Temperature Controller User's Guide: Hart Scientific
Temperature Controller User's Guide: Hart Scientific
7900
Temperature Controller
User’s Guide
Rev. 710801
Limited Warranty & Limitation of Liability
Each product from Fluke Corporation, Hart Scientific Division ("Hart") is warranted to be free from de-
fects in material and workmanship under normal use and service. The warranty period is 2 years for the
Temperature Controller. The warranty period begins on the date of the shipment. Parts, product repairs,
and services are warranted for 90 days. The warranty extends only to the original buyer or end-user cus-
tomer of a Hart authorized reseller, and does not apply to fuses, disposable batteries or to any other prod-
uct, which in Hart's opinion, has been misused, altered, neglected, or damaged by accident or abnormal
conditions of operation or handling. Hart warrants that software will operate substantially in accordance
with its functional specifications for 90 days and that it has been properly recorded on non-defective me-
dia. Hart does not warrant that software will be error free or operate without interruption. Hart does not
warrant calibrations on the Temperature Controller.
Hart authorized resellers shall extend this warranty on new and unused products to end-user customers
only but have no authority to extend a greater or different warranty on behalf of Hart. Warranty support is
available if product is purchased through a Hart authorized sales outlet or Buyer has paid the applicable
international price. Hart reserves the right to invoice Buyer for importation costs of repairs/replacement
parts when product purchased in one country is submitted for repair in another country.
Hart's warranty obligation is limited, at Hart's option, to refund of the purchase price, free of charge re-
pair, or replacement of a defective product which is returned to a Hart authorized service center within
the warranty period.
To obtain warranty service, contact your nearest Hart authorized service center or send the product, with
a description of the difficulty, postage, and insurance prepaid (FOB Destination), to the nearest Hart au-
thorized service center. Hart assumes no risk for damage in transit. Following warranty repair, the prod-
uct will be returned to Buyer, transportation prepaid (FOB Destination). If Hart determines that the
failure was caused by misuse, alteration, accident or abnormal condition or operation or handling, Hart
will provide an estimate or repair costs and obtain authorization before commencing the work. Following
repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for
the repair and return transportation charges (FOB Shipping Point).
THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL
OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IM-
PLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
HART SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL. OR CONSE-
QUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM
BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE OR ANY OTHER
THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or
limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not
apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court of com-
petent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
www.hartscientific.com
Subject to change without notice. • Copyright © 2005 • Printed in USA
Rev. 711801
Table of Contents
2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
4 Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.2 Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.2.1 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.2.2 Heater/Stirrer/LN Cooling . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2
6 General Operation . . . . . . . . . . . . . . . . . . . . . . . . 21
6.1 Control System . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
6.2 Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
6.3 Heater . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
6.4 LN Cooling . . . . . .
2 . . . . . . . . . . . . . . . . . . . . . . . 22
6.5 Temperature Controller . . . . . . . . . . . . . . . . . . . . . . . 22
i
7 Controller Operation . . . . . . . . . . . . . . . . . . . . . . . 23
7.1 Process Temperature . . . . . . . . . . . . . . . . . . . . . . . . 23
7.2 Reset Cut-out . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.3 Temperature Set-point . . . . . . . . . . . . . . . . . . . . . . . . 25
7.3.1 Programmable Set-points . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.3.2 Set-point Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.3.3 Set-point Vernier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
7.4 Temperature Scale Unit . . . . . . . . . . . . . . . . . . . . . . . 27
7.5 Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
7.5.1 Scan Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.5.2 Scan Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.6 Ramp and Soak Program . . . . . . . . . . . . . . . . . . . . . . 28
7.6.1 Number of Program Set-points . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.6.2 Set-points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.6.3 Program Soak Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
7.6.4 Program Function Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
7.6.5 Program Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7.7 Secondary Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7.8 Heater Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7.9 Proportional Band . . . . . . . . . . . . . . . . . . . . . . . . . . 32
7.10 Cutout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
7.11 Controller Configuration . . . . . . . . . . . . . . . . . . . . . . 34
7.12 Probe Parameters RTD Sensor . . . . . . . . . . . . . . . . . . . 34
7.12.1 R . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
0
7.12.2 ALPHA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
7.13 Operating Parameters . . . . . . . . . . . . . . . . . . . . . . . . 35
7.13.1 Cutout Reset Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
7.13.2 Soak Stability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
7.13.3 Boost Temperature Offset. . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
7.14 Serial Interface Parameters . . . . . . . . . . . . . . . . . . . . . 36
7.14.1 Baud Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
7.14.2 Sample Period. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
7.14.3 Duplex Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
7.14.4 Linefeed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
7.15 IEEE-488 Parameters . . . . . . . . . . . . . . . . . . . . . . . . 38
7.15.1 IEEE-488 Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
7.15.2 Termination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.16 Calibration Parameters . . . . . . . . . . . . . . . . . . . . . . . 39
