace and used to show the level of magnetization
flux vector in the – X direction a) In the X direction b) In the Y direction
therefore allowing the control of the magnitude and phase shift
significant difference between the fault and no fault cases. The of the Bx and By waveform in addition to data acquisition.
current density magnitude on the x axis in the middle of the These signals are later used to calculate core losses using (1).
lower sample was also plotted as shown in Fig. 8. The eddy
currents in the sample can be visualized from the current
1 § dB x dB y ·
density plots. These eddy currents increase in the presence of a
fault when compared to the no fault case. The increase is
P=
Tρ ³ ¨¨© H
T
x
dt
+ Hy ¸ dt
dt ¸¹
(1)

visible in the fault locations. The average current density over


the entire sample with no fault is 12.9kA/m2 compared to Where P is the total rotational core loss, T is the period, ȡ is

6105
directions and the total core losses obtained as an average of
both measurements. This average value reduces any sensor
Fault induced errors in the measurement.
A. One lamination verses two laminations
The first set of measurements is done to validate the ability
of the test bench to measure losses for two samples. This is
done by comparing the rotational core loss when one and two
laminations are used. Fig. 12 shows that the losses with one
and two samples is the same. Both measurements are at 60Hz.
However because of the limitation of the amplifiers the
maximum B for two laminations is restricted to 1T.
B. Faulted verses no fault laminations
The total core loss is obtained from calculating the area of
the measured B and H field quantities in the x and y. The
measured Bx,y waveforms at 1T in the CW and CCW direction
are shown in Fig. 13. For both the no fault and fault cases, the
a) b) measured flux density is sinusoidal and the presence of the
Fig. 9. Samples a) without fault b) with fault fault cannot be observed. This is mainly because in the
12
H coils
1 lam

Specific loss (W/Kg) 2 lams


8

B coils
4

0
0 0.2 0.4 0.6 0.8 1 1.2 1.4
Magnetizing Flux density (T)
coils Fig. 12. Comparison of core losses for one lamination and two laminations.
measurements, Bx,y are controlled to ensure the flux vector is
Fig. 10. Rotational test rig with the sample and B-H coils installed rotating. In addition, the B-coils measure the average induced
field over the entire sample which make the fault signatures on
the B unnoticeable. Fig. 14 shows the Hx and Hy in the CW

CW Bx - no fault
1
Flux density (T)

CW Bx - fault
CCW Bx - no fault
0 CCW Bx - fault

DSpace Amplifiers
-1

0 0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 0.09


Time(s)
CW By - no fault
1 CW By - fault
Flux density (T)

CCW By - no fault
Halbach set up 0
CCW By - fault

-1

Fig. 11. Rotational test setup 0 0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 0.09
Time(s)
the material mass density. Rotational core losses are measured Fig. 13. Flux density from the Bx,y sensors in the clockwise and
in the clockwise (CW) and counterclockwise (CCW) counterclockwise direction for the fault and no fault condition.

6106
and CCW directions for both the fault and no fault cases. The Fig 15 shows the associated core losses for fault C. Total
sample’s rolling direction is aligned with the x axis hence the rotational core losses are often considered as an average of the
lower values of Hx. The fault is aligned to the y axis hence the losses in the CW and CCW losses, this reduces the errors due
higher distortion in the Hy than the Hx fields. Generally, the to the sensor alignment. The average losses of the faulted case
faulted case shows a distortion in the H especially on the y are higher than the no fault case albeit by a 5% difference. The
axis where the fault is located. magnitude and effect of the fault on core losses is dependent
on the increased eddy currents. These eddy currents increase
The interference of the increased eddy currents in the
with the number of laminations with a fault and the operating
presence of the fault cause this distortion in the measured H.
frequency. The measurements were done for two samples, at
This distortion is dependent on the amount of eddy currents
60Hz. An investigation of a larger number of laminations and
the fault induces. Comparing simulated and measured results,
higher frequencies is required for further verification.
simulated results show a distortion in the B distribution across
the sample and minimum distortion in the H. The Increased eddy currents due to a fault caused an observable
experimental results show a distortion in the measured H and distortion in the measured H. This implies that techniques to
no distortion in the B. This difference is because the detect lamination faults can be developed based on the
simulation results show localized while in experiments, B is measured distorted H fields. In addition rotational core losses
determined from the induced voltage in the B coils. This EMF also increase with the presence of a fault.
from the coils averages the flux density along the length of B-
coils, hence B fault signatures are not visible. Moreover, the V. CONCLUSION
flux average flux density in the sample is controlled in The effect of size and position of faults on the rotating flux
experimental measurements. The local H measurements on the density has been examined using numerical simulations.
surface of the sample in simulation also shows some fault Results showed that the presence of the fault causes non
signatures. The experimentally measured H however shows uniform distribution of B in the sample depending on the size
more significant distortion since it is uncontrolled in relation and location of the fault. In addition the area around the fault
to B therefore responds to the presence of a fault. is also affected by the fault size and magnetization direction.
Interaction of the faults close to each other can also contribute
to more non-uniform flux density distributions. A fault was
Field intensity (A/m)

