Operating Instructions
Cerabar PMC51B
Process pressure measurement
HART
Cerabar PMC51B HART
Serial number
2. www.endress.com/deviceviewer Endress+Hauser
Operations App
3.
A0023555
• Make sure the document is stored in a safe place such that it is always available when
working on or with the device
• Avoid danger to individuals or the facility: read the "Basic safety instructions" section
carefully, as well as all other safety instructions in the document that are specific to
working procedures
The manufacturer reserves the right to modify technical data without prior notice. The
Endress+Hauser sales organization will supply you with current information and updates
to these instructions.
2 Endress+Hauser
Cerabar PMC51B HART Table of contents
Table of contents
identification . . . . . . . . . . . . . . . . . . . . . 13
11 Diagnostic and troubleshooting . . . . 45
4.1 Incoming acceptance . . . . . . . . . . . . . . . . . . . 13
11.1 General troubleshooting . . . . . . . . . . . . . . . . . 45
4.2 Product identification . . . . . . . . . . . . . . . . . . . 13
11.2 Diagnostic information on local display . . . . . . 49
4.3 Storage and transport . . . . . . . . . . . . . . . . . . 14
11.3 Diagnostic event in the operating tool . . . . . . . 50
11.4 Adapting the diagnostic information . . . . . . . 50
5 Installation . . . . . . . . . . . . . . . . . . . . . . . 15 11.5 Queued diagnostic messages . . . . . . . . . . . . . 50
5.1 Installation conditions . . . . . . . . . . . . . . . . . . 15 11.6 Diagnostic list . . . . . . . . . . . . . . . . . . . . . . . . 50
5.2 Mounting the measuring device . . . . . . . . . . . 17 11.7 Event logbook . . . . . . . . . . . . . . . . . . . . . . . . 53
5.3 Post-installation check . . . . . . . . . . . . . . . . . . 22 11.8 Resetting the measuring device . . . . . . . . . . . 54
11.9 Device information . . . . . . . . . . . . . . . . . . . . 54
11.10 Firmware history . . . . . . . . . . . . . . . . . . . . . . 55
6 Electrical connection . . . . . . . . . . . . . . 23
6.1 Connection conditions . . . . . . . . . . . . . . . . . . 23
6.2 Connecting the measuring device . . . . . . . . . . 23 12 Maintenance . . . . . . . . . . . . . . . . . . . . . . 56
6.3 Ensuring the degree of protection . . . . . . . . . 27 12.1 Maintenance tasks . . . . . . . . . . . . . . . . . . . . . 56
6.4 Post-connection check . . . . . . . . . . . . . . . . . . 27
13 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
7 Operation options . . . . . . . . . . . . . . . . . 29 13.1 General notes . . . . . . . . . . . . . . . . . . . . . . . . 57
7.1 Overview of operation options . . . . . . . . . . . . 29 13.2 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . 57
7.2 Operating keys and DIP switches on the 13.3 Replacement . . . . . . . . . . . . . . . . . . . . . . . . . 57
electronic insert . . . . . . . . . . . . . . . . . . . . . . . 29 13.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
7.3 Structure and function of the operating 13.5 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.4 Access to the operating menu via the local 14 Accessories . . . . . . . . . . . . . . . . . . . . . . . 59
display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
14.1 Device-specific accessories . . . . . . . . . . . . . . . 59
7.5 Access to the operating menu via the
14.2 Device Viewer . . . . . . . . . . . . . . . . . . . . . . . . 59
operating tool . . . . . . . . . . . . . . . . . . . . . . . . 31
15 Technical data . . . . . . . . . . . . . . . . . . . . 60
15.1 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Endress+Hauser 3
Table of contents Cerabar PMC51B HART
15.2 Environment . . . . . . . . . . . . . . . . . . . . . . . . . 62
15.3 Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
4 Endress+Hauser
Cerabar PMC51B HART About this document
1.2 Symbols
This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
serious or fatal injury.
WARNING
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
serious or fatal injury.
CAUTION
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
minor or medium injury.
NOTICE
This symbol contains information on procedures and other facts which do not result in
personal injury.
Ground connection:
Terminal for connection to the grounding system.
Permitted:
Procedures, processes or actions that are permitted.
Forbidden:
Procedures, processes or actions that are forbidden.
Endress+Hauser 5
About this document Cerabar PMC51B HART
Additional information:
Reference to documentation:
Reference to page:
Series of steps: 1. , 2. , 3.
Series of steps: 1. , 2. , 3.
