Professional Documents
Culture Documents
Syngistix 2.2 For ICP MS Software Guide
Syngistix 2.2 For ICP MS Software Guide
2
Software Reference Guide
April 2017
Release History
Release Publication Date
Any comments about this documentation should be addressed to User Assistance at:
PerkinElmer
710 Bridgeport Avenue
Shelton, CT 06484-4794
U.S.A.
Or emailed to: info@perkinelmer.com
Notices
This document is provided to customers who have purchased PerkinElmer, Inc. software. Software that
may be described in this document is furnished under a license agreement. It is against the law to copy,
modify, or distribute the software on any medium, except as specifically allowed in the license agreement.
Furthermore, the license agreement may prohibit the software from being disassembled, reverse
engineered, or decompiled for any purpose.
This document is controlled under the U.S. Export Administration Regulations. Unauthorized copy,
disclosure or use is strictly prohibited.
The information contained in this document is subject to change without notice.
Except as specifically set forth in its terms and conditions of sale, PerkinElmer, Inc. makes no
warranty of any kind with regard to this document, including, but not limited to, the implied
warranties of merchantability and fitness for a particular purpose.
PerkinElmer, Inc. shall not be liable for errors contained herein for incidental consequential damages in
connection with furnishing, performance or use of this material.
Copyright Information
This document contains proprietary information that is protected by copyright. Registered names,
trademarks, etc. used in this document, even when not specifically marked as such, are protected by law.
All products and company names mentioned herein may be the trademarks of their respective owners.
All rights are reserved. No part of this publication may be reproduced in any form whatsoever or translated
into any language without the prior, written permission of PerkinElmer, Inc.
AXIAL FIELD, DRC, DYNAMIC REACTION CELL, ENHANCED SECURITY, FIAS, PKI, QID,
REPORTER, SMARTTUNE, SMARTTUNE EXPRESS, SYNGISTIX, TOTALQUANT, UCT, and
UNIVERSAL CELL TECHNOLOGY are trademarks of PERKINELMER, Inc. or its subsidiaries.
PerkinElmer
710 Bridgeport Avenue
Shelton, CT 06484 USA
is ISO 9001 registered.
© 2017 edition PerkinElmer, Inc. PKI.
Documentation produced in Canada.
Software release date: April 2017
Syngistix for ICP-MS Software Reference Guide compiled April 11, 2017: DPS
Table of Contents
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Using this Software Reference Guide . . . . . . . . . . . . . . . . . . . . . . . .15
Additional Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Using the Help System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Contents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Search . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
To perform a search . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Favorites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
3
Software Reference Guide
4 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Table of Contents
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 5
Software Reference Guide
6 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Table of Contents
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 7
Software Reference Guide
8 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Table of Contents
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 9
Software Reference Guide
10 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Table of Contents
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 11
Software Reference Guide
Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 401
12 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Preface
This Software Reference Guide contains in-depth conceptual information about
the use of the application software, as well as software reference information
regarding screen components. This guide includes information for multiple
configurations of the NexION® ICP-MS instrument. Except where otherwise
specified, all functions are available for use on all instruments.
Additional Assistance
The system includes the following documentation and help materials:
• Installation Instructions containing comprehensive installation instructions.
This guide is available in PDF format in the installed program directory.
• A Help System providing task-level information for all software procedures.
This is available from the Syngistix ball menu Help submenu.
• A Maintenance Guide containing a detailed description of the instrument
hardware system, and instructions for installation, maintenance, and
troubleshooting.
• A Consumables and Supplies Reference Guide providing a list of the
cones, solutions, and other consumable items used by the system. This is
available from the Syngistix ball menu Help submenu.
• A Safety Manual providing important notices and warnings to guide you
through the safe operation and maintenance of the instrument.
• A Maintenance Log containing maintenance spreadsheets in PDF format.
15
Software Reference Guide
Contents
The Contents tab arranges help topics by subject, just like the Table of Contents in
a book. Click a book icon to display an overview of the topics in that section. Click
a topic icon to display a particular topic.
Index
Use the Index tab to search the help index for a word. Type the word you want to
look up in the text box and click Display, or click the topic in the alphabetical list.
The chosen topic is displayed.
Search
Use the Search tab to locate a word, phrase, or set of words. The help system
uses boolean, wildcard and nested expressions to refine a search. With the AND,
OR, NOT, and NEAR operators you can precisely define your search by creating a
16 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Preface
Note: The |, &, and ! characters do not work as boolean operators — you must use OR,
AND, and NOT.
To perform a search
1. Type the words or phrase you want to search for in the text box and click
List Topics. The topics containing the word or phrase are displayed,
including a Rank to help you choose the topic you want to view first.
2. Select a topic from the list and click Display. The topic you have chosen is
displayed in the main panel.
Favorites
Use the Favorites tab to quickly find frequently used help topics. You can store the
current help topic in a Favorites list by clicking Add at the bottom of the Favorites
tab, then return to the topic anytime by double-clicking the title on the tab.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 17
Software Reference Guide
18 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Chapter 1
19
Software Reference Guide
20 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Introduction to the Software
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 21
Software Reference Guide
22 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Introduction to the Software
Screen Components
• Syngistix ball menu: Accessed via the Syngistix ball icon in the upper left
corner of the screen, the Syngistix ball menu provides access to all your
Syngistix software administrative, configuration, and file management
commands. Functions include Workspace management; Event History
access; Service Mode access; Systems Tools; Help, Documentation, and
Video tutorial access; Print functions; and access to the Options
configuration dialog box. Not all functions are available all the time; some
are unavailable during particular processes.
If you are using the Enhanced SecurityTM software, the Set Current Project
Folder command is also available here.
• Measurement Status dialog box: The Measurement Status dialog box
appears onscreen whenever an analysis is in progress. It provides real
time updates on the progress of each individual measurement in the
analysis. See the Measurement Status Dialog Box on page 250.
• Status bar: The status bar displays information such as running mode
(working or service mode), plasma and ignition status, faults, and more.
If you are using the Enhanced SecurityM software, the status bar also
displays, in the bottom right corner, the full name of the user currently
logged into the system.
For UCT instruments, the status bar also displays the profile and analytic
state or mode in which the instrument is operating — Standard, KED, or
DRC mode. When the instrument is in the process of changing mode, the
status bar displays a message describing the change in progress, together
with the time remaining. Operations that cause delays, such as cell gate
changes or cell gas flow changes, also cause status messages to appear.
• Workspaces: In the Syngistix software, a Workspace is a discrete set of
working files and configurations saved together for a particular purpose.
The fileset for a given workspace can be reviewed in the Review dialog
box (see Review Files Dialog Box Software Reference on page 50) and
can include the following files and specifications:
Method file (.mth)
Dataset location
Sample (.sam)
Report Template (.rop)
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 23
Software Reference Guide
The arrow submenu provides access to a Mass Calibration view, which displays
MassCal and Realtime panels for controlling the automated spectrometer
hardware tuning process used to adjust mass calibration and resolution
The LogBook icon accesses the LogBook screen, where you can review
performance check data to track or troubleshoot issues or trends in your system
24 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Introduction to the Software
Analyze group
Icon Function
The Method icon accesses the Method screen, where you develop analytical
methods for performing determinations, and set up and manage QC (Quality
Control) parameters. This view includes the Method and Periodic Table panels
The arrow submenu provides controls for managing your method:
Sort — Click to define the order of the analyte entries on the Method panel
Timing tab. When you click this command, a list of sort options appears. To use
this command, click Sort, and then click a sort option
Define Group — Click to define an internal standard group in a quantitative
analysis method. To use this command, select the rows containing the analytes
you want to put into an internal standard group, and click Define Group. A
bracket appears in the Int Std column to indicate that the selected analytes are
now part of an internal standard group. This option is available only for the
Method panel Timing and Calibration tabs
Remove Group — Click to remove an internal standard group from a quantitative
analysis method. To use this command, click any analyte belonging to the group
you want to delete, and click Remove Group. The internal standard group is
deleted and the bracket in the Int Std column is removed. This option is available
only for the Method panel Timing and Calibration tabs
Set Internal Std — Click to specify the analyte that will serve as the internal
standard in an internal standard group. To use this command, you must first
specify an internal standard group using the Define Group command. With an
internal standard group defined, click the analyte that is the internal standard,
then click Set Internal Std. An arrowhead appears in the Int Std column to
indicate that the selected analyte is now designated as an internal standard. This
option is available only for the Method panel Timing and Calibration tabs, for
quantitative analysis methods
The Sample icon accesses the Samples screen, where you enter sample
information and control manual and batch determinations. The Sample screen
contains the Sample panel, as well as instances of the Report View and Realtime
panels, allowing you to monitor your results in graphic and tabular
form as they are run
The Dataset icon accesses the Dataset screen, where determination results are
stored, including method information and operating conditions
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 25
Software Reference Guide
Results group
Icon Function
The Reporter icon accesses the Report View screen, where you view detailed
determination results for the current sample, and scrolling data for each sample
in your batch
The arrow submenu also provides controls for managing QC (Quality Control)
functions. The following commands are available:
Restart QC ‐‐ Click to restart a quality control analysis that has stopped due to a
QC action
Clear QC Reagent Blank — Click to clear reagent blank data that is stored as part
of a QC analysis
Clear QC Logging Data — Click to clear all QC logging data that is stored with the
dataset
The Realtime icon accesses the Realtime graphics screen, where you can view
the raw data from an acquisition in real time
The Interactive icon accesses the Interactive graphics screen, where you can
view and analyze acquired data
The arrow submenu provides controls for managing calibration functions. The
following commands are available:
Clear Calibration — Click to erase any calibration information that is active in the
Calibration View panel and which will be used in any determinations or
reprocessing calculations
Clear Blank — Click to clear any signal data that is stored for the sample blank
Clear Calibration and Blank — Click to clear both calibration information and
blank reading data that would be used in completing a determination or
performing reprocessing
Workflow group
Icon Function
The Scheduler icon accesses the Scheduler screen, where you manage and
run routine tasks. Use the scheduler function to set up an automated run list of
daily tasks
26 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Introduction to the Software
Click to review all the active files to be used in an analysis. The Review Files
dialog box lists all the files related to the active or most recently active
workspace
Applications tab
Icon Function
The Nano App icon opens the Nano App software. Available only if the Nano
App is installed, and the instrument is not in a reprocessing session
The Nano App ICP‐MS nano analysis tool allows for the characterization of single
nanoparticles using the Syngistix software system. It is purchased separately
from the Syngistix for ICP‐MS base software, and must be installed on a
computer where the base software is already installed
The Single Cell icon opens the Single Cell App software. Available only if the
Single Cell App is installed, and the instrument is not in a reprocessing session
The Single Cell App performs ultra‐rapid, continuous analysis of discrete pulses to
foster the complete ionization of individual cells in a plasma in order to
determine the levels of metal particulate or dissolved metals within a given
sample
The Spotfire icon opens the Spotfire® data visualization module. Available
only if the Spotfire software is installed, and the instrument is not in a
reprocessing session
TIBCO Spotfire data visualization software is a dynamic, collaborative interface
that assimilates data from multiple sources — chemical structures, text,
numbers, images, chemical properties, biological assays, and more — and
empowers the user to perform complex analyses and create intuitive and
informative visual dashboards
Nano App ribbon
Icon Function
The Analysis icon accesses the Nano Analysis screen, which provides all the
controls needed to set up and run nanoparticle calibrations and sample
determinations, with realtime results panels for parsing your data on the fly
The Results icon accesses the Nano Results screen, where you can review in
detail the calculated histograms and related statistics calculated for each nano
sample or analysis. Here you can also reprocess your data by modifying the
variables used for these calculations
The Exit Nano icon returns you to the main Syngistix interface.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 27
Software Reference Guide
Single Cell App ribbon
Icon Function
The Analysis icon accesses the Single Cell Analysis screen, which provides all the
controls needed to set up and run calibrations and sample determinations, with
results panels for parsing your data in real time
The Results icon accesses the Single Cell Results screen, where you can review in
detail the calculated histograms and related statistics calculated for each single
cell sample or analysis. Here you can also reprocess your data by modifying the
variables used for these calculations
The Exit Single Cell icon returns you to the main Syngistix interface.
28 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Introduction to the Software
Note: This example shows the software for the 2000S instrument, which contains some advanced
functionality, but the software components largely look and function the same across all models.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 29
Software Reference Guide
File Locations
The software file management features include a specific folder setup for data,
with several subfolders unique to the software. These subfolders are:
• Autosampler: This subfolder contains the files for the different types of
autosampler that can be used with the instrument, including the ESI series
of autosamplers and prepFAST autodiluter; the PerkinElmer AS-XX series;
and the ASX-500 and ADX-500 systems.
• Conditions: This subfolder contains conditions .tun files. The information
contained in these files is used to perform the fine-tuning adjustments for
the hardware subsystems in the torch/quadrupole assembly of the
instrument.
• Dataset: This subfolder comes with a set of default data folders, wherein
files containing the data from your analyses are collected. You can also
create a new Dataset folder elsewhere on your computer if there is a more
convenient place for you to store your analytic data.
• MassCal: This subfolder contains all pre-defined and custom .tun mass
calibration files.
• Method: This subfolder contains any method .mth files developed for
analyses. It also contains a Service subfolder for service-representative-
specific information.
• Nano: This folder contains a variety of additional subfolders for all data
and setup files relevant to the Nano App ICP-MS nano analysis tool. This
folder exists only if the Nano App has been purchased and installed.
• ReportOptions: This subfolder contains the report options .rop templates
used to create a hardcopy or file reports. You can modify a report template
for your needs, name it, and save it to this folder for future use.
• ReportOutput: This subfolder is the destination for report files if no path is
specified in the method.
• Sample: This subfolder contains the sample .sam files for sample
analysis. These files identify the samples, control the order in which
samples are measured and initiate the measurements. In the software,
you fill in the Sample panel and save a file, with a name you select, to this
subfolder.
• Single Cell: This folder contains a variety of additional subfolders for all
data and setup files relevant to the Single Cell App cellular analysis tool.
This folder exists only if the Single Cell App has been purchased and
installed.
• System: This subfolder contains the calibration files and TotalQuant™
response files.
• Wizard: This subfolder contains all pre-defined and custom SmartTune™
.swz files. It also contains a Service subfolder for service-representative-
specific information.
• Workspace: This subfolder contains all pre-defined and custom .wrk
workspace files. It also contains a Service subfolder for service-
representative-specific information.
30 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Introduction to the Software
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 31
Software Reference Guide
You can create multiple profiles that use the same gas; they will be assigned to
the same channel. If you create more than three profiles for any one mode, not all
profiles will be displayed at once; rather, a scroll bar will appear, which you can
use to scroll up and down through the profiles. Note that only those that are active
(that is, that have been assigned to an available physical gas channel) can be
used to run optimizations and analyses.
Note: The numbers entered for each profile set the base cell gas flow and rejection
parameter values for any new methods created using those profiles. At least one KED
mode profile and one DRC mode profile must have a cell gas flow set to a value greater
than zero.
A number of default profiles are shipped with the software. You can alter the
settings for these profiles, but you cannot delete them from the system. Custom
profiles that you create yourself may be deleted from the system by right-clicking
the relevant profile tile and then clicking Remove Profile.
The pre-defined Standard profile, Ammonia DRC profile (assigned by default to
cell gas channel A), and Helium KED profile (assigned by default to cell gas
channel B) are all required to run SmartTune Express optimizations; they must all
be present and actively assigned to their default channels in order to use this
feature, regardless of which mode you are optimizing for.
Acquisition order
When the system runs an analysis, analytes are processed in a particular order,
as determined by the mode of each, its RF Power setting, and its position on the
Acquisition Profile tab. The order is determined as follows:
• If the RF Power setting changes within a sample, this order is run for the
first RF group, and then repeated for each subsequent RF group.
• Profiles are processed from the top-down and left-to-right as the tiles
appear on the Acquisition Profile tab, so that Analytes linked to KED mode
profiles are processed first, followed by those using DRC mode, and then
Standard mode.
• The acquisition order with each mode is left to right. You can drag and
drop tiles to reorder them within each mode grouping, with the exception of
the pre-defined DRC Ammonia profile, which is always placed last within
the DRC group, and cannot be moved forward. (Remember that if you
exceed three profiles in any one mode grouping, the additional tiles will be
visible only by using the scroll bar that appears to the right of the group.)
32 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Introduction to the Software
Screentip
1 2 3
4 5 6
7 8
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 33
Software Reference Guide
analytes would be processed during an analysis. There are the four pre-defined
tiles that come with the system:
• Helium KED — This is assigned to channel B by default.
• Oxygen DRC — This is assigned to channel C by default.
• Ammonia DRC — This is assigned to channel A by default. Channel A is
connected to the hardware Getter for the removal of impurities in the gas.
• Standard — Predefined profile present on all UCT systems.
...and four custom tiles created by the user:
• Helium KED Custom QID — Custom QID profile with a turquoise border.
• Xenon KED unlinked — Profile that is both Unlinked and has a Custom
QID; purple border denotes that both options were selected.
• Oxygen DRC Custom QID— Custom QID profile with a turquoise border.
• Standard unlinked — Unlinked profile with a blue border. (Note that you
cannot create a custom QID in Standard mode.)
34 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Introduction to the Software
Configuration tab
Use the Configuration tab to define basic system parameters such as deadtime
corrections, gas flow controls, and notification sounds.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 35
Software Reference Guide
In Service mode, you can adjust this parameter so that the peaks being
scanned move into a DAC range where they can be properly resolved.
Click Set to open the Calibration Factor dialog box and change the value.
Acceptable values are between 700 000 and 900 000 ppm.
• Audible Notifications: Select or clear the check boxes as required:
Enable Log Event Sound: When selected, the computer beeps
whenever a message is written to the event log.
Enable Fault Alarm: When selected, the computer beeps every five
seconds whenever a fault condition occurs. The beeping continues
until the fault is corrected.
Enable End of Analysis Sound: When selected, the computer beeps
whenever the computer finishes an analysis.
• Diagnostics Tab — Service Only: (Service mode only) Displays the
Diagnostics Refresh Rate. In the Refresh Rate field, type the frequency, in
seconds, at which the Instrument panel Diagnostics tab should refresh the
displayed information.
• Vacuum Subsystem: Displays the vacuum manifold type currently
installed on the instrument: the available options are (Version 1)
36 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Introduction to the Software
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 37
Software Reference Guide
38 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Introduction to the Software
Note: Keep in mind that the profile name will be used elsewhere in the
system, and that a short, meaningful name will be easier to read and
understand in these cases.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 39
Software Reference Guide
Note: If you leave the wash override parameters in the method empty, the software uses
the wash value specified during acquisition. If you set a wash override value, it overrides
that specified in the method. This lets you specify longer washes to prevent residue when
running high-concentration samples.
Figure 1-6 Syngistix Options dialog box Peristaltic Pump Defaults tab
Default Peristaltic Pump parameters
Sample Sample Read delay Delay and Wash
Wash (sec)
flush (sec) flush speed (sec) analysis speed speed
2000 series instruments
35 -42 15 -35 45 -42
300/350 series instruments
35 -24 15 -20 45 -24
40 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Introduction to the Software
Note: In this release, only some of the available options are editable.
Figure 1-7 Syngistix Options dialog box Performance Checks Defaults tab
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 41
Software Reference Guide
2000 Instruments
Std mode KED mode
Sensitivity (2000S instruments) Sensitivity
Be > 6 000 cps/1 ppb Co high > 25 000 cps/10 ppb
In > 110 000 cps/1 ppb Background
U > 80 000 cps/1 ppb Ar278 high < 30 cps
Sensitivity (2000C, B, & P) Kr83 high < 300 cps*
Be > 4 500 cps/1 ppb
In > 80 000 cps/1 ppb Oxides
U > 60 000 cps/1 ppb ClO‐high/Co‐high < 0.5%
Background ClO‐low/Co‐low < 2%
< 3 cps @mass 220
Doubly Charged Ions and Oxides
Ce++/Ce < 3% *If 83Kr > 300cps, the argon gas supply
contains high levels of Krypton gas. This
CeO/Ce < 2.5% amount of Krypton is sufficient to degrade
detection limits enough that you will not meet
DRC mode the 78Se detection limit specification and
<20cps @ mass 78 target background. As and
Sensitivity (2000S instruments) Se are performed with low flow conditions and
Fe > 60 000 cps/1 ppb V is performed with high flow conditions.
Sensitivity (2000C, B, & P)
Fe > 50 000 cps/1 ppb
42 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Introduction to the Software
300/350 UCT Instruments
Std mode KED mode
Sensitivity (300/350S instruments) Sensitivity (all UCT instruments)
Be > 4 000 cps/1 ppb Co high > 25 000 cps/10 ppb
In > 55 000 cps/1 ppb
Background
U > 35 000 cps/1 ppb
Ar278 high < 30 cps
Sensitivity Kr83 high < 300 cps*
(300/350X & D instruments)
Oxides
Be > 2 000 cps/1 ppb
ClO‐high/Co‐high < 0.5%
In > 40 000 cps/1 ppb
ClO‐low/Co‐low < 2%
U > 30 000 cps/1 ppb
CeO‐high/Ce‐high < 1%
Background
< 1 cps @mass 220
Doubly Charged Ions and Oxides *The Kr‐high criterion is for information
CeO/Ce < 2.5% purposes only, and has no pass or fail status
attached to it. If Kr83 > 300 cps, the Argon
Ce++/Ce < 3% contains enough Krypton to affect the Se78
DRC mode detection limit.
Sensitivity (300/350S instruments)
Fe > 40 000 cps/1 ppb
Sensitivity (300/350D instruments)
Fe > 25 000 cps/1 ppb
300/350Q Basic Quadrupole Instruments
Sensitivity
Be > 2 000 cps/1 ppb
In > 40 000 cps/1 ppb
U > 30 000 cps/1 ppb
Background
< 5 cps @mass 220
Doubly Charged Ions and Oxides
CeO/Ce < 2.5%
Ce++/Ce < 3%
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 43
Software Reference Guide
Note: If multiple users have been set up on the instrument and they do not share
project folders, files such as default.tun and default.dac are not shared between
users and the system administrator.
44 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Introduction to the Software
2. Check that the sample tubing and drain tubing leading from the peristaltic
pump to the spray chamber are correctly connected and in good working
condition.
3. If the pump tubing is new, gently stretch it. Attach the clips on the tubing to
the tubing stops. If the pump tubing has been used previously, check its
condition. Replace the tubing if it is discolored, brittle, crushed, or clogged.
Replace the tubing regularly, after every eight hours of use, or as needed.
If you routinely use organic solvents, you will require special tubing, as the
standard tubing is attacked and degraded by solvents. See the
Maintenance Guide for instructions on replacing peristaltic pump tubing.
4. Check that the pump rollers are clean and move freely. Replace the tubing
clamps for each channel and swing the cam levers over to apply tension to
the clamps.
5. Place the capillary tubing into a container of deionized water.
6. Check that the green system ready LED on the instrument control panel is
lit and steady.
If not, there may be a fault in the system and the plasma cannot be ignited.
If this occurs, check the diagnostic display to find which system
component is causing a fault.
7. On the Control screen Devices panel ICP-MS tab, click the Plasma Start
button to ignite the plasma.
The instrument initiates the ignition process — the progress of the ignition
sequence appears in the status bar. If the plasma does not ignite, wait 15
seconds and click the Plasma Start button again.
8. After the plasma ignites, let the instrument warm up for 45 minutes.
To start the instrument after an extended shutdown
This section presents additional information for restarting the instrument after a
period of more than two or three weeks. When the power and vacuum pumps are
off, the system requires additional time to initialize and pump down. For more
information about turning the instrument power on and off, see the Maintenance
Guide.
1. Plug the instrument and roughing pump into the power source, and turn on
the installed gases and the cooling water supply, chiller, or cooler.
2. 2000S SEMI instruments If the instrument is equipped with the optional
red Emergency Off (EMO) button, ensure that the EMO button is pulled out,
and then reset the circuit breakers. Then press the green Power On
button located on the left side of the instrument.
3. Ensure that the software is running.
4. Turn on the Instrument circuit breaker.
5. In the software, access the Control screen System View panel.
A stylized system layout provides a graphic display of the major operating
elements of your instrument, indicating the status of the plasma, argon
flow, cooling system, and interlocks.
6. On the Devices panel ICP-MS tab, click the Vacuum start button.
When a satisfactory vacuum is achieved, the vacuum status bar turns
green and the related status reads Ready, indicating that all the system
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 45
Software Reference Guide
hardware is operating properly and that the plasma can be ignited. If the
indicator reads Not Ready, check the schematic diagram to find out which
system is not prepared for operation.
7. Check the base vacuum pressure. For successful operation of the
instrument, the pressure should be in the 1x10-7 to 5x10-7 Torr range.
8. Turn on the RFG (RF Generator) circuit breaker.
9. Record the instrument readings in the appropriate section of the
instrument Maintenance Logs.
10. Turn on any peripheral devices connected to the instrument, such as
alternate sampling systems (flow injection, laser ablation, and so on).
Note: To guarantee that all standards have been met before the system is
turned on, refer to the manuals provided with each peripheral device.
11. Ensure that the green System Ready LED on the instrument is steady.
The instrument is ready for use.
Ignition Problems
The system indicates the progress of the ignition sequence on the Control screen.
If the plasma fails to ignite due to a component failure, an error message appears
together with a diagnostic display indicating which system component is causing
the failure. The following system interlocks must be satisfied before the plasma
can be ignited:
• The torch gas pressure must be achieved
• The vacuum chamber must be locked in the sampling position
• The operating vacuum must be achieved
• The ICP generator must be ready
Additional information is located on the System View panel Diagnostics tab, which
lists operating conditions for the component selected in the system graphic. The
following are a series of checks to assist in correcting failure conditions.
Argon Supply
Check the argon supply pressure and record the pressure on the Maintenance
Log sheet. Ensure that you have enough argon for the amount of time you plan to
use the instrument. For information about argon gas requirements, see the
Maintenance Guide.
• Check that the argon supply valve is open and the regulator is set at the
correct pressure.
• Check for leaks throughout the gas connections.
Cooling System
The cooling system must meet the required specifications and be installed
according to the information in the Maintenance Guide.
• Check that the cooling system is connected at the appropriate points on
the instrument.
• Check that all electrical connections have been made, and check the
external pressure.
46 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Introduction to the Software
Note: For 2000S SEMI instruments with EMO options kits only In an emergency,
you can extinguish the plasma by pressing the red override button on the
instrument.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 47
Software Reference Guide
5. Release the compression cams that secure the drain and sample tubing.
Loosen the drain and sample tubing. This increases the usable life of the
tubing.
Note: If you use your instrument on a daily basis, leave the system on with the
vacuum pumping. This permits quick startup the next time you want to run
analyses. When shutting down the instrument for more than two weeks, refer to
the disconnection procedures in the Maintenance Guide.
Task Scheduler
Use the Scheduler function to set up an automated run list of daily tasks. You can
automatically start and stop the plasma, and run optimizations and analyses
unattended. You can add multiples of any task type to your Scheduler list. The
following tasks are available:
• AutoStart: Use to start the plasma. Note that if the plasma is not on and
the AutoStart™ option is not selected, and the Scheduler is started, the
software alerts you with a Failed status.
• Warm-up: Use to prepare the instrument for analyses following plasma
ignition.
• SmartTune / SmartTune Express: Use to run performance checks and
optimizations. Ensure that all SmartTune files are correctly configured and
saved before adding them to the task list.
• Wait: Use this option to insert a pause into your run; for example, you
might add a Wait function after a SmartTune optimization, so that you can
check to see that the optimization passed all criteria before proceeding to
run your analyses. The wait function pauses your Scheduler workflow until
you start it again. You can specify a peristaltic Pump Speed for the
duration of the wait period.
• Analyze Samples: Use to add one or more sample or batch runs to the
task list. Ensure that all methods, samples, and sample batch files are
correctly configured and saved before adding them to the task list.
Note: When running a batch through the Scheduler function, it can take a
long time for the Measurement Status dialog box to appear. Click outside
of the Run List to reassert focus and hasten the process.
• Wash: Use to clean the system after an optimization or between sample
batches to prevent contamination.
48 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Introduction to the Software
• AutoStop: Use to shut down the plasma after all samples have been run.
Note: You can leave the Scheduler screen while it is running — the Scheduler status bar
provides task progress.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 49
Software Reference Guide
• Reset: Click to refresh the Status column, so you can start a new
Scheduler instance or list.
• Edit List: Click to access the Scheduler Edit List. Use this dialog box to
add and order tasks for your Scheduler list. Here you can select the tasks
you want the instrument to perform, and use the Move Up and Move Down
buttons to put them in the appropriate order.
50 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Introduction to the Software
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 51
Software Reference Guide
52 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Chapter 3
53
Software Reference Guide
54 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Instrument and Device Control
The XY plot and Z-depth torch graphic components are live; they can be
controlled via the Torch Position tab below, and reflect changes made via
the software optimization functions.
Diagnostics Tab
Use the Diagnostics tab to view the status and operating conditions for
checkpoints in the cooling, vacuum, gas, and plasma systems. Click on a section
of the instrument graphic, or one of the subsystem icons located below the
graphic, to view the diagnostics parameters for that component or subsystem.
Double-click any component to view its details.
Note: A dedicated Instrument Diagnostics Panel is also available via the Ribbon
Diagnostics icon to provide comprehensive diagnostics information for in-depth
troubleshooting.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 55
Software Reference Guide
Maintenance Tab
Use this tab to track usage and basic maintenance of key instrument components.
The system provides warnings and alerts to let you know when it is time to
inspect, clean, or replace a component. These reminders are configured here and
displayed in the System Status panel. Click any component to view or configure
settings; enable or disable an alert; or to view component statistics (available for
the ICP Power Tube (300/350 instruments only), Vacuum Gauge, and Detector
56 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Instrument and Device Control
only). You can also create custom maintenance reminders to your system to track
component life for tasks important to your lab.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 57
Software Reference Guide
Right-click anywhere in the maintenance table and click Add to create a custom
reminder. Note that Calendar Time (the number of days) is the only available
criteria for custom maintenance reminders
Note: The X position is somewhat dependent on the Y position due to the circular nature
of the possible torch path.
58 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Instrument and Device Control
Figure 2-6 Control screen System View panel Torch Position tab
Items on the Torch Position tab
• Description: This column describes the parameter.
• Status: This column indicates the current status or parameter for a given
parameter.
• Step: Type a step value, in millimeters, to indicate how big the intervals
between value settings should be.
• Value: The Value column indicates the current position for a given
parameter. Use the arrow controls to modify the current position in mm.
For the horizontal and vertical (X and Y) axes, the position is charted
visually on the XY graphic above. For the Z axis, torch depth is charted
visually on the Torch graphic above.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 59
Software Reference Guide
60 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Instrument and Device Control
• Vacuum Start/Stop: Use this toggle control to start and shut down the
vacuum system. Generally, you should leave the vacuum system on
unless you will not be using the instrument again for a period of three days
or longer. When you shut down the vacuum, the instrument takes
significantly longer to return to running pressure the next time you start it.
The status indicator bar is blue when the system is pumping, and the
vacuum is attempting to reach the minimum value. If the bar is yellow, it
indicates that the vacuum is on, but the high voltage power supply is off.
• Cone Access Open: Use this control to open the cone access door.
When fully open, the system is in service position; when fully closed, the
system is ready for XYZ calibration and general operation.
Note: While the Cone Access door can be opened via the software controls on
this panel, it must always be closed manually at the instrument. This is a
deliberate safety measure, designed to ensure that care is taken when closing the
door, so that no object or person is inadvertently trapped or pinched when the
internal mechanism slides into place and the door latches.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 61
Software Reference Guide
Pump Tab
Use this tab to control either the integrated peristaltic pump, or any external
peristaltic pump connected to your instrument.
62 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Instrument and Device Control
Autosampler Tab
Use this tab to select the type of autosampler used, the communications port, and
the autosampler tray.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 63
Software Reference Guide
64 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Instrument and Device Control
Probe Purge Pos.: This is used when the software is initializing the
autodiluter. In order to accurately measure the volumes of liquid
involved in each dilution, the dilution probe and associated tubing must
be full of diluent before any dilutions are done. To accomplish this, 10
mL of diluent are pumped into the dilution probe as part of the
initialization. To contain any excess diluent you can specify the
position of the large tube at the back right.
Selection of an autodiluter also activates autodilution functions in the
Sample Batch and Quality Control management sections of the software.
Slide the tube positioning scroll bars to select the X, Y, and Z coordinates
for the tube location. The corresponding tube number appears below the
scroll bar.
• Go to XYZ location: Moves the autosampler probe to the selected X, Y,
and Z coordinates.
• Go to rinse: Moves the autosampler probe to the wash solution.
• Go to standby: Click to move the probe to the standby position—the rinse
position with the probe raised. This is useful when you want to change the
probe or load a new tray.
• Interactive tray graphic: Use the interactive tray to move immediately to
the selected position.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 65
Software Reference Guide
FIAS Tab
This tab is displayed only if a FIAS system is installed and enabled, and selected
in an active method. Use this tab to establish communication with and manually
run the FIAS™ program without making measurements. Note that these controls
cannot execute a Read command, even if the FIAS step contains the Read.
66 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Instrument and Device Control
• Run Current Step: Runs the step. The software performs the step where
the cursor is located. There must be an entry in the Reference A/S
Location field to run this step.
• Run Current Program: Performs the FIAS program in the active method.