200 CW Hx - no fault
CW Hx - fault implemented on real laminations and the associated losses
0 measured at 60Hz up to 1T. A comparison of no fault and fault
conditions between two laminations showed less than 5%
-200 difference in the associated rotational core losses. However,
0 0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 0.09 the measured fields in the presence of a fault are distorted, this
Time(s)
can be used as an indication of the presence of a fault.
CW Hy - no fault
Field intensity (A/m)

500
CW Hy - fault

-500

0 0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 0.09


Time(s) CW - No Fault
8
400 CCW - No Fault
Field intensity (A/m)

CCW Hx - no fault Average - No Fault


200 CCW Hx - fault CW - Fault
CCW - Fault
0 6
Average - Fault
Loss (W/Kg)

-200

-400 4
0 0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 0.09
Time(s)
CCW Hy - no fault
Field intensity (A/m)

500
CCW Hy - fault
2
0

-500 0
0 0.2 0.4 0.6 0.8 1
0 0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 0.09 Time(s)
Time(s)
Fig. 15. Rotational core losses in the CW and CCW directions for fault and
Fig. 14. Magnetic field intensity for faulted and no fault laminations in the
no fault cases.
clockwise and counterclockwise direction.

6107
REFERENCES [8] G. B. Kliman, S. B. Lee, M. R. Shah, M. Lusted and N. K. Nair, "A new
method for synchronous generator core quality evaluation," IEEE
[1] P. Tavner and A. Anderson, "Core faults in large generators," IEE Transactions on Energy Conversion, vol. 19, no. 3, 2004.
Proceedings - Electric Power Applications, vol. 152, no. 6, pp. 1427-
[9] R. Romary, C. Demian, P. Schlupp and J. Roger, "Offline and Online
1439, 2005.
methods for stator core fault detection in large generators," IEEE
[2] G. Stone, I. Culbert, E. Boulter and H. Dhirani, Electrical insulation for Transactions on Industrial Electronics, vol. 60, no. 9, 2013.
rotating machines: Design, Evaluation, Aging,Testing and Repair, Wiley
[10] J. Akiror, A. Merkhouf, C. Hudon and P.Pillay, "Consideration of design
& Sons Inc., 2014.
and operation on rotational flux density distributions in hydro generator
[3] M. Znidarich, "Hydro generator stator cores part 1- constructional stators," in International Conference on Electrical Machines, Berlin,
features and core losses," in Australasian Universities Power Germany, 2014.
Engineering Confernce AUPEC, Sydney, 2008.
[11] M. Znidarich, "Hydro generator stator cores Part 2 - Core losses,
[4] H. Tomlinson, "Interlaminar insulation test for synchronous machine degradation mechanisms,testing and specification," in Australasian
stators," Power apparatus and systems, vol. 71, no. 1, 1952. universities power engineering conference (AUPEC), 2008.
[5] J. Sutton, "Method and apparatus for testing laminated cores of electrical [12] N. Alatawneh and P. Pillay, "Design of a novel test fixture to measure
machines". Patent US 20030117144 A1, 26 June 2003. rotational core losses in machine laminations," IEEE Industry
[6] R. Romary, S. Jelassi and J. Brudny, "Stator interlaminar fault detection Applications Society, vol. 48, no. 5, pp. 1467-1477, 2012.
using an external flux density sensor," IEEE Transactions on industrial
electronics, vol. 57, no. 1, 2010.
[7] S. Nazrulla, E. Strangas, J. Agapiou and T. A. Perry, "A device for the
study of electrical steel losses in stator lamination stacks," IEEE
Transactions on industrial electronics, vol. 61, no. 5, 2014.

6108

You might also like