Professional Documents
Culture Documents
H-NMR Determination of Ketone-Ethylene-Ester and Polyvinyl Chloride Contents in KEE-PVC Roofing Fabrics
H-NMR Determination of Ketone-Ethylene-Ester and Polyvinyl Chloride Contents in KEE-PVC Roofing Fabrics
for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: D8154 − 17
1. Scope 3. Terminology
1.1 This test method pertains to the determination of the 3.1 Definitions—Terminology D1079 and Specification
relative contents of Ketone-Ethylene-Ester (KEE) and Polyvi- D6754/D6754M shall apply to this test method.
nyl Chloride (PVC) after their extraction from reinforced
roofing membranes, or fabrics. Based on Proton Nuclear 4. Summary of Test Method
Magnetic Resonance Spectroscopy (1H-NMR), the method 4.1 Test Methods—There are two related test methods to
allows for the quantification of PVC with respect to an internal quantify PVC and KEE in roofing fabrics. In both methods the
standard. The KEE content is then obtained by difference. The polymer blend is extracted from the fabric and purified by a
test method is not applicable to membranes or blends that series of dissolution, centrifugation, and precipitation. After
contain high molecular weight polymers other than PVC and drying, the blend is dissolved in deuterated tetrahydrofuran
KEE. along with a 1H-NMR internal standard. In Method A, PVC
1.2 Units—The values stated in SI units are to be regarded content is quantified by comparing its 1H-NMR signal with that
as standard. No other units of measurement are included in this of the internal standard, and KEE is obtained by difference.
standard. Method B adds to Method A a calibration curve the slope of
which providing for a correction factor. This correction factor
1.3 This standard does not purport to address all of the is applied to the polymer content obtained by Method A to
safety concerns, if any, associated with its use. It is the provide for greater precision on the PVC and KEE contents.
responsibility of the user of this standard to establish appro-
priate safety, health, and environmental practices and deter- 5. Significance and Use
mine the applicability of regulatory limitations prior to use.
5.1 Determination of the percentage of KEE compound in
1.4 This international standard was developed in accor- sheet roofing is of concern to many specifiers and building
dance with internationally recognized principles on standard- owners. Method A, the faster method, is best used for quality
ization established in the Decision on Principles for the control. Method B may apply better to research and
Development of International Standards, Guides and Recom- development, or investigative work.
mendations issued by the World Trade Organization Technical
Barriers to Trade (TBT) Committee. 6. Interferences
2. Referenced Documents 6.1 The NMR analysis of the polymer blend extracted from
the sheet roofing is sensitive to impurities, including
2.1 ASTM Standards:2 plasticizer, filler, reinforcement, and trace solvent (other than
D1079 Terminology Relating to Roofing and Waterproofing the NMR solvent). The procedure here is meant to purify the
D6754/D6754M Specification for Ketone Ethylene Ester blend for proper analysis.
Based Sheet Roofing
7. Equipment and Apparatus
7.1 Glassware. Erlenmeyer flask and stopper, 125 mL.
1
This test method is under the jurisdiction of ASTM Committee D08 on Roofing Weighing bottle, 25 mL.
and Waterproofing and is the direct responsibility of Subcommittee D08.18 on
Nonbituminous Organic Roof Coverings. 7.2 Porcelain ware. Büchner funnel, 10 cm in diameter.
Current edition approved Nov. 1, 2017. Published November 2017. DOI:
10.1520/D8154-17. 7.3 Platform shaker. Must accept 125 mL Erlenmeyer, and
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or allow for swirling at 120 to 150 rpm.
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on 7.4 Metal spatula. To stir the suspension during the precipi-
the ASTM website. tation step.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
D8154 − 17
7.5 Pasteur pipettes. To transfer solubilized polymer and details. Centrifugation flasks must be resistant to the fluid
solvent from one flask to another. being used and must not swell.
7.6 Centrifuge. Capable of 10 000 RPM. 9.3 Magnetic Fields—NMR spectroscopy involves strong
7.7 Centrifuge tubes. 50 mL. Must be resistant to THF, for magnetic fields. Analysis by NMR requires safety and equip-
example, Fluorinated Ethylene Propylene. ment training. Follow local laboratory guidelines.