This will not execute a Read command. There must be an entry in the
Reference A/S Location field to run this program.
PC3 Tab
This tab is displayed only if a PC3 Peltier cooler is installed and enabled.Use this
tab to start and control the Peltier PC3 cooler.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 67
Software Reference Guide
• Plot: Displays the progress of the Peltier PC3 cooler, tracking temperature
in degrees Celcius against time in minutes.
• Current Temperature: This field displays the current temperature at the
Peltier cooler output in degrees Celcius.
• Setpoint Temperature: Type the desired temperature for the Peltier
cooler output in degrees Celcius, and then click Set.
• Set: In the Setpoint Temperature field, type the desired temperature for
the Peltier cooler output in degrees Celcius, and then click Set.
• Automatically turn PC3 on/off with plasma: Select this check box to
automatically turn the Peltier cooler on when the plasma is ignited, and off
when the plasma is extinguished.
Note: UCT instruments only. The Cell Voltages filter displays parameters related to the
DRC system. Many of these parameters are linked, so that if you change the value of one
parameter, the software automatically adjusts the value of one or more additional related
parameters.
68 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Instrument and Device Control
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 69
Software Reference Guide
70 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Instrument and Device Control
Getter Regeneration
UCT instruments only. You must regularly regenerate, or clean, the getter in the
DRC gas assembly to purge it of impurities from DRC gases. In most cases, you
should plan to run this process overnight or on the weekend, as regeneration
takes approximately eight hours to complete. You cannot run KED or DRC mode
methods during the regeneration process. Plasma ignition is not required for this
process.
The Getter Regeneration Wizard leads you through the regeneration process,
providing instructions and progress messages along the way.
Note: The getter is always installed on the physical gas channel A; it will never be
installed on a different channel.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 71
Software Reference Guide
72 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Chapter 2
73
Software Reference Guide
Key features
Security
• Administrators of the Syngistix ES software can configure Syngistix users
and assign them to different groups with each group assigned a specific
set of permissions. Your ES administrator defines these groups and
assigns permissions based on the needs of your users and lab
environment. Along with the use of passwords to gain access to the
software, these features permit overall access to be managed
appropriately in a regulated environment.
• The system administrator can determine whether different users can
normally access the same files such as method, mass calibration,
conditions, sample, and report files. Files can be located on the instrument
computer or on other computers connected to the controller across the
network.
Traceability
• Significant actions performed by the user are recorded in the Syngistix
Audit Trail with the date and time of the action, what was done, the name
of the user, and the reason the action was performed, together with any
relevant comments. The Syngistix Audit Trail can be filtered by several
data fields and then viewed onscreen or printed.
• New e-signature and file review functions can capture user credentials,
action reasons, and comments at every stage of the workflow for all key
files in the system, from the development of methods through sample
batch quality control. Challenge dialogs can collect e-signature information
every time a file is created, modified, or reviewed. Together with the
version control functions, this ensures complete traceability for all data
captured and decisions made.
• The software also provides file version control, automatically creating
version numbers for all key files and maintaining legacy versions in the
database. Differences between versions can be reviewed in the ES Tools
File History window.
• The Syngistix ES software does not permit analyses to be performed
without saving the resultant data. The security features also ensure that
analyses are conducted only with methods and other parameter files that
have been saved. These restrictions ensure that a proper audit trail is
maintained for all activities.
74 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
• Reprocessing does not change the stored data, but rather new data is
written to the dataset along with a notation that the data represents
reprocessed data rather than original data. This provides a permanent
record of all results regardless of whether they are from the initial analysis
or from post-analysis reprocessing.
• Data archival tools are also provided in the Syngistix ES software to
facilitate complete and accurate archival of data on other storage media.
Once archived, the analytical data along with the respective audit trail
entries can be restored at a future date for review or audit purposes.
Note: During the Enhanced Security installation process, either the Configurable
or Non-Configurable version of the software was selected. If the Configurable
option was chosen, some features can be disabled and enabled (see Selecting
Enhanced Security features on page 78). If the Non-Configurable option was
chosen, all features and functions are available at all times.
In this guide, many sections describe features which may or may not be available
to the user depending on the option selected during installation, and the
subsequent ES Setup configuration. If you are running the Non-Configurable
version of the software, disregard the conditional references and assume that all
functions are available to you.
Before users can log into the ES software, you must perform the following tasks to
set up the appropriate project folders and ensure proper privileges are assigned:
Windows tasks
• Set up new users in Windows
• Set up the Windows display properties for each user
• Enable the auditing of file deletion inside project folders
ES software tasks
• Log into ES Setup
• Select enhanced security features
• Define groups
• Create and set up users
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 75
Software Reference Guide
1. On the Windows Start menu, locate and select PerkinElmer Syngistix for
ICP-MS > Syngistix for ICP-MS (or double-click the Syngistix ICP-MS
icon on your desktop, if applicable). A login window appears.
2. At the prompt, type the user name and password for your Syngistix
administrator’s account. If this is the first time that you are logging into
Syngistix, you are prompted to change the password (initially 123456).
Type the new password and repeat the same password in the Confirm
Password box. Write down your new password.
3. Click OK. The Syngistix ICP-MS software starts and establishes
communication with the instrument.
Running as an Administrator
Depending on what functions you are performing, sometimes you will need to log
into the Syngistix ES software or the ES Tools utility as an administrator. For the
Syngistix ES software, you will need to run the program as an administrator only if
you are cleaning up data files, or installing an upgrade of the software. For the ES
Tools utility, you need log in as an administrator to use advanced functions, such
as accessing the security audit trail; viewing data for multiple users and folders;
repairing and compacting database files; and archiving data.
76 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
Note that both successful and failed attempts to log in as administrator are logged
in the audit trail.
Note: To run the Syngistix Enhanced Security software or the ES Tools utility as an
administrator, you must perform both parts of the following instructions (that is, you must
first be logged into Windows as an Administrator and then you must right-click and select
Run as Administrator); just logging in as a Windows Administrator is not sufficient in itself.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 77
Software Reference Guide
CAUTION! Be aware that turning off any of these features may compromise your audit
trail capabilities and your compliance with regulatory protocols.
78 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
2. In the Select Enhanced Security Features dialog box, select the desired
check boxes:
Rights Checking: When selected, users can only perform actions as
defined by their group membership.
File Versioning: When selected, the Syngistix ES software saves old
versions of key operational files to a database for full traceability,
ensuring that information is never obscured.
Audit Trail: When selected, an audit trail is kept of key user actions in
the Syngistix ES software; this is displayed as the ES Tools Syngistix
Audit Trail panel. Note that system administration tasks (such as
operations performed in ES Setup) are always logged in the Security
Audit Trail.
Challenge and Electronic Signatures: When selected, users are
required to provide their system ID credentials (their E-Signature)
and detail their reasons for performing selected actions.
Enabling this check box also adds Review, Approve, and Reject
buttons to the Syngistix software > Report View panel data tabs
to create additional signature points in the user workflow. Related
columns also appear on the Report View results tabs — along with
an e-signature credentials column — if the Display Signed By,
Display Reviewed By, and Display Approval Status check
boxes on the Advanced Report View Display Options dialog box
are also selected. These functions allow managers and lab
personnel to track e-signatures and review acquired data.
Quality Control Signoff: When selected, a QC Signoff button is
added to the Syngistix software > Report View panel QC tab to
create a QC-specific signature point for acquired data. This
replaces the Review button which appears on the other data tabs
(and which appears on the QC tab if the Challenge and Electronic
Signatures options is enabled, but the Quality Control Sign off
options is disabled). The signatory information from the QC Signoff
appears in the existing Reviewed By column on the QC tab. (The
Display Reviewed By check box on the Advanced Report View
Display Options dialog box must also be selected for this column to
appear.)
Because the electronic signature features are part of the audit trail
functionality, this check box is also disabled if the Audit Trail check
box is not selected.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 79
Software Reference Guide
3. Click OK.
Note: If this is the first time that you are logging in to the software, you are
prompted to change your password (initially set to a default of 123456). Type the
new password and repeat the same password in the Confirm Password box.
Record your new password somewhere safe.
80 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
6. On the Users tab, click New. The New User dialog box appears:
Note: The Enhanced Security software requires all new users to change their
passwords when they first log in. As a result, the User must change password
at next login check box is automatically selected and unavailable.
7. In the Available groups for user list, select a group (such as Chemists),
and click Add to move the group name to the User is a member of list.
Click Apply, and then click OK. See Setting up and managing groups on
page 85 for more information.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 81
Software Reference Guide
8. A message appears instructing you to assign a project folder for the new
user. Click OK.
9. Double-click Project Folders. The Setup dialog box appears.
10. Click the Users and Project Folders tab, and select the user’s name from
the drop-down list. Select a folder from the list of Available folders, and
then click Add to move it to the list of Assigned folders for the user. Click
Apply, and then click OK.
11. Close the ES Setup dialog box.
12. Write down the user’s name and the temporary password (123456) and
give it to the user. The user must change their password when they first
log into the software.
Note: You can specify the same project folder for all users until it becomes
necessary to use additional folders.
Note: You can set up additional administrator accounts using alternate names
and passwords, if desired.
To edit a user
1. On the Users tab, select a user from the Name drop-down list, and then
click Edit. The Edit User dialog box appears:
82 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
3. To force the user to change their password, select the User must change
password at next login check box.
Note: You cannot change the User name. You can specify the minimum length
of passwords when you manage password settings in the Password Control tab.
Note: Do not disable all users in the Administrator group; there must always be
at least one enabled Administrator.
Note: You cannot disable the user who is currently logged into ES Setup.
1. On the Users tab, select a user from the Name drop-down list, and then
click Delete.
2. When prompted to confirm the deletion, click Yes. The user is removed
from the Name drop-down list.
Note: You cannot delete all users from the Administrator group; there must
always be at least one active Administrator.
To unlock a user
If a user attempts unsuccessfully to log in to the software too many times, their
user name is locked, and the ES administrator must unlock it before the user can
log in and resume their work. The administrator sets the number of unsuccessful
login attempts after which the user is locked out. For more information, see
Managing password settings on page 92.
Note: The unlock function is unrelated to the user enable and disable functions.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 83
Software Reference Guide
5. In the Locked Out User dialog box, the user no longer appears. When you
are finished unlocking users, click OK. You do not have to unlock all
locked users to continue to use ES Setup.
To set up a new Enhanced Security user in Windows
1. If you are not logged in, log into the administrator account created for you
in Windows. If you did not supply a name, use the default Windows Admin
account. Enter your temporary password (initially 123456).
2. If this is the first time that you are logging into Windows, you are prompted
to change the password. Type your new password and repeat the same
password in the Confirm New Password box. Write down your new
password.
3. On the Windows Start menu, locate and select Control Panel.
4. In the Control Panel, select the Small Icons view, and then click
Administrative Tools > Computer Management.
5. Navigate to Local Users and Groups > Users.
6. Click Action > New User. The New User dialog box appears.
7. In the User Name field, type the user name. Type a default password of
123456 in the Password field; and then type it again in the Confirm
password field.
8. Click Create, and then click Close.
9. Now, in the list of users, right-click the new user name and then click
Properties.
10. On the General tab, ensure that the User must change password at
next logon check box is selected.
11. Click the Member of tab, and ensure that Users appears in the Member
of list. (If it does not, click Add and, in the Enter the object names to
select field, type Users and click OK. Users appears in the Members of
field.)
12. Click Apply, and then click OK.
13. Windows 10: Click the Windows button; click the User button; and then
click Sign Out. The new user is now available from the Start menu; click
the user’s icon. The user must change their password when they first log
into Windows.
OR
Windows 7: Click Start > Shut Down > Log Off. The Select user window
appears. The new user is now available. The user must change their
password when they first log into Windows.
Note: Do not create accounts with names that are difficult to remember. We
suggest that you use the exact same name for the Syngistix ES account as the
Windows account for the same user.
84 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
Administrators have full rights for activities pertaining to user and group
setup, permissions, and administrative functions. A user must have
administrative rights in order to access the security audit trail; view data for
multiple users and folders; repair and compact database files; and archive
data. See Running as an Administrator on page 76 for more information.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 85
Software Reference Guide
To add a group
1. In the ES Setup window, double-click Users & Groups. The Users,
Groups, Password Control and Summary dialog box appears. Click the
Groups tab:
86 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 87
Software Reference Guide
To select all of the permissions within a branch, click the check box
beside the parent of that branch.
To expand a branch, click the plus-sign icon. To collapse it, click the
minus-sign icon.
If you select a check box within a branch, the check box beside the
parent permission in the tree contains a grayed-out check mark.
3. When you are finished, click OK.
To delete a group
When a particular group is no longer needed, you can delete it.
Note: If you delete a user’s only group, that user cannot use the ES software until you
assign that user another group. Ensure that any affected users belong to a group.
1. On the Group tab, select a group from the Name drop-down list, and then
click Delete.
2. You are prompted to confirm the deletion. If the group you are trying to
delete has users assigned to it, the confirmation dialog box informs you of
this. To confirm deletion, click Yes. The group disappears from the Name
drop-down list.
88 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
2. On the Project Folders tab, type the full path of the project folder in the
New Project Folder box, or click Browse to locate the folder or create a
new one.
3. Click Add. The new project folder appears in the Project Folder list.
Note: If you add a project folder with a name identical to an existing project
folder (or import a similar folder from outside of the ES project folder), and this
existing project folder is not registered or defined, its read only default files get
overwritten by the added project folder. All other default files are preserved.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 89
Software Reference Guide
2. When prompted to confirm the deletion, click OK. The project folder
disappears from the Project Folder list.
Note: This operation does not delete the folder from your system. If you later
add a previously deleted project folder (or import a folder from outside the ES
project), the software overwrites only the default files, and leaves any files that
you have created intact.
90 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 91
Software Reference Guide
92 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
9. Click OK.
10. On the Password Control tab, to apply the changes you made and
continue working in this dialog box, click Apply. To apply changes and
close the dialog box, click OK.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 93
Software Reference Guide
94 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
Append Sample
Approve and Sign
File Save
Insert Priority Sample
Modify Configuration
Reject and Sign
Reprocess Data
Review and Sign
3. To require users to enter their e-signature credentials when they perform
this operation, enable the Signature Required check box.
4. To prompt users to enter relevant comments when they perform this
operation, enable the Prompt for Comments check box.
Note: If you wish to apply the same settings to all of the operations in the Name
drop-down list, click Update All. The Update All Signature Point Settings dialog
box appears. Here you can apply either option across all or no operations as
desired.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 95
Software Reference Guide
For each operation in the Signature Points Name drop-down list, you can select
one or both of the following options:
• Signature required — If this check box is enabled, User name and
Password fields appear at the top of the challenge dialog box for this
action. These are the shorter user ID (not the full name) and password
specified for this user in their ES Setup user profile (see Setting up and
managing users on page 80):
The e-signature fields are mandatory when enabled. A user will not be
allowed to continue with or complete the signature point action if a valid
user ID and password are not entered.
96 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
The Comment field is a freeform text field in which users can type any
notes; however, it does not constitute a mandatory field and no system
checks are run to determine whether or not a user has entered any
information here. If your SOPs require that users enter comments here,
this must be enforced either via the Review/Approve/Reject functions or
through management and training practices.
A challenge box with all options enabled looks like this:
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 97
Software Reference Guide
At the top of each box, a text field describes the action for the given signature
point.
A Reason drop-down list will always appear unless you delete all of the reasons
for a given signature point. For each signature point, there are one or more pre-
defined reasons in the system when you install the software. You can add your
own reasons to each list; and edit or delete the existing reasons. Arrow buttons
allow you to change the order of the reasons in the list; the first reason in the list
will always be the default selected on the challenge dialog box. If selection of a
valid reason is important to your company’s SOPs, this can be enforced via the
Review/Approve/Reject function or through management and training practices.
98 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 99
Software Reference Guide
100 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
The title bar of the window shows the location of the XML file generated. A series
of icons and filter drop-down selectors appear in the menu bar along the top of the
panel, and a calendar selection tool to the right of the menu bar lets you define a
date range for the data you wish to review:
Click the Refresh button to update the current view with the most
recent data.
Click the Clear Filters button to remove filtering options and
return to the complete file view.
In the date selection fields, select start and end dates for the date
range you want to view, and then click the Filter Date Range
button.
Note: The calendar shows only those dates for which there is existing
data.
For every event, the Syngistix Audit Trail table displays the relevant Record
Number, Time Stamp, User ID, User Name (their full name), the Action Type,
Action Details, Reasons, and Comments (if any). If the Audit Trail and Challenge
and Electronic Signatures options are enabled in the Select ES Features dialog
box, this panel also includes e-signature and reviewer/approver/file rejection
details. If the Quality Control Sign off check box is also selected, QC review
information is also displayed
You can change the view of the audit trail events in the following ways:
• To sort the rows alphanumerically by column, click the column heading
(except for Rec # or Reasons). The column by which the rows are sorted is
highlighted and an arrow appears beside the column heading. An arrow
pointing up indicates ascending sort order, while an arrow pointing down
means descending. The Time Stamp column defaults to descending, and
is selected by default when you first open the Syngistix Audit Trail window.
• When you first open the Syngistix Audit Trail window, all events are shown.
To display only events of a particular type, click one of the filters in the
Select a Filter drop-down list:
Syngistix log in/out
File Save
Syngistix System Event
Sample Analysis
Dataset Operations
Modify Configuration Parameters
Print Report
ES Setup
Archive/Restore Operation
Review
Approve
Reject
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 101
Software Reference Guide
QC Review
• To display only those events attributed to a project folder, click the folder
name in the Select a Project Folder drop-down list. Individual users can
see only their own project folder. Administrators can see all project folders.
• To display only those events attributed to a particular user, click the user
name in the Select a User drop-down list.
• To display only those events signed by a particular user, click the user
name in the Select Signed By drop-down list.
• To display only those events reviewed by a particular reviewer, click the
user name in the Select Reviewed By drop-down list.
• To display only those events approved by a particular user, click the user
name in the Select Approved By drop-down list.
• To display only those events rejected by a particular user, click the user
name in the Select Rejected By drop-down list.
• To display only those events with QC signoff attributed to a particular user,
click the user name in the Select QC Reviewed By drop-down list.
• display events between two particular dates, select the dates from the
drop-down boxes, and then click the Filter Date Range button.
• To remove filtering and display all events, click the Remove Filtering
button.
• To update the view with the most current information from the Syngistix
audit trail database, click the Refresh button.
• To print the filtered list of events, click File > Print.
102 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
Figure 3-15 Login History & Audit Trail dialog box Login History tab
The Login History tab shows a history of user login and logout actions, both
successful and failed, together with the following information:
• Full Name: The user’s full name, as defined in ES Setup
• User Name: The user’s shorter user name or ID, as defined in ES Setup
• Computer: the name or ID of the computer on which the function was
performed
• Status: An indication of whether the user’s login attempt was successful
or failed
• Logged In: The date and time at which they performed the login or failed
attempt
• Logged Out: The date and time at which they performed the logout
To print the contents of the current tab, click Print. To export the contents of the
current tab to a comma-separated or raw text file, click Export.To clear the
contents of the current tab, click Clear History (note that if the events have not
yet been archived, they are not cleared).
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 103
Software Reference Guide
Figure 3-16 Login History & Audit Trail dialog box security Audit Trail tab
The security Audit Trail tab shows the following information for administrative
actions performed in ES Setup:
• Function: The function performed
• Previous Value: the previous setting or value of the field that was
changed
• Current Value: the new setting or value to which the field was changed
• Full Name: the administrator’s full name, as defined in ES Setup
• User Name: the administrator’s shorter user name or ID, as defined in ES
Setup
• Computer: the name or ID of the computer on which the function was
performed
• Date Modified: the date and time when the function was performed.
To print the contents of the current tab, click Print. To export the contents of the
current tab to a comma-separated or raw text file, click Export.To clear the
contents of the current tab, click Clear Audit (note that if the events have not yet
been archived, they are not cleared).
104 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 105
Software Reference Guide
The File History window contains three panels with a row of filters and action
buttons running along the top of the window. The buttons provide the following
functionality:
• Project Folder — In the Project Folder drop-down list, select the project
folder containing the file you want to review. The list will contain only those
folders associated with your name in ES Setup.
• Refresh — To ensure that all the latest files and actions are reflected in
the File History window, click Refresh.
• Show Differences — Select two versions of a file in the File Versions
table and then click Show Differences to view a detail of the differences
between the two versions in the results panel at the bottom of the window.
• View File — Select a file version row in the File Versions table and then
click View File to display detailed information about this version of the
selected file (such as date created or modified, file location, and
configuration details) in the results panel at the bottom of the window.
• Signature Details— Select a file version row in the File Versions table and
then click Signature Details to display e-signature information for users
who have performed actions and provided e-signature credentials for this
version of the selected file in the results panel at the bottom of the window.
This information will also include e-signatures provided during Review,
Approve, and Reject functions either in this window, or in the Syngistix
software Report View panels.
Note: Default, pre-defined files provided with the software will list PerkinElmer
Inc. as the initial signatory, reviewer, and approver.
• Review — Select a file version row in the File Versions table and then click
Review to provide your e-signature credentials and comments (as
required) to indicate that you have reviewed the selected record.
• Approve — Select a file version row in the File Versions table and then
click Approve to provide your e-signature credentials and comments (as
required) to indicate that you approve the selected record.
• Reject — Select a file version row in the File Versions table and then click
Reject to provide your e-signature credentials and comments (as
required) to indicate that you reject the selected record.
The panels display the following information:
• The small File Types panel in the top-left corner provides an expandable
tree list of viewable files in your system. Expand the tree and select the file
you want to view.
• The larger panel in the top-right corner contains the File Version table,
listing all versions of the selected file.
• The bottom panel displays details or difference information for the file
version or versions selected, depending on the action button selected.
Select a row in the File Version table and then click an enabled button to
view details here; or select two rows and then click Show Differences to
display the delta here.
106 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
Note: The Review, Approve, and Reject buttons are available only if the Audit
Trail and Challenge and Electronic Signatures options are enabled in the ES
Setup window Select ES Features dialog box. These button are enabled only if
the user belongs to a group with the appropriate permissions for the given action
AND if the action is appropriate to the selected file version.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 107
Software Reference Guide
Multiple reviews can be performed on any data row (by one or more users); for
example, a reviewer can perform a subsequent review to add more details, add an
additional reason, or comment on another reviewer’s signature. Approve and
Reject functions cannot be performed on a row until at least one Review has been
logged.
If the Audit Trail > Quality Control Sign off option is also enabled on the Select
Enhanced Security Features window, the Review button on the Report View QC
tab is replaced by a QC Signoff button. You can configure a user Group for quality
assurance personnel to perform QC validation, and use this function to add it to
your workflow.
Because the Reporter currently displays information for quantitative analyses
only, acquired data records for other types of analyses — such as Isotope Ratio or
TotalQuant analyses — cannot have the review functions applied to them.
ES Tools File History Review, Approve, and Reject functions
All file types except data files can have Review, Approve, and Reject functions
applied to them only in the ES Tools > File History window. Here, the Review,
Approve, and Reject functions appear in the row of action buttons at the top of the
window. The File History window contains a comprehensive record of all key
workflow files in the system. Here users who are members of groups with the
appropriate permissions can select from any of their accessible project folders, to
isolate and examine all versions of any file contained therein. They can then click
on any file version that has not been closed to select a particular file version (or
CTRL + click to select multiple rows); and then indicate review, approval, or
rejection of the selected data.
Multiple reviews can be performed (by one or more users) on any file version that
has not been marked Version Closed. For example, a reviewer can perform a
subsequent review to add more details, add an additional reason, or comment on
another reviewer’s signature. Approve and Reject functions cannot be performed
on a file version until at least one Review has been logged.
Approved files are labeled Approved, and become the new in-use version of that
particular file in the system; all previous unapproved versions of that file are
marked Version Closed. When a file is rejected, it is labeled Rejected, and a
version of the file tagged with a _rejected suffix is created and kept in the project
folder for future examination; the previous version of the file is restored to use.
If your system has been configured to not require e-signature information (or
some subset of e-signature information) at certain points in the workflow, the
relevant cells in the File History table are marked NA (Not Applicable).
108 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 109
Software Reference Guide
110 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 111
Software Reference Guide
6. Click Archive.
7. When prompted to confirm the operation, click Yes. The Status area
shows the progress of the operation.
8. When the archive is complete, click Exit.
112 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Enhanced Security™ Software Administration
3. Choose a folder into which to restore the file by typing the full path of the
folder, or click Browse to locate a folder.
4. Click Restore. A confirmation dialog box appears, reminding you that any
files in the destination folder that have names identical to those in the
archive will be overwritten.
5. To confirm the restore operation, click Yes. The Status bar shows the
progress of the operation. When the operation is complete, a dialog box
appears reminding you to launch ES Setup to define the restored folder as
a project folder and assign it to a user.
6. Click Exit.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 113
Software Reference Guide
114 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Chapter 4
Optimization Processes
This section discusses the optimization and tuning functions of the software. The
SmartTune tools allow you to perform regular performance checks and
optimizations at varying levels of granularity. And each method you run includes
mass calibration and conditions files that specify the hardware parameters under
which determinations will occur.
Topics in this section include:
• The SmartTune™ Functions on page 115
• Custom Mass Calibrations on page 143
• Deadtime Corrections on page 144
• Mass Calibration panel Software Reference on page 145
• Conditions screen on page 132
• LogBook on page 147
115
Software Reference Guide
116 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 117
Software Reference Guide
Note: Some systems may not be able to meet the performance values indicated
in the pre-defined SmartTune files. These criteria reflect the factory specifications
to which your instrument was certified when manufactured. Laboratory conditions
and normal instrument use can affect performance over time, as can
modifications to instrument operating parameters, sample introduction systems,
and other hardware. Consequently, these criteria may no longer be achievable
following initial performance verification. In this event, you can still obtain
excellent analytical results in most cases by employing the appropriate QC
protocols.
118 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
required. Before doing this procedure, you must calibrate the detector
voltages.
• ICP RF Power: Use this procedure to optimize the ICP RF power only
when required. You will seldom need to perform this procedure. The ICP
RF power is set to a value of between 1000 and 1600 watts, depending on
the sample introduction system used and the nature of the sample matrix.
• Lab Performance Check: Use this procedure to set up a custom
performance check for your lab, and to validate that background count,
sensitivities, and oxides are all within the desired ranges.
Note: The data from any Lab Performance Check is automatically recorded in
the LogBook as part of the system’s performance tracking and diagnostics
function. You can use the Lab performance Check option to check system
performance for custom acquisition profiles and non-standard cell gases.
Note: The data from any Performance Check is automatically recorded in the
LogBook as part of the system’s performance tracking and diagnostics function.
• Plasma Gas Flow: Use this procedure to set up the plasma gas flow
optimization parameters.
• QID (Quadrupole Ion Deflector): Available on 300/350Q instruments
only. Use to optimize QID parameters to adjust the lens scanning system
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 119
Software Reference Guide
IMPORTANT! You must use the default method provided for all QID
calibrations. Also, ensure that the mass range used for your calibration includes
the range specified for your analytical determinations. Any deviation from these
protocols will compromise analytical performance.
• Torch Alignment: Use this procedure to optimize the X and Y axis torch
alignment. The instrument firmware uses an optimization algorithm to
determine the torch XY coordinates at which the maximum signal intensity
will be produced. The convergence criterion of the algorithm is that the
RSD between successive points must be less than 5%. The optimization
algorithm runs a maximum of 50 steps, after which, if the RSD criterion is
not yet met, the optimum is set to those coordinates where the maximum
intensity was recorded. The optimization fails if no intensities encountered
exceed 1000 cps.
• Torch Sampling Depth: Use this procedure to optimize the Z axis torch
depth alignment. This process may help to increase optimal performance
when determining for low oxide samples.
• [STD] Performance Check: Available on UCT instruments only. Use this
procedure to set up a STD mode performance check for your lab, and to
validate that background count, sensitivities, and oxides are all within the
desired ranges. Use this check when switching to STD mode from another
mode.
• [STD] QRO (Quadrupole Rod Offset): Available on UCT instruments
only. Use this procedure to optimize the quadrupole rod offset in standard
mode.
• [STD/DRC] QID (Quadrupole Ion Deflector): Available on UCT
instruments only. Use to optimize QID parameters to adjust the lens
scanning system for maximum efficiency measuring the isotopes of
interest. The QID function compensates for matrix suppression and space
charge effects in the ion optics, enabling the software to achieve optimum
sensitivity for each isotope.
When run as part of a SmartTune Express optimization, this function
employs a dynamic range-finding feature, wherein the software
automatically detects the optimum range settings for successive retries of
the optimization and configures them for you, decreasing the amount of
time spent on each run.
IMPORTANT! You must use the default method provided for all QID
calibrations. Also, ensure that the mass range used for your calibration includes
the range specified for your analytical determinations. Any deviation from these
protocols will compromise analytical performance.
120 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
IMPORTANT! You must use the default method provided for all QID
calibrations. Also, ensure that the mass range used for your calibration includes
the range specified for your analytical determinations. Any deviation from these
protocols will compromise analytical performance.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 121
Software Reference Guide
within the desired ranges. Use this check when switching to KED mode
profile from a profile in another mode.
Note: The data from any KED Performance Check is automatically recorded in
the LogBook as part of the system’s performance tracking and diagnostics
function.
Note: The data from any DRC Performance Check is automatically recorded in
the LogBook as part of the system’s performance tracking and diagnostics
function.
122 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
the DRC mode QRO, your aim is to select a value that results in minimum
background while maintaining maximum analyte signal.
• [Service] AC Rod Offset: Service mode only. Available on 300/350Q
instruments only. Use to optimize the AC Rod Offset.
• [Service] Discriminator Threshold: Service mode only. A bracket with
two capacitors connected in parallel sits on the ion detector high voltage
power supply board to reduce the ripple noise on the pulse stage positive
high voltage. This bracket has EMI fingers for grounding purposes. By
reducing the ripple, the instrument can operate at a lower discriminator
level and a lower multiplier gain, thereby extending the life of the detector.
The factory parameter for the discriminator threshold is printed on the data
sheet that is shipped with the instrument. The optimum discriminator
threshold is set so the instrument can distinguish between real pulses and
electronic noise. If, after a period of time, the noise increases, optimize the
discriminator. When you have finished, optimize the detector voltages.
• [Service] Torch XY Intensity Plot: Service mode only. Use this
procedure to set up a torch XY intensity plot at high or low resolution.
Note: When run, a Mass Calibration and Resolution should come first in a complex
optimization, but it is not normally a daily procedure. Run it daily only if you are using
certain EPA methods, such as US EPA method 200.8 or 6020a.
Procedure When to perform
Deadtime Correction When the detector is replaced, in which
case you should do it after the detector
optimization.
See Deadtime Corrections on page 144 for
more information.
Deflector Voltage When you require maximum sensitivity for
reading a single element.
Detector Voltages Monthly, or when sensitivity cannot be
recovered through other cleaning or
optimization methods, or when the
detector is replaced.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 123
Software Reference Guide
Procedure When to perform
Dual Detector Calibration When you require an extended dynamic
range above two million counts per second
in a quantitative analysis. A detector
calibration is always necessary if you have
selected the Dual Detector mode on the
Method panel Processing tab. You must
calibrate for each analyte you want to
quantitate. A minimum of two masses is
required for cross‐calibration.
Nebulizer Gas Flow Daily.
Performance Check Daily, or as required.
QID Daily if your methods use the QID™
function. Also, perform a QID calibration
when sample matrices are significantly
different to achieve the best performance.
Torch Alignment Daily, or after you clean or replace the
cones, or perform any maintenance
procedures in the torch chamber.
Procedure When to perform
Deadtime Correction When a detector has been replaced, and
after the new detector has been optimized.
See Deadtime Corrections on page 144 for
more information.
Deflector Voltages When you require maximum sensitivity for
reading a single element.
Detector Voltages Monthly, or when sensitivity cannot be
recovered through other cleaning or
optimization methods, or when the
detector is replaced.
Dual Detector Calibration When you require an extended dynamic
range above two million counts per second
in a quantitative analysis. A detector
calibration is always necessary if you have
selected the Dual Detector mode on the
Method panel Processing tab. You must
calibrate for each analyte you want to
quantitate. A minimum of two masses is
required for cross‐calibration.
124 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
Procedure When to perform
Mass Calibration When the sensitivity cannot be recovered
through optimization or when there are
changes made to the quadrupole power
source electronics.
Torch Alignment Daily, or after you clean or replace the
cones, or perform any maintenance
procedures in the torch chamber.
STD Performance Check Daily, or as required (i.e., when switching to
STD mode from another mode).
[STD] Cell Entrance/Exit Voltage When the performance check does not
meet specifications.
[STD] CRO (Cell Rod Offset) When the performance check does not
meet specifications.
[STD/KED] AMS Gas Flow Daily, if the AMS gas option is installed.
[STD/KED] Makeup Gas Flow Daily, if the makeup gas option is installed.
[STD/KED] Nebulizer Gas Flow Daily.
[STD/DRC] QID (Quadrupole Ion Daily if your methods use the QID™
Deflector) function. Also, perform a QID calibration
when sample matrices are significantly
different to achieve the best performance.
Procedure When to perform
Cell Gas Flow When measured detection limits do not
meet requirements.
See the help system for more information.
Detection Limits When measured detection limits do not
meet requirements.
See the help system for more information.
[DRC Profile] Cell Entrance/Exit When measured detection limits do not
Voltage meet requirements.
