Download as pdf or txt
Download as pdf or txt
You are on page 1of 13

Supplement-Checklist for the assessment in accordance with

ISO 13485 under CMDCAS


1. Purpose
This checklist is a supplement to Supplement Assessor Guideline MDD and CMDCAS 780E1, and the
Checklist for the Assessment in Accordance with the Standards ISO 13485 etc. 780E1_E. It describes the
auditing procedure for quality management systems according to ISO 13485 under the Canadian Medical
Devices Conformity Assessment System (CMDCAS) based on the Health Canada guidance documents GD 207
(November 2007) and GD 210 (January 2007).
This document replaces the previous version based on the previous versions of Health Canada guidance
documents.
2. Scope
This checklist is applicable for the certification and oversight of the quality systems of the manufacturers of class
II, III and/or IV medical device(s) according to the Canadian Medical Devices Regulations, who have submitted
or are intending to submit an application for Medical Device Licence(s) to Health Canada.
3. Responsibilities and authority
Lead auditor
The lead auditor is responsible for the examination and evaluation of the QM system in respect of the
customer’s system documentation, the related standards and the ruling according to ISO 13485 and the
additional requirements under the CMDCAS program.
He is responsible for the work of the audit team and for the observance of the DQS processes. In case where
the lead auditor is not an expert himself, he should accept the evaluation of the conformity under CMDCAS by
the expert without any protest.
Expert
The expert is responsible for the competent evaluation of the aspects of the QM system that are specific under
the CMDCAS program. The assessment of the quality system’s effectiveness in terms of ensuring observance
of the Canadian Medical Devices Regulations, especially those parts applicable to class II, III and/or IV medical
devices, is his essential task.
4. Approach to the evaluation
This checklist can be applied only in combination with the checklist 780E1_E. The requirements included in
this checklist should be audited along with the corresponding parts of the general ISO 13485 requirements
using the process approach, as outlined in the checklist 780E1_E, and not as a separate part of the audit
schedule.
All nonconformities shall be reported directly against a requirements of ISO 13485 and never directly
against a section of the MDR. If the manufacturer has not adequately addressed one or more applicable
requirements of the MDR, the nonconformity shall be cited against the corresponding ISO 13485
requirement, with reference to the missing section of MDR as objective evidence to support the
nonconformity. If the suitable ISO 13485 requirement cannot be found, the nonconformity can be
reported against the clause 4.2.1.
The present checklist must be used as instrument to perform the assessment.
The assessment results must be formulated in the audit report, where a reference to this checklist can be made.
5. Further applicable documents
 780E1 Assessors Guideline Supplement
 780E1_E-Checkliste-ISO13485-2003-9342EWG-9001
 Canadian Medical Devices Regulations (MDR)
(latest version available from Justice Canada at http://laws.justice.gc.ca/en/F-27/SOR-98-282/text.html)
 Health Canada guidance documents, especially GD 207 and GD 210
(available from Health Canada homepage at http://www.hc-sc.gc.ca/dhp-mps/md-im/qualsys/index-
eng.php)

412_06e_Checklist_CMDCAS.docx
Version: 1.0 1 / 13
 Health Canada Medical Devices Active Licence Listing at http://www.mdall.ca (MDALL), and
Medical Devices Establishment Licence (MDEL) Listing at http://www.hc-sc.gc.ca/dhp-mps/compli-
conform/licences/index-eng.php
 and other further applicable documents and standards referenced.
The original text of MDR, GD210, GD207, and the current manufacturer’s licence listings (MDALL, MDEL)
must be available to the audit team during the audit.
6. Application of the assessment checklist
The checklist serves for the evaluation of audit results. Each audit requirement should be evaluated separately.
The evaluation of the documentation and implementation of a standard’s requirement should be documented in
the column “Evaluation” in the following way:
1 = fulfilled
2 = partially fulfilled, still acceptable
3 = partially fulfilled, not acceptable
4 = not fulfilled
na = not applicable
The numbering of the questions corresponds to the numbers of the requirements as printed in ISO 13485 and
GD 210 (in the italic font).
7. Audit protocol
Please use the DQS form “Findings“ (“Feststellungen“). The findings must contain a reference to the standard’s
requirements. These can be entered in the column “Reference” (“Referenz”). Please use the numbering system
of the standard using at least two digits of the requirement’s number (e.g. 4.2 for “Documentation
Requirements”). Additional evidence, such as copies of manufacturer’s documentation, should be identified
clearly (e.g. using the numbers).
8. Abbreviations
CMDCAS Canadian Medical Devices Conformity Assessment System
MDR Canadian Medical Devices Regulations
HC Health Canada
SCC Standards Council of Canada
QMS quality management system

