User Manual
D968.131.00/05 en
RUBY (T40072)
Revision 2021-12
2 D968.131.00/05 en
RUBY (T40072)
Table of Contents
Table of Contents
Table of Contents........................................................................................................... 3
List of Figures.................................................................................................................5
List of Abbreviations......................................................................................................6
Important Information......................................................7
1 General Information.................................................................................................... 7
2 Safety Information.......................................................................................................9
Operating Manual...........................................................13
3 Intended Use..............................................................................................................13
3.1 RUBY Sets........................................................................................................................................... 15
4 Product Description.................................................................................................. 16
4.1 Phantom Body...................................................................................................................................... 17
4.2 Inserts................................................................................................................................................... 22
4.3 Dummy Plug.........................................................................................................................................29
4.4 Detector Holder.................................................................................................................................... 29
4.5 RUBY Tilted Base................................................................................................................................ 30
4.6 RUBY Carrying Case........................................................................................................................... 31
5 Application................................................................................................................. 32
5.1 Introducing an Insert............................................................................................................................ 32
5.2 Inserting a Detector..............................................................................................................................33
5.3 Positioning the RUBY Phantom on the RUBY Tilted Base.................................................................. 34
5.4 Loading a Film..................................................................................................................................... 35
5.5 Using the RUBY Set............................................................................................................................ 36
Technical Manual........................................................... 39
6 Technical Specifications...........................................................................................39
7 Ordering Information................................................................................................ 41
7.1 Suitable Detectors................................................................................................................................ 45
D968.131.00/05 en 3
RUBY (T40072)
Table of Contents
Service Manual...............................................................47
8 Cleaning......................................................................................................................47
9 Preventive Maintenance........................................................................................... 48
Literature........................................................................................................................49
Index...............................................................................................................................50
4 D968.131.00/05 en
RUBY (T40072)
List of Figures
List of Figures
Figure 1: Definition of the patient environment (example according to EN 60601-1 (IEC 60601-1))............. 11
Figure 2: Modules of RUBY product T40072................................................................................................. 16
Figure 3: RUBY phantom shown in the reference coordinate system according to IEC 61217..................... 17
Figure 4: Detail of the RUBY phantom - engraved lines............................................................................... 18
Figure 5: RUBY head phantom shown in the reference coordinate system according to IEC 61217............ 19
Figure 6: Detail of the RUBY head phantom - engraved lines...................................................................... 20
Figure 7: Stand................................................................................................................................................20
Figure 8: RUBY head phantom with stand.................................................................................................... 20
Figure 9: RUBY carrying case for head phantom.......................................................................................... 21
Figure 10: RUBY insert Patient QA - detector T40072.1.100.......................................................................... 23
Figure 11: RUBY insert Patient QA - detector T40072.1.100 (rear view)........................................................ 23
Figure 12: RUBY insert Patient QA - film T40072.1.400................................................................................. 24
Figure 13: RUBY insert Patient QA - film T40072.1.400 (rear view)............................................................... 24
Figure 14: RUBY insert Linac QA T40072.1.200............................................................................................. 25
Figure 15: RUBY insert Linac QA T40072.1.200 (rear view)...........................................................................25
Figure 16: RUBY insert System QA T40072.1.300..........................................................................................26
Figure 17: RUBY insert System QA T40072.1.300 (rear view)....................................................................... 26
Figure 18: MR cavities...................................................................................................................................... 26
Figure 19: CT cavities "Lung", "Brain", "Bone"................................................................................................ 27
Figure 20: RUBY insert System QA - MultiMet T40072.1.500.........................................................................28
Figure 21: RUBY insert System QA - MultiMet T40072.1.500 (rear view).......................................................28
Figure 22: Inhomogeneities...............................................................................................................................28
Figure 23: RUBY tilted base.............................................................................................................................30
Figure 24: RUBY carrying case........................................................................................................................31
Figure 25: RUBY insert Patient QA - detector with detector holder (without detector).................................... 33
Figure 26: Aligning the RUBY phantom on the RUBY tilted base................................................................... 34
Figure 27: Components of the RUBY insert Patient QA - film.........................................................................35
D968.131.00/05 en 5
RUBY (T40072)
List of Abbreviations
List of Abbreviations
MV megavolt
6 D968.131.00/05 en
RUBY (T40072)
Important Information - General Information
Important Information
1 General Information
– The product bears the CE-mark "CE" in accor- – PTW-Freiburg only considers itself responsi-
dance with the Council Directive 93/42/EEC ble for the safety, reliability, and performance
about Medical Devices and fulfills the essential of the product if the assembly, extension, read-
requirements of Annex I of this directive. The justment, modification, or repair is carried out
product is a class I device without a measuring by PTW-Freiburg or by persons authorized by
function (MDD). PTW-Freiburg, and if the product is used in
compliance with the technical documentation.