[DRC Profile] CRO When measured detection limits do not
meet requirements.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 125
Software Reference Guide
Procedure When to perform
[DRC Profile] Makeup Gas Flow When measured detection limits do not
[DRC Profile] Nebulizer Gas Flow meet requirements.
[DRC Profile] QRO (Quadrupole Rod When measured detection limits do not
Offset) meet requirements.
[DRC Profile] Performance Check Daily, or as required (i.e., when switching
to DRC mode from another mode).
Procedure When to perform
Cell Gas Flow When measured detection limits do not
meet requirements.
See the help system for more information.
[KED] QID (Quadrupole Ion Daily if your methods use the QID™
Deflector) function. Also, perform a QID calibration
when sample matrices are significantly
different to achieve the best performance.
[KED Profile] Cell Entrance Voltage When measured detection limits do not
meet requirements.
[KED Profile] Cell Exit Voltage When measured detection limits do not
meet requirements.
[KED Profile] CRO (Cell Rod Offset) When measured detection limits do not
meet requirements.
[KED Profile] Performance Check Daily, or as required (i.e., when switching
to KED mode from another mode).
[KED Profile] QRO (Quadrupole Rod When measured detection limits do not
Offset) meet requirements.
126 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 127
Software Reference Guide
Note: For UCT instruments, a drop-down list at the top of the Edit List dialog box
displays all acquisition profiles in the system. You can choose any profile (or
multiple profiles) when creating or saving a SmartTune optimization file, but you
can only run optimizations for which the selected profiles are currently assigned to
a gas channel.
Depending on the mode or profile chosen, some optimization types will have the
relevant profile name (gas and mode) added to the beginning of the optimization
name (for example: [Oxygen DRC] CRO or [Helium KED] Performance Check) to
allow you to optimize for the specific acquisition conditions.
• Mode: SmartTune Express only and UCT instruments only Displays the
mode or modes to which the optimization will apply. Click Set to specify
the modes you will be optimizing for when setting up a SmartTune Express
optimization. Here you can also indicate the appropriate cell gas flow for
each mode (in liters per minute).
128 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
• Optimization list box: In SmartTune Express, this box on the left side of
the SmartTune setup panel initially displays instructions for using
SmartTune Express; when an optimization is running, it lists the processes
being run as they are run.
In SmartTune Manual, this box on the left side of the SmartTune setup
panel shows a list of optimization procedures to be processed in
sequence. Click Edit List to add procedures. You can change the
optimization sequence by dragging entries in the list. You can also click an
entry in the list to show the details for that procedure. Or right-click a
procedure in the list, and then click:
Delete to remove a procedure from the list
Quick Optimize to run the selected procedure only
• Use Manual Sampling (No Autosampler): Select this check box if you do
not use an autosampler. Note that, if using manual sampling, you will be
prompted to aspirate each optimization solution in turn — Read and Flush
delays will not apply. (See also Use Smart Sampling, following.)
• Use Smart Sampling: For manual sample runs (not using an
autosampler), if selected, this intelligent sampling function detects and
determines when two sequential functions use the same solution. When
this is the case, you are no longer prompted between optimizations to
manually aspirate the sample.
• Stop if optimization fails: Select this check box if you want the
SmartTune process to stop immediately following any unsuccessful
optimization. Unavailable in the SmartTune Express function.
• Send Results to Printer: Select this check box if you want to
automatically print the result summary reports following optimization.
• Autosampler: These parameters specify autosampler locations for each
procedure:
Procedure displays read-only entries that are updated when entries in
the optimization procedure list are added, deleted, or moved
A/S Location lists the autosampler tray positions for the solution to be
used for each optimization. This entry is ignored if a manual analysis
with no autosampler is being performed
• Peristaltic Pump: This table provides peristaltic pump parameters
derived from the Global Pump parameters. Change these to override
these parameters when running a SmartTune optimization:
Sample Flush Time: Specifies the time in seconds during which the
sample tubing is flushed with sample solution. Type a value between 0
and 99 999
Sample Flush Speed: Specifies the pump speed in rpm during the
flush cycle. Any value between ±150 (2000 series instruments) or ±48
(300/350 series instruments) is valid
Read Delay Time: Specifies the time in seconds between the end of
the flush cycle and the beginning of data acquisition. Type a value
between 0 and 99 999
Read Delay Speed: Specifies the pump rate in rpm used during the
read delay cycle. Any value between ±150 (2000 series instruments)
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 129
Software Reference Guide
130 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
Control Function
Method Displays the method file used with this optimization
procedure. Use the default method displayed, or click
Browse to select a different file.
Criteria These options define the criteria against which the
instrument optimizes for this function. Typically, you
will click to choose one analyte for Intensity
optimization or one analyte Formula; use the formula
option when you want to optimize on a relationship
between two signals.
For some functions, you may choose All Analytes, and
the system will automatically select all the available
analytes in the related method, and provide a
secondary option for then excluding ratio and
background analytes as required.
Other functions employ a dynamic range‐finding
feature, wherein the software automatically detects
the optimum range settings for successive retries of
the optimization and configures them for you,
decreasing the amount of time spent on each run.
Specific criteria are listed with the individual
procedures.
Range These parameters define the range the instrument will
measure. During optimization, the system moves
through a series of values, with data acquired at each
value. You control the start and end flow values, and
the step size.
Ramp Select the Ramp check box to specify that the software
should conduct an optimization, but should return to
the original value when completed. Use this option
when you want to review the data before making any
parameter changes, or want to select a value that
does not result in the highest possible signal.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 131
Software Reference Guide
Conditions screen
Note: Generally, you should optimize and calibrate the instrument using the SmartTune
function. However, advanced users can use the Mass Calibration and Conditions panels
to perform high-precision or custom calibrations.
The Conditions screen is accessed via the Conditions icon on the ribbon, and
consists of a Conditions panel with multiple functional tabs, together with Sample
and Realtime panels, so that you can make changes and monitor your results
while optimizing the system. Use the Conditions panel to fine-tune the hardware
subsystems in the torch/quadrupole assembly. Here you can adjust control
parameters for the DACs (digital-to-analog converters) associated with these
functions. When opened, this panel displays the last file accessed. The Conditions
panel has several tabs:
• QID tab: Use this tab to view ion lens scan data. On UCT instruments, this
tab is further divided into subtabs for standard/DRC and KED modes, as
well as for any custom QID acquisition profiles.
• Dual Detector Calibration tab: Use this tab to view the performance of
each detector element (pulse and analog) and the crossover points
between the two detector systems.
132 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
• Manual Adjust tab: Use this tab to manually adjust the DAC values, so
you can make DAC adjustments without having to perform the full
optimization procedure.
• Advanced Optimize tab: Use this tab to control advanced optimization
routines for cell gases, rejection parameters, and Axial Field™ (AFT)
voltage. UCT instruments only.
• Cell Parameters tab: Use this tab to view advanced optimization routines
for cell gases and rejection parameters. You can also edit these values
here, and send them to the method. UCT instruments only.
QID Tabs
Use this tab to review lens scan data to ensure the instrument is experiencing
maximum efficiency measuring the isotopes of interest. The QID™ (Quadrupole
Ion Deflector) function compensates for matrix suppression and space charge
effects in the ion optics. It also helps the software to achieve optimum sensitivity
for each isotope in an analysis. You optimize the QID settings using the
SmartTune™ function. Use the QID tab only to review current QID calibration
data, or to reprocess a calibration using previously acquired data.
During acquisition profile creation, custom QIDs can also be specified, causing
additional tabs to appear here (the tab name will match the originating profile).
See Understanding Acquisition Profiles on page 31 for more information.
Note: On UCT instruments, this tab is further divided into subtabs for standard/DRC and
KED modes, as well as for any custom QID acquisition profiles. Although the algorithms
and parameters used for each mode are different, the controls on each of these subtabs
are identical.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 133
Software Reference Guide
134 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
• Max Intensity: Lists the maximum intensity value obtained for the listed
analyte during the QID calibration. This entry is read-only.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 135
Software Reference Guide
regardless of the original method type, as long as the data was acquired
using both detectors.
• Points Acquired: Specifies the number of unique voltages measured
during the dual detector calibration procedure.
Parameters in the Dual Detector Table
• Analyte: Specifies the analytes that will be used in the dual detector
calibration. The analytes listed here should match the list of analytes that
are being examined in the method. To automatically list all the analytes in
the active method, click Get Analyte List. When you require extended
dynamic range (above 2 million counts per second) in a quantitative
analysis, you must calibrate for each analyte you want to measure.
• Mass: Specifies the exact mass at which the dual detector calibration will
be performed for the analyte listed in the same row of the Dual Detector
Calibration data table. Populate the Analyte and Mass fields by clicking
Get Analyte List. This imports the analyte and mass from the method. To
add an analyte, add it to the Method panel Timing tab, and then click Get
Analyte List.
• #Points: Lists the number of crossover data points included in the dual
detector calibration curve. This entry is read-only. When the software
calibrates the dual detector system, response curves are generated for
both the pulse and analog stages of the detector. These two response
curves are integrated to yield a single curve which will serve as the
detector response curve for analytical determinations. For accurate dual
detector calibration, the system should have as many common data points
(that is, crossover points) as possible. These are data points at which the
system acquired both analog and pulse count data during the calibration
sequence.
• Coefficient: Lists the correlation coefficient for the composite response
curve that the software creates during dual detector calibration. This
correlation coefficient is generated during the linear regression of the data
points and is a measure of the quality of the final calibration curve. This
entry is read-only. A poor correlation coefficient is an indication that the
two response curves had very different slopes prior to their integration. In
this event, it is recommended that you recalibrate the system before
proceeding with analytical determinations.
• Gain: Lists the gain factor that will be used for the analyte in the same row
of the Dual Detector Calibration data table. This entry is read-only. The
gain value is used to decide how the software translates an analog current
signal into usable pulse count data during measurements.
• N(max): This read-only entry is the dynamic range of the detector using
the extended dynamic mode (dual mode), in counts per second. Signal
intensities above the N(max) value are considered to be in shutdown (S).
• Conversion Factor: Indicates the ratio of the analog signal to pulse
signal. It is the slope of the line on the dual detector calibration graphs.
These values are plotted in the Detector Response Function. This entry is
read-only.
136 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 137
Software Reference Guide
• View DAC Values: UCT instruments only. Click Active to view DAC values
for the selected profile only. Click All to view DAC values for all active
profiles (including Standard).
• DAC: Lists the current DAC value of the selected component. Use the
arrow controls to change the current value, or click and type directly in any
table cell.
Parameters in the Manual Adjust Table
• Profile/Value columns: UCT instruments only. Lists the active DAC value
of the hardware parameter selected for the profile indicated. Used when a
determination is run using the active Conditions file. To change a value,
click in the relevant table cell, delete the old entry, and type a new value.
The acceptable range for entries is listed in the Minimum Value and
Maximum Value fields.
138 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 139
Software Reference Guide
The AFT voltage does not require frequent optimization because it has
a very broad optimization peak. However, if you suspect that the AFT
voltage requires optimization, run this process. Because the AFT
voltage optimization value is dependent on the deflector voltage, cell
path voltage, and cell rod offset, you should optimize it after these
parameters. If the value changes by greater than 100 V, you should
then re-optimize the deflector voltage, cell path voltage, cell rod offset,
and AFT voltage.
RPa: Adjusts the DC voltage applied to the reaction cell quadrupole
RPq: Adjusts the RF voltage applied to the reaction cell quadrupole
• Analyte: Specifies the element-mass-profile (generally gas and mode) on
which you want to optimize. If no analytes are listed, click Get Analyte List
to retrieve the list of analytes from the active method.
• Start Value: Specifies the start value for the optimization.
• End Value: Specifies the final value for the optimization.
• Step Value: Specifies the amount the value will be increased during
successive measurements of the optimization procedure. During
optimization, the software ramps the DAC value between the Start Value
and the End Value, increasing in successive measurements by the Step
Value. The intensity for the selected analyte is then plotted as a function of
the parameter value to find the optimum parameter. Click Get Defaults to
retrieve the default system values for the Start, End, and Step values.
• Maximum Intensity: Specifies that the software should calculate the
parameter value which creates the highest signal intensity during an
optimization.
• Minimum Intensity: Specifies that the software should calculate the
parameter value that creates the lowest usable signal intensity during an
optimization.
• Ramp: Specifies that the software should conduct an optimization, but
should return to the original value when completed. This option is selected
when you want to review the data before making the parameter change, or
want to select a value which does not result in the highest possible signal.
• Formula: Use this option to optimize on a relationship between two
signals. The most common relationship is signal-to-noise ratio. This
involves placing the analyte of interest in the first window under the
Formula option. Then select an operation (+, -, × or / ). Select the second
analyte and determine the relationship of either greater than, less than or
max (<, >, or max). In the final field insert a value. For example: CeO
156/Ce 140 < 0.03 (formula used to optimize the oxide ratio).
• Spike Concentration (ppb): RPq optimizations only. Use this entry to
specify the concentration in ppb of the standard solution or spike when
performing an automated RPq optimization. All analytes in this solution
must have the same concentration.
• LOD (Limit of Detection): RPq optimizations only. Select to choose Limit
of Detection as the optimization criterion. The LOD, expressed as the
concentration or quantity, is derived from the smallest measure that can be
detected with reasonable certainty for a given analytical procedure.
140 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 141
Software Reference Guide
• Begin Mass (amu): (Data Only and TotalQuant methods only) Defines the
start of an atomic mass range for each measurement. The scanning range
defined by a single row in the table is the range between the Begin Mass
and End Mass values set in that specific row. This is a read-only
parameter whose value is configured in the Method panel Timing tab. To
populate a row with the default system cell parameter values, highlight the
row and then click Reset.
• End Mass (amu): (Data Only and TotalQuant methods only) Defines the
end of an atomic mass range for each measurement. This is a read-only
parameter whose value is configured in the Method panel Timing tab.
• Profile: Displays the acquisition profile to be used for the determination.
• [Gas] Flow columns (ml/min): These columns specify the flow rate in
ml/min at which the specified cell gas flows into the reaction cell.
• RPa (Rejection Parameter a): Specifies the DC voltage applied to the
reaction cell quadrupole for all methods. In normal operation, this entry is
automatically filled by the software when you type the atomic symbol or
isotope number in the Analyte column. To enter information in this column,
click the relevant cell and type the exact voltage at which the
measurement should occur.
• RPq (Rejection Parameter q): Specifies the RF voltage applied to the
reaction cell quadrupole for all methods. In normal operation, this entry is
automatically filled by the software when you type the atomic symbol or
isotope number in the Analyte column. To enter information in this column,
click the relevant cell and type the exact voltage at which the
measurement should occur.
142 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
The mass calibration .tun file manages the instrument's quadrupole mass filtering
performance. The mass calibration functions permit custom parameter
configuration for mass calibration and resolution. You should calibrate the
instrument whenever there are changes to the electronics, or if you need to modify
resolution for one or more elements.
Note: The instrument must run for at least 30 minutes with the plasma on before you
perform a mass calibration.
DAC Files
The software tuning, optimization, and calibration controls include several entries,
controls, and files that use the acronym DAC. A DAC is a digital-to-analog
converter channel connected to a specific hardware subsystem. An electronically
controlled hardware component is adjusted by varying the parameter of its related
DAC.
Calibration Adjustments
Mass calibration and resolution adjustments affect the DAC parameters for the
quadrupole subsystem. The quadrupole systems act as a mass filter; calibration
affects the timing of ions through this filter, and the shape of the resulting mass
spectral peaks. Use the mass calibration controls to perform the following
operations:
• A mass calibration to adjust the instrument's electronics and guarantee the
accuracy of the mass spectrometer
• A resolution check and adjustment to ensure that the resolution at each
mass of interest is within the defined range
When running either operation, you can view the results as they happen in the
accompanying Realtime graphics panel.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 143
Software Reference Guide
Mass Calibration
When a mass calibration is performed, the Interactive panel displays a series of
graphs corresponding to the individual elements measured during the calibration.
During calibration, the instrument performs a determination on a test solution,
typically a solution covering a broad mass range. The instrument finds the
maximum intensity of each isotope in the solution and its atomic mass relative to
its intensity — the measured mass of the isotope.
During this measurement, the system scans for a peak within a predefined scan
width region centered around the exact, or known, mass for that isotope. The
instrument can then compensate for the differences between the apparent and
exact mass.
Resolution Check or Adjustment
Resolution is measured as the peak width at 10% peak height. The hardware is
preset to scan the entire mass range at 0.7 amu resolution. This parameter
provides the best balance between peak separation and signal-to-noise ratio for
most applications. The resolution can be adjusted on an individual mass basis, for
example to correct for spectral interferences. Again, a test solution covering a
broad mass range is typically employed. However, when adjusting resolution, you
may want to use the analytes included in your analysis method.
The system scans the specified analytes within a predefined scan width region
centered around the known mass for each analyte isotope, then measures the
peak width at 10% peak height. This value should normally be approximately 0.7
amu. Resolution should be within ±0.10 amu of the expected resolution or an
adjustment should be made.
Deadtime Corrections
Deadtime correction is a mathematical equation that is used to correct for
erroneous readings at high count rates (pulse counting) caused by the finite width
of the detector output pulses. At these high count rates (> 1,000,000 cps
approximately) there is a finite probability that two pulses will arrive at the detector
output simultaneously. The deadtime correction function corrects for this anomaly.
Deadtime Calculation
Characteristic of all ion detection systems, deadtime (t) is the amount of time
required for the detector to process the signal from an incident ion. Following an
ion strike, the detector is unable to record another strike for a short period of
time—this deadtime is typically 50 - 60 ns. If two ions arrive at the detector within t
seconds of each other, the second ion is not counted. At low count rates of less
than 1e5 cps, only 0.5% ions arrive during this period, and the count rate
measured by the detector is 0.5% less than the count rate impinging upon the
detector. At higher count rates, however, the effect is more significant and follows
the relation:
Rtrue = R0 / (1 R0t)
where:
Rtrue xis the ion count rate arriving at the detector
R0 xis the ion count rate measured by the detector
t xis the deadtime
When the deadtime of the system has been measured, the above equation can be
applied to correct for it. The software uses the deadtime correction to compensate
144 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
for counting loss at high ion arrival rates. When measuring using signals above
2e5 cps, you must ensure that the deadtime correction is set accurately.
Note: Because changes made exclusively to the resolution DAC affect the mass
calibration, you should always adjust the resolution and mass calibration in
tandem.
• Peak Search Window (amu): Specifies the width of the search window in
which the software will look for a peak for the isotope being measured. The
value entered defines a window in atomic mass units centered around the
known mass for each isotope used in the calibration. For example: If you
are calibrating for Mg (24), with a known mass of 23.985 amu, and have
set the Peak Scan Width to 1 amu, then the search range extends from
23.485 amu to 24.485 amu. Reduce the value to avoid accidentally
calibrating on an interfering peak. If you need to use a value above 1 amu
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 145
Software Reference Guide
146 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
DAC entry, you should always rerun the resolution adjustment procedure
to verify its effect on the peak width value.
• Custom Resolution: Identifies any analyte that has a different resolution.
This entry does not affect the actual spectrometer controls, and is used
only as a reminder for the user.
LogBook
When any pre-defined SmartTune Express or Manual Performance Check is run,
the results of the optimization, together with all the relevant instrument parameters
used in that optimization, are recorded in the LogBook grid. This creates a
continuous performance history of the instrument, which you can review to track
instrument performance trends over time, and to troubleshoot problems. This can
be of particular use to lab managers and service personnel: you can export the
data to Microsoft Excel in .xls or .xlsx format, in order to perform further statistical
analyses, or send the data to an offsite PerkinElmer Customer Support
Representative to aid in remote diagnostics in the case of instrument malfunction.
In the tables, sample data is displayed in alternating rows of white and gray to
improve readability. Additional color cues employed in the tables include the use
of a red font to indicate performance check failures, and the Syngistix software
standard KED mode green and DRC mode yellow used in the column headers to
indicate analytes processed using profiles in these modes (UCT instruments
only). Note that all tables take their base colors from your system colors; if you
change from the default Windows system colors, you may change the table
background colors.
Note: Mass Calibration values are stored for every performance check; they cannot,
however, be displayed on the plot.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 147
Software Reference Guide
148 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 149
Software Reference Guide
150 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Optimization Processes
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 151
Software Reference Guide
Statistics panel
The Statistics panel displays historical trends for any analyte intensity or
parameter selected in the other two panes. Select one or more cells or columns in
the main LogBook or Instrument Parameter tables to display the related
information on this panel (press CTRL + click to make multiple selections). On the
plot, a floating Legend displays a reference of the colors used to plot the selected
parameters or analytes. Horizontal and vertical sliders allow you to adjust the area
of focus along each axis; click the square arrows button in the bottom right corner
of the graph to return to the original view. Click the small, square, gray button in
the upper right corner to expand this panel for better viewability.
152 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Chapter 5
Method Development
This section provides background information about creating analytical methods
for your determinations.
Topics in this section include:
• Modes of Operation on page 153
• Analytical Techniques on page 155
• Fundamentals of Data Acquisition on page 159
• Data Processing on page 164
• Fundamentals of Calibration on page 168
• Method Panel Software Reference on page 175
Modes of Operation
The NexION® 300/350Q instrument operates in a single mode (Standard mode)
to perform typical ICP-MS analyses of samples having no significant
interferences. NexION instruments with UCT™ functionality (all 2000 series
instruments; 300/350X, D, and S models) have three available operating modes
— Standard mode, DRC™ mode, and KED mode. The mode you choose to apply
depends on the samples you will be analyzing and the types of interference you
expect to encounter. In the software, you create acquisition profiles to pair mode
assignments with the cell gases used, and to define the related gas flow rates,
and other parameters. See Acquisition Profiles tab on page 37 for more
information.
Standard Mode
The NexION standard mode provides excellent ICP-MS analysis for most of the
isotopes in the periodic table where complex spectral interferences are not an
issue, or where it is not critical that you use the most abundant isotope of each
element. If an element experiences an interference, you can find and use an
alternate isotope. Or, if you are using an instrument with UCT functionality, you
can move on to KED mode or DRC mode.
KED Mode
This mode uses collisions with an inert gas together with kinetic energy
discrimination to reduce polyatomic interferences. While not as sensitive as DRC
mode, this method can be applied to all polyatomic interferences using a single
set of operating conditions, and does not require an in-depth knowledge of the
sample prior to analysis. This system discriminates between the kinetic energy of
analyte ions and that of their interferences by creating a positively-biased voltage
between the mass analyzer and the cell. The system then relies on the higher
probability of collisions of an inert gas (for example, He) with the polyatomic
interferences based on their relatively larger collisional cross-section —
polyatomic species cannot overcome the positive energy barrier established
between the mass analyzer and the cell. KED mode is especially useful in
applications where moderate detection limits are required for multi-element
analyses in complex matrices, such as in high salt samples.
153
Software Reference Guide
DRC™ Mode
DRC mode provides ultimately low detection limits and high-sensitivity for
complex spectral interferences including polyatomics and atomic isobars. The
DRC technology uses ion-molecular reactions and band-pass tuning to chemically
correct for interferences. This allows you to remove difficult-to-avoid interferences
(such as those formed in the plasma or present in the sample) before they reach
the mass spectrometer, without limiting the detection capabilities of the system.
This mode uses controlled gas-phase reaction chemistry inside an enclosed cell
containing an additional quadrupole mass filter equipped with Axial Field™
Technology. The result is a dramatically improved detection capability into the sub
parts-per-trillion range for most elements.
Ions extracted from the plasma are focused within the ion optics and directed into
the cell. The standard mode is achieved by turning off the reaction gas flow into
the cell and opening the vent to the analyzer chamber. In this mode, the cell is a
quadrupole device which transfers the ions to the mass spectrometer’s analyzer
chamber. DRC mode is achieved by closing the vent to the analyzer chamber and
introducing a reaction gas into the cell. In this mode, ion-molecule reactions are
promoted to selectively eliminate interferences.
The specificity of interference rejection is obtained through the selection of the
reaction gas and the operating conditions. Both RF and DC electrical components
are applied to provide control over the extent of ion-molecule reactivity in the cell.
This chemical band pass is used in concert with the selected reaction gas to
reduce interferences by interrupting the sequence of reactions that would
otherwise create an interference. The RF provides an intercept for reaction
intermediates of lower mass; the DC suppresses chemistry involving higher mass
reaction intermediates. The system introduces the reaction gas through a gas
assembly function; you create acquisition profiles to define the type of gas used
and the flow rate at which the cell operates.
154 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
Analytical Techniques
Developing a method involves defining your analysis goals, selecting the
analytical technique best-suited to achieving those goals, and configuring the
method parameters to provide the best possible combination of accuracy,
precision, and analysis speed.
The most important consideration when selecting an analytical technique is the
degree of precision required in your results. For survey measurements or semi-
quantitative testing, TotalQuant™ analyses provide the most convenient method,
providing both analytical effectiveness and fast results. However, with the addition
of standards in a TotalQuant determination, many quantitative requirements are
also satisfied. While Quantitative Analysis methods are generally selected when
accuracy and precision are the primary considerations, some applications may
benefit from the added precision achievable using an Isotope Dilution method.
The software provides several different ways of managing data acquisition:
• Quantitative Analysis
• TotalQuant III Analysis
• Isotope Dilution
• Isotope Ratio
• Data Only
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 155
Software Reference Guide
W A' – B'
c = Kx -------------------------------
MB – A
where:
K = natural atomic weight of the element
xxxx atomic weight of the spike isotope
W = weight of the isotopically enriched material added
M = weight or volume of the sample
a = element isotope of highest natural abundance
b = highest abundance isotope in enriched isotope solution
A,B = natural abundances of sample isotopes a and b
A’,B’ = isotopic abundances of a and b in the spike
ñ = ratio of isotope a to isotope b measured in the sample
156 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 157
Software Reference Guide
• Dil. to Vol.: Specifies the volume of the diluted samples the autodiluter will
create. This volume ranges from 2mL to 50 mL, with a default value of 10
mL.
• 1st Dil. Pos: Here you can specify which tubes should be used when
diluted samples are created. The first diluted sample is created in the tube
number corresponding to the value entered in the 1st Dil. Pos field, the
second in the next higher numbered tube, and so on. To use dilution
positions sequentially, only the first method in the batch must have a rack
position entered for the 1st Dil. Pos field on the Method panel Sampling
tab. All methods that follow must have a zero in this field.
• Probe Purge Pos: Here you can specify that an empty tube be used for
autodiluter initialization. In order to accurately measure the volumes of
liquid involved in each dilution, the dilution probe and associated tubing
must be full of diluent before any dilutions are done. Ten milliliters of
diluent are pumped through the probe into the tube at the Probe Purge
Position. The default value for this tube position is 10.
If the autodiluter is to be under QC control, on the QC Sample tab, right-click in
the Action 1 and Action 2 columns to view a menu of available actions. For
analyte upper limit measurements, the menu item Wash For X, Dilute, and Rerun
Current is available.
You can also dilute samples before measurement by right-clicking in the
Measurement Action column on the Sample panel Batch tab and choosing one of
the following actions:
• Run Diluted Sample
• Run Blank and Diluted Sample
• Run Stds. and Diluted Sample
• Run Blank, Stds. and Dil. Sample
Note: The following solution types cannot be pre-diluted: spike, duplicate and dilution
pairs, blanks, reagent blanks, calibration standards and QC standards.
158 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 159
Software Reference Guide
160 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 161
Software Reference Guide
162 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 163
Software Reference Guide
Data Processing
The raw data acquired during a determination must be processed before it is
passed along for the analytical technique-specific calculations. Making the correct
selections requires you to consider the mixture of the analytes, the sample matrix,
instrument conditions and analytical requirements. The processing functions are
divided into two broad classes:
• Functions used to process the mass spectral data
• Functions used to process the time-based signal profile
164 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 165
Software Reference Guide
166 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 167
Software Reference Guide
Fundamentals of Calibration
Calibration is a fundamental part of acquiring the best possible data from your
instrument. In Quantitative Analysis, you calibrate the instrument by analyzing
standards for all the elements you are interested in measuring. You analyze
standards at different concentrations to ensure instrument response is accurate
across the range of concentrations you are likely to encounter during your
determinations.
As you measure standard solutions, the software plots the measured intensity
versus the concentration for each element in the standard solution. These
individual calibration curves are updated as each subsequent standard is
measured. You view and edit calibration curves in the Calibration panel. If a point
seems aberrant, you can remove it and recalculate the curve. You can also store
any calibration data to use in future determinations.
In a TotalQuant determination, the instrument uses stored response tables to
correlate measured ion intensities with known concentration values. This internal
response data indicates how many ions per second should be observed for a
concentration of one milligram/Liter for each element in a sample. The response
table takes into account variations in instrument sensitivity to different elements.
However, changes in overall instrument sensitivity can occur from day to day.
Because instrument sensitivity for all elements generally changes to the same
degree, you can update the responses for all elements by measuring the
responses for a few calibration elements.
Both Isotope Dilution and Isotope Ratio are effectively self-calibrating techniques.
In Isotope Ratio, you compare the isotope of interest to a reference isotope of the
same element. For example, you might measure Pb204 compared against Pb206.
Because this ratio can be calculated from within a single sample measurement
calibration is not required, although it is recommended to run a standard of known
isotopic composition to verify that the instrument is not biasing the results due to
mass discrimination.
One additional calibration technique, internal standardization, is commonly
employed in methods. You can use internal standards to correct for changes in
instrument hardware response or for sample-to-sample variations in sensitivity. An
internal standard is a non-analyte isotope that is added to the standards and
samples before you analyze them. During the determination, the software uses
the ratio of analyte and internal standard intensities to adjust the final analyte
intensity values. Internal standards are used in both Quantitative Analysis and
TotalQuant methods. Isotope Ratio methods use the principles of internal
standardization as the basis for ratio measurement. Isotope Dilution
measurements do not require internal standards.
168 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
External Standardization
For samples that are not subject to sample-specific matrix interferences that
change the overall instrument sensitivity between measurement of the standards
and samples, external standardization is a simple and effective way to calibrate
your instrument.
External standardization involves measuring a blank solution followed by a set of
standard solutions to create a calibration curve over the concentration range of
interest. Typically, you run two or three standard solutions containing different
concentrations of all the elements you are determining. Increasing the number of
points on the calibration curve — that is, increasing the number of standard
solutions — may improve accuracy in circumstances where the calibration range
is very broad. However, it is seldom necessary to run more than five standard
solutions. After the standards have been measured, you continue with the
measurements of your unknowns.
The following figure summarizes the measurement procedure when using
external standardization:
Blank
Standard solution 1
Standard solution 2
Standard solution 3
Sample 1
Sample 2
Sample 3
and so forth
Standard Addition Calibration
Standard addition calibration provides an effective way to minimize sample-
specific matrix effects through the use of spiked samples.
In standard addition calibration, you first measure a blank solution. Next, you
measure the sample solution spiked with a known concentration of each element
you want to examine. The software measures the response for the spiked
samples and creates a calibration curve for each element for which a spike has
been added. This curve is based on the known concentration values which you
entered. The calibration curve plots the measured intensity of each spike element
against its concentration. Based on the slope of the calibration curves, the
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 169
Software Reference Guide
170 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
Addition Calibration
Addition calibration is a variant of standard addition. It is used when all samples
have a similar matrix. The following figure summarizes the measurement
procedure when using standard addition calibration:
Blank
Spiked sample 1 (spike conc. 1)
Spiked sample 1 (spike conc. 2)
Unspiked sample 1
Unspiked sample 2
Unspiked sample 3
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 171
Software Reference Guide
The software establishes the intensities for each of the calibration elements and
then calculates the response based on the concentrations you have specified. It
then adjusts the response factors in the response table for all elements in
proportion to the calibration elements.
For example, using cobalt and rhodium in the calibration solution, the software
finds the response for cobalt at mass 59 to be 5% greater than the internally
recorded response and the response for rhodium at mass 103 to be 15% greater.
It adjusts these elements by 5% and 15% respectively. For other elements that
have isotopes between cobalt and rhodium, the adjustment is based on the
relative mass difference between the measured analyte and the calibration
elements. The adjustment value for an individual element (c) is calculated as
follows:
Adjustment = a + [(b-a) * (c-d) / (e-d)]
where:
a = % difference for calibration element with lower mass
b = % difference for calibration element with higher mass
c = mass of element
d = mass of lower calibration element
e = mass of higher calibration element
For example, the adjustment for arsenic at mass 75 is:
Adjustment = 5 + [(15-5) * (75-59)/(103-59)] = 8.6%
For the most accurate results for all detectable elements, use a calibration
solution that contains five or six elements from a wide mass range, with the
elements concentrated at the lower masses. This provides reference points for
adjustments across the entire spectrum. A typical calibration solution might
contain 1µg/L of Li, Mg, Co, In, Tb, and Pb.
You can measure your standard solution once during a determination, after
measuring the blank, or periodically during the determination (for example, to
correct for time-dependent drift, or to re-standardize between groups of samples
with dissimilar matrices).
When you change the software's internal response data through external
calibration, the changes remain in effect until you update the response data. This
means the updated response information is used for all subsequent samples until
you perform another external calibration or load a different calibration file.