Ref. No.: Company:

Date(s) of Audit Auditor’s Name Auditor’s Signature

Initial / Re-Certification

st
1 Advancement

nd
2 Advancement

412_06e_Checklist_CMDCAS.docx
Version: 1.0 2 / 13
Checklist Section 2 Additional Guidance under GD210 and GD207

ISO QMS requirement ISO 13485:2003 MDR requirement SOR/98 -282 Auditor Considerations Evaluation
13485 Feb 26, 2006, consolidated
st nd
clause (re)- 1 2
cert. adv. adv.

1 Scope
1.1 General Most recent consolidated version of It says that ISO 13485:2003 shall be used by a Manufacturer to demonstrate
This International Standard specifies Part 1, Medical Devices Regulations that it can provide medical devices that meet customer as well as regulatory
requirements for a quality management (MDR) requirements.
system where an organization needs to
demonstrate its ability to provide medical
devices and related services that
consistently meet customer requirements
and regulatory requirements applicable to
medical devices and related services.

1.2 Application 32(2)(f) permits exclusion of Design Are any exclusions (e.g. ISO 13485 clause 7.3) and non-applicabilities (e.g.
If regulatory requirements permit and development controls (element clauses 7.5.1.2.2 and 7.5.1.2.3) of clauses from ISO 13485 supported with
exclusions of design and development 7.3) from the QMS for Class II medical details of the justification in the quality manual?
controls (see 7.3), this can be used as a devices.
justification for their exclusion from the The manufacturer cannot exclude Design and development controls if their
quality management system. These intention is to obtain a class III or IV device licence.
regulations can provide alternative Does the audit scope include, for Class II devices, manufacture [production], for
arrangements that are to be addressed in Class III or IV devices, design and manufacture, for all medical devices for
the quality management system. It is the which the Medical Device Licences exist or an application is planned?
responsibility of the organization to ensure
that claims of conformity with this Review with the MDALL database:
International Standard reflect exclusion of manufacturer on the Licence = manufacturer on the CMDCAS certificate
design and development controls [see licenced medical device’s type and class correspond to the medical
4.2.2 a) and 7.3]. device’s generic definition of the certificate
3 Terms and definitions This is not a requirement clause of ISO 13485:2003 but is meant to inform the
Existing ISO 9000:2005 and ISO reader. Where this a conflict of definitions, the MDR definitions take
13485:2003 definitions. precedence.
...definitions should be regarded as Part 1, Section 1 Interpretations The MDR holds the "Manufacturer" responsible for the safety and effectiveness
generic, as definitions provided in national In the MDR "Manufacturer" means a of the medical device. The Manufacturer shall be certified to ISO 13485:2003 by
regulations can differ slightly and take person who sells a medical device a Health Canada Recognized Registrar.
precedence. under their own name, or under a
trade-mark, design, trade name or QMS certificates must be issued to the organization that meets the MDR
other name or mark owned or definition of the Manufacturer.
controlled by the person, and who is Does the manufacturer meet his regulatory definition?
responsible for designing,
manufacturing, assembling,
processing, labelling, packaging,
refurbishing or modifying the device, or
for assigning to it a purpose, whether
those tasks are performed by that
person or on their behalf.

758E9_A-CHECKLIST CMDCAS.DOC Ausgabe 08/2008


412_06e_Checklist_CMDCAS.docx
Revision: 0a / 2014-10-01 3 / 134
Checklist Section 2 Additional Guidance under GD210 and GD207

ISO QMS requirement ISO 13485:2003 MDR requirement SOR/98 -282 Auditor Considerations Evaluation
13485 Feb 26, 2006, consolidated
st nd
clause (re)- 1 2
cert. adv. adv.