– The user manual is an integral part of the prod-
uct. It should always be kept near the product. – In case of any questions concerning the ser-
Observance of the manual is a prerequisite for vice, support, or warranty, please contact your
proper product performance and correct oper- supplier.
ation.
– This manual is in conformity with the product
– Operator safety, specified measuring accu- specifications and all applicable safety stan-
racy, and interference-free operation can dards that are valid as at the printing date. All
only be guaranteed if original products and rights are reserved for devices, circuits, tech-
parts are used. Furthermore, only the acces- niques, software, and names as referred to in
sories listed in this manual are approved by the manual.
PTW-Freiburg and only they or other acces-
sories whose use has been expressly permit- – PTW-Freiburg reserves the right to make mod-
ted by PTW-Freiburg may be used in con- ifications.
junction with the product. Safe operation and Please contact PTW or your local representa-
proper product performance are not guaran- tive for the most current information concern-
teed if accessories or consumables from other ing the products of interest.
manufacturers are used.
– No part of the technical documentation may
– PTW-Freiburg cannot be held liable for dam- be reproduced without written permission from
age resulting from the use of accessories PTW-Freiburg.
or consumables from other manufacturers or
when the user ignores the instructions and
information given in this manual.
D968.131.00/05 en 7
RUBY (T40072)
Important Information - General Information
8 D968.131.00/05 en
RUBY (T40072)
Important Information - Safety Information
2 Safety Information
NOTICE
Indicates a potential hazard. If not avoided, the
hazard can result in product damage.
HINT
Provides application tips or other useful infor-
mation to ensure that you get the most from
your equipment.
D968.131.00/05 en 9
RUBY (T40072)
Important Information - Safety Information
WARNING
Application for uses other than those stated in
the user manual.
10 D968.131.00/05 en
RUBY (T40072)
Important Information - Safety Information
CAUTION
Improper handling.
CAUTION
Hazards originating from other system compo-
nents.
D968.131.00/05 en 11
RUBY (T40072)
Important Information
Symbol Description
WARNING
Refer to the user manual! Improper handling.
Serial number
For a description of additional symbols, if Non-clinical testing has demonstrated that the
applicable, please refer to section 4 "Product product is "MR Conditional". The product may
Description". be safely used in an MR system under the fol-
lowing conditions.
Failure to observe may result in injuries and/
or loss of product functionality.
Static magnetic Do not exceed
field 3 T (30,000 Gs).
Spatial field gra- Do not exceed
dient 25 T/m (2,500 Gs/cm).
0027
12 D968.131.00/05 en
RUBY (T40072)
Operating Manual - Intended Use
Operating Manual
3 Intended Use
The product is intended for patient QA measure- The phantom is used for the following tests:
ment and for periodic quality assurance according
– Quality assurance tests of the radiotherapy
to the QA plan of the responsible medical physi-
system
cist.
– Verification of imaging systems at radiother-
apy systems
Product Description
– End-to-End tests
The phantom has a modular design. The phantom – Plan verification
consists of two different phantom bodies, several
inserts and optional components for the inserts.
Each insert can be combined with each phantom Contraindications
body.
The phantom is intended for QA purposes and
must not be used while a patient is present.
Operating Principle
It must not be used in diagnostic radiology.
Interchangeable inserts are placed in a geometric
or an anthropomorphic phantom body. The inserts The resulting measurement data is intended for
serve for housing a radiation detector or a film verification of a patient plan and must not be used
or serve for housing of different materials which to control the radiotherapy system by importing
could be analyzed with imaging systems. the data into the therapy patient plan.
D968.131.00/05 en 13
RUBY (T40072)
Operating Manual - Intended Use
Intended User
14 D968.131.00/05 en
RUBY (T40072)
Operating Manual - Intended Use
RUBY head set MultiMet L981654 consists For further information on application of the dif-
of the following main components (and further ferent RUBY Sets, refer to section 5.5 "Using the
accessories): RUBY Set".
– RUBY head phantom T40072.1.800
D968.131.00/05 en 15
RUBY (T40072)
Operating Manual - Product Description
4 Product Description
The RUBY product has a modular design. It is The product is composed of the following mod-
used to perform various tests for quality assur- ules:
ance in radiation therapy.