172 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
Blank Blank
Standard solution Standard solution
Sample 1 Sample 1
Sample 2 Sample 2
Sample 3 Sample 3
Sample 4 Sample 4
Sample 5 Standard solution
Sample 6 Sample 5
and so forth and so forth
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 173
Software Reference Guide
Internal Standards
An internal standard is a non-analyte isotope added to standards and samples
before a determination. You can use internal standards to correct for changes in
instrument hardware response or sample-to-sample variations in sensitivity.
The implementation protocol for internal standards varies according to the
analytical technique in which it is used. For Quantitative Analysis methods, the
Internal Standard entry clusters elements with similar ICP-MS response
characteristics into a standard group, then specifies which element within the
group will serve as the standard or reference isotope. The software assumes that
all elements within the standard group are similarly affected by instrument drift or
matrix interferences. Therefore, changes in the measured intensity of the internal
standard are used to create the ratios for correcting measured intensities of the
analytes.
174 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
For example, during the measurement of Pb in water, you must account for the
effect of salinity. High Na concentrations typically result in a matrix interference.
For these measurements, Bi (mass 209) can be used as an internal standard to
correct the measured intensities of Pb (mass 208).
Table 5-1 Quantitative Analysis Internal Standard Example
Tl Pb Bi
Standard solution
Intensity 4.078 6.117 2.039
Sample solution
Measured intensity 4.820 6.234 1.78
Adjusted intensity 5.521 7.141
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 175
Software Reference Guide
Timing Tab
Use this tab to specify the isotopes or range of isotopes to look for in your
samples. You also use this tab to identify the internal standards within the sample,
adjust the analysis time spent on an individual element and to specify the Mass
Calibration and Conditions files to use during determinations. On UCT
instruments, entries on this tab are color-coded: a yellow background indicates
that the element uses a DRC mode acquisition profile; a pale green background
indicates that the element uses a KED mode acquisition profile.
176 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 177
Software Reference Guide
UCT instruments
If the method contains elements defined using multiple profiles and modes
(that is, in more than one of standard, KED, and DRC mode), the estimate
is unavailable due to the delays required for gas flow changes during
mode switching. The delay time between flow rate changes will be
significant — that is, tens of seconds. Short settling times are possible in
single mode methods only.
• Est. Sample Time: Displays an estimate of the total time to analyze a
single sample based on the current timing parameters set in the table and
the values for the Sweeps/Reading, Readings/Replicate, and Replicates
entries. The values in this field are read-only. The calculation of estimated
time is updated whenever a change is made to the Replicates,
Readings/Replicate, Sweeps/Reading, or Dwell Time as soon as you click
another entry on the tab.
UCT instruments
The calculation of estimated time is also updated whenever a change is
made to the cell gas flow entries.
• MassCal File: Specifies the name of the mass calibration file used to
calibrate the instrument during determinations made with the method.
Every method must have a mass calibration file specified. The default
mass calibration file is named default.tun.
• Enable QC Checking: Starts the quality control functions. The software
includes a comprehensive set of quality control checks for verifying data
integrity when performing quantitative analysis measurements. This option
is available for Quantitative Analysis methods only.
• Conditions File: Specifies the name of the conditions file that will be used
during determinations made with the method. The conditions file contains
important hardware parameters for your instrument, including nebulizer
flow, ICP power, and dual detector calibration information. Every method
must have a conditions file specified. The default conditions file is named
default.dac.
The Timing Table
The Timing table lists the isotopes the software will measure, together with
specifics about the measurement timing conditions. This table must include at
least one analyte (and one Begin Mass/End Mass entry for Data Only and
TotalQuant analyses) and their associated measurement specifications. The
software automatically enters data in some of the table fields when you add an
analyte. These default parameters provide a standard starting point for developing
the method. The conditions yield reasonable data, permitting you to analyze the
results and define the type of modifications required to extract the best possible
data for your specific sample system. On UCT instruments, at the time of analysis,
analytes in the Timing table with the same profile and gas flow settings are
grouped together.
Parameters in the Timing Table
• Internal Standard: (Quantitative Analysis and TotalQuant methods only)
Lists the standard groups and internal standards that have been defined in
the method.
When using internal standards with TotalQuant methods, these entries are
used to calibrate the TotalQuant response table for optimal accuracy
during analytical measurements.
178 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
• Analyte (*): Specifies the element or isotope that the instrument will
examine in each measurement. The simplest way to specify entries is by
using the Periodic Table Panel. To use this panel, click in any cell in the
Analyte column, and then select the desired element and isotopes on the
Periodic Table panel.
You can also simply click in an analyte cell, and then type an element's
atomic symbol to indicate the intensity for that element in the sample.
Valid entries include the atomic symbols for all naturally occurring
elements, and the atomic symbol followed by “++”, such as Ca++, which
acquires data for the most abundant isotope, but at half the listed mass.
Molecular names for polyatomic species, such as ClO and ArN are also
valid entries. If you enter only the atomic symbol, the system automatically
enters the exact mass for the most abundant isotope (except for Ca, Cd,
Fe, Ni, Se, Sn, Ti, and Zn). For example, if you type Pb, the system will
enter 207.977 in the Mass column. You may also enter the atomic symbol
followed by the specific isotope of interest. For example, if you type
Pb206, the system will enter 205.975 for the mass.
• Mass (amu): Specifies the mass at which the isotope will be measured.
This entry normally does not need to be made, because the system
automatically fills in a mass parameter based on the analyte specified. If
the Scan mode is set to Scanning, this entry represents the center of the
measured peak.
• Begin Mass (amu): (Data Only and TotalQuant methods only) Defines the
start of an atomic mass range for each measurement. The scanning range
defined by a single row in the table is the range between the Begin Mass
and End Mass values set in that specific row. The total mass range that
the instrument will scan during an individual replicate is defined by all the
entries in the table. Valid entries are between 1 and 285 amu. To enter a
Begin Mass value, click in the relevant cell in the table and type a value. In
TotalQuant analyses, detectable isotopes for the naturally occurring
elements range from mass 6 (Li) to mass 238 (U). Because TotalQuant
methods use all the information available in the mass range, you must
gather data over the entire mass range that provides useful information in
your specific samples. The following additional rules should be observed
when parameter Begin Mass and End Mass values:
Because of the large amount of oxygen present in solution samples,
the measured intensities for the oxygen isotopes (16, 17, 18) are
usually beyond the working range of the instrument. It is generally best
to avoid acquiring data in this mass range, unless you are working with
laser sampling or other techniques that do not introduce water vapor
into the mass spectrometer.
Exclude masses 40 and 41 from the detection range. These masses
represent 40Ar and 41ArH, both of which are present in concentrations
beyond the working range of the instrument.
It is not usually necessary to collect data for masses above 209,
except for masses 232 to 238 where thorium and uranium occur,
because no stable element isotopes occur in that range.
If you specify a mass range that omits all the isotopes for an element,
you will receive no reported concentration for that element. However, if
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 179
Software Reference Guide
where:
Analyte Isotope is set using Analyte and Mass entries
Reference Isotope is set using the Reference Mass
• Scan Mode (*): Specifies whether the system will use either peak hop
data acquisition or peak scanning.Two options are available:
Peak Hopping: The instrument acquires data at the specified mass
only, then jumps directly to the mass for the next analyte in the method.
Peak hopping causes the software to spend more time at each mass,
reducing the number of readings required to achieve satisfactory
precision.
Scanning: The instrument acquires data at a number of data points
around the specified mass, based on the value of the MCA Channels
entry.
• MCA Channels: Indicates the number of multichannel analyzer channels
assigned to an analyte when using scanning mode. The MCA Channels
parameter is always 1 when using peak hopping. Valid entries are
between 1 and 20 for scanning. The default value is 1. Spectra are
scanned by rapidly and repetitively sweeping the quadrupole across all of
the channels within the mass range set by the Analyte/Mass entries.
During each sweep, the spectrometer acquires data for a length of time
determined by the dwell time, and the ion count rate is measured and
stored in an MCA channel buffer. The use of a multichannel analyzer
permits enhanced analytical precision by permitting rapid scanning of the
defined mass range, with all elements measured closely to one another in
time.
• Dwell Time per AMU (ms): Indicates the length of time spent measuring
the analyte during a single sweep. The Dwell Time per AMU is the total
time spent at the mass range corresponding to the individual analyte
listed. If more than one MCA channel is being used, this time is divided by
the number of MCA channels to determine the length of time spent at each
MCA channel. The acceptable range of parameters is between 0.1 and 1
800 000 ms. The default value is 50 in both peak hop and scanning
180 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
modes. The Dwell Time per AMU parameter applies only to the individual
analyte listed in the same row of the table. The optimum dwell time
parameter depends upon your analysis conditions and your sample.
Typical dwell time values for peak hopping are between 20 and 100 ms
using multiple sweeps per reading for a total integration time of 1000 ms. A
longer dwell time measures the analyte more precisely at the expense of
total analysis time. The use of shorter dwell times is recommended only for
transient analysis when determining a large number of analytes. However,
attempting to examine too many analytes using very short dwell times in a
short-duration transient signal will adversely affect accuracy and precision.
In Isotope Ratio determinations, it is important to collect enough analyte
ions of each isotope for an accurate calculation of the true ratio. The dwell
time of a minor isotope should be increased such that an equivalent
number of ions are counted for each isotope.
UCT instruments
For analytes using a DRC mode profile with an RPa value of 0, the
minimum dwell time is automatically raised to 10 ms. If an analyte uses a
DRC mode profile and the RPa value is not 0, the minimum dwell time is
automatically raised to 33 ms. If an analyte's flow rate or RPa value is
changed, putting the analyte in DRC mode, and the dwell time becomes
invalid, the dwell time is automatically changed by the software.
• Integration Time (ms): Displays the total amount of time that the analyte
listed will be measured during one replicate, based on the parameters for
Dwell Time per AMU, Sweeps/Reading, and Readings/Replicate as
follows:
Note: You can create and save methods using one or more unassigned
acquisition profiles (that is, profiles that are not currently assigned to a physical
gas channel), but you can run analyses and optimizations only when all of the
profiles in the methods involved have been assigned to an active channel.
• <Gas> Flow columns: UCT instruments only. These columns display the
gas flow rate for the reaction gases associated with the selected and
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 181
Software Reference Guide
assigned profiles. You can also specify multiple gas flows for a single
analyte to mix gases; the cell gas delays of the main gas will be applied.
The default value for each is 0.
• RPa (Rejection Parameter a): UCT instruments only. Displays the RPa
voltage for all the analytes as specified on the Conditions panel Cell
Parameters tab. The default value is 0.
• RPq (Rejection Parameter q): UCT instruments only. Displays the RPq
voltage for all the analytes as specified on the Conditions panel Cell
Parameters tab. The default value is 0.25
Periodic Table Panel
Use this panel to add analytes to your method by selecting them from a graphic
display of the periodic table.
182 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
Processing Tab
Use this tab to specify which detector signals to monitor, and to define the manner
in which raw signal data will be managed by the software. These parameters
include important handling options for both the signal profile and spectral peak
data.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 183
Software Reference Guide
184 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
define the manner in which signal data will be handled during ICP-MS
determinations. The available signal processing algorithms are as follows:
Average: Calculates the average intensity for all the readings in the
replicate.
Sum: Adds the total ion counts measured for all readings.
Amplitude: Indicates the maximum intensity point for all the readings
acquired.
None: Specifies that no signal processing should be performed on the
acquired data.
You can then set the number of Baseline Readings and a Smoothing
Factor, if applicable. The Baseline Readings entry defines the number of
readings that will be used in baseline correction; it is used to calculate a
baseline intensity value that is subsequently subtracted from all
measurements in the replicate. This baseline count can correct for solvent
effects or other sample induced intensity shifts.
Use the Apply Smoothing function to apply a Savitzky-Golay moving point
average smoothing function to the signal profile data. Smoothing tends to
reduce the effects of measurement noise in low signal-to-noise
measurements. The number of points used in the moving point average
calculation is determined by the value of the Factor entry. The minimum
number of points is five. The larger the value selected, the greater the
smoothing effects.
• Measurement Unit: The Measurement Unit group defines the manner in
which the raw data will be measured:
cps: Data is displayed using counts per second as the raw data.
counts: Data is displayed using total counts as the raw data.
Equation Tab
Use this tab to view isotope abundance information, to identify potential
interferences for your analytes, and to define interference corrections for use
during determinations.
Note: The fields that appear on this tab vary, depending on the type of method.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 185
Software Reference Guide
186 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
Analyte (*): Defines the atomic mass of the isotope that the instrument
will examine in each measurement. The simplest way to add analytes
is through the Periodic Table panel — click in any cell in the Analyte
column and select the required elements and isotopes. Alternatively,
you can click in a cell and manually type an element's atomic symbol to
determine the intensity for that element in the sample. Valid entries
include the atomic symbols for all naturally occurring elements. You
can also type the atomic symbol followed by “++”, such as Ca++, to
acquire data for the most abundant isotope, but at half the listed mass.
Molecular names for polyatomic species, such as ClO and ArN, are
also valid entries. If you type only the atomic symbol, the system
automatically enters the exact mass for the most abundant isotope
(except for Ca, Cd, Fe, Ni, Se, Sn, Ti, and Zn). For example, if you type
Pb, the system enters 207.977 in the Mass column. You can also type
the atomic symbol followed by the specific isotope of interest. For
example, if you type Pb206, the system shows 205.975 for the Mass.
Mass (amu): Here you specify the mass at which the isotope will be
measured. This entry normally does not need to be made, because the
system automatically populates the mass parameter based on the
analyte specified. If the Scan mode is set to Scanning, this entry
represents the center of the measured peak.
End Mass (amu): (Data Only methods only) Specifies the end of an
atomic mass range for each measurement. The scanning range
defined by a single row in the table is the range between the Begin
Mass and End Mass values set in that specific row. The total mass
range that the instrument will scan during an individual replicate is
defined by all the entries in the table. Valid entries are between 1 and
270 amu. To enter an End Mass value, click in the relevant cell in the
table and type a value.
Reference Mass: (Isotope Ratio methods only) Displays the reference
mass used in the ratio calculations. The Isotope Ratio compares the
isotope of interest to reference isotopes of the same element.
Isotope Ratio = Analyte Isotope/Reference Isotope
Elemental Equation: (TotalQuant methods only) Lists any
preprogrammed elemental equations included with the software.
Corrections: This parameter defines the elemental equations to be
used for interference correction. Many quantitation problems produced
by spectral interferences can be compensated for by using an
elemental equation.
Potential Interferences: This column lists the known interferences,
isotopes and polyatomic ions, for an isotope listed in the Equations
table. The software includes a comprehensive listing of the potential
interferences for a determination, helping you to define problems more
easily and to simplify correction. These standard equations correct for
isotope interferences, not for molecular interferences that may occur.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 187
Software Reference Guide
Calibration Tab
Use this tab to set up a calibration for use in your determinations. Here you can
define the composition of your standards and specify the concentrations of each
analyte in each standard solution. You can define up to 30 standards in a single
calibration. This tab is not available for the Data Only method type.
188 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 189
Software Reference Guide
maximum of four decimal places. If the Scan mode is set to Scanning, this
parameter represents the center of the measured peak.
• Curve Type (*): Specifies the type of calibration curve that will be
constructed from the calibration information for the specific analyte listed
on the same row of the Calibration table. To select a curve type, right-click
in the Curve Type field and then click a type. The following options are
available:
Simple Linear: Uses standard linear regression for curve calculation,
with equal weighting applied to all calibration solutions
Linear Thru Zero: Uses linear regression with a forced zero, that is, the
curve includes a point at the origin of the calibration graph
Weighted Linear: Uses a linear regression algorithm that incorporates
a standard, concentration-dependent weighting factor to emphasize
measurements in the low concentration region of the calibration curve.
The weighting factor (w) equals the inverse of the square of the
concentration of the standard.
• Sample Units (*): Specifies the concentration units for the sample
solutions. The following concentration units are available:
ppm: parts per million
ppb: parts per billion
ppt: parts per trillion
mg/L: milligrams per liter
µg/L: micrograms per liter
ng/L: nanograms per liter
µg/g: micrograms per gram
ng/g: nanograms per gram
mg/dL: milligrams per deciliter
µg/dL: micrograms per deciliter
ng/dL: nanograms per deciliter
µg/mL: micrograms per milliliter
ng/mL: nanograms per milliliter
pg/mL: picograms per milliliter
190 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
Sampling Tab
Use this tab to define the sampling device parameters for all analytic solutions
within the method. The information entered here includes general information
about the type of autosampler, plus information about pumping rates and times.
Note: Only one sampling device can be active in addition to the peristaltic pump.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 191
Software Reference Guide
When you select External, the External Read Trigger dialog box
192 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 193
Software Reference Guide
Report Tab
Use the Report tab to specify the type of report you want to produce and the
report template to be used. You can also specify the creation of a results file in tab
or comma delimited ASCII format.
194 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Method Development
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 195
Software Reference Guide
Notes Tab
Use the Notes tab to record notes about the method or about specific procedures
that may be required when using the method.
196 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Chapter 6
Interference Correction
This section provides information on interferences that can affect your ICP-MS
results; how to identify these problems in your data; and procedures for correcting
for these defects.
Topics in this section include:
• Interferences in ICP-MS on page 197
• Interference Identification on page 200
Interferences in ICP-MS
The ICP-MS technique offers numerous analytical benefits. However, during
method development, it is necessary to understand the influence of spectral
interferences that can affect your final results. An interference is anything that
causes the signal from an analyte to be different from the same signal for the
same concentration of that analyte in a calibration solution.
As with all other atomic spectrometric analytical techniques, interferences of
various types can occur during routine ICP-MS determinations. Interferences
associated with sample introduction using a nebulizer or spray chamber are
common to atomic absorption, optical emission ICP, and ICP-MS. These are
termed transport interferences. In addition, interferences can occur from the
presence of other isotopes or elements with the same atomic weight or mass
number as an analyte of interest. These are termed spectral interferences.
Matrix Interferences
Sample introduction effects when using a nebulizer or spray chamber system are
common in ICP-MS determinations. These interferences are generally the result
of either sample matrix effects that influence aerosol formation or the formation of
ions in the plasma (surface tension, viscosity). Complex sample matrices may
interfere with focusing ions into the sample and skimmer cone orifices, and
focusing and transporting ions through the ion lenses and quadrupole.
Matrix interferences can be identified by:
• Observing the intensity of the argon dimer, (Ar 40)(Ar 40), versus the
blank. The argon dimer peak and analyte element signals are usually
affected similarly by matrix interferences
• Determining if response is linear with dilution. Response dependence on
the absolute concentrations of analyte and matrix is an indicator of a
matrix interference
• Analyzing at low and high nebulizer argon flow rates
The most commonly observed result of matrix interference is the ion intensity if an
analyte element becomes dependent upon the total composition of the sample.
Typical signal suppression is observed. For example, a given concentration of
197
Software Reference Guide
selenium will exhibit a signal approximately 25% lower in a 1 000 mg/L NaCl
solution than the same concentration of selenium in water. This signal attenuation
is mass dependent. Light analytes in a predominantly high mass matrix (for
example, 1000 mg/L U) are more prone to signal attenuation than a heavy analyte
in the presence of a light mass matrix (for example, 1 000 mg/L Li).
Several methods can be used to compensate for matrix-induced signal
suppression. For a well-defined matrix, it is simple to match the composition of the
standard solutions with the sample solutions, and thus obtain the necessary
analytical accuracy. For a matrix of unknown composition, standard addition
calibration, in which known concentration standards are added directly to a
sample solution, provides a method for compensating for the sensitivity difference
between samples and standards.
Spectral Interferences
Spectral interferences are the result of other chemical species, such as isotopes
or ions, which are present at the same atomic mass as the analyte of interest.
Spectral interferences generally occur from the sources listed in the following
sections.
Isobaric Overlaps
Because most elements have more than one naturally occurring isotope, it is
possible for the mass spectrum of an isotope of one element to directly overlap
that of an isotope of another element. These interferences are termed isobaric
overlaps. All elements, except indium, have at least one isotope free from overlap.
Therefore, you can overcome this problem by selecting of isotopes carefully when
developing your methods.
Minor overlaps are corrected by measuring the intensity of the interfering element
at one of its major isotopes. Then, using published isotope ratios, it is possible to
subtract the intensity of the interfering isotope that is contributing to the analyte
signal. Using this principle, the software uses built-in calculation routines to
automatically correct for known isobaric overlaps when calculating analyte
concentrations in samples. In general, you do not need to make any further
adjustments to the system, assuming that the default calculations are not
modified.
Plasma-Induced Polyatomic Ions
The presence of atmospheric gases or the argon carrier are sources of potential
interfering ions as the result of reaction with other analyte or matrix components.
These interfering ions occur as polyatomic molecular ions. The following table lists
the most severe of the commonly occurring overlaps:
Of these, only the overlap of the most abundant isotope of argon with calcium
precludes analysis at the preferred isotope. In the case of calcium, determinations
198 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Interference Correction
Sulfur detection limits are affected both by poor ionization in the ICP (14%),
resulting in low ion intensity, and by the overlap of the primary sulfur ion by
diatomic oxygen. Unless the oxygen entering the plasma can be reduced to trace
levels, the overlap forces the use of a less abundant sulfur isotope S34 for
determinations.
Matrix Solvent-Induced Polyatomic Ions
The solvent or acid in which the sample is dissolved can also be a source of
interfering polyatomic ions. Chlorine from hydrochloric or perchloric acid and
sulfur from sulfuric acid all form polyatomic ions with argon and other plasma
gases. For organic solvents, carbon and oxygen can form polyatomic ions. The
most severe solvent-induced interference is from CO on silicon. All three silicon
isotopes are overlapped by carbon species, precluding the determination of
silicon in organic matrices. The major solvent-induced overlaps are listed in the
following table:
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 199
Software Reference Guide
Interference Identification
Whenever samples of unknown composition are received by an ICP-MS
laboratory, the method development and selection process involves a careful,
stepwise approach to determining potential interference problems and to defining
a solution to these interferences. This approach begins by performing a
TotalQuant™ analysis on the solution to determine the following:
• Elements present at measurable concentrations
• The concentration ranges for the elements of interest
• Interfering molecular ions
•Masses at which no elements are present, where internal standards could
be defined
This same information could also be determined by performing a full mass scan
and manually reviewing the spectral results in the Interactive panel. However, the
TotalQuant approach provides a fully automated means of gathering both spectral
and semi-quantitative data.
Using this information, you can determine which element isotopes to use to best
avoid potential interference problems, plus you can determine the best mass
ranges for internal standards.
Interference Correction
This section describes various interference correction strategies, and provides
relevant examples.
200 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Interference Correction
For example, if you want to examine the isotopes 56Fe and Ar40O16, you will
encounter an isotopic interference, as the mass of the two isotopes is identical. To
deal with this:
• In standard mode, select a less abundant isotope, such as 54Fe rather
than 56Fe, to avoid the interference
• In KED mode, helium gas introduced into the cell reduces the interferent
isotopes through collision and kinetic energy discrimination
• In DRC mode, the cell gas triggers an ion-molecule reaction that obviates
the interference, without affecting the isotope of interest:
NH3 + 56ArO+xxxxxNH3+ + 56Ar O
NH3 + 56Fe+xxxxxx no reaction
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 201
Software Reference Guide
202 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Interference Correction
(typically 1e9 cps), then the analyte goes into analog shutdown. Analog shutdown
values are also denoted by an S. In this case, the true value cannot be
determined unless the solution is diluted and reanalyzed. When the analyte
intensity cannot be accurately determined because a correction mass is in
shutdown, the intensity is denoted by an E on the plot to signal an elemental
equation error. This E point is included in the regression automatically.
When acquiring data in pulse mode, analytes whose correction equation has
some terms in shutdown are calculated by substituting the value of 2e6 for those
terms. The resulting value is displayed in brackets; for example:
.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 203
Software Reference Guide
204 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Interference Correction
specified the isotope to use as the base intensity measurement, you can enter
additional corrections as required by your sample.
For example, the major isotope of selenium occurs at mass 80. This mass is also
subject to interference from a plasma-induced molecular ion overlap caused by
(Ar 40)(Ar 40)+ ions. The argon intensity is generally much greater than the
selenium signal, so you would usually select an alternate isotope. For this
example, we will select Se 82, which has a relative abundance of 8.73.
First, we set the abundance factor by calculating the inverse of the normal natural
abundance, multiplying the quotient by 100.
[(1 / 8.73) * 100] = 11.45
Next, we set the determination mass to 82. However, the signal at mass 82 is also
subject to contribution from Kr 82, so a correction must be programmed.
To correct for Kr 82, we must find a non-interfering Kr signal. Here, we will use
Kr 83. Because the ratio of the natural abundance of Kr at mass 82 and mass 83
is nearly identical, the abundance factor is 1.00. And, because we must subtract
the contribution from Kr, we will set the factor to 1.00. Thus, our equation for the
calculations is:
I (Se) = 11.45 * [I (82) (abundance (Kr 82) / abundance (Kr 83) *I (83))]
where:
I = intensity of the specified mass
abundance = natural abundance of the isotope
The entries to be made in the Corrections column of the Equation tab are:
11.45 * (mass 82 1.00 * mass 83)
In practice, you would compensate for the Kr overlap by performing a blank
subtraction before analyzing the sample. The assumption is that the Kr comes
from the argon gas supply, and is the same for both samples and blanks.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 205
Software Reference Guide
206 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Chapter 7
Sampling Systems
In addition to pneumatic nebulization, the software incorporates control
parameters for different sample preparation and delivery options.
Topics in this section include:
• FIAS™ Sampling Systems on page 207
207
Software Reference Guide
208 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Sampling Systems
• Duration: Specifies the length of time for a step in the FIAS program. One
or both pumps may be utilized in any single program step. The acceptable
range is 0 - 99 seconds. A value of “-” specifies continuous operation.
• Pump 1 / Pump 2: Defines the rotation rate in rpm at which the FIAS
pumps will operate. A positive entry indicates clockwise rotation of the
pump, while a negative entry specifies counterclockwise rotation. The
acceptable range is -10 to -120, 0, and 10 to 120. The actual flow rate
depends on the inner diameter of the tubing used and the speed of
rotation of the pump. You should calibrate this rate every day.
• Valve: This is the position of the FIAS valve. Use Position 1 for filling the
sampling valve, and Position 2 to inject the sample into the ICP-MS.
• A/S (Autosampler) Location: In this column you will enter the rack
location to which the autosampler probe should move before the
corresponding row or step is executed. This location is the “start point” of
sampling and is available to all rows including the Pre-sample and Post-
run steps. An autosampler must be used in conjunction with the FIAS
sampling device. There are three ways to specify the autosampler location
in this column:
As an absolute number, such as 10 (indicates an actual rack location)
As a relative number, such as +5 or -5 (indicates a location relative to a
reference A/S location)
As a blank entry (the reference A/S location).
No pre-run error checking on the validity of the A/S location entered is
done by the software. If an incorrect A/S location is entered, the error
message Autosampler position is out of range appears during the run and
the FIAS program stops. Any partial data collected is not saved to the
dataset. The Reference A/S Location depends on the context in which the
FIAS program is executed:
If the FIAS program is applied to a manual or batch sample analysis,
the Reference A/S Location is the A/S Loc. entry in the Details dialog
box on the Sample panel Manual tab or Batch tab
If the FIAS program is applied to a Blank or Standard, the Reference
A/S Location is the net resultant location specified by the A/S Loc.
entry in the Method panel Sampling tab
If the FIAS program is applied to a QC Standard, the Reference A/S
Location is the A/S Loc. entry in the Method panel QC/Autosampler tab
Based on these rules, a net A/S location is calculated for each step of the
FIAS program. The A/S probe is moved to the net A/S location prior to the
execution of the corresponding FIAS step. If the net A/S location is the
same for two consecutive FIAS steps, the A/S probe is not moved
between the FIAS steps. You can also define the Reference A/S Location
in the Reference Autosampler Location field in the FIAS tab of the Device
Control panel. When Run Current Step or Run Current Program are
selected, this A/S location is used as the Reference Location for blank or
relative autosampler location entries on the Method panel Devices/FIAS
tab. You must fill in this field prior to selecting Run Current Step/Program
or an error message appears.
• Switches (2, 3, 4): Activates switch closures 2, 3 and 4 on the back of the
FIAS device at the beginning of the program step in which the action is
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 209
Software Reference Guide
210 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Sampling Systems
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 211
Software Reference Guide
212 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Chapter 8
Quality Control
This section describes the software quality control (QC) functions for quantitative
measurements.
Topics in this section include:
• Quality Control Overview on page 213
• Instrument Actions and Quality Control Functions on page 214
• Quality Control Software Reference on page 216
213
Software Reference Guide
214 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Quality Control
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 215
Software Reference Guide
Basic QC parameters
You can set up the software to automatically perform a variety of QC functions. In
order to take advantage of these functions, you must first configure several areas
of the software:
• The QC checking algorithms are active only during a batch analysis. The
sample batch file, which controls the autosampler for samples, must be
configured correctly.
• The QC Checking Enable check box on the Method panel Timing tab must
be selected.
• All true values and limits must be entered into the various screens in the
software for which QC checking is needed. It is important to note that the
QC checks are embedded in the individual methods, and not stored in a
separate file. This means that for each different set of QC standards or
limits, a new method must be created.
• The report options file must have the QC functions and the necessary
fields selected in order for you to actually see the results from the QC
checking calculations.
QC Software Conventions
The software employs several conventions regarding the QC functions:
• When a control label has a star in parentheses (*) in it, this indicates that
there are additional options available through a right-click menu. This
function also appears in other areas of the software.
• You must complete the calibration and blank information in the Method
panel for the standards used to calibrate the instrument; the QC
Autosampler tab provides the location for the calibration blank and
standards only. Note that, for QC purposes, it is not sufficient to simply
complete the calibration, blank, and standard locations on the Sample
panel Batch tab.
• For flagged QC samples (QC spike, QC duplicate, QC dilution, and so
forth), a detailed name or description may be entered in the Description
column of the Sample panel Batch tab. Also, if dilution factors are entered
using the Aliquot and Diluted to Volume columns on the Sample panel
Batch tab, the dilution factors for the reference sample and the flagged QC
sample must be the same.
216 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Quality Control
you can measure when the system's overall sensitivity has degraded
significantly. Intercepts can be monitored so that the zero point in the
calibration curve is never too far from the original value obtained. The
correlation coefficient can be monitored to assure those calibration
standard data are well correlated to the calibration algorithm chosen.
• QC Standards tab: Used to set concentration limits for the QC standards
to measure if the system continues to provide accurate answers for
solutions of known concentrations.
• QC Measurement Frequency tab: Used to specify the frequency with
which QC measurements are made.
• QC Standard Internal Standards tab: Used to set up the parameters for
measuring an internal standard element within the QC standards. This
feature enables the definition of limits on the Internal Standard Intensity
drift for QC standards.
• Calibration Standards tab: Used when calibration standards are
measuring how well the system is performing. The standard deviation of
the blank or RSD of the standards can both be monitored. If either the
standard deviation or RSD of the standards is too high, the system may
not provide sufficient precision for your determinations. This may indicate
a problem with the sample introduction system or other hardware
components.
• Sample Internal Standards tab: Used to set up the measurement of the
internal standard element that can be monitored for drift of the sample.
This feature enables the setup of limits on the Internal Standard Intensity
drift for samples.
• Sample tab: While samples are being measured, you can monitor the
sample concentration, sample standard deviation, and sample RSD.
Sample concentrations can be monitored to determine when the results
are below or above a default measurement limit. As a result of this
monitoring, certain actions can be automatically executed when a sample
is out of range. Monitoring the sample standard deviation is especially
useful for estimating a detection limit on samples with no concentrations.
Sample RSDs can be monitored to see if the instrument system is
providing satisfactory precision, a measure of the performance of the
sample introduction system and other hardware components.
• Spike tab: Used to set spike recovery limits for a determination. Spike
recovery is performed by spiking samples with a known concentration of
an analyte, then measuring how accurately the system determines this
spike concentration.
• Dilution tab: Used to monitor dilution measurements on sample solutions.
The percent difference between a sample and a dilution of that sample can
be measured during determinations. You can then set a maximum percent
difference and have out-of-range samples identified.
• Duplicate tab: Used to measure the relative percent difference between
two aliquots of the same sample solution.
• Spike Tables tab: Used to set up spiking tables. A spike table contains the
spike concentrations. Spike concentrations are set in up to 20 tables for
use during the QC measurements.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 217
Software Reference Guide
Control Description
Analyte Specifies the element that the instrument will determine
in each measurement. Read‐only. To add or delete an
analyte, go to the Method panel Timing tab
Mass (amu) Specifies the mass at which the element will be
measured. Read‐only. The system sets the mass based on
the analyte specified. To change the mass, go to the
Method panel Timing tab
Action 1 (*) / Action 2 Specifies what action the system should take if the
(*) corresponding measurement is out of limits. Action 1 is
executed first, and is repeated as often as indicated on
the QC Action Controls tab if results are beyond the
specified limits. Action 2 is then performed, and is also
executed as often as indicated if results are beyond the
specified limits. If the results are still out of limits, the
analysis is either continued or stopped, depending on
the QC Action Control parameters
When a measurement is out of limits, the text is entered
in the Message to Print column for all the elements out
of limits. Select one of the following from the list. Note
that some tabs will have a sub‐set of the options listed
here:
• Continue: The out‐of‐limits message is printed, and
the analysis continues. This is the default action.