3.3 Advisory notice Part 1, Section 1 Interpretations The intent of a regulatory "recall" is consistent with the ISO "advisory notice".
notice issued by the organization, "recall", in respect of a medical device
subsequent to delivery of the medical that has been sold, means any action
device, to provide supplementary taken by the manufacturer, importer or
information and/or to advise what action distributor of the device to recall or
should be taken in correct the device, or to notify its
the use of a medical device, owners and users of its
defectiveness or potential
the modification of a medical
defectiveness, after becoming aware
device,
that the device
the return of the medical device a. may be hazardous to health;
to the organization that supplied
b. may fail to conform to any
it, or
claim made by the
the destruction of a medical manufacturer or importer
device relating to its effectiveness,
NOTE Issue of an advisory notice might be benefits, performance
required to comply with national or regional characteristics or safety; or
regulations. c. may not meet the
requirements of the Act or
these Regulations.

4 Quality management system


4.1 General requirements 2, 3, 5 Application If some of the MDR requirements have been delegated by the legal
The organization shall...identify the 6, 7 (including Schedule 1) Manufacturer to another area of the organization, has the delegation of
processes needed for the quality Classification responsibility been clearly defined and documented?
management system and their application 26 Prohibition
throughout the organization (see 1.2)... 28 to 31 Devices deemed licensed Is there more than one geographic location which is used in the design and
32, 33 Application for a device licence manufacturing process?
34 Amendment to a device licence Is there a designated regulatory correspondent?
43 Obligation to inform
If YES → the regularory correspondent is to be involved in the audit!
43.1 Obligation to submit certificate
Offsite auditing using the document review and/or distant communication is
possible.
REGULATORY CORRESPONDENT is a person that is officially authorized by
the manufacturer to be its contact with HC with respect to its licenced devices.
Where an organization chooses to Does the Manufacturer have control over all outsourced processes and are
outsource any process that affects product these control mechanisms identified within the QMS?
conformity with requirements, the
organization shall ensure control over such
The MDR hold the "Manufacturer" responsible for the safety and effectiveness
processes. Control of such outsourced of the medical device including the design, fabrication, assembly, sterilization,
processes shall be identified within the shipping etc. The Manufacturer must demonstrate control over these processes.
quality management system (see 8.5.1). (A) Is the manufacturer responsible for its finished medical devices meeting
regulatory safety, effectiveness and QMS requirements regardless of whether
the manufacturer has outsourced the supply of parts, material, services or
758E9_A-CHECKLIST CMDCAS.DOC Ausgabe 08/2008
412_06e_Checklist_CMDCAS.docx
Revision: 0a / 2014-10-01 4 / 134
Checklist Section 2 Additional Guidance under GD210 and GD207

ISO QMS requirement ISO 13485:2003 MDR requirement SOR/98 -282 Auditor Considerations Evaluation
13485 Feb 26, 2006, consolidated
st nd
clause (re)- 1 2
cert. adv. adv.
finished medical devices?
(B) Does the supplier have a substantial involvement with the manufacture or
design of the medical device?
Is the supplier undertaking the supply of a part, material or service, which may
affect the conformity of the medical device with the Safety and Effectiveness
requirements of MDR?
If the answer is YES to (A) or (B) →
Is there enough objective evidence at the legal manufacturer’s location that
shows the competence of the supplier (subcontractor) to undertake supply of
the part, material or service in relation to the medical devices that are
manufactured or designed under the scope of the registered QMS?
Such evidence is to be especially evaluated in terms of the control and risk
mitigation by the manufacturer over the supplier and their products/services,
e.g.:
 design and product requirements
 personnel qualifications
 verification that subcontracted products meet specified requirements
 specifying a QMS under which the subcontracted products are
designed and manufactured
 performing the on-site inspections
 validating processes
 a QMS certificate with which DQS has confidence
o issued by a HC recognized registrar
(look up in the listing at the HC homepage)
o accredited (by SCC) and not expired
o covering the parts, material, service or finished medical
devices used by manufacturer
or a demonstration of manufacturer’s adequate controls / risk management of
the outsourced product / process.
4.2 Documentation requirements
4.2.1 General 9(2) A manufacturer shall keep Does the Manufacturer have files (often called the Technical File or Device
The quality management system objective evidence to establish that the Master File) that contain, or refers to the location of the evidence of safety and
documentation shall include...f) any other medical device meets those effectiveness required in Sections 10 to 20?
documentation specified by national or requirements. [safety and
regional regulations. effectiveness] These files, or parts of them, can be used to satisfy the :
Where this International Standard specifies 10 to 20 Safety and Effectiveness Class II licence application requirements of Sections 32(1) and (2);
that a requirement, procedure, activity or Requirements Class III licence application requirements of Sections 32(1) and (3); or
special arrangement be "documented", it 32 Application for a Medical device