2
D8154 − 17
NOTE 3—Thorough drying is required to remove traces of THF, which 12.3 Spectrum Processing—For quantitative NMR, the
will interfere with the calculated PVC-KEE content NMR analysis. THF peaks must be as narrow and symmetric as possible and the
will be readily visible in the 1H-NMR spectrum. Before dissolution of the
sample in deuterated-THF as per 11.2.1, the absence of protonated-THF
integral line must reflect the surface area of spectral lines as
may be verified by means of Fourier-transform infrared spectroscopy, or well as the absence of spectral lines (flat baseline). Note, for
thermo-gravimetric analysis. instance, the flat baseline on the TCNB integration in Fig.
11.1.9 Reserve the polymer dried as per 11.1.8 in a flask to X1.1.
protect it from contaminant until it is used to prepare the NMR 12.3.1 Phasing—It is first necessary to greatly expand the
sample. vertical scale of the spectrum to see the base of the peaks and
make them as symmetric as possible by adjusting the phase so
11.2 Preparation of NMR Sample: that the tails on each side of the peaks are as short and
11.2.1 Polymer Sample—Weigh in a small weighing pan symmetric as possible. Phasing is done by adjusting two
and with a precision of 0.001 mg about 20 mg of TCNB and parameters: the zero order and the first order phase and the
about 30 mg of polymer dried as per 11.1.8. Transfer these process vary depending on the NMR software.
solids to a 5 mL weighting glass bottle. Use about 1 mL of 12.3.2 Integration—Once properly phased, the spectrum
THF-d8 to wash residual solid in the weighting pan and pipet must be processed for baseline correction for the integral line
it out to the glass bottle. Close the weighing glass bottle with to reflect the baseline; the integral line should increase when a
its lid, and leave the polymer-TCNB solid mixture to dissolve peak is present and should be flat in the absence of peaks. Once
overnight. With a pipette, transfer the solution into a 5 mm again, the process is different depending on the software used
NMR tube. for data processing, but it is similar to phasing in the sense that
NOTE 4—Keep aside the weights of TCNB and the dried polymer for a zero and first order parameter must be adjusted.
use in the calculation of PVC and KEE content in Section 13.
11.2.2 Calibration Samples—Calibration samples have 13. KEE Calculations
PVC/KEE weight ratios close to 20/80, 30/70, 40/60, 50/50, 13.1 Two methods may be used to quantify the ratio of KEE
and 60/40. Samples are prepared by individually weighting and PVC in the extracted polymer blend. Both methods are
between about 5 mg to 25 mg of PVC and KEE for a total based on the NMR integration values of peaks from the internal
polymer content near 30 mg. Weigh in a small weighing pan standard TCNB and PVC. In Method A, the polymer ratio is
and with a precision of 0.001 mg about 20 mg of TCNB and obtained from the analysis of a single sample. In Method B, the
about 30 mg of the PVC/KEE mixture with a set ratio. Transfer polymer ratio is obtained after the development of a calibration
these solids to a 5 mL weighting glass bottle. Use about 1 mL curve.
of THF-d8 to wash residual solid in the weighting pan and pipet NOTE 7—The TCNB signal between 7.80 to 8.40 ppm arises from a
it out to the weighing bottle. Close the weighing bottle with its single proton, and it is assumed that the PVC signal between 4.12 to 4.75
lid, and leave the polymer-TCNB solid mixture to dissolve ppm also arises from a single proton, from CH2-CHCl.
overnight. With a pipette, transfer the solution into a 5 mm 13.1.1 Method A.