• Stop: The out‐of‐limits message is printed, and the
analysis stops
• Wash for X and Recalibrate: The out‐of‐limits
message is printed. The autosampler probe returns
to the wash position for X seconds. For X enter any
positive number. Then, all elements are recalibrated
by analyzing from the blank solution again
218 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Quality Control
Control Description
• Wash for X, Recalibrate and Rerun Current: The
autosampler probe returns to the wash position for X
seconds. For X enter any positive number. If zero is
entered, a wash is not done and only a recalibration
and rerun of the current standard will occur. The
failed item is recalibrated and rerun
• Wash for X, Recalibrate and Rerun Samples: The
autosampler probe returns to the wash position for X
seconds. For X enter any positive number. If zero is
entered, a wash is not done and only a recalibration
and rerun of the samples occurs. All the elements are
recalibrated. Recalibration is performed by analyzing
from the Blank solution again. For the Rerun Samples
portion of this action, the analysis resumes with the
first sample between the last passed QC and the
failed QC. For example, if you ran a blank, Standard 1,
Standard 2, QC1, samples 1 through 10, QC1 again,
samples 11 through 20 and QC1. If this QC1 now fails
the instrument recalibrates and resumes at sample
11
• Wash for X and Continue: The autosampler probe
returns to the wash position for X seconds. For X
enter any positive number. If zero is entered, a wash
is not done and the analysis continues. For the
Continued portion of this action, the analysis
continues with the next determined solution type
• Wash for X and Rerun Current: The autosampler
probe returns to the wash position for X seconds. For
X enter any positive number. If zero is entered, a
wash is not done and only a rerun of the current
blank or standard occurs. For the Rerun Current
portion of this action, the failed QC is rerun
• Wash for X and Rerun from Original: The
autosampler probe returns to the wash position for X
seconds. For X, enter any positive number. If zero is
entered, a wash is not done and only a rerun of the
original sample occurs. For the Rerun from Original
portion of this action, the analysis starts with the
original sample
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 219
Software Reference Guide
Control Description
• Wash for X, and Continue from Sample: The
autosampler probe returns to the wash position for X
seconds. For X, enter any positive number. If zero is
entered, a wash is not done and only a rerun of the
current blank or standard occurs. The analysis
continues with the sample specified. Select a sample
from where the analysis should resume
Action 1 Data / Action Displays any variable associated with Action 1 or Action
2 Data 2, such as the number of seconds specified to wash and
recalibrate
Message To Print Type a message to be printed in the summary section of
the report to warn that the element on that line is out of
limits. The message can be up to 78 characters in length.
For example, “Slope not within set limits”
QC Action Criteria (*) Specifies the conditions required for an action to be
taken. You can specify how many or which elements in
your QC standards must be out of limits before an action
is taken. For each element, right‐click and select one of
the following options from the list:
• Must Act: The action designated for that QC
standard will be carried out if that one element is out
of limits
220 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Quality Control
Control Description
• Do Not Act: The action designated for the QC
standard will not be carried out regardless of
whether it is out of limits. Any message specified in
the Message To Print column is printed so you can
monitor the element
• Act if X Elements Out: The action designated for
that QC standard will be carried out only if a specified
number of elements are out of limits. Any number
from 1 to the total number of elements can be
specified (this number must be the same for all
elements for which you specify this option)
QC Action Priority (*) Specify the relative priority of each analyte. The value
entered can be from 0 to 500. Enter a value of zero,
indicating Do Not Act, if the analyte is to be monitored
for information purposes only. Values from 1 to 500
indicate a Must Act priority. A value of 1 has the highest
priority and its actions are performed if that analyte is
out of limits. If no value is entered, the analyte is
assigned the default priority. The default sequence of
values starts from 1
The QC Action Priority column is used to tell the software
the priority of checking for the individual isotopes in the
checking routine. Only the action for the element out of
limits with the highest priority (lowest number) is carried
out to avoid multiple actions. In the event that multiple
isotopes have the same level of priority, action is taken
only if all are out of limits
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 221
Software Reference Guide
Calibration Tab
Use this tab to monitor calibration curve statistics, including the slopes, intercepts,
and correlation coefficients of the calibration curves for each analyte in your
method. The slope of the calibration curve is one measure of system sensitivity
and, by monitoring it over time, you can detect when the overall sensitivity has
degraded significantly. Monitor the intercepts to keep the zero point in the
calibration curve close to its original value. Track the correlation coefficient to
ensure calibration standards remain well-correlated to the calibration algorithm
chosen.
222 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Quality Control
or lower intercept can be entered for limit checking to be performed for that
analyte.
• Intercept Upper: Enter the upper intercept limit for each analyte except
internal standards. The upper and lower intercept limits determine the out-
of-limit conditions. Any positive or negative value is valid. Either an upper
or lower intercept can be entered for limit checking to be performed for that
analyte.
• Corr. Coef. (min): This limit determines the minimum acceptable
correlation coefficient for each analyte, except internal standards. Values
from 0.0000 to 1.0000 are valid. If no value is entered, limit checking is not
performed for that analyte.
• Measurement: Lists the types of measurement available, such as slope,
intercept, and correlation coefficient. Read-only.
QC Standards Tab
Use this tab to set concentration limits for the QC standards in order to track
system accuracy for solutions with known concentrations.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 223
Software Reference Guide
224 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Quality Control
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 225
Software Reference Guide
• Every # Samples: Specifies when the QC standards will be run during the
analysis by defining how many samples should be run between QC
standards. Type a number representing the number of samples to be run
between each QC standard.
Note: If the final QC Standard in this grouping happens to occur at the end of
your batch as the last sample, it will not be run. In order to run a QC Standard as
the final sample in a batch, you must explicitly designate it as the final sample, by
placing an X in the Final column on this tab.
• Before A/S Loc.: Specifies when the QC standards are run in relation to
the autosampler positions. For example, you can set a QC standard to run
before autosampler position 10. This tab has fifty Before A/S Location
columns. Use this to space the QC standards as best suits your analyses,
rather than in a fixed pattern like the Every # Samples column.
Note: When multiple QC standards are run (for example, QC Std 1, QC Std 2,
QC Std 3, and QC Std 4 have the Initial option selected), the QC Standards run in
numerical order: QC Std 1 then QC Std 2 then QC Std 3 and then QC Std 4.
226 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Quality Control
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 227
Software Reference Guide
228 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Quality Control
the Relative Value option and type a negative value in the Lower field (for
example, -10). Only numerical entries are accepted. After the entry is
accepted, the units appear on the Sample Int. Std tab as a percentage.
• Sample Int. Std Upper %: Right-click to display the Enter Percent Limit
Value dialog box. To enter an absolute value, select the Absolute Value
option and type a value (for example, 110). To enter a relative value, select
the Relative Value option and type a positive value in the Upper field (for
example, 10). Only numerical entries are accepted. After the entry is
accepted, the units appear on the Sample Int. Std tab as a percentage.
• Measurement: The list shown is determined by the internal standard
elements specified on the Timing tab. Read-only. To add or change an
entry, go to the tab on which it was made.
Sample Tab
You can monitor the sample concentration, sample standard deviation, and
sample RSD during a determination. Sample concentrations are monitored to
determine when the results are outside a default measurement limit. As a result of
this monitoring, certain actions can be automatically executed when a sample is
out of range. Monitoring the sample standard deviation is especially useful for
estimating a detection limit on samples with no concentrations. Sample RSDs are
monitored to determine if the system is providing satisfactory precision, which is a
measure of the performance of the sample introduction system and other
hardware components.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 229
Software Reference Guide
Any value is valid. The upper limit must be a larger value than the lower
limit. If the upper limit is exceeded the software will indicate either S for
shutdown, or EEE for elemental equation error.
• Sample Conc SD: Type an SD (standard deviation) value for the net
intensity. This determines the maximum acceptable value for each analyte.
Any positive value is valid. If a value is not entered, limit checking is not
performed.
• Sample Conc RSD: Type an RSD (relative standard deviation) value for
the net intensity. This determines the maximum acceptable value for each
analyte. Any positive value is valid. If a value is not entered, limit checking
is not performed.
• Measurement: Each element has four lines, listing the upper
concentration, lower concentration, SD and RSD. Read-only. To add or
change an entry, go to the tab on which it was made.
230 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Quality Control
Spike Tab
Use this tab to set spike recovery limits for a determination. Spike recovery
involves spiking samples with a known concentration of an analyte, then
measuring how accurately the system determines this spike concentration.
Note: If the spike recovery calculated for any spiked sample is below the lower
limit, a corresponding message is printed. To set a limit, type any positive value.
Spike % recovery is calculated as:
where:
US = unspiked sample result
SA = spike added
SS = spiked sample result.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 231
Software Reference Guide
If the original sample result is smaller than the detection limits specified in
the Spike Table used, zero is used in this equation instead of the original
result. This avoids subtracting negative sample results. Specifying the
detection limit is optional. If this field is empty, the original sample results
are used.
• Spike % Rec (upper) (*): To enter a spike recovery limit, right-click to
display the Enter Percent Limit Value dialog box. To enter an absolute
value, select the Absolute Value option and type a value (for example,
110). To enter a relative value, select the Relative Value option and type a
positive value in the Upper field (for example, 10). Only numerical entries
are accepted. After the entry is accepted, the units appear on the Spike tab
as a percentage.
Note: If the spike recovery calculated for any spiked sample is above the lower
limit a corresponding message is entered. To set a limit, type any positive value.
Spike % recovery is calculated as:
where:
US = unspiked sample result
SA = spike added
SS = spiked sample result
If the original sample result is smaller than the detection limits specified in
the Spike Table used, zero is used in this equation instead of the original
result. The defining detection limits are optional. If this field is empty, the
original sample results are used.
• Measurement: Lists the spikes for the elements specified on the Timing
tab. You cannot edit entries in this column. If you want to change an entry,
or add another measurement to monitor, you must change the entry on the
tab on which it was made, and then return.
232 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Quality Control
Dilution Tab
Use this tab to monitor dilution measurements on sample solutions. The percent
difference between a sample and a dilution of that sample can be measured
during determinations. You can then set a maximum percent difference to identify
out-of-range samples.
where:
OS = original undiluted sample result
DS = diluted sample result
DF = dilution factor
OS = original sample result (same name as dilution)
• Measurement: Lists dilutions for the elements specified on the Timing tab.
This column is read-only—to change an entry or add another
measurement to monitor, you must change the entry on the tab on which it
was made.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 233
Software Reference Guide
Duplicate Tab
Use this tab to measure the relative percent difference between two aliquots of the
same sample solution.
234 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Quality Control
where:
SS = spiked sample result
US = unspiked sample result
SA = spike added
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 235
Software Reference Guide
236 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Quality Control
QC actions Priority
Calibration limits: Correlation coefficient 1
Intercept 2
Slope 3
Sample limits: Lower concentration 2
Upper concentration 3
SD 4
RSD 5
Internal standard intensity 1
QC standard limits: Concentration or RSD 2
Internal standard intensity 1
Flagged sample Flagged calc. (Spike Rec. %Diff, 2
limits: Rel.%Diff.)
Internal standard intensity 1
Note: Some solution limit types in the QC Action Controls tab are further
prioritized internally by the software. When multiple limit conditions are exceeded
within a solution type, the action for the limit with the highest priority is executed.
You cannot modify these internal priorities.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 237
Software Reference Guide
Autosampler Tab
Use this tab to identify the location in the autosampler for each of the QC
standards.
238 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Chapter 9
Sample Analysis
Virtually all the parameters required for controlling a determination are contained
within the analytical method file. After selecting the method you want to use for a
determination, you set up and optimize the instrument. Next, you use the Sample
screen to identify samples, control the order in which they are measured, and
initiate the determination.
The main Sample screen consists of a single Sample panel divided into two tabs
to facilitate two different types of sample introduction:
• Manual: You introduce individual sample solutions into the spectrometer
• Batch: You group blanks, standards, and samples into batches to feed
them automatically into the spectrometer
The Sample screen is accessed via the Sample icon on the ribbon. It contains the
Sample panel, as well as instances of the Report View and Realtime panels,
allowing you to monitor your results in graphic and tabular form as they are run
(see Reporter™ Panels on page 261 and Realtime Panel on page 291 for details).
Following a determination, you can generate a report on the results of the
measurement, review and analyze the results in graphic form, reprocess the data,
and export the results.
239
Software Reference Guide
240 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Sample Preparation and Analysis
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 241
Software Reference Guide
Footer Section:
[End Samples]
[End SampleFile]
242 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Sample Preparation and Analysis
[Begin Samples]
0, "stability", "stab001", 1, 0, 0, 0, 0.1, 10, 100, 1000, 0.11, "Stability test",
"stability.mth", 0, n/a, 2
1, "stability", "stab002", 2, 1, 8, 0, 0.2, 20, 200, 2000, 0.22, "Stability test",
"stability.mth", 1, 0, 4
2, "stability", "stab003", 3, 2, 0, 77, 0.3, 30, 300, 3000, 0.33, "Stability test",
"stability.mth", 2, 1, n/a
3, "stability", "stab004", 4, 5, 0, 0, n/a, n/a, n/a, n/a, n/a, "Stability test",
"stability.mth", 3, n/a, 8
4, "stability", "stab005", 5, 4, 0, 0, n/a, n/a, n/a, n/a, n/a, "Stability test",
"stability.mth", 4, n/a, 10
5, "stability", "stab006", 6, 3, 0, 0, n/a, n/a, n/a, n/a, n/a, "Stability test",
"stability.mth", 5, 2, n/a
6, "stability", "stab007", 7, 6, 0, 0, n/a, n/a, n/a, n/a, n/a, "Stability test",
"stability.mth", 6, 3, 14
7, "stability", "stab008", 8, 0, 0, 0, n/a, n/a, n/a, n/a, n/a, "Stability test",
"stability.mth", 7, n/a, 16
[End Samples]
[End SampleFile]
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 243
Software Reference Guide
244 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Sample Preparation and Analysis
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 245
Software Reference Guide
• Summary: Specifies the summary report type and the destination to which
it will be directed.
• Build Run List: Click to open the Run List dialog box, where you can
review the sequence of samples prior to running them, set plasma
AutoStop options, insert priority samples into a running batch, or append
additional samples.
• Use Manual Sampling (No Autosampler): Select this check box if you do
not use an autosampler. Note that, if using manual sampling, you must
aspirate each sample in turn.
Items in the Batch Table
• A/S (Autosampler) Location: Lists the autosampler position for the
sample — the location that the solution occupies in the autosampler tray.
Type the number corresponding to the position in the autosampler tray
where the sample solution will be placed.
• Batch ID: Identifies a group of samples with a batch label. Type a name
(any alphanumeric combination) to enter an individual Batch ID or click the
column heading and, on the Edit menu, click Fill Down to automatically
enter a Batch ID for a group of samples.
• Sample ID: Identifies each sample with a unique name. Type a name (any
alphanumeric combination) to enter an individual Sample ID.
• Measurement Action (*): Specifies when to run standard and blank
solutions, in addition to specifying the running of the sample solution. To
select a calibration action, right-click the appropriate cell of the Samples
data table, and click a Measurement Action. The following options are
available:
Run Sample: Runs only the sample solution that is on the same line of
the Samples data table.
Run Blank, Stds. and Sample: Runs the solvent blank, the standards
specified on the Method panel Calibration tab, and the sample solution
that is on the same line of the Samples data table.
Run Blank and Sample: Runs the solvent blank solution and the
sample solution that is on the same line of the Samples data table.
Run Standards and Sample: Runs the standards specified on the
Calibration tab of the Method panel, and the sample solution that is on
the same line of the Samples data table. Typically used with Additions
calibration.
Run Diluted Sample: Runs the specified diluted sample. This is
applicable only if using an autodilution device.
Run Blank and Diluted Sample: Runs the solvent blank solution and
the specified diluted sample. This is applicable only if using an
autodilution device.
Run Standards and Diluted Sample: Runs the solvent standards and
the diluted sample as specified. This is applicable only if using an
autodilution device.
Run Blank, Standards and Diluted Sample: Runs the solvent blank and
the specified standards and diluted sample. This is applicable only if
using an autodilution device.
246 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Sample Preparation and Analysis
• Method (*): Specifies the method to be used for measuring the related
sample solution and for all following samples, until a different method is
specified. Right-click to browse for a method file.
• Description: Type additional information or comments about an individual
sample solution.
• Sample Type (*): Identifies samples used for a quality control
determination or as part of the TotalQuant calibration sequence. This entry
is required when using either a Quantitative Analysis method with a QC
function or when performing TotalQuant measurements using standard
additions calibration. Right-click the Sample Type field. The Sample Type
dialog box appears. Select the appropriate option for your sample:
Sample: Identifies the solution as a sample.
QC Spike: Identifies the sample as a spike. When selected, a dialog
box appears where you can select the Spike Table you want to use for
the spike, and the parent (Spike Of) sample reference location. You
create a Spike Table using the Spike Tables window in the method.
The Sample Type field displays Spike X of Y as the Sample Type,
where Y is the sample batch row number.
QC Dilution: Identifies the sample as a dilution. When selected, a
dialog box appears where you can select the Dilution Factor and the
parent (Dilution Of) sample reference location. The Sample Type field
displays Dilution DF:X of Y as the Sample Type, where Y is the
sample batch row number.
QC Duplicate: Identifies the sample as a duplicate. When selected, a
dialog box appears where you can select the parent (Duplicate Of)
sample reference location. The Sample Type field displays Duplicate
of Y as the Sample Type, where Y is the sample batch row number.
QC Reagent Blank: Identifies the sample as a reagent blank. If you
select this option, the system will subtract the concentration from all
subsequent samples until the next reagent blank. If you specify several
reagent blanks in succession, the system calculates the average
concentration of the set, then subtracts the average concentration from
all subsequent samples.
TQ Addition: Specifies that the sample contains specific added
calibration elements in known quantities, as required during a Sample
Addition calibration in TotalQuant analyses. This lets the instrument
calculate intensity values using the TotalQuant algorithms.
QC Duplicate Spike: Identifies the sample as a QC duplicate spike.
When selected, a dialog box appears for you to select the parent QC
spike sample (Duplicate Spike Of) reference location. The reference
location is the sample batch row number that the parent spike appears
on. The software performs an RPD (relative percentage difference)
calculation based on the results for the parent QC spike and the QC
duplicate spike. The Sample Type field displays Duplicate Spike of Y
as the Sample Type, where Y is the sample batch row number on
which the parent spike appears.
• Initial Sample Quantity (mg): Specifies the weight of the sample before
dilution.
• Sample Prep Volume (mL): Specifies the volume in which the initial
sample quantity has been diluted.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 247
Software Reference Guide
248 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Sample Preparation and Analysis
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 249
Software Reference Guide
Note: If you do not want the Measurement Status dialog box to appear during each scan,
on the Syngistix ball menu, click Help and then clear the check mark beside the
Show Measurement Status Dialog option.
250 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Sample Preparation and Analysis
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 251
Software Reference Guide
252 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Chapter 10
Data Reprocessing
This section discusses the use of the data reprocessing tools and describes the
typical circumstances under which reprocessing is performed. Reprocessing
involves reanalyzing the raw acquired data using either the same or different
processing parameters as those used when the data was initially acquired.
Topics in this section include:
• Previously Acquired Data on page 253
• Dataset File Organization on page 254
• Data Reprocessing Workflow on page 255
• Data and Parameter Considerations During Reprocessing on page 255
• Files Required for Reprocessing on page 256
• Dataset and Reprocessing Software on page 257
253
Software Reference Guide
Blank solutions: BLANK.000, BLANK.001, and so forth
STANDARD.001, STANDARD.002, and
Standards: so forth, or EXTERNAL STANDARD.002
for TotalQuant™ methods
SAMPLE.001, SAMPLE.002, and so
Samples: forth
Use the Dataset panel to view the contents of a dataset and reprocess the data
contained within.
254 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Data Reprocessing
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 255
Software Reference Guide
number of sweeps or dwell time than the method that was used to acquire the
data initially.
Note: If you previously used the NexION software, and you are reprocessing
data collected in the NexION software:
In the Reporter, the current standard's raw intensity is calculated and displayed; when
NexION data is reprocessed using original conditions and the next standard is
reprocessed, the software reads the calibration from the original dataset and, finding no
raw intensity value (as this measurement did not exist in the older software), reads the raw
intensity as zero. As a result, on the Reporter screen Calibration panel, the most recently
reprocessed standard displays a point with a non-zero intensity, and all earlier standards
are plotted with a zero calibration.
When reprocessing NexION data without original conditions, each point is simply
recalculated, and the Calibration panel is able to display points with correctly calculated
raw intensities.
To reprocess data using: The following files are needed:
the same method, but using modified original method file
data processing parameters
the same method, but using a different original method file; stored calibration
(stored) calibration file
a different method—include the new method file
standards during the reprocessing
a different method—use a stored new method file; stored calibration file
calibration and do not reprocess the
standards
any of the above combinations for a appropriate files from above, plus the
TotalQuant determination plus using a new response file
different response file
256 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Data Reprocessing
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 257
Software Reference Guide
is cleared when you click Reprocess, the conditions specified by the files
loaded in the workspace are used.
• Save Reprocessed Data: Used to specify whether or not you want to
save the reprocessed data in a new data file. If the Use Original Conditions
check box is selected, this check box is not available. If this check box is
selected when you click Reprocess, a new data file is created for each
reprocessed data file.
• Summary Report: Click to specify the summary report type and the
destination to which it will be directed.
• Load: This option loads one of the following file types from the open
dataset file into the current workspace. Select the file you want to open in
the dataset panel from the list, and then click Load.
Method: Loads the method (.mth) file used when the data was
acquired. This overwrites any current method parameters, so save the
current method if applicable.
Conditions: Loads the conditions (.dac) file that was used when the
data was acquired. If the dataset file is the result of a SmartTune™
optimization, it would contain the resultant values from the
optimization. Loading conditions from the dataset guarantees that the
dual detector calibration used at the time of data acquisition is also
used for reprocessing.
Note: If the method is also loaded, the .dac file associated with the
method will override this conditions file.
MassCal: Loads the mass calibration (.tun) file used when the data
was acquired.
Calibration: Loads the calibration (.cal) file used when the data was
acquired. If you want the calibration from a specific dataset file, you will
load the calibration that includes all calibration standards up to and
including that dataset file.
RSP: Loads the response (.rsp) file that was used when the data was
acquired. When a TotalQuant method is loaded in the workspace, the
TotalQuant.rsp file from the current project folder is opened.
Report Option: Loads the report options (.rop) file that was used when
the data was acquired.
Configuration: Loads the configuration parameters that were used
when the data was acquired. If you select this option, the System
Configuration From Dataset dialog box appears. To apply the
displayed parameters in the current session, click Apply. Those
parameters are then available in the System Configuration dialog box.
Columns in the Dataset table
• Batch ID: Specifies the alphanumeric name used to identify the batch of
samples. If no Batch ID was used, this field is empty. For some sample
types, such as SmartTune optimizations, the sample type is displayed.
• Sample ID: Lists the unique alphanumeric name used to identify each
sample solution run. Because this is a required field, an entry will always
appear in this column.
258 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Data Reprocessing
• Acquisition Date/Time: Specifies the exact month, day, year, and time at
which the original data was acquired. If the data was acquired on a
computer in the same time zone and daylight savings time bias as where
the time is being displayed, the local time is displayed. If the data was
acquired in a different time zone than the one where the time is being
displayed, the time zone name is also displayed. If the data was acquired
with a computer and operating system set to a language other than
English, the time zone is displayed in that language. The time and date are
displayed in the format specified in the regional parameters of Windows for
the current user.
• Method: Lists the name of the method file that was used to acquire the
data for the sample/standard/blank listed. The listing includes the
complete pathname for the method file; for example,
C:\Users\<Public>\Documents\PerkinElmer Syngistix\ICPMS\Method\pbtest.mth
• Description: Lists any notes or descriptive information entered in the
Sample panel at the time of determination.
• Read Type (*): Describes the original sample type designation for an
individual measurement and permits the operator to re-designate the
solution as a different type to permit analysis of the effect on a
determination. To view the list of available read types, right-click in the
appropriate cell of the data table. Select a sample type and click OK. The
following options are available:
Blank
Sample
Quant External Standard
Quant Standard Addition
Unspiked
TotalQuant External Standard
TotalQuant Sample Addition
Isotope Ratio Standard
Isotope Dilution Standard
QC Standard
QC Spike
QC Dilution
QC Duplicate
QC Reagent Blank
QC Duplicate Spike
Selection of the Quant External Standard or Quant Standard Addition
option includes arrows that are used to designate which standard number
the solution should be used as, because more than one standard may
have been used to construct the calibration.
• Sample File Name: Lists the name assigned to the individual data record
(the individual sample, standard, or blank solution) within the dataset. The
file name is created by the software, using the Sample ID from the Sample
panel when the data was acquired. If no entry was made at acquisition
time, the system describes the solution using a default name:
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 259
Software Reference Guide
Blank BLANK.000, BLANK.001, and so forth
solutions:
STANDARD.001, STANDARD.002, and so forth, or
Standards: EXTERNAL STANDARD.002 for TotalQuant methods
Samples: SAMPLE.001, SAMPLE.002, and so forth
260 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Chapter 11
Reports
This section describes the use of the Reporter™ functions. Included is information
on using the reporting templates to set up detailed sample reports; viewing live
sample batch data as you run your analyses; and producing hardcopy reports or
exporting data for analysis outside of the Syngistix™ software.
Topics in this section include:
• Reporter™ Panels on page 261
• Report Views on page 262
• Enhanced Security Report View functions on page 275
• Report Options on page 282
• Report Options Templates on page 282
• Report Appearance on page 283
• Summary Reports on page 287
• Intensity versus Time Reports on page 287
• Reported Shutdown Values on page 263
• Exporting Current Sample Reports on page 288
Reporter™ Panels
The software provides three main panels on two Reporter screens to help you
track your analyses and create informative and attractive reports.
The Reporter icon accesses the Report View screen, where you view detailed
determination results for the current sample, and scrolling data for each sample in
your batch. The arrow submenu provides access to the Report Options screen,
where you define summary report contents and layout.
The arrow submenu also provides controls for managing QC (Quality Control)
functions. The following commands are available:
Restart QC -- Click to restart a quality control analysis that has stopped
due to a QC action
Clear QC Reagent Blank -- Click to clear reagent blank data that is
stored as part of a QC analysis
Clear QC Logging Data -- Click to clear all QC logging data that is
stored with the dataset
The Reporter panels are as follows:
• Report View: Use this panel to view detailed information about your
current sample and to track data from multiple analyses as your sample
batches are processed. Scrolling results tables provide Raw and Net
Intensity views that display the measured signal data for each element
examined by the instrument, with or without related RSDs; a
Concentration view; an Unfactored Concentration view; a graphic
261
Software Reference Guide
Report Views
The Report View panel is made up of multiple tabs, each of which present the
running analytic data from your samples in a different way: these are the Current
Sample tab, Raw and Net Intensities tabs, Concentrations tab, Unfactored
Concentrations tab, Internal Standards tab, and QC tab.
Note: The Current Sample report is available for all analyses, while the Raw and Net
Intensities, Concentrations, Unfactored Concentrations, Internal Standards, and QC tabs
are populated only when running Quantitative Analyses.
When you initially start a new software session, or after you click Clear Data, the
Report View tabs are empty. They become populated as data is acquired,
regardless of whether or not the panel is open. As you run samples, the tabs are
updated in real time. The table-based tabs collect a scrolling list of sample results
collected during quantitative analyses to a maximum of 1000 samples, after which
the first samples scroll off the top of the list as new samples are added at the
bottom. Index numbering of the developing list goes up to a maximum count of
9,999 before rolling over to begin at 1 again.
To help you keep track of where you are in the data, your position in the scrolling
list on any table is retained when you move between the Raw and Net Intensities,
Concentrations, and Unfactored Concentrations tabs, so that you can compare
multiple types of data for a single sample without losing your place.
In the tables, sample data is displayed in alternating rows of white and gray to
improve readability. Additional color cues employed in the tables include the use
of a red font or cell background to indicate QC failures, and the Syngistix software
standard KED mode green and DRC mode yellow used in the column headers to
indicate analytes using profiles in these modes (UCT instruments only). On the
Concentrations tab, a purple font is used to denote internal standards, which are
also identified with an (IS) in the relevant analyte header. Note that all tables take
their base colors from your system colors; if you change from the default Windows
system colors, you may change the table background colors.
If you are using the Enhanced Security software, workflow change and traceability
management functionality can also be added, so that changes and analyses can
be reviewed and approved (or rejected), and e-signature and reason data
captured and displayed in the results tables.
262 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Reports
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 263
Software Reference Guide
264 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Reports
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 265
Software Reference Guide
266 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Reports
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 267
Software Reference Guide
Concentrations tab
Applies to Quantitative Analyses only. The Concentrations tab displays the
adjusted sample concentration for each element measured by the instrument.
This tab also consists of a tabular display with the sample name in the first column
and masses in subsequent columns with the concentration units directly beneath.
The column headings consist of the analytes as named in the related method
along with the mass in amu. The relevant units are also displayed, again as
defined in the method, below the analyte name in parentheses; samples and
standards may be measured in different units, as required. Each row represents a
single sample.
Analytes that are acting as Internal Standards are identified with an (IS) in the
relevant analyte header, and a purple font is used to highlight the internal standard
values within the table (where a valid blank with internal standards has been
applied). For each internal standard, the IS percent recovery value is displayed,
together with the related intensity RSD (if RSDs are displayed).
Note: If you are reprocessing data with original conditions, the percent recovery value
does not appear.
Select the Show RSDs check box to display the related Relative Standard
Deviation values for each sample. You can also display additional columns to
show the sample acquisition date and time, reprocessed data flags, and QC
function status indicators — click Advanced to configure these options.
268 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Reports
When the QC Status column is displayed, you can click on any Failed cell or the
corresponding Sample ID to go to the relevant detailed data section of the Report
View QC tab. (When the QC Status column is not displayed, any failed QC
measures can be identified by a red Sample ID cell; click on the Sample ID to
access the related item on the QC Tab.) Note that if a QC measure specifies that
there be a valid calibration in the current workspace, and none exists, the QC
Status column will appear blank, rather than showing Passed or Failed. The QC
status will also be blank if QC checking is not enabled in the given method; or
when it is enabled, but no QC limits have been specified.
If you are running the Enhanced Security™ software, depending on your software
configuration, workflow change and traceability management functionality may
also be added, so that sample runs can be reviewed and approved (or rejected),
and e-signature details captured and displayed in the acquired data results tables.
See Enhanced Security Report View functions on page 275 for more information.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 269
Software Reference Guide
270 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Reports
Note: If you are reprocessing data with original conditions, the percent recovery value
does not appear.
Up to ten distinct colors are plotted at any one time; additional internal standards
are plotted using those same colors again, so that a color may appear more than
once in a single plot.
When you move your mouse over a point in the graph a screentip appears
displaying the following additional information for that point: row index and sample
ID numbers; analyte name and mass; % recovery; intensity; and units used for the
intensity.
Note: The legend of the Internal Standards graph can display only the number of internal
standards that can fit into the screen; where the number of standards exceeds this, the
legend simply indicates More…. This is a static indication, and not a clickable link leading
to additional information. To see more standards in the legend, expand the screen.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 271
Software Reference Guide
272 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Reports
QC tab
Applies to Quantitative Analyses only. This tab displays relevant Quality Control
data for samples run for which QC criteria has been defined in the selected
method. The data displayed includes the basic sample information, such as the
originating row index number and analyte sample information. Analytes that are
acting as Internal Standards are identified with an (IS) in the relevant analyte
header. When a sample falls outside of the specified QC criteria, the relevant data
is displayed in red text. QC data is displayed for only as long as the corresponding
sample entries appear on the Raw and Net Intensities, Concentrations, and
Unfactored Concentrations tabs.
The data on the QC tab is organized into the following categories: QC Standards;
QC Spikes; QC Duplicates; QC Dilutions; and Samples, Blanks, and Calibration
Standards:
• For QC Standards, which are grouped by standard number, the original
sample concentration is displayed, followed by the RSD and % recovery.
• QC Spikes display the % recovery of the spike. QC Duplicate Spikes are
also listed here, showing the % difference. For all QC Spikes and QC
Duplicate Spikes listed, this tab also displays the original row index of the
parent sample.
• QC Duplicates and QC Dilutions display the % difference of the duplicate
or dilution. For all QC Duplicates and QC Dilutions listed, this tab also
displays the original row index of the parent sample.
• The Samples, Blanks, and Calibration Standards section displays the
named information only where a sample fails the QC criteria specified —
Samples display sample concentration, IS % recovery, RSD, and SD
information; Blanks display measured intensity SD data; and Calibration
Standards display net intensity RSD data.
The categories are each displayed in their own collapsible tree format — click on
the + and - signs beside each header to expand and view or collapse and hide the
available data.
All QC types from the Method panel QC tab are considered in this section.
However, because the QC Calibration sub-tab deals with limits on the calibration
as a whole (such as slope, intercept, and correlation coefficient limits) rather than
on individual samples, no QC failure is flagged if an individual calibration standard
goes outside specified limits, and no data, sample names or analyte headers
appear in red if a calibration fails a QC check. When a calibration fails, the
calibration information is displayed on the Intensities, Concentrations, and
Unfactored Concentrations tabs, inserted with a Sample ID of Calibration Curve,
and expanded upon on the QC tab with the addition of Slope, Intercept, and
Correlation Coefficient values for each analyte. Also note that repeated samples
and diluted samples as a result of QC failure actions are not added to this tab as
QC Duplicates or QC Dilutions.
