Professional Documents
Culture Documents
Patient-Ventilator Dyssynchrony in The Intensive Care Unit A Practical
Patient-Ventilator Dyssynchrony in The Intensive Care Unit A Practical
Abstract
Patient–ventilator dyssynchrony or asynchrony occurs when, for any parameter of respiration, discordance exists
between the patient’s spontaneous effort and the ventilator’s provided support. If not recognised, it may promote
oversedation, prolong the duration of mechanical ventilation, create risk for lung injury, and generally confuse the clinical
picture. Seven forms of dyssynchrony are common: (a) ineffective triggering; (b) autotriggering; (c) inadequate flow; (d)
too much flow; (e) premature cycling; (f) delayed cycling; and (g) peak pressure apnoea. ‘Reverse triggering’ also occurs
and may mimic premature cycling. Correct diagnosis of these phenomena often permits management by simple venti-
lator optimisation rather than by less desirable measures.
Keywords
Synchrony, patient-ventilator synchrony, vent synchrony, vent dyssynchrony, waveform, bucking
Background
diaphragmatic effort that is dissociated from effective
Mechanical ventilation of intubated patients is con- ventilator support can fatigue the respiratory muscles,
trolled by a combination of patient and ventilator- confound ventilator weaning, and even predispose to
determined variables. Except in the case of a complete- lung injury by inducing regional alveolar stress.5 More
ly passive (i.e. deeply sedated, paralysed, or comatose) subjectively, inability by clinicians to understand
patient, effective ventilation requires a close concor- unusual patterns during mechanical ventilation lends
dance between the patient’s effort and the ventilator’s to it a general air of mystery, obscuring its fundamental
contributions. Anomalies in this harmony—where var- nature as a physical, deterministic, and controllable
iables such as the patient’s intrinsic flow of inspiration process.
or desired inspiratory time do not match those offered
by the ventilator—produce the undesirable situation of
Basic concepts of mechanical ventilation
patient–ventilator dyssynchrony.
Does this phenomenon have a patient-relevant Failures of synchrony are most easily recognised and
impact? Outcome data are limited. Significant dyssyn- categorised in the context of a systematic approach
chrony (or asynchrony, an equivalent term) results in towards ventilation as a whole.
‘bucking’ of the ventilator and frank patient discom-
fort, but subtle dyssynchronies are often not obvious,
requiring a well-trained eye or even specialised studies
such as oesophageal manometry to detect. When pre- 1
Adult Critical Care, UConn Health, Farmington, USA
sent, patient distress and agitation may result in 2
Respiratory Therapy Department, UConn Health, Farmington, USA
increased sedation requirements, which are associated 3
Division of Pulmonary, Critical Care, and Sleep Medicine, UConn
with increased delirium and intensive care unit (ICU) Health, Farmington, USA
length of stay.1 However, observational studies suggest
Corresponding author:
that even subtle dyssynchrony is associated with mor- Brandon Oto, Adult Critical Care, UConn Health, 265 Farmington Ave,
tality and an increased duration of mechanical ventila- Farmington, CT 06030, USA.
tion.2–4 This may be due to a variety of effects, as Email: oto.brandon@gmail.com
Oto et al. 87
Most mechanical ventilation in the modern ICU is because the ventilator can control the respiratory pat-
performed using one of several basic modes. Although tern with little interference from the patient. This
nomenclature in mechanical ventilation has become maxim and its corollary hold true for nearly all dys-
highly variable, several modes are both clinically com- synchronies, with a few rare exceptions such as reverse
monplace and relatively standard in terminology. The triggering and ‘peak pressure apnoea’. Another conse-
first, used for full ventilatory support, is assist control quence is that modes of ventilation which directly con-
ventilation with either a volume or pressure target trol more variables of ventilation are generally at
(denoted respectively as volume control ventilation higher risk of producing dyssynchrony than those
(VCV) or pressure control ventilation (PCV)). The which allow more control by the patient; for example,
second is pressure support ventilation (PSV) and is VCV is more likely to yield certain dyssynchronies
used when spontaneous patient effort is encouraged. (such as flow dyssynchrony) than PCV, and PCV is
A variety of other modes now exist, such as synchron-
at higher risk than PSV.
ised intermittent mandatory ventilation, pressure regu-
lated volume control, volume support, and more, but
for the purposes of this discussion are merely amal- The common types of dyssynchrony
gams of existing principles. Several truly novel modal-
ities have also been introduced, usually proprietary to Seven types of dyssynchrony are commonly encoun-
specific ventilators (such as proportional assist ventila- tered. As terminology is inconsistent, these are best
tion and neurally adjusted ventilatory assist), but have recognised by their features rather than by a universal
not yet achieved widespread adoption. Many of the nomenclature.
same principles of invasive mechanical ventilation are
also applicable to non-invasive positive pressure venti- Triggering dyssynchrony
lation, with appropriate modifications and some differ-
ences in nomenclature. These relate to breath initiation.