NMR tube. Table X1.1 in Appendix X1 provides an example of 13.1.1.1 Calculation of PVC Content—In this method, the
calibration sample compositions. polymer and TCNB weights from 11.2.1, and the NMR
NOTE 5—Keep aside the weights of TCNB, PVC, and KEE for use in integration values from TCNB and PVC in 12.2 are used to
the calibration curve and calculation of PVC and KEE as per 13.1.2. calculate the PVC weight content as per Eq 1, where MW is the
molecular weight. The MW of TCNB is 260.89 g/mol, whereas
12. NMR Analysis that of PVC is 62.50 g/mol.
12.1 Acquisition Parameters—The proton-NMR probe tem- PVC mass
PVC content % 5 100* (1)
perature is to be set to 30°C and sample is to be left to stabilize sample mass
for 20 minutes in the probe prior to testing. Measure spin-
lattice relaxation, T1, by inversion recovery. Set the relaxation molar quantity PVC * MW of PVC
5100*
delay to about six times the value of T1 to ascertain that full sample mass
relaxation is achieved in the quantitative experiment. For
instance, a 60 s relaxation delay is used for a T1 of 9.9 s.
5100*
molar quantity TCNB* S integration PVC
integration TCNB D
*MW of PVC
NOTE 6—Probe tuning and 90 degree pulse width determination must sample mass
be done for every sample prior to acquisition, for example 9.3 µs at 53 dB.
12.1.1 Collect the quantitative 1H-NMR spectrum at a
spectral width of 8.0 kHz, acquisition time of 3.75 s, and 5100*
S
mass TCNB
MW of TCNB
* DS
integration PVC
integration TCNB
*MW of PVCD
co-add 32 scans (transients). sample mass
Taking for instance a TCNB mass of 0.019938 g, a polymer
12.2 Spectrum Analysis—Reference the spectrum at 3.570 mass of 0.029673 g, and a PVC integration of 242.61, and a
ppm with respect to the proton trace in THF-d8. Upon spectrum TCNB integration of 100 leads to a percent PVC content of
processing as per 12.3, set to 100 the integral regions from 7.80 39.0 %.
to 8.40 ppm for TCNB, and obtain the integration from 4.12 to PVC content % 5 100*
4.75 ppm for PVC. Fig. X1.1 in Appendix X1 provides an 0.011588 g
5100* 5 39.0
example of a spectrum with integration. 0.029673 g
3
D8154 − 17
13.1.1.2 Calculation of KEE Content—The KEE content is 14. Report
calculated as per Eq 2. 14.1 The report shall include the following information for
KEE percent content 5 100 2 percent PVC from Equation 1 (2) each material.
14.1.1 Complete identification of the sheet roofing material,
13.1.2 Method B:
including source, manufacturer, and lot number. Specify if the
13.1.2.1 Measure the PVC mass in the polymer blend
material is from standard production, or from a laboratory
extracted from the roof sheet sample as per Method A.
preparation without reinforcement, for instance.
13.1.2.2 Prepare model mixtures with pure KEE and PVC
14.1.2 The test method to quantify KEE, Method A or
as per 11.2.2 and keep the PVC content aside to build a
Method B.
calibration curve in 13.1.2.2(2).
14.1.3 The KEE content to the nearest tenth of one percent.
(1) Measure the PVC content of the model mixtures as per
NMR Method A and keep the percent PVC aside to build a
15. Precision and Bias
calibration curve in 13.1.2.2(2).