If a QC measure requires that there be a valid calibration in the current
workspace, and none exists, the QC Status column will appear blank, rather than
showing Passed or Failed, and the relevant cells on the QC tab will also appear
blank, as no data could be properly calculated.
You can also access the data on this tab from any of the other tabular report views
where a QC Status cell displays a Failed marker; simply click the Failed cell or
the corresponding Sample ID to go to the relevant row on the QC tab. Note that
QC failure messages are not displayed on this view — to view a failure message,
you must look at the relevant individual Current Sample report.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 273
Software Reference Guide
You can display additional columns to show the acquisition date and time, and
reprocessed data flags for relevant samples — click Advanced to configure this
option.
If you are running the Enhanced Security™ software, depending on your software
configuration, workflow change and traceability management functionality may
also be added, so that sample runs can be reviewed and approved (or rejected),
and e-signature and QC signoff details captured and displayed in the acquired
data results tables. See Enhanced Security Report View functions on page 275 for
more information.
274 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Reports
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 275
Software Reference Guide
Note: You do not need to have Microsoft Excel installed on the same computer
as your Syngistix ICP-MS software in order to perform a report view export. If
Excel is not installed, the export file will still be saved with all applicable data, and
with the Internal Standards graph drawn in .jpg form (rather than the usual
editable Excel graph object). You can then transfer the file to another computer, or
open the file with an Excel-compatible reader.
The default file format for the export function is .xlsx, but .xls is also generally
available. However, due to limitations in Microsoft Excel, if you are exporting more
than 256 columns of data, you can only export in .xlsx format.
Note: If you have a version of Microsoft Excel earlier than 2007 installed on your
computer, you will be allowed to save to .xlsx format, but you will not be able to
correctly open the resulting file in that version of Excel (the file will appear to
contain garbage information). In this case, you can choose to instead save to .xls
276 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Reports
format, which will open in any version of Excel, or you can open the resulting .xlsx
file in an Excel-compatible viewer or on another computer running Excel 2007.
Tip! (UCT instruments only) If you will be printing your exported reports to a
black and white printer, append the profile or mode you are using to each
analyte name in your method so that this information is included in your print-
outs (otherwise, mode is indicated by color only)
Tip! In some cases, when exporting a large amount of data from the Reporter
utility to Microsoft Excel, a message will appear indicating that the software is
not responding, and asking whether you wish to terminate the process or wait.
This occurs simply due to the time elapsed and is NOT an indication that the
software has crashed or frozen. Indicate that you would like to Wait for the
program to respond; the process will continue. The export may simply take
longer than expected.
• Clear Data: Click this button to clear all current data from the report view
tabs, except the Current Sample report. Note that you may wish to export
your data before clearing in order to preserve it for future examination.
Tip! In the Reporter panel, if you have a particularly large amount of data (for
example, data collected under many columns heads as a result of multiple
sample acquisition methods or calibration tables), it may take several seconds
to switch tab views, and several minutes to export that data to Microsoft Excel.
To combat this issue, clear your Reporter data frequently, either by clicking the
Clear Data button, or by restarting the Syngistix software.
A right-click menu is also available in the Report View tables to allow you to
designate a sort order for your analyte headers (choose the same order
designated in your method, or sort by analyte, mass, or mode), or to access the
Advanced Report View Display Options dialog box. Screentips provide additional
information throughout the Report Viewer; move your mouse over select controls
and data points to access these tips.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 277
Software Reference Guide
278 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Reports
Note: These options are available only if the Audit Trail and Challenge and
Electronic Signatures check boxes are selected in the ES Setup window Select
ES Features dialog box.
• Raw Intensity field: Select the number of decimal places to display in the
reporting tables for Raw Intensity values. Valid values = 0-4
• Net Intensity field: Select the number of decimal places to display in the
reporting tables for Net Intensity values. Valid values = 0-4
• Concentration field: Select the number of decimal places to display in the
reporting tables for Concentration values. Valid values = 0-4
• RSDs field: Select the number of decimal places to display in the
reporting tables for RSD (Relative Standard Deviation) values. Valid
values = 0-4
• SDs field: Select the number of decimal places to display in the reporting
tables for SD (Standard Deviation) values. Valid values = 0-4
• % Recoveries field: Select the number of decimal places to display in the
reporting tables for Percent Recovery values. This field also governs the
number of decimal places that will appear for Percent Difference values on
the QC tab. Valid values = 0-4
Note: If you are reprocessing data with original conditions, the percent
recovery value does not appear.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 279
Software Reference Guide
Calibration Panel
Use this panel to view graphic calibration curve data for the results listed on the
Report View panel. A calibration row appears after the last calibration standard
defined in the method is run.
On any table tab on the Report View panel, click a single calibration cell to display
that calibration only; select an entire Calibration Curves row to display the
complete set of calibrations available (a vertical scroll bar allows you to view all
the displayed calibrations). The single calibration view displays information in
regular whole numbers, while the multi-calibration view uses scientific shortened
notation to save space. As on other report graphs, S indicates a value in
shutdown, and NaN means “not a number” (no data available). Horizontal and
vertical sliders allow you to adjust the area of focus along each axis; click the
square arrows button in the bottom right corner of any calibration graph to return
to the original view. In multi-calibration view, click the small, square, gray button in
the upper right corner to expand this panel for better viewability.
280 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Reports
Note: If you previously used the NexION software, and you are reprocessing
data collected in the NexION software:
In the Reporter, the current standard's raw intensity is calculated and displayed; when
NexION data is reprocessed using original conditions and the next standard is
reprocessed, the software reads the calibration from the original dataset and, finding no
raw intensity value (as this measurement did not exist in the older software), reads the raw
intensity as zero. As a result, on the Reporter screen Calibration panel, the most recently
reprocessed standard displays a point with a non-zero intensity, and all earlier standards
are plotted with a zero calibration.
.
When reprocessing NexION data without original conditions, each point is simply
recalculated, and the Calibration panel is able to display points with correctly calculated
raw intensities.
The BEC and DL are calculated using the most recent Blank run up to
that point in time; if a subsequent Blank is run, existing calibration rows
are not updated; rather, a new calibration row is created, and the BEC
and DL are recalculated.
• Calibration Table: This section displays intensity and concentration
information for the specified analyte.
Net Intensity: Displays the net intensity value, blank corrected and
adjusted for internal standards.
Apparent Conc.: Displays the apparent concentration.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 281
Software Reference Guide
Report Options
Use the reporting options functions to create custom reports for your
determinations. You can define a report with the exact content you need and in a
style that best conveys the information content on the page. The following are the
basic components of a report options template:
• A report option template consists of a series of sections that are placed
consecutively on the page or in the file
• Each report section consists of a series of fields that are structured within
each section
• All the available determination data can be included in a report. You select
the sections and fields that you want to include, then you adjust the pre-
defined formats, including position on the page, field names and font style
and size, according to your individual requirements
282 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Reports
Report Appearance
To make report design fast and easy, the software identifies default location and
type specifications for each item that you include in a report template. However,
you may want to change these parameters to make the report fit your
requirements. You can customize a variety of options including the title page,
headers and footers, row and column sizing and spacing, font types and styles,
number formats, and page breaks.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 283
Software Reference Guide
284 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Reports
where:
The Standard Known Ratio is the user entry of the certified ratio for the
Standard.
• Ratio Normalized:
• Ratio Fraction:
• Mass Fraction: The Isotope Ratio mass fraction (wA) is the ratio of the
mass of isotope A to the total mass of all the isotopes:
mA
w A = -------------
-
i m
The mass (mA) of any isotope is the ratio fraction multiplied by its isotopic
mass. Note that the isotopic mass value for the calculation must be the
isotopic mass from the elemental library up to six decimal places.
Mean Values Section fields
• Reference Mass: The mass to which other masses of the same element
in the method are compared (ratioed).
• Analyte: The symbol for each analyte in the method.
• Mass: The mass of each isotope in the method.
• Measured Intensity Mean: The arithmetic mean of the replicate
measured intensity values.
• Blank Intensity: The arithmetic mean of the replicate measured intensity
values for the blank solution.
• Net Intensity Mean: The arithmetic mean of the replicate net intensity
values.
• Correction Factor Mean: The arithmetic mean of the replicate correction
factor values.
• Ratio Normalized Mean: The arithmetic mean of the replicate ratio
normalized values.
• Ratio Fraction Mean: The arithmetic mean of the replicate “ratio fraction”
values.
• Mode: (UCT instruments only) The operating mode.
• Gas: (UCT instruments only) The cell gas used.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 285
Software Reference Guide
Note: If you are using the default acquisition profile names, be aware that they
already include both the mode and cell gas as part of the name, so that adding a
combination of Mode / Gas / Profile fields may result in redundant information in
your report. However, if you have created custom profiles using a unique naming
or identification system that does not include both of these pieces of information,
you may wish to include a combination of these fields to ensure that all necessary
data is captured.
where:
n = the number of replicates for the sample
x = Ii or Ni or Ci or Ri
Ii = the instantiate for analyte i
Ci = the ratio correction factor for analyte i
Ni = the net intensity for analyte i (Ii - IO), blank intensity subtracted
IO = the intensity of the blank
Ri = the (calculated) corrected (isotope) ratio
286 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Reports
Summary Reports
Summary by Analyte and Summary by Sample reports capture the net value, SD,
and %RSD for the intensities and concentrations of analytes measured during
batch acquisition. The results can be sorted by analyte (Summary by Analyte) or
by sample (Summary by Sample).
There are two ways to generate a summary report: in real time after a batch
analysis or during the reprocessing of the data from a batch analysis.
Note: Summary reports are available for quantitative analysis methods only.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 287
Software Reference Guide
288 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Chapter 12
Graphic Data
This section describes the graphic viewing and manipulation tools.
Topics in this section include:
• Graphic Displays on page 289
• Graphics Display Toolbar on page 289
• Realtime Panel on page 291
• Realtime Graphics Options Dialog Boxes on page 292
• Interactive Panel on page 295
• Interactive Graphics Options Dialog Boxes on page 298
• Composite Signal and Spectral Options on page 301
• Cell Parameter Optimize Options on page 302
• Calibration View Panel on page 303
Graphic Displays
The software provides extensive displays of signal and spectral data in realtime or
post-run. There are three main panels through which you work with graphic data:
• Realtime Panel
• Interactive Panel
• Calibration View Panel
While each of these panels is designed to support different requirements, the
functions available in each are similar, and all are supported by a set of common
controls.
Linear/Logarithmic Use to select a linear or logarithmic value
Plot Y‐axis plot
Grid Off/On Use to select no grid (with tick marks) or a full
grid plot background
Curve/Bar/Line Plot Use to select a curve (spectral), bar, or line
plot
289
Software Reference Guide
Zoom Use to enlarge an area in a plot. Each time
you zoom, a number is assigned to the view,
with 0 the full display, 1 the first zoom, 2 the
second, and so on
To zoom in a plot, drag the cursor to draw a
box around the region you want to enlarge.
When you release the mouse button, the
graph is redrawn with the area inside the box
displayed at zoom scale
To switch between zoom levels, click the
Zoom button, and then click the number
associated with the zoom level you want to
use. To reset the zoom series, set the display
to zoom level 0, then redraw a different area
with the cursor. All remaining zoom levels in
the original series are deleted.
Print One/All Plots Use to control the output of plots to a printer
attached to the computer. To print a single
plot in a one‐plot display, click the Print One
button. To print a single plot in a multiple‐plot
display, click the plot you want to print, then
click the Print One button. To print all the
plots in a multiple‐plot display, click the Print
All button
Plot Options Use to define the colors used for plot axes,
labels, and tick marks, and to indicate
whether to display titles, labels, and sliders
Customize Fonts for Use to manage type parameters for plot
Titles and Labels labels and titles. Click and select Titles or
Labels from the list; the Font Options dialog
box appears. Select the preferred font
parameters and then click OK. The
parameters selected are applied to all plots
displayed
Zoom to Fit All Use to scale a plot. The software
automatically adjusts the y‐axis intensity to
maximize the display so that the largest peak
is on‐scale and shown at its maximum
intensity
Customize Line Use to select and apply a line width to the
Width displayed plot. Select from 1 pt, 2 pts, or 3
pts wide
290 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Graphic Data
Realtime Panel
Use the Realtime panel for realtime viewing and interpretation of signal,
spectral, and numeric displays of data during an acquisition or
optimization. Here you can adjust the axis range and display parameters,
manage the printing of graphics data, specify the colors used, and control
the font and type size for title and label text.
The Realtime screen is accessed via the Realtime icon on the ribbon.
When accessed in this way, it consists of the Realtime panel only. The
Realtime panel, however, is also displayed in several other multi-panel
views (including the Sample screen, Conditions screen, Instrument
Diagnostics screen, and the SmartTune screen). This allows you to
monitor the results of various processes without having to switch between
screens.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 291
Software Reference Guide
292 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Graphic Data
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 293
Software Reference Guide
294 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Graphic Data
Interactive Panel
Use this panel for post-acquisition viewing and interpretation of signal and
spectral displays. Here you can adjust the axis range and display parameters,
manage the printing of graphics data, control the colors used on the display, and
to control the font and type size for titles and label text. The Interactive panel is an
integral part of calibrating and optimizing your instrument, and is also useful in
post-acquisition analysis of analytical data.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 295
Software Reference Guide
analyte. The red section of each bar indicates the mass contribution of
the selected analyte.
Composite Signal: Overlays, adds, subtracts or sets up a ratio for two
time-varying signals. It is also useful for blank-subtracted data. This
option is available for individual elements/analytes only. The graph
displayed has an x-axis listing time in seconds and a y-axis listing
signal intensity in cps or total ion counts.
Composite Spectral: Overlays, adds, subtracts or sets up a ratio for
two mass spectra. It is also useful for blank-subtracted data. The graph
displayed has an x-axis listing mass range in amu and a y-axis listing
signal intensity in cps total ion counts.
Optimization: Displays the graph resulting from an optimization scan
for one of the digital-to-analog converters controlling the instrument
hardware. This option is used in conjunction with the Conditions panel
during hardware optimization. The graph displayed has an x-axis
listing the range of DAC values to be scanned and a y-axis listing
intensity in cps.
MassCal: Displays the graphs resulting from a mass calibration or
resolution adjustment performed during instrument optimization. This
option is used in conjunction with the Mass Calibration panel. Each
graph displayed has an x-axis listing the mass range in amu and a y-
axis listing intensity in cps. The individual graphs include a title bar
displaying the atomic symbol and the isotope mass for the analyte
results being plotted in that graph. When results are plotted, a line is
also automatically drawn to indicate the intensity at 10% of maximum
peak intensity (the point at which peak width is measured).
Dual Detector: Use with the Interactive panel as part of the dual
detector calibration procedure. Dual detector calibration is used to
adjust the response behavior of the pulse and analog stages of the
dual detector to provide optimum data integrity when signal intensity
varies widely during a run employing both pulse and analog signal
detection. By selecting Analytes, the graph can be displayed for up to
five analytes used in the calibration procedure. Each graph has an x-
axis displaying pulse intensity and a y-axis displaying analog intensity.
Detector Response Function graph shows the relationship of the
analog to pulse signals across the mass range. Each point represents
a mass which has a valid dual detector calibration. The straight lines
between points indicate interpolated conversion factors. For an
accurate quantitation you must have a measured conversion factor for
each isotope of interest. Using the interpolated conversion factor may
result in analytical inaccuracies. The detector response functions
1-
----
2
should have the shape of a curve. A flattening of the detector
x
response function is an indication that the detector is near the end of
its useful life.
QID function: Use with the Interactive panel as part of the QID™
calibration procedure. QID component scanning helps to adjust the
detection systems mass bias to provide optimum signal intensity
across the full mass range. You can configure this graph to display the
information for up to five analytes used during the scanning procedure.
The x-axis displays QID component voltage, and the y-axis displays
296 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Graphic Data
Note: You can reprocess cell parameter information only if the start, end, and
step values for the parameter and profile used during acquisition are identical for
both the Blank and the Sample. You cannot reprocess datasets created in a
previous (prior to v2.0) version of the software.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 297
Software Reference Guide
298 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Graphic Data
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 299
Software Reference Guide
to select analog mode. Click Dual to acquire input from both sources.
Dual mode is active only after the detector has been calibrated.
• Processing Options Group:
cps/counts: Set the processing function to display either total ion
counts or counts per second.
Spectral Peak: Use to specify how the data from the individual spectral
peaks is processed before it is displayed. There are three options:
average, sum, and maximum. Click Average to average the values of
the MCA Channels from each peak. Click Sum to add the values of all
the MCA Channels to create a total ion count for each peak. Click
Maximum to select the MCA channel with the maximum value from
each spectral peak. No processing can be performed on the spectral
profile when using Spectral Interactive Graphics mode.
Apply Smoothing Factor: This check box is available only in the
Interactive panel. Use to start spectral smoothing in order to reduce the
effects of noise in high sensitivity measurements. You must have
multiple readings per replicate to use this option. The smoothing
function is a Savitzky-Golay moving average of the readings along the
spectral profile. When the check box is selected, the list beside it is
made available. Click to display the smoothing factor options (the
number of moving average points), based on the number of readings
per replicate used for data acquisition.
Signal Profile: Use to specify how the signal profile is displayed before
further calculations are performed. There are three options: average,
sum, and maximum. Click Average to average the readings from a
profile. Click Sum to add the readings to create a total count for each
profile. Click Maximum to select the maximum reading from each
profile.
300 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Graphic Data
Note: You cannot overlay or reprocess datasets created in a previous version of the
Syngistix for ICP-MS software (prior to version 2.0) using these functions.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 301
Software Reference Guide
Note: You cannot overlay or reprocess datasets created in a previous version of the
Syngistix for ICP-MS software (prior to version 2.0) using these functions.
For each analyte, the intensity of both solutions at each time interval is plotted as
a function of cell parameter value. The composite line graph represents either the
estimated detection limit calculated at each time interval, or the estimated
background equivalent concentration calculated at each time interval.
• Detection limit option: The estimated detection limit is calculated at each
time interval according to the equation:
302 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Graphic Data
where:
I blk = the measured intensity for a blank solution, for example, deionized
water or matrix blank.
I std = the measured intensity for a standard solution, for example, 1 ppb
Iron.
Conc std = the concentration of the standard solution or spiked matrix.
Note: The unit of the detection limit is the same as that of the standard solution
concentration.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 303
Software Reference Guide
304 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Appendix A
305
Software Reference Guide
306 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
component code/message condition
Gas Fault Condition RB 1 Nebulizer gas flow fault: Check gas
This readback to the ICM connection
indicates a mass/pressure flow
controller fault condition.
2 Plasma gas flow fault: Check gas
connection
3 Auxiliary gas flow fault: Check gas
connection
4 Makeup gas flow fault: Check gas
connection
5 Oxygen gas flow fault: Check gas
connection
6 Channel A gas flow fault: Check cell gas
connection
7 Channel B gas flow fault: Check cell gas
connection
8 Getter temperature outside prescribed
range: Check the air intake filter, fan
operation, ambient temperature, and the
exhaust and ventilation systems
9 DRC vent strap position fault: The vent
may not be fully closed. Check LEDs and
the DRC vent mechanism
10 Nebulizer gas flow back pressure fault:
Check gas connection
11 Channel C gas flow fault: Check cell gas
connection
12 Cell gas manifold temperature outside
prescribed range
13 Argon gas manifold temperature outside
prescribed range
ICP Fault Condition RB 1 ICP plate voltage outside prescribed
This readback to the ICM range
indicates a ICP controller fault
condition.
2 ICP plate current outside prescribed
range
3 ICP grid current outside prescribed range
4 ICP filament voltage outside prescribed
range
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 307
Software Reference Guide
308 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
component code/message condition
16 DRC RF voltage outside prescribed range
17 DRC rod A voltage outside prescribed
range
18 DRC rod B voltage outside prescribed
range
19 DRC RF voltage (DAC) outside prescribed
range
20 DRC DC voltage (DAC) outside prescribed
range
21 DRC rod offset (DAC) fault
22 QPS current outside prescribed range
23 QPS RF detector compensator fault
24 QPS RF detector voltage outside
prescribed range
Monitor Fault Condition 1 Scanning relay failure
RB
This readback to the ICM
indicates a monitor fault
condition.
2 “Ready to scan” relay failure
3 Chiller relay failure (in ACDP)
6 Interface solenoid valve relay failed
7 Interface water solenoid valve failed
9 Interface open solenoid valve failed
10 Oxygen solenoid valve failed
11 DRC cell close solenoid valve failed
12 Interface close solenoid valve failed
13 Argon gas solenoid failed
14 DRC cell open solenoid valve failed
15 Exhaust fan control solenoid failed
16 Roughing pump purge valve failed
24 Purge valve failed
27 Opto‐coupler OUT_D2 failed (turbo
enable)
Motor Task State RB 0 Stopped Stopped
This readback to the ICM
indicates the state of the XYZ
motors.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 309
Software Reference Guide
310 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
component code/message condition
4 +20V DC power supply fault
5 ‐20V DC power supply fault
6 +15V DC power supply fault
7 ‐15V DC power supply fault
8 +5V DC power supply fault
9 ICM +3.3V regulator fault
10 ICM +2.5V regulator fault
11 ICM +1.2V regulator fault
13 QPS board +17V regulator fault
14 QPS board ‐17V regulator fault
15 QPS board +12V regulator fault
16 QPS board ‐12V regulator fault
17 QPS board +5V regulator fault
18 QPS board ‐5V regulator fault
19 QPS board +15V regulator fault
20 QPS board ‐15V regulator fault
21 QPS board +1.8V regulator fault
22 Lens board +550V DC/DC fault
23 Lens board ‐550V DC/DC fault
24 Lens board +275V DC/DC fault
25 Lens board ‐275V DC/DC fault
26 Lens board +5V regulator fault
27 Lens board ‐5V regulator fault
28 Lens board +15V regulator fault
29 Lens board ‐15V regulator fault
30 Vacuum gauge filament voltage outside
prescribed range
Pump Fault Condition RB 1 Turbo rotor speed outside prescribed
This readback to the ICM range
indicates a pump fault
condition.
2 Turbo motor temperature outside
prescribed range
3 Turbo controller temperature outside
prescribed range
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 311
Software Reference Guide
312 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
component code/message condition
16 Torch not in place
SS-RFG Error/Status RB 0 RFG and ICM not communicating
This readback to the ICM properly; check connections
indicates the status of the Solid
State RF generator.
(2000 series instruments)
1 RFG power transistor 1 is damaged
2 RFG power transistor 2 is damaged
3 RFG power transistor 1 is running over
the prescribed current
4 RFG power transistor 2 is running over
the prescribed current
5 The RFG power transistor is experiencing
an excessive current imbalance; the
current of transistor 1 is much greater
than the current of transistor 2
6 The RFG power transistor is experiencing
an excessive current imbalance; the
current of transistor 2 is much greater
than the current of transistor 1
7 RFG power transistor 1 cannot be biased
successfully
8 RFG power transistor 2 cannot be biased
successfully
9 The DC power being supplied to the
power transistor is incorrect; ensure that
the RFG breaker is on, and that the fuse
does not need replacing. Also check for
an issue with the DC Power Supply for the
power transistors
10 The value for the low voltage power
supply is incorrect
11 The RFG has overheated. Ensure that the
cooling system is operating properly. If
this error prevents the plasma from
igniting, manually run the coolant for 15
minutes before attempting to ignite the
plasma
12 The RFG has been overcooled
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 313
Software Reference Guide
314 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
component code/message condition
Roughing Pump Last Error DC bus overvoltage DC bus overvoltage
This readback from the
roughing pump controller to
the ICM indicates an error.
Overcurrent Overcurrent at drive output
IGBT overcurrent IGBT transistor overcurrent
Motor phase lost Loss of motor phase
Overspeed Overspeed
Static position Measured position does not vary; the
encoder is incorrectly connected or not
supplied with power or the shaft is not
turning
A, B, A\, B\ transposed A, B, A\, B\ signals wrong way round
u, v, w transposed u, v, and w commutation signals wrong
way round
U missing Some signals are present, but U signal is
missing
V missing Some signals are present, but V signal is
missing
W missing Some signals are present, but W signal is
missing
Pole pairs wrong The number of pairs of poles set is
incorrect; the revolutions measured
mechanically with A, B and electrically
with U, V, W are inconsistent, given the
number of pairs of poles entered
Autotune fault Autotune fault
Brake res. overload Braking resistor overload I x t
Motor overload Motor overload I x t
IGBT hot IGBT overheating detected by internal
sensor
Brake res. hot Internal braking resistor overheating
detected by thermal sensor
Thermal fault Motor thermal sensor has tripped
+24V PS overload Overload on the +24V power supply or
digital output
ADI1 current lost Loss of the current reference on analog
input ADI1
ADI2 current lost Loss of the current reference on analog
input ADI2
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 315
Software Reference Guide
316 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
component code/message condition
8 Drive output is currently limited
9 Regenerating
10 Braking IGBT active
11 Braking resistor alarm
12 Direction commanded
13 Direction running
14 Mains loss
Turbo Pump Error Code 0 Normal
This readback from the turbo
controller to the ICM indicates
an error code from the turbo
controller.
1 Turbo pump overspeed: Nominal speed
of the pump exceeded by over 10%. The
turbo pump may need to be serviced.
Contact your local service representative.
2 Turbo pump timeout: Maximum time for
passing through critical frequencies has
been exceeded; pump switched off. The
turbo pump may need to be serviced.
Contact your local service representative.
3 Turbo pump bearing over temperature:
Maximum bearing temperature
exceeded; pump switched off. The turbo
pump may need to be serviced. Contact
your local service representative.
4 Turbo pump short circuit: Short circuit in
the pump motor or the connecting cable;
pump switched off. The turbo pump may
need to be serviced. Contact your local
service representative.
5 Turbo pump over temperature:
Maximum temperature for the converter
has been exceeded; pump switched off.
Check the air intake filter, fan operation,
ambient temperature, and the exhaust
and ventilation systems. The turbo pump
may need to be serviced. Contact your
local service representative.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 317
Software Reference Guide
318 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
component code/message condition
106 Turbo pump too fast: Upper critical limit
frequency exceeded by overload; pump
switched off. The turbo pump may need
to be serviced. Contact your local service
representative.
116 Turbo pump too fast: Maximum overload
time exceeded; pump switched off. The
turbo pump may need to be serviced.
Contact your local service representative.
117 Turbo pump motor current low: No
motor current, or motor current too low;
pump switched off. The turbo pump may
need to be serviced. Contact your local
service representative.
118 Turbo pump motor connection error:
Error in the motor connection cable;
pump switched off.
125 Turbo pump current too high: Maximum
permissible permanent current
exceeded. The turbo pump may need to
be serviced. Contact your local service
representative.
126 Turbo pump bearing over temperature:
Error at the bearing temperature sensor.
Resistance out of range; pump switched
off. The turbo pump may need to be
serviced. Contact your local service
representative.
127 Turbo pump bearing over temperature:
Error at the bearing temperature sensor.
Resistance not in the plausible range;
pump switched off. The turbo pump may
need to be serviced. Contact your local
service representative.
128 Turbo pump motor temperature error:
Error at the motor temperature sensor.
Resistance not in the plausible range;
pump switched off. The turbo pump may
need to be serviced. Contact your local
service representative.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 319
Software Reference Guide
320 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
component code/message condition
213 Turbo pump power supply voltage too
high: Pump switched off. The turbo pump
may need to be serviced. Contact your
local service representative.
214 Turbo pump power supply voltage too
low: Pump switched off. The turbo pump
may need to be serviced. Contact your
local service representative.
216 Turbo pump memory error: Error in
external memory; pump switched off.
Contact your local service representative.
217 Turbo pump ID resistor error: Incorrect or
missing pump identification resistor;
pump switched off. Contact your local
service representative.
222 Turbo pump hardware test error: Pump
switched off. Contact your local service
representative.
223‐226 Turbo pump logical unit error: Error in
the programmed logical unit; pump
switched off. Contact your local service
representative.
227 Turbo pump model error: No set of
parameters defined for the recognized
pump model; pump switched off. Contact
your local service representative.
Turbo Pump Model RB 0 Turbo pump models TW 220/150, TW
This readback from the turbo 220/150/15
controller to the ICM indicates
the type of turbo pump.
1 Turbo pump models TW 400/300/25S
2 Turbo pump model TW 250S
3 Turbo pump model TW 70 H
4 Turbo pump models TW 300/TW 300 H
Turbo Pump Status RB 0 Normal
This readback to the ICM Pump status: In case of error, t0w has the
indicates the status of the value of 0 (not ready to be switched on)
turbo controller.
1 Ready to turn on
2 Speed increasing
3 Speed decreasing
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 321
Software Reference Guide
322 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
component code/message condition
XYZ Motor Control CYC_TST_ON Cycle test on
This software command to the
ICM controls the state of the
XYZ mechanism. When in
Service Mode, click the button
to access the XYZ torch control
dialog box
CYC_TST_OFF Cycle test off
Initialize Initialize
Stop Stop
Relax Relax
Lock X Lock X position
Lock Y Lock Y position
Lock Z Lock Z position
LOCK_XYZ Lock XYZ
Calibrate Calibrate
FACTORY_CAL Reset to factory calibration specifications
SET_IGN_POS Set plasma ignition position
CORR_CAL Correct calibration
SRV_POS 0x50 Send to service position (0x50)
IF_POS Send to interface position
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 323
Software Reference Guide
sub
system system component description units
Analyzer Aux I/O Aux Device Enable RB This readback from the Aux I/O device NA
to the ICM indicates whether scanning
is enabled or disabled.
Analyzer Aux I/O Ready To Scan This ICM command instructs a relay to NA
enter the Ready to Scan state.
Analyzer Aux I/O Scan Polarity Enable This readback from the Aux I/O device NA
RB to the ICM indicates whether the
external contact closure is normally
open or closed.
Analyzer Aux I/O Scanning This ICM command initiates scanning in NA
an external device.
Analyzer Control +3.3V DC RB This readback from the ICM indicates Volts
the 3.3V local supply reading.
Analyzer Control Cone Access LED This ICM command turns the yellow NA
Yellow Cone Access LED on the front panel On
or Off.
Analyzer Control Cone Access LED This ICM command turns the green NA
Green Cone Access LED on the front panel On
or Off.
Analyzer Control ICM Board Temp. RB This is a readback to the ICM from the C
temperature sensor on the board.
Analyzer Control ICM Interlock LED This ICM command turns all the NA
Test interlock LEDs on the ICM faceplate On
or Off.
Analyzer Control System LED Fault This ICM command turns the red NA
System LED on the front panel On or
Off.
Analyzer Control System LED Ready This ICM command turns the green NA
System LED on the front panel On or
Off.
Analyzer Detector Analog Current RB This readback from the high voltage µAmps
board to the ICM indicates the detector
analog (negative) current.
Analyzer Detector Analog Voltage This is a command from the ICM to the Volts
high voltage board. Adjust the analog
(negative) voltage as required.
324 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
system sub component description units
system
Analyzer Detector Analog Voltage RB This readback from the high voltage Volts
board to the ICM indicates the detector
analog (negative) voltage.
Analyzer Detector Discriminator This ICM command to the high voltage mVolts
Threshold board sets the discriminator threshold
value. Adjust the value as required.
Analyzer Detector Gate Voltage RB This readback from the high voltage Volts
board to the ICM indicates the detector
gate voltage.
Analyzer Detector Pulse Current RB This readback from the high voltage µAmps
board to the ICM indicates the detector
pulse (positive) current.
Analyzer Detector Pulse Voltage This ICM command to the high voltage Volts
board sets the pulse (positive) voltage.
Adjust the voltage as required.
Analyzer Detector Pulse Voltage RB This readback from the high voltage Volts
board to the ICM indicates the detector
pulse (positive) voltage.
Analyzer DRC AFT This ICM command to the lens board Volts
sets the DRC AFT voltage. Adjust the
voltage as required.
Analyzer DRC AFT RB This readback from the lens board to Volts
the ICM indicates the DRC AFT voltage.
Analyzer DRC Cell Entrance Voltage This ICM command to the lens board Volts
sets the cell entrance voltage. Adjust
the voltage as required.
Analyzer DRC Cell Entrance Voltage This readback from the lens board to Volts
RB the ICM indicates the cell entrance
voltage.
Analyzer DRC Cell Exit Voltage This ICM command to the lens board Volts
sets the cell exit voltage. Adjust the
voltage as required.
Cell Exit Voltage RB This readback from the lens board to Volts
the ICM indicates the cell exit voltage.
Analyzer DRC Channel A Gas Flow This ICM command to the channel A mL/mi
mass flow controller sets the gas flow n
rate. Adjust the flow rate as required.
Analyzer DRC Channel A Gas Flow This readback from the channel A mass mL/mi
RB flow controller to the ICM indicates the n
gas flow rate.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 325
Software Reference Guide
326 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
system sub component description units
system
Analyzer DRC DRC Power Supply This ICM command toggles the DRC NA
power supply between On and Off.
Analyzer DRC DRC Status RB This readback to the ICM indicates the NA
state of the DRC cell.
Analyzer DRC DRC Temp. Status RB This readback from the DRC power NA
supply to the ICM indicates whether
the temperature alarm is OK or has
been triggered. (300/350 series
instruments only).
Analyzer DRC DRC Venting Time This software command to the ICM s
indicates the delay time for switching
from DRC to STD mode. Adjust this
value as required.