1 2
3 4
3 5 6
16 D968.131.00/05 en
RUBY (T40072)
Operating Manual - Product Description
4.1 Phantom Body The different inserts can be introduced into the
RUBY phantom through an aperture. Its under-
side has been designed in such a way that
4.1.1 RUBY Phantom the RUBY phantom stands firmly on conven-
tional carbon fiber treatment couches of linear
The RUBY phantom T40072.1.001 is one of the accelerators. This means that the inserts can be
main modules of the RUBY product. Its main exchanged without any significant change in the
sections (coronal, sagittal, transversal) represent position of the RUBY phantom.
an octagon. The shape enables irradiation from
almost all directions and rotational shifting of the
RUBY phantom to position different planes at the
isocenter.
Figure 3: RUBY phantom shown in the reference coordinate system according to IEC 61217
D968.131.00/05 en 17
RUBY (T40072)
Operating Manual - Product Description
2 Black line
indicates the phantom center
4 3 Red line
indicates a defined translational and rota-
3
tional shift in all three spatial directions
– coronal: -12 mm
– rotation (Z-axis) = -1°
– transversal: -10 mm
– rotation (Y-axis) = 2.5°
Figure 4: Detail of the RUBY phantom - – sagittal: 15 mm
engraved lines – rotation (X-axis) = -1.5°
4 CT marker (3 pieces)
18 D968.131.00/05 en
RUBY (T40072)
Operating Manual - Product Description
Figure 5: RUBY head phantom shown in the reference coordinate system according to IEC 61217
D968.131.00/05 en 19
RUBY (T40072)
Operating Manual - Product Description
4.1.2.1 Stand
2 Black line
indicates the phantom center
3 CT marker (3 pieces)
20 D968.131.00/05 en
RUBY (T40072)
Operating Manual - Product Description
2 4
D968.131.00/05 en 21
RUBY (T40072)
Operating Manual - Product Description
4.2 Inserts
Five block-shaped inserts are available that are
introduced into the phantom body from the front.
The inserts are of identical size but their con-
struction differs depending on their function. The
inserts have three engraved CT markers for deter-
mination of the phantom center (see Figure 25).
22 D968.131.00/05 en
RUBY (T40072)
Operating Manual - Product Description
D968.131.00/05 en 23
RUBY (T40072)
Operating Manual - Product Description
24 D968.131.00/05 en
RUBY (T40072)
Operating Manual - Product Description
D968.131.00/05 en 25
RUBY (T40072)
Operating Manual - Product Description
4.2.4 RUBY Insert System QA The RUBY insert System QA has the following
cavities:
The RUBY insert System QA T40072.1.300 is a
– 3 cavities filled with MR-visible liquid ("MRI")
non-homogeneous insert.
HINT
The non-homogeneities (cavities) are cylinders
of different sizes positioned around the detector. The MR cavities contain air bubbles to com-
All cavities are interrupted at different points to pensate for volume expansion.
enable an image registration of the insert. – 3 cavities with CT-visible, tissue-equivalent
material according to ICRU Report No. 44 /
Report No. 46
("Lung", "Brain", "Bone")
26 D968.131.00/05 en
RUBY (T40072)
Operating Manual - Product Description
1 2
3 4 5 6
1 Lung - cavity 1
2 Lung - cavity 2
3 Bone - cavity 1
4 Bone - cavity 2
5 Brain - cavity 2
6 Brain - cavity 1
D968.131.00/05 en 27
RUBY (T40072)
Operating Manual - Product Description
28 D968.131.00/05 en
RUBY (T40072)
Operating Manual - Product Description
4.3 Dummy Plug The following five detector holders are available
(central position):
For the creation of a treatment planning CT, a
dummy plug is introduced in the appropriate insert – Detector holder T40072.1.110
in place of a detector. Its purpose is to provide a for microDiamond, Tx60019
homogeneous filling in the bore hole of the insert.
– Detector holder T40072.1.120
For the RUBY insert Patient QA - detector for PinPoint 3D chamber 0.016 cm³, Tx31022
and the RUBY insert System QA, dummy plug
T40072.1.150 is used. – Detector holder T40072.1.130
for Semiflex 3D chamber 0.07 cm³, Tx31021
For the RUBY insert System QA - Mul-
tiMet, dummy plugs T40072.1.150 (central), – Detector holder T40072.1.140
T40072.1.550 (caudal), and T40072.1.555 (cra- for microSilicon, Tx60023
nial) are used.
