Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

Heating Block Thermostat

Materials Testing & Laboratory Equipment

Heating Block Thermostat Vacuum Test


Apparatus for determining the thermal stability of nitro-cellulose and other nitro-compounds,
single and double base propellants according to the MIL-STD -286C method 403.1.3 and
STANAG 4556, Procedure 1, Vacuum Stability Test.(manometer method)

The equipment consists of:


• Heavy aluminium heating block, electrical heated;
15 holes 15.75 mm Ø, depth 125 mm;
temperature sensor: PT100 probe;
excess temperature safety device 150°C
• Programmable heating controller (PID) in a separate
control box, length of the power cable: appr. 2 m;
digital display for actual and set temperature;
control accuracy: ±0.1°C;
max. adjustable temperature: 135°C
• Set of 15 test tubes consisting of the cylindrical
test tube with mercury collar, capillary tube
with mercury cup;
dimensions of test tube: Ø 15 mm x 155 mm
• Rack for filling glass assembly with mercury
Mains connections: 230 V / 50 Hz 1800 VA

Optional Accessory:
• Vacuum pump IP44
Glass Assembly:
4 stage diaphragm pump
max. vacuum : 0.6 mbar 260

max. suction capacity: 28


80

13l/min 15
15
Grinding
(NS14.5/23)
155

O.d. 6,0 - 6.5 mm


125

i.d. 1,5 - 2,0 mm


850

45°

15
Grinding
Rack 22
80
ca
.1
8

Reichel & Partner GmbH Tel .: +49 6349 9630 981


Archenweyerer Weg 1 • D-76872 Steinweiler Fax: +49 6349 9630 983

You might also like