Disregarding idiosyncratic modes such as airway
pressure release ventilation (APRV) or oscillatory ven- 1. Failure to trigger: The patient is unable to trigger a
tilation, any mechanical breath is defined by three breath, or there is a delay between their effort and
parameters: the resulting breath.
2. Autotriggering: Undesired breaths are repeatedly
• The trigger: the variable that initiates inspiration. delivered due to a spurious trigger.
• The target or limit (terms used interchangeably): the
variable that defines or controls the inspiratory Flow dyssynchrony
phase.
• The cycling parameter: the variable that ends inspi- These relate to the delivery of the breath, in which
ration, after which passive expiration begins. mismatch typically involves flow.
Similarly, four variables are at work within the ven- 3. Inadequate flow: Patient flow demand exceeds
tilator to fill these parameters: delivered inspiratory ventilator-delivered flow, resulting in inadequate
volume, inspiratory pressure, inspiratory flow, and support.
inspiratory time, all of which are closely interrelated. 4. Flow overshoot: Ventilator-delivered flow exceeds the
By combining these features, the standard modes can patient’s flow demand.
be created (Table 1).
With this framework, it becomes more straightfor- Cycling dyssynchrony
ward to understand synchrony, using the following
maxim: These relate to the ending of the breath. Also known as
expiratory dyssynchrony.
Ventilatory dyssynchrony occurs when, for any param-
eter of respiration, discordance exists between the 5. Premature (early) cycling: The ventilator’s set inspi-
patient’s spontaneous effort and the ventilator’s pro- ratory time (“machine I-time”) is shorter than the
vided support. patient’s intrinsic inspiratory time (“neural I-
time”); the delivered breath therefore ends while
An obvious corollary is that dyssynchrony occurs the patient still desires continued flow.
most often in patients with a spontaneous respiratory 6. Delayed cycling: Machine I-time is longer than
drive. Completely apnoeic patients, such as the deeply neural I-time, resulting in a longer breath than the
sedated, are usually straightforward to ventilate, patient desires.
88 Anaesthesia and Intensive Care 49(2)
Autotriggering
Rapid initial expiratory flow
(also known as accidental triggering)
tapers to prolonged expiratory phase
Potential remedies.
• Eliminate the source of artefact: Address leaks or
drain excess fluids (cardiac oscillations usually
cannot be eliminated).
• Reduce trigger sensitivity: Trigger threshold can be
increased gradually until the contributory artefact
no longer triggers breaths.
• Switch to a pressure trigger: If flow triggering is in
use, switching to a pressure trigger may be
Figure 3. Pressure (top) and flow (bottom) scalars showing in attempted.
vivo examples of ineffective triggering. Here the wasted efforts
are mainly visible on the pressure tracing, and the effort that Inadequate flow
finally does trigger is strong enough to result in flow starvation
for the ensuing breath (see Inadequate flow). (also known as flow starvation, work shifting)
Pressure
good flow matching
Waveform manifestations. The pressure scalar should be Figure 6. Three breaths with increasing flow mismatch. Note
these are all patient-triggered breaths (proved by the negative
carefully inspected (Figures 6 and 7). A normal wave- deflection at the start of each inspiration), and hence this is not
form has the appearance of a flat or gently domed an example of poor compliance (positive ‘stress index’). The flow
structure. In the case of flow starvation, the patient is curve is unchanged, because inspiratory flow is fixed in volume
attempting to inspire faster than the flow provided, control ventilation (VCV), regardless of patient factors.
pulling the airway pressure more negatively. In mild
cases, the pressure curve becomes flattened and then
upwardly concave. In severe cases, that concavity variable or very brisk), consider switching to a pres-
drops below the baseline, indicating a negative airway sure mode.17
pressure and outright flow starvation. • In pressure control or pressure support: Shorten rise
This concave waveform may easily be mistaken for time.14
the scooped-out appearance that develops in passive • If patient effort is pathological (i.e. resulting in
breaths when tidal volume exceeds lung compliance; unnecessarily large tidal volumes and minute venti-
this has been described as a ‘stress index greater than lation): Attempt to resolve the cause of increased
1’ and may be associated with volutrauma.16 However, respiratory drive, such as managing pain and dis-
that phenomenon occurs during passive breaths, tress, controlling fever, etc. If this cannot be
whereas flow starvation occurs during active patient achieved, increase sedation.
effort. Although the concave appearance may be virtu-
ally identical in both, the distinction can be easily made Too much flow
by either observing the patient for effort, or by noting
on the ventilator whether the breath was triggered by (also known as pressure overshoot, flow overshoot)
the patient or by the passive time trigger.