(2) Prepare a calibration curve by plotting the actual PVC 15.1 Intra-laboratory Testing—The precision of intra-
weight percent of the model mixtures (13.1.2.2) against the laboratory test results are shown from triplicate analysis of
NMR content measured in 13.1.2.2(1). Fig. X1.2 shown an blends with 50 % and 20 % KEE in Table 1. The results show
example of a calibration curve. that with Method A the repeatability of the NMR methodology
(3) Note the slope of the calibration curve. It is used as a is good, with small relative errors ranging from –1.7 6 0.8 %
correction factor. for mixtures close to 50 % PVC and –2.8 6 0.1 % for mixtures
13.1.2.3 Calculations of PVC Content—Use Eq 3 to correct near 80 % PVC. With Method B, the relative error is reduced
the PVC mass from Eq 1 by means of the correction factor to +0.6 6 0.8 mixtures near 50 % PVC and –0.6 6 0.1 % for
from 13.1.2.2(3). mixtures near 80 % PVC. Given that the NMR quantification
method is developed first and foremost to quantify the KEE
PVC mass from Equation 1 * correction factor
PVC content ~%! 5 100* content of roofing compounds, the accuracy of Method 1 and
sample mass from Equation 1
Method 2 in this respect can be summarized as follows: (1)
(3) Method A provides KEE contents within about 11 % of the
Taking the same PVC and sample mass used earlier under
13.1.1.1 and the slope of the curve in Fig. X1.2, as an actual value, for example, 19.8 % vs. 22.0 %. (2) Method B
example, leads to provides KEE contents within 2.4 % of the actual value, for
example, 20.5 % vs. 21.0 %. In both cases, the accuracy is
0.011588 g * 1.0231
PVC content ~%! 5 100* 5 40.0 lowest for compounds with high PVC contents. For com-
0.029673 g
pounds with 50 % or less PVC, then the respective accuracies
13.1.2.4 Calculations of KEE Content—Calculate the re- in quantifying KEE are: (1) Method A: better than 2.5 %; (2)
vised KEE content from Eq 4. Method B: better than 1.3 %.
KEE percent content 5 100 2 percent PVC from Equation 3 (4) 15.2 Inter-laboratory Testing—To be determined.
4
D8154 − 17
TABLE 1 Composition Data for Nine Model PVC:KEE Blends Characterized by Quantitative Proton-NMR Method A and Method B
PVC KEE
model
PVC content (%) measured by: relative difference (%) KEE content (%) relative difference (%)
mixture A A
T A B A B T A B A B
1 39.9 39.1 40.0 –2.0 0.2 60.1 60.9 60.0 1.3 –0.2
2 49.0 48.3 49.4 –1.5 0.7 51.0 51.7 50.6 1.5 –0.7
3 49.9 49.4 50.6 –1.0 1.3 50.1 50.6 49.4 1.0 –1.3
4 49.8 48.6 49.7 –2.5 –0.2 50.2 51.4 50.3 2.5 0.2
5 58.8 57.8 59.1 –1.8 0.5 41.2 42.2 40.9 2.5 –0.7
6 69.3 68.3 69.9 –1.4 0.9 30.7 31.7 30.1 3.1 –2.1
7 80.1 77.8 79.6 –2.8 –0.5 19.9 22.2 20.4 11.2 2.2
8 79.5 77.2 79.0 –2.9 –0.7 20.5 22.8 21.0 11.3 2.6
9 80.2 78.0 79.8 –2.8 –0.5 19.8 22.0 20.2 11.3 2.2
average –2.1 0.2 average 5.1 0.2
standard 0.7 0.7 standard 4.7 1.7
dev. dev.
A
T: actual content.
APPENDIX
(Nonmandatory Information)
TABLE X1.1 Example of Calibration Samples Prepared as per 11.2.2 with their Target and Actual Compositions
Sample I.D. Target KEE : PVC Mass KEE Mass PVC TCNB mass Actual Content
% (mg) (mg) (mg) KEE PVC
% %
1 60 : 40 17.846 11.827 19.938 60.14 39.86
2 50 : 50 15.646 15.091 23.345 50.97 49.03
3 50 : 50 15.115 15.069 19.554 50.08 49.92
4 50 : 50 15.341 15.217 20.285 50.20 49.80
5 40 : 60 12.552 17.947 19.393 41.16 58.84
6 30 : 70 9.273 20.886 19.571 30.75 69.25
7 20 : 80 5.978 23.991 20.762 19.95 80.05
8 20 : 80 6.322 24.556 20.671 20.47 79.53
9 20 : 80 6.121 24.819 20.778 19.78 80.22
5
D8154 − 17
NOTE 1—The TCNB integral region from 7.80 to 8.40 ppm is set to 100, from which the PVC integral value between 4.12 to 4.75 ppm is obtained.
FIG. X1.1 Proton-NMR Spectra in THF-d8 for a Model 50:50 PVC:KEE Mixture (top) and a Polymer Blend Extracted from a Roofing Com-
pound with a PVC:KEE Ratio Close to 54:46 (bottom)
6
D8154 − 17
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/