Views: A, B, C, ...
Safety instructions:
Observe the safety instructions contained in the associated Operating Instructions.
6 Endress+Hauser
Cerabar PMC51B HART About this document
p
0
LRL LRV URV URL MWP OPL
A0029505
1 OPL: The OPL (over pressure limit = sensor overload limit) for the measuring device depends on the lowest-
rated element, with regard to pressure, of the selected components, i.e. the process connection has to be taken
into consideration in addition to the measuring cell. Pay attention to the pressure/temperature dependency.
2 MWP: The MWP (maximum working pressure) for the sensors depends on the lowest-rated element, with
regard to pressure, of the selected components, i.e. the process connection has to be taken into consideration
in addition to the measuring cell. Pay attention to the pressure/temperature dependency. The MWP may be
applied at the device for an unlimited period of time. The MWP can be found on the nameplate.
3 The maximum sensor measuring range corresponds to the span between the LRL and URL. This sensor
measuring range is equivalent to the maximum calibratable/adjustable span.
4 The calibrated/adjusted span corresponds to the span between the LRV and URV. Factory setting: 0 to URL.
Other calibrated spans can be ordered as customized spans.
p Pressure
LRL Lower range limit
URL Upper range limit
LRV Lower range value
URV Upper range value
TD Turn down. Example - see the following section.
1=2 3
LRL LRV URV URL
A0029545
1 Calibrated/adjusted span
2 Zero point-based span
3 Upper range limit (URL)
Endress+Hauser 7
About this document Cerabar PMC51B HART
Example:
• Sensor: 10 bar (150 psi)
• Upper range limit (URL) = 10 bar (150 psi)
• Calibrated/adjusted span: 0 to 5 bar (0 to 75 psi)
• Lower range value (LRV) = 0 bar (0 psi)
• Upper range value (URV) = 5 bar (75 psi)
URL
TD =
|URV - LRV|
In this example, the TD is 2:1. This span is based on the zero point.
1.5 Documentation
All available documents can be downloaded using:
• the serial number of the device (see cover page for description) or
• the data matrix code of the device (see cover page for description) or
• the "Downloads" area of the website www.endress.com
8 Endress+Hauser
Cerabar PMC51B HART Basic safety instructions
Endress+Hauser 9
Basic safety instructions Cerabar PMC51B HART
Hazardous area
To eliminate the risk of danger to persons or the facility when the device is used in the
approval-related area (e.g. explosion protection, pressure equipment safety):
‣ Check the nameplate to verify if the device ordered can be put to its intended use in the
approval-related area.
‣ Observe the specifications in the separate supplementary documentation that is an
integral part of these Instructions.
2.7 IT security
Endress+Hauser can only provide a warranty if the device is installed and used as
described in the Operating Instructions. The device is equipped with security mechanisms
to protect it against any inadvertent changes to the device settings. IT security measures in
line with operators' security standards and designed to provide additional protection for
the device and device data transfer must be implemented by the operators themselves.
10 Endress+Hauser
Cerabar PMC51B HART Product description
3 Product description
1 2 3 4
p
A0043088
The ceramic sensor is an oil-free sensor. The process pressure acts directly on the robust
ceramic process isolating diaphragm and deflects it. A pressure-dependent change in
capacitance is measured at the electrodes of the ceramic meter body and the process
isolating diaphragm. The measuring range is determined by the thickness of the ceramic
process isolating diaphragm.
Advantages:
• High overload resistance
• Thanks to ultrapure 99.9 % ceramic
• Extremely high chemical durability
• Resistant to abrasion and corrosion
• High mechanical durability
• Suitable for vacuums
p
h= ρ g
A0038343
h Height (level)
p Pressure
ρ Density of the medium
g Gravitation constant
Endress+Hauser 11
Product description Cerabar PMC51B HART
Advantages:
• Volume and mass measurements in any vessel shape with a freely programmable
characteristic curve
• Has a wide range of uses, e.g.
• For foam formation
• In vessels with agitators or screen fittings
• For liquid gases
12 Endress+Hauser
Cerabar PMC51B HART Incoming acceptance and product identification
DELIVERY NOTE
1=2
A0016870
• Is the order code on the delivery note (1) identical to the order code on the product
sticker (2)?
• Are the goods undamaged?
• Do the data on the nameplate correspond to the order specifications and the delivery
note?
• Is the documentation provided?
• If required (see nameplate): are the Safety Instructions (XA) provided?
If you can answer "no" to any of these questions, please contact Endress+Hauser.