– Detector holder T40072.1.160
for Semiflex chamber 0.125 cm³, Tx31010
4.4 Detector Holder In addition, the following ten detector holders are
available for the RUBY insert System QA - Multi-
The RUBY product includes 15 different detec-
Met (caudal and cranial position):
tor holders labeled at the front with the respec-
tive part numbers of the detectors. One detector
– Detector holder T40072.1.510
holder is required for the RUBY insert Patient QA
caudal for microDiamond, Tx60019
- detector and the RUBY insert System QA, three
detector holders are required for the RUBY insert – Detector holder T40072.1.515
System QA - MultiMet. cranial for microDiamond, Tx60019
The detector holder is screwed into the bore hole – Detector holder T40072.1.520
of the insert from the back. The corresponding caudal for PinPoint 3D chamber 0.016 cm³,
detector is introduced into the insert from the front Tx31022
and enters the detector holder in the middle of the
insert. – Detector holder T40072.1.525
cranial for PinPoint 3D chamber 0.016 cm³,
The holder design ensures that the detector refer- Tx31022
ence point is positioned at the center. Besides this
central reference point, there are two additional – Detector holder T40072.1.530
reference points for the RUBY insert System QA - caudal for Semiflex 3D chamber 0.07 cm³,
MultiMet. They are located in opposite directions Tx31021
from the center (caudal and cranial position).
– Detector holder T40072.1.535
A detector lock at the front of the insert prevents cranial for Semiflex 3D chamber 0.07 cm³,
the detector from slipping out. Tx31021
D968.131.00/05 en 29
RUBY (T40072)
Operating Manual - Product Description
30 D968.131.00/05 en
RUBY (T40072)
Operating Manual - Product Description
1 Cut-out for RUBY phantom 4 Cut-out for stand (RUBY head phantom)
2 Cut-out for RUBY head phantom 5 Cut-out for inserts
3 Cut-out for RUBY tilted base 6 Cut-out for detector holders and dummy plug
D968.131.00/05 en 31
RUBY (T40072)
Operating Manual - Application
5 Application
WARNING
Improper handling.
32 D968.131.00/05 en
RUBY (T40072)
Operating Manual - Application
1 2
Figure 25: RUBY insert Patient QA - detector with detector holder (without detector)
D968.131.00/05 en 33
RUBY (T40072)
Operating Manual - Application
• Introduce the respective insert into the phan- – The RUBY phantom is in an upright posi-
tom body from the front and as far as it will go tion, i. e., the rubber feet point downward.
(refer to section 5.1 "Introducing an Insert"). – The rubber feet of the RUBY phantom rest
in the recesses of the RUBY tilted base.
• Select the appropriate detector for the detector
holder. – The vertical, red lines on the RUBY phan-
tom continue on the RUBY tilted base.
• Introduce the detector holder into the insert
from the back as far as it will go and screw it • Slide the RUBY phantom in towards the back
tight. until it reaches the locating shims.
• Push the detector lock at the front of the insert ➔ The horizontal, red lines on the RUBY tilted
downward and slide the detector through the base will become visible.
bore hole in the insert into the phantom from
the front and as far as it will go.
34 D968.131.00/05 en
RUBY (T40072)
Operating Manual - Application
NOTICE
Observe the following points to avoid equip-
ment damage:
5
1
D968.131.00/05 en 35
RUBY (T40072)
Operating Manual - Application
36 D968.131.00/05 en
RUBY (T40072)
Operating Manual - Application
5.5.4 RUBY Head Set MultiMet 5.5.5 RUBY Head Set MultiMet
Premium
The RUBY head set MultiMet L981654 is intended
for the systematic quality assurance of treatment The RUBY head sets MultiMet Premium
planning techniques used for irradiation of multi- (L981655 / L981656 / L981657) are equivalent to
ple target volumes with one isocenter. the RUBY head set MultiMet and include the fol-
lowing additional products:
The three detector positions of the RUBY insert
System QA - MultiMet represent three possible – 3 UNIDOS Tango electrometers
positions of the target volumes and the CT mark-
– 3 extension cables
ers enable the isocenter to be defined at the cen-
ter of the RUBY head phantom. – 3 PinPoint 3D ionization chambers
D968.131.00/05 en 37
RUBY (T40072)
Operating Manual - Application
38 D968.131.00/05 en
RUBY (T40072)
Technical Manual - Technical Specifications
Technical Manual
6 Technical Specifications
Only values with specified tolerances or limits are guaranteed. Values without tolerances are for informa-
tional purposes only.
