Professional Documents
Culture Documents
AD229e Operation Manual P8
AD229e Operation Manual P8
AD229e Operation Manual P8
C) From the audiogram the lowest and the highest hearing loss is
determined. The noise level to be used is in the middle of the
two values (usually between 50 to 60 dB). Set the noise level
by means of “HL dB” (22).
The Békésy test in the AD229e is a fixed frequency Békésy test. The
test can be performed with either pure tone or narrow band noise as
stimuli (chosen in setup item no. 10). Before the Békésy mode is
entered the wanted stimuli must be chosen. As standard, continuous
tone is selected for the Békésy test. If pulsing tones are preferred,
“Pulse” (17) must be activated twice before Békésy test is selected.
Békésy Procedure:
A) Select desired test ear Right or Left (11) (12).
When the test is completed for the calculated ear the other ear will
be tested similarly and the test will stop.
After completing the test a Buzzer will start - (if it is selected in the
setup item no. 9).
Display Thresholds:
To view the thresholds stored, hold down “shift” (10) and while
activating “disp thr” (5). Then activate the “Auto Threshold” (7).
The thresholds for the various frequencies will now be
displayed on the LCD.
Use "HL dB" (22) and the "Frequency" (20), (21) to enter the
following data:
In the setup item no. 5 you can choose to have some of the
frequencies deactivated in the Auto Threshold mode.
Please note:
In the setup item no. 6 the Békésy test can be changed from
“Békésy” to Hughson Westlake (and “OSHA” in USA
versions), so that the Hughson Westlake (or OSHA). This
gives the operator the option to choose which threshold test
that he wants to power up when Auto Threshold is activated
Before the MHA mode is entered the wanted stimuli must be chosen.
To enter the MHA mode hold down "shift" (10) while pressing
“function” one or more times until “MHA” is displayed in the right
hand side of the screen.