Endress+Hauser 13
Incoming acceptance and product identification Cerabar PMC51B HART
4.2.2 Nameplates
Different nameplates are used depending on the device version.
Order code:
Ext. ord. cd.: 3 Ser. no.:
P: FW:
Cal./Adj.: Dev.Rev.:
MWP: 4 HW: 4 5
Fill Fluid:
Mat.:
1 2
5
Date:
2
1
Ord.cd.:
Ext.ord.cd.: 3 Ser. no.: 5
P:
Cal./Adj.:
5
MWP:
FW:
4
HW: Mat.:
Date:
Fill Fluid:
A0038163
14 Endress+Hauser
Cerabar PMC51B HART Installation
5 Installation
Always firmly tighten the housing cover and the cable entries.
1. Counter-tighten the cable entries.
2. Tighten the coupling nut.
Endress+Hauser 15
Installation Cerabar PMC51B HART
5.1.4 Orientation
NOTICE
Damage to the measuring device!
If a heated measuring device is cooled during a cleaning process (e.g. by cold water), a
vacuum develops for a short time. As a result of this, moisture can enter the sensor via the
pressure compensation element (1).
‣ Mount the measuring device with the pressure compensation element (1) pointing
downwards.
1
1
A0038723
16 Endress+Hauser
Cerabar PMC51B HART Installation
A0038730
1 Measuring device
2 Shutoff device
Mount the measuring device with the shutoff device above the tapping point so that any
condensate can flow into the process.
1
4
2
2
3 1
A0038731
1 Measuring device
2 Shutoff device
3 U-shaped siphon
4 O-shaped siphon
A siphon reduces the temperature to almost the ambient temperature. The defined water
column only causes minimal (negligible) measured errors and minimal (negligible)
thermal effects on the device.
Observe the maximum permitted ambient temperature of the transmitter!
• Ideally, mount the measuring device with the O-shaped siphon below the tapping point
The device may also be mounted above the tapping point
• Fill the siphon with liquid before commissioning
Endress+Hauser 17
Installation Cerabar PMC51B HART
3
A0038732
1 Measuring device
2 Shutoff device
3 Siphon
Mount the measuring device with the shutoff device below or at the same level as the
tapping point.
A0038733
• Always mount the measuring device below the lowest measuring point
• Do not install the measuring device at the following positions:
• In the filling curtain
• In the tank outlet
• In the suction area of a pump
• At a point in the tank which could be affected by pressure pulses from the agitator
• Mount the measuring device downstream from a shutoff device: the calibration and
function check can be carried out more easily
18 Endress+Hauser
Cerabar PMC51B HART Installation
1 2
A0017743
ø42...60 (1.65...2.36)
70 (2.76)
52 (2.05)
86 (3.39)
ø6 (0.24)
A0028493
Ordering information:
• Can be ordered via the Product Configurator
• Can be ordered as a separate accessory, part number 71102216
If the measuring device is ordered with a separate housing, the mounting bracket is
included in the delivery.
When mounting on a pipe, tighten the nuts on the bracket uniformly with a torque of at
least 5 Nm (3.69 lbf ft).
Endress+Hauser 19
Installation Cerabar PMC51B HART
2
1
3
4
5
6
6
7
r ³ 120 (4.72)
A0038728
3.
1.
4 mm
2. 4.
A0038224
20 Endress+Hauser
Cerabar PMC51B HART Installation
LWARNING
Supply voltage switched on!
Risk of electric shock and/or explosion!
‣ Switch off the supply voltage before opening the measuring device.
LCAUTION
When opening the terminal compartment cover of the dual compartment housing,
your finger can get caught between the cover and pressure compensation filter.
‣ Open the cover slowly.
1. If fitted: release the screw of the cover lock for the electronics compartment cover
using the Allen key.
2. Unscrew the electronics compartment cover from the transmitter housing and check
the cover seal.
3. Press the release mechanism and remove the display module.
4. Turn the display module to the desired position: maximum 4 × 90 ° in each direction.
Fit the display module on the electronics compartment in the desired position until it
clicks into place. Screw the electronics compartment cover back onto the transmitter
housing. If fitted: tighten the screw of the cover lock using the Allen key
0.7 Nm (0.52 lbf ft) ±0.2 Nm (0.15 lbf ft).
Housing thread
The thread of the electronics and connection compartment is coated with lubricant
varnish.
Avoid additional lubrication.
1
A0043807
1 Locking screw
Endress+Hauser 21
Installation Cerabar PMC51B HART
NOTICE
The housing cannot be unscrewed fully.