758E9_A-CHECKLIST CMDCAS.DOC Ausgabe 08/2008


412_06e_Checklist_CMDCAS.docx
Revision: 0a / 2014-10-01 5 / 134
Checklist Section 2 Additional Guidance under GD210 and GD207

ISO QMS requirement ISO 13485:2003 MDR requirement SOR/98 -282 Auditor Considerations Evaluation
13485 Feb 26, 2006, consolidated
st nd
clause (re)- 1 2
cert. adv. adv.
shall, in addition, be implemented and Licence Class IV licence application requirements of Sections 32(1) and (4).
maintained. For Class II devices refer to 32(1) and Have these regulatory requirements been documented in the Quality Manual?
For each type or model of medical device, (2) for minimum requirements.
the organization shall establish and → Refer to the original wording of MDR sec. 31(1) – 31(4) for the minimum
For Class III devices refer to 32(1) and
maintain a file either containing or requirements.
(3) for minimum requirements.
identifying documents defining product For Class IV devices refer to 32(1) and
specifications and quality management (4) for minimum requirements.
system requirements (see 4.2.3). These
documents shall define the complete
manufacturing process and, if applicable,
installation and servicing.

4.2.2 Quality manual


The organization shall establish and 32(2) f) Application for a Class II Applicable only to Class II medical device Manufacturers
maintain a quality manual that includes...a) device licence - QMS requirement
If the Manufacturer excludes element 7.3 Design and development from the
the scope of the quality management
system, including details of and justification
scope of the QMS, does the quality manual contain a detailed justification of its
for any exclusion and/or non-application exclusion?
(see 1.2)... Is the justification based on the Section 32(2)(f)?
4.2.3 Control of documents No specific requirement in MDR. Does the Manufacturer have the most recent consolidated version of the MDR
Existing ISO 13485:2003 QMS including any amendments published in Canada Gazette Part II?
requirements. http://laws.justice.gc.ca
http://canadagazette.gc.ca/
4.2.4 Control of records 55, 56 Distribution records Has the Manufacturer identified the lifetime of the medical devices?
The organization shall retain the records 66 to 68 Implant registration
Has the Manufacturer defined the record retention period for distribution records
for a period of time at least equivalent to
the lifetime of the medical device as
in respect of a medical device for the longer of (a) the projected useful life of the
defined by the organization, but not less device, or (b) two years after the date the device is shipped?
than two years from the date of product
release by the organization or as specified
by the relevant regulatory requirement.

5 Management responsibility
5.1 Management commitment No specific requirement in MDR. Does objective evidence exist of top management's commitment to meeting the
Top management shall provide evidence of requirements of the Canadian MDR? ( e.g. Device Licensing, Mandatory
its commitment to the development and Problem Reports, Recalls etc.)
implementation of the quality management
system and maintaining its effectiveness
by...communicating to the organization the
importance of meeting customer as well as
statutory and regulatory requirements....

758E9_A-CHECKLIST CMDCAS.DOC Ausgabe 08/2008


412_06e_Checklist_CMDCAS.docx
Revision: 0a / 2014-10-01 6 / 134
Checklist Section 2 Additional Guidance under GD210 and GD207

ISO QMS requirement ISO 13485:2003 MDR requirement SOR/98 -282 Auditor Considerations Evaluation
13485 Feb 26, 2006, consolidated
st nd
clause (re)- 1 2
cert. adv. adv.

5.2-5.5 No specific requirement in MDR.

5.6 Management review


5.6.2 Review input No specific requirement in MDR. Is a review of new or revised MDR part of the input to management review?

6 Resource management
6.2.2 Competence, awareness and training No specific requirement in MDR. Have the people who perform regulatory activities been trained in the
understanding of the MDR?
6.3 Infrastructure 14 Transport and storage Does the Manufacturer have and maintain the necessary infrastructure to
The organization shall determine, provide ensure that the characteristics and performance of their medical device(s)
and maintain the infrastructure needed to is(are) not adversely affected by transport or conditions of storage?
achieve conformity to product
requirements. Infrastructure includes, as
applicable
a. buildings, workspace and
associated utilities,
b. process equipment (both
hardware and software), and
c. supporting services (such as
transport or communication).