Analyzer DRC Getter Heater This ICM command turns the heater on NA
the getter On or Off.
Analyzer DRC Getter Heater This readback to the ICM indicates the s
Running Time RB conditioning time for getter
regeneration.
Analyzer DRC Getter Regeneration This readback to the ICM indicates the NA
RB status of the getter regeneration.
Analyzer DRC Getter Temp. This ICM command sets the getter C
temperature.
Analyzer DRC Getter Temp. RB This readback from the getter C
temperature sensor to the ICM
indicates the getter temperature.
Analyzer DRC Heatsink Temp. RB This readback from the DRC power C
supply to the ICM indicates the
heatsink temperature.
Analyzer DRC Mode This parameter indicates the current NA
operating mode of the software
(Standard, KED, or DRC mode). Select a
new mode as required.
Analyzer DRC RF Detector RB This readback from the DRC power Volts
supply to the ICM indicates the RF
voltage.
Analyzer DRC RF Voltage This ICM command to the DRC power Volts
supply sets a value for the RF voltage.
Analyzer DRC RF Voltage RB This readback from the DRC power Volts
supply to the ICM indicates the RF
voltage setpoint.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 327
Software Reference Guide
328 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
system sub component description units
system
Analyzer Mass Filter QPS +15V DC RB This readback from the QPS board to Volts
the ICM indicates the +15V local supply
reading.
Analyzer Mass Filter QPS +17V DC RB This readback from the QPS board to Volts
the ICM indicates the +17V local supply
reading.
Analyzer Mass Filter QPS +5V DC RB This readback from the QPS board to Volts
the ICM indicates the +5V local supply
reading.
Analyzer Mass Filter QPS ‐12V DC RB This readback from the QPS board to Volts
the ICM indicates the ‐12V local supply
reading.
Analyzer Mass Filter QPS ‐15V DC RB This readback from the QPS board to Volts
the ICM indicates the ‐15V local supply
reading.
Analyzer Mass Filter QPS ‐17V DC RB This readback from the QPS board to Volts
the ICM indicates the ‐17V local supply
reading.
Analyzer Mass Filter QPS ‐5V DC RB This readback from the QPS board to Volts
the ICM indicates the ‐5V local supply
reading.
Analyzer Mass Filter QPS Amplifier RB This readback from the QPS board to Amps
the ICM indicates the QPS amplifier
current.
Analyzer Mass Filter QPS DC Quad A RB This readback from the QPS board to Volts
the ICM indicates the DC mass voltage
on quad A of the analyzer quadrupole.
Analyzer Mass Filter QPS DC Quad B RB This readback from the QPS board to Volts
the ICM indicates the DC mass voltage
on quad B of the analyzer quadrupole.
Analyzer Mass Filter QPS Dithering This ICM command sets the dithering mAMU
Amplitude setpoint sent to the QPS board.
Analyzer Mass Filter QPS Dithering Switch This ICM command instructs the QPS NA
board to turn dithering On or Off.
Analyzer Mass Filter QPS Heatsink Temp. This readback from the QPS RF NA
Status heatsink temperature sensor to the
ICM indicates whether the
temperature is within normal range.
Analyzer Mass Filter QPS HVDC Quad A RB This readback from the QPS board to Volts
the ICM indicates the high voltage DC
on quad A of the analyzer quadrupole.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 329
Software Reference Guide
330 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
system sub component description units
system
Analyzer Optics Deflector Entrance This ICM command to the lens board Volts
Lens sets the deflector entry lens voltage.
Adjust the voltage as required.
Analyzer Optics Deflector Entrance This readback from the lens board to Volts
Lens RB the ICM indicates the deflector entry
lens voltage.
Analyzer Optics Deflector Exit Lens This ICM command to the lens board Volts
sets the deflector exit lens voltage.
Adjust the voltage as required.
Analyzer Optics Deflector Exit Lens RB This readback from the lens board to Volts
the ICM indicates the deflector exit
lens voltage.
Analyzer Optics Deflector Repellor This ICM command to the lens board Volts
sets the deflector repellor voltage.
Adjust the voltage as required.
Analyzer Optics Deflector Repellor RB This readback from the lens board to Volts
the ICM indicates the deflector repellor
voltage.
Analyzer Optics DPA Voltage RB This readback from the lens board to Volts
the ICM indicates the differential
pumping aperture voltage (300/350Q
instruments only).
Analyzer Optics Lens +15V DC RB This readback from the lens board to Volts
the ICM indicates the +15V local lens
board power supply reading.
Analyzer Optics Lens +5V DC RB This readback from the lens board to Volts
the ICM indicates the +5V local lens
board power supply reading.
Analyzer Optics Lens ‐15V DC RB This readback from the lens board to Volts
the ICM indicates the ‐15V local lens
board power supply reading.
Analyzer Optics Lens ‐5V DC RB This readback from the lens board to Volts
the ICM indicates the ‐5V local lens
board power supply reading.
Analyzer Optics Lens Board Temp. RB This readback from the lens board to C
the ICM indicates the lens board
temperature.
Analyzer Optics Lens PS +275V RB This readback from the lens board to Volts
the ICM indicates the +275V local lens
board power supply reading.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 331
Software Reference Guide
332 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
system sub component description units
system
Analyzer Vacuum Roughing Pump Ctrl This readback from the roughing pump C
Temp. RB controller to the ICM indicates the
controller temperature.
Analyzer Vacuum Roughing Pump This readback from the roughing pump Amps
Current RB controller to the ICM indicates the
motor current.
Analyzer Vacuum Roughing Pump Duty This readback from the roughing pump %
Cycle RB controller to the ICM indicates the
pump duty cycle as a percentage,
where 100% indicates full power.
Analyzer Vacuum Roughing Pump This readback from the roughing pump hrs &
Hrs.Min RB controller to the ICM indicates how min
long the pump has been running, (hh.mm)
measured in hours and minutes.
Analyzer Vacuum Roughing Pump Last This readback from the roughing pump NA
Error controller to the ICM indicates the last
error registered.
Analyzer Vacuum Roughing Pump This readback from the roughing pump Watts
Power RB controller to the ICM indicates the
motor power.
Analyzer Vacuum Roughing Pump This ICM command instructs the NA
Purge Valve roughing pump purge valve to Open or
Close.
Analyzer Vacuum Roughing Pump This ICM command to the roughing Hz
Speed pump controller sets the pump speed.
Analyzer Vacuum Roughing Pump This readback from the roughing pump RPM
Speed RB controller to the ICM indicates the
pump speed.
Analyzer Vacuum Roughing Pump This readback from the roughing pump NA
Status RB controller to the ICM indicates the
pump status.
Analyzer Vacuum Roughing Pump This readback from the roughing pump Volts
Voltage RB controller to the ICM indicates the
motor voltage.
Analyzer Vacuum Roughing Pump This readback from the roughing pump Years &
Yrs.Days RB controller to the ICM indicates how days
long the pump has been running, (Y.ddd)
measured in years and days.
Analyzer Vacuum Turbo Backing This ICM command turns the turbo NA
Isolation Valve backing isolation valve On or Off.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 333
Software Reference Guide
334 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
system sub component description units
system
Analyzer Vacuum Turbo Pump Speed SP This readback from the turbo controller Hz
RB to the ICM indicates the setpoint for
the turbo pump rotor speed.
Analyzer Vacuum Turbo Pump Start This readback from the turbo controller Starts
Counter to the ICM indicates the number of
starts of the turbo pump.
Analyzer Vacuum Turbo Pump Status This readback to the ICM indicates the NA
RB status of the turbo controller
Analyzer Vacuum Vacuum Chamber This ICM command instructs the NA
Vent Valve vacuum purge gas valve to Open or
Close.
Analyzer Vacuum Vacuum Gauge This ICM command turns the vacuum NA
gauge On or Off.
Analyzer Vacuum Vacuum Gauge This readback from the vacuum gauge Volts
Filament RB to the ICM indicates the filament
voltage.
Analyzer Vacuum Vacuum Gauge RB This readback from the vacuum gauge NA
to the ICM indicates whether the
vacuum gauge is OK or in a fault
condition.
Analyzer Vacuum Motorized Manifold This readback indicates the position or NA
State RB state of the motorized vacuum
manifold valve. (This will be blank if a
pneumatic vacuum manifold is
installed.)
Analyzer Vacuum Vacuum Pressure Full This readback from the vacuum gauge Torr
RB to the ICM indicates vacuum pressure
with a resolution of up to 1x10‐7 Torr.
Analyzer Vacuum Vacuum Pressure Low This readback from the vacuum gauge Torr
RB to the ICM indicates vacuum pressure
across a low range with high sensitivity,
with a resolution of up to 1x10‐8 Torr.
Analyzer Vacuum Vacuum Pressure This readback from the vacuum gauge NA
State RB to the ICM indicates whether the
vacuum pressure is OK or in a fault
state.
Analyzer Vacuum Vacuum Start Pressed This input from the system to the ICM NA
indicates whether the Vacuum Start
button is being pressed.
Analyzer Vacuum Vacuum State RB This readback to the ICM indicates the NA
state of the vacuum.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 335
Software Reference Guide
336 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
system sub component description units
system
Analyzer XYZ Torch Position Y This ICM command sets the Y mm
horizontal torch position in millimeters.
Adjust the position as required.
Analyzer XYZ Torch Position Y RB This readback to the ICM indicates the mm
Y vertical torch position in millimeters.
Analyzer XYZ Torch Position Z This ICM command sets the Z depth mm
torch position in millimeters.
Analyzer XYZ Torch Position Z RB This readback to the ICM indicates the mm
Z depth torch position in millimeters.
Analyzer XYZ Motor This readback indicates whether or not NA
Communication the XYZ motor system is
Status communicating correctly. (2000 series
instruments only)
Analyzer XYZ XYZ Motor Control This software command to the ICM NA
controls the state of the XYZ
mechanism. When in Service Mode,
click the button to access the XYZ torch
control dialog box.
Analyzer XYZ Motor Status X This readback indicates the status of NA
the X horizontal torch position motor.
(2000 series instruments only)
Analyzer XYZ Motor Status Y This readback indicates the status of NA
the Y vertical torch position motor.
(2000 series instruments only)
Analyzer XYZ Motor Status Z This readback indicates the status of NA
the Z depth torch position motor.
(2000 series instruments only)
Environment Cooling Coolant Flow State RB This readback to the ICM indicates the NA
state of the coolant flow.
Environment Cooling Exhaust Fan This ICM command turns the external NA
exhaust fan relay On or Off.
Environment Cooling Interface Water Flow This ICM command turns the interface NA
water flow On or Off.
Environment Cooling Recirculator This ICM command turns the NA
recirculator contact closure On or Off.
(300/350 series instruments only).
Environment Cooling Cooling Exhaust Flow This readback from the exhaust sensor F/M
RB interlock to the ICM indicates the
exhaust flow rate.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 337
Software Reference Guide
338 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
system sub component description units
system
Plasma Gas Auxiliary Gas Flow RB This readback from the auxiliary gas L/min
controller to the ICM indicates the
auxiliary gas flow rate.
Plasma Gas Auxiliary Gas Status This readback to the ICM indicates NA
whether the auxiliary gas is On or Off.
Plasma Gas Makeup Gas Flow This ICM command to the makeup gas L/min
mass flow controller sets the makeup
gas flow rate. Adjust the flow rate as
required (300/350 instruments only).
Plasma Gas Makeup Gas Flow RB This readback from the makeup gas L/min
mass flow controller to the ICM
indicates the makeup gas flow rate
(300/350 instruments only).
Plasma Gas Nebulizer Gas Flow This ICM command to the nebulizer gas L/min
controller sets the nebulizer gas flow
rate. Adjust the flow rate as required.
Plasma Gas Nebulizer Gas Flow This readback from the nebulizer gas L/min
RB mass flow controller to the ICM
indicates the nebulizer gas flow rate.
Plasma Gas Nebulizer Gas Status This readback to the ICM indicates NA
whether the nebulizer gas is On or Off.
Plasma Gas AMS Gas Flow This ICM command sets the AMS gas L/min
flow rate. Adjust the flow rate as
desired (2000 series instruments only).
Plasma Gas AMS Gas Flow RB This readback indicates the AMS gas L/min
flow rate (2000 series instruments
only).
Plasma Gas AMS Gas Solenoid This ICM command instructs the NA
secondary (AMS) gas solenoid to Open
or Close (2000 series instruments only).
Plasma Gas Oxygen Gas Flow This ICM command to the oxygen gas L/min
mass flow controller sets the oxygen
gas flow rate. Adjust the flow rate as
required.
Plasma Gas Oxygen Gas Flow RB This readback from the oxygen gas L/min
mass flow controller to the ICM
indicates the oxygen gas flow rate.
Plasma Gas Oxygen Gas Solenoid This ICM command instructs the NA
secondary (oxygen) gas solenoid to
Open or Close.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 339
Software Reference Guide
340 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
system sub component description units
system
Plasma Interlocks RFG Temp. RB This readback from the RF generator C
power tube temperature sensor to the
ICM indicates the RF generator
temperature. (300/350 series
instruments only)
Plasma Interlocks RFG Thermostat RB This is a readback to the ICM from the NA
RF generator thermostat interlock.
Plasma Interlocks Torch Box Closed RB This readback from the torch box NA
interlocks to the ICM indicates whether
the torch box/RFG door is closed, or
has rotated to the open position.
Plasma Interlocks Torch Box in Place 1 This readback from the torch box NA
RB interlocks to the ICM indicates whether
torch box interlock 1 is in position or in
a fault state.
Plasma Interlocks Torch Box in Place 2 This readback from the torch box NA
RB interlocks to the ICM indicates whether
torch box interlock 2 is in position or in
a fault state.
Plasma Interlocks Torch Box Rotation This ICM command locks or unlocks the NA
Lock door that allows access to the torch
box, cones, and RF generator. When
unlocked, the door can then be
manually opened. (300/350 series
instruments only)
Plasma Interlocks Torch Box Temp. RB This readback from the torch box C
temperature sensor to the ICM
indicates the torch box temperature.
Plasma Interlocks Torch Box Temp. This readback from the torch box to the NA
Status RB ICM indicates whether the
temperature is within normal range.
Plasma Interlocks Torch Mount Position This readback from the torch mount NA
RB interlock to the ICM indicates whether
the torch mount is in operating
position or in a fault state.
Plasma Peristaltic Pump Control This readback from the peristaltic NA
Register RB pump indicates the status of the
control register.
Plasma Peristaltic Pump Speed This ICM command sets the peristaltic rpm
pump speed. Adjust the speed as
required.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 341
Software Reference Guide
342 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
system sub component description units
system
Plasma RF SS‐RFG Q2 Current This readback from the far end of the A
Generator load coil indicates the Q2 value in
amps. This must be balanced with the
Q1 reading; they should generally be
within one amp of each other (around
22‐23 amps).
Plasma RF SS‐RFG Temp. This readback from the solid state RFG C
Generator glycol cooling block indicates the
temperature in degrees Celcius. It
should be roughly equal to the
incoming temperature.
Plasma RF SS‐RFG Plasma This optical sensor senses plasma
Generator Brightness Level brightness to determine whether or
not the plasma is lit. The value should
generally be in the 900 range; if it is
unusually low, the plasma may not be
lit.
Plasma RF SS‐RFG Plasma IR This infrared sensor indicates whether NA
Generator Sensor or not the plasma is lit. (Not used)
Plasma RF SS‐RFG VDD Supply This readback indicate s the voltage of Volts
Generator the small driver transistors that power
the main transistors.
Plasma RF SS‐RFG Q1 Gate This readback indicates the Q1 gate Volts
Generator Voltage voltage on the field effect transistors.
Increase the voltage for greater plasma
power.
Plasma RF SS‐RFG Q2 Gate This readback indicates the Q2 gate Volts
Generator Voltage voltage on the field effect transistors.
Increase the voltage for greater plasma
power.
Plasma RF SS‐RFG Driven Mode This readback indicates the voltage of Volts
Generator Gate Voltage the driven mode portion of the hybrid
gate system, which is used to light the
plasma. (Not used)
Plasma RF SS‐RFG Oscillation This readback indicates the voltage of Volts
Generator Mode Gate Voltage the oscillation portion of the hybrid
gate system, which is used to run the
plasma at a set frequency.
Plasma RF SS‐RFG CPU Firmware This readback indicates the firmware NA
Generator Version version number of the RF generator
CPU.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 343
Software Reference Guide
344 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Appendix B
Nano Application
This section provides background information about using the Nano App for
Syngistix™ ICP-MS software.
Note: The Nano App software must be purchased separately and installed together with
the Syngistix for ICP-MS base software. See the Syngistix Installation Guide for details.
Topics in this section include:
• Nano Application for Syngistix™ ICP-MS Software on page 345
• Understanding Nanoparticle Analyses on page 345
• Nano App Software Reference on page 346
• Analysis Screen on page 346
• Results Screen on page 361
345
Software Reference Guide
dissolved calibration so that it can be used for converting the particle net area
intensity data into size information through a mass flux calibration curve. If this is
required and unknown, you must run at least one standard nanoparticle solution
with a known size or particle concentration. If you are calculating the TE based on
particle size, you can run multiple standards to gather more precise data for your
calibration (it is ideal if one of your standards contains particles close to the size of
the nanoparticles you are looking for).
To reduce the effect of analyte interferences on nanoparticle detection when
developing a method to run in DRC mode, you must adjust three key parameters -
the Gas Flow, RPq, and AFT settings - to suppress noise while allowing
nanoparticle signals to be distinguished from random background fluctuations.
This is especially true when looking at the particles close to the size detection
limits. A high gas flow can reduce background substantially, but if the gas flow is
too high, fluctuations in the dissolved standard can appear as peaks in your
results due to increased disturbance in the ion movement through the pressurized
cell. We recommend that you tune your method by running your nanoparticle
blank and nanoparticle standards as samples, reviewing the results, and
repeating - adjusting the key parameters between each run - until you have
achieved sufficiently low background interference without introducing unwanted
peaks in your particle blank. You can then finalize your cell parameters and begin
your analysis.
Note: Cell gas delays defined in the Syngistix Options dialog box Acquisition Profile tab
are applied to all DRC mode methods used in your nano analyses. All other default
settings configured in the Syngistix Options tabs do not apply to the Nano App methods
and analyses. You must enter all other settings you wish to have apply to your
nanoparticle measurements in the Nano App itself, or within the methods created for it.
Analysis Screen
The Nano App Analysis screen provides all the controls needed to set up and run
dissolved and nanoparticle calibrations, and sample characterizations, with
realtime results panels for parsing your data on the fly. It is comprised of three
tabbed sections for Sample analyses, Batch analyses, and Transport Efficiency
(TE) calculations, each of which is divided into several panels:
• Acquisition panel: This panel has different controls for each of the three
tabs, but is generally used to control the acquisition, and run the required
blanks and standards for building calibrations, and samples.
• Method panel: The Method panel is similar across all of the analysis tabs.
Use this panel to load, create, or modify the analytical methods you will
use in your nanoparticle calibrations and analyses. This panel has three
tabs — a Parameters tab, Calibration tab, and a Pump Settings tab
where you configure the various settings.
• Calibration panel: This panel is also similar across all of the analysis
tabs. The Calibration panel displays data, either for an existing loaded
Sample or Transport Efficiency calibration, or in real time, as a calibration
is performed by the software (updated and recalculated continuously over
the course of your analyses).
346 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
• Realtime Signal panel: The Realtime Signal panel displays the data and
progress of your measurements as they occur in real time.
• Realtime Histogram panel: This panel displays the calculated intensity
histogram appropriate for the solutions as they are processed.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 347
Software Reference Guide
Acquisition Panel
Use this panel to control the acquisition, and run the required blanks and
standards for building calibrations, and samples. This panel is slightly different for
each of the three analysis tabs; each is illustrated and explained following.
Sample Tab Acquisition Panel
Figure B-2 Nano App Analysis screen Sample Tab Acquisition Panel
Items on the Sample Tab Acquisition Panel
• Start/Stop Pump: Click this toggle control to start or stop the peristaltic
pump.
• Pump speed selector: Type (or click the arrows to select) the peristaltic
pump speed in rpm. Range = ±150 rpm.
• Sample Flow Rate: Type the measured sample flow rate in mL/min,
based on the installed pump tubing size and the selected pump settings.
Range = 0.001 to 3.
• Calibration Group:
Calibration drop-down list: Select either a Dissolved or Particle
calibration type.
Analyze Blank: Click to begin analysis of the blank solution.
Analyze Standard: Click to begin analysis of the specified standard
solution. The number displayed corresponds to the standard solution
row in the Method panel.
348 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
• Analyze Sample: In the accompanying field, type the unique name of the
sample to be analyzed, and then click Analyze Sample to begin the
analysis.
• Transport Efficiency: Type the determined Transport Efficiency (TE) as a
percentage.
• Advanced: Click the Advanced button to access the Advanced Options
dialog box, where you can configure a counts override threshold, as well
as range and size settings for the related histogram display. See
Advanced Options Dialog Box on page 352 for more information.
TE Tab Acquisition Panel
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 349
Software Reference Guide
If you select this option, two additional fields appear at the bottom of
this panel:
— Mass Conc.: Type the mass concentration of the standard solution
in ng/L.
— Part. Conc.: Type the particle concentration of the standard
solution in parts/mL.
You can enter a value into only one of these fields; when you do so, the
other will become disabled.
• Calibration Group:
Calibration drop-down list: Select either a Dissolved or Particle
calibration type.
Analyze Blank: Click to begin analysis of the blank solution.
Analyze Standard: Click to begin analysis of the specified standard
solution.
TE: Displays the determined Transport Efficiency as a percentage.
• Advanced: Click the Advanced button to access the Advanced Options
dialog box, where you can configure a counts override threshold.
(Redundant range and size settings for the histogram also appear here,
but have no bearing on the transport efficiency.) See Advanced Options
Dialog Box on page 352 for more information.
350 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Batch Tab Acquisition Panel
Figure B-4 Nano App Analysis screen Batch Tab Acquisition Panel
Items on the Batch Tab Acquisition Panel
• Dataset Folder: This field indicates the name and location of the dataset
where the data from your batch analyses will be saved. Click Browse to
select a dataset folder location.
• Parameters in the Batch Table:
A/S Loc.: Lists the autosampler position for the sample — the location
that the solution occupies in the autosampler tray. Type the number
corresponding to the position of the solution in the autosampler tray.
Sample ID: Identifies each sample with a unique name. Type a name
(any alphanumeric combination) to enter an individual Sample ID.
Measurement Action: Specifies an analysis or control action when
running the given row of the batch. To select an action, click Select
Action. The following options are available:
Run Blank: Runs the specified blank solution. Select either a Particle
or Dissolved solution.
Run Particle Standard: Runs the specified particle standard solution.
Run Dissolved Standard: Runs the specified dissolved standard
solution.
Run Sample: Runs the specified unknown sample solution.
AutoStop: Turns off the plasma automatically following the batch
analysis; this action must be placed last in the batch.
Pause: Pauses the batch analysis until you choose to resume it. Use
this feature when analyzing multiple samples without an
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 351
Software Reference Guide
Figure B-5 Nano App Analysis screen Advanced Options Dialog Box
Items on the Advanced Options Dialog Box
• Override Threshold: To override the existing counts threshold, select this
check box and type a preferred Counts value for the new threshold.
• Size Histogram Bin Size: Type a size in nm to indicate the size of bin in
which to group displayed histogram data points. Range = 1-100.
• Fixed Range for Size Histogram: To specify a particular fitting range for
the size histogram graphic display, select this check box. Then, in the Start
352 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
field, type the lower end of the range you wish to display in nm (range = 0-
29,999). In the End field, type the upper end of the range you wish to
display in nm (range = 1-30,000).The Start value must be lower than the
End value.
Method Panel
Use the Method panel to load, create, or modify the analytical methods you will
use in your nanoparticle calibrations and analyses. This panel is similar across all
of the Analysis screen main tabs.
The Method panel is composed of three tabs — a Parameters tab, Calibration
tab, and a Pump Settings tab.
Figure B-6 Nano App Analysis screen Method panel Parameters tab
Items on the Method Panel Parameters Tab
• Dwell Time and Scan Time: The Dwell Time is the length of time to
measure the analyte during a single reading in microseconds; the Scan
Time is the total acquisition time in seconds. These values are dependent
on one another, but a typical combination is a Dwell Time of 50 µs
(microseconds) with a Scan Time of 100 seconds.
The Dwell Time has an absolute range of 10-50 000 micro seconds, and
the Scan Time has an absolute range of 1-18 000 seconds. However, the
two times are validated against one another to prevent processing lag, so
that the actual values are valid only when the scan time is within the
allowable range for the selected dwell time (and vice versa), and the total
measurements are less than or equal to 6 000 000 data points.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 353
Software Reference Guide
• Conditions File: Unless you specify otherwise, the Nano App uses the
same default.dac Conditions file used by the base Syngistix software. To
define different conditions for your nano analyses, you can modify the
default.dac in Syngistix or create a custom conditions file and save it with
your nano method. If you do this, ensure that you save it with a unique
name to avoid future confusion.
• Analyte information: Select your analyte of interest from the drop-down
list; the table is automatically populated with the information for the most
abundant isotope. Review the remaining columns; generally you will use
the default values as presented; however you can make changes to suit
your needs if required.
Analyte: Type or select the analyte of interest.
Mass (amu): The mass of the selected isotope of the analyte of
interest. The most abundant isotope is selected by default, but you
may select an alternate isotope from the drop-down list if desired.
Density: Type the density of the nanoparticle in g/cm3.
Mass Fraction (%): Type the mass fraction of the nanoparticle that is
composed of the analyte, expressed as a percentage. A default value
of 100% is typically applied.
Ionization Efficiency (%): Type the ratio of the ionization efficiency of
the nanoparticle to the ionization efficiency of the corresponding
dissolved analyte solution. A default value of 100% is typically applied.
RPq: (Rejection Parameter q) Specifies the RF voltage applied to the
reaction cell quadrupole. Range = 0.2 to 0.8.
• Cell Parameters information: UCT instruments only.
Profile: Select the acquisition profile to be used for the determination—
you can use the default Standard profile or any linked DRC mode
profile currently assigned to a physical gas channel. (You cannot use
unlinked profiles for this application.)
Gas Flow (mL/min.): Type the desired flow rate for the cell gas
associated with the selected profile. Valid values = 0.1 to 7.288
mL/min. The default value is 0.3 mL/min.
AFT (V): In DRC mode, the AFT system applies a linearly accelerating
axial field to the ion beam to decrease matrix effects, stabilize the
analyte signal, and increase the speed of the analysis during the
normal operation of the instrument. Type an axial field adjustment
value between ±500. The default value is 350.
354 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
• New: Click to create a new nano method. Default values are automatically
populated; make any necessary changes and then save the method under
a unique name.
Figure B-7 Nano App Analysis screen Method panel Calibration tab
Items on the Method Panel Calibration Tab
• Dissolved Standard list: Type the known mass concentration of analyte
for each of your standard dissolved solutions in µg/L. You can define up to
seven standards, as applicable.
• Particle Standard list: Type the known most abundant size of
nanoparticles for each of your standard particle solutions in nm (in the
case of nanoparticles with a Gaussian size distribution, this is also the
mean size). You can define up to seven standards, as applicable.
• Load: Click to Load an existing nano method.
• Save: Click to Save changes to the current nano method.
• New: Click to create a new nano method. Default values are automatically
populated; make any necessary changes and then save the method under
a unique name.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 355
Software Reference Guide
Figure B-8 Nano App Analysis screen Method panel Pump Settings tab
Items on the Method Panel Pump Settings Tab
This tab displays peristaltic pump parameters for the method. Change the default
settings as necessary to override the pump settings for this nano method:
• Peristaltic Pump table settings:
Sample Flush Time: Specifies the time in seconds during which the
sample tubing is flushed with sample solution. Type a value between 0
and 999.
Sample Flush Speed: Specifies the pump speed in rpm during the
flush cycle. Any value between ±150 (2000 series instruments) or ±48
(300/350 series instruments) is valid. A negative value indicates
counter-clockwise rotation of the pump head and a positive value
indicates clockwise rotation.
Read Delay Time: Specifies the time in seconds between the end of
the flush cycle and the beginning of data acquisition. Type a value
between 0 and 999.
Read Delay Speed: Specifies the pump rate in rpm used during the
read delay cycle. Any value between ±150 (2000 series instruments)
or ±48 (300/350 series instruments) is valid. A negative value indicates
counter-clockwise rotation of the pump head and a positive value
indicates clockwise rotation. The Read Delay Speed is the speed that
will be used during the sample analysis.
Wash Time: Specifies the time in seconds during which the sample
tubing is rinsed with wash solution following completion of data
acquisition. Type a value between 0 and 999.
356 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Wash Speed: Specifies the pump speed in rpm used during the wash
cycle. Any value between ±150±150 (2000 series instruments) or ±48
(300/350 series instruments) is valid. A negative value indicates
counter-clockwise rotation of the pump head and a positive value
indicates clockwise rotation. During a wash cycle, you can use a more
rapid pump rate to quickly wash sample solution from the sample
tubing.
• Peristaltic Pump Under Computer Control: Select this check box if you
want the peristaltic pump speeds to be automatically controlled by the
software. If this check box is cleared, you must control pump speeds
manually, using the pump speed control on the Acquisition panel.
• Load: Click to Load an existing nano method.
• Save: Click to Save changes to the current nano method.
• New: Click to create a new nano method. Default values are automatically
populated; make any necessary changes and then save the method under
a unique name.
Calibration Panel
The Calibration panel displays data, either for an existing loaded Sample or
Transport Efficiency calibration, or in real time, as a calibration is performed by the
software (updated and recalculated continuously over the course of your
analyses). This panel is similar across all of the Analysis screen main tabs.
Settings for the calibration display are configured on a floating Calibration Options
menu that overlays the panel.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 357
Software Reference Guide
Note: To remove an entry from the Sample table, right-click the item and then
click Remove.
Note: Ensure that you save your calibrations with a unique label. For future
reference, it may be useful to include a “particle” or “dissolved” marker to
differentiate the two types of calibration.
358 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Realtime Signal Panel
As you run each analysis, the Realtime Signal panel displays the data and
progress of your measurements as they occur.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 359
Software Reference Guide
360 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
•
Results Screen
The Nano App Results screen allows you to review in detail the calculated
histograms and related statistics calculated for each sample or analysis. Here you
can also reprocess your data by modifying the variables used for these
calculations. The following panels comprise the Results screen:
• File Options panel: The File Options panel displays the location of the
saved datasets, and provides some file handling options for each sample.
• Parameters panel: The Parameters panel displays key settings from the
method used for the selected sample or calibration.
• Calibrations panel: The Calibration panel displays the related dissolved
and particle calibration curves; if a transport efficiency value is specified,
the dissolved calibration is converted to a mass flux calibration curve.
• Results Table panel: This panel lists all the blank, standard, and sample
measurements for either the current sample set, or an existing dataset that
has been loaded in this space.
• Histogram panel: The Histogram panel displays the intensity or size
distribution for the selected sample.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 361
Software Reference Guide
362 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Parameters Panel
The Parameters panel displays key settings from the method used for the
selected sample or calibration.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 363
Software Reference Guide
Calibrations Panel
The Calibration panel displays the related dissolved and particle calibration
curves; if a transport efficiency value is specified, the dissolved calibration is
converted to a mass flux calibration curve.
For dissolved calibrations, this will subtract the blank from all points.
For particle calibrations, this will subtract the blank from the mean
dissolved intensity of the sample.
• Slope: Displays the rate of change of intensity as a function of
concentration or mass.
• Intercept: Displays the value at which the calibration curve crosses the
intensity axis.
• R2: The coefficient of determination (range =0-1). This defines how closely
your data aligns to the linear calibration plot line.
364 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Results Table Panel
This panel lists all the blank, standard, and sample determinations for either the
current sample set, or an existing dataset that has been loaded in this space.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 365
Software Reference Guide
Histogram Panel
The Histogram panel displays the intensity or size histogram for the selected
sample, with data grouped by the selected bin size and plotted linearly along the
x-axis. Settings for this display are configured on a floating Histogram Options
menu that overlays the panel.
366 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Switch to Dissolved / Switch to Particle toggle button: Click to show
either the dissolved or particle histogram data.
Fitting options drop-down list: These options allow you manage how
well the calculated curve fits the histogram data. Select the curve type
that best fits the selected region of the histogram data. The available
options are:
Gaussian: Fits a Gaussian Curve to the Histogram data in the
selected region and selects the maximum of the fitted curve as the
most abundant point observed.
LogNormal: Fits a LogNormal Curve to the histogram data in the
selected region, and selects the maximum of the fitted curve as the
most abundant point observed.
Max. Intensity: Selects the point with the maximum frequency in
the selected region as the most abundant point observed.
• Fitting window slider: Use the sliding window control to select the region of
interest and focus. A line plot of the histogram is plotted below the slider to
help guide you in selecting the region of interest.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 367
Software Reference Guide
368 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Appendix C
Note: The Single Cell App software must be purchased separately and installed together
with the Syngistix for ICP-MS base software. See the Syngistix Installation Guide for
details.
Topics in this section include:
• Single Cell Application for Syngistix™ ICP-MS Software on page 369
• Single Cell App Software Reference on page 371
• Analysis Screen on page 371
• Results Screen on page 385
Sample Preparation
Before running single cell analyses you must consider the composition of both the
media in which you have cultured your sample cells, and the standard you are
using for your analyses:
• Very complex media (such as RPMI), or those that are very high in salt can
suppress or interfere with the desired signals. In these cases, you may
need to transfer your cells to a less complex medium prior to analysis,
taking into account the amount of time your cells can survive in a less
complex media.