7.16.1 CTO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
7.16.2 CO and CG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
7.16.3 H and L . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
7.16.4 SCO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
ii
8 Digital Communication Interface . . . . . . . . . . . . . . . . 41
8.1 Serial Communications . . . . . . . . . . . . . . . . . . . . . . . 41
8.1.1 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
8.1.2 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
8.1.2.1 Baud Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
8.1.2.2 Sample Period. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
8.1.2.3 Duplex Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
8.1.2.4 Linefeed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
8.1.3 Serial Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
8.1.4 IEEE-488 Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
8.1.4.1 Setup and Address Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
8.1.4.2 IEEE-488 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
9 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . 49
9.1 RTD Probe Calibration . . . . . . . . . . . . . . . . . . . . . . . 49
9.1.1 Calibration Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
9.1.2 Measuring the Set-point Error . . . . . . . . . . . . . . . . . . . . . . . . . 49
9.1.3 Computing R and ALPHA . .
0 . . . . . . . . . . . . . . . . . . . . . . . . . 49
9.1.4 Calibration Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
10 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
11 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . 55
11.1 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
iii
Figures and Tables
iv
1 Before You Start
Symbols Used
Symbol Description
AC (Alternating Current)
AC-DC
Battery
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Off
On
1
7900 Temperature Controller
User’s Guide
Symbol Description
Canadian Standards Association
1.2.1 WARNINGS
To avoid personal injury, follow these guidelines.
GENERAL
DO NOT use the instrument for any application other than calibration work.
The instrument was designed for temperature calibration. Any other use of the
instrument may cause unknown hazards to the user.
DO NOT use the instrument in environments other than those listed in the
user’s guide.
Follow all safety guidelines listed in the user’s manual.
Calibration Equipment should only be used by Trained Personnel.
If this instrument is used in a manner not specified by the manufacturer, the
protection provided by the instrument may be impaired.
If the instrument is used to control a calibration heat source, insure the heater is
wired correctly to the corresponding position on the internal strip as labeled.
The heater will heat continually if the heater and stirrer power are swapped (see
Figure 3 on page 13).
2
1 Before You Start
Safety Information
BURN HAZARD
The instrument can be used to control instruments which generate extreme tem-
peratures. Precautions must be taken to prevent personal injury or damage to
objects. Probes may be extremely hot or cold when removed from a calibration
bath. Cautiously handle probes to prevent personal injury. Carefully place
probes on a heat resistant surface or rack until they are at room temperature.
Fires and severe burns may result if personnel fail to observe safety
precautions.
ELECTRICAL HAZARD
These guidelines must be followed to ensure that the safety mechanisms in this
instrument will operate properly. The instrument must be plugged into an ap-
propriate outlet as specified in Section 2.1, Specifications, on page 7. Also, the
current and voltage capability of the instrument must not be exceeded. The
power cord of the instrument is equipped with a three-pronged grounding plug
for your protection against electrical shock hazards. It must be plugged directly
into a properly grounded three-prong receptacle. The receptacle must be in-
stalled in accordance with local codes and ordinances. Consult a qualified elec-
trician. DO NOT use an extension cord or adapter plug.
DO use a ground fault interrupt device. This system that is controlled by this
instrument may contain a fluid. A ground fault device is advised in case fluid is
present in the electrical system and could cause an electrical shock.
3
7900 Temperature Controller
User’s Guide
Always replace the power cord with an approved cord of the correct rating and
type. If you have questions, contact a Authorized Service Center (see Section
1.3).
High voltage is used in the operation of this equipment. Severe injury or death
may result if personnel fail to observe the safety precautions. Before working
inside the instrument, turn off the power and disconnect the power cord.
FLUIDS
Fluids used in the system controlled by this instrument may produce noxious or
toxic fumes under certain circumstances. Consult the fluid manufacturer’s
MSDS (Material Safety Data Sheet). Proper ventilation and safety precautions
must be observed.