‣ Loosen the external locking screw by a maximum of 1.5 turns. If the screw is
unscrewed too much or completely (beyond the screw anchor point), small parts
(counter disk) can become loose and fall out.
‣ Tighten the securing screw (hex socket 4 mm (0.16 in)) with maximum
3.5 Nm (2.58 lbf ft) ± 0.3 Nm (0.22 lbf ft).
22 Endress+Hauser
Cerabar PMC51B HART Electrical connection
6 Electrical connection
A0045411
LWARNING
Risk of explosion!
‣ Please refer to the separate documentation on applications in hazardous areas for the
safety instructions.
A0043806
Housing thread
The thread of the electronics and connection compartment is coated with lubricant
varnish.
Avoid additional lubrication.
Endress+Hauser 23
Electrical connection Cerabar PMC51B HART
6.2.2 Terminals
• Supply voltage and internal ground terminal: 0.5 to 2.5 mm2 (20 to 14 AWG)
• External ground terminal: 0.5 to 4 mm2 (20 to 12 AWG)
1 2
3
Y + +
I - -
mA
4
A0028908
The HART communication resistor of 250 Ω in the signal line is always necessary in
the case of a low-impedance power supply.
Take the voltage drop into consideration:
Maximum 6 V for a 250 Ω communication resistor
Overvoltage category
Overvoltage category II
24 Endress+Hauser
Cerabar PMC51B HART Electrical connection
6.2.6 Wiring
LWARNING
Supply voltage might be connected!
Risk of electric shock and/or explosion!
‣ If the device is used in hazardous areas, make sure to comply with national standards
and the specifications in the Safety Instructions (XAs). The specified cable gland must
be used.
‣ The supply voltage must match the specifications on the nameplate.
‣ Switch off the supply voltage before connecting the device.
‣ If necessary, the potential matching line can be connected to the outer ground terminal
of the transmitter before the device is connected.
‣ A suitable circuit breaker should be provided for the device in accordance with IEC/EN
61010.
‣ The cables must be adequately insulated, with due consideration given to the supply
voltage and the overvoltage category.
‣ The connecting cables must offer adequate temperature stability, with due
consideration given to the ambient temperature.
‣ Only operate the measuring device with the covers closed.
Connect the device in the following order:
1. Release the cover lock (if provided).
2. Unscrew the cover.
3. Guide the cables into the cable glands or cable entries.
4. Connect the cables.
5. Tighten the cable glands or cable entries so that they are leak-tight. Counter-tighten
the housing entry.
6. Screw the cover securely back onto the connection compartment.
7. If fitted: tighten the screw of the cover lock using the Allen key 0.7 Nm (0.52 lbf ft)
±0.2 Nm (0.15 lbf ft).
3
-
2
1 +
A0042594
Endress+Hauser 25
Electrical connection Cerabar PMC51B HART
1 2
A0045413
1 Cable entry
2 Dummy plug
A0045413
The type of cable entry depends on the device version ordered.
Always route connecting cables downwards so that moisture cannot penetrate the
connection compartment.
If necessary, create a drip loop or use a weather protection cover.
A B – +
2 1
– 1 2 +
BU BN GNYE 3
– +
A0023097
4 3
1 2
A0011175
1 Signal +
2 Not assigned
3 Signal –
4 Ground
26 Endress+Hauser
Cerabar PMC51B HART Electrical connection
A B – +
+ – 8 7 6
1
2 5
3 4
Han7D
+ –
A0041011
NOTICE
Plug M12 and plug HAN7D: incorrect mounting can invalidate the IP protection class!
‣ The degree of protection only applies if the connecting cable used is plugged in and
screwed tight.
‣ The degree of protection only applies if the connecting cable used is specified according
to IP67, NEMA Type 4X.
‣ The IP protection classes are only maintained if the dummy cap is used or the cable is
connected.
Endress+Hauser 27
Electrical connection Cerabar PMC51B HART
28 Endress+Hauser
Cerabar PMC51B HART Operation options
7 Operation options
1
Zero/I
Alarm
2
Span/II
SW Max
off on
Display
43
A0039285
Endress+Hauser 29
Operation options Cerabar PMC51B HART
A0043059
4 Segment display
SmartBlue App
1. Scan the QR code or enter "SmartBlue" in the search field of the App Store or Google
Play.
A0039186
30 Endress+Hauser
Cerabar PMC51B HART Operation options
Prerequisites
System requirements
The SmartBlue App is available for download on a smartphone or tablet.