• If your cells must be separated from their exposure media, you will
generally need to wash and centrifuge them before resuspending them in
a fresh solution. We recommend a cycle in which you centrifuge and wash
the cells for at least three cycles, taking the delicacy of the given cells into
account: refer to the literature for your specific cells to determine the
appropriate procedures.
• It is the rapid, continuous measurement capabilities of the ICP-MS that
allows the Single Cell App to detect and analyze each cell individually.
Nevertheless, you may need to dilute samples having a particularly high
concentration of cells or particles, to avoid multiple cells being captured
369
Software Reference Guide
370 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Single Cell App Software Reference
The application has separate screens for cell analysis setup and data review and
processing.
Analysis Screen
The Single Cell App Analysis screen provides all the controls needed to set up
and run dissolved and particle calibrations, and sample characterizations, with
realtime results panels for viewing your data on the fly. It is comprised of three
tabbed sections for Sample analyses, Batch analyses, and Transport Efficiency
(TE) calculations, each of which is divided into several panels.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 371
Software Reference Guide
Acquisition Panel
Use this panel to control the acquisition, and run the required blanks and
standards for building calibrations, and samples. This panel is slightly different for
each of the three analysis tabs; each is illustrated and explained following.
Sample Tab Acquisition Panel
Figure C-2 Single Cell App Analysis screen Sample Tab Acquisition Panel
Items on the Sample Tab Acquisition Panel
• Start/Stop Pump: Click this toggle control to start or stop the peristaltic
pump.
• Pump speed selector: Type (or click the arrows to select) the peristaltic
pump speed in rpm. Range = ±150 rpm.
• Sample Flow Rate: Type the measured sample flow rate in mL/min,
based on the installed pump tubing size and the selected pump settings.
Range = 0.001 to 3.
• Calibration Group:
Calibration drop-down list: Select either a Dissolved or Particle
calibration type.
Analyze Blank: Click to begin analysis of the blank solution.
Analyze Standard: Click to begin analysis of the specified standard
solution. The number displayed corresponds to the standard solution
row in the Method panel.
372 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
• Analyze Sample: In the accompanying field, type the unique name of the
sample to be analyzed, and then click Analyze Sample to begin the
analysis.
• Transport Efficiency: Type the determined Transport Efficiency (TE) as a
percentage.
• Advanced: Click the Advanced button to access the Advanced Options
dialog box, where you can configure a counts override threshold, as well
as range and size settings for the related histogram display. See
Advanced Options Dialog Box on page 376 for more information.
TE Tab Acquisition Panel
Figure C-3 Single Cell App Analysis screen TE Tab Acquisition Panel
Items on the TE Tab Acquisition Panel
The Transport Efficiency function calculates the efficacy of the sample delivery,
taking into account loss due to sampling and nebulization, to provide the actual
delivery percentage.
• Start/Stop Pump: Click this toggle control to start or stop the peristaltic
pump.
• Pump speed selector: Type (or click the arrows to select) the peristaltic
pump speed in rpm. Range = ±150 rpm.
• Sample Flow Rate: Type the measured sample flow rate in mL/min,
based on the installed pump tubing size and the selected pump settings.
Range = 0.001 to 3.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 373
Software Reference Guide
If you select this option, two additional fields appear at the bottom of
this panel:
— Mass Conc.: Type the mass concentration of the standard solution
in ng/L.
— Part. Conc.: Type the particle concentration of the standard
solution in parts/mL.
You can enter a value into only one of these fields; when you do so, the
other will become disabled.
• Calibration Group:
Calibration drop-down list: Select either a Dissolved or Particle
calibration type.
Analyze Blank: Click to begin analysis of the blank solution.
Analyze Standard: Click to begin analysis of the specified standard
solution.
TE: Displays the determined Transport Efficiency as a percentage.
• Advanced: Click the Advanced button to access the Advanced Options
dialog box, where you can configure a counts override threshold.
(Redundant range and size settings for the histogram also appear here,
but have no bearing on the transport efficiency.) See Advanced Options
Dialog Box on page 376 for more information.
374 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Batch Tab Acquisition Panel
Figure C-4 Single Cell App Analysis screen Batch Tab Acquisition Panel
Items on the Batch Tab Acquisition Panel
• Dataset Folder: This field indicates the name and location of the dataset
where the data from your batch analyses will be saved. Click Browse to
select a dataset folder location.
• Parameters in the Batch Table:
A/S Loc.: Lists the autosampler position for the sample — the location
that the solution occupies in the autosampler tray. Type the number
corresponding to the position of the solution in the autosampler tray.
Sample ID: Identifies each sample with a unique name. Type a name
(any alphanumeric combination) to enter an individual Sample ID.
Measurement Action: Specifies an analysis or control action when
running the given row of the batch. To select an action, click Select
Action. The following options are available:
Run Blank: Runs the specified blank solution. Select either a Particle
or Dissolved solution.
Run Particle Standard: Runs the specified particle standard solution.
Run Dissolved Standard: Runs the specified dissolved standard
solution.
Run Sample: Runs the specified unknown sample solution.
AutoStop: Turns off the plasma automatically following the batch
analysis; this action must be placed last in the batch.
Pause: Pauses the batch analysis until you choose to resume it. Use
this feature when analyzing multiple samples without an
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 375
Software Reference Guide
Figure C-5 Single Cell App Analysis screen Advanced Options Dialog Box
Items on the Advanced Options Dialog Box
• Override Threshold: To override the existing counts threshold, select this
check box and type a preferred Counts value for the new threshold.
• Mass Histogram Bin Size: Type a size in ag to indicate the size of bin in
which to group displayed histogram data points. Range = 1-5,000 ag.
• Fixed Range for Mass Histogram: To specify a particular fitting range for
the mass histogram graphic display, select this check box. Then, in the
376 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Start field, type the lower end of the range you wish to display in ag (range
= 0-499,999,999). In the End field, type the upper end of the range you
wish to display in ag (range = 1-500,000,000). The Start value must be
lower than the End value.
Method Panel
Use the Method panel to load, create, or modify the analytical methods you will
use in your calibrations and analyses. This panel is similar across all of the
Analysis screen main tabs.
The Method panel is composed of three tabs — a Parameters tab, a Calibration
tab, and a Pump Settings tab.
Figure C-6 Single Cell App Analysis screen Method panel Parameters tab
Items on the Method Panel Parameters Tab
• Dwell Time and Scan Time: The Dwell Time is the length of time to
measure the analyte during a single reading in microseconds; the Scan
Time is the total acquisition time in seconds. These values are dependent
on one another, but a typical combination is a Dwell Time of 50 µs
(microseconds) with a Scan Time of 100 seconds.
The Dwell Time has an absolute range of 10-50,000 micro seconds, and
the Scan Time has an absolute range of 1-18,000 seconds. However, the
two times are validated against one another to prevent processing lag, so
that the actual values are valid only when the scan time is within the
allowable range for the selected dwell time (and vice versa), and the total
measurements are less than or equal to 6,000,000 data points.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 377
Software Reference Guide
• Conditions File: Unless you specify otherwise, the Single Cell App uses
the same default.dac Conditions file used by the base Syngistix software.
To define different conditions for your single cell analyses, you can modify
the default.dac in Syngistix or create a custom conditions file and save it
with your single cell method. If you do this, ensure that you save it with a
unique name to avoid future confusion.
• Analyte information: Select your analyte of interest from the drop-down
list; the table is automatically populated with the information for the most
abundant isotope. Review the remaining columns; generally you will use
the default values as presented; however you can make changes to suit
your needs if required.
Analyte: Type or select the analyte of interest.
Mass (amu): The mass of the selected isotope of the analyte of
interest. The most abundant isotope is selected by default, but you
may select an alternate isotope from the drop-down list if desired.
Density: Type the density of the particle in g/cm3.
Mass Fraction (%): Type the mass fraction of the particle that is
composed of the analyte, expressed as a percentage. A default value
of 100% is typically applied.
Ionization Efficiency (%): Type the ratio of the ionization efficiency of
the particle to the ionization efficiency of the corresponding dissolved
analyte solution. A default value of 100% is typically applied.
RPq: (Rejection Parameter q) Specifies the RF voltage applied to the
reaction cell quadrupole. Range = 0.2 to 0.8.
• Cell Parameters information: UCT instruments only.
Profile: Select the acquisition profile to be used for the determination—
you can use the default Standard profile or any linked DRC mode
profile currently assigned to a physical gas channel. (You cannot use
unlinked profiles for this application.)
Gas Flow (mL/min.): Type the desired flow rate for the cell gas
associated with the selected profile. Valid values = 0.1 to 7.288
mL/min. The default value is 0.3 mL/min.
AFT (V): In DRC mode, the AFT system applies a linearly accelerating
axial field to the ion beam to decrease matrix effects, stabilize the
analyte signal, and increase the speed of the analysis during the
normal operation of the instrument. Type an axial field adjustment
value between ±500. The default value is 350.
378 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
• New: Click to create a new single cell method. Default values are
automatically populated; make any necessary changes and then save the
method under a unique name.
Figure C-7 Single Cell App Analysis screen Method panel Calibration tab
Items on the Method Panel Calibration Tab
• Dissolved Standard list: Type the known mass concentration of analyte
for each of your standard dissolved solutions in µg/L. You can define up to
seven standards, as applicable.
• Particle Standard list: Type the known most abundant size of particles for
each of your standard particle solutions in nm (in the case of particles with
a Gaussian size distribution, this is also the mean size). You can define up
to seven standards, as applicable. The software converts each value
entered to the corresponding mass in ag, and displays both numbers.
• Load: Click to Load an existing single cell method.
• Save: Click to Save changes to the current single cell method.
• New: Click to create a new single cell method. Default values are
automatically populated; make any necessary changes and then save the
method under a unique name.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 379
Software Reference Guide
Figure C-8 Single Cell App Analysis screen Method panel Pump Settings tab
Items on the Method Panel Pump Settings Tab
This tab displays peristaltic pump parameters for the method. Change the default
settings as necessary to override the pump settings for this single cell method:
• Peristaltic Pump table settings:
Sample Flush Time: Specifies the time in seconds during which the
sample tubing is flushed with sample solution. Type a value between 0
and 999.
Sample Flush Speed: Specifies the pump speed in rpm during the
flush cycle. Any value between ±150 (2000 series instruments) or ±48
(300/350 series instruments) is valid. A negative value indicates
counter-clockwise rotation of the pump head and a positive value
indicates clockwise rotation.
Read Delay Time: Specifies the time in seconds between the end of
the flush cycle and the beginning of data acquisition. Type a value
between 0 and 999.
Read Delay Speed: Specifies the pump rate in rpm used during the
read delay cycle. Any value between ±150 (2000 series instruments)
or ±48 (300/350 series instruments) is valid. A negative value indicates
counter-clockwise rotation of the pump head and a positive value
indicates clockwise rotation. The Read Delay Speed is the speed that
will be used during the sample analysis.
Wash Time: Specifies the time in seconds during which the sample
tubing is rinsed with wash solution following completion of data
acquisition. Type a value between 0 and 999.
380 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Wash Speed: Specifies the pump speed in rpm used during the wash
cycle. Any value between ±150 (2000 series instruments) or ±48
(300/350 series instruments) is valid. A negative value indicates
counter-clockwise rotation of the pump head and a positive value
indicates clockwise rotation. During a wash cycle, you can use a more
rapid pump rate to quickly wash sample solution from the sample
tubing.
• Peristaltic Pump Under Computer Control: Select this check box if you
want the peristaltic pump speeds to be automatically controlled by the
software. If this check box is cleared, you must control pump speeds
manually, using the pump speed control on the Acquisition panel.
• Load: Click to Load an existing single cell method.
• Save: Click to Save changes to the current single cell method.
• New: Click to create a new single cell method. Default values are
automatically populated; make any necessary changes and then save the
method under a unique name.
Calibration Panel
The Calibration panel displays data, either for an existing loaded Sample or
Transport Efficiency calibration, or in real time, as a calibration is performed by the
software (updated and recalculated continuously over the course of your
analyses). This panel is similar across all of the Analysis screen main tabs.
Settings for the calibration display are configured on a floating Calibration Options
menu that overlays the panel.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 381
Software Reference Guide
Note: To remove an entry from the Sample table, right-click the item and then
click Remove.
Note: Ensure that you save your calibrations with a unique label. For future
reference, it may be useful to include a “particle” or “dissolved” marker to
differentiate the two types of calibration.
382 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Realtime Signal Panel
As you run each analysis, the Realtime Signal panel displays the data and
progress of your measurements as they occur.
Figure C-10 Single Cell App Analysis screen Realtime Signal panel
Figure C-11 Single Cell App Analysis screen Realtime Histogram panel
Items on the Realtime Histogram Panel
• Histogram Options floating menu:
Override: Select this check box to override the existing count
threshold with the value you type in the Override Threshold field.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 383
Software Reference Guide
Apply: Click this button to apply the new threshold to the histogram
graphic display.
Fitting options drop-down list: For particle standards only These
options allow you manage how well the calculated curve fits the
histogram data. Select the curve type that best fits the selected region
of the histogram data. The available options are:
Gaussian: Fits a Gaussian Curve to the Histogram data in the
selected region and selects the maximum of the fitted curve as the
most abundant point observed.
LogNormal: Fits a LogNormal Curve to the histogram data in the
selected region, and selects the maximum of the fitted curve as the
most abundant point observed.
Max. Intensity: Selects the point with the maximum frequency in
the selected region as the most abundant point observed.
Send to Results: Click this button to copy the data from this display
into the Results table.
• Zoom Out control: Click the zoom arrows button (arrows pointing out) to
the left of the horizontal slider to zoom out on the histogram display.
• Fitting window slider: For particle standards only Use the sliding window
control to select the region of interest and focus. A line plot of the
histogram is plotted below the slider to help guide you in selecting the
region of interest.
384 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Results Screen
The Single Cell App Results screen allows you to review in detail the calculated
histograms and related statistics calculated for each sample or analysis. Here you
can also reprocess your data by modifying the variables used for these
calculations.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 385
Software Reference Guide
Figure C-13 Single Cell App Results screen File Options panel
Items on the File Options Panel
• Dataset Folder: Displays the location to which single cell dataset files are
saved. Upon opening, it displays the default dataset location:
\\Users\<Public or My> Documents\PerkinElmer Syngistix\ICPMS \Single
Cell\Dataset. Click Browse to select a new location.
• View Dataset Files: Click to select a set of datasets to view in the current
results table.
• Dissolved Calibration File: Displays the name and location of the
currently loaded dissolved calibration file. Click Load to open an existing
dissolved calibration file for the selected sample.
• Particle Calibration File: Displays the name and location of the currently
loaded particle calibration file. Click Load to open an existing particle
calibration file for the selected sample.
• Export Current Sample: Click Export to export the data from the sample
file currently selected in the Results table, in Microsoft Excel .xls or .xlsx
format.
• Export Results Table: Click Export to export the data in the current
results table in Microsoft Excel .xls or .xlsx format.
386 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Parameters Panel
The Parameters panel displays key settings from the method used for the
selected sample or calibration.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 387
Software Reference Guide
Calibrations Panel
The Calibration panel displays the related dissolved and particle calibration
curves. If a transport efficiency value is specified, the dissolved calibration is
converted to a mass flux calibration curve.
For dissolved calibrations, this will subtract the blank from all points.
For particle calibrations, this will subtract the blank from the mean
dissolved intensity of the sample.
• Slope: Displays the rate of change of intensity as a function of
concentration or mass.
• Intercept: Displays the value at which the calibration curve crosses the
intensity axis.
• R2: The coefficient of determination (range =0-1). This defines how closely
your data aligns to the linear calibration plot line.
388 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Results Table Panel
This panel lists all the blank, standard, and sample determinations for either the
current sample set, or an existing dataset that has been loaded in this space.
Figure C-16 Single Cell App Results screen Results Table panel
Items on the Results Table Panel
• Sample: Displays the sample type (sample, standard, or blank) or ID.
• Analyte: Identifies the analyte of interest for this sample.
• Most Frequent Mass: The most abundant mass of the detected particles
in the selected histogram region, measured in ag.
• Mean Mass: The average mass of the detected particles in the selected
histogram region, measured in ag.
• No. of Peaks: The number of event peaks detected in the selected
histogram region.
• Mean Inten. (counts): The average area intensity observed for event
peaks in the selected histogram region, measured as counts per cell (with
the background signal subtracted).
• Cell Conc. (parts/mL): Calculated cell concentration in the sample based
on the transport efficiency and the number of peaks detected.
• Diss. Inten. (counts): Background dissolved intensity, calculated as the
average intensity per dwell time observed for detected dissolved
background.
• Diss. Conc. (ppb): Concentration of dissolved background signal in ppb,
calculated from the dissolved intensity, and based on the dissolved
calibration.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 389
Software Reference Guide
• Bin Size (ag): Displays the size of bin in which the data points are
grouped in ag.
• Start (ag): Indicates the lower end of the specified fitting range.
• End (ag): Indicates the upper end of the specified fitting range.
Histogram Panel
The Histogram panel displays the intensity or mass histogram for the selected
sample, with data grouped by the selected bin size and plotted linearly along the
x-axis. Settings for this display are configured on a floating Histogram Options
menu that overlays the panel.
390 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
LogNormal: Fits a LogNormal Curve to the histogram data in the
selected region, and selects the maximum of the fitted curve as the
most abundant point observed.
Max. Intensity: Selects the point with the maximum frequency in
the selected region as the most abundant point observed.
• Fitting window slider: Use the sliding window control to select the region of
interest and focus. A line plot of the histogram is plotted below the slider to
help guide you in selecting the region of interest.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 391
Software Reference Guide
392 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Appendix D
Note: Additional related sample information is appended at the end of the appendix.
Variables
A Slope
B Y‐axis intercept
C Number of calibration points (corrected for degrees of freedom)
R Linear correlation coefficient
SigmaA Standard deviation of A
SigmaB Standard deviation of B
Terms used in calculations
Delta Determinant used in calculating A and B
Variance Variance
SumX Sum of X (where X is concentration values of the calibration
standards)
SumY Sum of Y (where Y is intensity values of the calibration
standards)
SumX2 Sum of X*X for all calibration standards
SumY2 Sum of Y*Y for all calibration standards.
SumXY Sum of X*Y values
NumPts Total number of calibration points
Temp Temporary variable
SumW Sum of weights
393
Software Reference Guide
Terms used in calculations
SumWX Sum of W*X for all calibration standards
SumWX2 Sum of W*X*X for all calibration standards
SumWY Sum of W*Y for all calibration standards
SumWY2 Sum of W*Y*Y for all calibration standards
SumWXY Sum of W*X*Y for all calibration standards
Note: For Simple Linear fit, the weight is 1 for each standard and, therefore,
SumW = the number of calibration points.
The standard deviation of the slope and intercept are calculated as follows:
C = NumPts - 2
if (C > 0.0)
Variance =
(SumY2+ B * B * SumW + A * A * SumX2 -2.0*(B * SumY + A * SumXY - B
* A * SumX)) / C
394 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Linear Through Zero Calibration
The slope is calculated as follows:
A = SumXY/ SumX2
The standard deviation of the slope is calculated as follows:
C = NumPts - 1
Note: For Weighted Linear fit, weight for each standard is calculated as 1 / X2 .
SumW = the sum of the weights of all standards.
The standard deviation of the slope and intercept are calculated as follows:
C = NumPts - 2
if (C > 0.0)
Variance =
(SumWY2+ B * B * SumW + A * A * SumWX2 -2.0*(B * SumWY + A *
SumWXY - B * A * SumWX)) / C
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 395
Software Reference Guide
(if (Temp < 0.0) OR (if Temp = 0.0) then it is an invalid fit.)
where:
Standard Measured Ratio = Net intensity of isotope in Standard
Net intensity of reference isotope in Standard
The Standard Known Ratio is the user entry of the certified ratio for the
Standard
396 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Precision
All values in the software are carried in double precision (as double type), as
follows:
Standard Deviation
Standard deviation is calculated as the square root of the product between the
variance and the Bessel factor, as follows:
n 1 n 1 2
xk2 x k
n
SD( x ) k 0
k 0
n n n 1
where n = the number of replicates for the sample, and x = Ii or Ni or Ci or Ri
and, n>1
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 397
Software Reference Guide
where:
INet: intensity used for concentration calculation
Ianal: obtained intensity for the analyte
Istd: obtained intensity for the internal standard
Iblk: obtained blank intensity for the analyte
Iblk, std: obtained blank intensity for the internal standard
Concentration and Standard deviation calculation in Standard
Addition
Intensities are corrected using Step 1.
The Calibration curve is obtained with unspiked sample #1 (unknown #1) and
spiked samples using a simple linear algorithm (see Slope, Intercept, Standard
Deviation and Correlation Coefficient algorithms for software calibration
regressions on page 393 for the calculation of the Simple Linear Calibration
parameters.)
For Standard Addition, the concentration of the spiked samples are listed in the
method and the concentration of the unspiked sample is considered as 0 in order
to generate the calibration curve.
Sheet 2 of the included file Calibration Validation_for Std Add.xls (see the end of
this section for appended Excel sheets) provides an example of analyzing
Unspiked Sample #1 and Unspiked Sample #2 for Pb 208 using standard
addition. The Corrected Intensities in Column 3 Sheet 2 are the Net Intensity
Means from the Syngistix software quantitation reports. Sheet 3 compares the
calculation of the Net Int. Mean (based on 3 replicates) between the Syngistix
software and Excel for Unspiked Sample #1.
398 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
For the first unspiked sample:
Cunspiked sample #1 = Iintercept / slope
where:
Cunspiked sample #1: analyte concentration of the unspiked sample #1
Iintercept: intercept of the calibration curve.
For a second unspiked sample immediately analyzed after sample #1, (that is, the
addition calibration), the concentration is calculated using the obtained intensity
directly:
Cunspiked sample #N = Iunspiked sample #N / slope
where:
SDconc, unspiked: standard deviation of the concentration
Cunspiked: mean analyte concentration
SDnet intensity: standard deviation of the net intensity
Inet: mean net intensity
Sheet 3 of the included file Calibration Validation_for Std Add.xls compares the
calculation of the Concentration Standard Deviation between the Syngistix
software and Excel for Unspiked Sample #1.
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 399
Software Reference Guide
400 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Calibration Validation for Standard Addition
(DATA for Rh 103 SIMPLE LINEAR) (DATA for Rh 103 WEIGHTED LINEAR) (DATA for Rh 103 LINEAR Thru ZERO)
Std NumbeStd Conc. Corrrected InteW ELAN SW Std NumbeStd Conc. Corrected Intensity W WX WY ELAN SW Std NumbeStd Conc. Corrected Intensity ELAN SW
STD-1 1 42212.403 1 STD-1 1 42212.403 1.0000 1 42212.4 STD-1 1 42212.403
STD-2 10 399812.175 1 STD-2 10 399812.175 0.0100 0.100 3998.122 STD-2 10 399812.175
STD-3 20 827438.586 1 STD-3 20 827438.586 0.0025 0.0500 2068.596 STD-3 20 827438.586
C 1 C 1 C 2
Sigma B(inter) 11259.24102 11259.2 Sigma B(inter) 1376.753049 1376.75 Sigma B(inter) 0 0
Sigma A(slope) 871.2662284 871.266 Sigma A(slope) 799.8212446 799.821 Sigma A(slope) 394.5425471 394.543
Sheet 2: Std Add_IS_noIS
Calibration Type : STANDARD ADDITION Calibration Type : STANDARD ADDITION (Blk Sub. after Internal Std) for unspiked Sample #1
(DATA for Pb 208 SIMPLE LINEAR) (DATA for Pb 208 SIMPLE LINEAR)
Std NumbeStd Conc. Corrected Intensity W ELAN SW Std Number Std Conc. Corrected Intensit W ELAN SW
((STD-1)) 0 17914.1220743356 1 ((STD-1)) 0 0.1840311119 1.0000000000
STD-2 5 39813.4189591270 1 STD-2 5 0.4471350994 1.0000000000
STD-3 10 80417.0926497383 1 STD-3 10 0.8667962958 1.0000000000
Meas. Int. Mean Meas. Int. RSD Meas. Int. Mean Meas. Int. RSD
Rh 103 88316.20242741710 1.70397357750 88316.20242741710 1.70397357754
Pb 208 713.35116511290 4.28288639430 713.35116511293 4.28288639434
Rep Element Mass Measured Intensity Blank Intensity Net Intensity Net Intensity
UNSPIKED SAMPLE #1
1
Rh 103 99366.38102235590 88316.20242741710 11050.17859493880 11050.17859493880
Pb 208 18185.56765773140 713.35116511290 17472.21649261850 17472.21649261850
2
Rh 103 95079.35349491070 88316.20242741710 6763.15106749360 6763.15106749360
Pb 208 18826.39699415670 713.35116511290 18113.04582904370 18113.04582904380
3
Rh 103 96621.64974342930 88316.20242741710 8305.44731601220 8305.44731601220
Pb 208 18870.45506645750 713.35116511290 18157.10390134460 18157.10390134460
Intensities Meas. Int. Mean Meas. Int. RSD Blank Intensity Blank Int. RSD Meas. Int. Mean Meas. Int. RSD Blank Intensity Blank Int. RSD
Rh 103 97022.46142023200 2.23807662030 88316.20242741710 1.70397357750 97022.46142023200 2.23807662026 88316.20242741710 1.70397357754
Pb 208 18627.47323944850 2.05790097000 713.35116511290 4.28288639430 18627.47323944850 2.05790097003 713.35116511290 4.28288639434
Concentration
Net Int. Mean Conc. Mean Conc. SD Conc. RSD Net Int. Mean Conc. Mean Conc. SD Conc. RSD
Rh 103 8706.25899281490 -15.75163297460 -3.92863100910 24.94110302980 8706.25899281490 not calc. not calc not calc
Pb 208 17914.12207433560 2.36736363150 0.05065797930 2.13984783010 17914.12207433560 2.36736363131 0.05065797930 2.13984783013
STD-2
1 **(See Sheet 2 for Concentration Calculation)
Rh 103 89318.35171127300 88316.20242741710 1002.14928385590
Pb 208 39875.57458830370 713.35116511290 39162.22342319080
2
Rh 103 88817.24196834260 88316.20242741710 501.03954092560
Pb 208 39791.34020430250 713.35116511290 39077.98903918960
3
Rh 103 88954.08947612970 88316.20242741710 637.88704871270
Pb 208 41913.39558011370 713.35116511290 41200.04441500070
STD-3
1
Rh 103 93752.62828359060 88316.20242741710 5436.42585617350
Pb 208 82324.52494613220 713.35116511290 81611.17378101930
2
Rh 103 93587.55118824800 88316.20242741710 5271.34876083090
Pb 208 81662.74290040330 713.35116511290 80949.39173529040
3
Rh 103 90858.01536580880 88316.20242741710 2541.81293839180
Pb 208 79404.06359801810 713.35116511290 78690.71243290510
Calibration Validation with Internal Standard
Reported by
Reported by Syngistix Software Calculated by MS Excel Syngistix
401
Software Reference Guide
402 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Index
Equations Groups
correction factor 284 adding 86
isotope dilution 156 assigning users 87
isotope ratio 158 defining in Enhanced Security 85
sample concentration 156 deleting in Enhanced Security 88
ES Setup
logging in 77
ES Tools
opening 98
overview 98 H
running as administrator 99
Help system
E-Signatures 94
using 16
E-signatures 96
Excel
exporting reports 276
Exporting
current sample reports 288
Exporting reports 276
I
External Read Trigger dialog box 192, Icons 24
193 Ignition
External standards 169 unsuccessful 46
Instrument
diagnostics 55, 68
Instrument window 53
Maintenance tab 56
F Torch Position tab 58
maintenance 56
Faults monitoring 53
reference 305 shutdown 47
FIAS sampling system 207 starting 44
File History 105 starting after extended shutdown 45
Files supported instruments 19
archiving in ES 110 torch position 58
comparing in ES 109 Instrument Diagnostics panel 68
Elemental Library.xml 31 Instrument Parameters panel 149
enabling auditing of deletion in ES 90 Intensity vs Time reports 287
history in ES 105 Interactive window 295
locations 30 Interferences
required for analysis 240 correcting 197, 200
restoring 112
correction examples 202
reviewing in ES 105, 107 dissolved solids limitations 199
sample file format 241 identifying 200
version control in ES 105, 107
isobaric overlaps 198
Folders matrix 197
setting current 23 matrix solvent-induced polyatomic
ions 199
matrix-induced polyatomic ions 199
plasma-induced polyatomic ions 198
spectral 198
G Internal standards 174
Graphs Internal Standards tab 271
Graphics Display toolbar 289 blanks and normalization factors 272
Interactive window 295 Isobaric overlaps 198
Realtime window 291 Isotope dilution 156
working with 289 Isotope ratio methods 158
equation 158
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 403
Software Reference Guide
Modes
application 154
K DRC 154
example 154
KED mode 153 KED 153
Standard 153
Multiple software sessions 31
L
LogBook 147, 148
Instrument Parameters panel 149 N
Statistics panel 152 Net Intensities tab 267
Net intensity reports 263
NexION Audit Trail 100
Normalization Factors
in reports 272
M Notes
Maintenance 56 Notes tab 196
maintenance reminders 56
Manual Adjust tab 137
Mass calibration 144
Mass Calibration window 145
Mass discrimination 160 O
Matrix interferences 197 Online help 16
Matrix solvent-induced polyatomic Optimization Parameters
ions 199 unlinked 137
Matrix-induced polyatomic ions 199 Optimizations
Measurement Status dialog box 250 .DAC parameters 137
Menus scheduling 48
icons 24 SmartTune function 115
Syngistix ball menu 23 when to perform 123
Method panel Options
Periodic table 182 Options dialog box tabs 34
Methods 155 Options dialog box 34
analytical techniques 155
autodilution 157
data only 159
External Read Trigger dialog box 192
isotope dilution 156
isotope ratio 158
P
Method panel 175 Peak hop acquisition 160
Calibration tab 188 Peaks 164
Equation tab 185 spectral peak processing 164
External Read Trigger dialog Performance Checks
box 192 LogBook 147
Notes tab 196 Periodic table 182
Processing tab 183 Periodic Table panel 182
Report tab 194 Plasma-induced polyatomic ions 198
Sampling tab 191 Processing
Timing tab 176 data 164
Method panel tabs 175 signal profile processing 165
spectral peak processing 164
quantitative analysis 155
TotalQuant method 155 Profiles
signal profile processing 165
404 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Index
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 405
Software Reference Guide
Software reference
Calibration View window 304
S Conditions panel 132
Dataset panel 257
Sample addition calibration 173 FIAS sampling system setup 208
Sample concentration 156 Graphics Display toolbar 289
sample concentration equation 156 Instrument window 53
Sample Internal Standards tab Interactive window 295
quality control 228 LogBook panel 148
Sample tab Mass Calibration window 145
quality control 229 Measurement Status dialog box 250
Sample units 190 Method panel 175
Samples Quality Control tabs 216
analyzing 239 Realtime window 291
creating a sample file with a text Report panels 261
editor 241 Review Files dialog box 50
Sample window 244 Sample window 244
Batch tab 245 Scheduler screen 49
Manual tab 244 SmartTune window 127
Sampling tab 191 Software sessions
multiple 31
Sampling Devices 192
external 192 Spectral interferences 198
FIAS 192 Spectral peak processing 164
Sampling systems 207 Spectral scanning 160
checks 47 Spike tab
FIAS 207 quality control 231
Scanning acquisition 161 Spike Tables tab
Scheduler function 48 quality control 235
Scheduler screen 49 Standard addition calibration 169
Security Standard mode 153
introduction to Enhanced Security 73 Standards
Security Audit Trail 103 external 169
Service internal 174
diagnostics 68 Starting
Sessions 31 after extended shutdown 45
instrument 44
Set Current Project Folder 23
software 44
Shutdown
Statistics panel 152
instrument 47
software 48 Status
reference 305
Shutdown values 263
Status bar 23
Signal
steady state 162 Steady state 162
Signal profile 162 Summary reports 287
Signal profile processing 165 Syngistix ball menu 23
Signature dialog box 94 Syngistix LogBook 148
Signature Points Syngistix Options dialog box 34
defining 94 System View panel 54
Signatures 96 Diagnostics tab 55
SmartTune function 115
default files 117
software reference 127
understanding 117
Software T
exiting 48 Templates
file locations 30 report templates 282
starting 44 Time vs Intensity reports 287
406 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.
Timestamp
displaying in reports 278
Torch Position 58
Torches
checks 47
TotalQuant methods 155
calibration 171
Transient signal processing 163
Troubleshooting
ignition unsuccessful 46
Tuning
SmartTune function 115
U
UCT instruments
analysis order 241
Unfactored Concentrations tab 270, 273, 275
Units
concentrations 190
sample units 190
Unlinked profiles 137
Users
defining reasons for operations in ES 94
deleting in Enhanced Security 83
setting up in Enhanced Security 80
unlocking in Enhanced Security 83
Utilities
ES Setup 77
ES Tools 98
W
Workspaces
Review Files dialog box 50
Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited. 407
Software Reference Guide
408 Document controlled under U.S. Export Administration regulations. Unauthorized copy, disclosure, or use prohibited.