The instrument is equipped with a soft cutout (adjustable parameter) and a hard
cutout (set at the factory) Adjust the soft cutout according to fluid characteris-
tics or application. As a guideline, the soft cutout should be set 10°C to 15°C
below the flash point of the fluid (see Section 7.10, Cutout). Insure that the
flash point, boiling point, or other key fluid characteristics are not exceeded.
1.2.2 CAUTIONS
To avoid possible damage to the instrument, folow these guidelines.
DO use a ground fault interrupt device.
Operate the instrument in room temperatures as listed in Section 3.2, Environ-
mental Conditions.
Calibration constants should only be changed by trained personnel. The correct
setting of these parameters is important to the safety and proper operation of
the instrument.
The Factory Reset Sequence should be performed only by authorized personnel
if no other action is successful in correcting a malfunction. You must have a
copy of the most recent Report of Test to restore the test parameters.
The “cooling” outlet provided by the controller is for LN2 only. DO NOT con-
nect mechanical cooling to the controller.
DO NOT operate this instrument in an excessively wet, oily, dusty, or dirty en-
vironment or place in locations where hot or cold liquids are splashed on it.
Most probes have handle temperature limits. Be sure that the probe handle tem-
perature limit is not exceeded in the air above the system controlled by this
instrument.
The instrument and any thermometer probes used with it are sensitive instru-
ments that can be easily damaged. Always handle these devices with care. Do
not allow them to be dropped, struck, stressed, or overheated.
DO NOT use this instrument in a system that exceeds the current capability of
the instrument as listed in Section 3.1, Specifications.
4
1 Before You Start
Authorized Service Centers
DO NOT replace fuse(s) with one of a higher current rating or type. Always re-
place the fuse with one of the same rating, voltage, and type. The current rating
listed in Section 3.1, Specifications, is the maximum for the instrument. If the
instrument is connected to a system which uses less than the maximum capabil-
ity of this instrument, the fuses need to be changed in order to be correct for the
system. Once the instrument is connected to the system, the system current
needs to be measured or calculated and the appropriate fuse size and character-
istics selected. Section 4.2.6, Fuses, can be used as a guide for selecting a fuse.
Once the correct fuse type and rating is selected, the following information is
applicable. The instrument is equipped with operator accessible fuses. If a fuse
blows, it may be due to a power surge or failure of a component. Replace the
fuses once. If a fuse blows a second time, it is likely caused by failure of a
component. As a test disconnect the output device (heater) and apply power to
the rest of the system. Check to see if the fuse(s) blow. If the fuse(s) blow only
when an output device (heater, stirrer, etc.) is connected, the fault may be in the
system component. If not, contact an Authorized Service Center (see Section
1.3).
If a mains supply power fluctuation occurs, immediately turn off the instru-
ment. Power bumps from brown-outs and black-outs can damage the system.
Wait until the power has stabilized before re-energizing the instrument.
For best accuracy, the instrument needs to be calibrated with the system it
controls.
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: support@hartscientific.com
5
7900 Temperature Controller
User’s Guide
Phone: +31-402-675300
Telefax: +31-402-675321
E-mail: ServiceDesk@fluke.nl
Phone: +86-10-6-512-3436
Telefax: +86-10-6-512-3437
E-mail: xingye.han@fluke.com.cn
6
2 Introduction
2 Introduction
The Hart Scientific Model 7900 is a solid state temperature controller specifi-
cally designed to control the temperature of Rosemount designed temperature
calibration baths. The unique combination of analog and digital electronic cir-
cuitry provides exceptional accuracy and stability together with ease of opera-
tion and programmability.
Temperature sensing is done with a 4-wire 100-ohm platinum resistance probe
that plugs into the back of the controller. To maintain a constant temperature,
the controller adjusts the pulses of power supplied to the heater by means of a
solid state relay.
The 7900 controller can be easily programmed via the four-button front panel
or by the serial or IEEE interface. Programming allows the user to set the con-
trol temperature, temperature scale units to °C or °F, the proportional band, and
the calibration variables. The process or actual temperature is continuously dis-
played on a bright green LED panel. The percent heating power may also be
monitored.
An added safety device, the over-temperature cut-out, is also programmable.
This built-in feature protects the system from fault conditions causing excessive
temperature by disabling the heater if the temperature sensed by a separate
thermocouple probe exceeds the cut-out set-point. Before using the 7900 con-
troller you should understand the proper setup and operation.
