Professional Documents
Culture Documents
CC Lecture Reviewer - ppt3 (QC Part2)
CC Lecture Reviewer - ppt3 (QC Part2)
2
DIAGNOSTIC TEST CUMULATIVE SUM GRAPH (CUSUM)
QC Result
Patient result
STEP 1:
Most widey used QC chart in the clinical Establishing Mean Target Values Using Parallel
laboratory Testing
Allows the laboratorians to apply multiple rules
without the aid of a computer’ it is a graphic Parallel Testing is the name given to running the
representation of the acceptable limits of current QC lot and a new QC lot in tandem.
variation in the results of an analytical method. • To perform Parallel Testing:
It easily identifies random and systematic errors 1. The new QC lot should then be analysed 2-3
times daily for approximately 10 days,
OPERATION OF A QC SYTSTEM ensuring that at least 20 data points are
The QC program can be thought of as a three-stage generated.
process: 2. Review the data generated to ensure that
1. Establishing allowable statistical limits of assay precision (CV) is lower than the CV
variation for each analytic method obtained in the manufacturer’s
2. Using these limits as criteria for evaluating the specifications.
QC data generated for each test 3. Calculate the Mean and Standard Deviation
3. Taking action to remedy errors when indicated (SD) for each analyte.
a) Finding the cause(s) of error
b) Taking corrective action
c) Reanalyzing control and patient data STEP 2:
Establishing Historical CV (CVh)
Establishment of a Statistical Quality Control
New instrument or with new lots of control material: • CVh
control material must be analyzed for 20 days. the historical instrument precision for that
Exceptions: method measured as a percent
Machines that are highly precise (%CV = SD/Mean X 100)
(CV < 1%), best established over several lots (3 or more)
Eg. blood gases: 5 days is adequate average CV or more preferable
the highest acceptable CV over that period.
If you have no previous CV data, then use the
current CV as the CVh basis increase slightly
(~1- 2%) in order to establish QC ranges that
are not too narrow.
The calculated CVh should not exceed the
instrument-method CV provided by the
manufacturer(3).
The CVh should not exceed the average
assay-specific CV provided by EQA schemes.
EQA schemes (e.g. RIQAS)
QUALITY CONTROL P.2
STEP 3:
Establishing QC Ranges using CVh
• Parallel testing
• Calculate CV
• Calculate the SD
• Assigning Reference range
Shift
• Abrupt changes in the control mean
• Shifts in QC data represent a sudden and
dramatic positive or negative change in test
system performance
OUTLIERS
• Are control values that are far from the main set
of values
• Caused by systematic and random errors
• They are highly deviating values This is a warning rule that is violated when a
single control observation is outside the ±2s
limits.
This rule merely warns that random error or
systematic error may be present in the test
system.
The relationship between this value and other
control results within the current and previous
analytical runs must be examined.
If no relationship can be found and no source of
error can be identified, it must be assumed that a
single control value outside the ±2s limits is an
acceptable random error.
Random Error Patient results can be reported
• Is any deviation away from an expected result.
• There is acceptable (or expected) random error as
defined and quantified by standard deviation.
• There is unacceptable (unexpected) random error
that is any data point outside the expected
population of data (e.g., a data point outside the
±3s limits).
QUALITY CONTROL P.2