• iOS devices: iPhone 5S or higher from iOS11; iPad 5th Generation or higher from iOS11;
iPod Touch 6th Generation or higher from iOS11
• Devices with Android: from Android 6.0 and Bluetooth® 4.0
Initial password
The serial number of the device serves as the initial password when the connection is
established for the first time.
It is important to note the following if the Bluetooth module is removed from one
device and installed in another device: the password changed by the user is only
stored in the Bluetooth module and not in the device.
Endress+Hauser 31
System integration Cerabar PMC51B HART
8 System integration
The assignment of the measured values to the device variables can be changed in the
following submenu:
Application → HART output → HART output
In a HART Multidrop loop, only one device may use the analog current value for signal
transmission. For all other devices in the "Loop current mode" parameter, select the
Disable option.
Pressure 0
Scaled variable 1
Sensor temperature 2
Sensor pressure 3
Electronics temperature 4
Terminal current 5
Terminal voltage 6
32 Endress+Hauser
Cerabar PMC51B HART System integration
The device variables can be queried by a HART® master using HART® command 9 or
33.
0 mbar 8
1 bar 7
2 Pa 11
3 kPa 12
4 MPa 237
5 psi 6
6 torr 13
7 atm 14
8 mmH2O 4
10 mH2O 240
10 ftH2O 3
11 inH2O 1
13 mmHg 5
14 inHg 2
15 gf/cm² 9
16 kgf/cm² 10
Endress+Hauser 33
Commissioning Cerabar PMC51B HART
9 Commissioning
LWARNING
The settings of the current output are relevant for safety!
This can result in product overflow.
‣ The setting for the current output depends on the setting in the Assign PV parameter.
‣ After changing the setting for the current output, check the setting for the span (LRV
and URV) and readjust it if necessary!
LWARNING
Process pressure above or below permitted maximum/minimum!
Risk of injury if parts burst! Warning messages are generated if pressure is too high.
‣ If a pressure smaller than the minimum permitted pressure or greater than the
maximum permitted pressure is present at the device, a message is output.
‣ Use the device only within the sensor range limits!
34 Endress+Hauser
Cerabar PMC51B HART Commissioning
2 3 6 8
1 4 5 7
A0036169
1 2 3
A0039148
Endress+Hauser 35
Commissioning Cerabar PMC51B HART
9.6.2 FieldCare
In the "Extras" menu, click "Options".
Set the preferred language for FieldCare in the "Language" section.
36 Endress+Hauser
Cerabar PMC51B HART Commissioning
Endress+Hauser 37
Commissioning Cerabar PMC51B HART
If you cancel the Commissioning wizard before all the necessary parameters have
been entered, the device may be in an undefined state. In such situations, it is
advisable to reset the device to the factory default settings.
1) DeviceCare is available for download at www.software-products.endress.com. To download the software, it is necessary to register in the Endress
+Hauser software portal.
38 Endress+Hauser
Cerabar PMC51B HART Commissioning
I
[mA] B
20 450 mbar
20 mA
A
4 50 mbar
50 450 p
4 mA
[mbar]
A0039009
Adjustment:
1. Enter the pressure value for the 4 mA current via the Lower range value output
parameter (50 mbar (0.75 psi)).
2. Enter the pressure value for the 20 mA current via the Upper range value output
parameter (450 mbar (6.75 psi))
Result: The measuring range is set to 4 to 20 mA.
Endress+Hauser 39
Commissioning Cerabar PMC51B HART
Prerequisites:
• Measured variable in direct proportion to the pressure
• Due to the orientation of the measuring device, there may be pressure shifts in the
measured value (when the vessel is empty or partly filled, the measured value is not
zero)
Perform position adjustment if necessary
V
[l] B
1000 450 mbar
1000 l
A
0 50 mbar
50 450 p
0l
[mbar]
A0039010
A "Pressure value 1" parameter and "Scaled variable value 1" parameter
B "Pressure value 2" parameter and "Scaled variable value 2" parameter
The pressure present is displayed in the operating tool on the same settings page in
the "Pressure" field.