7 Product realization
7.1 Planning of product realization 32(4)(e) submit a quality plan with a Does the Manufacturer have a quality plan (which is also submitted to Health
The organization shall plan and develop Class IV device application Canada with Class IV medical device applications) that specifies the processes
the processes needed for product 9 Manufacturer's Obligations and resources for specific medical devices?
realization. 32(4)(d) submit risk assessment with Have risk assessment and control measures been performed throughout the
Planning of product realization shall be Class IV device application
consistent with the requirements of the product realization process for all classes of medical devices?
10 Manufacturer to perform risk
other processes of the quality management Note 1: See ISO 14971:2000 for guidance related to risk management.
analysis
system (see 4.1). Note 2: See GHTF SG3 guidance document "Risk Management as an Integral
The organization shall establish Part of the Quality Management System" for additional guidance.
documented requirements for risk www.ghtf.org
management throughout product
realization. Records arising from risk
management shall be maintained (see
4.2.4).

7.2 Customer-related processes


7.2.1 Determination of requirements related to the product

758E9_A-CHECKLIST CMDCAS.DOC Ausgabe 08/2008


412_06e_Checklist_CMDCAS.docx
Revision: 0a / 2014-10-01 7 / 134
Checklist Section 2 Additional Guidance under GD210 and GD207

ISO QMS requirement ISO 13485:2003 MDR requirement SOR/98 -282 Auditor Considerations Evaluation
13485 Feb 26, 2006, consolidated
st nd
clause (re)- 1 2
cert. adv. adv.

The organization shall determine ...c) 2 to 5 Application Has the Manufacturer determined the applicable statutory and regulatory
statutory and regulated requirements 6 to 7 Classification requirements that apply to the medical device?
related to the product 8 to 9 Application & Manufacturer's
Obligations Did the Manufacturer use the Classification rules in Schedule 1 to classify any
new medical devices?
10 to 20 Safety and Effectiveness
21 to 23 Labelling Does the Manufacturer track changes to the information and documents that
24 Contraceptive Devices Advertising had been supplied earlier to Health Canada but did not lead to a licence
26 to 27 Prohibition amendment? Are these changes reported during the annual licence renewal
28 to 31 Medical Devices Deemed process required in section 43(1) (b)?
Licensed Is the Manufacturer only selling licensed medical devices in Canada (see
32 Application for a Medical device
www.mdall.ca)?
Licence
34 Application for licence amendment → Refer to manufacturer’s MDALL listing to check the classification and
43 Obligation to inform compare manufacturer’s identification in the licence and in the (draft)
52 to 56 Distribution records certificate.
57, 58 Complaint handling (Note: Objective evidence to satisfy this requirement will be found either at the
59 to 62 Mandatory problem reporting
legal Manufacturers location or at another location that is part of the
63, 64 Recall
66 to 68 Implant registration Manufacturer's organization like a Canadian subsidiary that is controlled
Schedule 1 Classification rules for through the organization's QMS. See [GD210] Annex C for auditing simple and
medical devices complex multi-site organizations)
Schedule 2 list of Implants that must
comply with the implant registration
requirements of section 66

7.3 Design and development


7.3.2 Design and development inputs 10 Manufacturer to perform risk Has the Manufacturer determined the design and development inputs related to
Inputs relating to product requirements analysis the regulatory safety and effectiveness requirements?
shall be determined and records 11 Device must not adversely affect
maintained (see 4.2.4). These inputs shall the health or safety of a patient Has the Manufacturer determined the medical device licensing requirements for
include 13 Characteristics and performance of Class II, III or IV medical devices?
a) functional, performance and safety device must not deteriorate under
requirements, according to the intended normal use
use, 14 The characteristics and
b) applicable statutory and regulatory performance of a medical device shall
requirements... not be adversely affected by transport
e) output(s) of risk management (see 7.1). or conditions of storage
15 Compatibility - Materials used the
manufacture of the device
16 Minimum risks to a patient and
other person from foreseeable hazards
18 Compatibility between parts and
components
19 Measuring function to perform
within tolerance limits

758E9_A-CHECKLIST CMDCAS.DOC Ausgabe 08/2008


412_06e_Checklist_CMDCAS.docx
Revision: 0a / 2014-10-01 8 / 134
Checklist Section 2 Additional Guidance under GD210 and GD207

