Phosphorus, Reactive, TNT, PhosVer 3 Method 8048, 02-2009, 9th Ed
Phosphorus, Reactive, TNT, PhosVer 3 Method 8048, 02-2009, 9th Ed
Phosphorus, Reactive, TNT, PhosVer 3 Method 8048, 02-2009, 9th Ed
1. Enter the stored 2. Press: 82 ENTER 3. Insert the COD/TNT 4. Use a TenSette Pipet
program number for Adapter into the cell to add 5.0 mL of sample
The display will show
reactive phosphorus holder by rotating the to a Reactive
mg/L, PO4 and the
(PO43-), Test ’N Tube. adapter until it drops Phosphorus Test 'N Tube
ZERO icon.
into place. Then push Dilution Vial. Cap
Press: PRGM
Note: For alternate forms down to fully insert it. and mix.
The display will show: (P, P2O5), press the CONC
key. Note: A diffuser band Note: For samples with
PRGM ? covers the light path holes extreme pH, see the
on the adapter to give Interference section.
Note: For most accurate increased performance.The
results, perform a Reagent band should NOT be
Blank Correction using removed.
deionized water (see
Section 1).
* Procedure is equivalent to USEPA Method 365.2 and Standard Method 4500-P E for wastewater.
481
PHOSPHORUS, REACTIVE, continued
5. Clean the outside of 6. Place the sample vial 7. Tightly cover the 8. Press: ZERO
the vial with a towel. into the adapter. sample vial with the
The cursor will move to
instrument cap.
Note: Wiping with a damp Push straight down on the right, then the
towel, followed by a dry the top of the vial until it display will show:
one, will remove seats solidly into the
fingerprints or other marks. adapter. 0.00 mg/L PO4
Note: For multiple
Note: Do not move the vial
samples, zero only on the
from side to side as this can
first sample. Read the
cause errors.
remaining samples after
adding the PhosVer 3
Reagent.
9. Using a funnel, add 10. Cap the vial tightly 11. Press: 12. Immediately after
the contents of one and shake for 10-15 the timer beeps, place
TIMER ENTER
PhosVer 3 Phosphate seconds. the sample vial in the
Powder Pillow to the A 2-minute reaction adapter.
Note: The powder will not
vial. completely dissolve. time will begin. Push straight down on
Note: Read samples the top of the vial until it
between 2 and 8 minutes seats solidly into the
after the addition of the adapter.
PhosVer 3 reagent.
Note: Do not move the vial
Note: A blue color will from side to side as this can
develop if phosphate is cause errors.
present.
482
PHOSPHORUS, REACTIVE, continued
Accuracy Check
Note: Clean glassware Standard Additions Method
with 1:1 hydrochloric acid a) Fill three 25-mL graduated mixing cylinders with 25 mL of
solution. Rinse again with sample.
deionized water. Do not
use detergents containing
b) Snap the neck off a Phosphate PourRite Ampule Standard,
phosphates to clean
glassware. 50 mg/L as PO43-.
c) Use the TenSette Pipet to add 0.1 mL, 0.2 mL and 0.3 mL,
respectively, to the three 25-mL aliquots of sample
prepared in step a. Mix well.
483
PHOSPHORUS, REACTIVE, continued
d) Analyze each sample as described in the procedure; use
5.0 mL of the prepared standard additions for each test.
The concentration should increase as follows: 0.2 mg/L,
0.4 mg/L, 0.6 mg/L
PO43-, respectively.
484
PHOSPHORUS, REACTIVE, continued
Interferences
The following may interfere when present in concentrations
exceeding these listed below:
485
PHOSPHORUS, REACTIVE, continued
Summary of Method
Orthophosphate reacts with molybdate in an acid medium to
produce a phosphomolybdate complex. Ascorbic acid then
reduces the complex, giving an intense molybdenum blue color.
REQUIRED REAGENTS
Reactive Phosphorus Test ’N Tube Reagent Set.................................50 tests ............. 27425-45
Includes: (1) 21060-46, (50) Orthophosphate Dilution Vials*
Quantity Required
Description Per Test Unit Cat. No.
PhosVer 3 Phosphate Reagent Powder Pillows ..................1 ..............50/pkg ............. 21060-46
50 Orthophosphate Test ‘N Tube Dilution Vials ................1 ..............50/pkg.................................. *
REQUIRED APPARATUS
COD/TNT Adapter .............................................................1 ..................each ............. 48464-00
Funnel, micro ......................................................................1 ..................each ............. 25843-35
Pipet, TenSette, 1 to 10 mL.................................................1 ..................each ............. 19700-10
Pipet Tips, for 19700-10 TenSette Pipet .............................1 ..............50/pkg ............. 21997-96
Test Tube Rack ............................................................... 1-3 ..................each ............. 18641-00
OPTIONAL REAGENTS
Bromine Water, 30 g/L......................................................................... 29 mL ............... 2211-20
Drinking Water Standard, Inorganic, F–, NO3–N, PO43–, SO42–.... .............500mL.......... 28330-49
Hydrochloric Acid Standard Solution, 6.0 N (1:1) ............................ 500 mL ................. 884-49
Phenol Solution, 30 g/L ....................................................................... 29 mL ............... 2112-20
Phosphate Standard Solution, 1 mg/L as PO43- .................................. 500 mL ............... 2569-49
Phosphate Standard Solution, Voluette ampule,
50 mg/L as PO43-, 10 mL ...............................................................16/pkg ................. 171-10
Phosphate Standard Solution, PourRite ampule,
50 mg/L as PO43-, 2 mL .................................................................20/pkg .............. 171-20H
Wastewater Effluent Standard, Inorganic
(NH3–N, NO3–N, PO4, COD, SO4, TOC)......................................500 mL.......... 28332-49
Water, deionized........................................................................................ 4 L ................. 272-56
486
PHOSPHORUS, REACTIVE, continued
OPTIONAL APPARATUS
Ampule Breaker, Pour Rite (2-mL ampule)............................................ each..............24846-00
Ampule Breaker Kit ................................................................................ each..............21968-00
Aspirator, vacuum ................................................................................... each................2131-00
Cylinder, graduated, mixing, 25 mL (3 required) ................................... each..............20886-40
Dispenser, Repipet Jr., 2 mL ................................................................... each..............22307-01
Filter Holder, 47 mm, 300 mL, graduated............................................... each..............13529-00
Filter, membrane, 47 mm, 0.45 microns ........................................... 100/pkg..............13530-00
Flask, filtering, 500 mL........................................................................... each..................546-49
Flask, volumetric, Class A, 100 mL........................................................ each..............14574-42
pH Indicator Paper, 1 to 11 pH units ............................................. 5 rolls/pkg..................391-33
pH Meter, sension™1, portable with electrode ...................................... each..............51700-10
Pipet, TenSette, 0.1 to 1.0 mL ................................................................. each..............19700-01
Pipet Tips, for 19700-01...................................................................... 50 pkg..............21856-96
Pipet Tips, for 19700-01 ...................................................... .......... 1000/pkg..............21856-28
Pipet Filler, Safety Bulb .......................................................................... each..............14651-00
Pipet, volumetric, Class A, 5.00 mL ....................................................... each..............14515-37
Pipet, volumetric, Class A, 2.00 mL ....................................................... each..............14515-36
487