7
3 Specifications and Environmental Conditions
Specifications
3.1 Specifications
Temperature Range –100 °C to 400 °C (–148 °F to 752 °F)
Accuracy ± 1.0 °C
Readout Switchable °C or °F
9
7900 Temperature Controller
User’s Guide
10
4 Quick Start
Unpacking
4 Quick Start
This section provides a brief summary of the steps required to set up and oper-
ate the 7900 temperature controller. This section should be used as a general
overview and reference not as a substitute for the remainder of the manual.
Please read Sections 4.2 though 6 carefully before operating the controller.
4.1 Unpacking
Unpack the controller carefully and inspect it for any damage that may have oc-
curred during shipment. If there is shipping damage, notify the carrier
immediately.
An RTD control probe and a thermocouple cut-out probe should have been pur-
chased along with the controller.
Verify that the following components are present:
• 7900 Controller with attached power cord
• Control Probe
• Thermocouple Probe
• User’s Guide
• Probe Adapter (2)
• Extra power cord (15 Amp)
4.2 Installation
4.2.1 Setup
The 7900 controller is a precision instrument that should be located in an ap-
propriate environment. The location should be free from excessive dirt, mois-
ture, vibration, or temperature variations. There should be no present danger of
spilled liquids.
11
7900 Temperature Controller
User’s Guide
This controller can be used with the following heater assemblies. If the heater
assembly that you are using is not listed, please call Hart Scientific Customer
Service for assistance.
To connect the heater perform the following steps after removing the top cover
of the controller (four screws). Be sure the heater cable is adequate for the
amount of current required and that the heater is wired correctly and safely.
1. The heater cable has 4 wires labeled C, D, E, and G. Determine whether
the heater configuration uses G as a ground connection or G as one leg
of the High heater by measuring the resistance from the spade
connectors.
a) The PRESENT heater assembly configuration (Figure 2) uses G as a
ground connection
b) The OLD heater assembly configuration (Figure 1) uses G as one leg
of the High heater.
G D E C
G D E C
12
4 Quick Start
Installation
2. Feed the wires through the strain relief labeled “HEATER” and route
them to the terminal block.
3. If you have the PRESENT heater assembly configuration (Figure 2),
connect all the wires as labeled on the terminal block.
4. If you have the OLD heater assembly configuration (Figure 1), connect
the D and G wires together into the terminal slot labeled D. Connect the
rest of the wires as labeled. There is no direct ground connection through
the heater wires. Instead the assembly will be grounded through the flow
chute assembly.
5. The heater hookup is complete.
Heater
H
I L
G O
H W
LORTNOC RETAEH
EBORP XAM STTAW 0091
T R S A B G G D E C
GNILOOC RITS
ECIVED ROTOM
CT
EBORP
232- SR
13
7900 Temperature Controller
User’s Guide
CAUTION: Be sure that the combined current of the heater, stirrer, and
LN2 cooling does not exceed 20 amps.
4.2.5 Power
Connect the controller power cord to a power source of the appropriate voltage
and current rating (see Section 3.1, Specifications).
4.2.6 Fuses
CAUTION: Never use this instrument in a system that uses more power or
current as listed in Section 3.1, Specifications.
The controller is shipped from the factory with fast acting fuses rated for the
maximum capacity of the instrument.
If the controller is connected to a system which uses less than 10 amps, the
fuses will need to be changed in order to be correct for the system. Once the
controller is connected in the system, the system current needs to be measured
or calculated and the appropriate fuse size and characteristics selected. Gener-
ally, the fuse selected is rated at 125% of the maximum current of the system.
The time-current characteristics of the fuse are selected by the application.
Usually, fast acting fuses are selected systems without a high in-rush current,
i.e. "hot" calibration baths. Time-delay or slow blow fuses are selected for sys-
tems with a high in-rush current, i.e. "cold" calibration baths. Refer to the
fuseology section of your fuse catalog for help in determining fuse size and
characteristics or contact an Authorized Service Center (see Section 1.3) for as-
sistance. Once the correct fuse characteristics and rating of the fuses have been
selected and the appropriate fuses placed in the power entry module of the in-
strument, mark the instrument so the user can visibly see the fuse size and rat-
14
4 Quick Start
Setting the Temperature
ing for fuse replacement. Be sure to change both fuses to the new rating and
correct characteristic.
The controller uses 0.2 amps of current. This current should be taken into con-
sideration when calculating the system power.
Example when using the power of the system:
P = Power of the system (Total Watts)
V = Nominal line voltage (115 VAC)
I = Fuse current
P
I = 1.25 ×
0.9 × v
I F = 1.25 × I
15
7900 Temperature Controller
User’s Guide
16
5 Parts and Controls
Front Panel
17
7900 Temperature Controller
User’s Guide
(5) The heater power is either automatic or user selected. The heater power
switch is used to select the appropriate heater power levels for heating and con-
trolling at various temperatures. Selecting “HIGH” heat manually overrides the
automatic system.