D968.131.00/05 en 39
RUBY (T40072)
Technical Manual - Technical Specifications
40 D968.131.00/05 en
RUBY (T40072)
Technical Manual - Ordering Information
7 Ordering Information
Scope of Supply
D968.131.00/05 en 41
RUBY (T40072)
Technical Manual - Ordering Information
42 D968.131.00/05 en
RUBY (T40072)
Technical Manual - Ordering Information
RUBY Head Set MultiMet Premium Type number Set Set Set
L981655 L981656 L981657
con- con- con-
necting necting necting
system system system
M W N
The set includes:
UNIDOS Tango (connecting syst. M, 3x) TM10052 x
UNIDOS Tango (connecting system W, 3x) TW10052 x
UNIDOS Tango (connecting system N, 3x) TN10052 x
Extension cable 'M20m HD'
(connecting system M, 3x) T2954/K3-20 x
Extension cable 'W20w HD'
(connecting system W, 3x) T26002.1.002-20 x
Extension cable 'W20n HD'
(connecting system N, 3x) T26005.1.001-20 x
PinPoint 3D chamber 0.016 cm³
(connecting system M, 3x) TM31022 x
PinPoint 3D chamber 0.016 cm³
(connecting system W, 3x) TW31022 x
PinPoint 3D chamber 0.016 cm³
(connecting system N, 3x) TN31022 x
Inclined stand (3x) T10052.1.025 x x x
USB stick UNIDOS (3x) L991330 x x x
Patch cable RJ45 S/UTP Cat5e - 3 m gray (3x) L178072 x x x
RUBY head phantom T40072.1.800 x x x
RUBY insert for system QA – MultiMet T40072.1.500 x x x
RUBY detector holder T60019 T40072.1.110 x x x
RUBY detector holder T31022 T40072.1.120 x x x
RUBY detector holder T31021 T40072.1.130 x x x
RUBY detector holder T60023 T40072.1.140 x x x
RUBY detector holder T31010 T40072.1.160 x x x
RUBY detector holder T60019 MultiMet caudal T40072.1.510 x x x
RUBY detector holder T60019 MultiMet cranial T40072.1.515 x x x
RUBY detector holder T31022 MultiMet caudal T40072.1.520 x x x
RUBY detector holder T31022 MultiMet cranial T40072.1.525 x x x
RUBY detector holder T31021 MultiMet caudal T40072.1.530 x x x
RUBY detector holder T31021 MultiMet cranial T40072.1.535 x x x
RUBY detector holder T60023 MultiMet caudal T40072.1.540 x x x
RUBY detector holder T60023 MultiMet cranial T40072.1.545 x x x
RUBY detector holder T31010 MultiMet caudal T40072.1.560 x x x
RUBY detector holder T31010 MultiMet cranial T40072.1.565 x x x
D968.131.00/05 en 43
RUBY (T40072)
Technical Manual - Ordering Information
44 D968.131.00/05 en
RUBY (T40072)
Technical Manual - Ordering Information
D968.131.00/05 en 45
RUBY (T40072)
Technical Manual - Ordering Information
46 D968.131.00/05 en
RUBY (T40072)
Service Manual - Cleaning
Service Manual
WARNING 8 Cleaning
Improper execution of service tasks.
D968.131.00/05 en 47
RUBY (T40072)
Service Manual - Disposal of the Product
Below, you will find a list of the maintenance mea- The typical lifetime of the product is approx.
sures and inspections necessary for the product. 10 years.
• If damage or malfunctions are identified, the The product does not contain hazardous materi-
product must be repaired before it is used als.
again.
Repair
48 D968.131.00/05 en
RUBY (T40072)
Literature
Literature
D968.131.00/05 en 49
RUBY (T40072)
Index
Index
Film, load................................................................. 35
T
G Technical specifications........................................... 39
General information................................................... 7
I
Insert, introduce....................................................... 32
Inserts...................................................................... 22
Intended use............................................................ 13
L
List of figures............................................................. 5
Literature.................................................................. 49
M
Maintenance............................................................ 48
MRI safety information............................................ 12
O
Ordering information................................................ 41
P
Phantom body......................................................... 17
Product description.................................................. 16
R
Repair...................................................................... 48
RUBY Carrying Case.............................................. 31
RUBY carrying case for head phantom................... 21
RUBY head phantom.............................................. 19
RUBY head set MultiMet......................................... 37
RUBY head set MultiMet Premium.......................... 37
RUBY insert Linac QA............................................. 25
50 D968.131.00/05 en