Causes. This phenomenon is relatively unusual, as
Potential remedies. many clinicians are more likely to set flow rates too
• In volume control: Increase set inspiratory flow until low than too high. However, if flow on VCV has
the pressure waveform flattens out. If unable to been titrated up to prevent flow starvation, overcom-
match patient flow successfully (e.g. if effort is pensation may sometimes occur. In pressure modes, the
Oto et al. 93
Pressure
0
very high.
Pressure
Very steep (near vertical)
downslope
Pressure
Figure 11. In vivo examples of premature cycling. In the first
breath, continued inspiratory effort is visible at the start of
expiration (single line). In the second, this effort is sufficient to
trigger a stacked breath (double line). Note the prolonged 0
5. Yoshida T, Fujino Y, Amato MBP, et al. Fifty years of 15. Chiumello D, Pelosi P, Crosi M, et al. The effects of
research in ARDS. Spontaneous breathing during pressurization rate on breathing pattern, work of breath-
mechanical ventilation. risks, mechanisms, and manage- ing, gas exchange and patient comfort in pressure sup-
ment. Am J Respir Crit Care Med 2017; 195: 985–992. port ventilation. Eur Respir J 2001; 18: 107–114.
6. Aslanian P, El Atrous S, Isabey D, et al. Effects of flow 16. Sun XM, Chen GQ, Chen K, et al. Stress index can be
triggering on breathing effort during partial ventilatory accurately and reliably assessed by visually inspecting
support. Am J Respir Crit Care Med 1998; 157: 135–143. ventilator waveforms. Respir Care 2018; 63: 1094–1101.
7. Sassoon CS, Del Rosario N, Fei R, et al. Influence of 17. Kallet RH, Campbell AR, Alonso JA, et al. The effects of
pressure- and flow-triggered synchronous intermittent pressure control versus volume control assisted ventila-
mandatory ventilation on inspiratory muscle work. Crit tion on patient work of breathing in acute lung injury and
Care Med 1994; 22: 1933–1941. acute respiratory distress syndrome. Respir Care 2000;
8. Fernandez R, Mendez M and Younes M. Effect of ven- 45: 1085–1096.
tilator flow rate on respiratory timing in normal humans. 18. Akoumianaki E, Lyazidi A, Rey N, et al. Mechanical
Am J Respir Crit Care Med 1999; 159: 710–719. ventilation-induced reverse-triggered breaths: a frequent-
9. Thille AW, Cabello B, Galia F, et al. Reduction of ly unrecognized form of neuromechanical coupling. Chest
2013; 143: 927–938.
patient–ventilator asynchrony by reducing tidal volume
19. Delisle S, Charbonney E, Albert M, et al. Patient–venti-
during pressure-support ventilation. Intensive Care Med
lator asynchrony due to reverse triggering occurring in
2008; 34: 1477–1486.
brain-dead patients: clinical implications and physiologi-
10. Nava S, Bruschi C, Rubini F, et al. Respiratory response
cal meaning. Am J Respir Crit Care Med 2016; 194:
and inspiratory effort during pressure support ventilation
1166–1168.
in COPD patients. Intensive Care Med 1995; 21: 871–879.
20. Dianti J, Bertoni M and Goligher EC. Monitoring
11. Imanaka H, Nishimura M, Takeuchi M, et al.
patient–ventilator interaction by an end-expiratory occlu-
Autotriggering caused by cardiogenic oscillation during sion maneuver. Intensive Care Med 2020; 46: 2338–2341.
flow-triggered mechanical ventilation. Crit Care Med 21. Murias G, de Haro C and Blanch L. Does this ventilated
2000; 28: 402–407. patient have asynchronies? Recognizing reverse triggering
12. Gilstrap D and MacIntyre N. Patient–ventilator interac- and entrainment at the bedside. Intensive Care Med 2016;
tions. Implications for clinical management. Am J Respir 42: 1058–1061.
Crit Care Med 2013; 188: 1058–1068. 22. Epstein SK. Optimizing patient–ventilator synchrony.
13. Schwarz G, Errath M, Arguelles Delgado P, et al. Semin Respir Crit Care Med 2001; 22: 137–152.
Ventilator autotriggering: an underestimated phenome- 23. Tassaux D, Gainnier M, Battisti A, et al. Impact of expi-
non in the determination of brain death. Anaesthesist ratory trigger setting on delayed cycling and inspiratory
2019; 68: 171–176. muscle workload. Am J Respir Crit Care Med 2005; 172:
14. Chiumello D, Pelosi P, Taccone P, et al. Effect of differ- 1283–1289.
ent inspiratory rise time and cycling off criteria during 24. Oto B. Peak pressure apnea: an under-recognized, high-
pressure support ventilation in patients recovering from risk form of ventilator asynchrony. Heart Lung 2020; 49:
acute lung injury. Crit Care Med 2003; 31: 2604–2610. 426.