(6) The cooling power switch controls power to the LN2 cooling. Once the LN2
cooling is energized, the flow is automatically controlled.
18
5 Parts and Controls
Back Panel
(5) The heater connector is the source of controlled power for the system
heater. This power is switched by the solid-state relay to maintain a con-
stant temperature. To connect the heater, remove the top cover of the
controller. The heater cable is fed through the strain relief in the back
panel and the spade connectors are attached to the appropriate terminals
on the internal terminal block. See Figure 3. The heater control for tem-
perature range is automatic. The automatic system can be manually over-
ridden by turning the heater power to high using the front panel switch.
(6) The stirrer motor connector is an auxiliary power socket for the stirrer
motor. This power is constant. The combined current of the heater, stirrer
motor, and LN2 cooling must not exceed 20 amps. The stirrer motor ca-
ble is fed through the strain relief in the back panel and the spade con-
nectors are attached to the appropriate terminals on the terminal block
inside the controller. See Figure 3.
(7) The LN2 cooling device connector is the source of controlled power for
the system LN2 cooling.
CAUTION: The cooling power is for LN2 cooling only. Do not use this
connection for mechanical cooling.
This power is switched by the solid-state relay. The LN2 cooling alter-
nates with the heater power to maintain a constant temperature. The LN2
cooling cable is fed through the strain relief in the back panel and the
spade connectors are attached to the appropriate terminals on the internal
terminal block. See Figure 3.
(8) The control probe connector for the system, attaches the 100Ω control
RTD.
19
6 General Operation
Control System
6 General Operation
6.2 Power
Power to the controller is provided by an AC mains supply as specified in Sec-
tion 3.1, Specifications. Power to the controller passes through an internal filter
to prevent switching spikes from being transmitted to other equipment.
To turn on the controller switch the control panel power switch to the ON posi-
tion. The LED display will begin to show the process temperature and the
heater will turn on or off until the bath temperature reaches the programmed
set-point.
When powered on the control panel display will briefly show a four digit num-
ber. This number indicates the number of times power has been applied to the
controller. Also briefly displayed is data that indicates the controller hardware
configuration. This data is used in some circumstances for diagnostic purposes.
6.3 Heater
The power to the heater plugged into the controller is precisely controlled to
maintain a constant system temperature. Power is controlled by periodically
switching the heater on for a certain amount of time using a solid-state relay.
The front panel red/green control indicator shows the state of the heater. The
control indicator glows red when the heater is on and glows green when the
heater is off. The indicator will pulse constantly when the controller is main-
taining a stable temperature.
The heater power automatically switches between low and high when the bath
is being heated to a new temperature. The automatic system can be overridden
by turning the heater switch to “HIGH” power.
21
7900 Temperature Controller
User’s Guide
22
7 Controller Operation
Process Temperature
7 Controller Operation
This section discusses in detail how to operate the temperature controller using
the front control panel. Using the front panel key switches and LED display the
user may monitor the process temperature, set the temperature set-point in de-
grees C or F, monitor the heater output power, adjust the controller proportional
band, set the cut-out set-point, and program the probe calibration parameters,
operating parameters, serial and IEEE-488 interface configuration, and control-
ler calibration parameters. Operations of the functions are shown in the
flowchart summarized in Figure 6.
The temperature display function may be accessed from any other function by
pressing the “EXIT” button.
The message will continue to flash until the temperature is reduced and the
cut-out is reset. The cut-out has two modes, automatic reset and manual reset,
and determines how the cut-out is reset. In automatic mode, the cut-out resets
itself as soon as the temperature is lowered below the cut-out set-point. In man-
ual reset mode, the operator must reset the cut-out after the temperature falls
below the set-point. When the cut-out is active and the cut-out mode is set to
manual (“reset”), the display flashes “cut-out” until the user resets the
cut-out. To access the reset cut-out function press the “SET” button.
23
7900 Temperature Controller
User’s Guide
24
7 Controller Operation
Temperature Set-point
S Reset cut-out
This action also switches the display to the set temperature function. To return
to displaying the temperature, press the “EXIT” button. If the cut-out is still in
the over-temperature fault condition the display will continue to flash
“cut-out”. The bath temperature must drop a few degrees below the cut-out
set-point before the cut-out can be reset.