1. Enter the pressure value for the lower calibration point via the Pressure value 1
parameter: 50 mbar (0.75 psi)
Menu path: Application → Sensor → Scaled variable → Pressure value 1
2. Enter the volume value for the lower calibration point via the Scaled variable value
1 parameter: 0 l (0 gal)
Menu path: Application → Sensor → Scaled variable → Scaled variable value 1
3. Enter the pressure value for the upper calibration point via the Pressure value 2
parameter: 450 mbar (6.75 psi)
Menu path: Application → Sensor → Scaled variable → Pressure value 2
4. Enter the volume value for the upper calibration point via the Scaled variable value
2 parameter: 1 000 l (264 gal)
Menu path: Application → Sensor → Scaled variable → Scaled variable value 2
Result: The measuring range is set for 0 to 1 000 l (0 to 264 gal). Only the Scaled variable
value 1 parameter and Scaled variable value 2 parameter are set with this setting. This
setting has no effect on the current output.
40 Endress+Hauser
Cerabar PMC51B HART Commissioning
9.7.4 Linearization
In the following example, the volume in a tank with a conical outlet should be measured in
m3.
Prerequisites:
• Points for linearization table are known
• Level calibration is performed
• Linearization characteristic must continually increase or decrease
LWARNING
The settings of the current output are relevant for safety!
This can result in product overflow.
‣ The setting for the current output depends on the setting in the Assign PV parameter.
‣ After changing the setting for the current output, check the setting for the span (LRV
and URV) and readjust it if necessary!
V I
[m 3 ] [mA]
3.5 20
h V
0 4
[m] 0 3.5 [m 3 ]
0 3.0
A0038910
• Error message F435 "Linearization" and the alarm current appear as long as the
table is being entered and until the table is activated
• The 0% value (= 4 mA) is defined by the smallest point in the table
The 100% value (= 20 mA) is defined by the largest point in the table
• The assignment of the volume/mass values to the current values can be changed
with the Lower range value output parameter and Upper range value output
parameter
Endress+Hauser 41
Commissioning Cerabar PMC51B HART
Zero/I
Span/II
off on
Display
1
A0043441
42 Endress+Hauser
Cerabar PMC51B HART Operation
10 Operation
Navigation
"Application" menu → Measured values
Endress+Hauser 43
Operation Cerabar PMC51B HART
1. 2.
A A
B B
3. 4.
A A
B B
A0045038
A Pressure display
B Pressure input
44 Endress+Hauser
Cerabar PMC51B HART Diagnostic and troubleshooting
Endress+Hauser 45
Diagnostic and troubleshooting Cerabar PMC51B HART
46 Endress+Hauser
Cerabar PMC51B HART Diagnostic and troubleshooting
Endress+Hauser 47
Diagnostic and troubleshooting Cerabar PMC51B HART
Failure behavior current output Defines which current the output assumes in the case of an • Min.
error. • Max.
Min: < 3.6 mA
Max: >21.5 mA
48 Endress+Hauser
Cerabar PMC51B HART Diagnostic and troubleshooting
1 X
A0043759
1 Status signal
2 Status symbol with diagnostic event
Status signals
F
"Failure (F)" option
A device error has occurred. The measured value is no longer valid.
C
"Function check (C)" option
The device is in the service mode (e.g. during a simulation).
S
"Out of specification (S)" option
The device is being operated:
• Outside its technical specifications (e.g. during warmup or cleaning process)
• Outside the configuration carried out by the user (e.g. level outside configured span)
M
"Maintenance required (M)" option
Maintenance required. The measured value is still valid.
Endress+Hauser 49
Diagnostic and troubleshooting Cerabar PMC51B HART
50 Endress+Hauser
Cerabar PMC51B HART Diagnostic and troubleshooting
Diagnostic of sensor
Diagnostic of electronic
Diagnostic of configuration
Endress+Hauser 51
Diagnostic and troubleshooting Cerabar PMC51B HART
Diagnostic of process
52 Endress+Hauser
Cerabar PMC51B HART Diagnostic and troubleshooting
I1089 Power on
4) If operating via FieldCare, the event list can be displayed with the "Event List/HistoROM" FieldCare function.
Endress+Hauser 53
Diagnostic and troubleshooting Cerabar PMC51B HART
I1956 Reset
54 Endress+Hauser
Cerabar PMC51B HART Diagnostic and troubleshooting
Endress+Hauser 55
Maintenance Cerabar PMC51B HART
12 Maintenance
A0043756
56 Endress+Hauser
Cerabar PMC51B HART Repair
13 Repair
13.3 Replacement
LCAUTION
Data upload/download is not permitted if the device is used for safety-related
applications.
‣ After an entire device or an electronics module has been replaced, the parameters can
be downloaded to the device again via the communication interface. For this, the data
must have been uploaded to the PC beforehand using the "FieldCare/DeviceCare"
software.