ISO QMS requirement ISO 13485:2003 MDR requirement SOR/98 -282 Auditor Considerations Evaluation
13485 Feb 26, 2006, consolidated
st nd
clause (re)- 1 2
cert. adv. adv.
32 Application for a medical device
licence requirements

7.3.6 Design and development validation 12 device shall perform as intended Has design and development validation been carried out and does the medical
As part of design and development 32(3)(f) and 32 (4)(i) : studies on device perform as intended ?
validation, the organization shall perform Class III and IV devices to ensure
clinical evaluations and/or evaluation of
Has design and development validation been performed on initial production
safety and effectiveness
performance of the medical device, as medical devices or their equivalents?
20 validation of software used in a
required by national or regional regulations device Has the performance of any software used in the medical device been
validated?
7.3.7 Control of design and development 1 MDR definition of a "significant Does the Manufacturer have a process or procedure for identifying a "significant
changes change" change" to a Class III or IV medical device?
Significant change means a change Does the Manufacturer have a procedure for amending a Class III or IV medical
that could reasonably be expected to
device licence in the event of a significant change to a medical device?
affect the safety or effectiveness of a
medical device… (a) the Does the Manufacturer track changes to the information and documents that
manufacturing process, facility or had been supplied earlier to Health Canada but did not lead to a licence
equipment; (b) the manufacturing amendment? Are these changes reported during the annual licence renewal
quality control procedures, including process required in section 43(1)(b)?
the methods… to control the quality,
purity and sterility…; (c) the design of
Following any design and development changes, is the device master file
the device…; and (d) the intended use (technical file) updated ?
of the device, including … any change → Review the corresponding procedures and their application.
to the period used to establish the
expiry date.
34 Application for a medical device
licence amendment for Class III or IV
device in the event of a "significant
change"
43(1)(b) Obligation to inform minister
at time of annual licence renewal of
any change to information or
documents that do not meet Section
34 requirements.

7.4 Purchasing
7.4.1 Purchasing process No specific requirement in MDR. Do Manufacturers who use suppliers to provide finished medical devices have
The organization shall establish documented procedures in place to ensure that the finished medical device is
documented procedures to ensure that safe and effective?
purchased product conforms to specified
purchase requirements. The type and Do Manufacturers who use suppliers to provide parts, components or services
extent of control applied to the supplier and that could affect the safety and effectiveness of the finished medical device
the purchased product shall be dependent have documented procedures in place to ensure that the finished medical
upon the effect of the purchased product device is safe and effective?

758E9_A-CHECKLIST CMDCAS.DOC Ausgabe 08/2008


412_06e_Checklist_CMDCAS.docx
Revision: 0a / 2014-10-01 9 / 134
Checklist Section 2 Additional Guidance under GD210 and GD207

ISO QMS requirement ISO 13485:2003 MDR requirement SOR/98 -282 Auditor Considerations Evaluation
13485 Feb 26, 2006, consolidated
st nd
clause (re)- 1 2
cert. adv. adv.
on subsequent product realization or the → Also refer to clause 4.1 in this checklist!
final product.

7.4.2 Purchasing information No specific requirement in MDR.


Purchasing information shall describe the
product to be purchased, including where
appropriate .... quality management system
requirements.

7.5 Production and service provision


7.5.1 Control of production and service provision
7.5.1.1 General requirements 17 Validation and control of If the medical device is to be sold in a sterile condition, is it manufactured and
The organization shall plan and carry out sterilization process sterilized under appropriately controlled conditions, and is the sterilization
production and service provision under method used validated?
controlled conditions.

7.5.1.2 Control of production and service provision - Specific requirements


7.5.1.3 Particular requirements for sterile 17 Validation and control of
medical devices sterilization process

7.5.2 Validation of processes for production and service provision


7.5.2.2 Particular requirements for sterile 17 Validation and control of Does the Manufacturer have documented procedures for the validation of
medical devices sterilization process sterilizations processes?
The organization shall establish
documented procedures for the validation
Was the sterilization process validated prior to initial use?
of sterilization processes. Sterilization Does the Manufacturer have and maintain records of validation of each
processes shall be validated prior to initial sterilization process?
use. Records of validation of each → Review the objective evidence of appropriate manufacturer’s control
sterilization process shall be maintained
and validation of sterilization.
(see 4.2.4).