U Increment memory
Press “SET” to accept the new selection and access the set-point value.
25
7900 Temperature Controller
User’s Guide
pressing “SET”. The set-point value is displayed with the units, C or F, at the
left.
If the set-point value does not need to be changed then press “EXIT” to resume
displaying the temperature. To adjust the set-point value press “UP” or
“DOWN”.
U Increment display
When the desired set-point value is reached press “SET” to accept the new
value and access the set-point vernier. If “EXIT” is pressed any changes made
to the set-point are ignored.
To adjust the vernier press “UP” or “DOWN”. Unlike most functions the ver-
nier setting has immediate effect as it is adjusted allowing the user to continu-
ally adjust the system temperature with the vernier as the temperature is
displayed. “SET” does not need to be pressed.
U Increment display
Next press “EXIT” to return to the temperature display or “SET” to access the
temperature scale units selection.
26
7 Controller Operation
Temperature Scale Unit
S Access vernier
U Change units
Press “SET” to accept the new selection and resume displaying the bath
temperature.
7.5 Scan
The scan control can be turned on or off and the scan rate can be set and en-
abled so that when the set-point is changed the controller heats or cools at a
specified rate (degrees per minute) until the new set-point is reached. With the
27
7900 Temperature Controller
User’s Guide
scan disabled the controller heats or cools at the maximum possible rate.
28
7 Controller Operation
Ramp and Soak Program
Press “SET” to continue. Press “EXIT” to ignore any changes made to the
parameter.
7.6.2 Set-points
The next parameters are the program set-points.
Press “SET” to save the new set-point value. The other set-points can also be
set in the same manner. Once the set-points are programmed as desired press
“EXIT” to continue.
29
7900 Temperature Controller
User’s Guide
Function Action
1 Up-Stop
2 Up-Down-Stop
3 Up-Repeat
4 Up-Down-Repeat
30
7 Controller Operation
Secondary Menu
Press “SET” to activate the new program control command and return to the
temperature display.
To exit out of the secondary menu press “EXIT”. To continue on to the propor-
tional band setting function press “SET”.
31
7900 Temperature Controller
User’s Guide
32
7 Controller Operation
Cutout
D Decrement display
To store the new setting and access the cut-out set-point press “SET”. Press
“EXIT” to exit out of the secondary menu.
7.10 Cutout
As a protection against software or hardware fault, shorted heater triac, or user
error, the controller is equipped with an adjustable heater cutout device that
shuts power off to the heater if the system temperature exceeds a set value. This
protects the heater and system materials from excessive temperatures. The cut-
out temperature is programmable by the operator from the front panel of the
controller. The cutout temperature must always be set below the upper tempera-
ture limit of the bath fluid.
If the cutout is activated because of excessive temperature, power to the heater
shuts off and the system cools. The system cools until it reaches a few degrees
below the cutout set-point temperature. At this point the action of the cutout is
determined by the setting of the cutout mode parameter. The cutout has two
modes — automatic reset or manual reset. If the mode is set to automatic, the
cutout automatically resets when the system temperature falls below the reset
temperature allowing the system to heat up again. If the mode is set to manual,
the heater remains disabled until the user manually resets the cutout.
The cutout set-point may be accessed within the secondary menu. Press “SET”
and “EXIT” to enter the secondary menu and show the heater power. Then
press “SET” twice to access the cutout set-point.
33
7900 Temperature Controller
User’s Guide
D Decrement display
The next function is the configuration menu. Press “EXIT” to resume display-
ing the temperature.
Press “SET” to enter the menu. The probe parameters menu contains the pa-
rameters, R0 and ALPHA, which characterize the resistance-temperature rela-
tionship of the platinum control probe. These parameters may be adjusted to
34
7 Controller Operation
Operating Parameters
improve the accuracy of the bath. This procedure is explained in detail in Sec-
tion 9.
The probe parameters are accessed by pressing “SET” after the name of the pa-
rameter is displayed. The value of the parameter may be changed using the
“UP” and “DOWN” buttons. After the desired value is reached press “SET” to
set the parameter to the new value. Pressing “EXIT” causes the parameter to be
skipped ignoring any changes that may have been made.
7.12.1 R0
This probe parameter refers to the resistance of the control probe at 0°C. Nor-
mally this is set for 100.000 ohms.
7.12.2 ALPHA
This probe parameter refers to the average sensitivity of the probe between 0
and 100°C. Normally this is set for 0.00385°C.