Endress+Hauser 57
Repair Cerabar PMC51B HART
13.4 Return
The measuring device must be returned in the event of a factory calibration, or if the
wrong device has been ordered or delivered.
As an ISO-certified company and due to legal requirements,
Endress+Hauser is required to follow certain procedures when handling returned products
that have been in contact with a medium. To ensure swift, safe and professional device
returns, please read the return procedures and conditions on the Endress+Hauser website
at www.services.endress.com/return-material
‣ Select country.
The website of the responsible sales organization opens with all of the relevant
information relating to returns.
13.5 Disposal
58 Endress+Hauser
Cerabar PMC51B HART Accessories
14 Accessories
Endress+Hauser 59
Technical data Cerabar PMC51B HART
15 Technical data
15.1 Output
R Lmax
[W]
1065
848
3 R Lmax £ U – 10.5 V
23 mA
10.5 30 35 U
[V]
1
2
A0039232
Damping A damping affects all outputs (output signal, display). Damping can be enabled as follows:
Factory setting: 1 s
Ex connection data See the separate technical documentation (Safety Instructions (XA)) on
www.endress.com/download.
Linearization The device's linearization function allows the user to convert the measured value to any
units of height or volume. User-defined linearization tables of up to 32 value pairs can be
entered if necessary.
60 Endress+Hauser
Cerabar PMC51B HART Technical data
The assignment of the measured values to the device variables can be changed in the
following submenu:
Application → HART output → HART output
In a HART Multidrop loop, only one device may use the analog current value for signal
transmission. For all other devices in the "Loop current mode" parameter, select the
Disable option.
Supported functions
• Burst mode
• Additional transmitter status
• Device locking
Endress+Hauser 61
Technical data Cerabar PMC51B HART
15.2 Environment
Ambient temperature The following values apply up to a process temperature of +85 °C (+185 °F). At higher
range process temperatures, the permitted ambient temperature is reduced.
• Without LCD display:
Standard:–40 to +85 °C (–40 to +185 °F)
• With LCD display: –40 to +85 °C (–40 to +185 °F) with limitations in optical properties
such as display speed and contrast. Can be used without limitations up to
–20 to +60 °C (–4 to +140 °F)
• Separate housing: –20 to +60 °C (–4 to +140 °F)
Hazardous area
• For measuring devices for use in hazardous areas, see Safety Instructions, Installation or
Control Drawing
• Measuring devices with common explosion protection certificates (e.g. ATEX-/ IEC Ex,
etc.) can be used in hazardous areas up to the ambient temperature.
Storage temperature • Without LCD display: –40 to +90 °C (–40 to +194 °F)
• Without LCD display: –40 to +90 °C (–40 to +194 °F); optional –50 °C (–58 °F)
• With LCD display: –40 to +85 °C (–40 to +185 °F)
• Separate housing: –40 to +60 °C (–40 to +140 °F)
With M12 plug, elbowed: –25 to +85 °C (–13 to +185 °F)
Climate class Class 4K4H (air temperature: –20 to +55 °C (–4 to +131 °F), relative humidity: 4 to 100 %)
satisfied as per DIN EN 60721-3-4.
Condensation is possible.
Cable entries
• Gland M20, plastic, IP66/68 TYPE 4X/6P
• Gland M20, brass nickel plated, IP66/68 TYPE 4X/6P
• Gland M20, 316L, IP66/68 TYPE 4X/6P
• Thread M20, IP66/68 TYPE 4X/6P
• Thread G1/2, IP66/68 TYPE 4X/6P
If the G1/2 thread is selected, the device is delivered with an M20 thread as standard
and a G1/2 adapter is included with the delivery, along with the corresponding
documentation
• Thread NPT1/2, IP66/68 TYPE 4X/6P
• Dummy plug transport protection: IP22, TYPE 2
• Plug HAN7D, 90 deg. IP65 NEMA Type 4X
• Plug M12
When housing is closed and connecting cable is plugged in: IP66/67 NEMA Type 4X
When housing is open or connecting cable is not plugged in: IP20, NEMA Type 1
62 Endress+Hauser
Cerabar PMC51B HART Technical data
NOTICE
Plug M12 and plug HAN7D: incorrect mounting can invalidate the IP protection class!
‣ The degree of protection only applies if the connecting cable used is plugged in and
screwed tight.
‣ The degree of protection only applies if the connecting cable used is specified according
to IP67 NEMA Type 4X.
‣ The IP protection classes are only maintained if the dummy cap is used or the cable is
connected.