7.5.3 Identification and traceability


7.5.3.2 Traceability
7.5.3.2 General 21 c) & d) Labelling Do medical devices labels contain: a) the name of the medical device; b) the
.1 The organization shall establish 52 & 54 Distribution records name and address of the Manufacturer; c) the identifier of the medical device,
documented procedures for traceability. including the identifier of any medical device that is part of a system, test kit,
Such procedures shall define the extent of
medical device group, medical device family or medical device group family; d)
product traceability and the records
required (see 4.2.4, 8.3 and 8.5). Where in the case of a Class III or IV medical device, the control number?
traceability is a requirement, the → Refer to the original wording of MDR sec. 21(1) to review the labelling!
organization shall control and record the
unique identification of the product (see

758E9_A-CHECKLIST CMDCAS.DOC Ausgabe 08/2008


412_06e_Checklist_CMDCAS.docx
Revision: 0a / 2014-10-01 10 / 134
Checklist Section 2 Additional Guidance under GD210 and GD207

ISO QMS requirement ISO 13485:2003 MDR requirement SOR/98 -282 Auditor Considerations Evaluation
13485 Feb 26, 2006, consolidated
st nd
clause (re)- 1 2
cert. adv. adv.
4.2.4).

The organization shall establish 52 to 56 Distribution Records Do the Manufacturer's distributors have records of distribution of medical
documented procedures for traceability. devices and are these records kept for the longer of, the projected useful life of
Such procedures shall define the extent of the medical device as defined by the Manufacturer, or two years after the
product traceability and the records
medical device was shipped?
required (see 4.2.4, 8.3 and 8.5). Where
traceability is a requirement, the Do distribution records contain sufficient information to permit a complete and
organization shall control and record the rapid withdrawal of a Class II, III or IV medical device from the market?
unique identification of the product Is the projected useful life of the medical device defined?
66 Implant registration (Applicable to If the Manufacturer sells a medical device in Canada that is listed in Schedule 2
devices listed in Schedule 2 of the of the MDR, does the Manufacturer have procedures and records that satisfy
MDR that are : the traceability requirements of section 66?
1. Heart valves
→ Check the implant registration card’s contents against the requirement
2. Annuloplasty rings
3. Active implantable device of MDR sec. 66(1).
systems → Check if the implant registration card provides for recording of the
a. all models of implantable information specified in MDR sec. 66(2)
pacemakers and leads; → Check if the implant registration card is available in both English and
b. all models of implantable
defibrillators and leads;
French (MDR sec. 66(3))
c. artificial heart;
d. implantable ventricular
support system; and
e. implantable drug infusion
system
4. Devices of human origin
a. human dura mater; and
b. wound covering
7.5.3.2 Particular requirements for active 67(1) Implant registration Do the Manufacturer's agents or distributors of active implantable and
.2 implantable medical devices and 54(1),(2) Distribution records for implantable (as defined in section 3 of ISO 13485:2003) maintain distribution
implantable medical devices implants. records of these medical devices?
The organization shall require that its
agents or distributors maintain records of
the distribution of medical devices to allow
traceability and that such records are
available for inspection.... Records of the
name and address of the shipping package
consignee shall be maintained (see 4.2.4).

7.5.4 Customer property 67(3) Implant registration How does the Manufacturer safeguard the name and address of the patient,
The organization shall exercise care with unless disclosure is required by law?
customer property while it is under the
organization's control or being used by the
organization. The organization shall

758E9_A-CHECKLIST CMDCAS.DOC Ausgabe 08/2008


412_06e_Checklist_CMDCAS.docx
Revision: 0a / 2014-10-01 11 / 134
Checklist Section 2 Additional Guidance under GD210 and GD207

ISO QMS requirement ISO 13485:2003 MDR requirement SOR/98 -282 Auditor Considerations Evaluation
13485 Feb 26, 2006, consolidated
st nd
clause (re)- 1 2
cert. adv. adv.
identify, verify, protect and safeguard
customer property provided for use or
incorporation into the product. If any
customer property is lost, damaged or
otherwise found to be unsuitable for use,
this shall be reported to the customer and
records maintained (see 4.2.4).

7.5.5 Preservation of product 14 Conditions of storage and transport Are the characteristics and performance of the medical device protected during
The organization shall establish transport or storage, taking into account the Manufacturer's instructions and
documented procedures or documented information for transport and storage?
work instructions for preserving the
conformity of product during internal How does the manufacturer ensure such protection?
processing and delivery to the intended
destination.