Press “SET” to enter the menu. The operating parameters menu contains the
cutout reset mode and soak stability.
Press “SET” to access the parameter setting. Normally the cutout is set for
manual mode.
35
7900 Temperature Controller
User’s Guide
To change the parameter setting, press “UP” or “DOWN” and then “SET” to
store the value.
To change the parameter setting, press "UP" or "DOWN" and then "SET" to
store the value.
The serial interface parameters menu contains parameters which determine the
operation of the serial interface. The parameters in the menu are: baud rate,
sample period, duplex mode, and linefeed.
36
7 Controller Operation
Serial Interface Parameters
Press “SET” to choose to set the baud rate. The current baud rate value is then
displayed.
The baud rate of the serial communications may be programmed to 300 600,
1200, or 2400 baud. 2400 baud is the default setting. Use “UP” or “DOWN” to
change the baud rate value.
Press “SET” to set the baud rate to the new value or “EXIT” to abort the opera-
tion and skip to the next parameter in the menu.
Press “SET” to choose the sample period. The current sample period value is
displayed.
Adjust the value with “UP” or “DOWN” and then use “SET” to set the sample
rate to the displayed value. “EXIT” does not store the new value.
37
7900 Temperature Controller
User’s Guide
The mode may be changed using “UP” or DOWN” and pressing “SET”.
7.14.4 Linefeed
The final parameter in the serial interface menu is the linefeed mode. This pa-
rameter enables (on) or disables (off) transmission of a linefeed character (LF,
ASCII 10) after transmission of any carriage-return. The linefeed parameter is
indicated by,
The mode may be changed using “UP” or “DOWN” and pressing “SET”.
38
7 Controller Operation
Calibration Parameters
Adjust the value with “UP” or “DOWN” and then use “SET” to set the address
to the displayed value.
7.15.2 Termination
The transmission termination character can be set to carriage return only, line-
feed only, or carriage return and linefeed. Regardless of the option selected the
controller interprets either a carriage return or a linefeed as a command termi-
nation during reception. The termination parameter is indicated with,
39
7900 Temperature Controller
User’s Guide
7.16.1 CTO
Parameter CTO sets the calibration of the over-temperature cutout. This param-
eter is not adjustable by software but is adjusted with an internal potentiometer.
7.16.2 CO and CG
These parameters calibrate the accuracy of the controller set-point. These are
programmed at the factory when the controller is calibrated. Do not alter the
value of these parameters. If the user desires to calibrate the controller for im-
proved accuracy then calibrate R0 and ALPHA according to the procedure
given in Section 9.
7.16.3 H and L
These parameters set the upper and lower set-point limits of the controller.
These parameters should not be set beyond the safe operating temperature lim-
its of the system.
7.16.4 SCO
This parameter is used at the factory for testing purposes and SHOULD NOT
be altered by the user.
40
8 Digital Communication Interface
Serial Communications
8.1.1 Wiring
The serial communications ca-
ble attaches to the controller
through the DB-9 connector at
the back panel. Figure 8 shows
the pin-out of this connector
and suggested cable wiring. To
eliminate noise the serial cable
should be shielded with low re-
sistance between the connector
(DB-9) and the shield.
8.1.2 Setup
Figure 8 Serial Communications Cable Wiring
Before operation the serial in-
terface must first be set up by
programming the BAUD rate
41
7900 Temperature Controller
User’s Guide
8.1.2.4 Linefeed
The final parameter in the serial interface menu is the linefeed mode. This pa-
rameter enables (“On”) or disables (“OFF”) transmission of a linefeed charac-
ter (LF, ASCII 10) after transmission of any carriage-return. The default setting
is with linefeed on. The mode may be changed using “UP” or “DOWN” and
pressing “SET”.
42
8 Digital Communication Interface
Interface Commands
43
7900 Temperature Controller
User’s Guide
44
8 Digital Communication Interface
Power Control Functions
45
7900 Temperature Controller
User’s Guide
46
8 Digital Communication Interface
Power Control Functions
47
7900 Temperature Controller
User’s Guide
the parameter to ”1" turns the cooling on. “F2" with no value returns the cur-
rent state of the cooling power control.
Function Command 0 1
High Heat F1 low high
Cooling F2 off on
48
9 Calibration Procedure
RTD Probe Calibration
9 Calibration Procedure
In some instances the user may want to calibrate the controller to improve the
temperature set-point accuracy. Calibration is done by adjusting the controller
probe calibration constants R0 and ALPHA so that the process temperature as
measured with a standard thermometer agrees more closely with the set-point.