Process connection and process adapter when using the separate housing
FEP cable
• IP69 (on sensor side)
• IP66 TYPE 4/6P
• IP68 (1.83 mH2O for 24 h) TYPE 4/6P
PE cable
• IP69 (on sensor side)
• IP66 TYPE 4/6P
• IP68 (1.83 mH2O for 24 h) TYPE 4/6P
Endress+Hauser 63
Technical data Cerabar PMC51B HART
15.3 Process
°F Ta °C
212 100
176 80
140 60
104 40
68 20
32 0 °C
-40 -20 0 20 40 70 80 100 Tp
°F
-40 -4 32 68 104 158 176 212
A0043294
The process temperature ranges indicated here refer to the permanent operation of the
measuring device (maximum deviation of 5 °C (41 °F) is permitted)
Pay attention to the process temperature range of the seal. The temperatures indicated
depend on the resistance of the seal to the medium.
• FKM: –25 to +100 °C (–13 to +212 °F)
• FKM, cleaned for oxygen service: –10 to +60 °C (+14 to +140 °F)
• FFKM Perlast G75LT: –20 to +100 °C (–4 to +212 °F)
• FFKM Kalrez 6375: +5 to +100 °C (+41 to +212 °F)
• FFKM Chemraz 505: –10 to +100 °C (+14 to +212 °F)
• EPDM: –40 to +100 °C (–40 to +212 °F)
• HNBR: –25 to +100 °C (–13 to +212 °F)
64 Endress+Hauser
Cerabar PMC51B HART Technical data
Measuring devices with sensors, nominal value < 10 bar (150 psi)
• pmax: over pressure limit (OPL) of the sensor and depending on the process connection
used
• Measuring devices with PVDF thread or PVDF flange:
• Only mount with the enclosed mounting bracket!
• pmax: 15 bar (225 psi)
• Tmax environment: 60 °C (140 °F)
Measuring devices with sensors, nominal value ≥ 10 bar (150 psi)
• pmax: 40 bar (600 psi)
• Tmax environment: 60 °C (140 °F)
Ultrapure gas applications Endress+Hauser also offers measuring devices for special applications, such as for
ultrapure gas, which are cleaned from oil and grease. No special restrictions regarding the
process conditions apply to these measuring devices.
Steam applications and For steam and saturated steam applications: use a measuring device with a metal process
saturated steam membrane, or provide a siphon for temperature isolation when installing.
applications
Endress+Hauser 65
Index Cerabar PMC51B HART
Index
A N
Access authorization to parameters Nameplates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Read access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Write access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 O
Access code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Incorrect input . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
B P
Bluetooth® wireless technology . . . . . . . . . . . . . . . . . 30 Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
PV (HART variable) . . . . . . . . . . . . . . . . . . . . . . . . . . 32
C
CE mark (Declaration of Conformity) . . . . . . . . . . . . . 10 R
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 Read access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Reading measured values . . . . . . . . . . . . . . . . . . . . . . 43
D Repair concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
DD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Requirements concerning the staff . . . . . . . . . . . . . . . . 9
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . 10
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 S
Device description files . . . . . . . . . . . . . . . . . . . . . . . 32 Safety instructions
Device documentation Basic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Supplementary documentation . . . . . . . . . . . . . . . . 8 Service interface (CDI) . . . . . . . . . . . . . . . . . . . . . . . . 35
Device locking, status . . . . . . . . . . . . . . . . . . . . . . . . 43 Settings
Device Viewer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 Adapting the measuring device to the process
Diagnostic event conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
In the operating tool . . . . . . . . . . . . . . . . . . . . . . . 50 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Diagnostic events . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Diagnostic list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Status signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Diagnostic message . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Submenu
Diagnostics Event list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Measured values . . . . . . . . . . . . . . . . . . . . . . . . . 43
Display values SV (HART variable) . . . . . . . . . . . . . . . . . . . . . . . . . . 32
For locking status . . . . . . . . . . . . . . . . . . . . . . . . . 43
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 T
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
E Turning the display module . . . . . . . . . . . . . . . . . . . . 20
Event history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 TV (HART variable) . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Event list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 U
Use of the measuring device
F see Designated use
Filtering the event logbook . . . . . . . . . . . . . . . . . . . . 53 Using measuring devices
FV (HART variable) . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Borderline cases . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Incorrect use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
H
HART integration . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 W
HART protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 Workplace safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
HART variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Write access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
L
Local display
see Diagnostic message
see In alarm condition
M
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
66 Endress+Hauser