8 Measurement, analysis and improvement


8.2 Monitoring and measurement
8.2.1 Feedback 57 (1)(a) Complaint Handling Does the Manufacturer and its Canadian importer and distributor(s) maintain
The organization shall establish a records of reported problems or consumer complaints relating to the
documented procedure for a feedback performance characteristics or safety of the medical device?
system [see 7.2.3 c)] to provide early
warning of quality problems and for input Are these problem reports or consumer complaints used as input into the
into the corrective and preventive action corrective action system?
processes (see 8.5.2 and 8.5.3). Note 1 : for guidance on complaint handling and recall see http://www.hc-
sc.gc.ca/hpfb- dgpsa/inspectorate/medical _devices-eng.php
8.2.4- No specific requirement in MDR.
8.4
8.5 Improvement
8.5.1 General 64 & 65 Recall Do the Manufacturer and his Canadian importer have documented procedures
The organization shall establish (note the MDR definition of Recall for them to:
documented procedures for the issue and includes the actions of physically
implementation of advisory notices. These removing a device from the market or 1. carry out an effective and timely recall of the medical device following
procedures shall be capable of being notifying its owners and users of its a consumer complaint or reported problem related to medical device
implemented at any time. defectiveness or potential performance or safety;
defectiveness. See Section 1 of the 2. recall or correct a medical device, or to notify its owners and users in
MDR for official definition) Canada of its defectiveness or potential defectiveness after becoming
58 a) Complaint Handling aware that the medical device : (a) may be hazardous to health; (b)
59-62 Mandatory Problem Reporting may fail to conform to any claim made by the Manufacturer or importer
relating to its effectiveness, benefits, performance characteristics or
safety; or (c) may not meet the requirements of the Food and Drugs

758E9_A-CHECKLIST CMDCAS.DOC Ausgabe 08/2008


412_06e_Checklist_CMDCAS.docx
Revision: 0a / 2014-10-01 12 / 134
Checklist Section 2 Additional Guidance under GD210 and GD207

ISO QMS requirement ISO 13485:2003 MDR requirement SOR/98 -282 Auditor Considerations Evaluation
13485 Feb 26, 2006, consolidated
st nd
clause (re)- 1 2
cert. adv. adv.
Act (R.S., c. F-27, s.1) or the MDR (SOR/DORS/98-282)
Note 3 : for guidance on Product recall procedures see http://www.hc-
sc.gc.ca/hpfb-dgpsa/inspectorate/ medical_devices-eng.php
Records of all customer complaint 57(1)(b) Maintain records of actions Does the Manufacturer and its Canadian importer and distributors maintain
investigations shall be maintained (see taken following reported problems and records of reported problems or consumer complaints relating to the
4.2.4). If investigation determines that the customer complaint. performance characteristics or safety of the medical device ?
activities outside the organization
contributed to the customer complaint, Are these problem reports or consumer complaints used as input into the
relevant information shall be exchanged corrective and preventive action system?
between the organizations involved (see Note 1 : for guidance on complaint handling and recall see http://www.hc-
4.1). sc.gc.ca/hpfb-dgpsa/inspectorate/ medical_devices-eng.php
If any customer complaint is not followed
by corrective and/or preventive action, the
reason shall be authorized (see 5.5.1) and
recorded (see 4.2.4).

If national or regional regulations require Section 59 to 62: Mandatory Problem Does the Manufacturer and its Canadian importer have documented procedures
notification of adverse events that meet Reporting to inform Health Canada of incidents that meet the mandatory reporting criteria
specified reporting criteria, the organization found in Sections 59 to 62.
shall establish documented procedures to
notify the regulatory authorities. → Refer to the original wording of MDR sec. 59-62 to review these
procedures.
Preliminary report within 10 (30) days after the manufacturer or his distributor
becomes aware of the incident, if the incident has (has not) to the death or serious
deterioration in the state of health of the patient, user, or other person
Note 2 : For guidance on mandatory problem reporting see http://www.hc-
sc.gc.ca/hpfb-dgpsa/inspectorate/ medical_devices-eng.php
8.5.2 Corrective action 58 a), b) Complaint Handling See clause 8.5.1 of this checklist.

758E9_A-CHECKLIST CMDCAS.DOC Ausgabe 08/2008


412_06e_Checklist_CMDCAS.docx
Revision: 0a / 2014-10-01 13 / 134

You might also like