The thermometer used must be able to measure the temperature with higher ac-
curacy than the desired accuracy of the system.
49
7900 Temperature Controller
User’s Guide
The user should keep a record of these values in case they may need to be re-
stored in the future. The new values R0′ and ALPHA′ are computed by entering
the old values for R0 and ALPHA, the calibration temperature set-points tL and
tH , and the temperature errors errL and errH into the following equations,
′ ⎡ err t − errL tH ⎤
R0 = ⎢ H L ALPHA + 1⎥ R0
⎣ tH − tL ⎦
50
9 Calibration Procedure
RTD Probe Calibration
R0 = 100.000
ALPHA = 0.0038500
tL = 80.00°C
measured t = 79.843°C
tH = 120.00°C
measured t = 119.914°C
Compute errors,
errL = 79.843 - 80.00°C = -0.157°C
errH = 119.914 - 120.00°C = -0.086°C
Compute R0,
⎡ ( −0.086) x 80.0 − ( −0157
. ) x 120.0 ⎤
R0′ = ⎢ 0.00385 + 1 ⎥100.000 = 100115
.
⎣ 120.0 − 80.0 ⎦
Compute ALPHA,
. ) − (1 + 0.00385 x 80.0)( −0.086) ⎤
⎡ (1 + 0.00385x120.0)( −0157
ALPHA′ = ⎢ + 1⎥ 0.00385 = 0.0038387
⎣ 120.0 − 80.0 ⎦
51
10 Maintenance
10 Maintenance
The controller has been designed with the utmost care. Ease of operation and
simplicity of maintenance have been a central theme in the product develop-
ment. Therefore, with proper care the instrument should require very little
maintenance. Avoid operating the instrument in dirty or dusty environments.
• A battery is used to maintain operating parameters in the unit. All operat-
ing parameters, including calibration parameters should be checked on a
regular basis to insure accuracy and proper operation of the instrument.
See the troubleshooting section for the procedure on checking the status
of the battery.
• If the outside of the controller becomes soiled, it may be wiped clean with
a damp cloth and mild detergent. Do not use harsh chemicals on the sur-
face, which may damage the paint.
• If a hazardous material is spilt on or inside the equipment, the user is re-
sponsible for taking the appropriate decontamination steps as out-lined by
the national safety council with respect to the material. MSDS sheets ap-
plicable to all fluids used in the baths should be kept in close proximity to
the instrument.
• If the mains supply cord becomes damaged, replace it with a cord with
the appropriate gauge wire for the current of the system. If there are any
questions, call Hart Scientific Customer Service for more information.
• Before using any cleaning or decontamination method except those rec-
ommended by Hart, users should check with Hart Scientific Customer
Service to be sure that the proposed method will not damage the equip-
ment.
• If the instrument is used in a manner not in accordance with the equip-
ment design, the operation of the controller may be impaired or safety
hazards may arise.
53
11 Troubleshooting
Troubleshooting
11 Troubleshooting
This section contains information on troubleshooting, and a wiring diagram.
This information pertains to a number of bath models and certain specifics may
not pertain to your model.
11.1 Troubleshooting
In the event that the user of the 7900 temperature controller encounters diffi-
culty in operation, this section may help to find and solve the problem. Several
possible problem conditions are described along with likely causes and solu-
tions. If a problem arises, please read this section carefully and attempt to un-
derstand and solve the problem. Troubleshooting may require the assistance of
a technician with at least a basic understanding of electronics. If the controller
seems to be faulty or the problem cannot otherwise be solved then contact a
Hart Scientific Authorized Service Center (see Section 1.3).
55
7900 Temperature Controller
User’s Guide
56
11 Troubleshooting
Troubleshooting
57
12 Appendix - 910 Bath Probe Adapter Instructions
TC Cut-Out
12.1 TC Cut-Out
Your Rosemount bath was not originally provided with a TC cutout. Therefore,
provisions will have to be made to add this to the bath. May we suggest the
following:
1. There is adequate space on the cover plate of the bath exactly across
from the control probe to add the TC.
2. Remove the plate and flow chute from the bath (to keep metal shavings
out of the bath).
3. Drill a hole in the plate using at least a #265 drill bit (17/64).
4. Attach the probe adapter to the plate.
Adapter
Cover Plate Guide Tube
Spacer
Washer
Locking
Collar
59
7900 Temperature Controller
User’s Guide